Fluke 80T-150U User Manual

0 (0)
80T-150U-7001K Probe Kit
®
Installation and Calibration Instructions for the 80T-150U
Probe Installation for the 80T-150U
1. Remove the top cover from the 80T-150U temperature-to­voltage converter.
2. Note the location of the white, black, and bare cable leads (see Figure 1).
Caution
Use professional soldering practices when replacing the probe cable. Car eless soldering can damage the pca and void the warranty.
3. Clip the cable tie, and unsolder the three cable leads from the printed circuit assembly (pca).
4. Remove all remaining solder from the three plated t hru holes.
5. Firmly anchor the new cable assembl y to the pca using the cable tie provided in the probe kit.
6. Insert the black, white, and bare wires into t heir respective holes, and solder them to the pca.
7. Clip off the excess lead length on the bottom side of the pca.
8. Calibrate the temperature probe following t he instructions given below.
Calibration for the 80T-150U
NOTE
Values given in brackets apply to the performance test.
NOTE
Disregard the resist ors supplied with the probe kit.
1. Access the interior of the 80T-150U by removing the bottom case screw and separating the case halves.
2. Connect the 80T-150U to a DVM with 10 µV resol ution, and select mV dc range.
3. Select the °C position of the switch. Immerse the probe tip approximately two inches into a mercury thermometer­monitored 0°C bath, and allow 60 seconds for readings to stabilize.
4. Adjust R3 (see Figure 1) t o obtain the following reading:
0 ± 0.05 mV dc [0 ± 2 mV dc]
5. Select the °F position of the switch, and adjust R2 to obtain the following reading:
32.0 ± 0.1 mV dc [32.0 ± 4 mV dc]
6. Select the °C position, move the probe tip to a 70 °C to 90 °C bath and allow the readings to stabilize.
7. Adjust R4 to obtain a DVM reading that agrees with the bath temperature (BT) as monit ored by a mecury thermometer.
°C BT ± 0.05 mV dc [BT ± 2 mV dc]
8. Select the °F position, and verify that output is within ± 4 mV dc of the bath thermometer reading. If nececssary, change t he DVM range to obtain an on-scale reading.
9. Return the probe tip t o the 0 °C bath and check the output; if readjustment is necessary, repeat steps 6 through 10 until the readings can be obtained without adjestment.
10. Set the 80T-150U switch to the OFF position and remove the 80T-150U from the DVM.
11. Reassemble the 80T-150U
The 80T-150U is now calibrated.
Figure 1.
PN 433557 September 1976 Rev. 3, 4/02
1976, 1986, 2002 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companies.
Black
White
80T-150U-4011