
Nuclear Associates 37-001
TEL-ALIGN™ Teletherapy Alignment Gauge
March 2005
Manual No. 37-001-1 Rev. 2
©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies
Users Manual

Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms

Table of Contents
Section 1: Introduction................................................................................................ 1-1
1.1 Introduction .................................................................................................. 1-1
1.2 Equipment Required .................................................................................... 1-1
1.3 Test Procedure ............................................................................................ 1-1
1.4 Evaluation .................................................................................................... 1-1
1.5 Evaluation Results ....................................................................................... 1-2
i

Introduction
Introduction
Section 1
Introduction
1.1 Introduction
Precise radiation therapy requires treatment geometry that is accurate and reproducible. Since machine
parameters can easily become misaligned, they should be checked frequently. The TEL-ALIGN™
Teletherapy Alignment Gauge* permits a quick check of these parameters and facilitates the adjustment
of those which require realignment. It is especially useful in an overall QA program.
1.2 Equipment Required
The TEL-ALIGN consists of a rectangular plastic base, 17 cm x 16 cm x 5 cm, with a removable vertical
scale, 18 cm high. The base is placed on the treatment table and leveled, using the adjusting pads and
the circular bubble level. The upper surface of the base contains lead markers that form a square (10 cm
x 10 cm) for visualization on film. A crosshair in the center of the square lines up with two additional sets
of crosshairs, one on each outer edge of the base.
1
a. TEL-ALIGN Teletherapy Alignment Gauge
b. Ready-Pack Localization Film
1.3 Test Procedure
A. Place the central axis of the gantry in a vertical position. Set all angles (head rotation, gantry,
collimator and treatment couch at 0° position.
B. Set a 10 cm x 10 cm field size at isocenter.
C. Place the TEL-ALIGN base plate on the Mylar portion of the table. Position and level the plate
symmetrically with the 10 cm x 10 cm light field.
D. Position the top surface of the TEL-ALIGN at the isocenter distance.
1.4 Evaluation
A. Optical Distance Indicator (ODI)
Place the vertical ODI scale on the base plate. The coincidence of the machine's ODI scale can
then be checked with the TEL-ALIGN scale. The vertical scale can then be removed.
B. Collimator
Field size and crosshairs should coincide with the TEL-ALIGN, indicating proper alignment. The
symmetry of the collimator jaws, the alignment of the crosshairs, and angle indicators can be
checked at 90° increments by rotating the collimator.
*Designed and developed by the Medical Physics Department, Memorial Sloan Kettering Hospital, New York, NY
10021.
1-1

Nuclear Associates 37-001
Operators Manual
C. Head Rotation and Pitch
By exercising these motions and returning to the zero settings, the accuracy of the zero indicator or
interlock can be demonstrated by observing the coincidence of the machine crosshairs with the
TEL-ALIGN.
D. Isocenter
By rotating the gantry from 0° to 90°, 180° and 270°, any variation of coincidence can be tracked by
the projection of the crosshairs and corresponding crosses on the TEL-ALIGN.
E. Side Lights and/or Back Pointer
Projection and intersection of the corresponding crosses at 0°, 90°, 180° and 270° will demonstrate
alignment.
F. Light Field vs. Radiation Field
Ready-pack film is placed under the TEL-ALIGN gauge, and an appropriate exposure is given. The
coincidence of the radiation field edges, with the embedded lead markers in the gauge, can then be
verified on the developed film.
1.5 Evaluation Results
The allowed tolerance for alignment of the above parameters is usually of the order of 1-2 mm. However,
the manufacturer's specifications should be consulted for a specific machine. When the tests demonstrate
misalignment, adjustments can be made to the unit with the TEL-ALIGN in position on the table.
1-2

Fluke Biomedical
Radiation Management Services
6045 Cochran Road
Cleveland, Ohio 44139
440.498.2564
www.flukebiomedical.com/rms