Fluke 16 Multimeter
With Thermometer
It’s a thermometer
The Fluke 16 measures
temperature quickly and accurately.
Standard equipment includes an
80PK-1 bead probe thermocouple
and an 80AK temperature adapter,
which is compatible with a wide range of type K
thermocouples with miniconnectors.
It’s a flame rod circuit tester
A microamp function is included for
flame sensor measurements down
to 0.1 microamps.
It’s a multimeter with
capacitance
Use the Fluke test motor start
and run capacitors up to
10,000 microfarads. Rely on the
Fluke 16 to make your everyday
measurements - up to 600 volts ac
or dc, 40 megohms, diode test and
exceptionally fast continuity.
• Accurate temperature
measurement from -40
ο
(-40
to 400οC)
• Confidently test flame sensor
circuits with microamp measurement down to 0.1 microamps
• Capacitance up to
10,000 microfarads
• Min/Max recording with relative
time stamp records the highest
and lowest voltage readings
during a 100-hour period
• Auto-ranging
• High resistance ranges to 40 MΩ
• Millivolt range for compatibility
with other accessories
• V-Check™ mode is designed for
safety. When measuring ohms,
continuity, or diode test modes,
the meter automatically switches
to measuring ac or dc volts if more
than 4.5V is present
ο
to 750οF
• Continuity Capture™ locates
intermittent opens as brief as
250 microseconds and identifies
them in a symbolic display as shortto-open or open-to-short
• Sleep mode preserves battery life
• 4000 count digital display
• Meter includes: 80PK-1 Bead Temperature Probe, 80AK Temperature
Adapter, TL75 Test Leads, and C10
Holster
Safety
600V ac/dc, maximum voltage
between any terminal and earth
ground. Conforms to the
requirements of IEC-1010-1,
1990-09 and ANSI/ISA-S82.01-94 for
use in Overvoltage Category III
locations. Make ac or dc voltage
measurements on electrical
distribution systems to 600V where
transients do not exceed 6 kV and
only on circuits limited to 14 kVA if
the circuit voltage rating is above
250V. Certification agencies
(approvals/listings pending):
UL Per standard UL 3111
CSA Per standard CSA/CAN
C22.2 No.1010.1-92
TÜV Per standard EN61010
Part 1-1993
Drop test
ANSI/ISA-S82.01-94 and EN61010
Part1-1993. 10 ft. drop to hardwood
on concrete
Size
(HxWxL) 1.12 in x 2.95 in x 6.55 in,
(2.8 cm x 7.5 cm x 16.6 cm)
Battery life
650 hours with Alkaline
Weight
12 oz (340g)
Warranty
Three years

Specifications
Function Range Resolution Accuracy
DC Volts
AC Volts
Current ac/dc
Ohms
Capacitance
Temperature
4000 mV
4.000V 0.001V 0.9% + 2
40.00V 0.01V 0.9% + 1
400.0V 0.1V 0.9% + 1
600V 1V 0.9% + 1
4000 mV
4.000V 0.001V 1.9% + 3
40.00V 0.01V 1.9% + 3
400.0V 0.1V 1.9% + 3
600V 1V 1.9% + 3
200 µA 0.1 µA
400.0 Ω 0.1 Ω
4.000 kΩ 0.001 kΩ
40.00 kΩ 0.01 kΩ
400.0 kΩ 0.1 kΩ
4.000 MΩ 0.001 MΩ
40.00 MΩ 0.01 MΩ
1.000 µF 0.001 µF 1.9% + 2
10.00 µF 0.01 µF 1.9% + 2
100.0 µF 00.1 µF 1.9% + 2
1000 µF 001 µF 1.9% + 2
10,000 µF 0010 µF 10% + 90 typical
15 to 750° F 0.1° F 1% + 1.5° F typical
- 40 to 15° F 0.1° F 5% + 2.7° F
-10 to 400° C 0.1° C 1% + .8° C typical
-40 to -10° C 0.1° C 5% + 1.5° C
1
1
1 mV 0.9% + 2
1 mV 1.9% + 3
0.5% + 2
0.9% + 2
0.9% + 1
0.9% + 1
0.9% + 1
0.9% + 1
1.5% + 3
Accessories
and Ordering
Information
Included Accessories
80PK-1 Bead Temperature Probe
80AK Temperature Adapter
TL75 Test Leads
C10 Holster
Ordering Information
Fluke 16 Multimeter with
Temperature
Need Technical Assistance? Call:
1-800-44-FLUKE
Toll-free in the U.S.
(905) 890-7600 in Canada
(425) 356-5500 other countries
2
2
2
2
2
3
3
3
3
For more information to be sent to
you by fax, call 1-800-FLUKE-FAX
in the U.S. and Canada, or visit our
Website at http://www.fluke.com
1
The 4000 mV range can only be entered in manual range mode.
use the 4000 mV range with accessories.
2
The accuracy specification is for capacitors that have negligible dielectric absorption.
Capacitance measured with rc charge technique that evaluates C = Q/V. Capacitors
with non-negligible dielectric absorption can be measured, but measurement values
may be in error.
3
The accuracy specification for temperature do not include the error caused
by the thermocouple.