Appendix B: Model 4000M+ Communication Protocol.................................................B-1
B.1 Data Formats ............................................................................................... B-1
B.2 Setting Up for Operation .............................................................................. B-1
B.3 The Commands ...........................................................................................B-1
Appendix C: Schematic Diagram and Replacement Parts...........................................C-1
C.3 Schematic Diagram and Replacement Parts ............................................... C-1
Appendix D: kV Correction for Various Target/Filter Combinations...........................D-1
D.1 kV Correction for Various Target/Filter Combinations.................................. D-1
i
(Blank page)
General Information
Introduction
1
Section 1
General Information
1.1 Introduction
The Model 4000M+ is a self-contained, noninvasive X-ray Test Device. In a single exposure, it
simultaneously measures:
• kVp
• Exposure or Air Kerma
• Exposure Rate or Air Kerma Rate
• Time
The Model 4000M+ features a dual sensitivity preamplifier for compatibility with Radiographic,
Fluoroscopic, and Dental X-Ray Machines. In addition, it is calibrated for both tungsten anode (W/Al) and
molybdenum (Mo/Mo) anode X-Ray tubes, making it suitable for screen film mammography applications.
Its automatic waveform phase determination and extensive diagnostics minimize the potential for error.
The external ion chamber port accepts a variety of accessory ionization chambers for various
applications, including mammography, photo-timer calibration, and input phosphor image intensifier
measurements.
The Model 4000M+ uses proven technology to compute tube potential with +
separate, selectable filter pairs ensure optimum accuracy over the maximum range with minimum filtration
dependence. A separate internal ionization chamber measures tube output. Time is measured with
crystal quartz accuracy. A microprocessor controls the electronics and performs calculations to obtain the
displayed results.
2% accuracy. Five
1-1
Victoreen 4000M+
Operators Manual
1.2 Specifications
Measured Quantities
Kilovoltage
Accuracy: 1 kV Mo/Mo (22 to 35 kVp)
Time
Exposure
Detection Methods
kV CsI/photodiode pair measures x-ray transmission through
Time Computed from kV waveform stored in memory, against
Exposure Parallel plate ionization chamber (36 cm3 volume).
Calibration Check Reference to a calibrated voltage divider and a calibrated
Measured during the first 300 ms of exposure:
- kVp average
- kVp effective
- kVp maximum
Accuracy: +2%
Range:
W/Al: 27 to 155 kVp, calibrated to a tungsten anode tube with
4.5 mm Al total filtration;
Mo/Mo: 21-50 kVp, calibrated to a molybdenum anode tube
with 0.03 mm Mo total filtration;
(Mammo generators w/30 µ Mo)
Measured during entire exposure; referenced to 90% rise/fall
kV time.
Accuracy: Within 2% or 2 ms, which ever is greater.
Range: 1 ms to 10 sec.
Measured during entire exposure, kVp corrected.
Accuracy: +5%.
Range: 10 mR to 10 R
Accuracy: +5% Fluoroscopic Rate
Range: 0.5 to 200 R/min.
differential copper attenuator.
quartz crystal time base.
exposure monitor during irradiation.
1-2
General Information
Physical Characteristics
Display 16 character dot-matrix LCD.
Controls
Connectors
Power 120/230 VAC External Supply; Internal, rechargeable
Operating Conditions 10°-40°C max.; 90% relative humidity, non-condensing
Size 8 cm high, 22 cm wide, 23 cm deep (3 in. high, 8 ½ in. wide, 9
Six rocker switches:
ON/OFF: Turns power on/off.
RADIO/FLUORO: Selects radiographic or fluoroscopic mode
of operation.
EXP/ALL: Selects Exposure or All measurements.
(W/Al)/(Mo/Mo): Selects anode/filter.
HIGH/LOW: Selects High (Dental or Mammo)/Low sensitivity.
ROLL/RST: Rolls display through data; Resets exposure
measurement.
Power Supply/Charge Connector: 5.5 X 2.1 mm Male
Receptacle.
Scope Connector: BNC connector for oscilloscope connection.
Ion Chamber Signal Connector: BNC connector for external
ion chamber.
Ion Chamber Bias Connector: Banana jack supplying 300 volts
as Bias Supply.
RS-232 Port: 9-pin female connector configured as DCE.
batteries.
in. deep)
Specifications
1
1.3 Getting Started
Receiving Inspection
Upon receiving the unit:
1. Check the shipping carton(s) and their contents for in-shipment damage. If damage is evident, file
a claim with the carrier and contact Fluke Biomedical at 440.248.9300.
2. Check that all items listed on the packing slip are present and in good condition. If any items are
missing or damaged, contact Fluke Biomedical at 440.248.9300.
Storage
If necessary to store the unit prior to use, pack it in the original container(s) if possible, and store it in an
environment free of corrosive materials, fluctuations in temperature and humidity, and vibration and
shock.
1-3
Victoreen 4000M+
Operators Manual
(Blank page)
Operation
Power
Section 2
Operation
2.1 Power
Avoid shorting the battery. Do not mutilate or put a
fire. Use only approved charging methods.
The Model 4000M+ can be powered by the internal rechargeable nickel cadmium batteries or by AC line
power. An AC adapter is supplied with each unit. The connector for the adapter is located on the rear
panel and labeled POWER (see Figure 2-2).
The batteries charge whenever the unit is plugged in to an AC source. The batteries will fully charge in
approximately 6 hours, with the unit turned off, or in 24 hours, with the unit turned on. The batteries are
protected from overcharging.
2
2.2 Checking the Unit
Use the following procedure to check the Model 4000M+ for proper operation in the Radiographic Mode:
1. Place the unit on the x-ray table with the black plastic disk facing up.
2. Turn the Power Switch to ON. The Model No., software revision level, and filter setting will be
displayed, followed by a FLUORO Ready or RADIO Ready indication, depending on the position of
the FLUORO/RADIO Switch.
3. Position the Mode Selection Switch to RADIO.
4. Position the EXP/ALL Switch to ALL.
5. Position the Tube Target (Anode)/Filter Selection Switch to W/Al.
6. Position the Sensitivity Selection Switch to LOW.
7. Rotate the filter wheel to the 70-120 kVp setting. The RADIO Ready indication should appear on
the display.
If there is a hardware failure, an ERROR message
may be displayed instead of the RADIO Ready
condition. Error message indications and
meanings are discussed in Section 3 – Theory of
Operation.
8. Position the detector as discussed in Section 3 – Theory of Operation.
9. Make an exposure at 80 kVp and 100 mA and 100 mS.
10. Ready the display.
2-1
Victoreen 4000M+
Operators Manual
2.3 Front Panel Configuration
Model 4000M+
Noninvasive X-Ray Test Device
VICTOREEN
Figure 2-1. Front Panel Configuration Figure 2-1. Front Panel Configuration
The Model 4000M+ features a 16-character dot matrix LCD display and six (6) rocker switches on the
The Model 4000M+ features a 16-character dot matrix LCD display and six (6) rocker switches on the
front panel, as shown in Figure 2-1. The following paragraphs summarize their functions.
front panel, as shown in Figure 2-1. The following paragraphs summarize their functions.
Display Display
The 16-character dot matrix LCD displays the measurement results. The ROLL/RST Switch scrolls
The 16-character dot matrix LCD displays the measurement results. The ROLL/RST Switch scrolls
through the displayed values in the Radiographic Mode. In the Fluoroscopic Mode, displayed values are
through the displayed values in the Radiographic Mode. In the Fluoroscopic Mode, displayed values are
automatically toggled every two seconds.
automatically toggled every two seconds.
ROLL/RST HIGH
LOW
Mo/Mo
W/AI
EXPRADIOON
ALLFLUOROOFF
In the Radiographic Mode, displayed values include: In the Radiographic Mode, displayed values include:
• The average kVp • The average kVp
• The effective kVp • The effective kVp
• The maximum kVp • The maximum kVp
• The length of the exposure in seconds • The length of the exposure in seconds
• The tube output in Roentgens or Grays • The tube output in Roentgens or Grays
In the Fluoroscopic Mode, displayed values include: In the Fluoroscopic Mode, displayed values include:
• The real time exposure rate, in mR/min or mGy/sec • The real time exposure rate, in mR/min or mGy/sec
• The calculated kVp value • The calculated kVp value
Tube output is displayed in user-selected quantities of Roentgens or Grays. Display units are factory
Tube output is displayed in user-selected quantities of Roentgens or Grays. Display units are factory
preset according to the purchase order. The display units can be changed via an internal dip switch,
preset according to the purchase order. The display units can be changed via an internal dip switch,
located on the main PC Board. Use the following procedure:
located on the main PC Board. Use the following procedure:
1. Turn the Power Switch to OFF. 1. Turn the Power Switch to OFF.
2. Place the unit upside down on a flat surface. 2. Place the unit upside down on a flat surface.
3. Remove the four screws (and feet) securing the case bottom to the case top. 3. Remove the four screws (and feet) securing the case bottom to the case top.
4. Carefully remove the case bottom from the rest of the assembly. 4. Carefully remove the case bottom from the rest of the assembly.
5. Locate the four-position dip switch on the main PC board. 5. Locate the four-position dip switch on the main PC board.
6. Set the first (1) position as follows: 6. Set the first (1) position as follows:
• OFF to select Grays as the display units • OFF to select Grays as the display units
• ON to select Roentgens as the display units. • ON to select Roentgens as the display units.
2-2
Operation
Front Panel Configuration
Switch positions 2, 3, and 4 have been factory set
and MUST NOT be changed.
7. Carefully replace the case top.
8. Secure the case top to the rest assembly using the screws (and feet) removed in Step 3.
ON/OFF Switch
The ON/OFF Switch turns the instrument power on or off. As discussed earlier, the unit can be powered
with either the internal rechargeable batteries or with AC line power using the supplied AC adapter.
If an Invalid Filter message is displayed instead of
the Filter Wheel setting, the Filter Wheel is set to
CHK (or other illegal position) when the unit is
turned ON in the RADIO or FLUORO modes.
If the RADIO/FLUORO Switch is set to RADIO, and the EXP/ALL Switch is set to ALL, a RADIO Ready
message will be displayed.
2
If the RADIO/FLUORO Switch is set to FLUORO, and the EXP/ALL Switch is set to ALL, a FLUORO
Ready message will be displayed followed by an exposure rate value (in mGy/sec or mR/min) and a
calculated kVp value. The display will toggle between the exposure rate and kVp value.
The display will show exposure rate only (rather
than toggle between the exposure rate and
calculated kVp value) in the event that the input to
the detector is too low to provide an adequate
signal for a kVp calculation.
RADIO/FLUORO Switch
The RADIO/FLUORO Switch selects either the Radiographic Measurement Mode or the Fluoroscopic
Measurement Mode when the EXP/ALL Switch is set to ALL. When the EXP/ALL Switch is in EXP, the
RADIO/FLUORO Switch selects either the DOSE or RATE mode. Refer to Section 3 – Theory of
Operation for a discussion of applications for each mode.
In the Fluoroscopic Mode, a real time display of
tube output and kV are continuously updated. In
the Radiographic Mode, data are accumulated until
radiation ceases, and then the results are
displayed.
EXP/ALL Switch
The EXP/ALL Switch selects the Ion Chamber Mode, either the Exposure Dose/Rate Mode (EXP) or the
Radiographic/FLUORO Measurement Mode (ALL).
(W/Al)/(Mo/Mo) Switch
2-3
Victoreen 4000M+
Operators Manual
The (W/Al)/(Mo/Mo) Switch selects the target/filter of the X-Ray machine being used. Refer to Section 3
– Theory of Operation for a discussion of the applications for each setting.
HIGH/LOW Switch
The HIGH/LOW Switch selects the sensitivity of the detector. In the HIGH position, the kVp detector
sensitivity is increased by a factor of 20, allowing compatibility with dental and mammography machines.
The LOW position is suitable for most other applications.
When the unit is in the Fluoroscopic Mode
(FLUORO), the detector sensitivity is always high,
regardless of the front panel HIGH/LOW Switch
Setting.
ROLL/RST Switch
The ROLL/RST Switch, active only in the Radiographic Measurement and Exposure Modes, is a
momentary contact rocker that allows the user to change the display mode to Auto Roll or Manual Roll
and to scroll through measured quantities. The ROLL/RST Switch also zeros the display in the Exposure
Dose Mode. On power up (and whenever the Sensitivity, Measurement Mode, or filter wheel selection is
changed) the display mode defaults to Auto Roll.
In the Fluoroscopic Mode, the display automatically
toggles between the measured quantities. Values
are updated every two seconds.
In the Auto Roll Mode, the display scrolls through the measured quantities one after the other, as if the
ROLL button were being pressed once every two seconds. The display will continue to scroll until the
ROLL button is pressed again.
When a Radiographic exposure is complete, the kVp average value is displayed. Other measured
quantities can be displayed, one after another, by pressing the ROLL Switch to scroll through them.
Holding the ROLL Switch for longer than two seconds forces the display to enter the Auto Roll Mode.
Use the following procedure to change the Display Mode:
1. If the current Display Mode is Auto Roll (i.e., the display values are changing every two seconds)
and the unit is in the Radiographic Measurement Mode:
a. Press and release the ROLL Switch when the desired value is displayed (i.e., kVp Avg, kVp Eff,
kVp Max, Sec, or mR). The Display Mode will change to Manual Roll. The selected value will
remain displayed until ROLL is pressed again.
b. Perform Step 2 to scroll through the measured quantities, one after another.
2. If the current Display Mode is Manual Roll, press and hold the ROLL Switch for approximately two
seconds. After “roll “ appears in the bottom right corner of the display, the Display Mode will
change to Auto Roll. The display will be updated once every two seconds, scrolling through
measured quantities one after the other.
When the ROLL/RST Switch is pressed to zero the exposure, the PLEASE WAIT message displays for
approximately three seconds. The display will then read 0.0 mR or 0% Gy.
2-4
Operation
Rear Panel Configuration
2.4 Rear Panel Configuration
Power Supply/Charge Connector
The Power Supply/Charge Connector for the AC Adapter is located on the rear panel and labeled
POWER (see Figure 2-2). An AC adapter is supplied with each unit. The AC adapter supplies 9 volts,
500 mA power to the unit and recharges the internal batteries.
Use only the AC Adapter Supplied with the unit;
other adapters may damage the unit.
Scope Connector
The Scope Connector is a BNC connector labeled SCOPE and located on the rear panel of the unit (see
Figure 2-2). This connector allows the detector waveform to be displayed on a storage oscilloscope. It
provides a voltage signal proportional to the radiation waveform as measured by the lightly filtered
2
BIASSIGNAL
BEAM
DIRECTION
RS-232
detector.
Figure 2-2. Rear Panel Configuration
kVp
RANGE
W/AI
SCOPEPOWER
Filter Wheel
The filter wheel, located on the rear panel, allows selection of the kVp range. The wheel has six
positions:
• 27-42 (W/AI) or 21-50 (Mo/Mo)
• 35-60
• 50-85
• 70-120
• CHK
• 100-155
The numeric labels refer to the kVp range. The range associated with the position selected should
conform to the kVp setting of the X-ray machine. The CHK position is used for a calibration check.
Beam Direction Arrow
The Beam Direction Arrow shows how the unit must be positioned to accept x-rays from the x-ray
machine. For example, in the Fluoroscopic Mode, the unit must usually be turned upside down with the
black plastic disk facing the x-ray tube.
2-5
Victoreen 4000M+
Operators Manual
When the unit is upside down in the Fluoroscopic
Mode, the display automatically inverts so it can be
read from that position.
Ion Chamber Signal Connector
The Ion Chamber Signal Connector, labeled SIGNAL, is a BNC connector that accepts the input and
signal from the external ion chamber.
An electrical shock hazard exists on the ion
chamber bias (HV). Extreme care should be taken
when in close proximity to the ion chamber bias.
Ion Chamber Bias Connector
The Ion Chamber Bias Connector, labeled BIAS, is a banana-type jack supplying 300 DC volts as bias
supply for an external ion chamber.
RS-232 Port
The RS-232 Port is a 9-pin female connector configured as Data Communications Equipment (DCE).
See Appendix B for more information on this port.
2.5 Detector Positioning
The detector must be positioned according to the type of measurement required. Section 3 – Theory of
Operation includes a detailed discussion of detector positioning.
The following guidelines outline general detector positioning principles:
1. Locate the detector 46 to 102 centimeters (18 to 40 inches) away from the source of x-rays.
The optimum source-to-detector distance (SDD) is
66 centimeters (26 inches) for most measurements.
Refer to Appendix A for a discussion of source-todetector distance.
2. The black plastic disk (on the top panel of the Model 4000M+) must point toward the source, and
must be in the approximate center of the beam.
3. The beam must be parallel to, and in the same direction as, the BEAM DIRECTION arrow printed
on the rear panel of the Model 4000M+.
4. To eliminate measurement error due to the ‘heel’ effect, position the Model 4000M+ so that the
front-to-rear panel axis is parallel to the tube axis.
2-6
Operation
Calibration Check
2.6 Calibration Check
A calibration check should be performed periodically, or whenever a problem with the unit is suspected.
Use the following procedure to perform a calibration check on the Model 4000M+:
1. Turn the Power Switch to ON. The Model No., software revision level, and filter setting will be
displayed, followed by a FLUORO Ready or RADIO Ready indication, depending on the position
of the FLUORO/RADIO Switch.
2. Position the Mode Selection Switch to RADIO.
3. Position the Sensitivity Selection Switch to LOW.
4. Rotate the filter wheel to the CHK position.
5. Position the detector as discussed in Section 3 – Theory of Operation.
6. Set the x-ray machine to 100 kV, 100 mA, and 100 msec.
7. Make an exposure.
8. At the end of the exposure, one of the following messages will be displayed:
a. CHK OK, indicating that the Model 4000M+ has passed the calibration check.
b. CHK Error LOW or CHK Error HIGH, indicating that the Model 4000M+ has NOT passed the
calibration check. Repeat Steps 2 through 6.
2
If the unit continues to fail the calibration test, contact Fluke Biomedical at 440.248.9300 for further
instructions.
2.7 Radiographic Measurements
Use the following procedure to make kVp, exposure, and time measurements with the Model 4000M+ in
the Radiographic Mode:
1. Turn the Power Switch to ON. The Model No., software revision level, and filter setting will be
displayed, followed by a FLUORO Ready or RADIO Ready indication, depending on the position of
the FLUORO/RADIO Switch.
2. Position the Mode Selection Switch to RADIO.
3. Position the EXP/ALL Switch to ALL.
4. Position the Sensitivity Selection Switch to HIGH or LOW as required. (Refer to Sensitivity
Selection in Section 3 – Theory of Operation.)
5. Position the Tube Target (Anode)/Filter Selection Switch to Mo/Mo or W/Al as required. (Refer to
Tube Target (Anode)/Filter Selection in Section 3 – Theory of Operation.)
6. Rotate the filter wheel to the appropriate kVp setting, depending on the kVp setting of the X-Ray
machine. The RADIO Ready indication should appear on the display.
If there is a hardware failure, an ERROR message
may be displayed instead of the RADIO Ready
condition. Error message indications and
meanings are discussed in Section 3 – Theory of
Operation.
7. Position the detector as discussed in Section 3 – Theory of Operation.
8. Make an exposure.
2-7
Victoreen 4000M+
Operators Manual
9. At the end of the exposure, the value selected when the display was placed in the Manual Roll
Mode will be displayed. If the selected value is Average kVp, the following will be displayed:
XX.X kVp Avg
10. If the display is in Auto Roll Mode, the effective kVp maximum kVp, exposure time, and tube output
will be displayed in sequence. The display will continue to scroll through the values, changing once
every two seconds, until either:
a. Another exposure is made; when the exposure is complete, Analyzing... will be displayed.
Values for the new exposure will then be displayed, automatically changing once every two
seconds.
b. The ROLL/RST button is pressed and held until roll is displayed in the lower right corner of the
display. This puts the display in the Manual Roll Mode. Go to Step 11.
The default display mode is Auto Roll; i.e., on
power up, when the Measurement Mode Switch is
changed from FLUORO to RADIO, the sensitivity
selection (HIGH/LOW) is changed, or the filter
wheel is repositioned, the display will automatically
go into the Auto Roll Mode.
11. If the display is in the Manual Roll Mode:
a. Press ROLL/RST. Assuming, from Step 9, that the Average kVp value is displayed, pressing
ROLL will display the effective kVp as:
XX.X kVp Eff
b. Press ROLL/RST. The maximum kVp will be displayed as: XX.X kVp Max
c. Press ROLL/RST. The exposure time will be displayed as:
XX.XX Sec
d. Press ROLL/RST. The tube output will be displayed as:
XX.XX mR.
e. Press ROLL/RST. The average kVp will be displayed as: XX.X kVp Avg
f. Repeat Steps a through e as desired to scroll through the data.
g. Press and hold the ROLL/RST button for longer than two seconds to place the display in the
Autoroll Mode. The display will continuously scroll through the measured quantities as if the
ROLL/RST button were being pressed once every two seconds. Press ROLL/RST again to exit
the Autoroll Mode.
12. Repeat Steps 3 through 11 as required.
The Model 4000M+ is auto resetting; therefore,
another exposure can be made at any time. At the
end of each exposure, Analyzing... will appear on
the screen. The updated values will automatically
be displayed in the previously selected display
mode (Auto Roll or Manual Roll).
2-8
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.