Endress+Hauser Turbimax CUS71D Operating Instructions Manual

Products Solutions Services
Operating Instructions
Turbimax CUS71D
Ultrasonic interface sensor
BA00490C/07/EN/05.16 71353737
AUTHORIZED DISTRIBUTOR: InstrumentsAndControl.com Houston, Texas USA sales@InstrumentsAndControl.com 281-609-7170
Turbimax CUS71D Table of contents
Endress+Hauser 3
Table of contents
1 Document information ......... 4
1.1 Warnings ........................... 4
1.2 Symbols used ........................ 4
2 Basic safety instructions ....... 5
2.1 Requirements for personnel ........... 5
2.2 Designated use ...................... 5
2.3 Occupational safety .................. 6
2.4 Operational safety ................... 7
2.5 Product safety ....................... 7
3 Product description ............. 8
3.1 Product design ....................... 8
4 Incoming acceptance and
product identification .......... 8
4.1 Incoming acceptance ................. 8
4.2 Product identification ................. 9
5 Installation .................... 11
5.1 Installation conditions ............... 11
5.2 Mounting the sensor ................ 14
5.3 Post-installation check ............... 18
6 Electrical connection .......... 18
6.1 Connecting the sensor ............... 18
6.2 Post-connection check ............... 20
7 Commissioning ................ 20
7.1 Preparatory steps ................... 20
8 Operation ...................... 20
8.1 Sensor monitoring .................. 20
8.2 Cyclic cleaning ...................... 20
8.3 Automatic wiper function ............ 21
9 Diagnostics and
troubleshooting ............... 21
10 Maintenance .................. 21
10.1 Maintenance tasks .................. 22
11 Repairs ......................... 23
11.1 Spare parts ........................ 23
11.2 Return ............................ 23
11.3 Disposal ........................... 23
12 Accessories .................... 24
12.1 Assemblies ........................ 24
12.2 Holder ............................ 25
12.3 Cable extension ..................... 26
13 Technical data ................. 26
Index ................................. 28
Document information Turbimax CUS71D
4 Endress+Hauser
1 Document information
1.1 Warnings
Structure of information Meaning
L
DANGER
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation will result in a fatal or serious injury.
L
WARNING
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid the dangerous situation can result in a fatal or serious injury.
L
CAUTION
Causes (/consequences)
Consequences of non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or more serious injuries.
NOTICE
Cause/situation
Consequences of non-compliance (if applicable)
Action/note
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols used
Symbol Meaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
Turbimax CUS71D Basic safety instructions
Endress+Hauser 5
1.2.1 Symbols on the device
Symbol Meaning
Reference to device documentation
2 Basic safety instructions
2.1 Requirements for personnel
• Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and must follow the instructions contained therein.
• Measuring point faults may be repaired only by authorized and specially trained personnel.
Repairs not described in the Operating Instructions provided may only be carried out directly by the manufacturer or by the service organization.
2.2 Designated use
CUS71D is a sensor designed for interface measurement in water and wastewater. The sensor is particularly suited for use in the following applications:
• Wastewater treatment: primary clarifier, sludge thickener, secondary clarifier
• Water treatment: sedimentation tank after flocculant dosage, sludge height in contact sludge processes
• Static separation processes: with/without slow stirring and without introduction of air
Use of the device for any purpose other than that described, poses a threat to the safety of people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
NOTICE
Non-designated use
Incorrect measurements, malfunctions and even measuring point failure could result
Only use the product in accordance with the product specifications.
Pay particular attention to the technical data on the nameplate.
Basic safety instructions Turbimax CUS71D
6 Endress+Hauser
2.3 Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been connected in accordance with these Operating Instructions.
Turbimax CUS71D Basic safety instructions
Endress+Hauser 7
2.4 Operational safety
1. Before commissioning the entire measuring point, verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged.
2. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified: Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
2.5 Product safety
The product is designed to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. The relevant regulations and European standards have been observed.
We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional protection for the device and device data transfer must be implemented by the operators themselves.
Product description Turbimax CUS71D
8 Endress+Hauser
3 Product description
3.1 Product design
The sensor is designed for continuous in-situ determination of interfaces. The sensor includes all necessary modules:
• Power supply
• The ultrasonic source emits the measurement signals.
• The ultrasonic receiver receives the measurement signals, digitalizes the signals and converts them to a measurement value.
• The sensor microcontroller controls the internal operations and data transmission.
The preconfigured sensor can be used at a measuring point.
3.1.1 Measuring principle
A piezoelectric crystal is enclosed in a flat cylindrical plastic housing. When the crystal is excited by an electrical voltage, it generates a sonar signal. This causes ultrasonic waves to be transmitted at a frequency of approx. 650 kHz and at an angle of 6° to scan the separation zones.
The measured variable is the time the transmitted ultrasonic signal needs to reach the solid particles in the separation zone and return to the receiver. The separation zone is computed after the maximum slope and the maximum signal amplitude.
A sensor version featuring a wiper is available to reliably avoid the buildup of deposit on the sensor membrane.
4 Incoming acceptance and product identification
4.1 Incoming acceptance
1. Verify that the packaging is undamaged.  Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
2. Verify that the contents are undamaged.  Notify your supplier of any damage to the delivery contents.
Keep the damaged products until the matter has been settled.
3. Check the delivery for completeness.  Check it against the delivery papers and your order.
4. Pack the product for storage and transportation in such a way that it is protected against impact and moisture.
The original packaging offers the best protection.
The permitted ambient conditions must be observed (see "Technical data").
If you have any questions, please contact your supplier or your local sales center.
Turbimax CUS71D Incoming acceptance and product identification
Endress+Hauser 9
4.2 Product identification
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Safety information and warnings Compare the data on the nameplate with your order.
4.2.2 Product identification
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to the product page for your product on the Internet.
2. At the bottom of the page, click the link Online Tools and then select Access device specific information.
An additional window opens.
3. Enter the order code from the nameplate into the search field and then select Show details.
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 D-70839 Gerlingen
4.2.3 Scope of delivery
The delivery comprises:
• 1 Turbimax CUS71D sensor, version as ordered
• 1 set of Operating Instructions BA00490C/07/EN If you have any questions, please contact your supplier or your local sales center.
4.2.4 Certificates and approvals
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it complies with the legal specifications of the EU directives. The manufacturer confirms successful testing of the product by affixing to it the mark.
Incoming acceptance and product identification Turbimax CUS71D
10 Endress+Hauser
EAC
The product has been certified according to guidelines TP TC 004/2011 and TP TC 020/2011 which apply in the European Economic Area (EEA). The EAC conformity mark is affixed to the product.
Electromagnetic compatibility
Interference emission and interference immunity: Industrial environment as per
• EN 61326-1:2013
• EN 61326-2-3:2013
• NAMUR NE21: 2012