Fluke 368-FC Specifications

TECHNICAL DATA
Fluke 368/368 FC and 369/369 FC Leakage Current Clamp Meters
Measure leakage current to track small problems, avoid large problems
The Fluke 360 Series of true-rms leakage current clamp meters help users detect, document, record and compare leakage current readings over time as a means of preventing unplanned downtime, and identifying intermittent GFCI and RCD trips, without taking equipment off line.
The Fluke 368 FC and 369 FC have large diameter jaws (40 mm for the 368 FC, 61 mm for the 369 FC) for work with larger conduc­tors. The clamp’s jaw is fully shielded to accurately capture very small leakage signals, and to minimize external electromagnetic interference.
The 368 FC and 369 FC also offer wireless data connectivity via Fluke Connect. As part of Fluke Connect—the industry’s largest system of software and more than 40 wireless test tools—the 368FC and 369 FC allow you to track changes in leakage current over time, helping to identify potential problems before they turn into major failures.
FLUKE CONNECT COMPATIBLE (368 FC, 369 FC ONLY)
Download the free Fluke Connect® app from iTunes or Google Play to enable graphs for trending
LA RG E JAW
61 mm jaw (369 FC, 369) for use with large cables
WORKLIGHT, BACKLIGHT
Forward-facing worklight, plus backlit display for use in dark wiring cabinets
SPECIALIZED CURRECT MEASUREMENTS
Highest resolution of 1 μA, measure up to 60 A, true-rms for accurate measurements of complex signals
Technical specifications
Product overview
True-rms measurements for accuracy when measuring
complex, non-sinusoidal waveforms
61 mm large jaw design (369/369 FC), 40 mm jaw
(368/368 FC)
Highest resolution of 1 μA, measure up to 60 A
Selectable filter function eliminates unwanted noise
Max/Min/Average readings and hold function
Forward-facing LED worklight for use in dark wiring cabinets
Backlit display; auto backlight off and auto power off for
extended battery life
CAT III 600 V safety rating
Measurement function Alternating current
AC ampere range 3 mA, 30 mA, 300 mA, 3 A, 30 A, 60 A
Range selection mA A: manual selection
3 mA, 30 mA, 300 mA: automatic selection
3 A, 30 A, 60 A: automatic selection
Range/Resolution 3 mA
Accuracy – filter activated (40 to 70 Hz), filter off (40 to 1 kHz)
Accuracy – filter activated (40 to 70 Hz), filter off (40 to 1 kHz)
Frequency 40 Hz to 1 kHz
Crest factor 3
After calibration accuracy is valid for one year. Accuracy is expressed as ± (% of reading + digits). Reference conditions 23 ± 5 °C and a maximum relative humidity of 80 %.
[1]
The minimum value is 10 μA rms.
[2]
[2]
[2]
Outside of the TC/°C 18 °C to 28 °C, 0.02 + 1
[1]
/0.0 01 mA
30 mA/0.01 mA
300 mA/0.1 mA
3 A/0.001 A
30 A/0.01 A
60 A/0.1 A
368/368 FC
3 mA – 30 A 1 % + 5 digits
60 A 2 % + 5 digits
369/369 FC
3 mA to 30 A 1.5 % + 5 digits
60 A 2 % + 5 digits
2 Fluke Corporation Fluke 368/368 FC and 369/369 FC Leakage Current Clamp Meters
Loading...
+ 2 hidden pages