Endress+Hauser TSM487 Specifications

Technical Information
®
Easytemp TSM487
Compact thermometer with screw-in thread for universal applications
Application
The TSM487 compact thermometer is used for universal applications. Preferred applications are where no high process pressures and no extreme temperatures appear.
Function
The compact thermometer assembly includes a fiberglass insulated insert which is protected by a thermowell with process connection
. The terminal head is according to DIN 43729, form B, and is
G½” made of aluminum. The built-in head transmitter converts the resistance value into a temperature linear 4...20 mA analog output signal.
in vessels or in pipes,
Various measuring ranges selectable
• 2-wire technology, 4… 20 mA
• High accuracy of sensor and electronics
• Fiberglass insulated insert
• Replaceable electronics
Measuring ranges (selectable):
-30... +170 °C (-22... +338 °F)
0... +100 °C (32... +212 °F)
0... +200 °C (32... +392 °F)
Immersion lengths:
mm: 50, 100, 150, 250 (Ø 6) Inch: 2, 3.9, 5.9, 9.8 (Ø 0.24)
Accuracy:
£ 0.08%, Pt100 class A
£ 3.5 s (T ); s (T )
Operating conditions:
20 bar at +20 °C (290 PSI at +68 °F)
Electrical connection
Supply voltage and current output
£ 8
50 90
Application example
Pipe installation:
a) at elbows, against the flow
b) in smaller pipes,
leant against the flow
c) perpendicular to the flow
Ordering information
TSM487
TSM487- order code
Compact thermometer TSM487, RTD
Head transmitter: TMT187; non-replaceable insert in fiberglass insulation with diameter 6 mm (0.24”), 1.4404/SS316L Sensing element: 1xPt100 class A 4-wire; process connection G½”
Immersion length
A 50 mm B 100 mm
C 150 mm D 250 mm
Measuring range TMT187
DD 4... 20 mA; -30... 170 °C FE 4... 20 mA; 0... 100 °C FH 4... 20 mA; 0... 200 °C
TI149R/09/en
71095370
2
1
mA
-
+
8… 35 V
4… 20 mA
Easytemp TSM487
®
Technical data
Sensor
Sensing element
Platinum resistance element, 1x Pt100 (100 at 0 °C)W
Measuring range
-30... 170 °C (-22... 338 °F), 0... 100 °C (32... 212 °F), 0... 200 °C (32... 392 °F)
Accuracy Class A acc. to IEC 751: -50… +250 °C
Wiring
Insulation resistance 100 M , test voltage 250 V at ambient
4-wire connection, fiberglass insulated insert
³W
temperature
Response time T /3.5 s; T /8 s; according to IEC 751
Operating conditions
50 90
20 bar at +20 °C (290 PSI at +68 °F)
Sheat material SS 316L/1.4404
Process connection
Shape DIN 43772 form 2G
Material SS 316L/1.4404
Thread G½”
Terminal head
Type DIN 43729 form B
Protection class IP66/68
Cable entry M20x1.5
Material Aluminum, polyester powder coated
Electronics (replaceable)
Output
Output signal 4… 20 mA, temperature and resistance linear
Max. load (V -8 V)/0.022 A
power supply
Min. current
consumption 3.5 mA
Current limit £ 23 mA
Switch on delay 4 s (during power up I
= 3.8 mA)
a
Response time 1s
Signal on alarm
Under ranging Linear drop to 3.8 mA
Over ranging Linear rise to 20.5 mA
Sensor break/
Sensor short circuit
≥21mA
Electrical connection
Supply voltage U = 8... 35 V, reverse polarity protection
Galvanic isolation Û = 3.75 kV
Residual ripple
Reference operating
conditions
b
U 5 V at U 13 V, f = 1kHz
£³
ss b max.
Calibration temperature: +23°C(73°F)±5K(9°F)
Dimensions in mm (inch)
107.5 (4.23)
78 (3.2)
15.55 (0.6)
70 (2.76)
28
(1.1)
welded
15 (0.6)
G½”
L = Immersion length
Ø 6 (0.24)
Electronics (replaceable)
Accuracy
Influence of
supply voltage
£ ±0.01 %/V deviation from 24 V
Influence of load £W±0.02 %/100
Temperature drift
T = ±(15 ppm/K * max. meas. range +
d
50 ppm/K * preset meas. range) *
DJ
Pt100 0.2 K or 0.08 %
Environment conditions
Ambient temperature -
Climate class
Shock and
40… +85 °C (-58... +185 °F)
As per IEC 60 654-1, class C
vibration resistance 4g/2to150HzasperIEC60068-2-6
EMC Shock resistance and interference emission
as per IEC 61326 and NAMUR NE 21
L
TI149R/09/en/03.09
71095370 CDR X3
Loading...