Technical Information
®
Easytemp TSM187
Compact thermometer with screw-in thread
for challenging applications
Application
The TSM187 compact thermometer range covers a wide variety of
market needs worldwide. Typical application can be found in the
chemical and pharmaceutical industry, food, water and waste water
and power plants. Preferred applications are
where requirements are short response time mechanical strength.
Function
The compact thermometer assembly includes a replaceable insert in
mineral insulated cable which is protected by a thermowell with
process connection . The terminal head is according to DIN
43729, form B, and is thermally uncoupled by a extension neck.
The built-in head transmitter converts the resistance value into a
temperature linear 4...20 mA analog output signal.
in vessels or in pipes,
G½”
• Various measuring ranges selectable
• 2-wire technology, 4… 20 mA
• High accuracy of sensor and electronics
• Replaceable mineral insulated insert
• Replaceable electronics
Measuring ranges (selectable):
-30... +170 °C (-22... +338 °F)
•
0... +100 °C (32... +212 °F)
•
0... +200 °C (32... +392 °F)
•
0... +300 °C (32... +572 °F)
•
Immersion lengths:
mm: 120, 160, 250, 400 (Ø 9)
Inch: 4.7, 6.3, 9.9, 15.8 (Ø 0.35)
Accuracy:
£ 0.08%, Pt100 class A
Response time:
£ 18s(T ); 55s(T )
Operating conditions:
• 50 bar at +20 °C
(725 PSI at +68 °F)
• 1 bar at +400 °C
(14.5 PSI at +752 °F)
Electrical connection
Supply voltage and current output
2
-
8… 35 V
£
50 90
Application example
Pipe installation:
a) at elbows, against the flow
b) in smaller pipes,
leant against the flow
c) perpendicular to the flow
Ordering information
TSM187
TSM187- ⇐ order code
Compact thermometer TSM187, RTD
Head transmitter: TMT187; replaceable mineral (MgO) insulated
insert with diameter 6 mm (0.24”), 1.4404/SS316L
Sensor type: 1xPt100 class A 4-wire; process connection G½”
Immersion length
A 120 mm
B 160 mm
C 250 mm
D 400 mm
Measuring range TMT187
DD 4... 20 mA; -30... 170 °C
FE 4... 20 mA; 0... 100 °C
FH 4... 20 mA; 0... 200 °C
FI 4... 20 mA; 0... 300 °C
a)
b)
c)
TI148R/09/en
71095367
1
mA
+
4… 20 mA
Easytemp TSM187
®
Technical data
Sensor
• Sensing element
Platinum resistance element,
1x Pt100 (100 at 0 °C)W
• Measuring range
-30... 170 °C (-22... 338 °F), 0... 100 °C (32... 212 °F),
0... 200 °C (32... 392 °F), 0... 300 °C (32... 572 °F)
• Accuracy Class A acc. to IEC 751: -50… +250 °C
• Wiring
•
Insulation resistance 100 M , test voltage 250 V at ambient
4-wire connection, MgO isolated
³W
temperature
• Sheat diameter 6 mm (0.24”)
• Response time T /18 s; T /55 s; according to IEC 751
• Operating conditions
50 90
50 bar at +20 °C (725 PSI at +68 °F)
1 bar at +400 °C (14.5 PSI at +752 °F)
Thermowell
• Shape DIN 43772 form 2G
• Diameter 9 mm (0.36”)
• Material SS 316Ti/1.4571
Process connection
• Shape DIN 43772 form 2G
• Material
SS 316Ti/1.4571
• Thread G½”
Terminal head
• Type DIN 43729 form B
• Protection class IP 66/68
Cable entry M20x1.5•
• Material Aluminum, polyester powder coated
Electronics (replaceable)
Output
• Output signal 4… 20 mA, temperature and resistance linear
• Max. load (V -8 V)/0.022 A
power supply
• Min. current
consumption 3.5 mA
≤
• Current limit 23 mA≤
•
Switch on delay 4 s (during power up I
= 3.8 mA)
a
• Response time 1s
Signal on alarm
•
Under ranging Linear drop to 3.8 mA
•
Over ranging Linear rise to 20.5 mA
• Sensor break/
Sensor short circuit
21 mA≥
Dimensions in mm (inch)
107.5 (4.23)
78 (3.2)
15.5 (0.6)
70 (2.76)
28
(1.1)
145 (5.7)
15 (0.6)
G½”
L
L = Immersion length
Ø 9 (0.35)
Electronics (replaceable)
Electrical connection
•
Supply voltage U = 8... 35 V, reverse polarity protection
•
Galvanic isolation Û = 3.75 kV
• Residual ripple
Reference operating
•
conditions
b
£³
U 5 V at U 13 V, f = 1kHz
ss b max.
Calibration temperature:
+23°C(73°F)±5K(9°F)
Accuracy
• Influence of
supply voltage
£ ±0.01 %/V deviation from 24 V
• Influence of load £W±0.02 %/100
• Temperature drift
T = ±(15 ppm/K * max. meas. range +
d
50 ppm/K * preset meas. range) *
DJ
• Pt100 0.2 K or 0.08 %
Environment conditions
• Ambient temperature -
• Climate class
Shock and
•
40… +85 °C (-58... +185 °F)
As per IEC 60 654-1, class C
vibration resistance 4g/2to150HzasperIEC60068-2-6
• EMC Shock resistance and interference emission
as per IEC 61326 and NAMUR NE 21
TI148R/09/en/03.09
71095367
CDR X3