dB Measurements with Selectable Reference Impedances
of 2V to 8000V and Audio Power from 2V to 16V
Compare (Hi/Lo/Pass) Function for
Quick In-Tolerance Tests
0.05% dc Current Accuracy for 4-20 mA Current Loop Service
Touch Hold, Relative and Min Max
Audible Continuity and Diode Test
Optional Rechargeable Battery, Carrying Case, Rack Mount Kit & PC Software Package
Closed-Case Calibration
The Fluke 45 is a feature-rich 5 digit,
100,000 count DMM with a unique multifunction dual display, allowing measurement of two signal parameters from a
single test connection. The Fluke 45 offers
high performance and versatility for manufacturing test, depot and field service, and
research and development. A standard
RS-232C makes it ideal for PC instrument
applications.
Dual Display
The Fluke 45 is the first DMM with a
multifunction ‘‘dual display,’’ allowing the
user to select a wide variety of measurement combinations. It is particularly useful
in applications requiring two different
measurements of the same signal; i.e.
power supply testing, where Vdc output
can be viewed on the primary display
while the Vac ripple is shown on the
secondary display.
Standard RS-232C Interface
The RS-232C interface, standard in each
instrument, allows measurement data to
be filed, manipulated, printed or transmitted by modem. The print mode automatically formats measurement data for
printing on an RS-232C printer. Rates for
automated printing over RS-232C are
adjustable from 1 reading every 70 ms to
1 reading every 5.6 hours. The optional
QuickStart 45
automated communications and filing of
measurements with the Fluke 45 and an
IBM-PC or compatible via RS-232C.
TM
Software Package allows
dB Measurement
The Fluke 45 provides digital readout of
decibels with front panel selection of any
of twenty-one reference impedances from
2V to 8000V. For 2V, 4V, 8V and 16V
impedances, the meter automatically
calculates and displays audio power
in watts.
Compare Function
The Fluke 45 has a compare function for
fast in-tolerance limits testing. Upper- and
lower-limits are entered through the front
panel. Readouts show both a Hi/Lo/Pass
evaluation and measured value.
Touch Hold Relative and
Min/Max
Touch Hold captures the measurement,
beeps and locks it on the digital display
until you are ready to view it. It automatically updates with each new stable
measurement. The Relative mode remembers a reading and shows the change
(difference) between it and any readings
that follow. Min/Max records the highest
and lowest values measured. Either can be
recalled and displayed at any time.
Optional Battery and Case
An optional rechargeable battery and soft
carrying case are available for precision
field service applications. These options,
coupled with a 30 mA dc current range
and 0.05% accuracy, allow calibration of
4-20 mA loops in process control applications. The battery is available as a factory
installed option or can be user-installed at
a later date.
Optional GPIB/IEEE-488.2
and Rack Mount
The Fluke 45 may be used with GPIB/IEEE-
488.2 systems, including existing IEEE488 implementations. The IEEE- 488.2
option is available as a factory installed
option, or can be user installed and does
not require removal of the RS-232C interface. A rack mount kit is also available.
Closed-Case Calibration
Calibration can be performed via the
RS-232C (or optional IEEE-488.2) interface or manually from the front panel.
No internal adjustments are required.
Standard Equipment
Each Fluke 45 Dual Display Multimeter
includes an operator’s manual, quick
reference guide, line cord, and test leads.
Specifications
Technical Specifications
Accuracy specifications are given as ±
([% of reading] + [number of least significant digits]) at 18°C to 28°C with relative
humidity up to 90%, for a period of one
year after calibration. Six months specifications are also provided for dc volts. AC
inputs are ac-coupled and True-RMS
responding.
Display
Dual vacuum fluorescent displays, 99,999
counts each display. Annunciators: m, M,
k, V, A, V (ohms), Hz, ? + diode test,
(audible continuity), REMOTE, EXT TRIG,
SMF (reading rates), MAX, MIN, dB, HOLD,
REL, AUTO, – + (low battery). Resolution
and accuracy are dependent on selectable
reading rates of 2.5 (slow), 5 (medium) or
20 (fast) readings per second.
RateReadings
Slow2.599,999
Medium530,000
Fast203000
Per Second
Full Range
Display Counts
Page 2
Bench Multimeters
45 Dual Display Multimeter
RS-232C and IEEE-488
Reading Transfer Rates
Internal Trigger Operation (TRIGGER 1)
RateReadings Per Second
Slow2.5
Medium4.5
Fast4.5
External Trigger Operation (TRIGGER 4)
RateReadings Per Second
Slow1.5
Medium2.4
Fast3.8
Print Mode Operation (Print set at 1)
RateReadings Per Second
Slow2.5
Medium5.0
Fast13.5
Maximum Input: To be used in protected,
low-energy circuits only, not to exceed
250V or 4800 volt-amps
mA: 300 mA dc or ac rms. Protected
with a 500 mA, 250V, IEC 127 sheet 1
fast blow fuse
A: 10A dc or ac rms continuous or 20A
dc or ac rms for 30 seconds maximum.
Protected with a 15A, 250V, 10,000A
interrupt rating fast blow fuse.
Diode Test/Continuity
Accuracy
Slow999,99 mV10 mV
Medium2.5V100 mV
Fast2.5V1 mV
Test Current: Approximately 0.7 mA
when measuring forward bias junction
Audible Tone: Continuous tone for continuity. Brief tone for normal forward biased
diode or semiconductor junction
Open Circuit Voltage: 3.2V maximum
Continuity Capture Time: 50 ms maxi-
Input Impedance: 10 MV in parallel with <100 pF
Normal Mode Rejection Ratio: >80 dB at 50 or 60 Hz, slow and medium rates; >54 dB
for frequencies between 50 and 440 Hz, slow and medium rates; >60 dB at 50 Hz, fast
rate
Common Mode Rejection Ratio: >90 dB at dc, 50 or 60 Hz (1 kV unbalanced, slow &
medium rates)
*Error in power mode will not exceed twice the linear accuracy specification
Linear ResolutionLinear Resolutions
SlowMediumFastSlowMediumFast
Decibels Resolution
Linear AccuracydB Accuracy
SlowMediumFastSlow/MedFast
Accuracy specifications apply within the following limits, based on reading rate:
Slow: Between 15,000 counts and full range
Medium: Between 1,500 counts and full range
Fast: Between 150 counts and full range
Input Impedance: 1MVin parallel with <100 pF
Maximum Crest Factor: 3.0
Common Mode Rejection Ratio: >60 dB at 50 or 60 Hz (1 kV unbalanced medium rate)
Maximum Input: 750V rms, 1000V peak
7
volt-hertz product on any range, normal mode input
2 × 10
1 × 106volt-hertz product on any range, common mode input
Range
Power*
Max Input
Upper Freq
AC + DC Voltage (calculated)
Total measurement error will not exceed the sum of the separate ac and dc accuracy
specifications, plus one display count.
Page 3
Bench Multimeters
45 Dual Display Multimeter
Ohms
Range
300V—10 mV100 mV 0.05% + 2 + 0.02V0.251 mA
3kV—100 mV1V0.05% + 20.24120 mA
30 kV—1V10V0.05% + 20.2914 mA
300 kV—10V100V0.05% + 20.291.5 mA
3MV—100V1kV0.06% + 20.30150 nA
30 MV—1kV10 kV0.25% + 32.25320 nA
300 MV—100 kV1MV2%2.90320 nA
100V1mV——0.05% + 8 + 0.02V0.091 mA
1000V10 mV——0.05% + 8 + 0.02V0.10120 mA
10 kV100 mV——0.05% + 80.1114 mA
100 kV1V——0.05% + 80.111.5 mA
1000 kV10V——0.06% + 80.12150 nA
10 MV100V——0.25% + 61.50150 nA
100 MV100 kV——2% + 22.75320 nA
Resolution
SlowMediumFast
Accuracy
Typical Full
Scale Voltage
Open Circuit Voltage: 3.2V max on 100V, 300V, 30 MV, 100 MV & 300 MV ranges;
5 Hz-20 kHz 100 mA > 3 mA rms sinewave
45 Hz-2 kHz10A> 3A rms sinewave
Slow/MediumFast
Sensitivity of AC Voltage
Sensitivity Level of AC Current
Resolution
General Specifications
Power Requirements
Power: 90 to 264V ac (no switching
required), 50 and 60 Hz <15 VA minimum
Environmental Data
Maximum Common Mode Voltage:
1000V dc or peak ac from any input to
earth
Warmup Time: 1 hour to rated
specifications
Temperature Coefficient: <0.1 times the
applicable accuracy specifications per
degree C from 0°C to 18°C and from 28°C
to 50°C, (32°F to 64.4°F and 82.4°F to
122°F)
Operating Temperature: 0°C to 50°C
(32°F to 122°F)
Storage Temperature: −40°C to 70°C
(−40°F to 158°F). Elevated temperature
storage of battery will accelerate battery
self-discharge. Maximum storage time
before battery must be recharged:
20°C to 25°C1000 days
50°C180 days
70°C40 days
Relative Humidity (non-condensing): Up
to 90%, 0°C to 28°C (32°F to 82.4°F); up to
80%, 28°C to 35°C (82.4 °F to 95°F); up to
70%, 35°C to 50°C (95°F to 122°F), except
70%, 0°C to 50°C (32°F to 122°F) for the
1MV, 3 MV, 10 MV, 30 MV, 100 MV and
300 MV ranges.
Altitude: Operating, 0 to 10,000 feet;
non-operating, 0 to 40,000 feet
Vibration: 3G @ 55 Hz per MIL-T-28800D,
Class 3, Style E
Shock: Half-sine 40G per MIL-T-28800D,
Class 3, Style E. Bench handling.