Fluke 1630 User Manual

Manual Supplement
Manual Title: 1630 Users Supplement Issue: 4 Part Number: 2729710 Issue Date: 9/08 Print Date: October 2006 Page Count: 3 Revision/Date:
© 2006-2008 Fluke Corporation. All rights reserved. Printed in the Chinax
1630 Users Manual Supplement
Change #1, 39016, 39045
On page 4, Table 1, change the following numbering:
From:
To:
G, H, I
H, I, G
On page 9, under Using the High and Low Alarms, replace the first sentence in step 2 with the following:
Press the E or D button to increment/decrement the value by 1 Ω.
On page 14, under Electrical Specifications, change the following:
From: Protective Type…………………..IP23 according to IEC 60529/EN 60529
To: Protective Type…………………..IP30 according to IEC 60529/EN 60529
From: Power Requirement……………..9 V alkaline (type IEC
6F22, NEDA 1604)
To: Power Requirement……………..9 V alkaline (type IEC
6 LR 61 NEDA 1604A)
From: Accuracy of Calibration Plate…..± 0.5 %
To: Accuracy of Calibration Plate…..± 1 %
Under General Specifications, change the following:
From: Conductor Size…………………..35 mm (1.38 in)
approximately
To: Conductor Size…………………..33 mm (1.3 in)
approximately
11/06 1