viMicro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Chapter 1
Before You Begin
1.1Overview
This chapter provides an orientation to the use of this manual, and includes a pre-configuration
worksheet. This manual describes the procedures required to start, configure, use, maintain, and
troubleshoot the Model 1500 transmitter with the filling and dosing application.
1.2Safety
Safety messages are provided throughout this manual to protect personnel and equipment. Read each
safety message carefully before proceeding to the next step.
1.3Version
Different configuration options are available with different versions of the components. Table 1-1 lists
the version information that you may need and describes how to obtain the information.
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
Table 1-2 lists documentation sources for additional information.
Tabl e 1-2Flowmeter documentation resources
TopicDocument
Sensor installationSensor documentation
Transmitter installationTransmitter Installation: Model 1500 and 2500 Transmitters
Configuration and Use Manual1
Before You Begin
1.5Communication tools
Most of the procedures described in this manual require the use of a communication tool. To
configure and use the Model 1500 transmitter with the filling and dosing application, you must use
ProLink II v2.3 or later, or a customer-written program that uses the transmitter’s Modbus interface.
For certain features, ProLink II v2.5 or later is required; this is noted where applicable.
Basic information on ProLink II and connecting ProLink II to your transmitter is provided in
Chapter 2. For more information, see the ProLink II manual, installed with the ProLink II software or
available on the Micro Motion web site (www.micromotion.com).
For information on the transmitter’s Modbus interface, see:
•Using Modbus Protocol with Micro Motion Transmitters, November 2004, P/N 3600219,
Rev. C (manual plus map)
•Modbus Mapping Assignments for Micro Motion Transmitters, October 2004, P/N 20001741,
Rev. B (map only)
Both of these manuals are available on the Micro Motion web site.
1.6Planning the configuration
The pre-configuration worksheet in Section 1.7 provides a place to record information about your
flowmeter (transmitter and sensor) and your application. This information will affect your
configuration options as you work through this manual. Fill out the pre-configuration worksheet and
refer to it during configuration. You may need to consult with transmitter installation or application
process personnel to obtain the required information.
If you are configuring multiple transmitters, make copies of this worksheet and fill one out for each
individual transmitter.
2Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
Configuration and Use Manual3
Before You Begin
1.8Micro Motion customer service
For customer service, phone the support center nearest you:
•In the U.S.A., phone
800-522-MASS (800-522-6277) (toll-free)
•In Canada and Latin America, phone +1 303-527-5200
•In Asia:
-In Japan, phone 3 5769-6803
-In other locations, phone +65 6777-8211 (Singapore)
•In Europe:
-In the U.K., phone 0870 240 1978 (toll-free)
-In other locations, phone +31 (0) 318 495 670 (The Netherlands)
Customers outside the U.S.A. can also email Micro Motion customer service at
International.Support@EmersonProcess.com.
4Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Chapter 2
Connecting with ProLink II Software
2.1Overview
ProLink II is a Windows-based configuration and management tool for Micro Motion transmitters. It
provides complete access to transmitter functions and data.
This chapter provides basic information for connecting ProLink II to your transmitter. The following
topics and procedures are discussed:
•Requirements (see Section 2.2)
•Configuration upload/download (see Section 2.3)
•Connecting to a Model 1500 transmitter (see Section 2.4)
The instructions in this manual assume that users are already familiar with ProLink II software. For
more information on using ProLink II, or for detailed instructions on installing ProLink II, see the
ProLink II software manual, which is automatically installed with ProLink II, and is also available on
the Micro Motion web site (www.micromotion.com).
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
2.2Requirements
To use ProLink II with a Model 1500 transmitter with the filling and dosing application, the following
are required:
•ProLink II v2.3 or later, for access to the filling and dosing application
•ProLink II v2.5 or later, for access to meter verification
•The appropriate signal converter and cables: RS-485 to RS-232 or USB to RS-232
-For RS-485 to RS-232, the Black Box
Converter (Code IC521A-F) signal converter is available from Micro Motion.
-For USB to RS-232, the Black Box USB Solo (USB–>Serial) (Code IC138A-R2)
converter can be used.
•25-pin to 9-pin adapter (if required by your PC)
2.3ProLink II configuration upload/download
ProLink II provides a configuration upload/download function which allows you to save configuration
sets to your PC. This allows:
•Easy backup and restore of transmitter configuration
•Easy replication of configuration sets
Micro Motion recommends that all transmitter configurations be downloaded to a PC as soon as the
configuration is complete.
®
Async RS-232 <-> 2-wire RS-485 Interface
Parameters specific to the filling and dosing application are not included in the upload or download.
Configuration and Use Manual5
Connecting with ProLink II Software
To access the configuration upload/download function:
1. Connect ProLink II to your transmitter as described in this chapter.
2. Open the
File menu.
•To save a configuration file to a PC, use the
•To restore or load a configuration file to a transmitter, use the
option.
2.4Connecting from a PC to a Model 1500 transmitter
ProLink II software can communicate with a Model 1500 transmitter using Modbus protocol on the
RS-485 physical layer. There are two connection types:
Both connection types use the RS-485 terminals (terminals 33 and 34). These terminals are available
in service port mode for 10 seconds after transmitter power-up. After this interval, the terminals revert
to RS-485 mode.
•To make a service port connection, you must configure ProLink II appropriately and connect
during the 10-second interval after transmitter power-up. Once a service port connection is
made, the terminals will remain in service port mode. You may disconnect and reconnect as
often as required, as long as you continue to use service port mode.
•To make an RS-485 connection, you must configure ProLink II appropriately, wait for the
10-second interval to expire, then connect. The terminals will now remain in RS-485 mode,
and you may disconnect and reconnect as often as required, as long as you continue to use
RS-485 mode.
•To change from service port mode to RS-485 mode, or vice versa, you must cycle power to the
transmitter and reconnect using the desired connection type.
To connect a PC to the RS-485 terminals or an RS-485 network:
1. Attach the signal converter to the serial port of your PC, using a 25-pin to 9-pin adapter if
required.
2. To connect to the RS-485 terminals, connect the signal converter leads to terminals 33 and 34.
See Figure 2-1.
3. To connect to an RS-485 network, connect the signal converter leads to any point in the
network. See Figure 2-2.
4. For long-distance communication, or if noise from an external source interferes with the
signal, install 120-ohm, 1/2-watt resistors in parallel with the output at both ends of the
communication segment.
5. Ensure that the transmitter is disconnected from a host PLC.
Load from Xmtr to File option.
Send to Xmtr from File
6Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Connecting with ProLink II Software
Figure 2-1RS-485 terminal connections to Model 1500
PC
RS-485/B
RS-485/A
25-pin to 9-pin serial port
adapter (if necessary)
RS-485 to RS-232
signal converter
Figure 2-2RS-485 network connections to Model 1500
PC
DCS or PLC
25-pin to 9-pin serial port
adapter (if necessary)
RS-485 to RS-232
signal converter
Add resistance if necessary
(see Step 4)
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
RS-485/B
RS-485/A
6. Start ProLink II software. From the
Connection menu, click on Connect to Device. In the
screen that appears, specify connection parameters appropriate to your connection:
•For service port mode, set
value for your PC.
Baud rate, Stop bits, and Parity are set to standard values and cannot
Protocol to Service Port, and set COM port to the appropriate
be changed. See Table 2-1.
•For RS-485 mode, set the connection parameters to the values configured in your
transmitter. See Table 2-1.
Configuration and Use Manual7
Connecting with ProLink II Software
Tabl e 2-1Modbus connection parameters for ProLink II
Connection type
Connection parameterConfigurable (RS-485 mode)SP standard (service port mode)
ProtocolAs configured in transmitter
(default = Modbus RTU)
Baud rateAs configured in transmitter (default = 9600)38,400
Stop bitsAs configured in transmitter (default = 1)1
ParityAs configured in transmitter (default = odd)none
Address/TagConfigured Modbus address (default = 1)111
COM portCOM port assigned to PC serial portCOM port assigned to PC serial port
(1) Required value; cannot be changed by user.
Modbus RTU
(1)
(1)
(1)
(1)
(1)
7. Click the Connect button. ProLink II will attempt to make the connection.
8. If an error message appears:
a.Swap the leads between the two terminals and try again.
b.Ensure you are using the correct COM port.
c.If you are in RS-485 mode, you may be using incorrect connection parameters.
-Connect in service port mode and check the RS-485 configuration. If required, change
the configuration or change your RS-485 connection parameters to match the existing
configuration.
-If you are unsure of the transmitter’s address, use the
Poll button in the Connect
window to return a list of all devices on the network.
d.Check all the wiring between the PC and the transmitter.
8Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Chapter 3
Flowmeter Startup
3.1Overview
This chapter describes the procedures you should perform the first time you start the flowmeter. You
do not need to use these procedures every time you cycle power to the flowmeter.
The following procedures are discussed:
•Applying power to the flowmeter (see Section 3.2)
•Performing a loop test on the transmitter outputs (see Section 3.3)
•Trimming the mA output (see Section 3.4)
•Zeroing the flowmeter (see Section 3.5)
Note: All ProLink II procedures provided in this chapter assume that your computer is already
connected to the transmitter and you have established communication. All ProLink II procedures also
assume that you are complying with all applicable safety requirements. See Chapter 2 for more
information.
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
3.2Applying power
Before you apply power to the flowmeter, close and tighten all housing covers.
Turn on the electrical power at the power supply. The flowmeter will automatically perform
diagnostic routines. When the flowmeter has completed its power-up sequence, the status LED will
turn green if conditions are normal. If the status LED exhibits different behavior, an alarm condition is
present (see Section 5.4) or configuration of the filling and dosing application is not complete.
Configuration and Use Manual9
Flowmeter Startup
WARNING
Upon transmitter startup or abnormal power reset, any external device
controlled by a discrete output may be momentarily activated.
Upon transmitter startup or abnormal power reset, discrete output states are
unknown. As a result, an external device controlled by a discrete output may
receive current for a brief period.
When using Channel B as a discrete output:
•You can prevent current flow upon normal startup by setting Channel B polarity
to active low (see Section 4.6).
•There is no programmatic method to prevent current flow for Channel B upon
abnormal power reset. You must design the system so that a brief current flow to
the external device controlled by Channel B cannot cause negative
consequences.
When using Channel C as a discrete output, there is no programmatic method to
prevent current flow upon either transmitter startup or abnormal power reset. You
must design the system so that a brief current flow to the external device controlled
by Channel C cannot cause negative consequences.
3.3Performing a loop test
A loop test is a means to:
•Verify that the mA outupt is being sent by the transmitter and received accurately by the
receiving device
•Determine whether or not you need to trim the mA output
•Select and verify the discrete output voltage
•Read the discrete input
Perform a loop test on all inputs and outputs available on your transmitter. Before performing the loop
tests, ensure that your transmitter terminals are configured for the input/outputs that will be used in
your application (see Section 4.3).
ProLink II is used for loop testing. See Figure 3-1 for the loop test procedure. Note the following:
•The mA reading does not need to be exact. You will correct differences when you trim the mA
output. See Section 3.4.
10Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Flowmeter Startup
Figure 3-1ProLink II – Loop test procedure
ProLink Menu
Test
Fix Milliamp 1
Enter mA value
Fix mA
Read output at
receiving device
Correct?Correct?Correct?
YesNoYes No
Loop test successful
UnFix
Fix Discrete Out 1
Fix Discrete Out 2
ON or OFF
Verify state at
receiving device
Check output wiring
Troubleshoot receiving device
Read Discrete Input
Toggle remote input
Verify Present State LED
Loop test successful
device
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
at transmitter
Check input wiring
Troubleshoot input device
3.4Trimming the milliamp output
Trimming the mA output creates a common measurement range between the transmitter and the device
that receives the mA output. For example, a transmitter might send a 4 mA signal that the receiving
device reports incorrectly as 3.8 mA. If the transmitter output is trimmed correctly, it will send a
signal appropriately compensated to ensure that the receiving device actually indicates a 4 mA signal.
You must trim the mA output at both the 4 mA and 20 mA points to ensure appropriate compensation
across the entire output range.
ProLink II is used to trim the mA output. See Figure 3-2 for the mA output trim procedure. Note the
following:
•Any trimming performed on the output should not exceed ± 200 microamps. If more trimming
is required, contact Micro Motion customer support.
Configuration and Use Manual11
Flowmeter Startup
Figure 3-2ProLink II – mA output trim procedure
ProLink Menu
Calibration
Milliamp Trim 1
4 mA trim20 mA trim
Read mA output at
receiving device
Next
Enter receiving device
value in Enter Meas
Next
Read mA output at
No
receiving device
Equal?
3.5Zeroing the flowmeter
Zeroing the flowmeter establishes the flowmeter’s point of reference when there is no flow. The meter
was zeroed at the factory, and should not require a field zero. However, you may wish to perform a
field zero to meet local requirements or to confirm the factory zero.
Yes
Next
Read mA output at
receiving device
Enter receiving device
value in Enter Meas
Read mA output at
No
receiving device
Equal?
Next
Next
Yes
Finish
Note: Do not zero the flowmeter if a high severity alarm is active. Correct the problem, then zero the
flowmeter. You may zero the flowmeter if a low severity alarm is active. See Section 5.4 for
information on viewing transmitter status and alarms.
When you zero the flowmeter, you may need to adjust the zero time parameter. Zero time is the
amount of time the transmitter takes to determine its zero-flow reference point.
•A long zero time may produce a more accurate zero reference but is more likely to result in a
zero failure. This is due to the increased possibility of noisy flow, which causes incorrect
calibration.
•A short zero time is less likely to result in a zero failure but may produce a less accurate zero
reference.
The default zero time is 20 seconds. For most applications, the default zero time is appropriate.
You can zero the flowmeter with ProLink II or with the zero button on the transmitter.
If the zero procedure fails, see Section 11.6 for troubleshooting information.
12Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Flowmeter Startup
Additionally, if you have the enhanced core processor and you are using ProLink II to zero the
flowmeter, you can also restore the prior zero immediately after zeroing (e.g., an “undo” function), as
long as you have not closed the Calibration window or disconnected from the transmitter. Once you
have closed the Calibration window or disconnected from the transmitter, you can no longer restore
the prior zero.
3.5.1Preparing for zero
To prepare for the zero procedure:
1. Apply power to the flowmeter. Allow the flowmeter to warm up for approximately 20 minutes.
2. Run the process fluid through the sensor until the sensor temperature reaches the normal
3. Close the shutoff valve downstream from the sensor.
4. Ensure that the sensor is completely filled with fluid.
5. Ensure that the process flow has completely stopped.
process operating temperature.
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
CAUTION
If fluid is flowing through the sensor, the sensor zero calibration may be
inaccurate, resulting in inaccurate process measurement.
To improve the sensor zero calibration and measurement accuracy, ensure that
process flow through the sensor has completely stopped.
3.5.2Zero procedure
To zero the transmitter:
•With ProLink II, see Figure 3-3.
•With the zero button, see Figure 3-4. Note the following:
-You cannot change the zero time with the zero button. If you need to change the zero time,
you must use ProLink II.
-The zero button is located on the front panel of the transmitter. To press the zero button,
use a fine-pointed object that will fit into the opening (0.14 in [3.5 mm]). Hold the button
down until the status LED on the front panel begins to flash yellow.
Configuration and Use Manual13
Flowmeter Startup
Figure 3-3ProLink II – Flowmeter zero procedure
ProLink >
Calibration >
Zero Calibration
Modify zero time
if required
Perform Auto Zero
Calibration in Progress
LED turns red
Wait until Calibration in
Progress LED turns green
Red
Troubleshoot
Calibration
Failure LED
Figure 3-4Zero button – Flowmeter zero procedure
Press ZERO button
Status LED flashes
yellow
Status LED
Solid
Red
Troubleshoot
Green
Done
Solid Green or
Solid Yellow
Done
14Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Chapter 4
Required Transmitter Configuration
4.1Overview
This chapter describes the configuration procedures that are usually required when a transmitter is
installed for the first time. The procedures in this chapter should be performed in the order shown in
Figure 4-1.
Figure 4-1Required configuration procedures in order
Characterize the flowmeter
(Section 4.2)
Configure the channels
(Section 4.3)
Configure measurement units
(Section 4.4)
Configure mA output
(Section 4.5)
Configure discrete outputs
(Section 4.6)
Configure discrete input
(Section 4.7)
(2)
Done
(1)
(1) Only the input or outputs that have been assigned to
(1)
a channel need to be configured.
(2) If the meter verification option has been purchased,
the final configuration step should be to establish a
meter verification baseline (see Section 4.8).
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
This chapter provides basic flowcharts for each procedure. For more detailed flowcharts, see the
ProLink II flowcharts, provided in Appendix C.
Default values and ranges for the parameters described in this chapter are provided in Appendix A.
For optional transmitter configuration parameters and procedures, see Chapter 6. For configuration of
the filling and dosing application, see Chapter 7.
Note: All ProLink II procedures provided in this chapter assume that your computer is already
connected to the transmitter and you have established communication. All ProLink II procedures also
assume that you are complying with all applicable safety requirements. See Chapter 2 for more
information.
Configuration and Use Manual15
Required Transmitter Configuration
4.2Characterizing the flowmeter
Characterizing the flowmeter adjusts the transmitter to compensate for the unique traits of the sensor
it is paired with. The characterization parameters, or calibration parameters, describe the sensor’s
sensitivity to flow, density, and temperature.
4.2.1When to characterize
If the transmitter, core processor, and sensor were ordered together, then the flowmeter has already
been characterized. You need to characterize the flowmeter only if the core processor and sensor are
being paired together for the first time.
4.2.2Characterization parameters
The characterization parameters that must be configured depend on your flowmeter’s sensor type:
“T-Series” or “Other” (also referred to as “Straight Tube” and “Curved Tube,” respectively), as listed
in Table 4-1. The “Other” category includes all Micro Motion sensors except T-Series.
The characterization parameters are provided on the sensor tag. The format of the sensor tag varies
depending on your sensor’s date of purchase. See Figures 4-2 and 4-3 for illustrations of newer and
older sensor tags.
(1) See the section entitled “Density calibration factors.”
(2) On some sensor tags, shown as TC.
(3) See the section entitled “Flow calibration values.”
(4) Older T-Series sensors. See the section entitled “Flow calibration values.”
(5) Newer T-Series sensors. See the section entitled “Flow calibration values.”
(2)
T- S erie sO t he r
(1)
(1)
(1)
(1)
(1)
✓✓
(4)
(5)
(1)
(3)
16Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Required Transmitter Configuration
Figure 4-2Sample calibration tags – All sensors except T-Series
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
Density calibration factors
If your sensor tag does not show a D1 or D2 value:
•For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air.
•For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.
If your sensor tag does not show a K1 or K2 value:
•For K1, enter the first 5 digits of the density calibration factor. In the sample tag in Figure 4-2,
this value is shown as
12500.
•For K2, enter the second 5 digits of the density calibration factor. In the sample tag in
Figure 4-2, this value is shown as
14286.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration
factor. In the sample tag in Figure 4-2, this value is shown as
4.44.
Configuration and Use Manual17
Required Transmitter Configuration
Flow calibration values
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-character
FT value. Both values contain decimal points. During characterization, these are entered as a single
10-character string that includes two decimal points. In ProLink II, this value is called the Flowcal
parameter.
To obtain the required value:
•For older T-Series sensors, concatenate the FCF value and the FT value from the sensor tag, as
shown below.
Flow FCF X.XXXXFT X.XX
•For newer T-Series sensors, the 10-character string is represented on the sensor tag as the FCF
value. The value should be entered exactly as shown, including the decimal points. No
concatenation is required.
•For all other sensors, the 10-character string is represented on the sensor tag as the Flow Cal
value. The value should be entered exactly as shown, including the decimal points. No
concatenation is required.
4.2.3How to characterize
To characterize the flowmeter:
1. See the menu flowchart in Figure 4-4.
2. Ensure that the correct sensor type is configured.
3. Set required parameters, as listed in Table 4-1.
Figure 4-4Characterizing the flowmeter
ProLink Menu
Configuration
Device
·Sensor type
Straight
tube
Density
Sensor type?
Curved
tube
Density
Flow
T Series Config
18Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Flow
Required Transmitter Configuration
4.3Configuring the channels
The six input/output terminals provided on the Model 1500 are organized into three pairs. These pairs
are called Channels A, B, and C. The channels should be configured before doing any other I/O
configuration.
Changing the channel configuration without verifying I/O configuration can
produce process error.
When the configuration of a channel is changed, the channel’s behavior will be
controlled by the I/O configuration that is stored for the new channel type, which
may or may not be appropriate for the process. To avoid causing process error:
CAUTION
•Configure the channels before configuring the I/O
•When changing channel configuration, be sure that all control loops affected by
this channel are under manual control.
•Before returning the loop to automatic control, ensure that the channel's I/O is
correctly configured for your process. See Sections 4.5, 4.6, and 4.7.
.
The outputs and variable assignments are controlled by the channel configuration. Table 4-2 shows
how each channel may be configured and the power options for each channel.
(1) If set to external power, you must provide power to the outputs.
(1)
(1)
To configure the channels, see the menu flowchart in Figure 4-5.
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
Figure 4-5Configuring the channels
ProLink Menu
Configuration
Channel
Channel B
· Type assignment
· Power type
Channel C
· Type assignment
· Power type
Configuration and Use Manual19
Required Transmitter Configuration
4.4Configuring the measurement units
For each process variable, the transmitter must be configured to use the measurement unit appropriate
to your application.
To configure measurement units, see the menu flowchart in Figure 4-6. For details on measurement
units for each process variable, see Sections 4.4.1 through 4.4.5.
Figure 4-6Configuring measurement units
ProLink Menu
Configuration
Flow
· Mass flow units
· Vol flow units
Density
· Dens units
Temperature
· Temp units
Pressure
· Pressure units
4.4.1Mass flow units
The default mass flow measurement unit is
g/s. See Table 4-3 for a complete list of mass flow
measurement units.
If the mass flow unit you want to use is not listed, you can define a special measurement unit for mass
flow (see Section 6.4).
Tabl e 4-3Mass flow measurement units
ProLink II labelUnit description
g/sGrams per second
g/minGrams per minute
g/hrGrams per hour
kg/sKilograms per second
kg/minKilograms per minute
kg/hrKilograms per hour
kg/dayKilograms per day
mTon/minMetric tons per minute
mTon/hrMetric tons per hour
mTon/dayMetric tons per day
lbs/sPounds per second
lbs/minPounds per minute
lbs/hrPounds per hour
lbs/dayPounds per day
sTon/minShort tons (2000 pounds) per minute
sTon/hrShort tons (2000 pounds) per hour
sTon/dayShort tons (2000 pounds) per day
20Micro Motion® Model 1500 Transmitters with the Filling and Dosing Application
Required Transmitter Configuration
Tabl e 4-3Mass flow measurement units continued
ProLink II labelUnit description
lTon/hrLong tons (2240 pounds) per hour
lTon/dayLong tons (2240 pounds) per day
specialSpecial unit (see Section 6.4)
4.4.2Volume flow units
The default volume flow measurement unit is
L/s. See Table 4-4 for a complete list of volume flow
measurement units.
If the volume flow unit you want to use is not listed, you can define a special measurement unit for
volume flow (see Section 6.4).
Tabl e 4-4Volume flow measurement units
ProLink II labelUnit description
ft3/secCubic feet per second
ft3/minCubic feet per minute
ft3/hrCubic feet per hour
ft3/dayCubic feet per day
m3/secCubic meters per second
m3/minCubic meters per minute
m3/hrCubic meters per hour
m3/dayCubic meters per day
US gal/secU.S. gallons per second
US gal/minU.S. gallons per minute
US gal/hrU.S. gallons per hour
US gal/dayU.S. gallons per day
mil US gal/dayMillion U.S. gallons per day
l/secLiters per second
l/minLiters per minute
l/hrLiters per hour
mil l/dayMillion liters per day
Imp gal/secImperial gallons per second
Imp gal/minImperial gallons per minute
Imp gal/hrImperial gallons per hour
Imp gal/dayImperial gallons per day
barrels/secBarrels per second
barrels/minBarrels per minute
barrels/hrBarrels per hour
barrels/dayBarrels per day
specialSpecial unit (see Section 6.4)
(1)
(1)
(1)
(1)
Using ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You BeginUsing ProLink IIRequired ConfigurationFlowmeter StartupBefore You Begin
(1) Unit based on oil barrels (42 U.S gallons).
Configuration and Use Manual21
Required Transmitter Configuration
4.4.3Density units
The default density measurement unit is
measurement units.
Tabl e 4-5Density measurement units
ProLink II labelUnit description
SGUSpecific gravity unit (not temperature corrected)
g/cm3Grams per cubic centimeter
g/lGrams per liter
g/mlGrams per milliliter
kg/lKilograms per liter
kg/m3Kilograms per cubic meter
lbs/UsgalPounds per U.S. gallon
lbs/ft3Pounds per cubic foot
lbs/in3Pounds per cubic inch
degAPIAPI gravity
sT/yd3Short ton per cubic yard
g/cm3. See Table 4-3 for a complete list of density
4.4.4Temperature units
The default temperature measurement unit is
measurement units.