Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 1
Instruction Manual
IM-106-340, Rev 4.0 May 2006
Oxymitter 4000 Oxygen Transmitter
http://www.raihome.com
Page 2
Page 3
HIGHLIGHTS OF CHANGES
Effective May 31, 2006 Rev. 4.0
General Reformatted entire manual from a two column layout. Removed all references to JIS
specifications. Replaced SPS 4000 information with SPS 4001B information.
Cover Updated photo, revision number and date.
viii Removed Figure 3. Oxymitter 4000 with SPS 4000 Wiring Diagram from Quick Start
Guide.
1-2 Revised Figure 1-1 to show SPS 4001B and updated IMPS 4000.
1-5 Revised Figure 1-2 to show SPS 4001B.
1-8 Added step 11 and Figure 1-6.
1-10, 1-11 Revised Figure 1-8 and Figure 1-9 to show SPS 4001B.
1-12 Removed Components paragraph.
1-13 Removed Figure 1-9. SPS 4000.
1-15 Updated Figure 1-14. Abrasive Shield Assembly.
1-16, 1-17 Revised Specifications.
1-18, 1-19 Revised Product Matrix table.
2-1 Added two additional warnings.
2-5 Removed Figure 2-3. Oxymitter 4000 Installation (with SPS 4000).
2-12, 2-15 Revised Figure 2-9 and Figure 2-10 to show grounding locations.
2-16 Revised Install Interconnecting Cable paragraph. Removed Electrical Installation (For
Oxymitter 4000 with SPS 400).
2-18 Added SPS 4001B Connections.
3-1, 4-1 Revised Terminal Block Wiring text.
5-2 Added Reference Air information.
7-9 Added D/A Trim Procedure.
8-3 Revised Alarm Indications to include signal alarm levels.
8-23 Removed SPS 4000 Troubleshooting.
9-8, 9-9 Revised Figure 9-3 and Figure 9-4.
9-11, 9-13 Revised Figure 9-6 and Figure 9-7 with updated circuit board.
9-17 Revised Figure 9-10.
9-20 Removed SPS 4000 Maintenance and Component Replacement.
10-3, 10-4 Updated part numbers for the Cell Replacement Kit, ANSI 15’ and 18’.
10-4 Updated part numbers for the Contact and Thermocouple Replacement Assembly, 15’ and
18’.
Page 4
HIGHLIGHTS OF CHANGES (CONTINUED)
Effective May 31, 2006 Rev. 4.0 (Continued)
Page Summary
10-5 Removed Ceramic Diffuser Hub Assy. Changed part numbers 4851B89G04 and
4851B90G04 to 10 microns.
10-6 Revised Table 10-2. Removed Replacement Parts for SPS 4000 table.
11- 4 Revised Figure 11-4 to show the SPS 4001B.
11- 6 Added Figure 11-7 and explanation of the Oxybalance Display and Averaging System.
A-21 Added General Precautions for Handling and Storing High Pressure Gas Cylinders.
Page 5
Instruction Manual
IM-106-340, Rev 4.0 May 2006
Oxymitter 4000
Table of Contents
Essential Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
SECTION i Introduction
SECTION 1 Description and Specifications
Preface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
What You Need To Know. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
Can You Use the Quick Start Guide? . . . . . . . . . . . . . . . . . . . . . . . . . .vii
Quick Start Guide for Oxymitter 4000 Systems . . . . . . . . . . . . . . . . . . viii
Quick Reference Guide Manual Calibration Instructions . . . . . . . . . . . . ix
HART Communicator Fast Key Sequences . . . . . . . . . . . . . . . . . . . . . . x
Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
System Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Handling the Oxymitter 4000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
IMPS 4000 (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
SPS 4001B (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12
Model 751 Remote Powered Loop LCD Display . . . . . . . . . . . . . . . . 1-13
Probe Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Diffusion Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
SECTION 2 Installation
SECTION 3 Configuration of Oxymitter 4000 with Membrane Keypad
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Selecting Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Probe Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Remote Electronics Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Electrical Installation (with Integral Electronics) . . . . . . . . . . . . . . . . .2-10
Electrical Installation (with Remote Electronics) . . . . . . . . . . . . . . . . 2-13
Install Interconnecting Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
IMPS 4000 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
SPS 4001B Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Verify Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Terminal Block Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Oxymitter 4000 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Logic I/O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Recommended Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
TOC-1
Page 6
Oxymitter 4000
Instruction Manual
IM-106-340, Rev 4.0
May 2006
SECTION 4 Configuration of Oxymitter 4000 with LOI
SECTION 5 Startup and Operation of Oxymitter 4000 with Membrane Keypad
SECTION 6 Startup and Operation of Oxymitter 4000 with LOI
Verify installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Terminal Block Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Oxymitter 4000 Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Logic I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Recommended Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Start Up Oxymitter 4000 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Navigating the Local Operator Interface . . . . . . . . . . . . . . . . . . . . . . . 6-3
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
LOI Key Designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
LOI Menu Tree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Oxymitter 4000 Setup at the LOI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
LOI Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Oxymitter 4000 Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Remote Powered Loop LCD Display (Optional) . . . . . . . . . . . . . . . . 6-10
SECTION 7 HART/AMS
SECTION 8 Troubleshooting
SECTION 9 Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
HART Communicator Signal Line Connections. . . . . . . . . . . . . . . . . . 7-2
HART Communicator PC Connections . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Off-Line and On-Line Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Logic I/O Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
HART Communicator O
Defining a Timed Calibration via HART . . . . . . . . . . . . . . . . . . . . . . . . 7-9
D/A Trim Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Identifying and Correcting Alarm Indications . . . . . . . . . . . . . . . . . . . . 8-5
Calibration Passes, but Still Reads Incorrectly . . . . . . . . . . . . . . . . . 8-22
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Calibration with Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Semi-Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Manual Calibration with Membrane Keypad. . . . . . . . . . . . . . . . . . 9-3
Calibration with LOI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Oxymitter 4000 Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Removal and Replacement of Probe . . . . . . . . . . . . . . . . . . . . . . . 9-7
Cal Method . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
2
TOC-2
Page 7
Instruction Manual
IM-106-340, Rev 4.0 May 2006
Oxymitter 4000
SECTION 10 Replacement Parts
SECTION 11 Optional Accessories
APPENDIX A Safety Data
APPENDIX B Return of Material
Probe Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
Electronics Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
HART Handheld 375 Communicator . . . . . . . . . . . . . . . . . . . . . . . . .11-1
Asset Management Solutions (AMS) . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
By-Pass Packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
IMPS 4000 Intelligent Multiprobe Test Gas Sequencer . . . . . . . . . . .11-3
SPS 4001B Single Probe Autocalibration Sequencer . . . . . . . . . . . . 11-4
O
Calibration Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
2
Catalyst Regeneration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
OxyBalance Display and Averaging System . . . . . . . . . . . . . . . . . . . 11-6
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . . . . A-15
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
TOC-3
Page 8
Oxymitter 4000
Instruction Manual
IM-106-340, Rev 4.0
May 2006
TOC-4
Page 9
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Oxymitter Oxygen Transmitters
READ THIS PAGE BEFORE PROCEEDING!

ESSENTIAL INSTRUCTIONS

Emerson Process Management designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions
product.
• If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification.
Follow all warnings, cautions, and instructions supplied with the product.
Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount Analytical. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
prior to installing, operating, and servicing the
marked on and
to install,
http://www.processanalytic.com
The information contained in this document is subject to change without notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software within the Model 275/375 may require modification. If a software modification is required, please contact your local Rosemount Analytical Service Group or National Response Center at 1-800-433-6076 or 1-888-433-6829.
Page 10
Page 11
Instruction Manual
IM-106-340, Rev. 4.0 May 2006

Section i Introduction

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page iv
What You Need To Know . . . . . . . . . . . . . . . . . . . . . . . . . . page v
Can You Use the Quick Start Guide? . . . . . . . . . . . . . . . . page vii
Quick Start Guide for Oxymitter 4000 Systems . . . . . . . . page viii
Quick Reference Guide Manual Calibration Instructions page ix
HART Communicator Fast Key Sequences . . . . . . . . . . . page x
Oxymitter 4000
http://www.processanalytic.com
Page 12
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
May 2006

PREFACE The purpose of this manual is to provide information concerning the

components, functions, installation and maintenance of the Oxymitter 4000 Oxygen Transmitter.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS The following definitions apply to WARNINGS, CAUTIONS, and NOTES

found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

SYMBOLS

NOTE
Highlights an essential operating procedure, condition, or statement.
:
EARTH (GROUND) TERMINAL
:
PROTECTIVE CONDUCT OR TERMINAL
:
RISK OF ELECTRICAL SHOCK
:
WARNING: REFER TO INSTRUCTION MANUAL
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
iv
Page 13
Instruction Manual
IM-106-340, Rev. 4.0 May 2006

WHAT YOU NEED TO KNOW

Figure 1. Installation Options ­Oxymitter 4000 with Integral Electronics
Oxymitter 4000
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
BEFORE INSTALLING AND WIRING A ROSEMOUNT ANALYTICAL OXYMITTER 4000 OXYGEN TRANSMITTER
1. What type of installation does your system require?
Use the following drawings, Figure 1 and Figure 2, to identify which type of installation is required for your Oxymitter 4000 system.
LOGIC I/O
Cal. Gas
Ref. Air
OXYMITTER 4000
Line Voltage 4-20 mA
Cal. Gas Instr. Air (Ref. Air)
OXYMITTER 4000 WITH SPS 4001B
Line Voltage
4-20 mA
SPS
4001B
Line Voltage Cal. Gas 1 Cal. Gas 2
Instr. Air (Ref. Air)
OXYMITTER 4000 WITH REMOTE IMPS 4000 OPTION
Line Voltage
4-20 mA
LOGIC I/O
Cal. Gas
Ref. Air
IMPS
4000
Line Voltage Cal. Gas 1 Cal. Gas 2
Instr. Air (Ref. Air)
37260001
v
Page 14
Oxymitter 4000
Figure 2. Installation Options ­Oxymitter 4000 with Remote Electronics
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
OXYMITTER 4000
Line Voltage 4-20 mA
Cal. Gas Instr. Air (Ref. Air)
Calibration Gas
Reference Air
Cal. Gas 2 Cal. Gas 1
Instr. Air
Calibration Gas
Reference Air
Cal. Gas 2 Cal. Gas 1
Instr. Air
Logic I/O
SPS
4001B
Logic I/O
OXYMITTER 4000
WITH REMOTE ELECTRONICS AND SPS 4001B
Line Voltage
4-20 mA
OXYMITTER 4000
WITH REMOTE ELECTRONICS AND IMPS
Line Voltage
4-20 mA
IMPS
4000
37260049
vi
Page 15
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

CAN YOU USE THE QUICK START GUIDE?

Use this Quick Start Guide if...
1. Your system requires an Oxymitter 4000 with or without the SPS 4001B OPTION. Installation options for the Oxymitter 4000 are shown in Figure 1.
2. Your system does NOT require an IMPS 4000 OPTION installation.
3. Your system does NOT use a Remote Electronics as shown in Figure 2.
4. You are familiar with the installation requirements for the Oxymitter 4000 Oxygen Transmitter. You are familiar with the installation requirements for the Oxymitter 4000 Oxygen Transmitter with a SPS 4001B.
If you cannot use the Quick Start Guide, turn to Section 2: Installation, in this Instruction Manual.
vii
Page 16
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

QUICK START GUIDE FOR OXYMITTER 4000 SYSTEMS

Before using the Quick Start Guide, please read "WHAT YOU NEED TO KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT ANALYTICAL OXYMITTER 4000 OXYGEN TRANSMITTER" on the preceding page.
1. Install the Oxymitter 4000 in an appropriate location on the stack or duct. Refer to "Selecting Location" in Section 2: Installation, for information on selecting a location for the Oxymitter 4000.
2. If using an SPS 4001B, connect the calibration gasses to the appropriate fittings on the SPS 4001B manifold.
3. Connect reference air to the Oxymitter 4000 or SPS 4001B, as applicable.
4. If using an SPS 4001B, make the wiring connections as shown in the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual.
5. If NOT using an SPS 4001B, make the following wire connections as shown in Figure 3: line voltage, 4-20 mA, and logic I/O.
6. Verify the Oxymitter 4000 switch configuration is as desired. Refer to "Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in Section 3: Configuration of Oxymitter 4000 with Membrane Keypad, or "Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in Section 4: Configuration of Oxymitter 4000 with LOI.
7. Apply power to the Oxymitter 4000; the cell heater will turn on. Allow approximately one half hour for the cell to heat to operating temperature. Once the ramp cycle has completed and the Oxymitter 4000 is at normal operation, proceed with step 8 or 9.
8. If using an SPS 4001B, initiate a semi-automatic calibration.
9. If NOT using an SPS 4001B, perform a manual calibration. Refer to "Calibration with Keypad" or "Calibration with LOI" both in Section 9: Maintenance and Service, in this instruction manual.
viii
NOTE
If your system has a membrane keypad you can refer to the Quick Start Guide on the following pages.
Page 17
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 3. Oxymitter 4000 without SPS 4001B Wiring Diagram
AC Terminal
Cover
Terminal
Block
Oxymitter 4000
AC Line
Voltage Port
(85 to 264 VAC)

QUICK REFERENCE GUIDE MANUAL CALIBRATION INSTRUCTIONS

Line Voltage
Logic I/O
4-20 mA
Signal
Left Side of
Oxymitter 4000
AC L1
AC N
Ground Lugs
Signal
Port
29770003
4-20
+
-
+
-
Performing a Manual Calibration with a Membrane Keypad
1. Place the control loop in manual.
2. Press the CAL key. The CAL LED will light solid.
3. Apply the first calibration gas.
4. Press the CAL key. When the unit has taken the readings using the first calibration gas, the CAL LED will flash continuously.
5. Remove the first calibration gas and apply the second calibration gas.
6. Push the CAL key. The CAL LED will light solid. When the unit has taken the readings using the second calibration gas, the CAL LED will flash a two-pattern flash or a three-pattern flash. A two-pattern flash equals a valid calibration, three-pattern flash equals an invalid calibration.
7. Remove the second calibration gas and cap off the calibration gas port.
8. Press the CAL key. The CAL LED will be lit solid as the unit purges. When the purge is complete, the CAL LED will turn off.
9. If the calibration was valid, the DIAGNOSTIC ALARMS LEDs indicate normal operation. If the new calibration values are not within the parameters, the DIAGNOSTIC ALARMS LEDs will indicate an alarm.
10. Place the control loop in automatic.
ix
Page 18
Oxymitter 4000

HART COMMUNICATOR FAST KEY SEQUENCES

Perform Calibration
2 3 1 1
Instruction Manual
IM-106-340, Rev. 4.0
O Upper Range Value
2
3 2 1
May 2006
Trim Analog Output
2 4
Toggle Analog Output Tracking
2 3 1 2
Analog Output Lower Range Value
2 2
3
View O Value
2
1 1 1
View Analog Output
1 2 1
Technical Support Hotline:
For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-6076 1-440-914-1261
In addition to the CSC, you may also contact Field Watch. Field Watch coordinates Emerson Process Management’s field service throughout the U.S. and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768)
Rosemount Analytical may also be reached via the Internet through e-mail and the World Wide Web:
e-mail: GAS.CSC@.emersonprocess.com
38330001
World Wide Web: www.raihome.com
x
Page 19
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

Section 1 Description and Specifications

Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
IMPS 4000 (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-12
SPS 4001B (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-12
Model 751 Remote Powered Loop LCD Display . . . . . . . page 1-13
Probe Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-13
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-16

COMPONENT CHECKLIST

A typical Rosemount Analytical Oxymitter 4000 Oxygen Transmitter should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your system in the table located on the first page of this manual.
Also, use the product matrix in Table 1-1 at the end of this section to compare your order number against your unit. The first part of the matrix defines the model. The last part defines the various options and features of the Oxymitter
4000. Ensure the features and options specified by your order number are on or included with the unit.

SYSTEM OVERVIEW

Scope This Instruction Manual is designed to supply details needed to install, start

up, operate, and maintain the Oxymitter 4000. Signal conditioning electronics outputs a 4-20 mA signal representing an O keypad or fully functional Local Operator Interface (optional) for setup, calibration, and diagnostics. This same information, plus additional details, can be accessed with the HART Model 275/375 handheld communicator or Asset Management Solutions (AMS) software.
value and provides a membrane
2
http://www.processanalytic.com
Page 20
Oxymitter 4000
Analytical
Figure 1-1. Typical System Package
8
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
1
I n
s t r
u c
I
t
M
i
o
-
n
1
0
M
6
-
D
a
3
n
e
4
u
c
0
e
a
C
m
l
R
b
e
e
v
r
.
2
4
0
.
0
2
5
O
X
Y
M
H
IT
A
Z
TE
A
R
O
D
R
X
O
Y
40
U
G
S
E
0
N
A
0
R
T
E
R
A
A
N S
M
I T
T
E
R
2
A
n
a
l y
t
i c
a
l
Analytical
7
MAN4275A00
English
HART
Communicator
o
5
6
1. Instruction Manual
2. IMPS 4000 Intelligent Multiprobe Test Gas Sequencer (Optional)
3. Oxymitter 4000 with Integral Electronics
4. SPS 4001B Single Probe Autocalibration Sequencer (Optional) (Shown with reference air option)
5. HART® 275/375 Communicator Package (Optional)
6. Adapter Plate with Mounting Hardware and Gasket
7. Remote Electronics and Cable (Optional)
8. Reference Air Set (used if SPS 4001B without reference air option or IMPS 4000 supplied)
3
4
Analytical
37260002
1-2
Page 21
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

System Description The Oxymitter 4000 is designed to measure the net concentration of oxygen

in an industrial combustion processes process; i.e., the oxygen remaining after all fuels have been oxidized. The probe is permanently positioned within an exhaust duct or stack and performs its task without the use of a sampling system.
The equipment measures oxygen percentage by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
EMF = KT log10(P1/P2) + C
Where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
2. P1 is the partial pressure of the oxygen in the reference air on the opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry, instrument air (20.95% oxygen) as the reference air.
When the cell is at operating temperature and there are unequal oxygen concentrations across the cell, oxygen ions will travel from the high oxygen partial pressure side to the low oxygen partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade. The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the Oxymitter 4000 to provide exceptional sensitivity at low oxygen concentrations.
The Oxymitter 4000 measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
1-3
Page 22
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
May 2006

System Configuration Oxymitter 4000 units are available in seven length options, giving the user the

flexibility to use an in situ penetration appropriate to the size of the stack or duct. The options on length are 18 in. (457 mm), 3 ft (0,91 m), 6 ft (1,83 m), 9 ft (2,7 m), 12 ft (3,66 m), 15 ft (4,57 m), and 18 ft (5,49 m).
The electronics control probe temperature and provide an isolated output, 4-20 mA, that is proportional to the measured oxygen concentration. The power supply can accept voltages of 90-250 VAC and 48/62 Hz; therefore, no setup procedures for power are required. The oxygen sensing cell is main­tained at a constant temperature by modulating the duty cycle of the probe heater portion of the electronics. The electronics accepts millivolt signals gen­erated by the sensing cell and produces the outputs to be used by remotely connected user devices. The output is an isolated 4-20 mA linearized current.
The Oxymitter 4000 transmitter is available with an integral or remote elec­tronics package. Two calibration gas sequencers are available: the IMPS 4000 and the SPS 4001B (Figure 1-2).
Systems with multiprobe applications may employ an optional IMPS 4000 Intelligent Multiprobe Test Gas Sequencer. The IMPS 4000 provides auto­matic calibration gas sequencing for up to four Oxymitter 4000 units and accommodates autocalibrations based on the CALIBRATION RECOM­MENDED signal from the Oxymitter 4000, a timed interval set up in HART or the IMPS 4000, or whenever a calibration request is initiated.
For systems with one or two Oxymitter 4000 units per combustion process, an optional SPS 4001B Single Probe Autocalibration Sequencer can be used with each Oxymitter 4000 to provide automatic calibration gas sequencing. The SPS 4001B is fully enclosed in a NEMA cabinet suited for wall-mounting. The sequencer performs autocalibrations based on the CALIBRATION REC­OMMENDED signal from the Oxymitter 4000, a timed interval set up in HART, or whenever a calibration request is initiated.

System Features 1. The CALIBRATION RECOMMENDED feature detects when the sensing

cell is likely out of limits. This may eliminate the need to calibrate on a "time since last cal" basis.
2. The cell output voltage and sensitivity increase as the oxygen concentration decreases.
1-4
Page 23
Instruction Manual
Analytical
IM-106-340, Rev. 4.0 May 2006
Figure 1-2. Oxymitter 4000 AutoCalibration System Options
Oxymitter 4000
OXYMITTER 4000
Analytical
IMPS 4000
(1 to 4 Probes)
Analytical
SPS 4001B
(1 Probe)
37260039
3. Membrane keypad, Figure 1-3, and HART communication are standard. To use the HART capability, you must have either:
a. HART Model 275/375 Communicator.
b. Asset Management Solutions (AMS) software for the PC.
1-5
Page 24
Oxymitter 4000
Figure 1-3. Membrane Keypad
HEATER T/C
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
HEATER
02 CELL
CALIBRATION
02 CELL mV + 02 CELL mv ­HEATER T/C + HEATER T/C -
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Figure 1-4. Local Operator Interface (LOI)
INC INC
HIGH
LOW
GAS
GAS
DEC DEC
CAL
TEST GAS +
PROCESS -
%02
MEMBRANE
KEYPAD
37260003
4. An optional Local Operator Interface, Figure 1-4, allows continuous O display and full interface capability.
2
1-6
37260004
5. Field replaceable cell, heater, thermocouple, and diffusion element.
6. The Oxymitter 4000 is constructed of rugged 316 L stainless steel for all wetted parts.
7. The electronics are adaptable for line voltages from 90-250 VAC; therefore, no configuration is necessary.
Page 25
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
8. The Oxymitter 4000 membrane keypad is available in five languages:
English French German Italian Spanish
9. An operator can calibrate and diagnostically troubleshoot the Oxymitter 4000 in one of four ways:
a. Membrane Keypad. The membrane keypad, housed within the right
side of the electronics housing, provides fault indication by way of flashing LEDs. Calibration can be performed from the membrane keypad.
b. LOI. The optional LOI takes the place of the membrane keypad and
allows local communication with the electronics. Refer to Section 6 for more information.
c. Optional HART Interface. The Oxymitter 4000's 4-20 mA output line
transmits an analog signal proportional to the oxygen level. The HART output is superimposed on the 4-20 mA output line. This information can be accessed through the following:
i. Rosemount Analytical Model 275/375 Handheld
Communicator - The handheld communicator requires Device Description (DD) software specific to the Oxymitter 4000. The DD software will be supplied with many Model 275/375 units but can also be programmed into existing units at most Rosemount Analytical service offices. See Section 7, HART/ AMS, for additional information.
ii. Personal Computer (PC) - The use of a personal
computer requires AMS software available from Rosemount Analytical.
iii. Selected Distributed Control Systems - The use of
distributed control systems requires input/output (I/O) hardware and AMS software which permit HART communications.
d. Optional IMPS 4000. The Programmable Logic Controller (PLC) in
the IMPS 4000 provides fault indications using flashing LEDs and LCD display messages. Refer to the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer Instruction Manual for more information.
1-7
Page 26
Oxymitter 4000
+
+
++
Figure 1-5. Model 751 LCD Display Panel
Figure 1-6. OxyBalance Display Displaying Outputs
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
%
22220059
10. The optional Rosemount Analytical 751 remote-mounted LCD display panel shown in Figure 1-5 is loop-driven by the 4-20 mA output signal representing the O
11. Optional OxyBalance Display and Averaging System. Reviews up to eight 4-20 mA signals from individual probes. Trends individual outputs, calculates four programmable averages as additional 4-20 mA outputs.
percentage.
2

Handling the Oxymitter 4000

Analytical
38370013
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to prevent possible equipment damage.
The Oxymitter 4000 is designed for industrial applications. Treat each component of the system with care to avoid physical damage. Some probe components are made from ceramics, which are susceptible to shock when mishandled.
1-8
Page 27
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Figure 1-7. Oxymitter 4000 HART Communications and AMS Application
HART
Model 275/375
Handheld
Interface
Hazardous Area
Oxymitter 4000
with Integral Electronics
2 Calibration Gas Lines
by Customer
[ ( ) max]300 ft 90 m
Line Voltage
4-20 mA Output
(Twisted Pairs)
Termination in Control Room
Asset Management Solutions

System Considerations Prior to installing your Oxymitter 4000, make sure you have all the

components necessary to make the system installation. Ensure all the components are properly integrated to make the system functional.
37260005
After verifying that you have all the components, select mounting locations and determine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considerations, convenience, and serviceability.
Figure 1-7 shows a typical system wiring.
A typical system installation for an Oxymitter 4000 with integral electronics is shown in Figure 1-8. A typical system installation for an Oxymitter 4000 with remote electronics is shown in Figure 1-9.
A source of instrument air is optional at the Oxymitter 4000 for reference air use. Since the unit is equipped with an in place calibration feature, provisions can be made to permanently connect calibration gas bottles to the Oxymitter 4000.
If the calibration gas bottles will be permanently connected, a check valve is required next to the calibration fittings on the integral electronics.
This check valve is to prevent breathing of the calibration gas line and subsequent flue gas condensation and corrosion. The check valve is in addition to the stop valve in the calibration gas kit or the solenoid valves in the IMPS 4000 or SPS 4001B.
NOTE:
The electronics is rated NEMA 4X (IP66) and is capable of operation at temperatures up to 185°F (85°C).
The optional LOI is also rated for operation at temperatures up to 185°F (85°C). The infrared keypad functionality will degrade at temperatures above 158°F (70°C).
Retain the packaging in which the Oxymitter 4000 arrived from the factory in case any components are to be shipped to another site. This packaging has been designed to protect the product.
1-9
Page 28
Oxymitter 4000
Figure 1-8. Typical System Installation - Oxymitter 4000 with Integral Electronics
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Gases
Stack
Oxymitter
4000
Line
Voltage
Logic I/O
4to20mA
Signal
STANDARD
Duct
Adapter Plate
Flowmeter
Calibration Gas
SPS 4001B* SINGLE PROBE
AUTOCALIBRATION OPTION
(WITH REFERENCE AIR OPTION)
Gases
Pressure
Regulator
Voltage
Instrument Air Supply (Reference Air)
Gases
Stack
Adapter
Plate
Oxymitter
4000
Line
IMPS 4000* MULTIPROBE
AUTOCALIBRATION
OPTION
Duct
Calibration Gas
Inst. Air Supply
Calibration Gas 2
Calibration Gas 1
1-10
Adapter Plate
Line Voltage
4to20mA
Oxymitter
4000
Signal
Stack
Calibration Gas
Logic I/O
Duct
Calibration Gas 2
Calibration Gas 1
Reference
Air
4to20mA
Signal
Inst. Air Supply
SPS 4001B
Logic I/O
Reference
Air
IMPS 4000
37260047
Page 29
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 1-9. Typical System Installation - Oxymitter 4000 with Remote Electronics
Oxymitter 4000
Oxymitter 4000
Remote
Electronics
Logic I/O
4to20mA
Signal
Gases
Stack
Line Voltage
STANDARD
Duct
Adapter Plate
Flowmeter
Calibration Gas
Instrument Air Supply
(Reference Air)
Pressure
Regulator
Adapter Plate
Gases
Stack
IMPS 4000* MULTIPROBE
AUTOCALIBRATION
OPTION
Duct
SPS 4001B* SINGLE PROBE
AUTOCALIBRATION OPTION
(WITH REFERENCE AIR OPTION)
Gases
Duct
Stack
Adapter Plate
Oxymitter
Remote
Electronics
Line Voltage
4to20mA
Signal
4000
Calibration Gas
Reference Air
Calibration Gas 1
Logic I/O
Line Voltage
Inst. Air Supply
Calibration Gas 2
Electronics
SPS 4001B
Oxymitter 4000
Remote
Line Voltage
Calibration Gas
Reference Air
Inst. Air Supply
Calibration Gas 2
Calibration Gas 1
Logic I/O
4to20mA
Signal
Line Voltage
IMPS 4000
*Note: The IMPS 4000 or SPS 4001B must
be installed in a non-hazardous, explosive-free environment.
37260006
1-11
Page 30
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
May 2006

IMPS 4000 (OPTIONAL) Information on the IMPS 4000 is available in the IMPS 4000 Intelligent

Multiprobe Test Gas Sequencer Instruction Manual.

SPS 4001B (OPTIONAL) The SPS 4001B Single Probe Autocalibration Sequencer provides the

capability of performing automatic, timed or on demand, calibrations of a single Oxymitter 4000 without sending a technician to the installation site.

Mounting The SPS 4001B is fully enclosed in a NEMA cabinet suited for wall-mounting.

This cabinet provides added protection against dust and minor impacts. The SPS 4001B consists of a manifold and a calibration gas flowmeter. The mani­fold provides electrical feedthroughs and calibration gas ports to route power and signal connections and calibration gases to and from the sequencer. In addition, the manifold houses two calibration gas solenoids that sequence the gases to the Oxymitter 4000, a pressure switch that detects low calibration gas pressure, and two PC boards. A terminal strip housed within the terminal cover provides convenient access for all user connections.
Components optional to the SPS 4001B include a reference air flowmeter and pressure regulator. The reference air flowmeter indicates the flow rate of reference air continuously flowing to the Oxymitter 4000. The reference air pressure regulator ensures the instrument air (reference air) flowing to the Oxymitter 4000 is at a constant pressure [20 psi (138 kPa)]. The regulator also has a filter to remove particulates in the reference air and a drain valve to bleed the moisture that collects in the filter bowl.
Brass fittings and Teflon tubing are standard. Stainless steel fittings and tubing are optional. Also, disposable calibration gas bottles are available as an option or can be purchased through a local supplier.

Operation The SPS 4001B works in conjunction with the Oxymitter 4000's CALIBRA-

TION RECOMMENDED feature to perform an autocalibration. This feature automatically performs a gasless calibration check every hour on the Oxymit­ter 4000. If a calibration is recommended and its contact output signal is set for "handshaking" with the sequencer, the Oxymitter 4000 sends a signal to the sequencer. The sequencer automatically performs a calibration upon receiving the signal. Thus, no human interface is required for the automatic calibration to take place. For further SPS 4001B information, refer to the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual.
1-12
Page 31
Instruction Manual
+
+
++
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

MODEL 751 REMOTE POWERED LOOP LCD DISPLAY

Figure 1-10. Model 751 Remote Powered Loop LCD Display

PROBE OPTIONS

The display (Figure 1-10) provides a simple, economical means to obtain accurate, reliable, and remote indication of important process variables. This display operates on the 4-20 mA line from the Oxymitter 4000. Refer to Model 751 remote powered loop LCD manual for calibration and wiring.
%
22220059A

Diffusion Elements Ceramic Diffusion Assembly

Figure 1-11. Ceramic Diffusion Assembly
The ceramic diffusion assembly, Figure 1-11, is the traditional design for the probe. Used for over 25 years, the ceramic diffusion assembly provides a greater filter surface area. This element is also available with a flame arrestor, and with a dust seal for use with an abrasive shield.
19280010
1-13
Page 32
Oxymitter 4000
Figure 1-12. Snubber Diffusion Assembly
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Snubber Diffusion Assembly
The snubber diffusion assembly, Figure 1-12, is satisfactory for most applications. This element is also available with a flame arrestor, and with a dust seal for use with an abrasive shield.
36210012
Figure 1-13. Hastelloy Cup-Type Diffusion Assembly
Cup-Type Diffusion Assembly
The cup-type diffusion assembly, Figure 1-13, is typically used in high temperature applications where frequent diffusion element plugging is a problem. It is available with either a 10 or 40 micron, sintered, Hastelloy element.
This element is also available with a dust seal for use with an abrasive shield.
36150001
Abrasive Shield Assembly
1-14
The abrasive shield assembly, Figure 1-14, is a stainless steel tube that surrounds the probe assembly. The shield protects against particle abrasion, provides a guide for ease of insertion, and acts as a position support, especially for longer probes. The abrasive shield assembly uses a modified diffuser and vee deflector assembly, fitted with dual dust seal packing.
Page 33
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 1-14. Abrasive Shield Assembly
A
15
3.584
3.554
90
A
Oxymitter 4000
B
o
o
.45 min
VIEW A
0.75 thru 4 pls,
eq sp on 4.75 B.C.
6.00
VIEW B
22.5
B
o
16860033
.745
Dia on a 7.50 Dia B.C. (ref)
.755
NOTE
In highly abrasive applications, rotate the shield 90 degrees at normal service intervals to present a new wear surface to the abrasive flow stream.
1-15
Page 34
Oxymitter 4000

SPECIFICATIONS

Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Oxymitter Specifications
O
Range
2
Standard 0 to 10% O2, 0 to 25% O2, 0 to 40% O2 (via HART) Accuracy ±0.75% of reading or 0.05% O System Response to Calibration
Gas
Temperature Limits
Process 32° to 1300°F (0° to 704°C) up to 2400°F (1300°C)
Electronics Housing -40° to 158°F (-40° to 70°C) ambient Electronics Package -40° to 185°F (-40° to 85°C) [Operating temperature
Local Operator Interface -40° to 158°F (-40° to 70°C), [above 158°F (70°C) the
Probe Lengths 18 in. (457 mm) 12 ft (3,66 m)
Mounting and Mounting Position Vertical or horizontal;
Materials
Probe Wetted or welded parts - 316L stainless steel (SS)
Electronics Enclosure Low-copper aluminum Calibration Manual, semi-automatic, or automatic Calibration Gas Mixtures
Recommended Calibration Gas Flow 2.5 l/m (5 scfh) Reference Air 2 scfh (1 l/m), clean, dry, instrument-quality air
Electronics NEMA 4X, IP66 with fitting and pipe on reference
Electric Noise EN 61326-1, Class A Certifications General Purpose
Initial – less than 3 seconds, T90 – less than 8 seconds
with optional accessories
of electronics inside of instrument housing, as measured by a HART communicator, Rosemount Analytical Asset Management Solutions software.]
infrared keypad will cease to function, but the Oxymitter 4000 will continue to operate properly.]
3 ft (0,91 m) 15 ft (4,57 m) 6 ft (1,83 m) 18 ft (5,49 m) 9 ft (2,74 m)
a spool piece, (P/N 3D39761G02), is available to offset transmitter housing from hot ductwork.
Non-wetted parts - 304 SS, low-copper aluminum
0.4% O2, Balance N 8% O2, Balance N
(20.95% O
exhaust port to clear dry atmosphere
), regulated to 34 kPa (5 psi)
2
2
2
, whichever is greater
2
1-16
C
US
APPROVED
Line Voltage 90-250 VAC, 48/62 Hz. No configuration necessary.
3/4 in. -14 NPT conduit port
Table continued on next page
Page 35
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Oxymitter Specifications
Signals
Analog Output/HART 4-20 mA isolated from power supply, 950 ohms
maximum load
Logic I/O Two-terminal logic contact configurable as either an
alarm output or as a bi-directional calibration handshake signal to IMPS 4000 or SPS 4001B, self-powered (+5 V) in series with 340 ohms Conduit ports — 3/4 in.-14 NPT (for analog output and logic I/O signal lines)
Power Requirements:
Probe Heater 175 W nominal
Electronics 10 W nominal
Maximum 500 W
1-17
Page 36
Oxymitter 4000
Table 1-1. Product Matrix
OXT4A Oxymitter 4000 In Situ Oxygen Transmitter
Oxygen Transmitter - Instruction Book
Code Sensing Probe Type
1 ANSI (N. American Std.) Probe with Ceramic Diffuser 2 ANSI Probe with Flame Arrestor and Ceramic Diffuser 3 ANSI Probe with Snubber Diffuser 4 DIN (European Std.) Probe with Ceramic Diffuser 5 DIN Probe with Flame Arrestor and Snubber Diffuser 6 DIN Probe with Snubber Diffuser
Code Probe Assembly
0 18 in. (457 mm) Probe 1 18 in. (457 mm) Probe with Abrasive Shield 2 3 ft (0,91 m) Probe 3 3 ft (0,91 m) Probe with Abrasive Shield 4 6 ft (1,83 m) Probe 5 6 ft (1,83 m) Probe with Abrasive Shield 6 9 ft (2,74 m) Probe 7 9 ft (2,74 m) Probe with Abrasive Shield 8 12 ft (3,66 m) Probe 9 12 ft (3,66 m) Probe with Abrasive Shield A 15 ft (4,57 m) Probe with Abrasive Shield B 18 ft (5,49 m) Probe with Abrasive Shield
Code Mounting Hardware- Stack Side
0 No Mounting Hardware ("0" must be chosen under "Mounting Hardware - Probe Side" below) 1 New Installation - Square weld plate with studs 2 Mounting to Model 218 Mounting Plate (with Model 218 Shield Removed) 3 Mounting to Existing Model 218 Support Shield 4 Mounting to Other Mounting 5 Mounting to Model 132 Adapter Plate
Code Mounting Hardware- Probe Side
0 No Mounting Hardware 1 Probe Only (ANSI) (N. American Std.) 2 New Bypass or New Abrasive Shield (ANSI) 4 Probe Only (DIN) 5 New Bypass or New Abrasive Shield (DIN)
Code Electronic Housing & Filtered Customer Termination - NEMA 4X, IP66
12 HART Integral Electronics, Transient Protected Filtered Termination, ATEX
Certification
14 Remote Electronics with Transient Protected Filtered Termination
(requires cable)
OXT4C331112 Example
(1)
(1)
(1)
(1)
(1)
(1)
(1)
(2)
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
1-18
Page 37
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Cont’d
Code Operator Interface
1 HART with Membrane Keypad - blind cover 2 HART with Membrane Keypad - window cover 3 HART with Local Operation Interface, window cover, English only
Code Language
1English 2German 3French 4 Spanish 5Italian
Oxymitter 4000
(3)
Code Termination Filtering
00 Specified as part of Electronic Housing
Code Calibration Accessories
00 No Hardware 01 Calibration Gas Flowmeter and Reference Air Set 02 Intelligent Multiprobe Sequencer (Refer to Table 1-3)
Code Electronics to Probe Cable
00 No Cable 10 20 ft (6 m) Cable 11 40 ft (12 m) Cable 12 60 ft (18 m) Cable 13 80 ft (24 m) Cable 14 100 ft (30 m) Cable 15 150 ft (45 m) Cable 16 200 ft (61 m) Cable
Cont’d 1 3 00 01 00 Example
NOTES:
(1)
Recommended uses: High velocity particulates in flue stream, installation within 11.5 ft (3,5 m) of soot blowers or heavy salt cake buildup. Applications: Pulverized coal, recovery boilers, lime kiln.
(2)
Where possible, specify ANSI or DIN designation; otherwise, provide details of the existing mounting plate as follows:
Plate with studs Bolt circle diameter, number, and arrangement of studs; stud thread; and stud height above mounting plate. Plate without studs Bolt circle diameter, number, and arrangement of holes; thread; and depth of stud mounting plate with accessories.
1-19
Page 38
Oxymitter 4000
Table 1-2. Calibration Components
Table 1-3. Intelligent Multiprobe Test Gas Sequencer Versions
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Part Number Description
1A99119G01 Two disposable calibration gas bottles - 0.4% and 8% O
nitrogen - 550 liters each* 1A99119G02 Two flow regulators for calibration gas bottles 1A99119G03 Bottle rack
Notes:
*Calibration gas bottles cannot be shipped via airfreight.
When the bottles are used with CALIBRATION RECOMMENDED features, the bottles should provide 2 to 3 years of calibrations in normal service.
Part Number Description Number of Oxymitters
3D39695G01 IMPS 1 3D39695G02 IMPS 2 3D39695G03 IMPS 3 3D39695G04 IMPS 4 3D39695G05 IMPS w/115 V Heater 1 3D39695G06 IMPS w/115 V Heater 2 3D39695G07 IMPS w/115 V Heater 3 3D39695G08 IMPS w/115 V Heater 4 3D39695G09 IMPS w/220V Heater 1 3D39695G10 IMPS w/220V Heater 2
3D39695G11 IMPS w/220V Heater 3
3D39695G12 IMPS w/220V Heater 4
, balance
2
1-20
Page 39
Instruction Manual
IM-106-340, Rev. 4.0 May 2006

Section 2 Installation

Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Installation (with Integral Electronics) . . . . . . . page 2-10
Electrical Installation (with Remote Electronics) . . . . . . . page 2-13
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-16
IMPS 4000 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-18
SPS 4001B Connections . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-18
Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" at the front of this Instruction Manual. Failure to follow safety instructions could result in serious injury or death.
Oxymitter 4000
Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
The Oxymitter 4000 (OXT4A) can be installed in general purpose areas only. Do not install the OXT4A in hazardous areas. For hazardous areas use the OXT4C.
http://www.processanalytic.com
Page 40
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
May 2006

MECHANICAL INSTALLATION

Selecting Location 1. The location of the Oxymitter 4000 in the stack or flue is most important

for maximum accuracy in the oxygen analyzing process. The Oxymitter 4000 must be positioned so the gas it measures is representative of the process. Best results are normally obtained if the Oxymitter 4000 is positioned near the center of the duct (40-60% insertion). Longer ducts may require several Oxymitter 4000 units since the O
stratification. A point too near the wall of the duct, or the inside radius of a bend, may not provide a representative sample because of the very low flow conditions. The sensing point should be selected so the process gas temperature falls within a range of 32° to 1300°F (0° to 704°C). Figure 2-1 through Figure 2-8 provide mechanical installation references. The ambient temperature of the integral electronics housing must not exceed 185°F (85°C). For higher ambient temperatures, we recommend the remote mounted electronics option.
2. Check the flue or stack for holes and air leakage. The presence of this condition will substantially affect the accuracy of the oxygen reading. Therefore, either make the necessary repairs or install the Oxymitter 4000 upstream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance access to the membrane keypad or LOI. Allow adequate clearance for removal of the Oxymitter
4000.
can vary due to
2
Do not allow the temperature of the Oxymitter 4000 electronics to exceed 185°F (85°C) or damage to the unit may result.

Probe Installation 1. Ensure all components are available to install the Oxymitter 4000. If

equipped with the optional ceramic diffusion element, ensure it is not damaged.
2. The Oxymitter 4000 may be installed intact as it is received.
NOTE
An abrasive shield is recommended for high velocity particulates in the flue stream (such as those in coal-fired boilers, kilns, and recovery boilers). Vertical and horizontal brace clamps are provided for 9 ft and 12 ft (2,75 m and 3,66 m) probes to provide mechanical support for the Oxymitter 4000. Refer to Figure 2-6.
3. Weld or bolt adapter plate (Figure 2-5) onto the duct.
4. If using the optional ceramic diffusion element, the vee deflector must be correctly oriented. Before inserting the Oxymitter 4000, check the direction of gas flow in the duct. Orient the vee deflector so the apex points upstream toward the flow (Figure 2-7). This may be done by loosening the setscrews and rotating the vee deflector to the desired position. Retighten the setscrews.
2-2
Page 41
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 2-1. Oxymitter 4000 Probe Installation
Oxymitter 4000
85.8
49.8
31.8
(808)
16
(406)
18 in.
(1265)
34
(864)
3ft
70
REF.
GAS
E
V
I
-
L
A
E
R
T
E
I
H
P
U
S
O
M
T
A
E
V
I
S
O
C
R
-
I
C
G
N
I N
R
A
W
L
P
X
N
E
H
W
-
T
H
E
G
I
T
N
I
P
-
E
E
K
Elec Conn
3/4 NPT
DIM "B"
DIM "A"
Table 2. Installation/Removal
PROBE
(2179)
(1778)
6ft
121.8
(3094)
106
(2692)
9ft
157.8
(4008)
142
(3607)
12 ft
193.8
(4923)
178
(4521)
15 ft
229.8
(5837)
214
(5436)
18 ft
Dimensions are in
inches with millimeters
in parentheses.
Note:
3535B18H02
3535B45H01
0.062 THK Gasket
ANSI
DIN
Insulate if exposed to
Ambient weather conditions
2.27 (58)
Dia Max
500 VA
5 Amps
R
TM
HART
SMART FAMILY
800-433-6076
Rosemount Analytical Inc.
Orrville, OH 44667-0901
R
TM
4-20 mA
85-264 VAC 48-62 Hz
OXYMITTER 4000
SERIAL NO.
VOLTS: WATTS:
TAG NO.
OUTPUT: LINE FUSE:
E
V
I
-
L
A
E
R
T
E
I
H
P
U
S
O
M
T
A
E
V
I
S
O
C
R
-
I
C
G
N
I N
R
A
W
L
P
X
N
E
H
W
-
T
H
E
G
I
T
N
I
P
-
E
E
K
CAL.
GAS
Cal Gas
Ref Air
Dim "B”
ANSI 1/4 (6.35) Tube
DIN 6 mm Tube
12
(305)
6.52
(39)
1.55
(166)
2.89
(73)
Cover Removal and Access
12
(305)
12.50 (318) Removal Envelope
Bottom View
6.02 (153)
4.77 (121)
With
Dim "A"
Snubber
Standard
Diffuser
Process flow must be in
this direction with respect
to deflector 3534B48G01
for probe
3.80(96)
Add to Dim “A”
Diffuser
5.14(131)
with Ceramic
Add to Dim “A”
for probe with
and Flame
Ceramic Diffuser
Arrestor
DIN
4512C19H01
ANSI
Table 1. Mounting Flange
4512C17H01
7.28
(185)
6.00
(153)
Flange
Dia
0.71
0.75
Hole
(18)
(20)
Dia
5.71
(145)
4.75
(121)
Eq Sp
(4) Holes
on BC
36920001
2-3
Page 42
Oxymitter 4000
Figure 2-2. Oxymitter 4000 Remote Electronics Installation
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
REMOTE ELECTRONICS
WITH MEMBRANE KEYPAD AND BLIND COVER
2.44
(62,0)
DIA.
2.21
(56,0)
6.48
(164,6)
8.72 (221,5)
3.33
(84,6)
REMOTE ELECTRONICS
WITH LOI AND WINDOW COVER
7.47
(189,8)
5.52
(140,2)
2.62
(66,5)
2.68 (68,1)
PIPE MOUNT
CONFIGURATION
Note:
Dimensions are in inches with millimeters in parentheses.
WALL MOUNT
CONFIGURATION
37260046
2-4
Page 43
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 2-3. Oxymitter 4000 with Abrasive Shield
12.50
(318)
Oxymitter 4000
E
V
I
-
L
A
E
R
E
H
P
S
O
M
-
T
G
A
N
I
N
E
R
V
A
I
W
S
O
-
L
P
X
E
N
T
I
U
C
R
I
C
N
E
H
W
T
H
G
I
T
I
P
-
E
E
K
CAL.
GAS
Connection
3/4 NPT Electrical
Cal Gas*
Ref Air
1/4 in. Tube
ANSI
6 mm Tube
DIN
*Add check valve in Cal Gas line
Dim "B"
Removal Envelope
7.00
Dim "A"
(178)
3.6 (91) Dia Nominal
6.02
(153)
4.77
(121)
Seal Assy
Diffuser/Dust
TABLE 4. Abrasive Shield -3D39003
32.5
Dim "B"
DIN
ANSI
Flange
50.5
(826)
9.25
(235)
9.00
(229)
Flange
Dia
86.5
(1283)
(24)
0.94
(19)
0.75
Hole
Dia
122.5
(2197)
7.48
7.50
(8) Holes
(3112)
(190)
(190)
Eq Sp
on BC
158.5
(4026)
194.5
(4940)
230.5
(5855)
1. These flat-faced flanges are manufactured to ANSI, DIN bolt patternsand
Notes:
and are not pressure rated.
2. Dimensions are in inches with millimeters in parentheses.
3.9
(99)
Snubber/Dust
Seal Assembly
(5)
0.2
Deflector Assy
31
TABLE 3.
13
Dim "A"
Installation/Removal Table
18 in.
Probe
(330)
(787)
3ft
67
(1702)
6ft
103
9ft
139
(2616)
(3531)
12 ft
175
15 ft
211
(4445)
(5359)
18 ft
36920002
2-5
Page 44
Oxymitter 4000
Figure 2-4. Oxymitter 4000 Adapter Plate Dimensions
8 threaded holes
equally spaced on
Instruction Manual
IM-106-340, Rev. 4.0
D Dia B.C.
Abrasive Shield
Flange O.D.
May 2006
4000 WITH ABRASIVE SHIELD
TABLE 6. ADAPTER PLATE* DIMENSIONS FOR OXYMITTER
ADAPTOR PLATE OUTLINE
DIN
(P/N 3535B58G06)
ANSI
(P/N 3535B58G02)
in.
(mm)
Dimensions
DIN
(P/N 4512C36G01)
9.25
9.00
"A"
7.5
(235)
(229)
(191)
3.94
4.75
"B"
(M-16 x 2)
(100)
(121)
Dia
(M-16 x 2)
0.625-11
"C"
Thread
(145)
5.708
7.48
7.50
"D"
(190)
(191)
Dia
attaching hardware.
*Part numbers for adapter plates include
o
C
22.5
A
B
A
Crosshatched area in 4
corners may be used to
provide additional holes for
field bolting of plate to
outside wall surface.
Adapter plate for 3, 6, 9,
and 12 ft. Abrasive Shield
Installations. See figure 2-3.
4 studs,
lockwashers and
nuts equally
spaced on
A
o
C Dia B.C.
45
2-6
TABLE 5. ADAPTER PLATE* DIMENSIONS FOR OXYMITTER 4000
ANSI
(P/N 4512C34G01)
in.
(mm)
Dimensions
6.00
"A"
(153)
0.625-11
"B"
Thread
4.75
"C"
(121)
Dia
attaching hardware.
*Part numbers for adapter plates include
A
C
B
(63,5)
2.500 Dia
Adapter plate
for Oxymitter 4000
Installation. See
Figure 2-1.
35830004
Note: Dimensions are
in inches with
millimeters in
parentheses.
Page 45
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 2-5. Oxymitter 4000 Adapter Plate Installation
Oxymitter 4000
INSTALLATION FOR METAL WALL
STACK OR DUCT CONSTRUCTION
0.50 13()
3.75 95()
Mounting holes shown rotated 45 out of true position
Weld or bolt mounting plate to metal wall of stack or duct; Joint must be air tight.
o
Minimum Dia hole in wall
Stack or duct metal wall
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
Bolt mounting plate to
outside wall surface
Field weld pipe
to adapter plate
Mounting Holes
shown rotated 45
out
of true position
o
Joint must
be air tight
Outside wall
surface
0.50 13()
4.50 114() O.D. Ref
Pipe 4.00 in. Sched 40 Pipe Sleeve (Not by Rosemount Analytical) length by Customer
Masonry stack wall
Weld or bolt adapter plate to metal wall of stack or duct. Joint must be air tight.
Notes:
1. Dimensions are in with millimeters
2. All masonry stack work and joints except adaptor
plate are not furnished by Rosemont Analytical.
2.50 (63,5)
Minimum Dia hole in wall
Stack or duct metal wall
inches
in parentheses.
Bolt adapter plate to
outside wall surface
Joint must
be air tight
Outside wall
surface
Field weld pipe to adapter plate
3.5 (89) O.D. Ref
Pipe 3.00 in. Sched 40 Pipe Sleeve (Not by Rosemount Analytical) length by Customer
Masonry stack wall
35830005
2-7
Page 46
Oxymitter 4000
Figure 2-6. Oxymitter 4000 Abrasive Shield Bracing Installation
o
60 Max
o
Min
30
4.12
(105)
Brace Bars (Not by Rosemount Analytical)
2.00 (51)
1.00 (25)
2 Holes - 0.625 (16) Dia for
0.50 (12) Dia Bolt
Note: Dimensions are in inches with
5.62
(143)
Instruction Manual
IM-106-340, Rev. 4.0
millimeters in parentheses.
Vertical brace clamp assy. Horizontal brace clamp assy.
(Both brace clamp assemblies are the same. Installation and location of clamp assemblies and brace bars to be done in field.)
Abrasive Shield
By Rosemount Analytical
}
May 2006
4.12
(105)
0.375 (10)
1.00 (25) Max
Note: Bracing is for vertical and horizontal Oxymitter 4000
Installation. External bracing required for 9 ft. through 18 ft. (2,75 m through 5,49 m) probes as shown above.
5.62
(143)
36.00 (914)
5. In vertical installations, ensure the system cable drops vertically from the Oxymitter 4000 and the conduit is routed below the level of the electronics housing. This drip loop minimizes the possibility that moisture will damage the electronics (Figure 2-8).
6. If the system has an abrasive shield, check the dust seal gaskets. The joints in the two gaskets must be staggered 180°. Also, make sure the gaskets are in the hub grooves as the Oxymitter 4000 slides into the 15° forcing cone in the abrasive shield.
7. Insert probe through the opening in the mounting flange and bolt the unit to the flange. When probe lengths selected are 9 to 18 ft (2.74 to
5.49 m), special brackets are supplied to provide additional support for the probe inside the flue or stack (Figure 2-6).
36920003
2-8
NOTE
If process temperatures will exceed 392°F (200°C), use anti-seize compound on stud threads to ease future removal of Oxymitter 4000. For ambient temperatures that will exceed 185°F (85°C), we recommend the remote mounted electronics option.
Page 47
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Figure 2-7. Orienting the Optional Vee Deflector
Gas Flow Direction
Vee
Deflector
Apex
Diffusion
Filter
Uninsulated stacks or ducts may cause ambient temperatures around the electronics to exceed 185°F (85°C), which may cause overheating damage to the electronics.
Element
Setscrew
Vee
Deflector
22220020

Remote Electronics Installation

8. If insulation is being removed to access the duct work for Oxymitter 4000 mounting, make sure the insulation is replaced afterward (Figure 2-8).
For an Oxymitter 4000 equipped with remote electronics, install the probe according to the instructions in "Probe Installation". Install the remote electronics unit on a stand pipe or similar structure, Figure 2-2.
2-9
Page 48
Oxymitter 4000
Figure 2-8. Installation with Drip Loop and Insulation Removal
Replace Insulation
after Installing
Oxymitter 4000
Insulation
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Line
Voltage
E
V
I
-
L
A
E
R
T
E
H
P
S
O
-
M
T
G
A
N
I
N
E
R
V
A
I
W
S
O
L
P
X
I
U
C
R
I
C
N
E
H
-
E
W
T
H
G
I
T
N
I
P
-
E
E
K
CAL.
GAS
Logic I/O, 4-20 mA Signal
Drip Loop
Adapter
Plate

ELECTRICAL INSTALLATION (WITH INTEGRAL ELECTRONICS)

Stack or Duct Metal Wall
29340005
For Oxymitter 4000 with Integral Electronics
All wiring must conform to local and national codes.
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault situation. This circuit breaker should also include a mechanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.
2-10
Page 49
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
NOTE
To maintain proper earth grounding, ensure a positive connection exists between the sensor housing, the electronics housing, and earth. The connecting ground wire must be 14 AWG minimum. Refer to Figure 2-9.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 221°F (105°C).
1. Remove cover (27).
2. Connect Line Voltage. a. Connect the line, or L1 wire to the L1 terminal and the neutral, or
L2 wire, to the N terminal (Figure 2-9). The Oxymitter 4000 automatically will configure itself for 90-250 VAC line voltage and 50/60 Hz. The power supply requires no setup.
3. Connect 4-20 mA Signal and Calibration Handshake/Logic I/O Leads. Use individual shielded twisted wire pairs. Terminate the shield only at the electronics housing.
a. 4-20 mA Signal. The 4-20 mA signal represents the O
can also operate the Model 751 Remote Powered Loop LCD Display or any other loop powered display. Superimposed on the 4-20 mA signal is HART information that is accessible through a Model 275/375 Handheld Communicator or AMS software.
b. Calibration Handshake/Logic I/O. The output can either be an alarm
or provide the handshaking to interface with an IMPS 4000 or SPS 4001B. For more information, refer to "Logic I/O" in Section 4: Configuration of Oxymitter 4000 with LOI, and either the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer Instruction Manual or the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual.
c. If autocalibration is not utilized, a common bi-directional logic contact
is provided for any of the diagnostic alarms listed in Table 8-1. The assignment of alarms which can actuate this contact can be modified to one of seven additional groupings listed in Table 3-1 and Ta bl e 4- 1 .
value and
2
The logic contact is self-powered, +5 VDC, 340 ohm series resistance. An interposing relay will be required if this contact is to be utilized to annunciate a higher voltage device, such as a light or horn, and may also be required for certain DCS input cards. A Potter & Brumfield R10S-E1Y1-J1.0K 3.2 mA DC or an equal interposing relay will be mounted where the contact wires terminate in the control/relay room.
d. Install cover (27, Figure 9-3).
2-11
Page 50
Oxymitter 4000
Figure 2-9. Electrical Installation - Oxymitter 4000 with Integral Electronics
INTEGRAL ELECTRONICS WITHOUT SPS 4001B
4-20 mA Signal
Logic I/O + Logic I/O –
4-20 mA 4-20 mA
90-250 VAC, 50-60 Hz Line Voltage Input
+ –
Ground
Line 1
Neutral
Ground
-
+
-
Ground
+
Lugs
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Calibration Handshake/ Logic I/O
Line Voltage
AC L1
AC N
Terminal Block
4-20 mA
+
Signal
-
Ground
90-250 VAC, 50-60 Hz Line Voltage Input
INTEGRAL ELECTRONICS WITH REMOTE SPS 4001B
Line 1 Neutral Ground
Calibration
Handshake
Line 1
Neutral
Ground
INTEGRAL ELECTRONICS WITH REMOTE IMPS 4000
Line 1 Neutral Ground
90-250 VAC, 50-60 Hz Line Voltage Input
90-250 VAC, 50-60 Hz Line Voltage Input
Refer to SPS 4001B Instruction Manual
2-12
Calibration Handshake
Refer to IMPS 4000
Instruction Manual
4-20 mA
+
4-20 mA
Ground
37260051
Page 51
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

ELECTRICAL INSTALLATION (WITH REMOTE ELECTRONICS)

For Oxymitter 4000 with Remote Electronics
All wiring must conform to local and national codes.
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault situation. This circuit breaker should also include a mechanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.
NOTE
To maintain proper earth grounding, ensure a positive connection exists between the sensor housing, the electronics housing, and earth. The connecting ground wire must be 14 AWG minimum. Refer to Figure 2-10.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 221°F (105°C).
1. Remove cover (27) from remote electronics.
2. Connect Line Voltage. a. Connect the line, or L1 wire to the L1 terminal and the neutral, or
L2 wire, to the N terminal (Figure 2-10). The Oxymitter 4000 automatically will configure itself for 90-250 VAC line voltage and 50/60 Hz. The power supply requires no setup.
3. Connect 4-20 mA Signal and Calibration Handshake/Logic I/O Leads (Figure 2-10). Use individual shielded twisted wire pairs. Terminate the shield only at the electronics housing.
a. 4-20 mA Signal. The 4-20 mA signal represents the O
value and
2
can also operate the Model 751 Remote Powered Loop LCD Display or any other loop powered display. Superimposed on the 4-20 mA signal is HART information that is accessible through a Model 275/375 Handheld Communicator or AMS software.
2-13
Page 52
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
b. Calibration Handshake/Logic I/O. The output can either be an alarm
or provide the handshaking to interface with an IMPS 4000 or SPS 4001B. For more information, refer to "Logic I/O" in Section 4: Configuration of Oxymitter 4000 with LOI, and either the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer Instruction Manual or the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual.
c. If autocalibration is not utilized, a common bi-directional logic contact
is provided for any of the diagnostic alarms listed in Table 8-1. The assignment of alarms which can actuate this contact can be modified to one of seven additional groupings listed in Table 4-1.
The logic contact is self-powered, +5 VDC, 340 ohm series resistance. An interposing relay will be required if this contact is to be utilized to annunciate a higher voltage device, such as a light or horn, and may also be required for certain DCS input cards. A Potter & Brumfield R10S-E1Y1-J1.0K 3.2 mA DC or an equal interposing relay will be mounted where the contact wires terminate in the control/relay room.
4. Install cover (27, Figure 9-4).
5. Install Interconnecting Cable. a. Remove cover (3) from junction box (5). Connect the electronics end
of the interconnecting cable to the "FROM PROBE" side of the terminal block (Figure 2-10).
b. Remove housing cover (27).
c. Connect the heater power leads, the thermocouple leads and the
oxygen signal leads at the terminal block. The leads are tagged for polarity.
2-14
Page 53
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 2-10. Electrical Installation - Oxymitter 4000 with Remote Electronics
Oxymitter 4000
ELECTRONICS
Interconnecting
Cable
PROBE
REMOTE
(BY FACTORY)
TO ELECTRONICS
GRN
ORN
YEL
RED
BLK
Oxygen
4-20 mA
BLK
Signal
Signal
BLU
-
-
4-20
ORN
+
+
RED
YEL
+
-
Ground
Lugs
+
-
Type K Thermocouple Signal
WHT
Heater Power
(Below Cover)
BLK
Calibration
Handshake/
Logic I/O
Line Voltage
AC L1
AC N
Terminal Block
Terminal Block
1234 56
BLU
To IMPS 4000 or SPS 4001B
if used. Refer to IMPS 4000
or
SPS 4001B
Instruction Manual
90-250 VAC, 50-60 Hz
2
T/C
O
CELL
ORN
YEL
RED
FROM PROBE
GRN GRN
Logic I/O + Logic I/O -
4-20 mA + 4-20 mA -
Ground
Line Voltage Input
HTR
WHT
GRN
BLK
Neutral
Ground
Ground Lugs
Line 1
37260052
2-15
Page 54
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

Install Interconnecting Cable

PNEUMATIC INSTALLATION

NOTE
If interconnect cable was not purchased with the Oxymitter 4000, consult the factory for the proper wire type and gauge.
1. Remove cover (27, Figure 9-4) from junction box (5). Connect the electronics end of the interconnecting cable (9) to the "FROM PROBE" side of terminal block (Figure 2-10).
2. Remove cover (27).
3. See (Figure 2-10). Connect the heater power leads, the thermocouple leads, and the oxygen signal leads of the interconnecting cable to the terminal block. The cable leads are tagged for polarity. To avoid a shock hazard, the heater power terminal cover must be installed.
4. Install covers (27, Figure 9-3 and Figure 9-4).
OXYMITTER 4000
Reference Air Package
After the Oxymitter 4000 is installed, connect the reference air set to the Oxymitter 4000. Refer to Figure 2-11.
Instrument Air (Reference Air): 10 psig (68.95 kPag) minimum, 225 psig (1551.38 kPag) maximum at 2 scfh (56.6 L/hr) maximum; less than 40 parts per million total hydrocarbons. Regulator outlet pressure should be set at 5 psi (35 kPa). Reference air can be supplied by the reference air set of the IMPS 4000 or SPS 4001B.
If using an IMPS 4000, refer to the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer Instruction Manual for the proper reference air connec­tions.
If using an SPS 4001B, refer to the SPS 4001B Single Probe Autocalibra­tion Sequencer Instruction Manual for the proper reference air connec­tions.
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero) be between 0.4% and 2.0% O than 40 parts per million. Failure to use proper gases will result in erroneous readings.
Calibration Gas
Two calibration gas concentrations are used with the Oxymitter 4000, Low Gas - 0.4% O 4000 connections.
and High Gas - 8% O2. See Figure 2-12 for the Oxymitter
2
. Do not use gases with hydrocarbon concentrations of more
2
2-16
Page 55
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 2-11. Air Set, Plant Air Connection
0.125-27 NPT Female Outlet Connection
1
2
Outlet
Oxymitter 4000
3.12 (79,25) Max
3
2.250 (57,15)
Note: Dimensions are in inches with
millimeters in parentheses.
4.81 (122,17)
Flow Set
Point Knob
0.25-18 NPT Female Inlet Connection
1.19 (30,22)
Drain Valve
(38,10)
1.50
2.0
(50,80)
2 Mounting Holes
3.19 (81,03) Lg
through Body for
0.312 (7,92) Dia Bolts
10.0
(254)
REF
Replacement Parts
1 Flowmeter 0.2-2.0 scfh 771B635H02
2 2" Pressure Gage 0-15 psig 275431-006
3 Combination Filter-Reg. 0-30 psig 4505C21G01
TM
OXYMITTER 4000 SERIAL NO. TAG NO.
85-264 VAC 48-62 Hz
4-20 mAR5 Amps
Rosemount Analytical Inc. Orrville, OH 44667-0901
800-433-6076
SMART FAMILY
To Electronics
TM
HART
500 VA
WATTS:VOLTS:
FUSE:LINEOUTPUT:
Instrument Air Supply 10-225 psig Max Pressure
0.250 or 6 mm O.D. Tubing (Supplied by Customer)
Ref Air Set
263C152G01
Schematic Hookup for Reference Air Supply on Oxymitter 4000 Probe Head.
8.50
(215,90)
Max
26170035
Figure 2-12. Oxymitter 4000 Gas Connections Calibration Gas Connections
Calibration Gas
R
RosemountAnalyticalInc. Orrville,OH 44667-0901
800-433-6076
TM
OXYMITTER4000 SERIALNO. TAGNO. VOLTS: WATTS:
85-264VAC 48-62 Hz
4-20 mA
OUTPUT: LINE FUSE:
SMARTFAMILY HART
R
TM
500VA
5Amps
Reference Air
26170025
2-17
Page 56
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

IMPS 4000 CONNECTIONS

SPS 4001B CONNECTIONS

See the IMPS 4000 Intelligent Multiprobe Sequencer Instruction Manual for wiring and pneumatic connection.
See the SPS 4001B Single Probe Autocalibration Sequencer Instruction Manual for wiring and pneumatic connection.
NOTE:
Upon completing installation, make sure that the Oxymitter 4000 is turned on and operating prior to firing up the combustion process. Damage can result from having a cold Oxymitter 4000 exposed to the process gases.
During outages, and if possible, leave all Oxymitter 4000 units running to prevent condensation and premature aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the Oxymitter 4000 units and remove them from the wash areas.
2-18
Page 57
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Section 3 Configuration of Oxymitter 4000
with Membrane Keypad
Verify Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Logic I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-5

VERIFY INSTALLATION

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.

Mechanical Installation Ensure the Oxymitter 4000 is installed correctly. See Section 2: Installation.

Terminal Block Wiring 1. Remove cover (27) to expose terminal block (25).

2. Check the terminal block wiring (Figure 3-1). Be sure the power, 4-20 mA signal, and the logic outputs are properly connected and secure. To avoid a shock hazard, the power terminal cover must be installed. For units with remote electronics, check the terminal block wiring at the probe and at the remote electronics unit.
3. Install housing cover (27, Figure 9-3 or Figure 9-4) on terminal block (25).
http://www.processanalytic.com
Page 58
Oxymitter 4000
Figure 3-1. Electronics Housing Terminals and Membrane Keypad
4-20
+
+
-
-
4-20 mA Signal
Logic I/O
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Oxymitter 4000 Electronics Housing
Rosemount Analytical Inc. Orrville, OH 44667-0901
800-433-6076
TM
OXYMITTER 4000 SERIAL NO. TAG NO.
85-264 VAC 48-62 Hz
4-20 mA
OUTPUT:
AC L1
AC N
Terminal Block
R
SMART FAMILY
TM
HART
500 VA
WATTS:VOLTS:
5 Amps
FUSE:LINE
HEATERT/C
SW2
HEATER
DIAGNOSTIC
ALARMS
CALIBRATION
CALIBRATIONRECOMMENDED
TEST
POINTS
INC INC
HIGH
LOW
GAS
GAS
DEC DEC
02 CELL
02 CELLmV + 02 CELLmv -
HEATERT/C +
HEATERT/C -
CAL
TESTGAS +
PROCESS -
%02
O N
J1
TP1
TP2
TP3
TP4
N R
RG
YEL
RED
G
O
TP5
TP6

Oxymitter 4000 Configuration

Ground Lugs
37260009
Located on the microprocessor board, the top board, are two switches that configure outputs for the Oxymitter 4000 (Figure 3-2). SW1 determines if the 4-20 mA signal is internally or externally powered. SW2 determines:
1. Oxymitter 4000 status, HART or LOCAL.
2. Oxygen range, 0 to 10% O
or 0 to 25% O2. (0 to 40% O2 is also
2
configurable only through HART/AMS.)
3. The 4-20 mA signal, at fault or power up, 3.5 mA or 21.6 mA.
Remove power from the Oxymitter 4000 before changing defaults. If defaults are changed under power, damage to the electronics package may occur.
SW1 Setting
The two settings are internally or externally powering the 4-20 mA signal. The factory setting is for the 4-20 mA signal to be internally powered.
3-2
Page 59
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
SW2 Setting
The factory sets this switch as follows:
1. Position 1 is HART/LOCAL. This switch setting controls the configura­tion of the Oxymitter 4000. The defaults cannot be changed via HART/AMS unless the switch is in the HART position. Placing SW2, position 1 in the LOCAL position forces the O
position 2. The position 1 switch must be placed in the LOCAL position or changes in SW2, position 2 will have no effect.
2. Position 2 determines the O O
or 0 to 25% O2. The factory setting is 0 to 10% O2. If necessary, the
2
O
range can be configured from 0 to 40% O2. To select values within
2
range. This can be set to either 0 to 10%
2
this range, set SW2, position 1 to HART and then enter the range via HART/AMS. Do not change SW2, position 1 to LOCAL unless you want to operate in the range specified by SW2, position 2.
Typically, the probe's sensing cell, in direct contact with the process gases, is heated to approximately 1357°F (736°C). The external temperature of the probe body may exceed 842°F (450°C). If operating conditions also contain high oxygen levels and combustible gases, the Oxymitter 4000 may self-ignite.
range to the setting of
2
3. Position 3 determines the output at startup or at an alarm. The settings are 3.5 mA or 21.6 mA. The factory setting is 3.5 mA. At startup, the current at the analog output is 3.5 mA or 21.6 mA.
4. Position 4 can be used to set the heater for 115 or 220 VAC operation. This switch is functional only when the software is set for manual voltage selection (Auto Tune = No). Otherwise, the internal electronics auto detect the input line voltage and sets the heater voltage accordingly (Auto Tune = Yes).
3-3
Page 60
Oxymitter 4000
Figure 3-2. Defaults - Oxymitter 4000 with Membrane Keypad
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
4-20 mA
is internally
powered (Default)
4-20 mA requires
external power
O Range set by HART/AMS
0to10%O/
0to25%O:
3.5 mA/21.6 mA:
SW1
DIAGNOSTIC
HART:
Local:
2
(From 0 to 40% O ) O Range set by Pos 2
2
2
O Range
2
2
When alarm exists, or on power­up, output current goes to this value
HEATER T/C
ALARMS
CALIBRATION
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
GAS
LOW
GAS
DEC DEC
2
HEATER O2 CELL
O2 CELL mV +
O2 CELL mV -
HEATER T/C +
HEATER T/C -
CAL
TEST GAS +
PROCESS -
% O2
0 to 10% O
SW2
1 2 3 4
TP1
TP2
TP3
TP4
TP5
TP6
OFF
HART
2
3.5 mA
220 V 115 V
position
(Ex-factory)
ON
J1
YEL
RED
GRN
ORG
ON
Default
Note:
The 115 V option at switch SW2 position 4 is active only when the heater voltage option is set to manual in the software (auto tune = no).
Local
0 to 25% O
21.6 mA
2
3-4
Read O
Once the cell is up to operating temperature, the O
Concentration
2
percentage can be read:
2
1. Access TP5 and TP6 next to the membrane keypad. Attach a multimeter across TP5 and TP6. The calibration and process gases can now be monitored. Pressing the INC or DEC once will cause the output to switch from the process gas to the calibration gas. Pressing INC or DEC a second time will increase or decrease the calibration gas parameter. If the keys have been inactive for one minute, the output reverts to the process gas. If the keys have been inactive for one minute, the output reverts to the process gas.
37260010
Page 61
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
When a calibration has been initiated, the value at TP5 and TP6 is the % O
seen by the cell.
2
Oxygen levels, as seen on the multimeter, are:
8.0% O
0.4% O
= 8.0 VDC
2
= 0.4 VDC
2
2. HART/AMS.
3. Model 751. The loop-driven LCD display.

LOGIC I/O This two-terminal logic contact can be configured either as a solid-state

relay-activated alarm or as a bi-directional calibration handshake signal to an IMPS 4000 or SPS 4001B. The configuration of this signal depends on the setting of the LOGIC I/O PIN MODE via HART/AMS or LOI. The ten different modes available are explained in Table 3-1.
Table 3-1. Logic I/O Configuration (as set at HART/AMS or LOI)
Mode Configuration
0 The unit is not configured for any alarm condition. 1 The unit is configured for a Unit Alarm. 2 The unit is configured for Low O 3 The unit is configured for both a Unit Alarm and Low O2. 4 The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.
5* The unit is configured for both a Unit Alarm and a High AC
Impedance/CALIBRATION RECOMMENDED.
6 The unit is configured for both a Low O
RECOMMENDED.
7 The unit is configured for a Unit Alarm, a Low O2, and a High AC
Impedance/CALIBRATION RECOMMENDED.
8** The unit is configured for a calibration handshake with IMPS 4000 or SPS 4001B.
CALIBRATION RECOMMENDED will initiate the calibration cycle.
9 The unit is configured for a calibration handshake. CALIBRATION
RECOMMENDED will not initiate the calibration cycle with the IMPS 4000 or SPS 4001B.
*The default condition for an Oxymitter 4000 without an IMPS 4000 or SPS 4001B.
**The default condition for an Oxymitter 4000 with an IMPS 4000 or SPS 4001B.
.
2
and High AC Impedance/CALIBRATION
2
Alarm
When configured as an alarm, this signal alerts you to an out-of-spec condition. The output is 5 V in series with a 340 ohm resistor. For optimum performance, Emerson Process Management recommends connecting the output to a Potter & Brumfield 3.2 mA DC relay (P/N R10S-E1Y1-J1.0K).
Of the ten modes in Table 3-1, mode 1 through mode 7 are the alarm modes. The factory default is mode 5 for Oxymitter 4000 units without an IMPS 4000 or SPS 4001B. In this mode, the output will signal when a unit alarm or a CALIBRATION RECOMMENDED indication occurs.
3-5
Page 62
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Calibration Handshake Signal
If using an optional IMPS 4000 or SPS 4001B, the logic I/O must be configured for calibration handshaking. Of the ten modes in Table 3-1, only modes 8 and 9 are configured for calibration handshaking. For an Oxymitter 4000 with an IMPS 4000 or an SPS 4001B, the factory sets the default to mode 8. In this mode, the logic I/O will be used to communicate between the Oxymitter 4000 and sequencer and to signal the sequencer when a CALIBRATION RECOMMENDATION indication occurs.

Recommended Configuration

4-20 mA Signal Upon Critical Alarm
Rosemount Analytical recommends that the factory default be utilized. The 4-20 mA signal will go to the 3.5 mA level upon any critical alarm which will cause the O the failure setting if normal operations cause O % O
(3.5 mA) level.
2
If the O
reading to be unusable. Customer can also select 21.6 mA as
2
measurement is being utilized as part of an automatic control loop,
2
readings to go below the zero
2
the loop should be placed into manual upon this failure event or other appropriate action should be taken.
Calibration
Rosemount Analytical recommends utilizing an autocalibration system, actuated by the "calibration recommended" diagnostic. New O
cells may
2
operate for more than a year, but older cells may require recalibration every few weeks as they near the end of their life. This strategy ensures that the O reading is always accurate, and eliminates many unnecessary calibrations based on calendar days or weeks since previous calibration. When utilizing the SPS 4001B or IMPS 4000, consider wiring some or all associated alarm contacts.
1. CALIBRATION INITIATE. Contact from the control room to an SPS 4001B or IMPS 4000 (one per probe) provides the ability to manually initiate a calibration at any time from the control room. Note that calibrations can also be initiated from a HART handheld communicator, from Asset Management Solutions software, or from the keypad on the Oxymitter 4000.
2. IN CALIBRATION. One contact per probe provides notification to the control room that the "calibration recommended" diagnostic has initiated an automatic calibration through the SPS 4001B or IMPS 4000. If the O signal is being utilized in an automatic control loop, this contact should be utilized to place the control loop into manual during calibration.
3. CALIBRATION FAILED. One contact per probe from an SPS 4001B or IMPS 4000 to the control room for notification that the calibration procedure failed. Grouped with this alarm is an output from a pressure switch which indicates when the calibration gas bottles are empty.
4. 4-20 mA SIGNAL DURING CALIBRATION. The 4-20 mA signal can be configured to respond normally during any calibration, or it can be configured to hold the last O
value upon the initiation of calibration. The
2
factory default is for the 4-20 mA signal to operate normally throughout calibration. Holding the last O
value may be useful if several probes are
2
being averaged for the purpose of automatic control. Unless several probes are being averaged, always place control loops that are using the O
signal into the manual mode prior to starting the calibration.
2
2
2
3-6
Page 63
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Section 4 Configuration of Oxymitter 4000
with LOI
Verify installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Logic I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4

VERIFY INSTALLATION

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.

Mechanical Installation Ensure the Oxymitter 4000 is installed correctly. See Section 2: Installation.

Terminal Block Wiring 1. Remove cover (27) to expose terminal block (25).

2. Check the terminal block wiring, Figure 4-1. Be sure the power, 4-20 mA signal, and logic outputs are properly connected and secure.
3. Install housing cover (27, Figure 9-3 or Figure 9-4) on terminal block (25).
http://www.processanalytic.com
Page 64
Oxymitter 4000
Figure 4-1. Electronics Housing Terminals with LOI
4-20
+
+
-
-
4-20 mA Signal
Logic I/O
Ground Lugs
AC N
AC L1
Terminal Block
OXYMITTER 4000 SERIAL NO. TAG NO.
85-264 VAC 48-62 Hz
4-20 mA
OUTPUT:
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Oxymitter 4000 Electronics Housing
Rosemount Analytical Inc. Orrville, OH 44667-0901
800-433-6076
TM
R
SMART FAMILY
TM
HART
500 VA
WATTS:VOLTS:
5 Amps
FUSE:LINE
LOI
37260011

Oxymitter 4000 Configuration

Located on the microprocessor board are two switches that configure outputs (Figure 4-2). To access these switches, the LOI module must be removed. SW1 determines if the 4-20 mA signal is internally or externally powered. SW2 determines:
1. Range control, HART or LOCAL.
2. Oxygen range, 0 to 10% O
or 0 to 25% O2. (0 to 40% O2 is also
2
configurable only through HART/AMS.)
3. The 4-20 mA signal, at fault or power up, 3.5 mA or 21.6 mA.
Remove power before changing defaults. If defaults are changed under power, damage to the electronics package may occur.
SW1 Setting
The two settings are internally or externally powering the 4-20 mA signal. The factory setting is for the 4-20 mA signal to be internally powered.
4-2
Page 65
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
SW2 Setting
The factory sets this switch as follows:
1. Position 1 is HART/LOCAL. This switch setting controls the configura­tion of the Oxymitter 4000. The defaults cannot be changed via HART/AMS or the LOI unless the switch is in the HART position. Placing SW2, position 1 in the LOCAL position forces the O of position 2. The position 1 switch must be in the LOCAL position or changes in SW2, position 2 will have no effect.
2. Position 2 determines the O O
or 0 to 25% O2. The factory setting is 0 to 10% O2. If necessary, the
2
O
range can be configured from 0 to 40% O2. To select values within
2
range. This can be set to either 0 to 10%
2
this range, set SW2, position 1 to HART and then enter the range via HART/AMS or the LOI. Do not change SW2, position 1 to LOCAL unless you want to operate in the range specified by SW2, position 2.
Typically, the probe's sensing cell, in direct contact with the process gases, is heated to approximately 1357°F (736°C). The external temperature of the probe body may exceed 842°F (450°C). If operating conditions also contain high oxygen levels and combustible gases, the Oxymitter 4000 may self-ignite.
range to the setting
2
3. Position 3 determines the output at startup or at an alarm. The settings are 3.5 mA or 21.6 mA. The factory setting is 3.5 mA. At startup, the current at the analog output is 3.5 mA or 21.6 mA.
4. Position 4 can be used to set the heater for 115 or 220 VAC operation. This switch is functional only when the software is set for manual voltage selection (Auto Tune = No). Otherwise, the internal electronics auto detect the input line voltage and sets the heater voltage accordingly (Auto Tune = Yes).
Read O
Once the cell is up to operating temperature, the O
Concentration
2
percentage can be read:
2
1. To access TP5 and TP6 under the LOI module (Figure 4-2), power down the Oxymitter 4000 and remove the LOI module. Attach alligator leads from a multimeter across TP5 and TP6. Install the LOI module and power up the Oxymitter 4000. Allow time for the cell to reach operating temperature. The calibration and process gases can now be monitored. When a calibration has been initiated, the value at TP5 and TP6 is the % O
seen by the cell. Oxygen levels, as seen on the
2
multimeter, are:
8.0% O
0.4% O
= 8.0 VDC
2
= 0.4 VDC
2
2. HART/AMS.
3. Model 751. The loop-driven LCD display.
4-3
Page 66
Oxymitter 4000
Figure 4-2. Defaults - Oxymitter 4000 with LOI
4-20 mA
is internally
powered (Default)
4-20 mA requires
external power
HART:
Local: 0to10%O/ 0to25%O:
3.5 mA/21.6 mA
O Range set by HART/AMS
2
(From 0 to 40% O ) O Range set by Pos 2
2
2
O Range
2
2
When alarm exists, or
2
on power up, current output goes to this value
OFF
HART
0 to 10% O
2
3.5 mA
220 V 115 V
Default
position
(Ex-factory)
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
ON
Local
0 to 25% O
21.6 mA
2
SW1
1 2 3 4
SW2
Note:
The 115 V option at switch SW2 position 4 is active only when the heater voltage option is set to manual in the software.
ORG
TP1
TP2
TP3
TP4
TP5
TP6
J1
RED
YEL
GRN

LOGIC I/O This two-terminal logic contact can be configured either as a solid-state

relay-activated alarm or as a bi-directional calibration handshake signal to an IMPS 4000 or SPS 4001B. The configuration of this signal depends on the setting of the LOGIC I/O PIN MODE via HART/AMS or LOI. The ten different modes available are explained in Table 4-1.
37260012
4-4
Alarm
When configured as an alarm, this signal alerts you to an out-of-spec condition. The output is +5 Vdc in series with a 340 ohm resistor.
For optimum performance, Rosemount Analytical recommends connecting the output to a Potter & Brumfield 3.2 mA DC relay (P/N R10S-E1Y1-J1.0K).
Page 67
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Table 4-1. Logic I/O Configuration (as set at HART/AMS or LOI)
Oxymitter 4000
Of the ten modes in Table 4-1, mode 1 through mode 7 are the alarm modes. The factory default is mode 5 for Oxymitter 4000 units without an IMPS 4000 or SPS 4001B. In this mode, the output will signal when a unit alarm or a CALIBRATION RECOMMENDED indication occurs.
Calibration Handshake Signal
If using an optional IMPS 4000 or SPS 4001B, the logic I/O must be configured for calibration handshaking. Of the ten modes in Table 4-1, only modes 8 and 9 are configured for calibration handshaking. For an Oxymitter 4000 with an IMPS 4000 or an SPS 4001B, the factory sets the default to mode 8. In this mode, the logic I/O will be used to communicate between the Oxymitter 4000 and the sequencer and to signal the sequencer when a CALIBRATION RECOMMENDED indication occurs.
Mode Configuration
0 The unit is not configured for any alarm condition. 1 The unit is configured for a Unit Alarm. 2 The unit is configured for Low O 3 The unit is configured for both a Unit Alarm and Low O2. 4 The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.
5* The unit is configured for both a Unit Alarm and a High AC
Impedance/CALIBRATION RECOMMENDED.
6 The unit is configured for both a Low O
RECOMMENDED.
7 The unit is configured for a Unit Alarm, a Low O2, and a High AC
Impedance/CALIBRATION RECOMMENDED.
8** The unit is configured for a calibration handshake with IMPS 4000 or SPS 4001B.
CALIBRATION RECOMMENDED will initiate the calibration cycle.
9 The unit is configured for a calibration handshake. CALIBRATION
RECOMMENDED will not initiate the calibration cycle with the IMPS 4000 or SPS 4001B.
*The default condition for an Oxymitter 4000 without an IMPS 4000 or SPS 4001B.
**The default condition for an Oxymitter 4000 with an IMPS 4000 or SPS 4001B.
.
2
and High AC Impedance/CALIBRATION
2
4-5
Page 68
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

Recommended Configuration

Recommended Configuration
4-20 mA Signal
The 4-20 mA signal is linear, and can be ranged by the user for O 0-40% O
. It should be noted that the 4-20 mA signal will go to one of several
2
range from
2
default values under certain conditions:
Condition Default Option
In calibration 4-20 mA signal responds
normally to cal gas
O2 over preset range
Critical alarm 4-20 mA signal defaults to
Dead Instrument 4-20 mA signal goes to
4-20 mA signal defaults to
20.5 mA
3.5 mA
zero mA
4-20 mA signal holds last value during cal. (recommended only if several O probes are being averaged together)
none
4-20 mA signal dfaults to 21.1 mA
none
2
It is very important that the control system be configured to recognize these various signal levels, and operators be briefed as to their meaning.
Calibration
Rosemount Analytical recommends utilizing an autocalibration system, actuated by the "calibration recommended" diagnostic. New O
cells may
2
operate for more than a year, but older cells may require recalibration every few weeks as they near the end of their life. This strategy ensures that the O reading is always accurate, and eliminates many unnecessary calibrations based on calendar days or weeks since previous calibration. When utilizing the SPS 4001B or IMPS 4000, consider wiring some or all associated alarm contacts.
2
1. CALIBRATION INITIATE. Contact from the control room to an SPS 4001B or IMPS 4000 (one per probe) provides the ability to manually initiate a calibration at any time from the control room. Note that calibrations can also be initiated from a HART handheld communicator, from Asset Management Solutions software, or from the keypad on the Oxymitter 4000.
2. IN CALIBRATION. One contact per probe provides notification to the control room that the "calibration recommended" diagnostic has initiated an automatic calibration through the SPS 4001B or IMPS 4000. If the O signal is being utilized in an automatic control loop, this contact should be utilized to place the control loop into manual during calibration.
3. CALIBRATION FAILED. One contact per probe from an SPS 4001B or IMPS 4000 to the control room for notification that the calibration procedure failed. Grouped with this alarm is an output from a pressure switch which indicates when the calibration gas bottles are empty.
4. 4-20 mA SIGNAL DURING CALIBRATION. The 4-20 mA signal can be configured to respond normally during any calibration, or it can be configured to hold the last O
value upon the initiation of calibration. The
2
factory default is for the 4-20 mA signal to track (operate normally) throughout calibration. Holding the last O
value may be useful if
2
several probes are being averaged for the purpose of automatic control. Unless several probes are being averaged, always place control loops that are using the O
signal into the manual mode prior to starting the
2
calibration.
2
4-6
Page 69
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Section 5 Startup and Operation of
Oxymitter 4000 with Membrane Keypad
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-2

POWER UP Startup Display

When power is applied to the probe, the cell heater turns on. It takes approximately one half hour for the cell to heat to operating temperature. This condition is indicated by the top four LEDs (DIAGNOSTIC ALARMS) on the membrane keypad (Figure 5-1). Starting with the CALIBRATION LED, the LEDs light in ascending order until all four LEDs are on. At this point, all four turn off and the cycle starts again. This ramp cycle continues until the cell is up to operating temperature.
Operating Display
The ramp cycle turns into a cycle where the diagnostic LEDs light in sequence from the top to the bottom, one at a time. After the bottom LED turns on, the sequence starts again at the top with the HEATER T/C LED (Figure 5-1).
Figure 5-1. Startup and Normal Operation
HEATER T/C
CALIBRATION
LOW GAS
HEATER O2 CELL
O2 CELL mV +
O2 CELL mv ­HEATER T/C + HEATER T/C -
CAL
TEST GAS +
PROCESS -
% O2
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
GAS
DEC DEC
TP1
TP2
TP3
TP4
TP5
TP6
SW2
HEATER T/C
HEATER
O CELL
2
ON
J1
CALIBRATION
1 2 3 4 1 2 3 4
Lighting sequence during warm-up
(Startup display)
YEL
RED
GRN
ORG
HEATER T/C
HEATER
O CELL
2
CALIBRATION
1 2 3 4 1 2 3 4
Lighting sequence during normal operation
(Operating display)
22220056
http://www.processanalytic.com
Page 70
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
Error
If there is an error condition at startup, one of the diagnostics LEDs will be blinking. Refer to Section 8: Troubleshooting, to determine the cause of the error. Clear the error, cycle power, and the operating display should return.
Keypad
The five membrane keys on the membrane keypad are only used during calibration to adjust the high and low gas and to initiate the calibration sequence (Figure 5-2).
Reference Air
Ensure reference air, if used, is set to 0.25 l/min (0.5 scfh)
May 2006

OPERATION

Overview Ensure the Oxymitter 4000 is at normal operation. The diagnostic LEDs will

display the operating cycle. All other LEDs should be off (Figure 5-1).
DIAGNOSTIC ALARM LEDs
If there is an error in the system, one of these LEDs will flash various blink codes (Section 8: Troubleshooting). In the case of multiple errors, only one will be displayed based on a priority system. Correct the problem and cycle power. The operating display will return or the next error will be displayed. The alarms are:
HEATER T/C HEATER O
CELL
2
CALIBRATION
CALIBRATION RECOMMENDED LED
Turns on when the system determines that a calibration is recommended. Further information is available in Section 9: Maintenance and Service.
TEST POINTS
Test points 1 through 6 will allow you to monitor with a multimeter: the heater thermocouple, O
1. TP1 and TP2 monitor the oxygen cell millivolt output which equates to the percentage of oxygen present.
2. TP3 and TP4 monitor the heater thermocouple.
3. TP5 and TP6 monitor the process gas or the calibration gas parameter.
CAL LED
cell millivolt, and the process O2.
2
5-2
The CAL LED is on steady or flashing during calibration. Further information is available in Section 9: Maintenance and Service.
Page 71
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Figure 5-2. Calibration Keys
Diagnostic
LEDs
Membrane
Keys
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
GAS
DEC DEC
Membrane
Keys
HEATER T/C
CALIBRATION
02 CELL mV + 02 CELL mv ­HEATER T/C + HEATER T/C -
LOW GAS
HEATER
02 CELL
CAL
TEST GAS +
PROCESS -
% 02
Membrane
Key
22220023
Keys
INC and DEC. The INC and DEC keys are used to set the values of the calibration gases. Attach a multimeter across TP5 and TP6. The calibration and process gases can now be monitored. Pressing the INC or DEC once will cause the output to switch from the process gas to the calibration gas. Pressing INC or DEC a second time will increase or decrease the calibration gas parameter. If the keys have been inactive for one minute, the output reverts to the process gas. When a calibration has been initiated, the value at TP5 and TP6 is the % O
seen by the cell.
2
Oxygen levels, as seen on the multimeter, are:
8.0% O
0.4% O
= 8.0 volts DC
2
= 0.4 volts DC
2
CAL
The CAL key can:
• Initiate a calibration.
• Sequence through calibration.
• Abort the calibration.
NOTE
Refer Section 9: Maintenance and Service, for calibration instructions.
Model 751 Remote Powered Loop LCD Display (Optional)
Refer to Remote Powered Loop LCD manual for calibration and operation.
5-3
Page 72
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
5-4
Page 73
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Section 6 Startup and Operation of
Oxymitter 4000 with LOI
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Start Up Oxymitter 4000 Calibration . . . . . . . . . . . . . . . . . page 6-3
Navigating the Local Operator Interface . . . . . . . . . . . . . page 6-3
LOI Key Designations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-4
LOI Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-4
Oxymitter 4000 Setup at the LOI . . . . . . . . . . . . . . . . . . . . page 6-6
LOI Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-9
Oxymitter 4000 Test Points . . . . . . . . . . . . . . . . . . . . . . . . page 6-10
Remote Powered Loop LCD Display (Optional) . . . . . . . . page 6-10

POWER UP Startup Display

When power is applied to the probe, the cell heater turns on. It takes approximately one half hour for the cell to heat to operating temperature. This condition is indicated by a "warm up" display on the LOI (Figure 6-1). This message will continue to display until the cell is up to operating temperature.
Operating Display
The normal operating display is the % O is shown in Figure 6-2.
Error
If there is an error condition at startup, an alarm message will be displayed. Refer to Section 8: Troubleshooting, to determine the cause of the error. Clear the error, cycle power, and the % O
LOI
The Local Operator Interface can be used to change the software and alarm settings, to adjust the high and low gas settings, and to initiate the calibration sequence. Refer to the LOI menu (Figure 6-4).
concentration. The "normal" display
2
display should return.
2
http://www.processanalytic.com
Page 74
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Figure 6-1. Startup Display
Figure 6-2. O2 Concentration Display
O2: Ø.ØØ% LK warm up 367dgC
37260035
6-2
O2: 2.59% LK normal
37260036
Page 75
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 6-3. LOI Features
Selection
Arrow
Touch
Confirmation
LED
Oxymitter 4000
Selection
Arrow
LCD
Display
Window
START UP OXYMITTER
Selection
Arrows
Refer to Section 9: Maintenance and Service, for calibration instructions.
37260054
4000 CALIBRATION

NAVIGATING THE LOCAL OPERATOR INTERFACE

Overview

The Local Operator Interface (LOI), shown in Figure 6-3, utilizes a bright blue gas-fluorescent display. Intensity is adjustable. There is an Infrared LED source and a detector for each key. The detectors can detect a finger placed above the button through the glass window. There is no need to open the instrument in bad weather in order to access the electronics.
It should be noted that the Oxymitter 4000 also utilizes HART communica­tions, permitting access to all instrument functionality anywhere the 4-20 mA signal terminates via a HART model 275/375 handheld communicator.

Lockout The Local Operator Interface (LOI) has a lockout feature that prevents nui-

sance actuation by someone brushing against the glass window, raindrops, dirt, insects, etc. This lockout mode is automatically established when no but­tons are pushed for 30 seconds (default). This countdown to lockout is con­figurable.
6-3
Page 76
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
In order to unlock the display, input a "Z" pattern. First, push the top left (gray) arrow, then the top right, followed by the bottom left and finally the bottom right. The "LK" notation in the upper right corner of the display will now dis­appear. Push the gray arrow at the top left hand corner once more to enter into the menu structure. Once one moves deeper into the menu structure, additional time is provided to the user so that the lockout initiation does not become a nuisance. This additional "revert" time is defaulted at one hour and is also user configurable.
May 2006

LOI KEY DESIGNATIONS The gray key (top left) will move one level higher in the menu structure. When

entering numbers, this key will move the cursor to the left. This key also dou­bles as an "Enter" key, once numbers are entered, and when the cursor is moved to it's left-most position. The new data entry value will appear in the top line of the LOI display once it is accepted.
The blue key (bottom left) acts as a selector when choosing from among a number of menu items. This key also will move the cursor to the right when entering numbers.
Up/Down keys (to the left side of the keypad) are used to increment up and down when selecting from a series of menu picks. They are also used for incrementing values up and down for data input.

LOI MENU TREE This LOI menu for the Oxymitter 4000 is shown in Figure 6-4. This menu tree

is specific to the Oxymitter 4000. The menu tree will assist in navigating the LOI.
Menu items in normal text display information, only. Menu Items in italics per­mit data entry. Menu items in bold text are procedures.
6-4
Page 77
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 6-4. Menu Tree for Local Operator Interface on the Oxymitter 4000 (Sheet 1 of 2)
Temperatures
Oxymitter 4000
O2 Temp _____dgC O2 Temp-MAX _____dgC Board Temp _____dgC Board Temp-MAX _____dgC
O2 2.59% LK
%
normal
NOTE Use Z pattern touch command to unlock menu.
SENSOR
DATA
CALIBRATION
(CONTINUED ON
SHEET 2)
Voltages
Output Values
Start Calib
Abort Calib
Cal Constants
Cal Status
Current Calib
Previous
Failed Calib
O2 Sensor _____mV O2 Sensor T/C _____mV Board Temp IC _____mV
O2 Analog % _____% O2 Analog mA ____mA
O2 Slope ____ mV/D O2 Constant ____ mV O2 Cell Imped ____ ohm
Pre O2 Slope _____mV/D Pre O2 Constant _______mV
Bad O2 Slope _____mV/D Bad O2 Constant _______mV
Calib Step _______
Calib Time Next O2 Cal
For this menu column, the selections in are user configurable. All other parameters are display only.
NOTE
Italics
___Sec.
___H
Idle Recommend Cal Apply Gas 1 Flow Gas 1 Read Gas 1 Done Gas 1 Apply Gas 2 Flow Gas 2 Read Gas 2 Done Gas 2 Cal Abort Stop Gas Purge
37260017
6-5
Page 78
Oxymitter 4000
Figure 6-4. Menu Tree for Local Operator Interface (LOI) on the Oxymitter 4000 (Sheet 2 of 2)
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
(CONTINUED FROM
SHEET 1)
SYSTEM
Calib Setup
Input/Output
Parameters
Software
Analog
Digital
O2 Gas 1 O2 Gas 2 O2-Reset Vals O2
Out Tracks O2 Cal Intervl O2-
Next Cal Gas Time Purge Time Auto Calib?
O2 Type O2 Range O2 Alarm Leve
Do O2 Trim
Logic IO Mode Low O2 Alarm
Calibration” in Section 9.
Input State Force Output
O2 Slope O2 Constant O2 T90 Time Auto Tune? Lockout Time Revert Time Luminance
Version xxx Checksum xxx Build Number xxx Build Date xxxxxx Test Code xx SW Err File xx SW Err Line xx
_____% _____%
Yes/No Yes/No
____H
____H ___Sec. ___Sec.
Yes/No
_______
______%
l _____mA
See Table 4-1
See “Abort
____ mV/D
____ mV
0:00
Yes/No
0:00 0:00
______
In column four of this menu, the selections in are user configurable. text selections are procedures; related instructions are displayed on the LOI. All other parameters are display only.

OXYMITTER 4000 SETUP AT THE LOI

Status
NOTE
Italics Bold
Alarms __________ PID Parameters 115/220
Reset Device?
(Cal. required after reset)
Yes/No
37260018
In setting up the Oxymitter 4000 from the LOI, it is best to start at the SYSTEM/Calibration Setup menu, Figure 6-4.
SYSTEM/Calibration Setup
O2 Gas #1 - Enter the high or low cal gas value (the order is not important).
O2 Gas #2 - Enter the second cal gas value.
NOTE
Refer to Section 9: Maintenance and Service, for calibration instructions.
6-6
Page 79
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
NOTE
Rosemount Analytical recommends 0.4% O2 and 8% O2 for calibration gases.
O2 Reset Values - Resets factory default values.
O2 Output Tracks - 4 to 20 mA signal can be held at the last value during
calibration, or the signal can be left to track the cal gases.
O2 Cal Interval - If automatic calibration is selected, this selects the interval between calibrations.
O2 Next Cal - If automatic calibration is selected, this selects the time until the first initial calibration takes place.
Gas Time - How long should each cal gas flow. Factory default is 300 seconds, but the user may want to vary this depending upon the length of calibration gas tubing runs.
Purge Time - Used if the O
output is selected to hold the last value during
2
calibration. After the second cal gas is removed, how long until the sensor comes back to the normal process reading, and the 4-20 mA signal can be released.
Auto Calib? - Select "Yes" if an SPS or IMPS autocalibration system is part of the system.
SYSTEM/Input/Output
Analog
Pertaining to the analog 4-20 mA signal representing O
.
2
O2 Type - 4-20 mA signal may be configured to increase with increasing O
or the reverse.
2
O2 Range - Upper O
range is user selectable.
2
O2 Alarm Level - User can configure the digital output to alarm at a given O
level.
2
Do O2 Trim - Procedure for calibrating the 4-20 mA signal to a precision mA source. Procedure is intuitive.
Digital
A bi-directional logic signal may be configured as an alarm, or as a calibration handshake signal.
Logic I/O Mode - One of 9 different sets of conditions can be set for the digital signal. See Table 8-2.
Low O2 Alarm - If any of the conditions noted above include a low O
2
process alarm, set the value here.
Input State - Notes the current condition of the bi-directional digital signal.
Force Output - Forces the output state of the signal to either open or
closed. This is used primarily when diagnosing potential problems with this signal.
6-7
Page 80
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
SYSTEM/Parameters
O2 Slope - O
slope is data regarding the strength of the sensing cell output.
2
This information is automatically calculated after a calibration, and the user does not normally input this data.
O2 Constant - O
constant is the amount of voltage a cell generates with
2
ambient air as the calibration gas. Again, this is normally calculated as a result of calibration, and is not normally input by the user.
O2 T90 Time - Some users may feel that the O certain processes. This feature permits the user to dampen the O
reading is too active for
2
signal. The
2
default value is zero seconds dampening.
Auto Tune - The electronics detects the line voltage powering the instrument automatically, and picks proper algorithms for heater control. User can force a high voltage algorithm, or a low, but Auto Tune is the default, and is recommended.
Lockout Time - Keypad lockout time default is 30 sec., but it is user configurable. A "Z" keypad pattern will unlock the keypad.
Revert Time - Once a user goes one level deep into the menu structure, an additional "revert time" is provided to prevent nuisance lockouts. One hour is the default, and it is user configurable.
Luminance - Gas fluorescence brightness is user adjustable.
SYSTEM/Status
Alarms - Diagnostic alarms. Section 8: Troubleshooting.
PID Parameter - Displays the line voltage, powering the Oxymitter, and infers
the temperature control algorithm being used to control heater temperature.
Reset Device - Device can be reset here as opposed to re-powering. Calibration parameters will be lost.
SYSTEM/Software
This is data regarding the Oxymitter 4000 software version, and errors that may have occurred.
SENSOR DATA
Displays information about the O
cell and thermocouple.
2
6-8
Page 81
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Temperatures
O2 Temp - Indicates the thermocouple temperature at the sensing cell; this should always be 1357°F (736°C).
O2 Temp Max - Maximum temperature the cell has seen. (Some process temperatures can exceed the 1357°F (736°C) setpoint temperature, and this will indicate this condition.)
Board Temp - The temperature inside the Oxymitter electronics housing 185°F (85°C), is the max.
Board Temp Max - This is the maximum temperature that the electronics has experienced over time.
Voltages
The raw mV signals feeding the temperature indications listed in the previous paragraph.
Output Values
Indication of the current readings for O
and mA.
2

LOI INSTALLATION The LOI module connects to the top of the electronic assembly in the

electronics housing. There are four matching connectors (Figure 6-5) on the back of the LOI module that allow the user to orient (rotate) the LOI as desired.
Figure 6-5. LOI Module Connectors
Connector
Receptacles
LOI Module
Rear View
37260055
6-9
Page 82
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

OXYMITTER 4000 TEST POINTS

REMOTE POWERED LOOP LCD DISPLAY (OPTIONAL)

Figure 6-6. Oxymitter 4000 ­Tes t P o i n ts
Refer to Figure 6-6. System test points are located on the board below the LOI module. Test points 1 through 6 allow you to monitor with a multimeter: the heater thermocouple, the O
cell millivolt, and the process O2.
2
• TP1 and TP2 monitor the oxygen cell millivolt output which equates to the percentage of oxygen present.
• TP3 and TP4 monitor the heater thermocouple.
• TP5 and TP6 monitor the process gas or the calibration gas parameter.
Refer to Remote Powered Loop LCD manual for calibration and operation.
TP1
TP2
TP3
TP4
TP5
TP6
J1
RED
YEL
GRN
ORG
37260037
6-10
Page 83
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

Section 7 HART/AMS

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-1
HART Communicator Signal Line Connections . . . . . . . page 7-2
HART Communicator PC Connections . . . . . . . . . . . . . . . page 7-2
Off-Line and On-Line Operations . . . . . . . . . . . . . . . . . . . page 7-4
Logic I/O Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-4
HART Communicator O
Defining a Timed Calibration via HART . . . . . . . . . . . . . . page 7-9
D/A Trim Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 7-9

OVERVIEW The HART Communicator is a handheld communications interface device. It

provides a common communications link to all microprocessor-based instru­ments that are HART compatible. The handheld communicator contains an 8 x 21 character liquid crystal display (LCD) and 25 keys. A pocket-sized manual, included with the HART Communicator, details the specific functions of all the keys.
Cal Method . . . . . . . . . . . . . . . . page 7-8
2
To interface with the Oxymitter 4000, the HART Communicator requires a ter­mination point along the 4-20 mA current loop and a minimum load resistance of 250 ohms between the communicator and the power supply.
The HART Communicator accomplishes its task using a frequency shift key­ing (FSK) technique. With the use of FSK, high-frequency digital communica­tion signals are superimposed on the Oxymitter 4000's 4-20 mA current loop. The HART communicator does not disturb the 4-20 mA signal, since no net energy is added to the loop.
The HART Communicator may be interfaced with a personal computer (PC), providing that special software has been installed. To connect the HART Com­municator to a PC, an interface adapter is required. Refer to the proper HART Communicator documentation in regard to the PC interface option.
http://www.processanalytic.com
Page 84
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

HART COMMUNICATOR SIGNAL LINE CONNECTIONS

The HART Communicator can connect to the Oxymitter 4000's analog output signal line at any wiring termination in the 4-20 mA current loop. There are two methods of connecting the HART Communicator to the signal line. For applications in which the signal line has a load resistance of 250 ohms or more, refer to method 1. For applications in which the signal line load resis­tance is less than 250 ohms, refer to method 2.
Method 1, For Load Resistance
Refer to Figure 7-1 and the following steps to connect the HART Communica­tor to a signal line < 250 ohms or more of load resistance.
Explosions can result in death or serious injury. Do not make connections to the HART Communicator's serial port, 4-20 mV signal line, or NiCad recharger jack in an explosive atmosphere.
Using the supplied lead set, connect the HART Communicator in parallel with to the Oxymitter 4000. Use any wiring termination points in the analog output 4-20 mA signal line.
250 Ohms
Method 2, For Load Resistance < 250 ohms
Refer to Figure 7-2 and the following steps to connect the HART Communica­tor to a signal line with < 250 ohms load resistance.

HART COMMUNICATOR PC CONNECTIONS

Explosions can result in death or serious injury. Do not make connections to the HART Communicator's serial port, 4-20 mV signal line, or NiCad recharger jack in an explosive atmosphere.
1. At a convenient point, break the analog output 4-20 mA signal line and install the optional 250 ohm load resistor.
2. Plug the load resistor into the loop connectors (located on the rear panel of the HART Communicator).
There is an option to interface the HART Communicator with a personal com­puter. Load the designated AMS software into the PC. Then link the HART Communicator to the PC using the interface PC adapter that connects to the serial port (on the communicator rear panel).
Refer to the proper HART Communicator documentation in regard to the PC interface option.
7-2
Page 85
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 7-1. Signal Line Connections, 250 Ohms Load Resistance
AC L1
AC N
+
4-20
-
+
-
4-20 mA Signal Line
Oxymitter 4000
RL 250
Terminal Block
Loop Connectors
SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS
LOOP CONNECTORS
USE INTERFACE 00275 0013 ONLY
SERIAL PORT
HART
Communicator
HART Communicator
Rear Panel
Figure 7-2. Signal Line Connections, < 250 Ohms Load Resistance
AC L1
AC N
+
4-20
-
+
-
4-20 mA Signal Line
Analog Output Device
Lead Set
23230001
RL < 250
Terminal Block
Loop Connectors
SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS
LOOP CONNECTORS
HART Communicator
Rear Panel
USE INTERFACE 00275 0013 ONLY
SERIAL PORT
HART
Communicator
Analog Output Device
250 Ohm Load
Resistor
(Note)
Note: The signal loop must be broken
to insert the optional 250 Ohm load resistor.
23230002
7-3
Page 86
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

OFF-LINE AND ON-LINE OPERATIONS

LOGIC I/O CONFIGURATIONS

The HART Communicator can be operated both off-line and on-line.
Off-line operations are those in which the communicator is not connected to the Oxymitter 4000. Off-line operations can include interfacing the HART Communicator with a PC (refer to applicable HART documentation regarding HART/PC applications.
In the on-line mode, the communicator is connected to the 4-20 mA analog output signal line. The communicator is connected in parallel to the Oxymitter 4000 or in parallel to the 250 ohm load resistor.
NOTE
If the HART Communicator is turned on while connected to the 4-20 mA analog output signal line, an undefined status indication appears while the communicator warms up. Wait until the warm-up period ends to continue.
The opening menu displayed on the LCD is different for on-line and off-line operations. When powering up a disconnected (off-line) communicator, the LCD will display the Main Menu. When powering up a connected (on-line) communicator, the LCD will display the On-line Menu. Refer to the HART Communicator manual for detailed menu information.
The Oxymitter 4000 logic I/O output can be configured for ten different modes through HART/AMS. The factory default condition is Mode 5. A list of possible configurations appear in Table 7-1.
The Unit Alarm configuration available for Modes 1, 3, 5, and 7 refers to the diagnostic alarm faults in Table 8-1.

HART/AMS MENU TREE This section consists of a menu tree for the HART Communicator. This menu

is specific to Oxymitter 4000 applications.
Table 7-1. Logic I/O Configuration (as set at HART/AMS or LOI)
Mode Configuration
0 The unit is not configured for any alarm condition. 1 The unit is configured for a Unit Alarm. 2 The unit is configured for Low O 3 The unit is configured for both a Unit Alarm and Low O2. 4 The unit is configured for a High AC Impedance/CALIBRATION
5* The unit is configured for both a Unit Alarm and a High AC
6 The unit is configured for both a Low O
7 The unit is configured for a Unit Alarm, a Low O2, and a High AC
8** The unit is configured for a calibration handshake with IMPS 4000 or SPS 4001B.
9 The unit is configured for a calibration handshake. CALIBRATION
*The default condition for an Oxymitter 4000 without an IMPS 4000 or SPS 4001B.
**The default condition for an Oxymitter 4000 with an IMPS 4000 or SPS 4001B.
RECOMMENDED.
Impedance/CALIBRATION RECOMMENDED.
RECOMMENDED.
Impedance/CALIBRATION RECOMMENDED.
CALIBRATION RECOMMENDED will initiate the calibration cycle.
RECOMMENDED will not initiate the calibration cycle with the IMPS 4000 or SPS 4001B.
.
2
and High AC Impedance/CALIBRATION
2
7-4
Page 87
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 7-3. HART/AMS Menu Tree (Sheet 1 of 3)
Oxymitter 4000
DEVICE SETUP PV PV AO PV LRV PV URV
PROCESS VARIABLES
DIAG/SERVICE
VIEW FLD DEV VARS
VIEW OUTPUT VARS
VIEW FLD DEV mV
STATUS
O2 value O2 cell temp CJ temp
VIEW PV-Aout
VIEW SV
VIEW TV
VIEW 4V
Cell mV Cell TC mV CJ mV
Status Group 1
Status Group 2
Status Group 3
Operate Model
AO Saturated
AO Fixed
Max Case Temp
PV is PV O2 value PV % rnge PV AO
SV is Cold Junct SV __ mV
TV is Cell TV __ mV
4V is Cell TC 4V __ mV
Open T/C Shorted T/C Reversed T/C Heater Open Cell Temp Very HI High Case Temp
Cell Temp Low Cell Temp High Cell Open High Cell Imp. CK.ER - EEPROM Cal.Error Slope Cal. Recommended
Cal. Error Const. Last Cal. Failed
AO1 Out Rnge lim
(CONTINUED ON
SHEET 2)
LOOP TEST
O2 CALIBRATE
D/A TRIM
Loop test method...
PERFORM O2 CAL
O2 CAL STATUS
LAST CALCONSTANTS
RESET CALCONSTANTS
D/A trim method
O2 Cal
Optrak TG?
CalState
CalState TimeRemain Present O2
Cal slope Cal const Cell Imp
Reset CalConstants method...
O2 Cal method...
Refer to “HART Communicator O2 Cal Method” for the complete O2 calibration method using the HART Communicator.
26170028
7-5
Page 88
Oxymitter 4000
Figure 7-3. HART/AMS Menu Tree (Sheet 2 of 3)
(CONTINUED FROM
SHEET 1)
Ta g
ASSIGN PV&SV
PV is Oxygen SV is Cold Junct TV is Cell 4V is Cell TC
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
DEVICE SETUP PV PV AO PV LRV PV URV
BASIC SETUP
DETAILED SETUP
SELECT O2 RANGE
DEVICE INFORMATION
S/W VERSION INFO
SENSORS
SIGNAL CONDITION
OUTPUT CONDITION
O2
URV ____% LRV ____%
Dev id Descriptor Message Date Final asmbly num Snsr s/n
Ver Chk sum Bld num Bld date
O2 O2 CELLTEMP COLDJUNCTEMP O2 CELL MV O2 CELLTCMV COLD JUNC MV
PV URV PV LRV PV % rnge
ANALOG OUTPUT
HART OUTPUT
ALARM OUTPUT
SLOPE
LIMITS DISPLAYED
PV AO ____ mA PV AO Alrm typ
LOOP TEST
D/A TRIM
Poll addr Num req preams
Logic I/O Pin State Logic I/O Pin Mode
Loop test method...
D/A trim method...
No Alarm Unit Alarm Low O2 Alarm Low O2/Unit Alarm Cal Rec Cal Rec/Unit Alarm Low O2/Cal Rec Low O2/Unit/Cal Rec Cal Rec/Handshake Handshake
7-6
(CONTINUED ON
SHEET 3)
O2 CALIBRATION
O2 ALARMS
CONSTANT
HighTG LowTG OPtrak TG? TGtime PurgeTime Cal Mode Cal Intrvl h Next CalTime h
LoAlarmSP
OP locks OP tracks
Manual Auto
35830006
Page 89
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Figure 7-3. HART/AMS Menu Tree (Sheet 3 of 3)
(CONTINUED FROM
SHEET 2)
REVIEW
DEVICE INFORMATION
CAL INFO
Oxymitter 4000
Manufacturer Model Dev id Ta g Descriptor Message Date Final asmbly num Snsr s/n Fld dev rev Hardware rev Software rev Universal rev
Optrak TG? Tgtime PurgeTime LowTG HighTG Cal slope Cal const CellImp Imp Delta
DEVICE SETUP PV PV AO PV LRV PV URV
DEVICE CONFIG
OUTPUTS CONFIG
Slope Constant
URV LRV LoAlarmSP I/O Pin Mode Poll addr Num req preams
26170030
7-7
Page 90
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006

HART COMMUNICATOR O2 CAL METHOD

Use the following procedure to perform a calibration using the HART Commu­nicator. If necessary, use the menu tree in Figure 7-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press the left arrow key.
1. From the PERFORM O access the O
Failure to remove the Oxymitter 4000 from automatic control loops prior to performing this procedure may result in a dangerous operating condition.
2. In the first O2 CAL screen, a "Loop should be removed from automatic control" warning appears. Remove the Oxymitter 4000 from any auto­matic control loops to avoid a potentially dangerous operating condition and press OK.
3. The next several screens indicate the calibration status. At each of the following status prompts, select menu item 2, NEXT CAL STEP: COMPLETE CAL RECOMMENDED APPLY GAS 1 GAS 1 FLOW
4. At this point, select menu item 4, EXIT, to leave the O2 CAL procedure.
5. From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to monitor the calibration status as it updates. Or, access the O2 CALI­BRATE screen and select menu item 2, O2 CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.
6. When CALSTATE displays APPLY GAS 2, return to the O2 CAL procedure.
7. When the "Loop should be removed from automatic control" warning appears, return the Oxymitter 4000 to the automatic control loops previ­ously removed and press OK.
8. At the STOP GAS status prompt, select menu item 2, NEXT CAL STEP. When the status displays PURGING, select menu item 4, EXIT, to leave the O2 CAL procedure.
9. From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to monitor the calibration status as it updates. Or, access the O2 CALI­BRATE screen and select menu item 2, O2 CAL STATUS, to view menu item 1, CALSTATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.
10. When CALSTATE displays STOP GAS, return to the O2 CAL proce­dure.
calibration procedure.
2
CAL screen, select menu item 1, O2 CAL, to
2
7-8
Page 91
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
11. When the "Loop should be returned to automatic control" message appears, return the Hazardous Area Oxymitter 4000 to the automatic control loops previously removed and press OK.
12. At the STOP GAS status prompt, select menu item 2, NEXT CAL STEP When the status displays PURGING, select menu item 4, EXIT, to leave the O2 CAL procedure.
13. From the PERFORM O2 CAL screen, view menu item 3, CALSTATE, to monitor the calibration status as it updates. Or, access the O2 CALI­BRATE screen and select menu item 2, O2 CAL STATUS, to view menu item 1, CAL-STATE; menu item 2, TIMEREMAIN; and menu item 3, PRESENT O2, as the calibration status updates.
14. When CALSTATE displays COMPLETE, the calibration is finished.

DEFINING A TIMED CALIBRATION VIA HART

Use the following procedure to specify a time interval (in hours) at which the Oxymitter 4000 will be automatically calibrated. If necessary, use the menu tree in Figure 7-3 (Sheet 2 of 3) for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press the left arrow key.
1. From the DEVICE SETUP screen, select DETAILED SETUP.
2. From the DETAILED SETUP screen, select O2 CALIBRATION.
3. From the O2 CALIBRATION screen, select menu item 6, CAL MODE. Set the CAL MODE to AUTO.
4. Return to the O2 CALIBRATION screen and select menu item 7, CAL INTRVL.
5. At the prompt, input a time interval (in hours) at which an automatic calibration will occur; then press ENTER.

D/A TRIM PROCEDURE The D/A trim procedure is used to calibrate the 4-20 mA output signal to a

precision mA measurement device (calibrated digital ammeter, etc.). The procedure is interactive and stored in the Oxymitter software.
Use one of the following communication methods to access the D/A trim procedure:
LOI Menu
1. Use the "Z" pattern key entry to access the LOI menu.
2. Press the down key two times to access the SYSTEM menu.
3. Press the down key once to access the Input/Output menu.
4. From the Analog selection, press the right-pointing key to display the Analog submenu listing.
5. Press the down key as needed to access Trim O2 Out.
6. Press the Enter key to start the trim procedure. Follow the LOI display prompts to perform the trim procedure.
7-9
Page 92
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
7-10
Page 93
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000

Section 8 Troubleshooting

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-3
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-3
Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 8-4
Identifying and Correcting Alarm Indications . . . . . . . . . page 8-5
Calibration Passes, but Still Reads Incorrectly . . . . . . . . page 8-22

OVERVIEW While the Oxymitter 4000 electronics provides a significant number of

diagnostic alarms to assist in troubleshooting potential problems, it is good to place these alarms in perspective with respect to the instrument's operating principles:
When the Zirconium Oxide sensing cell is heated to its setpoint [1357°F (736°C)], the cell will generate a voltage that represents the difference between the process O O
ambient air).
2
Test points, Figure 8-1, are provided to read the raw millivolt value generated by the thermocouple that controls the cell temperature and also the raw cell signal.
% and the reference O2% inside the probe (20.95%
2
The cell temperature at test points 3 and 4 should always be stable at approximately 29 to 30 millivolts, which represents the [1357°F (736°C)] setpoint temperature.
When flowing calibration gasses, the raw cell millivolt value at test points 1 and 2 should represent the levels on the chart in Figure 8-1. Note that the raw cell millivolt value increases logarithmically as the O decreases.
concentration
2
http://www.processanalytic.com
Page 94
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Figure 8-1. O Reading vs. % O
Sensor mV
2
at 1357°F
2
(736°C) (Reference Air, 20.9% O
)
2
200
150
100
EMF (mV)
50
O Sensor Performance at 1357
2
(736
O
C)
O
F
0
Concentration O (%)
O2 % EMF(mV)
O2 % EMF(mV)
100 20 15 10 9 8 7 6 5 4
-34 1.0 7.25 16.1 18.4 21.1 23.8 27.2 31.2 36.0
3 2 1 0.8 0.6 0.5 0.4 0.2 0.1 0.01
42.3 51.1 66.1 71.0 77.5 81.5 86.3 101.4 116.6 166.8
2
1001010.01 0.10
37260043
8-2
Page 95
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Install all protective equipment covers and safety ground leads after troubleshooting. Failure to install covers and ground leads could result in serious injury or death.
Oxymitter 4000

GENERAL The troubleshooting section describes how to identify and isolate faults that

may develop in the Oxymitter 4000. When troubleshooting, reference the following.
Grounding
It is essential that adequate grounding precautions are taken when installing the system. Thoroughly check both the probe and electronics to ensure the grounding quality has not degraded during fault finding. The system provides facilities for 100% effective grounding and the total elimination of ground loops.
Electrical Noise
The Oxymitter 4000 has been designed to operate in the type of environment normally found in a boiler room or control room. Noise suppression circuits are employed on all field terminations and main inputs. When fault finding, evaluate the electrical noise being generated in the immediate circuitry of a faulty system. Ensure all cable shields are connected to earth.
Loose Integrated Circuits
The Oxymitter 4000 uses a microprocessor and supporting integrated circuits (IC). If the electronics are handled roughly during installation or located where subjected to severe vibration, the ICs could work loose. Before troubleshoot­ing the system, ensure all ICs are fully seated.
Electrostatic Discharge
Electrostatic discharge can damage the ICs used in the electronics. Before removing or handling the processor board or the ICs, ensure you are at ground potential.

ALARM INDICATIONS The first indication of a problem at the analyzer usually comes from the Oper-

ators running the process. Critical alarms that render the O unusable will force the 4-20 mA analog output signal representing O a default condition, as follows:
4-20 mA Signal Alarm Levels
4-20 mA signal Analyzer Condition
0 mA Analyzer unpowered, or completely failed
3.5 mA Critical Alarm - analyzer reading unusable (factory default)
3.8 mA
4 to 20 mA Normal Operation
20.5 mA Reading Over Range (Example - range is 0-10%. Current reading is 12%)
>21 mA
Reading Under Range (Example - user sets range to 2-10%. Current reading is 1.9%)
Critical Alarm - analyzer reading is unuasble (user can choose this alarm level instead of the factory default level of 3.5 to 3.6 mA)
measurement
2
to go to
2
8-3
Page 96
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
NOTE
Make sure that the Control System is configured to interpret these signal levels correctly!
Once an alarm condition is indentified, the Oxymitter 4000 electronics offers a number of diagnostics to interpret the specific alarm.
If the Oxymitter 4000 has the simple keypad operator interface, the majority of fault conditions will be indicated by one of the four LEDs referred to as diagnostic, or unit alarms on the operator's keypad (Figure 8-2). An LED will flash a code that will correspond to an error message. Only one LED will blink at a time. An alarm code guide is provided inside the screw-on cover for the electronics.
Alarm indications will also be available via the optional LOI or the HART Model 275/375 hand-held communicator and Rosemount Analytical's Asset Management software. When the error is corrected and/or power is cycled, the diagnostic alarms will clear or the next error on the priority list will appear.
Figure 8-2. Diagnostic LEDs
Diagnostic
LEDs
HEATER T/C
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
LOW
GAS
GAS
DEC DEC
HEATER
02 CELL
CALIBRATION
02 CELL mV + 02 CELL mv ­HEATER T/C + HEATER T/C -
CAL
TEST GAS +
PROCESS -
% 02
37260044

ALARM CONTACTS If autocalibration is not utilized, a common bidirectional logic contact is pro-

vided for any of the diagnostic alarms listed in Table 8-1. The assignment of alarms which can actuate this contact can be modified to one of seven addi­tional groupings (mode 0 through mode 7) listed in Table 7-1.
8-4
The logic contact is self-powered, +5 VDC, with a 340 ohm series resistance. An interposing relay will be required if this contact is to be utilized to annunci­ate a higher voltage device, such as a light or horn. An interposing relay may also be required for certain DCS input cards.
Page 97
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
A Potter & Brumfield R10S-E1Y1-J1.0K 3.2 mA DC or an equal interposing relay will be mounted where the contact wires terminate in the control/relay room.
If autocalibration systems are utilized, the bidirectional logic contact is utilized as a "hand-shake" signal between the autocalibration system (SPS 4001B or IMPS 4000) and is unavailable for alarming purposes. Additional contacts are provided through the autocalibration systems, noted below.
SPS 4001B and IMPS 4000, 1-4 probes
• One contact closure per probe from the control room to the SPS 4001B or IMPS 4000 for "calibration initiate".
• One contact output per probe from the SPS 4001B or IMPS 4000 to the control room for "in calibration" notification.
• One contact output per probe from the SPS 4001B or IMPS 4000 to the control room for "calibration failed" notification. (Includes output from pressure switch indicating "cal gas bottles empty").
Additional IMPS 4000 Alarm Contacts
• One contact per IMPS 4000 for "low calibration gas flowing".
• One contact per IMPS 4000 for "high calibration gas flowing".

IDENTIFYING AND CORRECTING ALARM INDICATIONS

NOTE
The 4-20 mA signal can be configured to respond normally during any calibration, or can be configured to hold the last O calibration. Factory default is for the 4-20 mA signal to operate normally throughout calibration.
NOTE
Holding the last O for the purpose of automatic control. Unless several probes are being averaged, always place any control loops using the O prior to calibrating.
For an Oxymitter 4000 with a membrane keypad, faults are indicated by four diagnostic, or unit, alarm LEDs. A pattern of repeating blinks define the problem. A condensed table of the errors and the corresponding blink codes can be found on the inside right cover of the electronics housing. Table 8-1 also identifies the blink code and fault status of each LED as well as the output of the 4-20 mA signal line and a fault number that corresponds to the troubleshooting instructions provided in this section.
For an Oxymitter 4000 with the optional LOI, alarm messages are displayed on the LOI display window when the alarm status display is accessed via the LOI menu. A listing of the alarm/fault messages and the related fault status descriptions and fault numbers are shown in Table 8-2.
value may be useful if several probes are being averaged
2
value upon the initiation of
2
signal into manual
2
8-5
Page 98
Instruction Manual
IM-106-340, Rev. 4.0
Oxymitter 4000
Table 8-1. Diagnostic/Unit Alarm Fault Definitions ­Membrane Keypad Only
LED Flashes Status 4-20 mA Line Fault Recoverable
HEATER T/C 1
HEATER 1
O
CELL 1
2
CALIBRATION 1
*Critical alarm conditions will render the O2 measurement as unusable, and any of these events will cause the 4-20 mA signal to go to a
user-selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value is 3.5 mA. Alarms which are not self-clearing (Self-Clearing = NO) will require a reset. Perform the Reset Procedure in Section 3: Configuration of Oxymitter 4000 with Membrane Keypad to continue operation.
**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended LED on the operator's keypad.
Open Shorted
2
Reversed
3
A/D Comm Error
4
Open
2
High High Temp
3
High Case Temp
4
Low Temp
5
High Temp High mV
Bad
3
EEprom Corrupt
4
Invalid Slope
2
Invalid Constant
3
Last Calibration Failed
**
Calibration Recommended
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)* Track O
2
3.5 mA (factory default)* Track O
2
Track O
2
Track O
2
Track O
2
1 2 3 4
5 6 7 8 9
10 11 12
13 14 15
May 2006
No No No No
No
No Yes Yes Yes
Yes Yes
No Yes
Yes Yes Yes
Table 8-2. Diagnostic/Unit Alarm Fault Definitions - LOI
Message Status Fault Number Self Clearing
O2 T/C OPEN Heater T/C Open 1 No O2 T/C SHORTED Heater T/CShorted 2 No O2 T/C REVERSED Heater T/C Polarity Reversed 3 No ADC ERROR A/D Comm Error 4 No O2 HEATER OPEN O2 Heater Open 5 No VERY HI O2 TEMP Very High Process Temperature 6 No BOARD TEMP HI Electronics Overheated 7 Yes O2 TEMP LOW Low Process Temperature 8 Yes O2 TEMP HI High Process Temperature 9 Yes O2 CELL OPEN O2 Cell Open 10 Yes O2 CELL BAD O2 Cell Failed 11, 13, 14 Yes EEPROM CORRUPT EEprom Failed 12 No CALIB FAILED Last Calibration Failed 15 Yes LINE FREQ ERROR Incorrect Input Line Frequency Detected on Power Up No
8-6
Page 99
Instruction Manual
IM-106-340, Rev. 4.0 May 2006
Oxymitter 4000
Figure 8-3. Fault 1, Open Thermocouple
HEATER T/C
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
LOW
GAS
GAS
DEC DEC
HEATER O2 CELL
CALIBRATION
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
CAL
TEST GAS +
PROCESS -
% O2
KEYPAD
TP1
TP2
TP3
TP4
TP5
TP6
SW2
Fault 1, Open Thermocouple
Figure 8-3 shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and a Oxymitter 4000 with an LOI (lower view). The upper view also shows J1 and test points TP1 through TP6, located on the
ON
microprocessor board, below the membrane keypad or the LOI module.
Membrane Keypad
J1
When Fault 1 is detected, the HEATER T/C LED flashes once, pauses for three seconds, and repeats.
YEL
RED
GRN
ORG
1. Check connector J1. Ensure the connector is properly seated.
2. Using a multimeter, measure the voltage from TP3+ to TP4-. If the reading is 1.2 VDC ±0.1 VDC, the thermocouple is open.
3. Remove power. Disconnect J1. Measure the resistance across the red and yellow thermocouple leads. The resistance should be approximately 1 ohm.
4. If the thermocouple is open, see "Heater Strut Replacement" in Section 9: Maintenance and Service.
Alarms
O2 T/C Open
LOI
LOI
When Fault 1 is detected, the LOI displays the "O2 T/C Open" message.
1. Remove power. Unscrew and remove the LOI module from the electronic assembly.
2. Reconnect power to the Oxymitter 4000.
3. Perform the diagnostic steps 1 through 4 shown for the membrane keypad.
37260019
8-7
Page 100
Oxymitter 4000
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Figure 8-4. Fault 2, Shorted Thermocouple
HEATER T/C
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC INC
HIGH
LOW
GAS
GAS
DEC DEC
HEATER
O2 CELL
CALIBRATION
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
CAL
TEST GAS +
PROCESS -
% O2
KEYPAD
TP1
TP2
TP3
TP4
TP5
TP6
SW2
Fault 2, Shorted Thermocouple
Figure 8-4 shows the electronic assembly for an Oxymitter 4000 with a membrane keypad (upper view) and an Oxymitter 4000 with an LOI (lower view). The upper view also shows J1 and test points TP1 through TP6, located on the
ON
microprocessor board, below the membrane keypad or the LOI module.
Membrane Keypad
J1
When Fault 2 is detected, the HEATER T/C LED flashes twice, pauses for three seconds, and repeats.
YEL
RED
GRN
ORG
1. Using a multimeter, measure the voltage from TP3+ to TP4-. If the reading is 0 ±0.5 mV, then a shorted thermocouple is likely.
2. Remove power and disconnect J1.
3. Measure the resistance from TP3+ to TP4-. The reading should be approximately 20K ohms.
4. If so, the short is not on the PC board. The thermocouple wiring or the thermocouple is shorted. See "Heater Strut Replacement" in Section 9: Maintenance and Service.
Alarms
O2 T/C Shorted
LOI
LOI
When Fault 2 is detected, the LOI displays the "O2 T/C Shorted" message.
1. Remove power. Unscrew and remove the LOI module from the electronic assembly.
2. Reconnect power to the Oxymitter 4000.
3. Perform the diagnostic steps 1 through 4 shown for the membrane keypad.
37260020
8-8
Loading...