Escort MU-IN-D-8-L, MU-OE-D-8-L, MU-OE-D-16-L, MU-ST-D-8-L, MU-IN-D-16-L User Manual

Page 1
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Gu i de v1.1
Version 1.1
Escort iMiniPlus PDF Data Logger
Quick User Guide
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066 Ph: 540-254-1433 Fax: 540-254-2433 Email: techsupport@escortdataloggers.com
Product code:
MU-IN-D-8-L
MU-OE-D-8-L
MU-ST-D-8-L
MU-IN-D-16-L
MU-OE-D-16-L
Page 2
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Gu i de v1.1
Version 1.1
Contents
1 Specifications ........................................................................................... 3
2 Getting Started – What you need ............................................................... 4
3 Front view ............................................................................................... 5
Rear view ................................................................................................ 5
4
5 How iMiniPlus PDF Logger works ................................................................ 6
6 Installing Escort Console Software to PC ..................................................... 6
7 FDA 21 CFR Part 11 compliance ................................................................. 8
8 Connecting logger to PC ............................................................................ 8
9 PDF report contents .................................................................................. 9
10 Programming logger with Escort Console ................................................... 12
10.1 Customizing PDF report and changing details ..................................... 12
10.2 Programming password ................................................................... 13
10.3 Configure Sensors and Alarm Settings............................................... 14
10.3.1 Specification and Alarm .................................................................... 14
10.3.2 LCD Display ................................................................................ 16
10.4 Checking the time clock .................................................................. 16
10.5 Setting the Start, Stop and logging duration ...................................... 17
10.5.1 Duration of trip, at least .......................................................... 17
10.5.2 Interval between each reading ................................................. 18
10.5.3 Start new log trip ................................................................... 18
10.5.4 Finish log trip ......................................................................... 18
10.5.5 Continuous logging ................................................................. 19
10.5.6 Enable stop button in this logger(s) .......................................... 19
11 Viewing logger data with PDF Reader Softwa re .......................................... 20
12 Downloading logger readings via Escort Console Pro ................................... 20
12.1 Download wizard ............................................................................ 21
13 Does Console work with PDF file? ............................................................. 22
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066 Ph: 540-254-1433 Fax: 540-254-2433 Email: techsupport@escortdataloggers.com
Page 3
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
iMiniPlus PDF 8K iMiniPlus PDF 16K Operating Temperature -40°C to +70°C ( -40°F to +158°F) -40°C to +70°C ( -40°F to +158° F) Product Code MU-IN-D-8-L MU-IN-D-16-L MU-OE-D-8-L MU-OE-D-16-L MU-ST-D-8-L Sensors Internal or External options Internal or External options LCD display Large 5 digits Large 5 digits Start option “START” button “START” button *"RESTART" feature with ‘START’ button Stop option “STOP” button or PC “STOP” button or PC Book marking YES YES Continuous Logging mode YES YES Display resolution 0.1 ºC 0.1 ºC Accuracy ±0.5 ºC[-40 ºC to-10 ºC] ±0.5 ºC[-40 ºC to-10 ºC] ±0.3 ºC[-10 ºC to+70ºC] ±0.3 ºC[-10 ºC to+70ºC] ±1.0 ºF [-40°F to +14°F] ±1.0 ºF [-40°F to +14°F] ±0.6 ºF [+14° F to +158°F] ±0.6 ºF [+14° F to +158°F] Sensor response time INT – T90 of 20 min in moving air INT – T90 of 20 min in moving air
EXT – T90 of 5 min in movi ng air EXT – T90 of 5 min in moving air Log interval 5 seco n ds to 18 hours 5 seconds to 18 hours Battery Low icon on LCD YES YES Memory capacity 8048 16240 Communications USB USB Trip counter YES YES Lease counter YES YES High / Low Alarms Consecutive Consecutive Non Consecutive Non Consecutive Catastrophe Alarm Catastrophe Alarm Integrator Alarm Integrator Alarm Start delay after “STAR T”
button pressed File formats PDF, LCF, LRF, CSV, SYLK, Text PDF, LCF, LRF, CSV, SYLK, Text
Program Password protection LCD Display Latest, Highest, Lowest, Average Latest, Highest, Lowest, Average
Time over Specifications Time over Specifications Time under Specifications Time under Specifications Low/ High Alarm Low/ High Alarm Degrees ºF or ºC Degrees ºF or ºC
*Restart fe at u re with 'START' button only applicable if pr e v ious logged tr i p has not trigger ed 'HIGH ALARM' and/or 'LO W ALARM'
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
Version 1.1
1 Specifications
1 minute to 4 hours 15 minutes 1 minute to 4 hours 15 minutes
YES YES
- 3 - Email: techsupport@escortdataloggers.com
Page 4
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
2 Getting Started – What you need
A B C D
A: Escort Console Pro Software (Software Disk)
Version 1.1
Software is also available at www.escortdataloggers.com
B: PC Workstation or Notebook/Laptop
PDF Reader software installed
C: USB to Mini-B plug (5-pin) D: Escort iMiniPlus PDF Data Logger
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 4 - Email: techsupport@escortdataloggers.com
Page 5
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
3 Front view
Display & buttons
Version 1.1
4 Rear view
USB connector, Product label and battery
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 5 - Email: techsupport@escortdataloggers.com
Page 6
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
The iMiniPlus PDF Logger comes with water-proof covers fo r th e USB connection and the battery.
Make sure that when logger is in use these covers are in place. This is important in wet or dusty environments. If covers are lost please contact your distributor for a replacement.
When replacing the battery align one triangular mark on the battery cover against similar shaped mark on the logger base to avoid damaging plastic lock feature of the battery cover part.
5 How iMiniPlus PDF Logger works
iMiniPlus PDF Logger needs to be programmed by Escort Console for logging trip before use. (16K iMiniPlus PDF loggers can be res tarted by 'START' button, if during the program 'Restart' option is enabled only applicab le if previous logged trip has not triggered 'HIGH ALARM' and/or 'LOW AL A RM')
There is no need to install USB drivers on PC. iMiniP lus PDF Logger is configured as generic USB Mass Storage Device. All operating systems should install this device automatically using default USB dr ivers.
iMiniPlus PDF does not require any proprietary software installation to download data from logger after logging trip is finished. iMiniPlus PDF normally generates 3 files once connected to USB.
PDF report. This file is used to view logger data. This file is not present if the logger has been configured not to produce it.
CONTROL.BIN file is used for communication purposes with Escort software
.LRF file is used by Escort Console to create secure .LCF file for traceable data records.
iMiniPlus PDF Logger does not consume battery power when connected to USB and cannot
measure or recognize change in Battery status.
iMiniPlus PDF Logger checks battery status every 15 minutes when not connected to USB and
updates Battery Low icon on LCD.
Escort Console checks battery status of the logger when programming the logger for new trip.
Please note that battery low detect circuit is only accurate when electronics and battery of the is at room temperatures from 15 °C to 25 °C or from 59 °F to 77 °F. If logger is used in cold environment it may report battery low prematurely on LCD display. In this case remove logger from cold and check again battery status after 15 minutes or when the temperature of electronics has risen to room temperature range.
Version 1.1
6 Installing Escort Console Software to PC
Install the free Escort Console Pro Software to your PC/ Laptop using either
1. The Escort Console Software CD or
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 6 - Email: techsupport@escortdataloggers.com
Page 7
Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
2. Through the escort website www.escortdataloggers.com See instruct ion below for Website Installation
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 7 - Email: techsupport@escortdataloggers.com
Page 8
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
3. Now that you are registered, return to the downloading process by clicking on one of the download buttons (same as above); you may be required to log in again. From there you should be able to follow box steps during the process.
Once you have downloaded and installed the Escort Console Pro Software
4.
5. Install Escort Console pro software on your workstation
For a more in-depth guide to Escort Console Installation, please refer to the user manual provided on the website:
http://escortdataloggers.com/manuals/Software/Escort_Console_User_Manual_Engl ish_US.pdf
www.escortdataloggers.com or
7 FDA 21 CFR Part 11 compliance
The PDF report should not be used to provide traceable and auditable data record. The P DF report can be chan ged by the user. This i s contrary t o the FDA 21 CFR Part 11 requirements to data security. In order to produce secure and traceable rec ord, users h ave to open LRF with Escort Console and create an LCF file. There is no need to physically connect logger to PC and download again with
Escort Console i f logger finished trip. You can just mail LRF and open it with Escort Console.
Version 1.1
8 Connecting logger to PC
Unpack the logger. You will find a USB cable in the box with the logger.
The cable has two different ends. The smaller one i s type Mi ni B and t he bigg er one is type A. Connect USB ca ble with Mini B (smaller) connector to the logger and A size to PC or laptop computer as illustrated below.
Simply connect the logger to your PC and wait until the Operating System installs device drivers. Plea se allow Windows sufficient time to install the device. Please do not connect multiple loggers at the same time since it may cause unexpected behavior.
Once the device is installed, you can see the logger as External Removable Drive named ESCORT MPP in “My Computer” and in Escort Console.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 8 - Email: techsupport@escortdataloggers.com
Page 9
Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
Once iMiniPlus PDF data logger is plugged in to wo rk station, A new Drive appears in ‘My Computer’
If double clicked ESCORT MPP drive
9 PDF report contents
Open (double click) RECORD.PDF You can disable PDF report generation by programming logger with Escort Console software. You can also en able or di sable certa in parts of the repo rt as desc ribed by the document named “iMiniPlus PDF Logger User guide - working with Console”.
Accuracy of statistical values displayed in PDF report in Fahrenheit temperature scale is± 0.1 °F. Accuracy for time out specifications values is ± 1 second. If first or last reading exceeds specification only half of the log interval is added to the time out of specifications.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 9 - Email: techsupport@escortdataloggers.com
Page 10
Issue Date: 23 May 2011
•Alarm limits programmed to the logger
triggered
Escort iMiniPlus PDF Logger Quick User Guide v1.1
Page1 of PDF report
Header: Each header displays logger serial number and date /
time information from logger without time zone and daylight saving
Version 1.1
•Product Code
•Serial Number of the logger
•Hardware version – this is actually firmware version
•Description string programmed to the logger by Escort Console
•Battery status last measured by the logger before it has been connected to PC
•Time Zone and daylight savings programmed to the logger
•Date / Time when logging trip started
•Date / Time when logging trip finished
•Current number of readings a nd total duration of the trip
•Logger temperature range
Statistics part includes:
•Highest temperature reading for the current trip
•Average temperature reading for the current trip. Please note:
if the logger is configured for continuous logging, logger
Device Specification
memory is getting overwritten and logger only holds the last 8048 readings. Logger computes average that in cl ud es al l rea di n gs even the ones already overwri tt en . Es co rt C on s ole computes average based on readings retrieved from logger memory and result may be diffe r ent.
•Lowest temperature reading for the current trip
•High Alarm status
•Low Alarm status
•Both High and Low Alarm can h ave the following status: Disabled – Alarm was not enabled when logger was
programmed
Activated – Alarm has been programmed and triggered Not activated – Alarm was programmed but has not been
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
This part of the report is a graph of readings for the current tr ip. This part of the report can be disabled when programming. You can change date display format to display month first by programming logger. Date and Time is d isplayed along the X axis. Temperature values are displayed up the Y axis. If Alarms are disabled Alarm limits are not displayed on the graph
- 10 - Email: techsupport@escortdataloggers.com
Page 11
Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
Page2 of PDF report
Out of specifications
Out of specification is a repor t of how long the logger spent outside of alarm limits This part of the report can be disabled when p rogramming Each line in this report can be read as :” Log ge r has spe nt 49 min utes and 40 se conds between 25 °C and 25.9 °C” In this case the value of 1 degree per report line is configured by programmable parameter named “Increment”.
Note: if Fahrenheit te m perature unit scale is selected, the accuracy of “Out of specification” report summary increment limits are ±0.1°F due to the complexity of conversion between Fahrenheit and Celsius. The same report produced by Escort Console may differ from PDF report produced by the logger in term of temperature by ±0.1°F. Duration of out of specification and the list of readings will read the same.
Page3 of PDF report
List of readings
List of readings
This part of the report always has two columns.
This part of the report can be disabled when programming
You can change date display format to display month first by programming logger
Marked readings are displayed with asterisk sign.
If logger has 8048 readings, the list of readings will be 73 pages long.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 11 - Email: techsupport@escortdataloggers.com
Page 12
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
10 Programming logger with Escort Console
You have to use Escort Console Professional version 2.10.03 or later to program the logger. (For 16K logger MU-IN-D-16-L Version 2.12.07 or above)
Check this check box and click on “Program and Configure” button on the toolbar.
Program Wizard should start. Then click “Next”. Logger should be recognized and displayed.
Version 1.1
10.1 Customizing PDF report and changing details
Click on “Next” butto n. You can see editabl e “Description” field and options for customizing PDF report:
You can customise PDF report generated by the logger, by checking or un checking these boxes.
If you check all boxes, the logger will generate a full report that can be several pages long depending on the nu m ber of readings.
If you uncheck “Create PDF file” checkbox, the logger will not generate PDF report at all. You may need this option to prevent unwanted, unauthorized or incidental access to the data from the logger without Escort Software.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 12 - Email: techsupport@escortdataloggers.com
Page 13
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
If you uncheck “Show summary in PDF file” the logger will not generate the section of the report that summarizes alarm violations of the logging trip
If you uncheck “Show Graph in PDF” the logger will not generate the graph
If you uncheck “Show list of readings in PDF file”, the logger will not generate
the list of readings report
You can configure the logger to display date in month / day / year or day / month / year format by choosing one option from drop down box on this tab as shown below:
10.2 Programming password
The iMiniPlus PDF Logger supports a password option that, when enabled, results in users being required to provide the correct password to gain further access to the iMiniPlus PDF Logger. When the password option is enabled, a window similar to the following picture will be displayed allowing th e user to enter the correct password.
The Program and Configure process will only continue successfully if the correct password is supplied by the user. Passwords are digits only, up to 4 characters. You can enter a password to safeguard your logger from unauthorized tampering. Make sure that you keep a note of your password and logger serial number in a safe place. If you lose or forget the password, the logger will have to be returned to your nearest distributor or Escort Data Logging Systems to restore default password.
Change Password – if you have password protected the iMiniPlus PDF Logger or want to add a password for the first time you can do this here.
Version 1.1
Protect program settings - Tick this option if you want to protect your programming settings for the current logger.
Protect retrieva l of readings- Tick this option to disallow viewing of data from the current logger without a password.
After complete the password section, click OK Select the Next button or Sensors tab.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 13 - Email: techsupport@escortdataloggers.com
Page 14
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
10.3 Configure Sensors and Alarm Settings
Sensors Tab – this displays the sensor information some of which can be changed by using Configure Sensor(s) and others that remain fixed i.e. range ­40°C to 70°C. The description “Internal” refers to the type of sensor as opposed to “External” which will display on t he external sensor mo del.
Configure Sensor(s)… - click on this button to set up the alarm and LCD
Version 1.1
10.3.1 Specification and Alarm
Raise alarm on below lower spe c re adingstick this box if you want to
activate the low alarm and select the value in the box immediately to the right. Note if the lower spec value is 0°C, then the alarm is triggered at -0.1°C, not
0.0°C as this value is consid ered within s pecification.
Raise alarm on above upper spec readin gs
to activate the high alarm and select the value in the box immediately to the right. Note if the upper spec value is 50°C, then the alarm is triggered at 50.1°C, not 50.0°C as this value is considered within specification.
Increment - the grouping size used for listing out-of-specification readings.
Alarm Type
This feature allows the user to specify the co nditions that will trigger an alarm.
Standard Alarm
Number of consecutive ou t of specification readingsthe alarm is
triggered i f the readi ngs are con secuti vely ove r or under th e speci fied th reshold . For instance if the upper spec is 50°C and the value for the number of consecutive out-of-spec readings is 3 and the logging interval is 10minutes, the readings will need to be in excess of 50°C for 30 minutes before the alarm is
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
tick this b ox if you want
- 14 - Email: techsupport@escortdataloggers.com
Page 15
Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
triggered. If the readings fluctuate outside upper and lower threshold temperatur es, t h e sa me rule applies (providing th e ap p rop riate “raise al arm ” b ox is ticked). On ce the readi ng(s) return insi de alarm limit s, the consecuti ve alarm is reset.
Total number of out of specification readings (cumulative) – the
alarm is triggered if the total number of out of spec readings equals or exceeds the entered value. In this ca se th e a larm is NOT reset if readings retu rn t o w ithin the specified parameters.
Do not set the Total number of out-of-spec readings value to be less than the Consecutive out-of-spec readings as this will mean that the Consecutive out-of-spec readings alarm will never be triggered.
Degree-Minutes Alarm (Integrator Alarm)
The Degree-Minutes Alarm is a special feature that “integrates” temperature and time parameters into one value.
Total number of degree-minutes below specs exceeds -
triggered when the total number of temperature degrees times the minute’s duration below the low alarm spec is greater than the value entered.
the alarm is
Total number of degree-minutes above specs exceeds - the alarm is
triggered when the total number of temperature degrees times the minute’s duration above the upper alarm spec is greater than the value entered.
Catastrophe Alarm
This alarm is always additional to the Standard and Degree-Minute alarms and there is no delay. Do not set Catastrophe Alarm inside normal alarm limits because other alarms will never be triggered.
Instantly raise alarm on read ings below - tick this box if you want to
immediately trigger the low alarm and select the valu e i n t h e b ox immediat el y t o the right. Not e if the lower spec value i s 0 °C, then the a larm is trig gered a t -0.1 °C, not 0.0 C as this value is conside red within specification.
Instantly raise alarm on readin gs above - tick this box if you want to
immediatel y trigger th e high ala rm and select the value in the box imm ediately to the right . Note i f the upp er spec val ue i s 50 °C, then the alarm is triggered at
50.1 °C, not 50.0 °C as this value is considered within specification.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 15 - Email: techsupport@escortdataloggers.com
Page 16
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
The table below demonstrates arbitrary data from an iMiniPlus PDF Logger and how different alarm sources will be triggered. This is complicated scenario for alarm settings based on:
Upper spec alarm: +20°C Number of Consecutive out of spec r ea d ings: 4 Total number of out of spec readings: 5Total number of degree-minutes abov e sp ecs exceeds: 8 Instantly raise alarm on readings above: +23°C
10.3.2 LCD Display
The iMiniPlus PDF Logger LCD can be configured to display or not display the temperature information.
Enable Display
Tick this box if you want to enable the display to show the temperature readings and statistics. If the box is un-ticked, all other features will be grayed out. The iMiniPlus PDF Logger will show “d-off” instead of a temperature value. Additionally you will not be able to scroll through the highest lowest, average and ou t of spec readin g s. Th e w o rd “LATEST” will flash and if any alarm had been triggered, “HIGH ALARM” or “LOW ALARM” will be displayed. The bookmarking function can still be used when the display is disabled. When the iMiniPlus PDF is stoppe d using the Stop button or it stops because it has reached the end of a logging trip, the words “d-off” will disappear and the word “STOPPED” will appear.
Version 1.1
Display Units of measurement
You can select the LCD to display the temperature in Celsius or Fahrenheit.
Display change rate
You can select a frequency or 1, 2, 3 or 4 seconds that the LCD scrolls between the Highest, Lowest, Average and out-of-spec information. Note that the
“LATEST”
So if the iMiniPlus PDF is program med to log every 1 minute that will be the rate at which the
reading will only change at a frequency equal to the logging interval.
“LATEST” reading changes.
10.4 Checking the time clock
There maybe times when the time clock on the iMiniPlus PDF Logger may alter. This is the case when the battery is removed and not replaced within 30seconds.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 16 - Email: techsupport@escortdataloggers.com
Page 17
Issue Date: 23 May 2011
conditions.
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
Date/Time Tab
- Check that the date and time is correct.
If you have removed batteries from the iMiniPlus PDF Logger for any length of time make sure you either select:
Change to the date and time on thi s c omputer (default) Set to
The software will always default to the “Change to the date and time on this computer.” option.
10.5 Setting the Start, Stop and l ogging duration
Startup Tab - Configure the start, recording and stop
8K iMiniPlus PDF logger
16K iMiniPlus PDF with 'auto-restart' feature
10.5.1 Duration of trip, at least
If you are using the iMiniPlus PDF Logger to record the temperature of a trip (shipment) for instance, you have the option of setting up the parameters yourself or let the Console software calculate these automatically for you. The software takes the total memory available (8048 readings for 8K loggers and 16240 readings for 16K loggers) and calculates the maximum interval
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 17 - Email: techsupport@escortdataloggers.com
Page 18
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
between each reading and the finish log trip time and date based on the trip duration you select. Yo u can select 1 day to 60day trip duration this way. You will notice that each time you select trip duration, the logging interval will change.
10.5.2 Interval between each reading
This is sometimes called the logging interval or logging rate and is described as the period o f ti me betw een each l ogged readi ng . If th e in terval is a low number, then the iMiniPlus PDF memory will be fill ed qui ckly comp ared t o a hi gh num ber. If you want to r ec ord acute chang es in temperat u re o v e r short periods of t ime, it is best to select a low num ber or faster l oggin g interval. If y ou want a logger to record readings over a long period of time, then you will be best advised to use a lower logging interval or high number. The fastest logging interval the iMiniPlus PDF Logger can log at is every 5 seconds. The slowest logging interval the iMiniPlus PDF Lo gg er can log is every 18 hours.
10.5.3 Start new log trip
The only way to start an iMiniPlus PDF Logger is by using the START button. With other Escort products it is possible to have a timed or date start as well as a switch start. You can also program a delay into the starting process of up to 4 hours 15 minutes. This will show immediately after the Start button is pressed within the iMiniPlus PDF with the words “IN DELAY”.
Version 1.1
10.5.4 Finish log trip
There are several ways to cease a logging trip or period. You can program the iMiniPlus PDF Logger to log until all the memory is used up (8048 readings) or you may wish to log only 100 readings for instance. Alternatively you can select a certain number of days at which the iMiniPlus PDF will stop logging. For instance you may only want to record the temperature for a discrete period like a few days, after which you are not interested in any further data recording.
At the prescribed time the iMiniPlus PDF L ogger will stop logging and “STOPPED” will be displayed on the LCD.
You can also enable the “STOP” button on the logger. This is disabled by default, because in most applications it is better to take too many readings than too few. Downloading the logger (reading the stored results into the computer) does not interfere wi th a running tri p. However, if you Program the log ger with new tri p settings while it is still running, the previous trip data are lost. In ot her words, be sure to download and save the previous trip before setting up a new one. In this case “READY” is displayed on the LCD.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 18 - Email: techsupport@escortdataloggers.com
Page 19
Issue Date: 23 May 2011
Program tab
Escort iMiniPlus PDF Logger Quick User Guide v1.1
10.5.5 Continuous logging
If you are intending to place the iMiniPlus PDF in an environment that you want to record temperatures 24/7 (for example, laboratory refrigerator), you can select the iMiniPlus PDF Logger to log continuously. This means that the iMiniPlus PDF Logger will log until the memory it filled up at which time it will start overwriting the first logged data with the latest. Users must be careful to routinely download data from the iMiniPlus PDF Logger otherwise da ta may be l ost. Careful sel ection of the l ogging interval is required in this case. It is possible to download at anytime while the iMiniPlus PDF Logger is logging, without changing the program settings.
10.5.6 Enable stop button in this logger(s)
The iMiniPlus PDF Logger Stop butto n can be activated through programming by simply ticking the appropriate box.
Select Next button or Summary tab.
Version 1.1
Select Program button or
Click “Finish” button. If the upload has failed either select the
Cancel button to close the software and start again or select the Restart button to start the sequence again. If the iMiniPlus PDF Logger still fails to upload contact your local distributor or Escort Da ta Loggers.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
Program settings are being uploaded to iMiniPlus PDF Logger
The iMiniPlus PDF Logger has been programmed successfully
- 19 - Email: techsupport@escortdataloggers.com
Page 20
Issue Date: 23 May 2011
Escort iMiniPlus PDF Logger Quick User Guide v1.1
11 Viewing logger data with PDF Reader Software
You can download data from the logger by connecting the logger to a USB port. Once the logger hardware is installed and logger can be recognized by Operating System as Mass S t o rage Device you can op en R EC ORD.PDF file. You can validate logger data by simply looking at PDF report.
Please note that PDF Reader software is not a b le to save opened PDF file to other location if logger has been disconnected from PC after PDF report was opened with PDF Reader.
Please note if you can not save PDF Report opened in PDF Reader software using “Save As” menu item you can always copy file using Windows Explorer and paste it to the desired location.
12 Downloading logger readings via Escort Console Pro
You can download data from the logger by connecting the logger to Console.
Make sure the USB cable is connected to your PC and to the iMiniPlus PDF Logger.
Version 1.1
There are two ways to retrieve the data fro m an iMiniP lus PDF Logger, both of which will achieve the same result.
1. Click the Download Information icon on the toolbar
2. Click the "Download Information..." item located under the Action menu.
The process of retrieving data stored within an iMiniPlus PDF Logger does not erase it from the memory. If the iMiniPlus PDF is currently recording while you retrieve the information, it will continue to record without interruption. The Download Information feature initiates the Download Wizard.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 20 - Email: techsupport@escortdataloggers.com
Page 21
Issue Date: 23 May 2011
any new start conditions to the logger(s)”
Escort iMiniPlus PDF Logger Quick User Guide v1.1
12.1 Download wizard
During the progress of the Download Wizard, you will see a series of 5 buttons at the bottom of the wizard box.
TheCancel” button immediately stops and closes the wizard. To start again you will have to select the Download icon. The start the Wizard progress from the beginning without closing the feature.
“Back” enables you to go back one step in the ste p-by- step progress andNext”
“Download”
about the step you are in. Please note that this feature may other Escort products from time to time. Add itionally some Help documentation may not be present due to the software version you are currently using. Consult
takes you to the next step. The Next button is preceded by the
button. The Help button will provide description and instruction
Restart” button enables you to
make reference to
this iMiniPlus PDF Logger User guide if
Version 1.1
Download Wizard’s start page
Displays Information
You have the option of programming new start conditions. This may be h elp for instance when you are have to reset a large number of iMiniPlus PDF loggers for a new application. Otherwise leave the Logger option at its default setting “ Do not upload
you are not sure about any description.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 21 - Email: techsupport@escortdataloggers.com
Select
Page 22
Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger Quick User Guide v1.1
Click on the Finish button.
Readings & Summary displayed
13 Does Console work with PDF file?
Escort Console neither opens logger PDF report nor creates one. Escort Console only works with LRF and LCF files .
In order to ensure that the PDF file has not been tampered with you need to visually compare it with Console view of the LRF file.
Escort Verification Tec hnologies, Inc. dba ESCORT Data Loggers
PO Box 309, 120 Parkway Drive Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 22 - Email: techsupport@escortdataloggers.com