These Operating Instructions contain all the information that is required in various phases
of the life cycle of the device: from product identification, incoming acceptance and
storage, to mounting, connection, operation and commissioning through to
troubleshooting, maintenance and disposal.
1.2 Symbols
1.2.1 Safety symbols
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
This symbol contains information on procedures and other facts which do not result in
personal injury.
1.2.2 Electrical symbols
SymbolMeaning
Direct current
Alternating current
Direct current and alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
1.2.3 Symbols for certain types of information
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
4Endress+Hauser
iTHERM MultiSens Bundle TMS31About this document
,…,
SymbolMeaning
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
Series of steps
Result of a step
Help in the event of a problem
Visual inspection
1.2.4 Documentation
DocumentPurpose and content of the document
iTHERM TMS31 MultiSens Bundle
(TI1443T/01/xx)
Planning aid for your device
The document contains all the technical data on the device and provides
an overview of the accessories and other products that can be ordered for
the device.
The document types listed are available:
In the Download Area of the Endress+Hauser Internet site: www.endress.com →
Downloads
1.2.5 Registered trademarks
• FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus Foundation, Austin, Texas, USA
• HART®
Registered trademark of the HART® FieldComm Group
• PROFIBUS®
Registered trademark of the PROFIBUS Nutzerorganisation e.V. (Profibus User
Organization), Karlsruhe - Germany
Instructions and procedures in the operating instructions may require special precautions
to ensure the safety of the personnel performing the operations. Information that
potentially raises safety issues is indicated by safety pictograms and symbols. Please refer
to the safety messages before performing an operation preceded by pictograms and
symbols. Although the information provided herein is believed to be accurate, be advised
that the information contained herein is NOT a guarantee of satisfactory results.
Specifically, this information is neither a warranty nor guarantee, expressed or implied,
regarding performance. Please note that the manufacturer reserves the right to change
and / or improve the product design and specifications without notice.
2.1 Requirements for the personnel
The personnel for installation, commissioning, diagnostics and maintenance must fulfill
the following requirements:
Trained, qualified specialists must have a relevant qualification for this specific function
‣
and task
Are authorized by the plant owner/operator
‣
Are familiar with federal/national regulations
‣
Before beginning work, the specialist staff must have read and understood the
‣
instructions in the Operating Instructions and supplementary documentation as well as
in the certificates (depending on the application)
Following instructions and basic conditions
‣
The operating personnel must fulfill the following requirements:
Being instructed and authorized according to the requirements of the task by the
‣
facility's owner-operator
Following the instructions in these Operating Instructions
‣
2.2 Designated use
The product is intended to measure the temperature profile inside a tank, a silo or any
storage system, through RTD or thermocouple technologies.
The manufacturer is not liable for damage caused by improper or non-designated use.
The product has been designed according to the following conditions:
ConditionDescription
Internal pressureThe design of joints, threaded connections and sealing elements has been
executed as a function of the maximum allowable pressure inside the storage
vessel.
Operating temperatureThe materials used were chosen according to the operating and design
minimum and maximum temperatures. Thermal displacement has been taken
into account to avoid intrinsic stresses and to ensure proper integration
between the instrument and the plant. Specific care has to be taken when the
instrument’s sensing elements are fixed to the internals.
Stored materialDimensions and choice of materials minimize:
distributed and localized corrosion.
FatigueCyclic loads during operations have been taken into consideration.
VibrationsDuring normal operation the multipoint is not subjected to vibration
phenomena. In case of external vibration induced by other equipment near the
multipoint, the rope system is able to compensate for them.
Mechanical stressThe maximum stress on the measuring device is guaranteed to stay below the
yielding stress of the material, for every working condition.
External environmentThe junction box (with and without head transmitters), wires, cable glands and
other fittings have been selected to work within the allowed ranges in terms of
external temperature.
2.3 Workplace safety
The external installation area must be free of interferences to avoid any injury during
installation, and to avoid any damage to the measuring device.
2.4 Operational safety
Operate the device in proper technical condition and fail-safe condition only.
‣
The operator is responsible for interference-free operation of the device.
‣
Hazardous area
To eliminate a danger for persons or for the facility when the device is used in the
hazardous area (e.g. explosion protection or safety equipment):
Based on the technical data on the nameplate, check whether the ordered device is
‣
permitted for the intended use in the hazardous area. The nameplate can be found on
the side of the junction box or on the extension neck.
Observe the specifications in the separate supplementary documentation that is an
‣
integral part of these Instructions.
Electromagnetic compatibility
The measuring system complies with the general safety requirements in accordance with
EN 61010-1, the EMC requirements of IEC/EN 61326 and NAMUR Recommendation NE
21 and NE 89.
NOTICE
The unit must only be powered by a power supply that operates using an energy-
‣
limited electric circuit that is compliant with IEC 61010-1, "SELV or Class 2 circuit".
2.5 Product safety
The unit is constructed using the most up-to-date production equipment and complies
with the safety requirements of the local guidelines. The temperature measuring system is
fully factory tested according to the specifications indicated on the order and/or any
additional test which is considered as safety-relevant. However, if it is installed incorrectly
or is misused, certain application dangers can occur. Installation, wiring and maintenance
of the unit must only be done by trained, skilled personnel who are authorized to do so by
the plant operator. This skilled staff must have read and understood these instructions and
must follow them. The plant operator must make sure that the measurement system has
been installed by tightening the threaded components (e.g bolts and nuts) with the
predefined torques and tools → 12, and correctly wired according to the wiring
diagrams → 17.
Endress+Hauser7
Product descriptioniTHERM MultiSens Bundle TMS31
1
2b
4
8
3
5
2a
6
7
10b
9
9
11
10a
9
11
2c
3 Product description
3.1 Equipment architecture
The multipoint thermometer belongs to a range of modular product configuration for
multipoint temperature detection with a design where subassemblies and components can
be managed individually for easy maintenance and spare part ordering.
The temperature probe-only version consists of many sub-assemblies:
• Insert
• Rope
• Weight
• Process connection
• Neck (see below for a more detailed description)
In general the instrument measures the temperature profile inside the process
environment by means of many sensors wrapped around a rope, jointed to a suitable
process connection which ensures the right tightness level.
The temperature probe + diagnostic version combines the temperature probe with a head
transmitter, which is available with enhanced accuracy and reliability compared to directly
wired sensors. Output communication protocols available are: Analog output 4 to 20 mA,
HART®, PROFIBUS® PA, FOUNDATION Fieldbus™. Externally the extension cables are
wired into the junction box, which can be directly mounted or remotely as an option.
Design
1MultiSens Bundle design with open support on the left side, MultiSens Bundle design with open frame
with protection in the middle and tube neck design on the right side
A0038296
8Endress+Hauser
iTHERM MultiSens Bundle TMS31Product description
Description and available options
Hinged cover junction box for electrical connections. It includes
components such as electrical terminals, transmitters and cable
1: Head
2a: Open supporting frame
2b: Supporting frame with cover
2c: Tube neck
3: Compression fitting
4: Process connection
5: Eyebolt
6: Toggle joint
7: Ogives
8: Insert
9: Rope
10a: Swage eye
10b: Metric swage thread
11: Weight
glandes.
• 316/316L
• Other materials on request
Modular frame support that is adjustable for all available junction
boxes.
316/316L
Modular support that is adjustable for all available junction boxes
and ensures extension cable inspection.
316/316L
Modular tube frame support adjustable for all available junction
boxes
316/316L
High reliability for tightness between process and external
environment, for a wide range of process fluids concentration and
severe combination between temperature and pressure.
• 316L
• 316H
Represented by a flange according to international standards, or
engineered to satisfy specific process requirements. → 43
Lifting device for easy handling during installation phase.
316
Connection between the rope and the process connection.
316
Insert guide for the correct positioning of the measuring sensing
element.
• 316
• 316L
Thermocouple (type J, K) grounded and ungrounded execution or
RTD (Pt100 wire wound).
Metallic rope.
316
Ring-bolt end connection.
316
Threaded end connection.
316
Weight to maintain the rope pretensioned and in a straight position
during working condition (i.e. tank filling).
• 316
• 316L
Endress+Hauser9
Incoming acceptance and product identificationiTHERM MultiSens Bundle TMS31
Made in Italy2018
20060 Pessano con Bornago-MI
iTHERM MultiSens Bundle
Ser.no.: XXXXXXXXXXX
Ord.cd.: T-XXXX/XXMS31
2xPt100/TF/4Cl.A
no. of inserts: 8
TMT182-B
No of transmitter:12
Ref to BAT/0 /en019551
Measuring range: 0...250 °C
0044
5
6
TSV00XXXX-XXXXX
4 Incoming acceptance and product
identification
4.1 Incoming acceptance
Before proceeding with the installation the following incoming acceptance procedures are
suggested:
• Once the device is received it is always suggested to verify the integrity of the packaging
and possible damages. Non-compliances should be immediately reported to the
manufacturer. Damaged material shall not be installed: in these conditions, in fact, the
manufacturer cannot guarantee the original safety requirements and cannot be
considered responsible for any consequential effect.
• Compare the scope of delivery with the order content.
• Carefully remove all packaging/protection related to the freight.
4.2 Product identification
The following options are available for identification of the device:
• Nameplate specifications
• Enter the serial number from the nameplate in the W@M Device Viewerwww.endress.com/deviceviewer: All data relating to the device and an overview of the
Technical Documentation supplied with the device are displayed.
The following nameplate layout serves to identify the specific product information from
the serial number, design conditions, sizes, configuration to approvals:
A0038307
2Nameplate of the multipoint thermometer (example as landscape format)
Field
number
1Order code and serial numberTMS31-xxxxx
2TSV drawing numberTSV301237-XXXXX
3Sensor and product configuratione.g. number of measuring points
4Assembled transmitter-
5Sensor measuring temperature range-
10Endress+Hauser
DescriptionExamples
iTHERM MultiSens Bundle TMS31Incoming acceptance and product identification
Field
number
6CE marking-
-Approval number, hazardous area classification
DescriptionExamples
e.g.–50 to 60 °C (–58 to 140 °F) for hazardous
and Ex logo (if applicable)
Safety instructions number (if applicable)
Ambient temperature (if hazardous area
classification is applicable)
area application
Compare and check the data on the nameplate of the device against the requirements
of the measurement enviroment.
4.3 Storage and transport
Carefully remove each package and protection relative to the transportation packaging.
NOTICE
Transportation of the device to the installation area
Handle the device by always using the provided eyebolt as the main lifting part.
‣
Handle with care. During mounting phases avoid each load to welded or threaded parts,
‣
under the action of the weight of the device.
When necessary, use additionally ropes to keep the device straight in the area between
‣
the junction box an the first part of the metallic rope.
It is strictly requested to avoid bumping against obstacles nearby the place where the
‣
device is to be installed.
Avoid any friction between the device and the other surrounding bodies.
‣
Avoid twisting of the sensing element twisting.
‣
For ease, keep the thermometer bundle rope in its coil configuration during lifting
‣
operation → 15.
Pack the device in such a way as to protect it reliably against impact for storage (and
transportation). The original packaging provides optimum protection.
For permitted storage temperature → 36
Endress+Hauser11
MountingiTHERM MultiSens Bundle TMS31
5 Mounting
5.1 Mounting requirements
WARNING
L
Failure to follow these installation guidelines could result in death or serious injury
Make sure only qualified personnel perform the installation.
‣
WARNING
L
Explosions could result in death or serious injury
Do not remove the junction box cover in explosive atmospheres when the circuit is live.
‣
Before connecting any additional electric and electronic device in an explosive
‣
atmosphere, make sure the instruments in the loop are installed in accordance with
intrinsically safe or non incendive field wiring practices.
Verify that the operating atmosphere of the transmitters is consistent with the
‣
appropriate hazardous locations certifications.
All covers and threaded components must be fully engaged to meet explosion-proof
‣
requirements.
WARNING
L
Process leaks could result in death or serious injury
Do not release screwed parts while in operation. Install and tighten the fittings before
‣
applying pressure.
NOTICE
Additional loads and vibrations from other plant components can affect the operation
of the sensor elements.
Is it not allowed to apply additional loads or external moments to the system coming
‣
from the connection with another system not foreseen from installation plan.
The system is not suitable for being installed in locations where vibrations are present.
‣
The deriving loads can undermine the sealing of the junctions and damage the
operation of the sensing elements.
It will be care of the final user to verify the installation of suitable devices in order to
‣
avoid the overcoming of the admitted limits.
For the environment conditions please refer to the technical data → 36
‣
While installing the measurement system, avoid any friction during installation,
‣
specifically avoid sparks generation.
Make sure that the load of the stored material (like grain, clinker, pellet, etc.) does not
‣
lead to deformation or stress to the probes or welds (if the probe is fixed on internals).
5.2 Mounting the assembly
The bundle rope thermometer is packed in a coil setup for a simple and compact
transportation. It is recommended to keep it in its coiled configuration until the
thermometer is close to the storage connection; a long and straight vertical rope would
result in a more difficult lifting and installing operations.
5.2.1 Junction box directly mounted on board
The following instructions must be followed for the proper installation of the device
(please note that this applies to “Open supporting frame”, “Supporting frame with covers”
and “Tube neck” version).
12Endress+Hauser
iTHERM MultiSens Bundle TMS31Mounting
1.
2.
3.
4.
Assembly sequence
1.
A0038308
Place the gasket between the flanged nozzle and the flange of the device (after
checking the cleanliness of gasket seats on the flanges).
2.Bring the device to the nozzle, inserting the thermoelements bundle rope through
the nozzle, avoiding interlacing and deformation of the thermoelement probes and
twisting of the rope system as well.
3.
Start the bolts insertion through the flanges ’ holes and tighten them with the nuts
by using a suitable wrench tool - but do not tighten them completely.
4.
Complete the bolts insertion through the flanges’ holes and tight them with the
crossed method by means of an appropriate equipment (i.e. controlled tensioning
according to the applicable standards).
A0038309
A0038310
Endress+Hauser13
MountingiTHERM MultiSens Bundle TMS31
+
+
-
-
+
+
-
-
+
+
-
-
1.2.
Wiring sequence (Connection customer side)
1.
A0038311
In case of direct wiring, introduce completely the extension or compensating cables
through the respective cable glands in the junction box.
2.Tighten the cable glands on the junction box.
3.After having opened the cover of the junction box, connect the compensating cables
to the terminals of the junction box following the wiring instruction provided,
ensuring the right matching between the cable tag numbers and the terminals tag
numbers.
4.Close the cover ensuring the right gasket position to avoid any impact on the IP
degree of protection.
5.In case of using the supporting frame with covers, check if all its components are still
coupled properly one each other.
5.2.2 Junction box remote connection
Junction box not supplied. Assembly sequence
Please refer to → 13 for a correct assemble process.
Conduit connection
Make sure the cable gland is properly tightened after the wiring connections.
14Endress+Hauser
A0038312
iTHERM MultiSens Bundle TMS31Mounting
Wiring sequence (connection customer side)
Please refer to → 14 for a correct wiring sequence.
Junction box supplied but not connected to the multipoint. Assembly sequence
Before any assemble and wiring operation, please make sure to fix the junction box to a
stable metallic support according to your needs and in a place easy to access.
Please refer to → 13 for a correct assemble process.
Conduit connection
Please refer to → 14 for a correct assemble process.
Wiring sequence (connection customer side)
Please refer to → 14 and → 19 for a correct wiring sequence.
Junction box supplied and connected to the multipoint.
Assembly sequence
Before any assemble and wiring operation, please make sure to fix the junction box to a
stable metallic support according to your needs and in a place easy to access.
Please refer to paragraph 5.2.1.1 for a correct assemble process.
Wiring sequence (connection customer side)
Please refer to paragraph 5.2.1.1 for a correct assemble process.
NOTICE
After the mounting, perform few simple checks on the thermometric system
installed.
Check the tightness of the threaded connections. If any part is loosened, tight it
‣
applying the proper torque.
Check the proper straight tensioning of the rope bundle to avoid any inconvenient
‣
bending which may result in an improper positioning of the thermoelements inside the
storage system.
Check the proper positioning of the weight at the rope.
‣
Check the proper connection of the swage eye to the selected anchor point inside the
‣
vessel (no weight version).
Check for correct wiring, test the electrical continuity of the sensors (by warming up
‣
the tip, when feasible) and then verify the absence of short circuits.
5.3 Post-mounting check
Before commissioning the measuring system make sure that all final checks have been
carried out:
Device conditions and specifications
Is the device undamaged (visual inspection)?
Do the ambient conditions match the device specification?
For example:
• Ambient temperature
• Proper conditions
Are the threaded components undeformed?
Are the gaskets not permanently deformed?
Installation
Endress+Hauser15
MountingiTHERM MultiSens Bundle TMS31
Is the equipment aligned with the nozzle axis?
Are the gasket seats of flanges clean?
Is the coupling between the flange and its counter flange reached?
Are the thermoelements not interlaced, undeformed or twisted?
Is the rope bundle in a proper straight tensioned configuration with no twisting or wrapping?
Is the toggle joint connected properly to the flange eyebolt?
Are the bolts completely inserted in the flange? Make sure the flange is completely attached to
the nozzle?
Are the cable glands tightened on the extension cables?
Are the extension cables connected to the junction box terminals?
Failure to observe this may result in destruction of parts of the electronics.
Switch off power supply before installing or connecting the device.
‣
When installing Ex-approved devices in a hazardous area please take special note of
‣
the instructions and connection schematics in the respective Ex documentation added
to these Operating Instructions. The local Endress+Hauser representative is available
for assistance if required.
When wiring to a transmitter also observe the wiring instructions in the enclosed
Brief Operating manuals of the relevant transmitter.
For wiring the device proceed as follows:
1.Open the housing cover on the junction box.
2.Open the cable glands on the sides of the junction box.
3.Feed the cables through the opening in the cable glands.
4.Connect the cables as shown on → 17
5.On completion of the wiring, screw the screw terminals tight. Tighten the cable
glands again. In doing so, also pay particular attention to → 21. Close the
housing cover again.
6.In order to avoid connection errors always take note of the hints given in the post
connection check! → 21
6.1 Quick wiring guide
Terminal assignment
NOTICE
Destruction or malfunction of parts of the electronics through ESD - electrostatic
discharge.
Take measures to protect the terminals from electrostatic discharge.
‣
To avoid incorrect measuring values an extension or compensation cable for direct
wiring of thermocouple and RTD sensors for the signal transmission has to be used.
The polarity indication on the respective terminal block and the wiring scheme has to
be observed.
The planning and the installation of the bus connection cables of the plant is not to be
concerned of the manufacturer of the device. Therefore the manufacturer cannot be
considered to be responsible for possible damages due to the choice of materials that
are not suitable for that application or to a faulty installation.
Power supply
head transmitter and
analog output 4 to 20 mA,
or bus connection
(red)(red)
(red)(red)
(white)(white)
(white)
+
+
-
-
+
+
-
-
+
+
-
-
+
+
-
-
1B
1B
1A
1A
A0016712-EN
5Wiring diagram of the single sensor input head transmitters (TMT18x)
Thermocouple cable colors
According to IEC 60584According to ASTM E230
• Type J: Black (+), white (-)
• Type K: Green (+), white (-)
• Type J: White (+), red (-)
• Type K: Yellow (+), red (-)
6.2 Connecting the sensor cables
Each sensor is marked with an individual TAG number. As default configuration, all
wires are always connected to the installed transmitters or terminals and generally
checked in house before final shipment. In case of remote junction box it might be
that the following steps have to be executed for the multipoint side aswell.
The wiring is done in consecutive order, which means that the input channel(s) of
transmitter no. 1 are connected to the insert wires starting from insert no. 1. Transmitter
no. 2 is not used until all channels of transmitter no. 1 are fully connected. The wires of
each insert are marked with consecutive numbers starting from 1. If double sensors are
used the internal marking has a suffix to distinguish the two sensors, e.g. 1A and 1B for
double sensors in the same insert or measuring point no. 1.
Endress+Hauser19
6Direct wiring on the mounted terminal block. Example for the internal sensor wires marking with 2 x TC
7Mounted and wired head transmitter. Example for the internal sensor wires marking with 2 x TC
Sensor typeTransmitter typeWiring rule
1 x RTD or TC• Single input (one channel)
• Double input (two channel)
• Multi-channel input (8 channel)
2 x RTD or TC• Single input (one channel)
• Double input (two channel)
• Multi-channel input (8 channel)
• 1 Head transmitter per insert
• 1 Head transmitter for 2 inserts
• 1 Multi-channel transmitter for 8 inserts
• Not available, wiring excluded
• 1 Head transmitter per insert
• 1 Multi-channel transmitter for 4 inserts
6.3 Connecting the power supply and signal cables
Cable specification
• A shielded cable is recommended for fieldbus communication. Take the plant grounding
concept into consideration.
• The terminals for connecting the signal cable (1+ and 2-) are protected against reverse
polarity.
• Conductor cross-section:
– Max 2.5 mm2 (14 AWG) for screw terminals
– Max 1.5 mm2 (16 AWG) for spring terminals
Always observe the general procedure on → 17.
A0033290
8Connecting the signal cable and power supply to the installed transmitter
1External ground terminal
2Terminals for signal cable and power supply
3Internal ground terminal
4Shielded signal cable, recommended for fieldbus connection
For any specific electrical shielding and grounding regarding the transmitter wiring
please refer to the appropriate operating manual of the installed transmitter.
Where applicable, national installation regulations and guidelines must be observed
during the installation! Where there are large differences in potential between the
individual grounding points, only one point of the shielding is connected directly with the
reference ground. In systems without potential equalization, therefore, cable shielding of
fieldbus systems should only be grounded on one side, for example at the supply unit or at
safety barriers.
NOTICE
If the shielding of the cable is grounded at more than one point in systems without
potential matching, power supply frequency equalizing currents can occur that
damage the signal cable or have a serious effect on signal transmission.
In such cases the shielding of the signal cable is to be grounded on only one side, i.e. it
‣
must not be connected to the ground terminal of the housing (terminal head, field
housing). The shield that is not connected should be insulated!
6.5 Degree of protection
The device components can be conform to the requirements up to IP 68 ingress protection.
In order to fulfil the degree of protection after installation or service, the following points
must be taken into consideration: → 9, 21
• The housing seals must be clean and undamaged before they are replaced in the sealing
rebate. If they are found to be too dry, they should be cleaned or even replaced.
• All housing screws and covers must be tightened.
• The cables used for connection must be of the correct specified outside diameter (e.g.
M20 x 1.5, cable diameter from 0.315 to 0.47 in; 8 to 12 mm).
• Tighten cable gland.
• Loop the cable or conduit before placing into the entry ("Water sack"). This means that
any moisture that may form cannot enter the gland. Install the device so that the cable
or conduit entries are not facing upwards.
• Entries not used are to be blanked off using the blanking plates provided.
• The protective grommet must not be removed from the NPT fitting.
A0011260
9Connection hints to maintain IP protection
6.6 Post-connection check
Is the device undamaged (internal equipment inspection)?
Electrical connection
Does the supply voltage match the specifications on the nameplate?
Are the power supply and signal cables correctly connected? → 17
Are all the screw terminals well tightened and have the connections of the spring terminals been
checked?
Are all the cable glands installed, tightened and sealed?
Are all the housing covers installed and tightened?
Does the marking of the terminals and cables match?
Is the electrical continuity of the thermocouple verified?
22Endress+Hauser
iTHERM MultiSens Bundle TMS31Commissioning
7 Commissioning
7.1 Preliminaries
Set-up guidelines of Standard, Extended and Advanced Commissioning for Endress
+Hauser instruments in order to guarantee the function of the instrument according to:
• Endress+Hauser operating manual
• Customer set up specification, and/or
• Application conditions, when applicable under process conditions
Both the operator and the person responsible for the process should be informed that a
commissioning job will be carried out, observing the following actions:
• If applicable, before disconnecting any sensor that is attached to the process, determine
what solid or fluid is being measured (observe safety data sheet).
• Be aware of the temperature conditions.
• Never open a process fitting or loosen flange bolts before you have confirmed it is safe to
do so.
• Be sure not to disturb the storage system when disconnecting inputs/outputs or when
simulating signals.
• Ensure our tools, equipment and the customer storage area are protected from cross
contamination. Consider and plan necessary cleaning steps.
• When commissioning requires chemicals (e.g. as reagents for standard operation or for
cleaning purposes), always follow and observe the safety regulations.
7.1.1 Reference documents
• Endress+Hauser Standard Operating Procedure for Health and Safety (see
documentation code: BP01039H)
• Operating Manual of relevant tools and equipment to perform the commissioning job.
• Relevant Endress+Hauser Service Documentation (operating manual, work instructions,
service info, service manual, etc.).
• Calibration certificates of the quality relevant equipment if available.
• If applicable, safety data sheet.
• Customer specific documents (safety instructions, setup points, etc.).
7.1.2 Tools and equipment
Multimeter and instrument related configuration tools as necessary from the above
mentioned action list.
7.2 Function check
Before commissioning the device make sure that all final checks have been carried out
• "Post-mounting check” checklist
• “Post-connection check” checklist
The commissioning should be performed according to our commissioning segmentation
(Standard, Extended and Advanced).
7.2.1 Standard commissioning
Visual inspection of device
1.Check the instrument(s) for damage which may have been caused during transport/
shipping or mounting/wiring
2.Check that the installation is done according to the operating manual
Endress+Hauser23
CommissioningiTHERM MultiSens Bundle TMS31
3.Check that the wiring is done according to the operating manual and the local
regulations (e.g. grounding)
4.Check the dust/water tightness of the instrument(s)
1.Check that the environmental conditions are appropriate for the instrument(s):
Ambient temperature, humidity (ingress protection IPxx), vibrations, hazardous
areas (Ex, Dust-Ex), RFI/EMC, sun protection, etc.
2.Check access to the instrument(s) for utilization and maintenance
Configuration parameters
Configure the instrument(s) according to the Operating Manual with the parameters
‣
specified by the customer or mentioned on the design specification
Output signal value check
Check and confirm that the local display and the output signals of the instrument(s)
‣
conform with the customer’s display
7.2.2 Extended commissioning
In addition to the steps of Standard Commissioning, the following should be additionally
completed:
Instrument Conformity
1.Check the received instrument(s) with the purchase order or design specification
including accessories, documentation and certificates
2.Check Software Version (e.g. application software such as “Batching”) when provided
3.Check that the documentation has the correct issue and version
Functional test
1.Test of the instrument outputs, including switching points, auxiliary inputs/outputs
with the internal or an external simulator (e.g. FieldCheck)
2.Compare the measuring data/results with a reference from the customer. (e.g.
laboratory result in case of an analyzer, weight scale in the case of a batching
application, etc.)
3.Adjust the instrument(s) if necessary and as described in the operating manual
7.2.3 Advanced commissioning
The Advanced Commissioning provides a loop test in addition to the steps covered in the
Standard and Extended Commissioning.
Loop test
1.Simulate a minimum of 3 output signals from the instrument(s) to the control room
2.Read out/note the simulated and indicated values and check for linearity
24Endress+Hauser
iTHERM MultiSens Bundle TMS31Diagnostics and troubleshooting
7.3 Switching on the device
Once the final checks have been successfully completed, it is time to switch on the supply
voltage. Afterwards the multipoint thermometer is operational. If there are Endress
+Hauser temperature transmitter in use, please refer to the enclosed Brief Operating
Instructions for commissioning.
8 Diagnostics and troubleshooting
8.1 General troubleshooting
Always start troubleshooting with the checklists below if faults occur after start up or
during operation. This takes you directly (via various queries) to the cause of the problem
and the appropriate remedial measures.
NOTICE
Repair of parts of the device
In the event of a serious fault, a measuring device might have to be replaced. In the
‣
case of replacement see section 'Return' → 26.
It is always important to check the connection between the cables and terminals, in
‣
order to guarantee the proper strain relief to the cables, and the tightening and the
sealing of the screw terminals.
Before commissioning the measuring system make sure that all final checks have been
carried out:
• Follow the checklist in section 'Post-mounting check' → 15
• Follow the checklist in section 'Post-connection check' → 21
If transmitters are used, please refer to the documentation of the transmitter installed for
diagnostic and troubleshooting procedures .
Endress+Hauser25
RepairiTHERM MultiSens Bundle TMS31
9 Repair
9.1 General notes
The accessibility around the device for maintenance has to be guaranteed. Each
component that is part of the device must be – in the case of replacement – replaced by an
original spare part of Endress+Hauser which guarantees the same characteristics and
performance. To ensure continued operational safety and reliability it is suggested to carry
out repairs on the device only if they are expressly permitted by Endress+Hauser,
observing federal/national regulations pertaining to repair of an electrical device.
9.2 Spare parts
When ordering spare parts, please specify the serial number of the unit!
Spare parts of the Multipoint thermometer assembly are:
• Cable glands
• Transmitters or electrical terminals
• Junction box and related accessories
• Ferrule sets of the compression fittings
9.3 Endress+Hauser services
ServiceDescription
CertificationsEndress+Hauser is able to fulfill requirements belonging to the design, product
manufacturing, tests and commissioning according to specific approvals by
handling or suppling individual certified components and by checking the
integration on the whole system.
MaintenanceAll Endress+Hauser systems are designed for an easy maintenance due to a
modular design, allowing the replacement of old or wear out parts. Standardized
parts ensure fast reaction for maintenance.
CalibrationEndress+Hauser’s range of calibration services covers on-site verification tests,
accredited laboratory calibrations, certificates and traceability to ensure
compliance.
InstallationEndress+Hauser helps you commission plants while minimizing costs. Fault free
installation is decisive for the quality and longevity of the measurement system
and plant running. We provide the right expertise at the right time to meet
project deliverables.
TestsIn order to ensure product quality and to guarantee efficiency during the entire
lifetime the following tests are available:
• Dye penetration test according to ASME V art. 6, UNI EN 571-1 and ASME VIII
Div. 1 App 8 standards
• PMI test accoding to ASTM E 572
• X-ray test according to ASME V art. 2, art. 22 and ISO 17363-1 (requirements
and methods) and ASME VIII div. 1 and to ISO 5817 (acceptance criteria).
Thickness up to 30 mm
9.4 Return
The requirements for safe device return can vary depending on the device type and
national legislation.
1.Refer to the website for more information:
http://www.endress.com/support/return-material
2.Return the device if repairs or a factory calibration are required, or if the wrong
device was ordered or delivered.
26Endress+Hauser
iTHERM MultiSens Bundle TMS31Repair
9.5 Disposal
9.5.1 Removing the measuring device
1.Switch off the device.
WARNING
L
Danger to persons from process conditions.
Beware of hazardous process conditions such as pressure in the measuring device, high
‣
temperatures or aggressive fluids.
2.Carry out the mounting and connection steps from the "Mounting the measuring
device" and "Connecting the measuring device" sections in reverse order. Observe the
safety instructions.
9.5.2 Disposing of the measuring device
WARNING
L
Danger to personnel and environment from fluids that are hazardous to health.
Ensure that the measuring device and all cavities are free of fluid residues that are
‣
hazardous to health or the environment, e.g. substances that have permeated into
crevices or diffused through plastic.
Observe the following notes during disposal:
Observe valid federal/national regulations.
‣
Ensure proper separation and reuse of the device components.
‣
Endress+Hauser27
AccessoriesiTHERM MultiSens Bundle TMS31
10 Accessories
Various accessories, which can be ordered with the device or subsequently from Endress
+Hauser, are available for the device. Detailed information on the order code in question is
available from your local Endress+Hauser sales center.
28Endress+Hauser
iTHERM MultiSens Bundle TMS31Accessories
10.1 Device-specific accessories
AccessoriesDescription
Anchor weight
The installation of the anchor weight ensures a straight
vertical position of the rope, please make sure to have
enough space for a correct weight positioning inside the
storage system. The dimensions will be established
during the order development according to the rope
multipoint dimension.
• Left side - Removable/Replaceable
• Right side - Fixed
Ogives
Toggle joint terminal
A0038304
Ogives are integrated in the multipoint rope, they provide
a correct positioning of the probe thermoelement along
the rope length and maintain them in position under
working condition.
A0038305
Toggle joint connection between rope and flange to allow
reciprocal rotation.
A0038306
Endress+Hauser29
AccessoriesiTHERM MultiSens Bundle TMS31
10.2 Communication-specific accessories
Configuration kit TXU10Configuration kit for PC-programmable transmitter with setup software and
interface cable for PC with USB port
Order code: TXU10-xx
Commubox FXA195
HART
Commubox FXA291Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
HART Loop Converter
HMX50
Wireless HART adapter
SWA70
Fieldgate FXA320Gateway for the remote monitoring of connected 4-20 mA measuring devices via a
Fieldgate FXA520Gateway for the remote diagnostics and remote configuration of connected HART
For intrinsically safe HART communication with FieldCare via the USB interface.
For details, see "Technical Information" TI00404F
Common Data Interface) and the USB port of a computer or laptop.
For details, see "Technical Information" TI00405C
Is used to evaluate and convert dynamic HART process variables to analog current
signals or limit values.
For details, see "Technical Information" TI00429F and Operating Instructions
BA00371F
Is used for the wireless connection of field devices.
The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
For details, see Operating Instructions BA061S
Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00053S
measuring devices via a Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00051S
Field Xpert SFX100Compact, flexible and robust industry handheld terminal for remote configuration
and for obtaining measured values via the HART current output (4-20 mA).
For details, see Operating Instructions BA00060S
10.3 Service-specific accessories
AccessoriesDescription
ApplicatorSoftware for selecting and sizing Endress+Hauser measuring devices:
• Calculation of all the necessary data for identifying the optimum measuring
device: e.g. pressure loss, accuracy or process connections.
• Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters over the entire life cycle of a project.
Applicator is available:
• Via the Internet: https://wapps.endress.com/applicator
• On CD-ROM for local PC installation.
30Endress+Hauser
iTHERM MultiSens Bundle TMS31Accessories
W@MLife cycle management for your plant
W@M supports you with a wide range of software applications over the entire
process: from planning and procurement, to the installation, commissioning and
operation of the measuring devices. All the relevant device information, such as
the device status, spare parts and device-specific documentation, is available for
every device over the entire life cycle.
The application already contains the data of your Endress+Hauser device. Endress
+Hauser also takes care of maintaining and updating the data records.
W@M is available:
• Via the Internet: www.endress.com/lifecyclemanagement
• On CD-ROM for local PC installation.
FieldCareFDT-based plant asset management tool from Endress+Hauser.
It can configure all smart field units in your system and helps you manage them. By
using the status information, it is also a simple but effective way of checking their
status and condition.
For details, see Operating Instructions BA00027S and BA00059S
Endress+Hauser31
Technical dataiTHERM MultiSens Bundle TMS31
11 Technical data
11.1 Input
11.1.1 Measured variable
Temperature (temperature linear transmission behavior)
11.1.2 Measuring range
RTD:
InputDesignationMeasuring range limits
RTD as per IEC 60751Pt100–200 to +600 °C (–328 to +1112 °F)
Thermocouple:
InputDesignationMeasuring range limits
Thermocouples (TC) as per IEC
60584, part 1 - using an
Endress+Hauser - iTEMP
temperature head transmitter
–40 to +520 °C (–40 to +968 °F)
–40 to +800 °C (–40 to +1472 °F)
Thermocouples (TC) - flying
leads - as per IEC 60584 and
ASTM E230
1)Limited by jacket material of insert
Type J (Fe-CuNi)
Type K (NiCr-Ni)
–210 to +520 °C (–346 to +968 °F), typical sensitivity
above 0 °C ≈ 55 μV/K
–270 to +800 °C (–454 to +1472 °F)
sensitivity above 0 °C ≈ 40 μV/K
1)
, typical
11.2 Output
11.2.1 Output signal
Generally, the measured value can be transmitted in one of two ways:
• Directly-wired sensors - sensor measured values forwarded without a transmitter.
• Via all common protocols by selecting an appropriate Endress+Hauser iTEMP
temperature transmitter. All the transmitters listed below are mounted directly in the
junction box and wired with the sensory mechanism.
11.2.2 Family of temperature transmitters
Thermometers fitted with iTEMP transmitters are an installation-ready complete solution
to improve temperature measurement by significantly increasing accuracy and reliability,
when compared to direct wired sensors, as well as reducing both wiring and maintenance
costs.
PC programmable head transmitters
They offer a high degree of flexibility, thereby supporting universal application with low
inventory storage. The iTEMP transmitters can be configured quickly and easily at a PC.
Endress+Hauser offers free configuration software which can be downloaded from the
Endress+Hauser Website. More information can be found in the Technical Information.
32Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
HART® programmable head transmitters
The transmitter is a 2-wire device with one or two measuring inputs and one analog
output. The device not only transfers converted signals from resistance thermometers and
thermocouples, it also transfers resistance and voltage signals using HART
®
communication. It can be installed as an intrinsically safe apparatus in Zone 1 hazardous
areas and is used for instrumentation in the terminal head (flat face) as per DIN EN
50446. Swift and easy operation, visualization and maintenance by PC using operating
software, Simatic PDM or AMS. For more information, see the Technical Information.
PROFIBUS® PA head transmitters
Universally programmable head transmitter with PROFIBUS® PA communication.
Conversion of various input signals into digital output signals. High accuracy over the
complete ambient temperature range. Swift and easy operation, visualization and
maintenance using a PC directly from the control panel, e. g. using operating software,
Simatic PDM or AMS. For more information, see the Technical Information.
FOUNDATION Fieldbus™ head transmitters
Universally programmable head transmitter with FOUNDATION Fieldbus™
communication. Conversion of various input signals into digital output signals. High
accuracy over the complete ambient temperature range. Swift and easy operation,
visualization and maintenance using a PC directly from the control panel, e.g. using
operating software such as ControlCare from Endress+Hauser or NI Configurator from
National Instruments. For more information, see the Technical Information.
Advantages of the iTEMP transmitters:
• Dual or single sensor input (optionally for certain transmitters)
• Unsurpassed reliability, accuracy and long-term stability in critical processes
• Mathematical functions
• Monitoring of the thermometer drift, sensor backup functionality, sensor diagnostic
functions
• Sensor-transmitter matching for dual sensor input transmitter, based on Callendar/Van
Dusen coefficients
Endress+Hauser33
Technical dataiTHERM MultiSens Bundle TMS31
A
AA
-200 -100 0 100 200 300 400 500 600°C
0.5
1.0
1.5
2.0
B
2.5
3.0
- 0.5
- 1.0
- 1.5
- 2.0
- 2.5
- 3.0
B
A
AA
Max. deviation (°C)
Max. deviation (°C)
11.3 Performance characteristics
11.3.1 Accuracy
RTD resistance thermometer as per IEC 60751
ClassMax. tolerances (°C)Characteristics
Cl. AA, former 1/3
± (0.1 + 0.0017 · |t|
Cl. B
Cl. A± (0.15 + 0.002 · |t|
Cl. B± (0.3 + 0.005 · |t|
Temperature ranges for compliance with the
tolerance classes
Wire wound
sensor (WW):
Thin-film version
Cl. ACl. AA
–100 to +450 °C –50 to +250 °C
Cl. ACl. AA
(TF):
Standard–30 to +300 °C0 to +150 °C
1)
)
1)
)
1)
)
1)|t| = absolute value °C
In order to obtain the maximum tolerances in °F, the results in °C must be multiplied
by a factor of 1.8.
Permissible deviation limits of thermoelectric voltages from the standard characteristic for
thermocouples as per IEC 60584 or ASTM E230/ANSI MC96.1:
StandardTypeStandard toleranceSpecial tolerance
IEC 60584Class DeviationClass Deviation
J (Fe-CuNi)2±2.5 °C (–40 to 333 °C)
±0.0075 |t|
K (NiCr-NiAl) 2±2.5 °C (–40 to 333 °C)
1)|t| = absolute value °C
±0.0075 |t|
1)
(333 to 750 °C)
1)
(333 to 1200 °C)
1±1.5 °C (–40 to 375 °C)
±0.004 |t|
1±1.5 °C (–40 to 375 °C)
±0.004 |t|
1)
(375 to 750 °C)
1)
(375 to 1000 °C)
A0008588-EN
34Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
StandardTypeStandard toleranceSpecial tolerance
ASTM E230/ANSI
MC96.1
1)|t| = absolute value °C
J (Fe-CuNi) ±2.2 K or ±0.0075 |t|
K (NiCrNiAl)
Deviation, the larger respective value applies
1)
(0 to 760 °C)±1.1 K or ±0.004 |t|
(0 to 760 °C)
±2.2 K or ±0.02 |t|
±2.2 K or ±0.0075 |t|
1)
(–200 to 0 °C)
1)
(0 to 1260 °C)
±1.1 K or ±0.004 |t|
(0 to 1260 °C)
1)
1)
11.3.2 Influence of ambient temperature
Depends on the head transmitter used. For details, see the Technical Information.
11.3.3 Response time
Response time for the sensor assembly without transmitter. It refers to inserts in
direct contact with process.
RTD
Calculated at an ambient temperature of approx. 23 °C by immersing the insert in running
water (0.4 m/s flow rate, 10 K excess temperature):
Insert diameterResponse time
Mineral-insulated cable, 3 mm (0.12 in)t
RTD insert StrongSens, 6 mm (¹⁄₄ in)t
50
t
90
50
t
90
2 s
5 s
< 3.5 s
< 10 s
Thermocouple (TC)
Calculated at an ambient temperature of approx. 23 °C by immersing the insert in running
water (0.4 m/s flow rate, 10 K excess temperature):
Insert diameterResponse time
Grounded thermocouple:
3 mm (0.12 in), 2 mm (0.08 in)
Ungrounded thermocouple:
3 mm (0.12 in), 2 mm (0.08 in)
t
50
t
90
t
50
t
90
0.8 s
2 s
1 s
2.5 s
11.3.4 Shock and vibration resistance
• RTD: 3G / 10 to 500 Hz according to IEC 60751
• RTD iTHERM StrongSens Pt100 (TF, vibration resistant): Up to 60G
• TC: 4G / 2 to 150 Hz according to IEC 60068-2-6
Endress+Hauser35
Technical dataiTHERM MultiSens Bundle TMS31
11.3.5 Calibration
Calibration is a service that can be performed on each individual insert, either in order
phase, or after multipoint installation.
When calibration shall be performed once the multipoint is installed, please contact
the Endress+Hauser service to get full support. Together with the Endress +Hauser
service any further activity can be organised to achieve the calibration of the target
sensor. In any case it is forbidden to unscrew any threaded component on the process
connection under operating conditions = running process.
Calibration involves comparing the measured values of the sensing elements of the
multipoint inserts (DUT device under test) with those of a more precise calibration
standard using a defined and reproducible measurement method. The aim is to determine
the deviation of the DUT measured values from the true value of the measured variable.
Two different methods are used for the inserts:
• Calibration at fixed-point temperatures, e.g. at the freezing point of water at 0 °C (32 °F).
• Calibration compared against a precise reference thermometer.
Evaluation of inserts
If a calibration with an acceptable uncertainty of measurement and transferable
measurement results is not possible, Endress+Hauser offers an insert evaluation
measurement service, if technically feasible.
11.4 Environment
11.4.1 Ambient temperature range
Junction boxNon-hazardous areaHazardous area
Without mounted transmitter–50 to +85 °C (–58 to +185 °F) –50 to +60 °C (–58 to +140 °F)
With mounted head
transmitter
With mounted multi-channel
transmitter
11.4.2 Storage temperature
Junction box
With head transmitter–50 to +95 °C (–58 to +203 °F)
With multi-channel transmitter–40 to +80 °C (–40 to +176 °F)
With DIN rail transmitter–40 to +95 °C (–40 to +203 °F)
11.4.3 Humidity
–40 to +85 °C (–40 to +185 °F) Depends on the respective hazardous area
approval. Details see Ex documentation.
–40 to +85 °C (–40 to +185 °F) –40 to +70 °C (–40 to +158 °F)
Condensation according to IEC 60068-2-33:
• Head transmitter: Permitted
• DIN rail transmitter: Not permitted
Maximum relative humidity: 95% according to IEC 60068-2-30
36Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
11.4.4 Climate class
Determined when the following components are installed into the junction box:
• Head transmitter: Class C1 according to EN 60654-1
• Multi-channel transmitter: Tested as per IEC 60068-2-30, meets the requirements
regarding class C1-C3 in accordance with IEC 60721-4-3
• Terminal blocks: Class B2 according to EN 60654-1
11.4.5 Degree of protection
• Specification for conduit: IP68
• Specification for the junction box: IP66/67
11.4.6 Electromagnetic compatibility (EMC)
Depending on the head transmitter used. For detailed information see the related
Technical Information, listed at the end of this document.
11.5 Mechanical construction
11.5.1 Design, dimensions
The overall rope assembly is made of different parts. The rope joint ensures sufficient
degree of freedom to the rope system allowing movements during filling and emptying
operations. This guarantees low stresses (no extra tensioning) on the rope due to possible
lateral force acting on it, therefore a lateral sag of 30cm per 10m rope length is
recommended. The transition between the inserts and the extension cable is obtained by
the usage of compression fittings, ensuring the declared IP degree protection.
Endress+Hauser37
Technical dataiTHERM MultiSens Bundle TMS31
I
H
I
C
H
A
B
F
L
MPx
A
A
B
L
MPx
F
MPx
L
B
L
MPx
F
MPx
L
U
U
U
L
MPx
A0038299
10Design of the modular multipoint thermometer, with tube neck on the left side, frame neck on the middle
or with tube neck design as option on the right side. All dimensions in mm (in)
A, B,CDimensions of the junction box, see following figure
MPx Numbers and distribution of measuring points: MP1, MP2, MP3 etc.
L
Immersion length of sensing elements or thermowells
MPx
I, H Encumbrance of the junction box and support system
FExtension neck length
LDevice length
UImmersion length
Extension neck F in mm (in)
Standard 250 (9.84)
Specifically customized extension necks are available on request.
Immersion lengths MPx of sensing elements/thermowells:
Based on customer requirements
38Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
A
B
C
Rope maximum load:
Rope
Ø mm
ConstructionWeight
kg/m
MBL
kNkg
61x190,178629,53000
81x190,322535400
101x190,502848500
A0038300
• Stainless steel AISI 316
• Rope according to EN
10264-4
• Rope grade
1.570 N/mm2
Junction box (directly mounted)
A0028118
1Cable glands
2Junction box
3Frame
A0038301
11Open design on the left side, with cover design in the middle and tube neck design on the right side
Endress+Hauser39
Technical dataiTHERM MultiSens Bundle TMS31
A
A0038302
12Remote junction box design
The junction box is suited for chemical agents environments. Sea water corrosion
resistance and extreme temperature variation stability is guaranteed. Ex-e Ex-i terminals
can be installed.
Possible junction box dimensions (A x B x C) in mm (in):
ABC
Stainless SteelMin.170 (6.7)170 (6.7)130 (5.1)
Max.500 (19.7)500 (19.7)240 (9.5)
AluminiumMin.100 (3.9)150 (5.9)80 (3.2)
Max.330 (13)500 (19.7)180 (7.1)
Type of specificationJunction boxCable glands
MaterialAISI 316/AluminiumNiCr Plated brass
AISI 316 / 316L
Ingress protection (IP)IP66/67IP66
Ambient temperature range–50 to +60 °C
(–58 to +140 °F)
ApprovalsATEX, FM, UL, CSA approval
for use in hazardous area
IEC
Marking• ATEX II 2 GD Ex e IIC /Ex ia
Ga IIC Ex tb IIIC Db
T6/T5/T4
• UL913 Class I, Division 1
Groups B, C, D T6/T5/T4
• FM3610 Class I, Division 1
Groups B, C, D T6/T5/T4
CoverHinged-
Maximum sealing diameter-6 to 12 mm (0.24 to 0.47 in)
• CSA C22.2 No. 157 Class 1,
Division 1 Groups B, C, D
T6/T5/T4
The neck extension ensures the connection between the flange and the junction box. The
design has been developed to ensure several mounting layouts to deal with possible
obstacles and constraints that can be met in any plant such as the storage tank
infrastructure (step ways, loading structures, stairs, etc.) and an eventual thermal
insulation. The neck extension design allows easy access for monitoring extension cables.
It guarantees a high stiffness connection for the junction box and vibration loads. No
closed volumes are present in the neck extension (not for tube neck design). This avoids
the accumulation of waste and potentially dangerous fluids coming from the environment
that can damage the instrumentation allowing continuous ventilation.
Endress+Hauser41
Technical dataiTHERM MultiSens Bundle TMS31
Inserts
Different insert types are available. For any different requirement that is not
described here, please contact the Endress+Hauser sales department.
Thermocouple
Diameter in mm (in)TypeStandardHot junction typeSheath material
1x type K
3 (0.12)
2x type K
1x type J
2x type J
IEC 60584 /ASTM E230Grounded/UngroundedAISI 316L
RTD
Diameter in mm (in)TypeStandardSheath material
1x Pt100 WW
3 (0.12)
6 (¹⁄₄)
2x Pt100 WW
1x Pt100 TF
2x Pt100 TF
IEC 60751AISI 316L
11.5.2 Weight
The weight can vary depending on the configuration: Dimension and content of the
junction box, neck length, dimensions of process connection, the number of inserts and the
weight of the rope end. The approximate weight of a typically configured multipoint rope
(number of inserts = 12, flange size = 3", medium size junction box) = 55 kg (121 lb)
11.5.3 Materials
It refers to insert sheath, neck extension, junction box and all wetted parts.
The temperatures for continuous operation specified in the following table are only
intended as reference values for use of the various materials in air and without any
significant compressive load. The maximum operation temperatures are reduced
considerably in some cases where abnormal conditions such as high mechanical load occur
or in aggressive media.
Material
name
AISI
316/1.4401
AISI 316L/
1.4404
1.4435
Short formRecommended
max.
temperature for
continuous use
in air
X5CrNiMo 17-12-2650 °C (1202 °F)• Austenitic, stainless steel
X2CrNiMo17-12-2
X2CrNiMo18-14-3
650 °C (1202 °F)• Austenitic, stainless steel
Properties
• High corrosion resistance in general
• Particularly high corrosion resistance in chlorinebased and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)
• High corrosion resistance in general
• Particularly high corrosion resistance in chlorinebased and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)
• Increased resistance to intergranular corrosion
and pitting
• Compared to 1.4404, 1.4435 has even higher
corrosion resistance and a lower delta ferrite
content
42Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
Material
name
Alloy600/
2.4816
AISI
304/1.4301
AISI 304L/
1.4307
AISI 316Ti/
1.4571
AISI
321/1.4541
AISI
347/1.4550
Short formRecommended
max.
temperature for
continuous use
in air
NiCr15Fe1100 °C
(2012 °F)
X5CrNi18-10850 °C (1562 °F)• Austenitic, stainless steel
X2CrNi18-9850 °C (1562 °F)• Good welding properties
X6CrNiMoTi17-12-2 700 °C (1292 °F)• Addition of titanium means increased resistance
X6CrNiTi18-10815 °C (1499 °F)• Austenitic stainless steel
X6CrNiNb10-10800 °C (1472 °F)• Austenitic stainless steel
Properties
• A nickel/chromium alloy with very good
resistance to aggressive, oxidizing and reducing
atmospheres, even at high temperatures
• Resistance to corrosion caused by chlorine gases
and chlorinated media as well as many oxidizing
mineral and organic acids, sea water etc.
• Corrosion from ultrapure water
• Not to be used in sulfur-containing atmospheres
• Well usable in water and lowly pollute waste
water
• Only at relatively low temperatures resistant to
organic acids, saline solutions, sulphates,
alkaline solutions, etc.
• Impervious to intergranular corrosion
• High ductility, excellent drawing, forming, and
spinning properties
to intergranular corrosion even after welding
• Broad range of uses in the chemical,
petrochemical and oil industries as well as in coal
chemistry
• Can only be polished to a limited extent, titanium
streaks can form
• High resistance to intergranular corrosion even
after welding
• Good welding characteristics, suitable to all
standard welding methods
• It is used in many sectors of the chemical
industry, petrochemical, and pressurized vessels
• Good resistance to a wide variety of
environments in the chemical, textile, oilrefining, dairy and food industries
• Added niobium makes this steel impervious to
intergranular corrosion
• Good weldability
• Main applications are furnace fire walls, pressure
vessels, welded structures, turbine blades
11.5.4 Process connection
Standard process connection flanges are designed according to the following standards:
1)Flanges according to GOST standard are available on request.
Endress+Hauser43
1)
SizeRatingMaterial
Technical dataiTHERM MultiSens Bundle TMS31
11.6 Certificates and approvals
11.6.1 CE Mark
The complete assembly is provided with individual components CE marked, to ensure safe
use in hazardous areas and pressurized environments.
11.6.2 Hazardous area approvals
The Ex approval applies to individual components like junction box, cable glands,
terminals. For further details on the available Ex versions (ATEX, CSA, FM, IEC-EX, UL,
NEPSI, EAC-EX), please contact your nearest Endress+Hauser sales organization. All
relevant data for hazardous areas can be found in separate Ex documentation.
11.6.3 Certification HART
The HART® temperature transmitter is registered by the FieldComm Group. The device
meets the requirements of the HART® Communication Protocol Specifications.
11.6.4 Certification FOUNDATION Fieldbus
The FOUNDATION Fieldbus™ temperature transmitter has successfully passed all test
procedures and is certified and registered by the Fieldbus Foundation. The device thus
meets all the requirements of the following specification:
• Certified according to FOUNDATION Fieldbus™ specification
• FOUNDATION Fieldbus™ H1
• Interoperability Test Kit (ITK), up to date revision status (device certification no.
available on request): the device can also be operated with certified devices of other
manufacturers
• Physical layer conformance test of the FOUNDATION Fieldbus™
11.6.5 Certification PROFIBUS® PA
The PROFIBUS® PA temperature transmitter is certified and registered by the PNO
(PROFIBUS® Nutzerorganisation e. V.), PROFIBUS user organization. The device meets all
the requirements of the following specifications:
• Certified according to FOUNDATION Fieldbus™ specification
• Certified in accordance with PROFIBUS® PA Profile (the up to date profile version is
available on request)
• The device can also be operated with certified devices of other manufacturers
(interoperability)
11.6.6 Other standards and guidelines
• EN 60079: ATEX certification for hazardous areas
• IEC 60529: Degree of protection of housing (IP code)
• IEC 60584 and ASTM E230/ANSI MC96.1: Thermocouples
11.6.7 Material certification
The material certificate 3.1 (according to EN 10204) can be requested separately. The
certificate includes a declaration related to the materials used to produce the
thermometer. It guarantees the traceability of the materials through the identification
number of the rope multipoint thermometer.
44Endress+Hauser
iTHERM MultiSens Bundle TMS31Technical data
11.6.8 Test report and calibration
The "Factory calibration" is carried out according to an internal procedure in a laboratory of
Endress+Hauser accredited by the European Accreditation Organization (EA) to ISO/IEC
17025. A calibration which is performed according to EA guidelines (SIT/Accredia) or
(DKD/DAkkS) may be requested separately. The calibration is performed on the inserts of
the multipoint.
11.7 Documentation
• Operating manuals iTEMP temperature transmitters:
– TMT180, PC-programmable, single-channel, Pt100 (KA00118R/09/a3)
– TMT181, PC programmable, single-channel, RTD, TC, Ω, mV (KA141R/09/a3)
– HART® TMT182, single-channel, RTD, TC, Ω, mV (KA142R/09/c4)
– HART® TMT82, two-channel, RTD, TC, Ω, mV (BA01028T/09/en)
– PROFIBUS® PA TMT84, two-channel, RTD, TC, Ω, mV (BA00257R/09/en)
– FOUNDATION FieldbusTM TMT85, two-channel, RTD, TC, Ω, mV (BA00251R/09/en)
– FOUNDATION FieldbusTM TMT125, 8-channel, RTD, TC, Ω, mV (BA00240R/09/en)
– Safety requirements: DIN EN 61010-1:2011-07
– EMC requirements : DIN EN 61326-1:2013-07
– RSG45 DIN RAIL
– TMT162
– TMT142
– Field Display (FD188)
• Technical Information of inserts:
– Resistance thermometer insert Omnigrad T TST310 (TI00085T/09/en)
– Thermocouple insert Omnigrad T TSC310 (TI00255t/09/en)
• Technical Information application example:
– RN221N active barrier, for supplying loop-powered 2-wire transmitters
(TI073R/09/en)
– HAW562 surge arresters, (TI01012K/09/en)
Endress+Hauser45
www.addresses.endress.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.