Technical Information
Orbisint CPS12/CPS12D/CPS13
ORP electrodes, analog and digital with Memosens technology and
reference electrode
Each with dirt-repellent PTFE diaphragm for standard applications in
process and environmental technology
Application
• Long-term monitoring and limit monitoring of processes with
stable process conditions
– Paper industry
– Plastics chemistry
– Power plants (e.g. flue gas washers)
– Incineration plants
– Food industry
– Breweries
• Water treatment
– Drinking water
– Cooling water
– Well water
With ATEX, FM1 and CSA1 approval
0
for application in hazardous areas
Your benefits
• Robust electrode requiring low maintenance thanks to large
PTFE ring diaphragm
• Application under pressures of up to 16 bar (232 psi)
• Long service life due to double junction system of metal lead
and thus long electrode poison diffusion path
• Various measuring elements for application in oxidising and
reducing media
• Three lengths: 120, 225 and 360 mm (4.72, 8.86,
14.17 inches)
Further benefits offered by Memosens technology
• Maximum process safety through contactless inductive signal
transmission
• Data safety through digital data transmission
• Easy handling due to storage of sensor-specific data
• Predictive maintenance possible thanks to registration of
sensor load data
1
approval for digital sensors pending
TI367C/07/en/05.05
51513586
Function and system design
Measuring principle Redox measurement
The redox potential is a unit of measurement for the state of equilibria between oxidising and reducing
components of a medium. Redox potential is measured similarly to the pH value. A platinum or gold electrode
is used instead of pH-sensitive membrane glass. Analog to the pH measurement, an integrated Ag/AgCl
reference system is used as a reference electrode.
General properties • Low maintenance
The dirt-repellent, sterilisable PTFE ring diaphragm of the CPS12 prevents blocking and assures long-time
stability and accuracy.
• Long service life
The double junction system of the metal lead offers better protection from electrode poisons and guarantees
a considerably longer service life.
• Durability
Depending on the ordered version, the electrode is pressure proof up to 16 bar / 232 psi and can be applied
at temperatures of up to 135 °C / 275 °F.
Important properties CPS12D Maximum process safety
The inductive and non-contacting measured value transmission of Memosens guarantees maximum process
safety and offers the following benefits:
• All problems caused by moisture are eliminated.
– The plug-in connection is free from corrosion.
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.
• The transmitter is galvanically decoupled from the medium. The result: No more need to ask about
"symmetrically high-impedance" or "unsymmetrical" or an impedance converter.
• EMC safety is guaranteed by screening measures for the digital measured value transmission.
Data safety through digital data transfer
The Memosens technology digitalizes the measured value in the sensor and transfers it to the transmitter via a
contactless connection. The result:
• An automatic error message is generated if the sensor fails or the connection between sensor and transmitter
is interrupted.
• The availability of the measuring point is dramatically increased by immediate error detection.
• The digital signals are suitable for application in hazardous areas; the integrated electronics are intrinsically
safe.
Easy handling
Sensors with Memosens technology have integrated electronics that allow for saving calibration data and
further information such as total hours of operation and operating hours at very high temperatures. When the
sensor is mounted, the calibration data are automatically transferred to the transmitter and used to calculate
the current redox potential. Storing the calibration data in the sensor allows for calibration and adjustment
away from the measuring point. The result:
• The sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather
do neither affect the calibration quality nor the operator.
• The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated
sensors.
• The transmitter does not need to be installed close to the measuring point but can be placed in the control
room.
• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive
maintenance is possible.
• The sensor history can be documented on external data carriers and evaluation programs at any time. Thus,
the current application of the sensors can be made to depend on their previous history.
Communication with the transmitter
Always connect the digital sensor to a digital transmitter with Memosens technology. Data transmission to an
analog transmitter is not possible.
Orbisint CPS12 / CPS12D / CPS13
2 Endress+Hauser
Orbisint CPS12 / CPS12D / CPS13
Data storage of CPS12D Digital sensors are able to store the following system data in the sensor.
• Manufacturing data
– Serial number
–Order code
– Date of manufacture
• Calibration data
– Calibration date
– Calibrated offset (operating mode "mV")
– % slope (operating mode "%")
– Number of calibrations
– Serial number of the transmitter used for the last calibration
• Application data
– Temperature application range
– Redox application range
– Date of first commissioning
– Operating hours
These system data can be displayed with the Mycom S or the Liquiline M CM42 transmitter.
Measuring system A complete measuring system comprises:
• CPS12 or CPS12D ORP electrode
• transmitter, e.g. Liquisys M CPM223/253 (with Memosens technology for CPS12D)
• special measuring cable, e.g. CPK9 or Memosens data cable CYK10 for CPS12D
• immersion, flow or retractable assembly, e.g. Cleanfit P CPA472
Measuring system for measurement of redox potential
1 Retractable Cleanfit P CPA472 assembly
2 CPS12 / CPS12D ORP electrode
3 CPK9 special measuring cable (for electrodes with TOP68 plug-in head) / CYK10 for digital sensors
4 Liquisys M CPM253 transmitter
a0003690
Endress+Hauser 3
Orbisint CPS12 / CPS12D / CPS13
Input
Measured variables Redox potential
Measuring range -1500 ... 1500 mV
Caution!
"
Please note the process operating conditions.
Installation
Installation instructions Do not install the electrode upside down. The inclination angle must be at least 15° from the horizontal. A
smaller inclination angle is not permitted as such an inclination results in air cushion forming. This might impair
the contact of reference and metal lead.
Caution!
"
• Make sure that the assembly’s threaded connection for the electrode is clean and well running before
installing the electrode.
• Hand tighten the electrode (3 Nm)! (Given value only applies to installation in Endress+Hauser assemblies.)
• Make sure to follow the installation instructions in the operating instructions of the used assembly.
Electrode installation; inclination angle min. 15° from the horizontal
A Permitted inclination angle
B Non-permitted inclination angle
Environment
Ambient temperature
Caution!
"
Danger of frost damage
Do not use the electrode at temperatures below -15 °C / 5 °F.
Storage temperature 0 ... 50 °C / 32 ... 122 °F
Ingress protection
IP 67: GSA plug-in head (with closed plug-in connection)
IP 68: TOP 68 plug-in head (1 m / 3.28 ft water column, 50 °C / 122 °F, 168 h)
IP 68: Memosens plug-in head (10 m / 32.8 ft water column, 25 °C /77 °F, 45 days, 1 M KCl)
15°
A
15°
B
a0003686