
Technical Information
Oxymax W COS71
Potentiostatic amperometric three-electrode sensor
Long-term stable sensor for trace measurement
Application
Continuous measurement of the dissolved oxygen
concentration plays an important role in monitoring boiler feed
water:
• Trace quantities
Oxygen may only be present in trace quantities. Excessive
concentrations of oxygen can result incorrosion damage in the
circuit.
• Thermal resp. chemical degassing
Boiler feed water is thermally and chemically degassed to
achieve oxygen-free water. This state must be maintained in
service under process conditions.
• Laboratory measurement vs. continuous process monitoring
Spot laboratory measurements to measure oxygen involve a
risk of error due to sampling errors. Continuous measurement
directly in the process eliminates this risk and provides
reliable information about the state of the medium at any
time.
Your benefits
• Maximum measurement accuracy:
– Longterm-stable measurement thanks to potentiostatic
amperometric three-electrode system
– Long maintenance intervals
– Intelligent sensor self monitoring
• Calibration to any transmitter or location and subsequent
installation at measuring point (in digital mode with Liquisys
M COM 223/253) as calibration data is stored in the sensor
• Membrane covered sensor, i.e.:
air
selectivity
2
– high O
– Minimum maintenance effort
– Minimum calibration effort thanks to simple calibration in
TI286C/07/en/02.07
51506697

Oxymax W COS71
Function and system design
Measuring principle The oxygen molecules diffused through the membrane are reduced to hydroxide ions (OH-) at the cathode.
Silver is oxidized to silver ions (Ag+) at the anode (this forms a silver halogenide layer).
A current flows due to the electron donation at the cathode and the electron acceptance at the anode. Under
constant conditions, this flow is proportional to the oxygen content of the medium.
This current is converted in the transmitter and indicated on the display as an oxygen concentration in mg/l,
as a saturation index in % SAT or as an oxygen partial pressure in hPa.
Potentiostatic amperometric
three-electrode system
The high-impedance, current-free reference electrode plays an important role.
The formation of a silver bromide or silver chloride coating on the anode uses up the bromide or chloride ions
dissolved in the electrolyte
In the case of conventional membrane-covered sensors working with the two-electrode system, this causes an
increase in signal drift.
This is not the case with the three-electrode system:
The change in bromide or chloride concentration is registered by the reference electrode and an internal control
circuit holds the working electrode potential constant. The advantages of this principle are significantly
increased accuracy of the signal and considerably extended calibration intervals.
Measuring system A complete measuring system comprises at least:
• Oxygen sensor
• Transmitter, e.g. Liquisys M COM223/253-WX/WS
• Special measuring cable
• Assembly, e.g. flow assembly COA260 or retractable assembly COA451
Optional:
• Junction box VS (with cable extension)
Measuring system (example)
1 Flow assembly
2 Junction box VS (optional)
3 Transmitter Liquisys M COM253
4 Oxygen sensor
a0004102
2 Endress+Hauser

Oxymax W COS71
Input
Measured variable dissolved oxygen [mg/l, ppm, % SAT or hPa]
Measuring range with Liquisys M COM223/253-WX/WS:
0.001 to 20.000 mg/l
0.0 to 200.0 % SAT
0 to 400 hPa
Wiring
Electrical connection Direct connection to the transmitter (field device)
Connect the sensor directly to the transmitter (COM253-WX/WS) by using the special measuring cable with
SXP plug.
SXP plug
a0004118
Direct connection to the transmitter (panel mounting device)
• Remove the SXP connector (transmitter side!) from the cable.
• Refer to the following table for the cable assignment and the assigned terminals for
Liquisys M COM223-WX/WS.
• Please note that the cable assignment changes depending on the sensor version (fixed cable or TOP68
connection).
Terminal
COM223
87 YE +U
0GY0V WH0V
96 PK NTC (analog) or Com. (digital) GN Communication (digital)
97 BU NTC (analog) or Com. (digital) BN Communication (digital)
88 BN –U
19 GN Alarm
18 WH Sensor signal
Sensor with fixed cable (OMK) Sensor with TOP68 connection
(CYK71)
Core Assignment Core Assignment
B
B
YE +U
Koax innen –U
B
B
Endress+Hauser 3