Direct current and alternating currentGround connection
Protective ground connection
A terminal which must be connected
to ground prior to establishing any
other connections.
Symbols for certain types of information
Proline Promass P 100
A grounded terminal which, as far as
the operator is concerned, is
grounded via a grounding system.
Equipotential connection
A connection that has to be connected
to the plant grounding system: This
may be a potential equalization line
or a star grounding system depending
on national or company codes of
practice.
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
Visual inspection
Symbols in graphics
SymbolMeaningSymbolMeaning
1, 2, 3,...Item numbers
A, B, C, ...ViewsA-A, B-B, C-C, ... Sections
Hazardous area
Series of steps
Safe area (non-hazardous area)
Flow direction
4Endress+Hauser
Page 5
Proline Promass P 100
12
3
Function and system design
Measuring principle
The measuring principle is based on the controlled generation of Coriolis forces. These forces are
always present in a system when both translational and rotational movements are superimposed.
Fc = 2 · ∆m (ν · ω)
Fc = Coriolis force
∆m = moving mass
ω = rotational velocity
ν = radial velocity in rotating or oscillating system
The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity ν in the system and
thus on the mass flow. Instead of a constant rotational velocity ω, the sensor uses oscillation.
In the sensor, an oscillation is produced in the measuring tube. The Coriolis forces produced at the
measuring tube cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow (i.e. when the fluid stands still), the oscillation measured at points A and B has
the same phase (no phase difference).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the
outlet (3).
Measuring system
A0016772
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the
tube oscillations at the inlet and outlet. System balance is created by exciting an eccentrically
arranged swinging mass to antiphase oscillation. The measuring principle operates independently of
temperature, pressure, viscosity, conductivity and flow profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and
thus the density of the oscillating system (comprising measuring tube and fluid) results in a
corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a
function of medium density. The microprocessor utilizes this relationship to obtain a density signal.
Volume measurement
Together with the measured mass flow, this is used to calculate the volume flow.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor
due to temperature effects. This signal corresponds to the process temperature and is also available
as an output signal.
The device consists of a transmitter and a sensor. If a device with Modbus RS485 intrinsically safe is
ordered, the Safety Barrier Promass 100 is part of the scope of supply and must be implemented to
operate the device.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
• Additionally for device version with local display:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with 4-20 mA HART, pulse/frequency/switch
output:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with EtherNet/IP output:
– Via Web browser (e.g. Microsoft Internet Explorer)
– Via Add-on Profile Level 3 for automation system from Rockwell
Automation
– Via Electronic Data Sheet (EDS)
• Also for device version with PROFINET output:
– Via Web browser (e.g. Microsoft Internet Explorer)
– Via device master file (GSD)
Proline Promass P 100
Sensor
Promass P• For use in sterile applications in heavily regulated industries
– Process connections: stainless steel, 1.4435 (316L); 1.4404
(316/316L)
– Surface quality: Ra
Ra
0.38 µm (15 µin) (electropolished)
max
0.76 µm (30 µin)
max
– Delta ferrite <1%
Safety Barrier Promass 100
• Dual-channel safety barrier for installation in non-hazardous locations
or zone 2/div. 2:
– Channel 1: DC 24 V power supply
– Channel 2: Modbus RS485
• In addition to current, voltage and power limitation, it offers galvanic
isolation of circuits for explosion protection.
• Easy top-hat rail mounting (DIN 35 mm) for installation in control
A0016763
cabinets
6Endress+Hauser
Page 7
Proline Promass P 100
23
7
8
6
9
10
11
4
1
5
Equipment architecture
A0016779
1Possibilities for integrating measuring devices into a system
1Automation system (e.g. PLC)
2EtherNet/IP
3PROFIBUS DP
4Modbus RS485
54-20 mA HART, pulse/frequency/switch output
6Safety Barrier Promass 100
7Modbus RS485 intrinsically safe
8Non-hazardous area
9Non-hazardous area and Zone 2/Div. 2
10 Intrinsically safe area and Zone 1/Div. 1
SafetyIT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.
Endress+Hauser7
Page 8
Input
Measured variableDirect measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring rangeMeasuring ranges for liquids
Proline Promass P 100
DNMeasuring range full scale values
[mm][in][kg/h][lb/min]
8³⁄₈0 to 20000 to 73.50
15½0 to 65000 to 238.9
2510 to 180000 to 661.5
401½0 to 450000 to 1654
5020 to 700000 to 2573
Recommended measuring range
"Flow limit" section → 55
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value are not overridden by the electronics unit, with the result
that the totalizer values are registered correctly.
Input signalExternal measured values
To increase the accuracy of certain measured variables or to calculate the corrected volume flow for
gases, the automation system can continuously write different measured values to the measuring
device:
• Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure
measuring device for absolute pressure, e.g. Cerabar M or Cerabar S)
• Medium temperature to increase accuracy (e.g. iTEMP)
• Reference density for calculating the corrected volume flow for gases
Various pressure transmitters and temperature measuring devices can be ordered from Endress
+Hauser: see "Accessories" section → 99
It is recommended to read in external measured values to calculate the following measured variables:
• Mass flow
• Corrected volume flow
min(F)
to
max(F)
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Digital communication
The measured values can be written from the automation system to the measuring via:
• PROFIBUS DP
• Modbus RS485
• EtherNet/IP
• PROFINET
8Endress+Hauser
Page 9
Proline Promass P 100
Output
Output signalCurrent output
Current output4-20 mA HART (active)
Maximum output values• DC 24 V (no flow)
Load0 to 700 Ω
Resolution0.38 µA
DampingAdjustable: 0.07 to 999 s
Assignable measured
variables
• 22.5 mA
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
FunctionCan be set to pulse, frequency or switch output
VersionPassive, open collector
Maximum input values• DC 30 V
• 25 mA
Voltage dropFor 25 mA: ≤ DC 2 V
Pulse output
Pulse widthAdjustable: 0.05 to 2000 ms
Maximum pulse rate10000 Impulse/s
Pulse valueAdjustable
Assignable measured
variables
Frequency output
Output frequencyAdjustable: 0 to 10000 Hz
DampingAdjustable: 0 to 999 s
Pulse/pause ratio1:1
Assignable measured
variables
Switch output
Switching behaviorBinary, conductive or non-conductive
Switching delayAdjustable: 0 to 100 s
• Mass flow
• Volume flow
• Corrected volume flow
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Endress+Hauser9
Page 10
Proline Promass P 100
Number of switching
cycles
Assignable functions• Off
Unlimited
• On
• Diagnostic behavior
• Limit value
– Mass flow
– Volume flow
– Corrected volume flow
– Density
– Reference density
– Temperature
– Totalizer 1-3
• Flow direction monitoring
• Status
– Partially filled pipe detection
– Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
PROFIBUS DP
Signal encodingNRZ code
Data transfer9.6 kBaud…12 MBaud
Signal on alarm
Modbus RS485
Physical interfaceIn accordance with EIA/TIA-485-A standard
Terminating resistor• For device version used in non-hazardous areas or Zone 2/Div. 2: integrated
and can be activated via DIP switches on the transmitter electronics module
• For device version used in intrinsically safe areas: integrated and can be
activated via DIP switches on the Safety Barrier Promass 100
EtherNet/IP
StandardsIn accordance with IEEE 802.3
PROFINET
StandardsIn accordance with IEEE 802.3
Depending on the interface, failure information is displayed as follows:
Current output
4-20 mA
Failure modeChoose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
HART
Device diagnosticsDevice condition can be read out via HART Command 48
10Endress+Hauser
Page 11
Proline Promass P 100
Pulse/frequency/switch output
Pulse output
Failure modeChoose from:
• Actual value
• No pulses
Frequency output
Failure modeChoose from:
• Actual value
• 0 Hz
• Defined value: 0 to 12500 Hz
Switch output
Failure modeChoose from:
• Current status
• Open
• Closed
PROFIBUS DP
Status and alarm
messages
Diagnostics in accordance with PROFIBUS PA Profile 3.02
Modbus RS485
Failure modeChoose from:
• NaN value instead of current value
• Last valid value
EtherNet/IP
Device diagnosticsDevice condition can be read out in Input Assembly
PROFINET
Device diagnosticsIn accordance with "Application Layer protocol for decentral device periphery and
distributed automation", version 2.3
Local display
Plain text displayWith information on cause and remedial measures
BacklightRed backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Operating tool
• Via digital communication:
– HART protocol
– PROFIBUS DP
– Modbus RS485
– EtherNet/IP
– PROFINET
• Via service interface
• Via Web server
Endress+Hauser11
Page 12
Plain text displayWith information on cause and remedial measures
Additional information on remote operation → 89
Web browser
Plain text displayWith information on cause and remedial measures
Light emitting diodes (LED)
Status informationStatus indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
• PROFINET network available
• PROFINET connection established
• PROFINET blinking feature
Proline Promass P 100
Ex connection data
These values only apply for the following device version:
Order code for "Output", option M "Modbus RS485", for use in intrinsically safe areas
Safety Barrier Promass 100
Safety-related values
Terminal numbers
Supply voltageSignal transmission
2 (L-)1 (L+)26 (A)27 (B)
U
nom
U
= AC 260 V
max
= DC 24 V
U
nom
U
= AC 260 V
max
= DC 5 V
Intrinsically safe values
Terminal numbers
Supply voltageSignal transmission
20 (L-)10 (L+)62 (A)72 (B)
Uo = 16.24 V
Io = 623 mA
Po = 2.45 W
With IIB
1)
: Lo = 92.8 µH, Co = 0.433 μF, Lo/Ro = 14.6 μH/Ω
1)
: Lo= 372 µH, Co = 2.57 μF, Lo/Ro = 58.3 μH/Ω
With IIC
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
1)The gas group depends on the sensor and nominal diameter.
12Endress+Hauser
Page 13
Proline Promass P 100
Transmitter
Intrinsically safe values
• Option BM: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• Option 85: ATEX II2G + IECEx Z1 Ex ia + CSA C/US
IS Cl. I, II, III Div. 1
Low flow cut off
Galvanic isolation
Protocol-specific dataHART
The switch points for low flow cut off are user-selectable.
The following connections are galvanically isolated from each other:
• Outputs
• Power supply
Manufacturer ID0x11
Device type ID0x4A
HART protocol revision7
Device description files
(DTM, DD)
HART loadMin. 250 Ω
Order code for
"Approval"
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Information and files under:
www.endress.com
Supply voltageSignal transmission
20 (L-)10 (L+)62 (A)72 (B)
Terminal numbers
Ui = 16.24 V
Ii = 623 mA
Pi = 2.45 W
Li = 0 µH
Ci = 6 nF
Endress+Hauser13
Page 14
Dynamic variablesRead out the dynamic variables: HART command 3
The measured variables can be freely assigned to the dynamic variables.
Measured variables for PV (primary dynamic variable)
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary
dynamic variable)
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
The range of options increases if the measuring device has one or more
application packages.
Heartbeat Technology Application Package
Additional measured variables are available with the Heartbeat Technology
application package:
• Carrier pipe temperature
• Oscillation amplitude 0
Device variablesRead out the device variables: HART command 9
The device variables are permanently assigned.
A maximum of 8 device variables can be transmitted:
• 0 = mass flow
• 1 = volume flow
• 2 = corrected volume flow
• 3 = density
• 4 = reference density
• 5 = temperature
• 6 = totalizer 1
• 7 = totalizer 2
• 8 = totalizer 3
• 13 = target mass flow
• 14 = carrier mass flow
• 15 = concentration
Proline Promass P 100
PROFIBUS DP
Manufacturer ID0x11
Ident number0x1561
Profile version3.02
Device description files (GSD,
DTM, DD)
Information and files under:
• www.endress.com
On the product page for the device: Documents/Software → Device drivers
• www.profibus.org
14Endress+Hauser
Page 15
Proline Promass P 100
Output values
(from measuring device to
automation system)
Input values
(from automation system to
measuring device)
Supported functions• Identification & Maintenance
Configuration of the device
address
Analog input 1 to 8
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Density
• Reference density
• Concentration
• Temperature
• Carrier pipe temperature
• Electronic temperature
• Oscillation frequency
• Oscillation amplitude
• Frequency fluctuation
• Oscillation damping
• Tube damping fluctuation
• Signal asymmetry
• Exciter current
Digital input 1 to 2
• Partially filled pipe detection
• Low flow cut off
Totalizer 1 to 3
• Mass flow
• Volume flow
• Corrected volume flow
Analog output 1 to 3 (fixed assignment)
• Pressure
• Temperature
• Reference density
Digital output 1 to 3 (fixed assignment)
• Digital output 1: switch positive zero return on/off
• Digital output 2: perform zero point adjustment
• Digital output 3: switch switch output on/off
Totalizer 1 to 3
• Totalize
• Reset and hold
• Preset and hold
• Stop
• Operating mode configuration:
– Net flow total
– Forward flow total
– Reverse flow total
Simplest device identification on the part of the control system and
nameplate
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS
upload/download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing
diagnostic messages that occur
• System units
– Mass flow
– Mass
– Volume flow
– Volume
– Corrected volume flow
– Corrected volume
– Density
– Reference density
– Temperature
– Pressure
• Concentration application package
– Coefficients A0 to A4
– Coefficients B1 to B3
Order code for "Output":
Option M: Modbus RS485, for use in intrinsically safe areas (connection via Safety Barrier Promass 100)
Endress+Hauser27
Page 28
Proline Promass P 100
L
L
1
2
+
_
1
2
EtherNet/IP connection version
Order code for "Output", option N
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Options
A, B
Options
A, B, C
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, hygienic, stainless
• Option C ultra-compact, hygienic, stainless
6EtherNet/IP terminal assignment
1Power supply: DC 24 V
2EtherNet/IP
Connection methods available
Output
Device plugs
→ 30
Device plugs
→ 30
Power
supply
Terminals• Option L: plug M12x1 + thread NPT ½"
Device plugs
→ 30
Possible options for order code
"Electrical connection"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Option Q: 2 x plug M12x1
A0017054
Order code for
"Output"
Option NDC 24 VEtherNet/IP
Order code for "Output":
Option N: EtherNet/IP
Terminal number
Power supplyOutput
2 (L-)1 (L+)Device plug M12x1
28Endress+Hauser
Page 29
Proline Promass P 100
L
L
1
2
+
_
1
2
PROFINET connection version
Order code for "Output", option R
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Options
A, B
Options
A, B, C
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, hygienic, stainless
• Option C ultra-compact, hygienic, stainless
7PROFINET terminal assignment
1Power supply: DC 24 V
2PROFINET
Connection methods available
Output
Device plugs
→ 30
Device plugs
→ 30
Power
supply
Terminals• Option L: plug M12x1 + thread NPT ½"
Device plugs
→ 30
Possible options for order code
"Electrical connection"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Option Q: 2 x plug M12x1
A0017054
Order code for "Output":
Option R: PROFINET
Terminal number
Order code for
"Output"
Option RDC 24 VPROFINET
Power supplyOutput
2 (L-)1 (L+)Device plug M12x1
Endress+Hauser29
Page 30
Safety Barrier Promass 100
A
Safe area
1
L+2L
27
26
A B
Power
supply
24V
DC
Modbus
RS485
10207262
L+LA B
Power
supply
Modbus
RS485
Hazardous area
Safety Barrier
Promass 100
Power
Communication
Lift panel for
bus termination
1
2
A
1
2
4
3
5
8Safety Barrier Promass 100 with terminals
1Non-hazardous area and Zone 2/Div. 2
2Intrinsically safe area
Proline Promass P 100
A0016922
Pin assignment, device plug
Order codes for the M12x1 connectors, see the "Order code for electrical connection" column:
• 4-20 mA HART, pulse/frequency/switch output → 24
• PROFIBUS DP→ 25
• Modbus RS485 → 26
• EtherNet/IP → 28
• PROFINET→ 29
Supply voltage
For all connection versions except MODBUS RS485 intrinsically safe (device side)
Device plug MODBUS RS485 intrinsically safe with supply voltage → 31
PinAssignment
1L+DC 24 V
2Not assigned
3Not assigned
4L-DC 24 V
A0016809
5Grounding/shielding
CodingPlug/socket
APlug
The following is recommended as a socket:
• Binder, series 763, part no. 79 3440 35 05
• Alternatively: Phoenix part no. 1669767 SAC-5P-M12MS
– With the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
– With the order code for "Output", option N: EtherNet/IP
• When using the device in a hazardous location: Use a suitably certified socket.
30Endress+Hauser
Page 31
Proline Promass P 100
3
2
4
1
5
3
2
4
1
5
1
2
4
3
5
4-20 mA HART with pulse/frequency/switch output
Device plug for signal transmission (device side)
PinAssignment
1+4-20 mA HART (active)
2-4-20 mA HART (active)
3+Pulse/frequency/switch output (passive)
4-Pulse/frequency/switch output (passive)
A0016810
• Recommended plug: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location, use a suitably certified plug.
5Grounding/shielding
CodingPlug/socket
ASocket
PROFIBUS DP
For use in the non-hazardous area and Zone 2/Div. 2.
Device plug for signal transmission (device side)
PinAssignment
1Not assigned
2APROFIBUS DP
3Not assigned
4BPROFIBUS DP
A0016811
• Recommended plug: Binder, series 763, part no. 79 4449 20 05
• When using the device in a hazardous location, use a suitably certified plug.
5Grounding/shielding
CodingPlug/socket
BSocket
MODBUS RS485
Device plug for signal transmission with supply voltage (device side), MODBUS RS485 (intrinsically
safe)
PinAssignment
1L+Supply voltage, intrinsically safe
A0016809
2A
3B
4L-Supply voltage, intrinsically safe
5Grounding/shielding
CodingPlug/socket
APlug
Modbus RS485 intrinsically safe
• Recommended socket: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location: Use a suitably certified socket.
Endress+Hauser31
Page 32
Proline Promass P 100
3
2
4
1
5
3
2
4
1
3
2
4
1
Device plug for signal transmission (device side), MODBUS RS485 (not intrinsically safe)
For use in the non-hazardous area and Zone 2/Div. 2.
PinAssignment
1Not assigned
2AModbus RS485
3Not assigned
4BModbus RS485
A0016811
• Recommended plug: Binder, series 763, part no. 79 4449 20 05
• When using the device in a hazardous location, use a suitably certified plug.
EtherNet/IP
Device plug for signal transmission (device side)
5Grounding/shielding
CodingPlug/socket
BSocket
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0016812
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
CodingPlug/socket
DSocket
PROFINET
Device plug for signal transmission (device side)
PinAssignment
1+TD +
2+RD +
3-TD –
4-RD –
A0016812
CodingPlug/socket
DSocket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
Supply voltage
The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
32Endress+Hauser
Page 33
Proline Promass P 100
Transmitter
For device version with communication type:
• HART, PROFIBUS DP, EtherNet/IP: DC 20 to 30 V
• Modbus RS485, device version:
– For use in the non-hazardous area and Zone 2/Div. 2: DC 20 to 30 V
– For use in the intrinsically safe area: power supply via Safety Barrier Promass 100
Safety Barrier Promass 100
DC 20 to 30 V
Power consumptionTransmitter
Order code for "Output"
Option B: 4-20 mA HART with pulse/frequency/switch output3.5 W
Option L: PROFIBUS DP3.5 W
Option M Modbus RS485, for use in non-hazardous areas and Zone 2/
Div. 2
Option M: Modbus RS485, for use in intrinsically safe areas2.45 W
Option N: EtherNet/IP3.5 W
Option R: PROFINET3.5 W
Safety Barrier Promass 100
Order code for "Output"
Option M: Modbus RS485, for use in intrinsically safe areas4.8 W
Option M Modbus RS485, for use in non-hazardous
areas and Zone 2/Div. 2
Option M: Modbus RS485, for use in intrinsically safe
areas
Option N: EtherNet/IP145 mA18 A (< 0.125 ms)
Option R: PROFINET145 mA18 A (< 0.125 ms)
Maximum
Power consumption
Maximum
Power consumption
Maximum
Current consumption
90 mA10 A (< 0.8 ms)
145 mA16 A (< 0.4 ms)
switch-on current
3.5 W
Maximum
Safety Barrier Promass 100
Power supply failure
Order code for "Output"
Option M: Modbus RS485, for use in intrinsically safe
areas
• Totalizers stop at the last value measured.
Maximum
Current consumption
230 mA10 A (< 0.8 ms)
Maximum
switch-on current
• Depending on the device version, the configuration is retained in the device memory or in the
plug-in memory (HistoROM DAT).
• Configuration is retained in the plug-in memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Endress+Hauser33
Page 34
Electrical connectionConnecting the transmitter
121234
ABC
4
4...20 mA
+
–
5
2
1
3
6
+
_
AHousing version: compact, aluminum coated
BHousing version: compact hygienic, stainless
1Cable entry or device plug for signal transmission
2Cable entry or device plug for supply voltage
CHousing version: ultra-compact, hygienic, stainless, M12 device plug
3Device plug for signal transmission
4Device plug for supply voltage
Proline Promass P 100
A0016924
• Terminal assignment → 23
• Pin assignment, device plug → 30
In the case of device versions with a connector, the transmitter housing does not need to be
opened to connect the signal cable or power supply cable.
Connection examples
Current output 4-20 mA HART
9Connection example for 4-20 mA HART current output (active)
1Automation system with current input (e.g. PLC)
2Cable shield, observe cable specifications
3Connection for HART operating devices
4Resistor for HART communication (≥ 250 Ω): observe maximum load
5Analog display unit: observe maximum load
6Transmitter
A0016800
34Endress+Hauser
Page 35
Proline Promass P 100
1
+
_
12345
2
+
–
+
–
3
1
+
_
+
_
2
+
_
3
Pulse/frequency output
A0016801
10Connection example for pulse/frequency output (passive)
1Automation system with pulse/frequency input (e.g. PLC)
2Power supply
3Transmitter: observe input values → 9
Switch output
11Connection example for switch output (passive)
1Automation system with switch input (e.g. PLC)
2Power supply
3Transmitter: observe input values
A0016802
Endress+Hauser35
Page 36
Proline Promass P 100
2
1
A
B
A
B
A
B
3
3
2
1
A
B
A
B
A
B
3
4
4
PROFIBUS DP
A0021429
12Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2
1Control system (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Transmitter
If baud rates > 1.5 MBaud an EMC cable entry must be used and the cable shield must continue
as far as the terminal wherever possible.
Modbus RS485
Modbus RS485, non-hazardous area and Zone 2/Div. 2
A0016803
13Connection example for Modbus RS485, non-hazardous area and Zone 2/Div. 2
1Control system (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Distribution box
4Transmitter
36Endress+Hauser
Page 37
Proline Promass P 100
21
A
B
A
B
3
L+
L-
A
B
L+
L-
L- L+
A
B
567
8
4
124
5
5
3
Modbus RS485 intrinsically safe
A0016804
14Connection example for Modbus RS485 intrinsically safe
1Control system (e.g. PLC)
2Cable shield, observe cable specifications
3Safety Barrier Promass 100
4Observe cable specifications
5Non-hazardous area
6Non-hazardous area and Zone 2/Div. 2
7Intrinsically safe area
8Transmitter
17Connection example for HART input (burst mode) via current output (active)
1Cable shield, observe cable specifications
2Resistor for HART communication (≥ 250 Ω): observe maximum load
3Connection for HART operating devices
4Analog display unit
5Transmitter
6Sensor for external measured variable
A0019828
38Endress+Hauser
Page 39
Proline Promass P 100
3
4...20 mA
4
521
+
-
6
4...20 mA
+
7
3
4
52
+
-
-
+
A0019830
18Connection example for HART input (master mode) via current output (active)
1Automation system with current input (e.g. PLC).
Prerequisite: automation system with HART version 6, HART commands 113 and 114 can be processed.
2Cable shield, observe cable specifications
3Resistor for HART communication (≥ 250 Ω): observe maximum load
4Connection for HART operating devices
5Analog display unit
6Transmitter
7Sensor for external measured variable
Potential equalizationRequirements
No special measures for potential equalization are required.
Please consider the following to ensure correct measurement:
• Same electrical potential for the fluid and sensor
• Company-internal grounding concepts
For devices intended for use in hazardous locations, please observe the guidelines in the Ex
documentation (XA).
Terminals
Transmitter
Spring terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Safety Barrier Promass 100
Plug-in screw terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Cable entries
• Cable gland: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
– M20
Cable specificationPermitted temperature range
• –40 °C (–40 °F) to +80 °C (+176 °F)
• Minimum requirement: cable temperature range ≥ ambient temperature +20 K
Power supply cable
Standard installation cable is sufficient.
Signal cable
Current output
For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Endress+Hauser39
Page 40
Proline Promass P 100
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance<30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
>0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance<30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
>0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
PROFINET
Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT
5e and CAT 6 are recommended.
For more information on planning and installing PROFINET networks, see: "PROFINET Cabling
and Interconnection Technology", Guideline for PROFINET
Connecting cable between Safety Barrier Promass 100 and measuring device
Cable typeShielded twisted-pair cable with 2x2 wires. When grounding the cable shield,
observe the grounding concept of the plant.
Maximum cable resistance 2.5 Ω, one side
Comply with the maximum cable resistance specifications to ensure the operational reliability
of the measuring device.
40Endress+Hauser
Page 41
Proline Promass P 100
The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-sectionMaximum cable length
[mm2][AWG][m][ft]
0.52070230
0.7518100328
1.017100328
1.516200656
2.514300984
Performance characteristics
Reference operating
conditions
Maximum measured error
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool → 98
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals → 44
Mass flow and volume flow (liquids)
±0.10 % o.r.
Density (liquids)
Under
reference operating conditions
[g/cm³][lbs/in³][g/cm³][lbs/in³][g/cm³][lbs/in³]
±0.0005±0.00097±0.01±0.019±0.002±0.0039
1)Valid over the entire temperature and density range
2)Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
3)Order code for "Application package", option EF "Special density and concentration "
Standard density
calibration
1)
Wide-range
density specification
2) 3)
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DNZero point stability
[mm][in][kg/h][lb/min]
8³⁄₈0.200.007
15½0.650.024
2511.800.066
Endress+Hauser41
Page 42
Proline Promass P 100
DNZero point stability
[mm][in][kg/h][lb/min]
401½4.500.165
5027.00.257
Flow values
Flow values as turndown parameter depending on nominal diameter.
In the case of analog outputs, the output accuracy must also be considered for the measured
error; in contrast, this need not be considered in the case of fieldbus outputs (e.g. Modbus
RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
Current output
AccuracyMax. ±5 µA
Pulse/frequency output
o.r. = of reading
AccuracyMax. ±50 ppm o.r. (across the entire ambient temperature range)
Repeatability
o.r. = of reading; 1 g/cm3 = 1 kg/l; T = medium temperature
42Endress+Hauser
Page 43
Proline Promass P 100
[kg/m ]
3
[°C]
-40
0
-80
40 80 120 160 200 240 280 320 360 400
[°F]
-50
0
50
100150
200
0
2
4
6
8
10
14
12
16
1
2
18
20
Base repeatability
Mass flow and volume flow (liquids)
±0.05 % o.r.
Design fundamentals → 44
Response time
Influence of ambient
temperature
Influence of medium
temperature
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
The response time depends on the configuration (damping).
Current output
o.r. = of reading
Temperature coefficientMax. ±0.005% o.r./°C
Pulse/frequency output
Temperature coefficientNo additional effect. Included in accuracy.
Mass flow and volume flow
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error of the sensor is ±0.0002 % of the full scale value/°C
(±0.0001 % of the full scale value/°F).
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is ±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F).
Field density calibration is possible.
Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 41) the measured error is
±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F)
A0016610
1Field density calibration, for example at +20 °C (+68 °F)
2Special density calibration
Endress+Hauser43
Page 44
Temperature
ZeroPoint
BaseAccu
⋅ 100
³
± BaseAccu
ZeroPoint
BaseAccu
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
±
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
³
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
½ ⋅±
E [%]
0102030405060
0
0.5
1.0
1.5
2.0
2.5
708090100
Q [%]
±0.005 · T °C (± 0.005 · (T – 32) °F)
Proline Promass P 100
Influence of medium
pressure
Design fundamentals
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
o.r. = of reading
DN[% o.r./bar][% o.r./psi]
[mm][in]
8³⁄₈–0.002–0.0001
15½–0.006–0.0004
251–0.005–0.0003
401½–0.007–0.0005
502–0.006–0.0004
o.r. = of reading, o.f.s. = of full scale value
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rateMaximum measured error in % o.r.
A0021332
A0021339
A0021333
Calculation of the maximum repeatability as a function of the flow rate
Flow rateMaximum repeatability in % o.r.
A0021335
A0021336
Example for max. measured error
A0021334
A0021340
A0021337
EError: Maximum measured error as % o.r. (example)
QFlow rate as %
44Endress+Hauser
A0024176
Page 45
Proline Promass P 100
1
2
3
4
5
Mounting location
Installation
No special measures such as supports are necessary. External forces are absorbed by the construction
of the device.
A0023344
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter
prevent the sensor running empty while measurement is in progress.
19Installation in a down pipe (e.g. for batching applications)
1Supply tank
2Sensor
3Orifice plate, pipe restriction
4Valve
5Batching tank
A0015596
DNØ orifice plate, pipe restriction
[mm][in][mm][in]
8³⁄₈60.24
15½100.40
251140.55
401½220.87
502281.10
Endress+Hauser45
Page 46
Proline Promass P 100
12
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
flow direction (direction of medium flow through the piping).
OrientationRecommendation
AVertical orientation
A0015591
BHorizontal orientation, transmitter
head up
→ 20, 46
A0015589
CHorizontal orientation, transmitter
head down
→ 20, 46
A0015590
DHorizontal orientation, transmitter
head at side
A0015592
1)Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
2)Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
1)
Exceptions:
2)
Exceptions:
Inlet and outlet runs
Special mounting
instructions
If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to
the fluid properties.
A0014057
20Orientation of sensor with curved measuring tube
1Avoid this orientation for fluids with entrained solids: Risk of solids accumulating.
2Avoid this orientation for outgassing fluids: Risk of gas accumulating.
No special precautions need to be taken for fittings which create turbulence, such as valves, elbows
or T-pieces, as long as no cavitation occurs → 55.
Complete drainability guaranteed
When the sensor is installed in a horizontal line, eccentric clamps can be used to ensure complete
drainability. When the system is pitched in a specific direction and at a specific slope, gravity can be
used to achieve complete drainability. The sensor must be mounted in the correct position to ensure
full drainability in the horizontal position. Markings on the sensor show the correct mounting
position to optimize drainability.
46Endress+Hauser
Page 47
Proline Promass P 100
1
4
3
1
2
5
A
B
C
A0016583
1Eccentric clamp connection
2"This side up" label indicates which side is up
3Slope the device in accordance with the hygiene guidelines. Slope: approx. 2 ° or 35 mm/m (0.42 in/feet)
4Line on the underside indicates the lowest point of the eccentric process connection.
5Transmitter
Securing with mounting clamp in the case of hygiene connections
It is not necessary to provide additional support for the sensor for operational performance purposes.
If, however, additional support is required for installation purposes, the following dimensions must
be observed.
Use mounting clamp with lining between clamp and measuring instrument.
A0016588
DNABC
[mm][in][mm][in][mm][in][mm][in]
8³⁄₈29811.73331.3281.1
15½40215.83331.3281.1
25154221.34331.3381.5
401 ½65825.9136.51.44562.2
50277230.3944.11.74752.95
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration
takes place under reference conditions → 41. Therefore, a zero point adjustment in the field is
generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or very highviscosity fluids).
Endress+Hauser47
Page 48
Mounting Safety Barrier
> 22.5
(> 0.89)
> 22.5
(> 0.89)
Promass 100
Proline Promass P 100
A0016894
21Minimum distance between additional Safety Barrier Promass 100 or other modules.Engineering unit mm
(in)
Ambient temperature range
Environment
Measuring device Non-Ex–40 to +60 °C (–40 to +140 °F)
Ex na, NI version –40 to +60 °C (–40 to +140 °F)
Ex ia, IS version• –40 to +60 °C (–40 to +140 °F)
• –50 to +60 °C (–58 to +140 °F) (order code for "Test, certificate",
option JM))
Readability of the local display–20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside
the temperature range.
Safety Barrier Promass 100–40 to +60 °C (–40 to +140 °F)
If operating outdoors:
‣
Avoid direct sunlight, particularly in warm climatic regions.
Weather protection covers can be ordered from Endress+Hauser: see "Accessories" section
Temperature tables
In the following tables, the following interdependencies between the maximum medium
temperature Tm for T6 to T1 and the maximum ambient temperature Ta apply when operating the
device in hazardous areas.
1)The following applies for specified sensors with a maximum medium temperature Tm = 401 °F: Tm = 338 °F
2)The following applies for specified sensors with a maximum medium temperature Tm = 401 °F: Tm = 401 °F
Endress+Hauser49
Page 50
Proline Promass P 100
Ta
[°C]
T1
[450°C]
35140
50140
60140
35140
45140
50140
5085120140140
–85120140140
––120140140
5085120140140
–85120140140
––120140140
T6
[85° C]T5[100°C]T4[135°C]T3[200°C]T2[300°C]
Explosion hazards arising from gas and dust
Determining the temperature class and surface temperature with the temperature table
• In the case of gas: Determine the temperature class as a function of the ambient temperature T
and the medium temperature Tm.
• In the case of dust: Determine the maximum surface temperature as a function of the maximum
ambient temperature Ta and the maximum medium temperature Tm.
Example
• Measured maximum ambient temperature: Tma = 47 °C
• Measured maximum medium temperature: Tmm = 108 °C
a
Storage temperature
Climate class
Degree of protection
A0019758
22Procedure for determining the maximum surface temperature
1.Select device (optional).
2.In the column for the maximum ambient temperature Ta select the temperature that is
immediately greater than or equal to the measured maximum ambient temperature Tma that is
present.
Ta = 50 °C.
The row showing the maximum medium temperature is determined.
3.Select the maximum medium temperature Tm of this row, which is larger or equal to the
measured maximum medium temperature Tmm.
The column with the temperature class for gas is determined: 108 °C ≤ 120°C → T4.
4.The maximum temperature of the temperature class determined corresponds to the maximum
surface temperature for dust: T4 = 135 °C
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) (standard version)
–50 to +80 °C (–58 to +176 °F) (Order code for "Test, certificate", option JM)
DIN EN 60068-2-38 (test Z/AD)
Transmitter and sensor
• As standard: IP66/67, type 4X enclosure
• With the order code for "Sensor options", option CM: IP69K can also be ordered
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Safety Barrier Promass 100
IP20
Vibration resistance
Compact version
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g2/Hz
50Endress+Hauser
– 200 to 2000 Hz, 0.001 g2/Hz
– Total: 1.54 g rms
Page 51
Proline Promass P 100
PN40
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
200
400
600
0
800
[bar][psi]
Shock resistance
Shock resistance
Interior cleaning
Electromagnetic
compatibility (EMC)
Compact version
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Compact version
Rough handling shocks according to IEC 60068-2-31
• Cleaning in place (CIP)
• Sterilization in place (SIP)
• Cleaning with pigs
Options
Oil- and grease-free version for wetted parts, without inspection certificate
Order code for "Service", option HA
• Depends on the communication protocol:
– HART, PROFIBUS DP, Modbus RS485, EtherNet/IP:
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
– PROFINET: as per IEC/EN 61326
• Complies with emission limits for industry as per EN 55011 (Class A)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
For details, refer to the Declaration of Conformity.
Medium temperature range
Density
Pressure-temperature
ratings
Process
Sensor
• –50 to +150 °C (–58 to +302 °F)
• –50 to +205 °C (–58 to +401 °F) with extended temperature (order code for "Measuring tube mat."
option TD, TG)
Seals
No internal seals
0 to 5000 kg/m3 (0 to 312 lb/cf)
The following pressure-temperature ratings refer to the entire device and not just the process
connection.
Flange according to EN 1092-1 (DIN 2501)
23With flange material 1.4404 (F316/F316L).
Endress+Hauser51
A0027769-EN
Page 52
Proline Promass P 100
Class 150
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
20K
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
DN 8...40
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
[bar][psi]
200
400
600
0
≥DN50
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Flange according to ASME B16.5
A0026540-EN
24With flange material 1.4404 (F316/F316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Flange JIS B2220
25With flange material 1.4404 (F316/F316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Flange DIN 11864-2 Form A
26With flange material 1.4435 (316L)
52Endress+Hauser
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
A0026541-EN
A0027781-EN
Page 53
Proline Promass P 100
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
[bar][psi]
200
400
0
DN8…40
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
[bar][psi]
200
400
600
0
≥DN50
DN 8...40
DN 50
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
[bar][psi]
200
400
600
0
Neumo BioConnect, BBS flange
A0027782-EN
27With flange material 1.4435 (316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Thread DIN 11851
A0027783-EN
28With flange material 1.4435 (316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
DIN 11851 allows for applications up to +140 °C (+284 °F) if suitable sealing materials are used.
Please take this into account when selecting seals and counterparts, as these components can limit
the pressure and temperature range.
Thread DIN 11864-1 Form A
Endress+Hauser53
29With connection material 1.4435 (316L)
A0027784-EN
Page 54
Proline Promass P 100
[bar][psi]
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
200
400
0
0
2
6
[bar]
100
50
0
[psi]
300[°F]350250200150100500-50
[°C]200150100500-50
400
8
DN 8...40
DN 50
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
[bar][psi]
200
400
600
0
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Thread ISO 2853
A0027785-EN
30With connection material 1.4435 (316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
Thread SMS 1145
31With connection material 1.4435 (316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please
take this into account when selecting seals and counterparts, as these components can limit the
pressure and temperature range.
Clamp connections DIN 11864-3 Form A
54Endress+Hauser
32With connection material 1.4435 (316L)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TD, TG
A0027786-EN
A0027784-EN
Page 55
Proline Promass P 100
Tri-Clamp; clamp connection ISO 2852, DIN 32676, BBS, Neumo BioConnect
The clamp connections are suitable up to a maximum pressure of 16 bar (232 psi). Please observe
the operating limits of the clamp and seal used as they can be over 16 bar (232 psi). The clamp and
seal are not included in the scope of supply.
Secondary containment
pressure rating
The sensor housing is filled with dry nitrogen and protects the electronics and mechanics inside.
The following secondary containment pressure rating is only valid for a fully welded sensor housing
and/or a device equipped with closed purge connections (never opened/as delivered).
DNSecondary containment
pressure rating
(designed with a safety factor
≥ 4)
[mm][in][bar][psi][bar][psi]
8³⁄₈253621902755
15½253621752535
251253621652930
401½253621522204
502253621031493
Secondary containment burst pressure
If there is a risk of measuring tube failure due to process characteristics, e.g. with corrosive
fluids, we recommend the use of sensors whose secondary containment is equipped with special
pressure monitoring connections (order code for "Sensor option", option CH "Purge connection").
With the help of these connections, the fluid collected in the secondary containment can be
bled off in the event of tube failure. This is especially important in high-pressure gas
applications. These connections can also be used for gas purging (gas detection).
Do not open the purge connections unless the containment can be filled immediately with a
dry, inert gas. Use only low gauge pressure to purge. Maximum pressure: 5 bar (72.5 psi).
If a device fitted with purge connections is connected to the purge system, the maximum
nominal pressure is determined by the purge system itself or by the device, depending on which
component has the lower nominal pressure.
Dimensions: → 58
Flow limit
Pressure loss
System pressure
Select the nominal diameter by optimizing between the required flow range and permissible pressure
loss.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):
flow velocity < 1 m/s (< 3 ft/s).
To calculate the pressure loss, use the Applicator sizing tool → 98
It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
Endress+Hauser55
Page 56
Proline Promass P 100
t
a
0
10
30
20
40
[mm][in]
8090100110
[°C]
[°F]
200250290
0.5
1.0
1.5
0
T
40(104)
T
60(140)
t
120130140
T
m
A0015594
Thermal insulation
In the case of some fluids, it is important that the heat radiated from the sensor to the transmitter is
kept to a minimum. A wide range of materials can be used for the required insulation.
A0019919
aMinimum distance to insulation
tmaximum Insulation thickness
The minimum distance between the transmitter housing and the insulation is 10 mm (0.39 in) so
that the transmitter head remains completely exposed.
Maximum recommended insulation thickness
33Maximum recommended insulation thickness depending on the temperature of the medium and the
ambient temperature
tInsulation thickness
T
m
T
40(104)
T
60(140)
Medium temperature
Maximum recommended insulation thickness at an ambient temperature of Ta = 40 °C (104 °F)
Maximum recommended insulation thickness at an ambient temperature of Ta = 60 °C (140 °F)
Maximum recommended insulation thickness for the extended temperature range and
insulation
For the extended temperature range, version with long extension neck, order code for "Measuring
tube material", option TD, TG or extension neck for insulation, order code for "Sensor option", option
CG:
56Endress+Hauser
A0023173
Page 57
Proline Promass P 100
0
20
60
40
80
[in]
1
2
3
0
t
100
4
[mm]
T
40(104)
T
60(140)
80100120140
200300400
160180200
[°F]
T
m
[°C]
34Maximum recommended insulation thickness depending on the temperature of the medium and the
ambient temperature
tInsulation thickness
T
m
T
40(104)
T
60(140)
Medium temperature
Maximum recommended insulation thickness at an ambient temperature of Ta = 40 °C (104 °F)
Maximum recommended insulation thickness at an ambient temperature of Ta = 60 °C (140 °F)
A0023177
Heating
NOTICE
Danger of overheating with insulation
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F)
NOTICE
The insulation can also be thicker than the maximum recommended insulation thickness.
Prerequisite:
Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣
Ensure that a sufficiently large area of the housing support remains exposed. The uncovered part
‣
serves as a radiator and protects the electronics from overheating and excessive cooling.
Some fluids require suitable measures to avoid loss of heat at the sensor.
Heating options
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Heating jackets for the sensors can be ordered as accessories from Endress+Hauser → 97.
NOTICE
Danger of overheating when heating
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F).
Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣
Ensure that a sufficiently large area of the housing support remains exposed. The uncovered part
‣
serves as a radiator and protects the electronics from overheating and excessive cooling.
Vibrations
Endress+Hauser57
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
Page 58
Mechanical construction
A
DC
B
E
H
F
L
G
K
Dimensions in SI unitsCompact version
Order code for "Housing", option A "Compact coated aluminum"
Proline Promass P 100
A0016525
DN
[mm]A[mm]B[mm]C[mm]D[mm]
8136147.593.5541912998.3110892
15136147.593.55419129912.0010892
25136147.593.55419131217.6012192
40136147.593.55423140726.00176141
50136147.593.55425651638.00260168
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +70 mm
2)If using a display, order code for "Display; Operation", option B: values +28 mm
3)Dependent on the particular process connection
1) 2)
E
[mm]
1) 2)
F
[mm]G[mm]H[mm]K[mm]L[mm]
3)
3)
3)
3)
3)
58Endress+Hauser
Page 59
Proline Promass P 100
G
K
DC
B
A
E
H
F
L
Order code for "Housing", option B "Compact hygienic, stainless"
A0016526
DN
[mm]A[mm]B[mm]C[mm]D[mm]
1) 2)
E
[mm]
8133.5136.87858.81862948.3110892
15133.5136.87858.818629412.0010892
25133.5136.87858.818630717.6012192
40133.5136.87858.822640226.00176141
50133.5136.87858.825151138.00260168
1) 2)
F
[mm]G[mm]H[mm]K[mm]L[mm]
3)
3)
3)
3)
3)
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +70 mm
2)If using a display, order code for "Display; Operation", option B: values +14 mm
3)Dependent on the particular process connection
Endress+Hauser59
Page 60
Order code for "Housing", option C "Ultra-compact hygienic, stainless"
G
K
ADC
B
E
H
F
L
Proline Promass P 100
A0016527
DN
[mm]A[mm]B[mm]C[mm]D[mm]
1) 2)
E
[mm]
8111.4123.667.755.91862948.3110892
15111.4123.667.755.918629412.0010892
25111.4123.667.755.918630717.6012192
40111.4123.667.755.922640226.00176141
50111.4123.667.755.925151138.00260168
1) 2)
F
[mm]G[mm]H[mm]K[mm]L[mm]
3)
3)
3)
3)
3)
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +70 mm
2)If using a display, order code for "Display; Operation", option B: values +14 mm
3)Dependent on the particular process connection
60Endress+Hauser
Page 61
Proline Promass P 100
C
D
A
B
E
L
+1.5 (+0.06)
−2.0 (–0.08)
±3.5 (±0.14)
DN 125 (5")≥DN 100 (4")≤
Flange connections
Fixed flange EN 1092-1, ASME B16.5, JIS B2220
A0015621
35Engineering unit mm (in)
Flange according to EN 1092-1 (DIN 2501): PN 40
1.4404 (316/316L)
Order code for "Process connection", option D2W
DN
[mm]
1)
8
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
95654 × Ø1417.017.3336
1595654 × Ø1420.017.3440
25115854 × Ø1419.028.5580
401501104 × Ø1821.043.1707
501651254 × Ø1825.054.5828
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
1)DN 8 with DN 15 flanges as standard
Flange according to ASME B16.5: Cl 150
1.4404 (316/316L)
Order code for "Process connection", option AAW
DN
[mm]
1)
8
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
9060.34 × Ø15.711.215.7336
159060.34 × Ø15.711.215.7440
2511079.44 × Ø15.714.226.7580
4012598.44 × Ø15.717.540.9707
50150120.78 × Ø19.119.152.6828
Surface roughness (flange): Ra 3.2 to 6.3 µm
L
[mm]
L
[mm]
1)DN 8 with DN 15 flanges as standard
Endress+Hauser61
Page 62
Proline Promass P 100
Flange JIS B2220: 20K
1.4404 (316/316L)
Order code for "Process connection", option NEW
DN
[mm]
1)
8
1595704 × Ø1516.015440
25125904 × Ø1917.525580
401401054 × Ø1920.040707
501551208 × Ø1927.550828
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)DN 8 with DN 15 flanges as standard
A
[mm]
95704 × Ø1516.015336
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
62Endress+Hauser
Page 63
Proline Promass P 100
C
D
A
B
E
XX
L
+1,5 (+0.06)
-2,0 (-0.08)
Fixed flange DIN 11864-2
A0015627
36Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Flange DIN 11864-2 Form A, for pipe according to DIN 11866 series A, flat flange
1.4435 (316L)
Order code for "Process connection", option KJW
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
859.0424 × Ø91016.00384
1559.0424 × Ø91016.00488
2570.0534 × Ø91026.00626
4082.0654 × Ø91038.00753
5094.0774 × Ø91050.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
Endress+Hauser63
Page 64
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
C
D
A
B
E
X
X
BBS fixed flange
A0016910
37Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
BBS flange small (sterile orbital), for pipe according to DIN 11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BSK
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
859424 × Ø91010.00384
1559424 × Ø91016.00488
2570534 × Ø91026.00626
4082654 × Ø91038.00753
5094774 × Ø91050.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS flange small (sterile orbital), for pipe according to DIN 11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BSL
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
859424 × Ø91014.00384
1562454 × Ø91018.10488
2574574 × Ø91029.70626
4088714 × Ø91044.30753
50103854 × Ø91056.30877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
64Endress+Hauser
Page 65
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
C
D
A
B
E
X
X
Neumo BioConnect fixed flange
A0016907
38Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Neumo BioConnect flange, for pipe according to DIN 11866 series A, flange form R
1.4435 (316L)
Order code for "Process connection", option BSB
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
865454 × Ø91010.00384
1575554 × Ø91016.00488
2585654 × Ø91226.00626
40100804 × Ø91238.00753
50110904 × Ø91450.00877
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
Endress+Hauser65
Page 66
Clamp connections
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
≥(1")DN 25DN 8, 15 (¾", ½")
A
B
Tri-Clamp
39Engineering unit mm (in)
Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FHW
DN
[mm]
8½25.09.40362
15¾25.015.75466
251
401½
502
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp
[in]
1)
1)
1)
A
[mm]
50.422.10606
50.434.80731
63.947.50853
Proline Promass P 100
B
[mm]
A0023342
L
[mm]
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
½" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FCW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
15½25.09.40466
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
¾" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FFW
DN
[mm]
Clamp
[in]
A
[mm]
B
[mm]
L
[mm]
8¾25.015.75362
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
66Endress+Hauser
Page 67
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
1" Tri-Clamp, for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FPW
DN
[mm]
81
151
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp
[in]
1)
1)
A
[mm]
B
[mm]
50.422.10362
50.422.10466
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
Tri-Clamp eccentric, for pipe according to DIN 11866 series C
1.4435 (316L)
DN
[mm]
Order code for
"Process connection",
Clamp
[in]
A
[mm]
B
[mm]
option
8FEB½25.09.40362
15FED¾25.015.75466
25FEF1
40FEH1½
50FEK2
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
1)
1)
1)
50.422.10606
50.434.80738
63.947.50860
Additional information on "Eccentric clamps"
L
[mm]
L
[mm]
1)The connection complies with the hygienic clamp dimensions as per ASME BPE.
Clamp connection DIN 32676, ISO 2852
A0015625
40Engineering unit mm (in)
Clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KPW
DN
[mm]
834.016.00362
1534.016.00466
2550.526.00606
A
[mm]
B
[mm]
L
[mm]
Endress+Hauser67
Page 68
Clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KPW
DN
[mm]
A
[mm]
B
[mm]
4050.538.00732
5064.050.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp ISO 2852 , for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSA
DN
[mm]
A
[mm]
B
[mm]
850.522.6362
1550.522.6466
2550.522.6606
4050.535.6731
5064.048.6853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Proline Promass P 100
L
[mm]
L
[mm]
Clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JSC
DN
[mm]
A
[mm]
B
[mm]
834.014.00362
1534.018.10466
2550.529.70606
4064.044.30731
5077.556.30853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
68Endress+Hauser
Page 69
Proline Promass P 100
A
B
X
L
+1,5 (+0.06)
-2,0 (-0.08)
X
L
+1,5 (+0.06)
-2,0 (-0.08)
X
A
B
X
Clamp connection DIN 11864-3
A0016903
41Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Clamp DIN 11864-3 Form A, for pipe according to DIN 11866 series A, with notch
1.4435 (316L)
Order code for "Process connection", option KLW
DN
[mm]
834.016.05370
1534.016.05474
2550.526.05614
4064.038.05738
5077.550.05853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
B
[mm]
L
[mm]
BBS clamp connection
A0016908
42Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Endress+Hauser69
Page 70
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
X
A
B
X
BBS Quick Connect (sterile orbital), for pipe according to DIN 11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BSE
DN
[mm]
825.010.00362
1550.516.00466
2550.526.00606
4064.038.00732
5077.550.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS Quick Connect (sterile orbital) for pipe according to DIN11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BSJ
DN
[mm]
825.014.00362
1550.518.10466
2550.529.70606
4064.044.30738
5077.556.30860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
A
[mm]
A
[mm]
B
[mm]
B
[mm]
L
[mm]
L
[mm]
Neumo BioConnect clamp connection
A0016905
43Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
70Endress+Hauser
Page 71
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
Clamp, Neumo BioConnect, for pipe according to DIN 11866 series A, clamp form R
1.4435 (316L)
Order code for "Process connection", option BSA
DN
[mm]
A
[mm]
B
[mm]
825.010.00362
1525.016.00466
2550.426.00606
4064.038.00732
5077.450.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Eccentric clamp connection DIN 32676, ISO 2852
L
[mm]
A0016543
44Engineering unit mm (in)
Eccentric clamp DIN 32676, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KRW
DN
[mm]
A
[mm]
B
[mm]
L
[mm]
834.010.00362
1534.016.00466
2550.526.00606
5064.050.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JEC
DN
[mm]
A
[mm]
B
[mm]
L
[mm]
834.010.30362
1534.014.00466
2534.018.10606
4050.529.70738
Endress+Hauser71
Page 72
Proline Promass P 100
X
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
X
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B
1.4435 (316L)
Order code for "Process connection", option JEC
DN
[mm]
5064.044.30853
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
Eccentric clamp ISO 2852, for pipe according to DIN11866 series B, for connection to DN15 pipes
1.4435 (316L)
Order code for "Process connection", option JED
DN
[mm]
2550.518.10606
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
A
[mm]
B
[mm]
B
[mm]
L
[mm]
L
[mm]
Eccentric clamp connection DIN 11864-3
A0016904
45Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Eccentric clamp DIN 11864-3 Form A, for pipe according to DIN 11866 series A, with notch
1.4435 (316L)
Order code for "Process connection", option KNW
DN
[mm]
834.010.00370
1534.016.00474
2550.526.00624
5077.550.00869
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
L
[mm]
72Endress+Hauser
Page 73
Proline Promass P 100
X
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
X
BBS eccentric clamp connection
A0016909
46Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
BBS Quick Connect (sterile orbital), eccentric, for pipe according to DIN11866 series A, female
1.4435 (316L)
Order code for "Process connection", option BEJ
DN
[mm]
825.010.00362
1550.516.00466
2550.526.00606
5077.550.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
L
[mm]
BBS Quick Connect (sterile orbital), eccentric, for pipe according to DIN11866 series B, female
1.4435 (316L)
Order code for "Process connection", option BEK
DN
[mm]
A
[mm]
B
[mm]
L
[mm]
825.010.30362
1550.514.00466
2550.518.10606
4050.529.70738
5064.044.30860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
Endress+Hauser73
Page 74
Proline Promass P 100
X
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
X
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
Neumo BioConnect eccentric clamp connection
A0016906
47Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Neumo BioConnect eccentric clamp connection
1.4435 (316L)
Order code for "Process connection", option BEA
DN
[mm]
82510.00362
152516.00466
252526.00610
502550.00859
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
A
[mm]
B
[mm]
L
[mm]
Cable glands
Thread DIN 11851, DIN 11864-1, SMS 1145, BBS
48Engineering unit mm (in)
A0015628
74Endress+Hauser
Page 75
Proline Promass P 100
Thread DIN 11851, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KCW
DN
[mm]
A
[mm]
B
[mm]
8Rd 34 × ¹⁄₈16362
15Rd 34 × ¹⁄₈16466
25Rd 52 × ¹⁄₆26606
40Rd 65 × ¹⁄₆38738
50Rd 78 × ¹⁄₆50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Thread DIN 11851, Rd 28 × ¹⁄₈", for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KAW
DN
[mm]
A
[mm]
B
[mm]
8Rd 28 × ¹⁄₈"10.00362
15Rd 28 × ¹⁄₈"10.00466
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
L
[mm]
Thread DIN 11864-1 Form A, for pipe according to DIN 11866 series A
1.4435 (316L)
Order code for "Process connection", option KGW
DN
[mm]
A
[mm]
B
[mm]
8Rd 34 × ¹⁄₈"16362
15Rd 34 × ¹⁄₈"16466
25Rd 52 × ¹⁄₆"26620
40Rd 65 × ¹⁄₆"38738
50Rd 78 × ¹⁄₆"50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Thread SMS 1145
1.4435 (316L)
Order code for "Process connection", option SAW
DN
[mm]
A
[mm]
B
[mm]
8Rd 40 × ¹⁄₆"22.50362
15Rd 40 × ¹⁄₆"22.50466
25Rd 40 × ¹⁄₆"22.50606
40Rd 60 × ¹⁄₆"35.50742
50Rd 70 × ¹⁄₆"48.50864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
L
[mm]
L
[mm]
Endress+Hauser75
Page 76
BBS thread (sterile orbital), for pipe according to DIN 11866 series A
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
1.4435 (316L)
Order code for "Process connection", option BSC
DN
[mm]
A
[mm]
B
[mm]
8M22 ×1.510.00362
15M30 × 216.00466
25M42 × 226.00606
40M52 × 238.00732
50M68 × 250.00854
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
BBS thread (sterile orbital), for pipe according to DIN 11866 series B
1.4435 (316L)
Order code for "Process connection", option BSD
DN
[mm]
A
[mm]
B
[mm]
8M26 × 1.510.00362
15M30 × 216.00466
25M42 × 226.00606
40M56 × 238.00738
50M68 × 250.00860
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
max
Proline Promass P 100
L
[mm]
L
[mm]
Thread ISO 2853
A0015623
49Engineering unit mm (in)
Thread ISO 2853, for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSD
DN
[mm]
837.1322.60370
1537.1322.60474
2537.1322.60614
4050.6535.60742
A
[mm]
B
[mm]
L
[mm]
76Endress+Hauser
Page 77
Proline Promass P 100
H
A
G
L
B
45°
C
A
Thread ISO 2853, for pipe according to ISO 2037
1.4435 (316L)
Order code for "Process connection", option JSD
DN
[mm]
5064.1048.60864
Ra
0.76 µm: order code for "Measuring tube material", option BB, TD
max
Ra
0.38 µm electropolished: order code for "Measuring tube material", option BC, TG
Order code for "Housing", option A "Compact coated aluminum"
Proline Promass P 100
A0016525
DN
[in]
A
[in]
B
[in]C[in]
D
[in]
E
[in]
1) 2)
³⁄₈5.355.813.682.137.5211.770.334.253.62
½5.355.813.682.137.5211.770.474.253.62
15.355.813.682.137.5212.280.694.763.62
1½5.355.813.682.139.0916.021.026.935.55
25.355.813.682.1310.0820.311.510.246.61
F
[in]
1) 2)
G
[in]
H
[in]
K
[in]
L
[in]
3)
3)
3)
3)
3)
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +2.76 in
2)If using a display, order code for "Display; Operation", option B: values +1.1 in
3)Dependent on the particular process connection
78Endress+Hauser
Page 79
Proline Promass P 100
G
K
DC
B
A
E
H
F
L
Order code for "Housing", option B "Compact hygienic, stainless"
A0016526
DN
[in]
A
[in]B[in]
C
[in]D[in]
E
1) 2)
[in]
³⁄₈5.265.393.072.317.3211.570.334.253.62
½5.265.393.072.317.3211.570.474.253.62
15.265.393.072.317.3212.090.694.763.62
1½5.265.393.072.318.915.831.026.935.55
25.265.393.072.319.8820.121.510.246.61
F
[in]
1) 2)
G
[in]
H
[in]
K
[in]
L
[in]
3)
3)
3)
3)
3)
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +2.76 in
2)If using a display, order code for "Display; Operation", option B: values +0.55 in
3)Dependent on the particular process connection
Endress+Hauser79
Page 80
Order code for "Housing", option C "Ultra-compact hygienic, stainless"
G
K
ADC
B
E
H
F
L
C
D
A
B
E
L
+1.5 (+0.06)
−2.0 (–0.08)
±3.5 (±0.14)
DN 125 (5")≥DN 100 (4")≤
Proline Promass P 100
A0016527
DN
[in]
A
[in]
B
[in]C[in]
D
[in]
E
[in]
1) 2)
³⁄₈4.394.872.672.27.3211.570.334.253.62
½4.394.872.672.27.3211.570.474.253.62
14.394.872.672.27.3212.090.694.763.62
1½4.394.872.672.28.915.831.026.935.55
24.394.872.672.29.8820.121.510.246.61
F
[in]
1) 2)
G
[in]
H
[in]
K
[in]
L
[in]
3)
3)
3)
3)
3)
1)If using an extension neck for the extended temperature range, order code for "Sensor option", option CG
and order code for "Measuring tube material", option TD, TG: values +2.76 in
2)If using a display, order code for "Display; Operation", option B: values +0.55 in
3)Dependent on the particular process connection
Flange connections
Fixed flange ASME B16.5
80Endress+Hauser
50Engineering unit mm (in)
A0015621
Page 81
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
C
D
A
B
E
X
X
Flange according to ASME B16.5: Cl 150
1.4404 (316/316L)
Order code for "Process connection", option AAW
DN
[in]
³⁄₈
1)
A
[in]
B
[in]
C
[in]
D
[in]
E
[in]
3.542.374 × Ø0.620.440.6213.23
½3.542.374 × Ø0.620.440.6217.32
14.333.134 × Ø0.620.561.0522.83
1½4.923.874 × Ø0.620.691.6127.83
25.914.754 × Ø0.750.752.0732.6
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)DN ³⁄₈" with DN ½" flanges as standard
Neumo BioConnect fixed flange
L
[in]
A0016907
51Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Neumo BioConnect flange; for pipe according to DIN 11866 series A, flange form R
1.4435 (316L)
Order code for "Process connection", option BSB
DN
[in]
A
[in]
B
[in]
C
[in]
D
[in]
E
[in]
L
[in]
³⁄₈2.561.774 × ⌀0.350.390.3915.12
½2.952.174 × ⌀0.350.390.6319.21
13.352.564 × ⌀0.350.471.0224.65
1½3.943.154 × ⌀0.350.471.529.65
24.333.544 × ⌀0.350.551.9734.53
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Endress+Hauser81
Page 82
Clamp connections
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
≥(1")DN 25DN 8, 15 (¾", ½")
A
B
Tri-Clamp
52Engineering unit mm (in)
Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FHW
DN
[in]
³⁄₈½0.980.3714.25
½¾0.980.6218.35
111.980.8723.86
1½1 ½1.981.3728.78
222.521.8733.58
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Clamp
[in]
A
[in]
B
[in]
Proline Promass P 100
A0023342
L
[in]
½" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FCW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
L
[in]
½½0.980.3718.35
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
¾" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FFW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
L
[in]
³⁄₈¾0.980.6214.25
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
82Endress+Hauser
Page 83
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
X
A
B
X
1" Tri-Clamp; for pipe according to DIN 11866 series C
1.4435 (316L)
Order code for "Process connection", option FPW
DN
[in]
Clamp
[in]
A
[in]
B
[in]
³⁄₈11.980.8714.25
½11.980.8718.35
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Tri-Clamp eccentric; for pipe according to DIN 11866 series C
1.4435 (316L)
DN
[in]
Order code for
"Process connection",
Clamp
[in]
A
[in]
B
[in]
option
³⁄₈FEB½0.980.3714.25
½FED¾0.980.6218.35
1FEF11.980.8723.86
1½FEH1½1.981.3729.06
2FEK22.521.8733.86
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
L
[in]
L
[in]
Neumo BioConnect clamp connection
A0016905
53Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Clamp; Neumo BioConnect, for pipe according to DIN 11866 series A; clamp form R
1.4435 (316L)
Order code for "Process connection", option BSA
DN
[in]
A
[in]
³⁄₈0.980.3914.25
½0.980.6318.35
11.981.0223.86
1½2.521.528.82
B
[in]
L
[in]
Endress+Hauser83
Page 84
Clamp; Neumo BioConnect, for pipe according to DIN 11866 series A; clamp form R
X
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
X
1.4435 (316L)
Order code for "Process connection", option BSA
DN
[in]
A
[in]
B
[in]
23.051.9733.62
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Neumo BioConnect eccentric clamp connection
Proline Promass P 100
L
[in]
A0016906
54Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier. Engineering
unit mm (in).
Neumo BioConnect eccentric clamp connection
1.4435 (316L)
Order code for "Process connection", option BEA
DN
[in]
A
[in]
B
[in]
L
[in]
³⁄₈0.980.3914.25
½0.980.6318.35
10.981.0224.02
10.981.9743.39
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
max
Additional information on "Eccentric clamps"
84Endress+Hauser
Page 85
Proline Promass P 100
L
+1,5 (+0.06)
-2,0 (-0.08)
A
B
H
A
G
L
Cable glands
Thread SMS 1145
55Engineering unit mm (in)
Thread SMS 1145
1.4435 (316L)
Order code for "Process connection", option SAW
DN
[in]
³⁄₈Rd 40 × ¹⁄₆0.8914.25
½Rd 40 × ¹⁄₆0.8918.35
1Rd 40 × ¹⁄₆0.8923.86
1½Rd 60 × ¹⁄₆1.429.21
2Rd 70 × ¹⁄₆1.9134.02
Ra
30 µin: order code for "Measuring tube material", option BB, TD
max
Ra
15 µin electropolished: order code for "Measuring tube material", option BC, TG
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg].
Weight in US units
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [lbs].
DN
[mm]
811
1513
2519
4035
5058
Weight [kg]
DN
[in]
3/824
½29
142
1½77
2128
Weight [lbs]
86Endress+Hauser
Page 87
Proline Promass P 100
1
2
3
Safety Barrier Promass 100
49 g (1.73 ounce)
MaterialsTransmitter housing
• Order code for "Housing", option A "Compact, aluminum coated":
Aluminum, AlSi10Mg, coated
• Order code for "Housing", option B "Compact, hygienic, stainless":
Hygienic version, stainless steel 1.4301 (304)
• Order code for "Housing", option C "Ultra-compact, hygienic, stainless":
Hygienic version, stainless steel 1.4301 (304)
• Window material for optional local display (→ 89):
– For order code for "Housing", option A: glass
– For order code for "Housing", option B and C: plastic
Cable entries/cable glands
56Possible cable entries/cable glands
1Cable entry in transmitter housing with internal thread M20 x 1.5
2Cable gland M20 x 1.5
3Adapter for cable entry with internal thread G ½" or NPT ½"
Order code for "Housing", option A "Compact, coated aluminum"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Cable gland M20 × 1.5Nickel-plated brass
Adapter for cable entry with internal thread G ½"
Adapter for cable entry with internal thread NPT ½"
Order code for "Housing", option B "Compact, hygienic, stainless"
The various cable entries are suitable for hazardous and non-hazardous areas.
• Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / according to JIS B2220:
Stainless steel, 1.4404 (316/316L)
• All other process connections:
Stainless steel, 1.4435 (316L)
List of all available process connections → 88
Seals
Welded process connections without internal seals
Safety Barrier Promass 100
Housing: Polyamide
Process connections
• Fixed flange connections:
– EN 1092-1 (DIN 2501) flange
– EN 1092-1 (DIN 2512N) flange
– ASME B16.5 flange
– JIS B2220 flange
– DIN 11864-2 Form A flange, DIN11866 series A, flat flange
– BBS small flange (sterile, orbital), DIN11866 series A, grooved flange
– BBS small flange (sterile, orbital), DIN11866 series B, grooved flange
• Clamp connections
– Tri-Clamp (OD tubes), DIN 11866 series C
– DIN 11864-3 Form A Clamp, DIN11866 series A, aseptic liner
– DIN 32676 clamp, DIN11866 series A
– ISO 2852 clamp, ISO2037
– ISO 2852 clamp, DIN11866 series B
– BBS Quick-Connect (sterile, orbital), DIN11866 series A, female
– BBS Quick-Connect (sterile, orbital), DIN11866 series B, female
– Neumo BioConnect Clamp, DIN11866 series A, clamp form R
• Excentric clamp connection:
– Excen. Tri-Clamp, DIN 11866 series C
– DIN 11864-3 Form A Clamp, DIN11866 series A, aseptic liner
– DIN 32676 clamp, DIN11866 series A
– ISO 2852 clamp, DIN11866 series B
– BBS Quick-Connect (sterile, orbital), DIN11866 series A, female
– BBS Quick-Connect (sterile, orbital), DIN11866 series B, female
– Neumo BioConnect Clamp, DIN11866 series A, clamp form R
• Threaded hygienic connection:
– DIN 11851 threaded hygienic connection, DIN11866 series A
– SMS 1145 threaded hygienic connection
– ISO 2853 threaded hygienic connection, ISO2037
– DIN 11864-1 Form A threaded hygienic connection, DIN11866 series A
– BBS threaded adapter (sterile, orbital), DIN11866 series A
– BBS threaded adapter (sterile, orbital), DIN11866 series B
For information on the different materials used in the process connections → 87
Surface roughness
All data relate to parts in contact with fluid.
• Ra
= 0.76 µm (30 µin)
max
• Ra
= 0.38 µm (15 µin) electropolished
max
• Delta ferrite <1%
88Endress+Hauser
Page 89
Proline Promass P 100
Operability
Operating concept
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Quick and safe commissioning
• Individual menus for applications
• Menu guidance with brief explanations of the individual parameter functions
Reliable operation
• Operation in the following languages:
– Via "FieldCare" operating tool:
English, German, French, Spanish, Italian, Chinese, Japanese
– Via integrated Web browser (only available for device versions with HART, PROFIBUS DP,
PROFINET and EtherNet/IP):
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Bahasa (Indonesian), Vietnamese, Czech, Swedish, Korean
• Uniform operating philosophy applied to operating tools and Web browser
• If replacing the electronic module, transfer the device configuration via the plug-in memory
(HistoROM DAT) which contains the process and measuring device data and the event logbook. No
need to reconfigure.
For devices with Modbus RS485, the data recovery function is implemented without the plug-in
memory (HistoROM DAT).
• Troubleshooting measures can be called up via the operating tools and Web browser
• Diverse simulation options
• Status indicated by several light emitting diodes (LEDs) on the electronic module in the housing
compartment
Local display
Remote operationVia HART protocol
A local display is only available for device versions with the following communication protocols:
HART, PROFIBUS-DP, PROFINET, EtherNet/IP
The local display is only available with the following device order code:
Order code for "Display; Operation", option B: 4-line; lit, via communication
Display element
• 4-line liquid crystal display with 16 characters per line.
• White background lighting; switches to red in event of device errors.
• Format for displaying measured variables and status variables can be individually configured.
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F). The readability of
the display may be impaired at temperatures outside the temperature range.
This communication interface is available in device versions with a HART output.
Endress+Hauser89
Page 90
Proline Promass P 100
1235
7
46
12
3
444
57Options for remote operation via HART protocol
1Control system (e.g. PLC)
2Field Communicator 475
3Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)
4Commubox FXA195 (USB)
5Field Xpert SFX350 or SFX370
6VIATOR Bluetooth modem with connecting cable
7Transmitter
A0016948
Via PROFIBUS DP network
This communication interface is available in device versions with PROFIBUS DP.
58Options for remote operation via PROFIBUS DP network
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Measuring device
A0020903
Via Ethernet-based fieldbus
This communication interface is available in device versions with EtherNet/IP.
90Endress+Hauser
Page 91
Proline Promass P 100
4
213
555
3
12
444
A0016961
59Options for remote operation via Ethernet-based fieldbus
1Control system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Add-on Profile Level 3 for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
Via PROFINET network
This communication interface is available in device versions with PROFINET.
A0026545
60Options for remote operation via PROFINET network
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
Service interfaceVia service interface (CDI-RJ45)
This communication interface is present in the following device version:
• Order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
• Order code for "Output", option L: PROFIBUS DP
• Order code for "Output", option N: EtherNet/IP
• Order code for "Output", option R: PROFINET
Endress+Hauser91
Page 92
Proline Promass P 100
1
2
3
1
2
3
HART
A0016926
61Connection for the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
1Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
PROFIBUS DP
A0021270
62Connection for order code for "Output", option L: PROFIBUS DP
1Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
92Endress+Hauser
Page 93
Proline Promass P 100
1
2
3
1
2
3
EtherNet/IP
A0016940
63Connection for order code for "Output", option N: EtherNet/IP
1Service interface (CDI -RJ45) and EtherNet/IP interface of the measuring device with access to the integrated
Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
PROFINET
A0016940
64Connection for order code for "Output", option R: PROFINET
1Service interface (CDI -RJ45) and PROFINET interface of the measuring device with access to the integrated
Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
Via service interface (CDI)
This communication interface is present in the following device version:
Order code for "Output", option M: Modbus RS485
Endress+Hauser93
Page 94
Proline Promass P 100
1
2
3
Modbus RS485
1Service interface (CDI) of the measuring device
2Commubox FXA291
3Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
A0016925
CE mark
C-Tick symbol
Ex approval
Certificates and approvals
The measuring system is in conformity with the statutory requirements of the applicable EC
Directives. These are listed in the corresponding EC Declaration of Conformity along with the
standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex ia
Category (ATEX)Type of protection
II2GEx ia IIC T6…T1 Gb
II2GEx ia IIC T6…T1 Gb or Ex ia IIB T6…T1 Gb
II1/2G, II2DEx ia IIC T6…T1 Ga/Gb or Ex ia IIB T6…T1 Ga/Gb
Ex tb IIIC Txx °C Db
II2G, II2DEx ia IIC T6…T1 Gb or Ex ia IIB T6…T1 Gb
Ex tb IIIC Txx °C Db
Ex nA
Category (ATEX)Type of protection
II3GEx nA IIC T6…T1 Gc or Ex nA IIC T5-T1 Gc
94Endress+Hauser
Page 95
Proline Promass P 100
CSA
C
US
Currently, the following versions for use in hazardous areas are available:
IS (Ex i)
• Class I Division 1 Groups ABCD
• Class II Division 1 Groups EFG and Class III
NI (Ex nA)
Class I Division 2 Groups ABCD
Sanitary compatibility
• 3-A approval
• EHEDG-tested
HART certificationHART interface
The measuring device is certified and registered by the FieldComm Group. The measuring system
meets all the requirements of the following specifications:
• Certified according to HART 7
• The device can also be operated with certified devices of other manufacturers (interoperability)
Certification PROFIBUSPROFIBUS interface
The measuring device is certified and registered by the PROFIBUS User Organization (PNO). The
measuring system meets all the requirements of the following specifications:
• Certified in accordance with PROFIBUS PA Profile 3.02
• The device can also be operated with certified devices of other manufacturers (interoperability)
Certification PROFINETPROFINET interface
The measuring device is certified and registered by the PNO (PROFIBUS User Organization ). The
measuring system meets all the requirements of the following specifications:
• Certified according to:
– Test specification for PROFINET devices
– PROFINET Security Level 1 – Net load test
• The device can also be operated with certified devices of other manufacturers (interoperability)
EtherNet/IP certification
The measuring device is certified and registered by the ODVA (Open Device Vendor Association).
The measuring system meets all the requirements of the following specifications:
• Certified in accordance with the ODVA Conformance Test
• EtherNet/IP Performance Test
• EtherNet/IP PlugFest compliance
• The device can also be operated with certified devices of other manufacturers (interoperability)
Modbus RS485 certification
The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed
all the test procedures carried out and is certified by the “MODBUS/TCP Conformance Test
Laboratory” of the University of Michigan.
Pressure Equipment
Directive
The devices can be ordered with or without a PED approval. If a device with a PED approval is
required, this must be explicitly stated in the order. For devices with nominal diameters less than or
equal to DN 25 (1"), this is neither possible nor necessary.
• With the PED/G1/x (x = category) marking on the sensor nameplate, Endress+Hauser confirms
compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure
Equipment Directive 97/23/EC.
• Devices bearing this marking (PED) are suitable for the following types of medium:
– Media in Group 1 and 2 with a vapor pressure greater than, or smaller and equal to0.5 bar
(7.3 psi)
– Unstable gases
• Devices not bearing this marking (PED) are designed and manufactured according to good
engineering practice. They meet the requirements of Art.3 Section 3 of the Pressure Equipment
Directive 97/23/EC. The range of application is indicated in tables 6 to 9 in Annex II of the
Pressure Equipment Directive.
Endress+Hauser95
Page 96
Proline Promass P 100
Other standards and
guidelines
• EN 60529
Degrees of protection provided by enclosures (IP code)
• IEC/EN 60068-2-6
Environmental influences: Test procedure - Test Fc: vibrate (sinusoidal).
• IEC/EN 60068-2-31
Environmental influences: Test procedure - Test Ec: shocks due to rough handling, primarily for
devices.
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use general requirements
• IEC/EN 61326
Emission in accordance with Class A requirements. Electromagnetic compatibility (EMC
requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment
• NAMUR NE 32
Data retention in the event of a power failure in field and control instruments with
microprocessors
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 80
The application of the pressure equipment directive to process control devices
• NAMUR NE 105
Specifications for integrating fieldbus devices in engineering tools for field devices
• NAMUR NE 107
Self-monitoring and diagnosis of field devices
• NAMUR NE 131
Requirements for field devices for standard applications
• NAMUR NE 132
Coriolis mass meter
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com → Select your
country → Products → Select measuring technology, software or components → Select the product
(picklists: measurement method, product family etc.) → Device support (right-hand column):
Configure the selected product → The Product Configurator for the selected product opens.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Application packages
Many different application packages are available to enhance the functionality of the device. Such
packages might be needed to address safety aspects or specific application requirements.
96Endress+Hauser
Page 97
Proline Promass P 100
The application packages can be ordered with the device or subsequently from Endress+Hauser.
Detailed information on the order code in question is available from your local Endress+Hauser sales
center or on the product page of the Endress+Hauser website: www.endress.com.
Detailed information on the application packages:
• Special Documentation for the device
• Special Documentation for the device
Heartbeat Technology
Concentration
PackageDescription
Heartbeat Verification
+Monitoring
PackageDescription
Concentration
measurement and special
density
Heartbeat Monitoring
Continuously supplies data, which are characteristic of the measuring principle, to
an external condition monitoring system for the purpose of preventive
maintenance or process analysis. These data enable the operator to:
• Draw conclusions - using these data and other information - about the impact
process influences (such as corrosion, abrasion, buildup etc.) have on the
measuring performance over time.
• Schedule servicing in time.
• Monitor the process or product quality, e.g. gas pockets.
Heartbeat Verification
Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter
7.6 a) "Control of monitoring and measuring equipment".
• Functional testing in the installed state without interrupting the process.
• Traceable verification results on request, including a report.
• Simple testing process via local operation or other operating interfaces.
• Clear measuring point assessment (pass/fail) with high test coverage within the
framework of manufacturer specifications.
• Extension of calibration intervals according to operator's risk assessment.
Calculation and outputting of fluid concentrations
Many applications use density as a key measured value for monitoring quality or
controlling processes. The device measures the density of the fluid as standard and
makes this value available to the control system.
The "Special Density" application package offers high-precision density
measurement over a wide density and temperature range particularly for
applications subject to varying process conditions.
With the help of the "Concentration Measurement" application package, the
measured density is used to calculate other process parameters:
• Temperature-compensated density (reference density).
• Percentage mass of the individual substances in a two-phase fluid.
(Concentration in %).
• Fluid concentration is output with special units (°Brix, °Baumé, °API, etc.) for
standard applications.
The measured values are output via the digital and analog outputs of the device.
Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Device-specific accessoriesFor the sensor
AccessoriesDescription
Heating jacketIs used to stabilize the temperature of the fluids in the sensor.
Water, water vapor and other non-corrosive liquids are permitted for use as fluids.
If using oil as a heating medium, please consult with Endress+Hauser.
For details, see Operating Instructions BA00099D
Endress+Hauser97
Page 98
Proline Promass P 100
Communication-specific
accessories
AccessoriesDescription
Commubox FXA195
HART
Commubox FXA291Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
HART Loop Converter
HMX50
Wireless HART adapter
SWA70
Fieldgate FXA320Gateway for the remote monitoring of connected 4-20 mA measuring devices via a
Fieldgate FXA520Gateway for the remote diagnostics and remote configuration of connected HART
Field Xpert SFX350Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
For intrinsically safe HART communication with FieldCare via the USB interface.
For details, see "Technical Information" TI00404F
Common Data Interface) and the USB port of a computer or laptop.
For details, see the "Technical Information" document TI405C/07
Is used to evaluate and convert dynamic HART process variables to analog current
signals or limit values.
For details, see "Technical Information" TI00429F and Operating Instructions
BA00371F
Is used for the wireless connection of field devices.
The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
For details, see Operating Instructions BA00061S
Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00053S
measuring devices via a Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00051S
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area.
For details, see Operating Instructions BA01202S
Service-specific accessories
Field Xpert SFX370Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area and the Ex area.
For details, see Operating Instructions BA01202S
AccessoriesDescription
ApplicatorSoftware for selecting and sizing Endress+Hauser measuring devices:
• Calculation of all the necessary data for identifying the optimum flowmeter: e.g.
nominal diameter, pressure loss, accuracy or process connections.
• Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters throughout the entire life cycle of a project.
Applicator is available:
• Via the Internet: https://wapps.endress.com/applicator
• On CD-ROM for local PC installation.
W@MLife cycle management for your plant
W@M supports you with a wide range of software applications over the entire
process: from planning and procurement, to the installation, commissioning and
operation of the measuring devices. All the relevant device information, such as
the device status, spare parts and device-specific documentation, is available for
every device over the entire life cycle.
The application already contains the data of your Endress+Hauser device. Endress
+Hauser also takes care of maintaining and updating the data records.
W@M is available:
• Via the Internet: www.endress.com/lifecyclemanagement
• On CD-ROM for local PC installation.
98Endress+Hauser
Page 99
Proline Promass P 100
FieldCareFDT-based plant asset management tool from Endress+Hauser.
It can configure all smart field units in your system and helps you manage them. By
using the status information, it is also a simple but effective way of checking their
status and condition.
For details, see Operating Instructions BA00027S and BA00059S
DeviceCareTool for connecting and configuring Endress+Hauser field devices.
For details, see Innovation brochure IN01047S
Commubox FXA291Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
Common Data Interface) and the USB port of a computer or laptop.
For details, see "Technical Information" TI00405C
System components
AccessoriesDescription
Memograph M graphic
display recorder
The Memograph M graphic display recorder provides information on all relevant
measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
iTEMPThe temperature transmitters can be used in all applications and are suitable for
the measurement of gases, steam and liquids. They can be used to read in the fluid
temperature.
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
Standard documentationBrief Operating Instructions
Brief Operating Instructions containing the most important information for standard
commissioning are supplied with the device.
For details, see "Technical Information" TI00133R and Operating Instructions
BA00247R
For details, see "Fields of Activity", FA00006T
Operating Instructions
Documentation code
Measuring device
Promass P 100BA01192DBA01253DBA01059DBA01067DBA01431D