Fluke 5220A Data Sheet

The 5520A/5500A
Multi-Product Calibrators
Extended Specifications
December 1997
General Specifications
5520A/5500A
Warmup Time Twice the time since last warmed up, to a maximum of 30 minutes Settling Time Less than 5 seconds for all functions and ranges except as noted Standard Interfaces IEEE-488 (GPIB), RS-232, 5725A (5500A only) Temperature Performance Operating: 0°C to 50°C
Temperature Coefficient Temperature coefficient for temperatures outside tcal ±5°C is 10% of the 90-day specification
Relative Humidity Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
Altitude Operating: 3,050m (10,000 ft) maximum
Safety Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
Analog Low Isolation 20V EMC 5520A: Designed to comply with IEC 61326-1/1997
Line Power Line Voltage (selectable): 100V, 120V, 220V, 240V
Power Consumption 5520A: 600 VA
Dimensions Height:17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Weight (without options) 5520A: 22 kg (49 lb)
Absolute Uncertainty Definition 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity,
Specification Confidence Interval >99%
Calibration (tcal): 15°C to 35°C Storage: -20°C to 70°C
(or 1-year, as applicable) per °C
Storage: <95%, noncondensing
Non-operating: 12,200m (40,000 ft) maximum
CAN/CSA-C22.2 No. 1010.1-92
5500A: Designed to comply with FCC Rules Part 15
Line Frequency: 47 to 63 Hz Line Voltage Variation: ±10% about line voltage setting
5500A: 300 VA
Width: 43.2 cm (17 in), standard rack width Depth: 47.3 cm (18.6 in) overall
5500A: 20 kg (44 lb)
line and load regulation and the traceability of the external standards used for calibration.
You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
2 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications
DC V oltage Specifications
5520A
Absolute Uncertainty, tcal ±5°C
±(ppm of output + µV)
Ranges
90 Days 1 Year
0 to 329.9999 mV 20 + 1.5 25 + 1.5 3 + 1 0.1 µV50 0 to 3.299999V 9 + 2 11 + 2 2 + 1.5 1 µV 10 mA 0 to 32.99999V 10 + 15 12 + 15 2 + 10 10 µV 10 mA 30 to 329.9999V 15 + 150 18 + 150 2.5 + 100 100 µV5 mA 100 to 1000V 15 + 1500 18+ 1500 3 + 300 1000 µV5 mA
Auxiliary Output (dual output mode only)
0 to 329.999 mV 300 + 350 400 + 350 30 + 100 1 µV5 mA
0.33 to 3.29999V 300 + 350 400 + 350 30 + 100 10 µV5 mA
3.3 to 7V 300 + 350 400 + 350 30 + 100 100 µV5 mA
TC Output and Measure in Linear 10 µV/°C and 1 mV/°C Modes
0 to 329.999 mV 40 + 3 50 + 3 5 + 2 0.1 µV10
2
5500A
Absolute Uncertainty, tcal ±5°C
±(ppm of output + µV)
Ranges
90 Days 1 Year
0 to 329.9999 mV 50 + 3 60 + 3 5 + 1 0.1 µV50 0 to 3.299999V 40 + 5 50 + 5 4 + 3 1 µV 10 mA 0 to 32.99999V 40 + 50 50 + 50 4 + 30 10 µV 10 mA 30 to 329.9999V 45 + 500 55 + 500 4.5 + 300 100 µV5 mA 100 to 1000V 45 + 1500 55 + 1500 4.5 + 900 1000 µV5 mA
Auxiliary Output (dual output mode only)
0 to 329.999 mV 300 + 350 400 + 350 30 + 100 1 µV5 mA
0.33 to 3.29999V 300 + 350 400 + 350 30 + 100 10 µV5 mA
TC Output and Measure in Linear 10 µV/°C
0 to 329.999 mV 50 + 3 60 + 3 5 + 1 0.1 µV10
1
Remote sensing is not provided. Output resistance is <5 m for outputs 0.33V. The AUX output has an output resistance of <1. TC output has an output impedance of 10 ±1.
2
Two channels of dc voltage output are provided.
Note: Maximum voltage output is 1020V.
2
Stability
24 Hours, ±1°C
±(ppm of output
+ µV)
Stability
24 Hours, ±1°C
±(ppm of output
+ µV)
Resolution
Resolution
Maximum
Burden
Current
Maximum
Burden
Current
1
1
5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 3
DC Current Specifications
5520A
Absolute Uncertainty, tcal ±5°C
Ranges
0 to 329.999 µA 120 + 0.02 150 + 0.02 1 nA 10V 400 µH 0 to 3.29999 mA 80 + 0.03 100 + 0.03 0.01 µA 10V 400 µH 0 to 32.9999 mA 80 + 0.2 100 + 0.2 0.1 µA 7V 400 µH 0 to 329.999 mA 80 + 2 100 + 2 1 µA 7V 400 µH 0 to 2.99999A 160 + 40 200 + 40 10 µA 6V 400 µH 0 to 10.9999A 380 + 330 500 + 330 100 µA 4V 400 µH 11 to 20.5A
1
At 20°C, full 20.5A output is available continuously for 20 minutes.
1
±(ppm of output + µA)
90 Days 1 Year
800 + 750 1000 + 750 100 µA 4V 400 µH
Resolution
5500A
Absolute Uncertainty, tcal ±5°C
Ranges
0 to 3.29999 mA 0.010 + 0.05 0.013 + 0.05 0.01 µA 4.5V 1 µH 0 to 32.9999 mA 0.008 + 0.25 0.01 + 0.25 0.1 µA 4.5V 200 µH 0 to 329.999 mA 0.008 + 3.3 0.01 + 3.3 1 µA 3.0V 200 µH 0 to 2.19999A 0.023 + 44 0.03 + 44 10 µA 3.4V 200 µH 0 to 10.9999A 0.038 + 330 0.06 + 330 100 µA 2.5V 200 µH
5725A Amplifier
0 to 11A 0.03 + 330 0.04 + 330 100 µA 4V 400 µH
±(% of output + µA)
90 Days 1 Year
Resolution
Compliance
Voltage
Compliance
Voltage
Maximum Inductive
Load
Maximum Inductive
Load
4 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications
Resistance Specifications
5520A
Floor
2
Resolution
Allowable Current
Absolute Uncertainty, tcal ±5°C ±(ppm of output + floor)
1
Ranges
ppm of output
90 Days 1 Year
Time, Temperature Since Ω Zero
12 Hours ±1°C 7 Days ±5°C
0 to 10.9999 35 40 0.001 0. 01 0.0001 1 to 125 mA 11 to 32.9999 25 30 0.0015 0.015 0.0001 1 to 125 mA 33 to 1 09.9999 22 28 0.0015 0.015 0.0001 1 to 70 mA 110 to 329.9999 22 28 0.002 0.02 0.0001 1 to 40 mA 330 to 1.099999 k 22 28 0.002 0.02 0.001 250 µA to 18 mA
1.1 to 3.299999 k 22 28 0.02 0.2 0.001 250 µA to 5 mA
3.3 to 10.99999 k 22 28 0.02 0.1 0.01 25 µA to 1.8 mA 11 to 32.99999 k 22 28 0.2 1 0.01 25 µA to 0.5 mA 33 to 109.9999 k 22 28 0.2 1 0.1 2.5 µA to 0.18 mA 110 to 329.9999 k 25 32 2 10 0.1 2.5 µA to 0.05 mA 330 k to 1.099999 M 25 32 2 10 1 250 nA to 0.018 mA
1.1 to 3.299999 M 40 60 30 150 1 250 nA to 5 µA
3.3 to 10.99999 M 110 130 50 250 10 25 nA to 1.8 µA 11 to 32.99999 M 200 250 2500 2500 10 25 nA to 500 nA 33 to 109.9999 M 400 500 3000 3000 100 2.5 nA to 180 nA 110 to 329.999 M 2500 3000 100000 100000 1000 2.5 nA to 50 nA 330 to 1100 M 12000 15000 500000 500000 10000 1 nA to 13 nA
1
Continuously variable from 0 to 1.1 G.
2
Applies for 4-WIRE compensation only. For COMP OFF and 2-WIRE COMP, add 5 µV per amp of stimulus current to the floor specification. For example, in 2-wire mode, at 1 k, the floor specification within 12 hours of an ohms zero cal for a measurement current of 250 µA is 0.002 + 5 µV/250 µA = 0.022.
5500A
Absolute Uncertainty, tcal ±5°C
Ranges
1
±(% of output + Ω)
90 Days 1 Year
0 to 10.999 0.009 + 0.008 11 to 32.999 0.009 + 0.015 33 to 109.999 0.007 + 0.015 110 to 329.999 0.007 + 0.015
3 3 3 3
330 to 1.09999 k 0.007 + 0.06 0.009 + 0.06 0.01 250 µA to 18 mA
1.1 to 3.29999 k 0.007 + 0.06 0.009 + 0.06 0.01 250 µA to 5 mA
3.3 to 10.9999 k 0.007 + 0.6 0.009 + 0.6 0.1 25 µA to 1.8 mA 11 to 32.9999 k 0.007 + 0.6 0.009 + 0.6 0.1 25 µA to 0.5 mA 33 to 109.999 k 0.008 + 6 0.011 + 6 1 2.5 µA to 0.18 mA 110 to 329.999 k 0.009 + 6 0.012 + 6 1 2.5 µA to 0.05 mA 330 k to 1.09999 M 0.011 + 55 0.015 + 55 10 250 nA to 0.018 mA
1.1 to 3.29999 M 0.011 + 55 0.015 + 55 10 250 nA to 5 µA
3.3 to 10.9999 M 0.045 + 550 0.06 + 550 100 25 nA to 1.8 µA 11 to 32.9999 M 0.075 + 550 0.1 + 550 100 25 nA to 500 nA 33 to 109.999 M 0.4 + 5500 0.5 + 5500 1000 2.5 nA to 180 nA 110 to 329.999 M 0.4 + 16500 0.5 + 16500 1000 2.5 nA to 60 nA
1
Continuously variable from 0 to 330 M.
2
Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
3
The floor added is improved to 0.006 (0-10.99 range) and 0.010 (11-329.999 range) if the 5500A is zeroed within 8 hours and temperature is ±1°C of zeroing ambient temperature.
2
0.012 + 0.008
0.012 + 0.015
0.009 + 0.015
0.009 + 0.015
Resolution
3 3 3 3
0.001 1 to 125 mA
0.001 1 to 125 mA
0.001 1 to 70 mA
0.001 1 to 40 mA
Allowable Current
5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 5
AC V oltage (Sinewave) Specifications
5520A
Absolute Uncertainty, tcal ±5°C
Ranges
1.0 to 32.999 mV 10 to 45 Hz 600 + 6 800 + 6 1 µV50
33 to 329.999 mV 10 to 45 Hz 400 + 8 500 + 8 1 µV50
0.33 to 3.29999V 10 to 45 Hz 250 + 50 300 + 50 10 µV 10 mA
3.3 to 32.9999V 10 to 45 Hz 250 + 650 300 + 650 100 µV 10 mA
33 to 329.999V 45 Hz to 1 kHz 150 + 2000 190 + 2000 1 mV 5 mA, except
330 to 1000V 45 Hz to 1 kHz 250 + 10000 300 + 10000 10 mV 2 mA, except
Auxiliary Output (dual output mode only)
10 to 329.999 mV 10 to 20 Hz 0.15% + 370 0.2% + 370 1 µV5 mA
0.33 to 3.29999V 10 to 20 Hz 0.15% + 450 0.2% + 450 10 µV5 mA
3.3 to 5V 10 to 20 Hz 0.15% + 450 0.2% + 450 100 µV5 mA
1
Remote sensing is not provided. Output resistance is <5 m for outputs 0.33V. The AUX output resistance is <1. The maximum load capacitance is 500 pF, subject to the maximum burden current limits.
2
There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz for the 5520A.
Note: Maximum voltage output is 1020V.
Frequency
45 Hz to 10 kHz 120 + 6 150 + 6
10 to 20 kHz 160 + 6 200 + 6 20 to 50 kHz 800 + 6 1000 + 6 50 to 100 kHz 3000 + 12 3500 + 12
100 to 500 kHz 6000 + 50 8000 + 50
45 Hz to 10 kHz 110 + 8 130 + 8
10 to 20 kHz 120 + 8 150 + 8 20 to 50 kHz 300 + 8 350 + 8 50 to 100 kHz 600 + 32 800 + 32
100 to 500 kHz 1600 + 70 2000 + 70
45 Hz to 10 kHz 100 + 25 120 + 25
10 to 20 kHz 160 + 50 190 + 50 20 to 50 kHz 250 + 50 300 + 50 50 to 100 kHz 550 + 125 700 + 125
100 to 500 kHz 2000 + 600 2400 + 600
45 Hz to 10 kHz 125 + 200 150 + 200
10 to 20 kHz 220 + 600 240 + 600 20 to 50 kHz 300 + 600 350 + 600 50 to 100 kHz 750 + 1600 900 + 1600
1 to 10 kHz 160 + 6000 200 + 6000 20 mA for 10 to 20 kHz 220 + 6000 250 + 6000 45 to 65 Hz 20 to 50 kHz 600 + 20000 800 + 20000 50 to 100 kHz 1600 + 50000 2000 + 50000
1 to 5 kHz 200 + 10000 250 + 10000 6 mA for
5 to 10 kHz 250 + 10000 300 + 10000 45 to 65 Hz
2
20 to 45 Hz 0.08% + 370 0.1% + 370
45 Hz to 1 kHz 0.08% + 370 0.1% + 370
1 to 5 kHz 0.15% + 450 0.2% + 450
5 to 10 kHz 0.3% + 450 0.4% + 450 10 to 30 kHz 0.6% + 9 00 0.8% + 900
20 to 45 Hz 0.08% + 450 0.1% + 450
45 Hz to 1 kHz 0.07% + 450 0.09% + 450
1 to 5 kHz 0.15% + 1400 0.2% + 1400
5 to 10 kHz 0.3% + 1400 0.4% + 1400 10 to 30 kHz 0.6% + 2800 0.8% + 2800
20 to 45 Hz 0.08% + 450 0.1% + 450
45 Hz to 1 kHz 0.07% + 450 0.09% + 450
1 to 5 kHz 0.15% + 1400 0.2% + 1400
5 to 10 kHz 0.3% + 1400 0.4% + 1400
±(ppm of output + µV)
90 Days 1 Year
Resolution
Maximum
Burden
1
6 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications
5500A
Absolute Uncertainty, tcal ±5°C
Ranges
Frequency
±(% of output + µV)
Resolution
90 Days 1 Year
1.0 to 32.999 mV 10 to 45 Hz 0.26 + 20 0.35 + 20 1 µV50 45 Hz to 10 kHz 0.11 + 20 0.15 + 20
10 to 20 kHz 0.15 + 20 0.2 + 20 20 to 50 kHz 0.19 + 20 0.25 + 20 50 to 100 kHz 0.26 + 33 0.35 + 33
100 to 500 kHz 0.75 + 60 1 + 60
33 to 329.999 mV 10 to 45 Hz 0.19 + 50 0.25 + 50 1 µV50
45 Hz to 10 kHz 0.04 + 20 0.05 + 20
10 to 20 kHz 0.08 + 20 0.1 + 20 20 to 50 kHz 0.12 + 40 0.16 + 40 50 to 100 kHz 0.17 + 170 0.24 + 170
100 to 500 kHz 0.53 + 330 0.7 + 330
0.33 to 3.29999V 10 to 45 Hz 0.11 + 250 0.15 + 250 10 µV 10 mA 45 Hz to 10 kHz 0.02 + 60 0.03 + 60
10 to 20 kHz 0.06 + 60 0.08 + 60 20 to 50 kHz 0.1 + 300 0.14 + 300 50 to 100 kHz 0.17 + 1700 0.24 + 1700
100 to 500 kHz 0.38 + 3300 0.5 + 3300
3.3 to 32.9999V 10 to 45 Hz 0.11 + 2500 0.15 + 2500 100 µV 10 mA 45 Hz to 10 kHz 0.03 + 600 0.04 + 600
10 to 20 kHz 0.06 + 2600 0.08 + 2600 20 to 50 kHz 0.14 + 5000 0.19 + 5000 50 to 100 kHz 0.17 + 17000 0.24 + 17000
33 to 329.999V 45 Hz to 1 kHz 0.04 + 6.6 mV 0.05 + 6.6 mV 1 mV
1 to 10 kHz 0.06 + 15 mV 0.08 + 15 mV 20 mA for
10 to 20 kHz 0.07 + 33 mV 0.09 + 33 mV
330 to 1000V 45 Hz to 1 kHz 0.04 + 80 mV 0.05 + 80 mV 10 mV
1 to 5 kHz 0.15 + 100 mV 0.20 + 100 mV 6 mA for 5 to 10 kHz 0.15 + 500 mV 0.20 + 500 mV
5725 Amplifier
10 to 329.999 mV 45 Hz to 1 kHz 0.04 + 80 mV 0.05 + 80 mV 50 mA
1 to 20 kHz 0.06 + 100 mV 0.08 + 100 mV 10 mV 70 mA
20 to 30 kHz 0.08 + 100 mV 0.1 + 100 mV 70 mA
100 to 750V 30 to 100 kHz 0.38 + 500 mV 0.5 + 500 mV 10 µV 70 mA
Auxiliary Output (dual output mode only)
2
10 to 329.999 mV 10 to 20 Hz 0.15 + 370 0.2 + 370 1 µV5 mA
20 to 45 Hz 0.08 + 370 0.1 + 370
45 Hz to 1 kHz 0.08 + 370 0.1 + 370
1 to 5 kHz 0.15 + 450 0.2 + 450 5 to 10 kHz 0.3 + 450 0.4 + 450
0.33 to 3.29999V 10 to 20 Hz 0.15 + 450 0.2 + 450 10 µV5 mA
20 to 45 Hz 0.08 + 450 0.1 + 450
45 Hz to 1 kHz 0.07 + 450 0.09 + 450
1 to 5 kHz 0.15 + 1400 0.2 + 1400 5 to 10 kHz 0.3 + 1400 0.4 + 1400
1
Remote sensing is not provided. Output resistance is <5 m for outputs 0.33V. The AUX output resistance is <1. The maximum load capacitance is 500 pF, subject to the maximum burden current limits.
2
There are two channels of voltage output. The maximum frequency of the dual output is 10 kHz for the 5500A.
Note: Maximum voltage output is 1020V.
Maximum
1
Burden
5 mA, except
45 to 65 Hz
2 mA, except
45 to 65 Hz
5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 7
AC Current (Sinewave) Specifications
5520A (LCOMP Off)
Ranges
29.00 to 329.99 µA 10 to 20 Hz 0.16% 0.1 µA 0.2% 0.1 µA 200 µH
0.33 to 3.2999 mA 10 to 20 Hz 0.16% 0.15 µA 0.2% 0.15 µA 200 µH
3.3 to 32.999 mA 10 to 20 Hz 0.15% 2 µA 0.18% 2 µA 50 µH
33 to 329.99 mA 10 to 20 Hz 0.15% 20 µA 0.18% 20 µA 50 µH
0.33 to 2.99999A 10 to 45 Hz 0.15% 100 µA 0.18% 100 µA 2.5 µH
3 to 10.9999A 45 to 100 Hz 0.05% 2000 µA 0.06% 2000 µA1 µH
11 to 20.5A
1
At 20°C, full 20.5A output is available continuously for 20 minutes.
2
For compliance voltages greater than 1V, add 1 mA/V to the floor specification from 1 to 5 kHz.
3
For compliance voltages greater than 1V, add 5 mA/V to the floor specification from 5 to 10 kHz.
1
Frequency
20 to 45 Hz 0.12% 0.1 µA 0.15% 0.1 µA
45 Hz to 1 kHz 0.1% 0.1 µA 0.125% 0.1 µA
1 to 5 kHz 0.25% 0.15 µA 0.3% 0.15 µA 5 to 10 kHz 0.6% 0.2 µA 0.8% 0.2 µA
10 to 30 kHz 1.2% 0.4 µA 1.6% 0.4 µA
20 to 45 Hz 0.1% 0.15 µA 0.125% 0.15 µA
45 Hz to 1 kHz 0.08% 0.15 µA0.1% 0.15 µA
1 to 5 kHz 0.16% 0.2 µA 0.2% 0.2 µA 5 to 10 kHz 0.4% 0.3 µA 0.5% 0.3 µA
10 to 30 kHz 0.8% 0.6 µA 1.0% 0.6 µA
20 to 45 Hz 0.075% 2 µA 0.09% 2 µA
45 Hz to 1 kHz 0.035% 2 µA 0.04% 2 µA
1 to 5 kHz 0.065% 2 µA 0.08% 2 µA 5 to 10 kHz 0.16% 3 µA 0.2% 3 µA
10 to 30 kHz 0.32% 4 µA 0.4% 4 µA
20 to 45 Hz 0.075% 20 µA 0.09% 20 µA
45 Hz to 1 kHz 0.035% 20 µA 0.04% 20 µA
1 to 5 kHz 0.08% 50 µA 0.10% 50 µA 5 to 10 kHz 0.16% 100 µA 0.2% 100 µA
10 to 30 kHz 0.32% 200 µA 0.4% 200 µA
45 Hz to 1 kHz 0.05% 100 µA 0.06% 100 µA
1 to 5 kHz 5 to 10 kHz
100 to 1 kHz 0.08% 2000 µA 0.10% 2000 µA
1 to 5 kHz 2.0% 2000 µA 2.5% 2000 µA
45 to 100 Hz 0.1% 5000 µA 0.12% 5000 µA1 µH
100 Hz to 1 kHz 0.13% 5000 µA 0.15% 5000 µA
1 to 5 kHz 2.0% 5000 µA 2.5% 5000 µA
2
3
0.5% 1000 µA 0.6% 1000 µA
2.0% 5000 µA 2.5% 5000 µA
Absolute Uncertainty, tcal ±5°C ±(% of output + µA)
90 Days 1 Year
Maximum
Inductive Load
8 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications
Loading...