Supplement
Title: 374, 375, 376 Users Supplement Issue: 5
Part Number: 3608883 Issue Date: 10/12
Print Date: July 2010 Page Count: 2
Revision/Date:
This supplement contains information necessary to ensure the accuracy
of the above Users Manual.
© 2010-2012 Fluke Corporation.
All rights reserved.

374, 375, 376 Users Manual Supplement
Change #1
On page 3, under Warning add the following:
When batteries are changed, ensure that the calibration
seal in the battery compartment is not damaged. If
damaged, the Meter may not be safe to use. Return the
Meter to Fluke for replacement of the seal.
Change #2, 56750
On page 27, under Operating Altitude:
Change
To: Operating Altitude 2000 meters
: Operating Altitude 3000 meters
Change #3
On page 4, add:
Measure a known voltage to verify product operation.
5/12 1

Manual Supplement 374, 375, 376 Users
Change #4,
63774
On page 7, add the following to the Symbols table:
Conforms to relevant South Korean EMC Standards.
Conforms to CAN/CSA-C22.2 No. 61010-1, second edition, including
Amendment 1.
On page 27, add the following to EMC:
Applies to use in Korea only……Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[1] This product meets requirements for
industrial (Class A) electromagnetic wave
equipment and the seller or user should take
notice of it. This equipment is intended for
use in business environments and is not to
be used in homes.
On page 28, replace Agency Approvals with:
Agency Approvals……………………, , ,
2 10/12
[1]