Emerson Process Management designs, manufactures and tests its products
to meet many national and international standards. Because these
instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within
their normal specifications. The following instructions MUST be adhered to
and integrated into your safety program when installing, using, and
maintaining Rosemount Analytical products. Failure to follow the proper
instructions may cause any one of the following situations to occur: Loss of
life; personal injury; property damage; damage to this instrument; and
warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the
product.
• If you do not understand any of the instructions, contact your Emerson Process Management representative for clarification.
• Follow all warnings, cautions, and instructions marked on and
supplied with the product.
• Inform and educate your personnel in the proper installation, operation,
and maintenance of the product.
• Install your equipment as specified in the Installation Instructions
of the appropriate Instruction Manual and per applicable local and
national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install,
operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use
replacement parts specified by Emerson Process Management.
Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire,
electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by
qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
http://www.raihome.com
NOTES:
The 375 Field Communicator must be upgraded to System Software 2.0 with
Graphic License for operation with the Xi Electronics. The AMS software must
be upgraded to AMS 8.0 or above.
Contact Emerson Process Management’s Global Service Center (GSC) at
1-800-833-8314 to upgrade the 375 Field Communicator software to System
Software 2.0 with Graphic License.
PREFACEThe purpose of this manual is to provide information concerning components,
functions, installation and maintenance of the Xi Electronics.
Some sections may describe equipment not used in your configuration. The
user should become thoroughly familiar with the operation of this module
before operating it. Read this instruction manual completely.
DEFINITIONSThe following definitions apply to WARNINGS, CAUTIONS, and NOTES
found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of
effectiveness.
NOTE
Highlights an essential operating procedure, condition, or statement.
http://www.raihome.com
Instruction Manual
RISKOFELECTRICAL SHOCK
WARNING:REFER TOINSTRUCTIONMANUAL
PROTECTIVECONDUCT OR TERMINAL
EARTH(GROUND) TERMINAL
:
:
:
:
IM-106-910Xi, Original Issue
Xi Advanced Electronics
November 2010
SYMBOLS
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is
a manual illustration number. It is not a part number, and is not related to the
illustration in any technical manner.
OVERVIEWThe Xi is specifically designed to control a zirconium oxide probe for
measuring oxygen, usually the O
the Rosemount Analytical Customer Support Center (CSC) in Solon, Ohio, to
get recommendations for other oxygen probes. 800-433-6076 (US and
Canada).
remaining from a combustion process. Call
2
The Xi electronics has several main functions:
1. Heater Control - The electronics receives a type K thermocouple input
from an O
probe and switches power on and off to the probe's heater in
2
order to maintain a temperature setpoint of 736 degrees C.
2. Signal Conditioning - The electronics receives the raw millivolt signal
from the O
sensing cell, then linearizes and amplifies the signal to
2
provide a linear 4-20 mA output signal used for recording or as an input
into a DCS system for control purposes.
3. Calibration - A bottled calibration gas of known value is typically flowed
into the probe's sensor to verify that it is reading correctly. If the signal is
out of calibration, the calibration gas is used to adjust the 4-20 mA
output signal. During calibration the Xi prompts the technician to flow
two calibration gases into the probe and, with the calibration gases
flowing, automatically adjusts the O
signal. With the addition of a Single
2
Probe Sequencer (SPS), the Xi electronics can also switch the
calibration gases on and off.
4. Diagnostics - Multiple alarms are available for display. The alarm
displays are intended to assist a technician in locating where an
instrument problem may reside.
The Xi electronics has been verified to operate the following probes:
• Westinghouse 218 and World Class
• Rosemount Analytical Oxymitter
• Yokogawa
iv
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Technical Support
Hotline
For assistance with technical problems, please call the Customer Support
Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-60761-440-914-1261
In addition to the CSC, you may also contact Field Watch. Field Watch
coordinates Emerson Process Management’s field service throughout the
U.S. and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768)
Emerson Process Management may also be reached via the Internet through
e-mail and the World Wide Web:
e-mail: GAS.CSC@emerson.com
World Wide Web: www.raihome.com
A typical Rosemount Analytical O2 Combustion Flue Gas Transmitter should
contain the items shown in Figure 1-1. A complete Oxygen Analyzer system
will include some or all of the equipment shown. However, this manual
describes item 8 only. Record the part number, serial number, and order
number for the Xi Electronics in the table located on the back cover of this
manual.
Also, use the product matrix (Table 1-1) at the end of this section to compare
your order number against your unit. The first part of the matrix defines the
model. The last part defines the various options and features. Ensure the
features and options specified by your order number are on or included with
the unit.
This Instruction Manual is designed to supply details needed to install, start
up, operate, and maintain the Xi Electronics. Signal conditioning electronics
outputs a 4-20 mA signal representing an O
additional details, can be accessed with the handheld HART Model 375/475
Field Communicator or Asset Management Solutions (AMS) software.
value. This information, plus
2
http://www.raihome.com
Xi Advanced Electronics
Figure 1-1. Typical System Package
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
1. Instruction Manual
2. Weld Plate
3. Traditional Architecture Cable
4. O2 Probe
5. Reference Air Set (not used if SPS 4001B or IMPS 4000 is used)
6. HART® 375/475 Field Communicator Package (Optional)
7. Optional SPS 4001B or IMPS 4000 Autocalibration Sequencer
8. Xi Advanced Electronics
1-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
System ConfigurationsTraditional Architecture, HART and Xi Communications
Some customers prefer not to mount electronics onto the probe, so a
"traditional architecture" version is offered. This probe sends raw millivolt
signals via a 7-conductor cable to the Xi electronics, Figure 1-2, which does
all heater control and signal conditioning in addition to its display/keypad
functions. The Xi Advanced Electronics is offered to support direct replacement probes with either 120 volt or 44 volt heaters.
Figure 1-2. Direct Replacement Probe with Traditional Architecture Electronics
Figure 1-3. O
Probe with Xi Electronics and Autocalibration Sequencer
2
1-3
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Automatic Calibration
Communication Options
Calibrations consist of introducing bottled gases of known value into the probe
so the electronics can make automatic adjustments to the O2 readings to
match the bottled gas value. 0.4% O2 and 8% O2 (balance nitrogen) gases
are recommended. Never use nitrogen or instrument air as calibration gases.
Flowmeters (for calibration gases) and regulators and flowmeters (for
reference air) are available as loose components, mounted into an optional
manual calibration switching panel, or as a fully automatic calibration system,
Figure 1-3, where calibration solenoids are switched from the Xi Electronics.
See IM-106-340AC, SPS 4000B Single Probe Autocalibration Sequencer or
IM-106-400IMPS, IMPS 4000 Intelligent Multiprobe Test Gas Sequencer, for
additional details.
Data Communications
An operator can configure and troubleshoot the O2 Probe system in one of
two ways:
1. Using the Xi Advanced Electronics the Xi carries the following optional
advanced features:
• Fully automatic calibration
• Flame safety interface
• High temperature operation [above 700°C (1292°F) standard
temperature].
• Stoichiometer feature provides the ability to indicate O2 efficiency
when the combustion process goes into reducing conditions
(0% O2).
• Programmable reference provides enhanced accuracy when
measuring at or near O2 level (20.95% O2).
2. Using the HART Interface. The Xi’s 4-20 mA output line transmits an
analog signal proportional to the oxygen level. The HART output is
superimposed on the 4-20 mA output line. This information can be
accessed through the following:
• Rosemount Analytical Model 375/475 Field Communicator - The
handheld communicator requires Device Description (DD) software
specific to the Xi. The DD software will be supplied with many Model
375/475 units but can also be programmed into existing units at most
Emerson Process Management service offices. See Section 4,
Startup and Operation, for additional information.
• Personal Computer (PC) - The use of a personal computer requires
AMS software available from Emerson Process Management.
• Delta V and Ovation Distributed Control System (DCS) with
AMS-inside capability.
1-4
NOTE:
The 375 Field Communicator must be upgraded to System Software 2.0 with
Graphic License for operation with the Xi. The AMS software must be
upgraded to AMS 8.0 or above.
Contact Emerson Process Management’s Global Service Center (GSC) at
1-800-833-8314 to upgrade the 375 Field Communicator software to System
Software 2.0 with Graphic License.
Instruction Manual
38890063
Analytical
IM-106-910Xi, Original Issue
November 2010
Figure 1-4. Wireless THUM
Adapter
Xi Advanced Electronics
3. The Xi can also transmit HART information wirelessly via a wireless
THUM Adapter, Figure 1-4. The THUM Adapter threads into the Xi
conduit port and converts the 4-20 mA O2 signal to a wireless protocol.
All other HART information is also transmitted.
In addition to the wireless THUM Adapter, a hard-wire connection of the
4-20 mA signal to the DCS may be used at the same time. More
detailed information regarding the application of the THUM Adapter is
available in Product Data Sheet 00813-0100-4075. Note that the THUM
Adapter may also be used with the Oxymitter, mounted integral to the
probe, or to the Oxymitter remote electronics.
Figure 1-5. OxyBalance
Displays
Optional OxyBalance Display and Averaging System
Receives up to eight 4-20 mA signals from individual Xi units. Trends
individual outputs and calculates four programmable averages as additional
4-20 mA outputs. OxyBalance graphic displays are shown in Figure 1-5. See
IM-106-4050, OxyBalance Oxygen Display and Averaging System, for
additional details.
1-5
Xi Advanced Electronics
SPECIFICATIONS
Measurement Specifications
Net O2 Range:0 to 50% O2 user scalable
Lowest Detectable Limit:0.01% O
Signal Stability:±0.03% O
Accuracy in Reducing Conditions:±10% of reading or 0.1% O
System Response in Reducing
Conditions:
Ambient Temperature Effect on Xi
4-20 mA Signal:
Environmental Specifications
Xi Advanced Electronics:Type 4X/IP66, Polycarbonate Material
Ambient Temperature Limits:-20°C to 50°C (-4°F to 122°F)
Xi LCD display:
Ambient Temperature Limits-20°C to 55°C (-4°F to 131°F)
General Purpose Certifications:
Installation Specifications
Mounting:Panel, wall, or pipe.
Reference Air:0.5 scfh (0,25 /min), clean, dry, instrument-quality air
Calibration:Semi-automatic or automatic
Cal Gases:0.4% O2 and 8% O2, balance N
Traditional Architecture Cable200 ft (61 m) maximum length
Transmitter Electrical Power:
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
-2 to 50% O2 user scalable with stoichiometer
2
2
going from oxidizing to reducing -T90 in 120 seconds
going from reducing to oxidizing -T90 in 30 seconds
less than 0.0025% O2 per degree Celsius
-20°C to 70°C (-4°F to 158°F) as measured by
electronics
(20.95% O2), regulated to 5 psi (34 kPa)
12 - 24 VDC (loop-powered from control room or Xi)
2
2
1-6
Electrical Power for Xi:100-240VAC ±10%, 50/60 Hz
Power Consumption of Xi:12 VA maximum or
776 VA maximum with Traditional Architecture,
120V Probes
450VA maximum with Traditional Architecture,
44V Probes
Alarm Relay Outputs: Two provided - 2 Amperes, 30 VDC, Form-C
Optional Loss of Flame Input:Internally powered input to remove heater power
actuated via dry contact output from user’s* flame
scanner
Emerson Process Management has satisfied all obligations from the
European legislation to harmonize the product requirements in Europe. 1All static
performance characteristics are with operating variables constant. Specifications
subject to change without notice.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Table 1-1. Product Matrix, Xi Advanced Electronics
XiXi Advanced Electronics
CodeRemote Type
01Single Channel
02Single Channel, accepting a loss-of-flame input to remove heater power with flame status relay
03Dual Channel
04Single Channel Traditional Architecture for 120V probes
05Single Channel Traditional Architecture for 44V probes
CodeMounting
00No Hardware
01Panel Mount Kit with Gasket
022" Pipe/Wall Mount Kit
Xi010100010101Example
(1)
(1)
CodeCable
00No Cable
1020’ (6 m) Cable
1140’ (12 m) Cable
1260’ (18 m) Cable
1380’ (24 m) Cable
14100’ (30 m) Cable
15150’ (45 m) Cable
16200’ (60 m) Cable
CodeStoichiometer Function
00None
01Single Channel (Stoichiometer cell also required in probe)
02Dual Channel (Stoichiometer cell also required in probe)
00None
01SPS 4001B Single Probe Sequencer, general purpose NEMA 4X, includes check valve for probe
CodeIntelligent Multiprobe Sequencers (IMPS)
00None
01IMPS single-probe, general purpose NEMA 4X, includes check valve for probe
02IMPS two-probe, general purpose NEMA 4X, includes check valve for probe
03IMPS three-probe, general purpose NEMA 4X, includes check valve for probe
04IMPS four-probe, general purpose NEMA 4X, includes check valve for probe
05IMPS single-probe, 115V heated general purpose NEMA 4X, includes check valve for probe
06IMPS two-probe, 115V heated general purpose NEMA 4X, includes check valve for probe
07IMPS three-probe, 115V heated general purpose NEMA 4X, includes check valve for probe
08IMPS four-probe, 115V heated general purpose NEMA 4X, includes check valve for probe
09IMPS single-probe, 220V heated general purpose NEMA 4X, includes check valve for probe
10IMPS two-probe, 220V heated general purpose NEMA 4X, includes check valve for probe
11IMPS three-probe, 220V heated general purpose NEMA 4X, includes check valve for probe
12IMPS four-probe, 220V heated general purpose NEMA 4X, includes check valve for probe
O2CAL0004Example
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Table 1-3. Calibration Gases
Part NumberDescription
1A99119G01Two disposable calibration gas bottles - 0.4% and 8% O2, balance
nitrogen - 550 liters each*
1A99119G02Two flow regulators for calibration gas bottles
1A99119G03Bottle rack
Note:
*Calibration gas bottles cannot be shipped via airfreight.
Before installing this equipment read the "Safety instructions for the wiring and installation of
this apparatus" at the front of this Instruction Manual. Failure to follow safety instructions
could result in serious injury or death.
Xi Advanced Electronics
Install all protective equipment covers and safety ground leads after installation. Failure to
install covers and ground leads could result in serious injury or death.
The Xi Advanced Electronics can be installed in general purpose areas only. Do not install
the Xi in hazardous areas or in the vicinity of flammable liquids.
If external loop power is used, the power supply must be a safety extra low voltage (SELV)
type.
NOTE
All unused ports on the probe housing and Xi enclosure should be plugged
with a suitable fitting.
http://www.raihome.com
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
SYSTEM
CONSIDERATIONS
Figure 2-1. Typical System
Installation
A typical system installation for a Xi and O2 Probe is shown in Figure 2-1.
MECHANICAL
INSTALLATION
Xi Advanced Electronics
The Xi Advanced Electronics is available in a panel mounting, wall mounting,
or pipe mounting configuration. Refer to Figure 2-2 or Figure 2-3 for the panel,
wall, or pipe mounting details.
1. Ensure all components are available to install the Xi.
2. Select a mounting location near or removed from the O
Consider the temperature limitations of the Xi (see "Specifications")
when selecting the mounting location.
3. Mount the Xi at a height convenient for viewing and operating the
interface. Approximately 5 ft (1,5 m) is recommended.
4. The keypad window on the Xi may have interior and exterior protective
membranes. Remove the protective membranes prior to use of the Xi
enclosure. Failure to remove the protective membranes may cause the
display to appear distorted. The membrane may be difficult or
impossible to remove after extended use at elevated temperatures.
Probe.
2
2-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Figure 2-2. Xi Advanced Electronics - Panel Mounting Details
Xi Advanced Electronics
2-3
Xi Advanced Electronics
Figure 2-3. Xi Advanced Electronics - Wall/Surface and Pipe Mounting Details
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
2-4
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
ELECTRICAL
INSTALLATION
Xi Advanced Electronics
All wiring must conform to local and national codes. Multiple wiring diagrams
are shown in this section. Always refer to the diagrams that apply to your
transmitter configuration and disregard all other wiring diagrams.
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers
and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation
of this equipment, connection to the main electrical power supply must be made through a
circuit breaker (min 10A) which will disconnect all current-carrying conductors during a fault
situation. This circuit breaker should also include a mechanically operated isolating switch.
If not, then another external means of disconnecting the supply from the equipment should
be located close by. Circuit breakers or switches must comply with a recognized standard
such as IEC 947.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105
°C
(221°F).
NOTE
If metal conduit is used with the Xi the conduit should be reliably bonded to
protective earth. The grounding plate inside the Xi is not bonded to PE and
does not provide adequate grounding.
1. Remove cover screws from the front cover of the Xi. Swing down the
front cover of the interface box.
2. Pull out the I/O board on the right-hand side of the card rack inside the
Xi. If your system is configured to operate two transmitter probes there
are two I/O interface boards.
3. See Figure 2-4. Connect the 4-20 mA signal wires at J4 of the I/O
board. Attach the supplied ferrite clamp over the 4-20 mA OUT wires
that extend past the shield.
NOTE
Installation of the ferrite clamp over the 4-20 mA OUT wires is required for
compliance with the European EMC Directive.
2-5
Xi Advanced Electronics
Figure 2-4. Signal Connections at I/O Board
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Figure 2-5. Alarm Indicator Relay Terminals
4. Terminate the shield of the 4-20 mA signal wires at the designated
5. Connect the signal wires from the SPS or IMPS (if used) to the
6. Connect the signal wires for the flame status input (if used) to the
7. Connect the customer’s alarm indicator devices to the alarm indicator
8. Reinstall the I/O board in the card rack of the Xi.
ground terminal of the Xi. Do not allow bare shield wires to contact the
circuit boards. Insulate the shield wires prior to termination.
applicable terminals of J3. Refer to the SPS or IMPS instruction manual
for wiring details.
applicable terminals of J2. The flame status sensing device is supplied
by the customer. Refer to the applicable OEM documents for signal
wiring details.
relay terminals. See Figure 2-5 for the alarm indicator relay terminals.
2-6
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Figure 2-6. Power Connections Xi Advanced Electronics
Xi Advanced Electronics
Optional Flame Safety
Interlock
Traditional Architecture
Cable Connections
9. If your system is configured for two channel operation, repeat steps 2
through 7 to connect the other signal wires.
10. Remove the connector from the power supply board located on the
left-hand side of the card rack inside the Xi.
11. See Figure 2-6. Connect the line, or L1 wire to the L1 terminal and the
neutral, or L2 wire, to the N terminal.
12. Reinstall the power supply connector in the power supply board.
13. Close and fasten the Xi cover.
A flame safety interlock by Emerson Process Management is available for
heater power disconnect whenever there is a loss of the process flame or a
heater runaway condition (heater over-temperature) in the O
simplified wiring diagram for the flame safety interlock is shown in Figure 2-7.
This input is internally powered by the Xi and is actuated via a dry contact
output from the user’s flame scanner. A closed contact indicates a flame is
present. An open contact indicates a loss of flame.
A traditional architecture configuration is used to provide for remote location
of the transmitter electronics. All electronics are housed inside the Xi. A
multi-conductor power/signal cable connects between the probe and the Xi.
Use the following procedure to connect the traditional architecture probe to
the Xi.
NOTE
The Traditional Architecture cable is provided at the specified length and is
ready for installation. The cable glands must be properly terminated to
maintain EMC/EMI noise protection.
Probe. A
2
1. Run the 7-conductor cable between the traditional architecture probe
and the installation site for Xi. Use new cable conduit or trough as
needed.
Install all protective equipment covers and safety ground leads before equipment startup.
Failure to install covers and ground leads could result in serious injury or death.
If external loop power is used, the power supply must be a safety extra low voltage (SELV)
type.
http://www.raihome.com
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Configuration
Refer to Figure 3-1 for the configuration of jumpers JP1 through JP8. The
jumper configuration for your I/O board depends on the system design and
system components used in your installation.
The setting of switch SW4 and the configuration of jumpers JP1 through JP8
must be verified on the I/O board in the Xi. All four dip switches on switch
SW4 must be set to the OFF position, as shown.
Figure 3-1. I/O Board Jumper Configuration
SET TEST GAS VALUES
3-2
Use a Field Communicator or the Xi to set test gas values for calibration.
A Xi shipped from the factory has test gas values for low and high set to 0.4%
and 8.0% respectively. This same process must be performed any time a
replacement Transmitter Board, I/O Board or DR Board is installed.
Setting Test Gas Values Using Xi Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select CAL SETUP.
5. From CAL SETUP, select Cal Gas 1. Enter the percent O
6. From CAL SETUP, select Cal Gas 2. Enter the percent O
low O
high O
test gas.
2
test gas.
2
used for the
2
used for the
2
7. Press the LEFT arrow key several times to return to the main menu.
8. Repeat steps 2 through 6 for PROBE 2 if configured for dual channel.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Setting Test Gas Values Using Field Communicator
1. Use the Field Communicator software to access the HART menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select CAL SETUP.
4. From CAL SETUP, select Cal Gas 1. Enter the percent O2 used for the
5. From CAL SETUP, select Cal Gas 2. Enter the percent O2 used for the
low O
test gas.
2
high O2 test gas.
ALARM RELAY OUTPUT
CONFIGURATION
Table 3-1. Alarm Relay Output
Configurations
The Xi has two dry contact Form-C alarm relay output signals that can be
configured in eight different modes through the Xi keypad display or the
375/475 Field Communicator. A list of possible configurations is shown in
Table 3-1. Each alarm relay output can be configured separately.
If the Xi is configured with the optional Flame Safety Interlock, Alarm 2 is
configured with "Heater Relay" and prewired to the AC Relay Board. In this
condition the relay configuration cannot be changed to any other setting. If
the Xi is not configured with the optional Flame Safety Interlock, "Heater
Relay" is not valid and cannot be chosen for Alarm 2.
ModeConfiguration
No Alarm*The output is not configured for any alarm condition.
Unit AlarmThe output is configured for a Unit Alarm.
Low O2 AlarmThe output is configured for a Low O2 alarm.
Low O2/Unit AlmThe output is configured for a Unit alarm and a Low O2 alarm.
Cal RecommendedThe output is configured for a Calibration Recommended
display.
Cal Rec/Unit Alm**The output is configured for a Unit alarm and a Calibration
Recommended display.
Low O2/Cal RecThe output is configured for a Low O2 alarm and a Calibration
Recommended display.
Low O2/Unit/Cal RecThe output is configured for a Low O2 alarm, a Unit alarm, and a
Calibration Recommended display.
Heater RelayThe output is configured for Flame Safety Interlock.
* The default configuration for Alarm 2
** The default configuration for Alarm 1
Configuring Alarm Relays with the Xi Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select ALARM RELAY.
5. From ALARM RELAY, select as follows:
Alm Relay1 - Alarm 1 mode
Alm Relay2 - Alarm 2 mode
Low O2 Alm SP - Low O2 alarm setpoint
High Temp Alm SP - High temperature alarm setpoint
6. Press the LEFT arrow key several times to return to the main menu.
7. Repeat steps 2 through 6 for PROBE 2 if configured for dual channel.
3-3
Xi Advanced Electronics
Configuring Alarm Relays with the Field Communicator
1. Use the 375/475 Field Communicator software to access the HART
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select ALARM RELAY.
4. From ALARM RELAY, select as follows:
5. From CAL SETUP, select Cal Gas 2. Enter the percent O2 used for the
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
menu.
Alm Relay1 - Alarm 1 mode
Alm Relay2 - Alarm 2 mode
Low O2 Alm SP - Low O2 alarm setpoint
High Temp Alm SP - High temperature alarm setpoint
high O2 test gas.
ANALOG OUTPUT
CONFIGURATION
The analog output signal from the Xi can be configured for the 4-20 mA range
and fault condition.
A Xi shipped from the factoryhas the analog outputs set to a 4 to 20 mA
range with a 3.5 mA alarm level. This same process must be performed any
time a replacement Transmitter Board or I/O Board is installed.
Configuring the Analog Output with the Xi Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select ANALOG OUTPUT.
5. From ANALOG OUTPUT, set the following parameters:
O2 LRV - O2 value at the lower analog output value (0 mA or 4 mA)
O2 URV - O2 value at the upper analog output value (20 mA)
AO Range - Range of the analog output (0-20 mA or 4-20 mA)
Signal Alarm Level - O2 alarm level (3.5 mA or 21.1 mA)
Configuring the Analog Output with the Field Communicator
1. Use the 375/475 Field Communicator software to access the HART
menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select ANALOG OUTPUT.
4. From ANALOG OUTPUT, set the following parameters:
3-4
O2 LRV - O2 value at the lower analog output value (0 mA or 4 mA)
O2 URV - O2 value at the upper analog output value (20 mA)
O2 AO Range - Range of the analog output (0-20 mA or 4-20 mA)
Signal Alarm Level - O2 alarm level (3.5 mA or 21.1 mA)
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
AUTOCALIBRATION
SETUP
If autocalibration is desired, the Xi must be used with either an SPS 4001B or
IMPS 4000. The Xi must be properly configured before autocalibration can
take place. Refer to the applicable SPS 4001B or IMPS 4000 instruction
manual for details on performing autocalibration. Refer to Section 4, Startup
and Operation for details on manual calibration procedures.
A Xi is shipped from the factory without autocalibration configured. This same
process must be performed any time a replacement I/O Board is installed.
Configuring Autocalibration with the Xi Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select CAL SETUP.
5. From CAL SETUP, select Auto Cal.
6. Press the RIGHT arrow key to change the state from NO to YES.
Configuring Autocalibration with the Field Communicator
1. Use the 375/475 Field Communicator software to access the HART
menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select CAL SETUP.
4. From CAL SETUP, select Auto Cal.
5. Press the RIGHT arrow key to change the state from NO to YES.
3-5
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
OPTIONAL ADVANCED
FEATURES INSIDE
THE Xi
Advanced features available inside the Xi are typically ordered as part of the
initial package. However, these advanced features are also available for field
retrofit.
A Xi is shipped from the factory with the optional enhanced software features
enabled based on the configuration.
The I/O Board is shipped from the factory without any of the enhanced software features
activated. These features must be activated once the new board has been installed and
before the Remote Interface is put into service.
If the existing I/O Board has been operated with the Stoichiometric enhanced software
feature, this feature must be activated in the new board before the Remote Interface is put
back into service. Failure to do so will cause a false analog output signal to the DCS.
NOTE
For enhanced software feature option upgrades or to enable the feature to
duplicate the existing configuration, contact Emerson Process Management
at 1-800-433-6076. Reference the following:
Extended Process
Temperature Range to
800°C (1472°F)
6A00269G03Enhanced Software Option Upgrade, 800°C Process
Function
The Oxygen Analyzer employs a heater and thermocouple to maintain a
temperature setpoint at 736°C (1357°F). Temperature control is maintained
within ±1°C to process temperatures of about 705°C (1300°F). This is
satisfactory for most applications, but excursions to higher temperatures can
occur in some processes. In these instances, the heater is turned off and the
process temperature is used to heat the sensing cell.
The oxygen reading is adjusted immediately to compensate for the varying
process temperatures. It should be noted that cell life will be reduced by
continuous operation at temperatures above 705°C (1300°F). If process
temperatures are expected to be continuously above 705°C, we recommend
the use of an optional bypass or probe mounting jacket accessory. The
extended temperature range feature is selected in the Xi product matrix, but
may also be purchased as a field retrofit.
3-6
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Stoichiometer
Programmable
Reference
Process upsets can sometimes cause a combustion process to go into
sub-stoichiometric or reducing conditions. The oxygen readings from one or
more probes may decline all the way to zero. The stoichiometer cell will
measure the amount of oxygen deficiency during these reducing conditions.
The trends in your DCS can be set up for a lower range limit of -1 or -2%
oxygen to depict the level of oxygen deficiency.
The operator can see if his control recovery actions are having the desired
effect. These types of events do not occur frequently, but knowing the
parameters of the situation prevents over-correcting while coming out of the
reducing condition.The stoichiometer feature requires purchasing the acid
resistant stoichiometer cell and the stoichiometer feature inside the Xi.
NOTE
Make sure the DCS is configured for the same range as the Xi. For instance:
-1% O2 to 10% O2.
The zirconium oxide sensing technology has historically measured process
oxygen by using ambient or instrument air as a reference (20.95% oxygen).
The sensor develops most of its signal at the low oxygen levels typically found
in combustion flue gasses (2-4% oxygen), and is most accurate at these
levels. When measuring near 20.95% O2, the sensor develops only a few
millivolts of signal, and accuracy degrades.
The programmable reference feature permits the user to use a bottled
reference gas of low oxygen value (0.4% oxygen recommended). When
measuring at or near 21% oxygen, a strong negative oxygen signal results,
with much improved accuracy. A bottle of reference gas typically lasts about a
month at the low flows required. Typical applications would be:
Flue gas recirculation - controlling the mixing of flue gasses into the burner
windbox prior ahead of the burner to reduce NOx emissions.
Moisture monitoring - measuring the amount of moisture coming off of
industrial dryers by noting the dilution effect water vapor has on the normal
20.95% ambient drying air. (Non-combustion drying processes only.)
Enriched oxygen concentration - pure oxygen is sometimes mixed in with
the combustion air to increase heat at the flame. This is used in steel and
other metals reduction processes and in some catalyst regenerators.
Interface to the Xi for setup, calibration and diagnostics can be via a 375/475
Field Communicator or Asset Management System.
Setup, calibration and diagnostic operations will differ depending on the
selected interface for communications with the transmitter.
The O2 Probe will take approximately 45 minutes to warm up to the 736°C
heater setpoint. The 4-20 mA signal will remain at a default value of 3.5 mA
through this warm-up period. Once warm, the O2probe will be reading
oxygen, and the 4-20 mA signal will be reading based on the default range
of 0-10% O2.
NOTE
The Xi offers optional advanced features such as elevated process
temperature capability to 800°C, autocalibration via an SPS solenoid box, a
stoichiometer feature for indicating the level of oxygen deficiency in reducing
conditions, and programmable reference to enhance accuracy at near
ambient levels of O2.
http://www.raihome.com
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Operation via Xi
Startup Display
Figure 4-1. Xi Display (Typical)
The following procedures describe operations using the Xi to set up and
calibrate the system. Additional operating instructions are included in the SPS
4001B or IMPS 4000 instruction manual, if applicable to your system.
The O2 Probe will take approximately 45 minutes to warm up to the 736°C
heater setpoint. The 4-20 mA signal will remain at a default value of 3.5 mA
through this warm-up period. Once warm, the probe will be reading oxygen
and the 4-20 mA signal display will be the 0 to 10% O2 value.
Error Conditions
Xi Controls
If there is an error condition at startup, an alarm message will be displayed.
Refer to Section 5: Troubleshooting, to determine the cause of the error. Clear
the error and cycle power. The %O2 and temperature display should return
less the alarm message.
The Xi can be used to change the software and alarm settings, to adjust the
high and low gas settings, and to initiate the calibration sequence. Refer to
the following control descriptions. Use the control keys on the front panel of
the Xi, Figure 4-1, to navigate and edit the Xi menu, Figure 4-2.
MENU toggles between three Main menu options: System, Probe1, and
Probe2 (if available). The top level of the selected main menu is displayed.
DIAG toggles between the Alarms list of the three main menus. All faults and
warnings related to the selected main menu device are displayed.
ENTER saves newly entered data and returns you to previous menu level.
EXIT returns you to the previous menu level without saving newly entered
data. When navigating the menu tree, pressing EXIT returns you to the Main
menu.
UP/DOWN keys scroll up and down through menu items. During data entry
the Up/Down keys increment and decrement the data values.
LEFT arrow key returns you to the previous menu level. During data entry, the
left arrow key moves the cursor one digit to the left.
4-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
RIGHT arrow key advances you to the next menu level and, when a menu
item is highlighted, selects the item from a list of menu options. During data
entry, the right arrow key moves the cursor one digit to the right.
Password Protection
Beginning with Xi system software version 1.05 or higher the main display
and diagnostic screens of the Xi can be viewed at any time, but further access
and unauthorized configuration changes can be prevented by enabling a
password protection feature. However, the Xi is shipped with password
protection disabled.
Password protection can be enabled by selecting: System Main Menu>
Configure> LCD>Enable Password (see the Xi Menu, Figure 4-2).
The factory default upon enabling the password protection is ROSE, but the
password can consist of any 4 alpha/numeric characters.
If the user forgets the password, call Rosemount Analytical technical support
at 800-433-6076 to gain access to a master password.
A "Lock" icon will be displayed at the top right corner of the main display when
password protection is in effect.
The password protection will relock itself after a certain number of seconds
with no button pushes (defined as "revert time" in the same "LCD setup"
menu). Users can also force the front panel to be locked by selecting System
Main Menu >Log Off. The Log Off selection will perform no function if the
password feature is disabled.
The Xi has a "Reset" function that reestablishes all factory default conditions,
including the password protection feature, i.e. the password protection will fall
back to a disabled condition after a reset.
4-3
Xi Advanced Electronics
Figure 4-2. Xi Menu (Sheet 1 of 3)
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
4-4
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Figure 4-2. Xi Menu (Sheet 2 of 3)
Xi Advanced Electronics
4-5
Xi Advanced Electronics
Figure 4-2. Xi Menu (Sheet 3 of 3)
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
4-6
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
SYSTEM PARAMETER
DESCRIPTIONS
Among the parameters available through the Xi and 375/475 Field
Communicator menus are a number of "System Parameters". The system
parameters define variables that configure the Xi in the transmitter system.
System parameters are described in the following table.
CPU
YVersion--Software version number for the CPU board.
YBuild Num--Software build number for the CPU board.
YComm Status--Communication status between the CPU and I/O boards.
YRestart Cntr--Software restarts counter for the CPU board.
NPoll Addr--Polling Address - Address used to identify a Field Device; changeable by the user to control.
NSerial Number--I/O board serial number.
NFeatures--Available advanced software features for the I/O board.
NEE Erase Count--This is for nonvolatile memory diagnostic only.
YNum Probe--Number of probes configuration:
YProbe Sel--This parameter is applicable only if the Num Probe configuration is set to 2. It defines which
YProbe 1 Enable--Probe 1 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
YProbe 2 Enable--Probe 2 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
YLineX Probe--Probe number for line x. (1, 2)
YLine 1 Data--Main display, line 1 data: 0 displays PV (O2), 1 displays SV (O2 Temp)
YLine 2 Data--Main display, line 1 data: 0 displays PV (O2), 1 displays SV (O2 Temp)
YLine 3 DataL--Main display, line 3 left hand side data:
YLine 3 DataR--Main display, line 3 right hand side data:
YLine 4 DataL--Main display, line 4 left hand side data:
YLine 4 DataR--Main display, line 4 right hand side data:
YRevert TimeminXi display reverts to main display time. Also locks screen if password protection is enabled.
YLanguage--0 displays English
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
4-7
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
PROBE PARAMETER
DESCRIPTIONS
Among the parameters available through the Xi and 375/475 Field
Communicator menus are a number of "Probe Parameters". The probe
parameters define variables that configure a specific probe in the transmitter
system. Probe parameters are described in the following table.
TXI/O
YYO2%Current oxygen concentration value (O2%). The value should reflect the last good O2 value if it
NYElec TempdegCCurrent electronic temperature measured at the I/O board.
YYO2 CellmVRaw mV value for ZrO2 sensor.
NYTC VoltmVO2 T/C voltage.
YYCell ImpOhmCell impedance/sensor resistance measured.
YYHtr VoltVoltHeater voltage.
YYCPU VoltVoltTransmitter CPU voltage.
YYO2 AOmAAnalog output value represents the O2 concentration measurement.
NYO2 AO%%O2 analog output percentage for O2 AO.
YYO2 Temp MaxdegCThis is the highest O2 sensor temperature reached since last reset.
YYCJC Temp MaxdegCThis is the highest temperature reached at the cold junction since last reset.
NYElec Temp MaxdegCThis is the highest temperature reached at the I/O board since last reset.
YYHtr Volt MaxdegCThis is the highest heater voltage reached since last reset.
YYHtr Duty Cycle--O2 heater duty cycle. Value between 0 and 1.
YYPID SPdegCPID temperature set point.
YYHtr Ramp RatedegC/s Heater ramp rate calculated in degree C per second.
NYFlame Stat In--Flame status input state.
NYSPS/IMPS In--SPS/IMPS input state.
NYSPS/IMPS Out--SPS/IMPS output state.
NYAlm Relay 1 Out--Alarm Relay 1 output state.
NYAlm Relay 2 Out--Alarm Relay 2 output state.
YYOP Mode--Device operating mode:
YYTag--Device tag.
YYDevice ID--Unique Device ID number. (HART)
YYPV is--Primary variable assignment. (HART)
YYSY is --Secondary variable assignment. (HART)
YYTY is--Third variable assignment. (HART)
YYQV is--Fourth variable assignment. (HART)
YYCal SlopemV/Dec Current calibration slope. This is the slope value that was calculated as a result of the last
YYCal ConstmVCurrent calibration constant. This is the constant value that was calculated as a result of the last
YYCal ImpOhmCell Impedance. This is the sensor resistance that was calculated as a result of the last
NYPrev SlopemV/Dec Previous calibration slope. There are ten calibration results. 1 is the most recent and 10 is the
NYPrev ConstmVPrevious calibration constant. There are ten calibration results. 1 is the most recent and 10 is
PARAMETER
NAMEUNITSPARAMETER DESCRIPTION
is in the "Lock" state during calibration.
(OFF/ON)
(OFF/ON)
(OFF/ON)
(OFF/ON)
(OFF/ON)
PO=Power up; WU=Warm Up (analog output is railed); NM=Normal operation;
CA=Calibrating (analog output can be tracking or locked at last good value based on
"AO Tracks" configuration); AL=Alarm detected (recoverable); SF=Alarm detected
(non-recoverable)
successful calibration.
successful calibration. It is valid between -4mV and +10mV.
successful calibration.
least recent calibration slope.
the least recent calibration constant.
4-8
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
TXI/O
NYPrev Cal ImpOhmPrevious Cell Impedance. This is the sensor resistance that was calculated as a result of
NYFailed SlopemV/Dec Failed calibration slope.
NYFailed ConstmVFailed calibration constant.
YYCal ResultmVCalibration result.
NYDelta Imp--Delta impedance since last calibration.
NYCal Step--This represents the step of the calibration cycle is in.
NYTime RemainsecTime remaining in the present calibration cycle state.
YYO2 SlopemV/Dec O2 slope. This is the slope value that will be used to calculate O2.
YYO2 ConstmVO2 constant. This is the constant value that will be used to calculate O2.
NYT90 FiltersecAnalog output T90 time. It represents the time to take a step change in oxygen to reach 90% of
NYO2 Cell Ref%O2 sensor reference gas percentage. It allows using a sensor reference gas other than air.
NYO2 Temp SPdegCO2 sensor temperature set point. It allows measurement of oxygen with an elevated sensor
NYAlarm Relay 2--Alarm Relay 2 mode. (no alarm; unit alarm; low O2 alarm; low O2/unit alarm
NYLow O2 Alm%Low O2 alarm threshold.
NYHi Temp Alm%High temperature alarm threshold.
NYAO Tracks--Analog output track O2 sensor measurement during a calibration. (No, Yes).
NYAuto Cal--Enable/disable automatic calibration.
NYStart On CalRec--Start automatic calibration on Cal Recommended state. (No, Yes)
YYCal Gas 1%Test Gas 1 value. This is the actual value of the gas being applied during the Test Gas 1 phase
YYCal Gas 2%Test Gas 2 value. This is the actual value of the gas being applied during the Test Gas 2 phase
YYGas TimesecTest Gas application time. This is the length of time test gases are applied to the O2 probe
YYPurge Time secTest Gas purge time. This is the length of time before the output will be returned to the process
NYCal IntervalhrAutomatic calibration interval. The number 9999 disables the automatic timed calibration.
NYNext Cal TimehrTime remaining until the next automatic periodic calibration. The number 9999 disables the next
YYT EE Val--Transmitter board nonvolatile memory diagnostic.
NYIO EE Val--I/O board nonvolatile memory diagnostic.
YYVersion--Software version number for the Transmitter.
NYVersion--Software version number for the I/O board.
YYT Restart Cntr--Software restarts count for the Transmitter.
NYIO Restart Cntr--Software restarts count for the I/O board.
YYAlarms--Current Alarms (See section 5, Troubleshooting, ”Alarm Indications”.)
PARAMETER
NAMEUNITSPARAMETER DESCRIPTION
previous successful calibration. There are ten calibration results. Index 1 is the most recent and
Index 10 is the least recent sensor resistance measured.
the final value at the filter output.
temperature. (0=736 degC set point; 1=834 degC set point)
(no alarm; unit alarm; low O2 alarm; low O2/unit alarm;
Cal recommended; Cal recommended/unit alarm; low O2/Cal recommended;
low O2/unit alarm/Cal recommended)
Cal recommended; Cal recommended/unit alarm; low O2/Cal recommended;
low O2/unit alarm/Cal recommended; Heater relay)
No = Set to Manual calibration mode.
Yes = Set to Automatic calibration mode.
of a calibration.
of a calibration.
during low or high Test Gas phase of a calibration.
reading after a calibration.
automatic timed calibration.
4-9
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
OPERATION VIA
HART/AMS
Field Communicator
Signal Line Connections
The 375/475 Field Communicator is a handheld communications interface
device. It provides a common communications link to all microprocesor-based
instruments that are HART compatible. The handheld communicator contains
a liquid crystal display (LCD) and 21 keys. A pocket-sized manual, included
with the 375/475 Field Communicator, details the specific functions of all the
keys.
The 375/475 Field Communicator accomplishes its task using a frequency
shift keying (FSK) technique. With the use of FSK, high-frequency digital
communication signals are superimposed on the Xi's 4-20 mA current loop.
The 375/475 Field Communicator does not disturb the 4-20 mA signal, since
no net energy is added to the loop.
NOTES
The 375 Field Communicator must be upgraded to System Software 2.0 with
Graphic License for operation with the Xi. The AMS software must be
upgraded to AMS 8.0 or above for operation with the Xi.
Contact Emerson Process Management’s Global Service Center (GSC) at
1-800-833-8314 to upgrade the 375 Field Communicator software to System
Software 2.0 with Graphic License.
When working at the Xi, the 375/475 Field Communicator can be connected
directly to test points TP21 and TP22 on the Xi I/O Board as shown in
Figure 4-3. The AM+ and AM- test points are provided to monitor the 4-20 mA
signal without breaking into the loop.
Figure 4-3. 375/475 Field Communicator Connection at the Xi
4-10
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Field Communicator
Menu Trees
OFF-LINE AND ON-LINE
OPERATIONS
Connect the 375/475 Field Communicator in the Xi (Xi-to-DCS) 4-20 mA
signal loop or to the Xi terminals as shown in Figure 4-3 and refer to
Figure 4-4 for the 375/475 Field Communicator Xi menu tree.
The 375/475 Field Communicator can be operated both off-line and on-line.
Off-line operations are those in which the communicator is not connected to
the O2 Probe. Off-line operations can include interfacing the 375/475 Field
Communicator with a PC (refer to applicable HART documentation regarding
HART/PC applications.)
In the on-line mode, the 375/475 Field Communicator is connected to the
4-20 mA analog output signal line. The communicator is connected in parallel
to the O2 Probe or in parallel to the 250 ohm load resistor.
4-11
Xi Advanced Electronics
Figure 4-4. 375/475 Field Communicator Xi Menu Tree (Sheet 1 of 2)
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
4-12
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Figure 4-4. 375/475 Field Communicator Xi Menu Tree (Sheet 2 of 2)
Xi Advanced Electronics
4-13
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
CALIBRATION GENERAL
O2 CALIBRATION
New O2 cells may operate for more than a year without requiring calibration,
but older cells may require recalibration every few weeks as they near the end
of their life.
A
CALIBRATION RECOMMENDED
is required. This strategy ensures that the O2 reading is always accurate and
eliminates many unnecessary calibrations based on calendar days or weeks
since previous calibration.
The O2 Probe(s) can be calibrated manually through the handheld 375/475
Field Communicator or the Xi. Fully automatic calibration can be performed
automatically using the Xi and the SPS 4001B Single Probe Autocalibration
Sequencer or the IMPS 4000 Intelligent Multiprobe Sequencer.
This section covers manual calibration. For automatic calibration details see
the Instruction Manual for the SPS 4001B Single Probe Autocalibration
Sequencer or the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer.
Calibration can be performed using three basic calibration methods. The
methods available to you for use depend on the configuration of your system.
The paragraphs that follow describe how to perform a calibration for three
basic system configurations shown in Figure 4-5.
Included in the calibration procedures are instructions for setting up the calibration parameters. Setup of the calibration parameters should be performed
before the first O2 calibration. Thereafter, perform calibration setup only as
needed to change the calibration parameters or to reset the parameters following the replacement of primary system components.
alarm provides notice of when a calibration
Figure 4-5. Calibration Methods,
Simplified
O2 Calibration with Xi
Before calibrating verify that the configuration gas parameters are correct.
Refer to Section 3: Configuration of Xi Electronics.
For systems with configuration 1, shown in Figure 4-5, use the following procedure to perform a calibration using the Xi. If necessary, use the Xi menu
tree in Figure 4-2 for reference.
4-14
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key to select the menu item. To return to
a preceding menu press the left arrow key.
1. From the Main Menu, select SYSTEM, to access the Xi System menu.
2. From the Xi SYSTEM menu, scroll down and select DETAILED SETUP.
3. Select menu item 4, CAL SETUP, to input the cal gas and flow times.
4. Return to the SYSTEM menu and select the last menu item,
CALIBRATION, to access the CALIBRATION menu.
5. From the O2 CALIBRATION options, select Start Cal to start the O2
calibration procedure.
Failure to remove the Xi from automatic control loops prior to performing this procedure may
result in a dangerous operating condition.
O2 Calibration with Xi
and Field Communicator
6. In the first Start Cal screen, a "Loop should be removed from automatic
control" warning appears. Remove the Xi from any automatic control
loops to avoid a potentially dangerous operating condition and press
OK.
7. Follow the Xi display prompts to perform the O2 cal procedure.
For systems with configuration 2, shown in Figure 4-5, use the following procedure to perform a calibration of the system using the 375/475 Field Communicator. If necessary use the menu tree in Figure 4-4 for reference.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number. To return to a preceding menu, press the left arrow
key.
1. Select DEVICE SETUP.
2. From the DEVICE SETUP screen select menu item 4, DETAILED
SETUP.
3. Select menu item 3, CAL SETUP, to input the cal gas and gas flow
times.
4. Return to the DEVICE SETUP screen and select menu item 2,
DIAG/SERVICE.
5. From the DIAG/SERVICE screen, select menu item 3, CALIBRATION,
to access the O2 CALIBRATION screen.
6. From the O2 CALIBRATION screen, select menu item 1, O2 CAL, to
access the O2 calibration procedure.
4-15
Xi Advanced Electronics
Failure to remove the Xi from automatic control loops prior to performing this procedure may
result in a dangerous operating condition.
7. In the first O2 CAL screen, a "Loop should be removed from automatic
8. Follow the handheld 375/475 Field Communicator display prompts to
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
control" warning appears. Remove the Xi from any automatic control
loop to avoid a potentially dangerous operating condition and press OK.
perform the O2 cal procedure.
D/A TRIM
Figure 4-6. D/A Trim Methods,
Simplified
The D/A trim procedure is used to calibrate the 4-20 mA output signal to a
precision mA measurement device.
The two paragraphs that follow describe how to perform a D/A trim for two of
the three basic system configurations shown in Figure 4-10. Only the signal to
the DCS needs to be trimmed.
D/A Trim with Xi
4-16
For systems with configuration 1, shown in Figure 4-6, use the handheld
375/475 Field Communicator to access the D/A trim procedure according to
the instructions that follow. Refer to the 375/475 Field Communicator Xi Menu
Tree in Figure 4-2.
1. From the DEVICE SETUP screen select menu item 2, DIAG/SERVICE,
to access the diagnostics and service menu options.
2. Select menu item 3, CALIBRATION, to access the calibration menu
options.
3. Select menu item 3, 4-20 mA SIG, to access the 4-20 mA SIGNAL
screen.
4. Select menu item 1, D/A Trim, to start the trim procedure.
Calibration Passes, but Still Reads Incorrectly . . . . . . . . page 5-4
Probe Passes Calibration, O
Probe Passes Calibration, O
How do I detect a plugged diffuser? . . . . . . . . . . . . . . . . .page 5-6
Can I calibrate a badly plugged diffuser? . . . . . . . . . . . . page 5-6
Still Reads High . . . . . . . .page 5-5
2
Still Reads Low . . . . . . . . page 5-6
2
OVERVIEW OF
OPERATING
PRINCIPLES
When the Zirconium Oxide sensing cell is heated to its setpoint [736°C
°F)], the cell will generate a voltage that represents the difference
(1357
between the process O
instrument air).
O
2
When flowing calibration gases, the raw cell millivolt value should represent
the levels on the chart in Figure 5-1. Note that the raw cell millivolt value
increases logrithmically as the O
% and the reference O2% inside the probe (20.95%
2
concentration decreases.
2
http://www.raihome.com
IM-106-910Xi, Original Issue
Xi Advanced Electronics
Figure 5-1. O2 Sensor mV Reading vs %O2 at 736°C (1357°F) (Reference Air, 20.95% O2)
Instruction Manual
November 2010
GENERAL
Grounding
5-2
Install all protective equipment covers and safety ground leads after troubleshooting. Failure
to install covers and ground leads could result in serious injury or death.
Consider the following equipment conditions, features, and requirements
when troubleshooting a problem.
It is essential that adequate grounding precautions are taken when installing
the system. Thoroughly check both the probe and electronics to ensure the
grounding quality has not degraded during fault finding. The system provides
facilities for 100% effective grounding and the total elimination of ground
loops.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Electrical Noise
Electrostatic Discharge
ALARM INDICATIONS
The Xi has been designed to operate in the type of environment normally
found in a boiler room or control room. Noise suppression circuits are
employed on all field terminations and main inputs. When fault finding,
evaluate the electrical noise being generated in the immediate circuitry of a
faulty system. Ensure all cable shields are connected to earth.
Electrostatic discharge can damage the ICs used in the electronics. Before
removing or handling the circuit boards, ensure you are at ground potential.
The first indication of a problem at the O2 measuring system usually comes
from the operators running the process. Critical alarms that render the O2
measurement unusable will force the 4-20 mA analog output signal representing O2 to go to a default condition, as follows:
20.5 mAReading Over Range (Example - range is 0-10%. Current reading is 12%)
>21 mA
Transmitter Condition
Reading Under Range
(Example - user sets range to 2-10%. Current reading is 1.9%)
Critical Alarm - transmitter reading is unusable (user can choose this alarm
level instead of the factory default level of 3.5 to 3.6 mA)
IDENTIFYING AND
CORRECTING FAULT
INDICATIONS
NOTE
To ensure correct operation you should make sure that the Digital Control
System is configured to interpret these signal levels correctly.
Once an alarm condition is identified, the Xi offers a number of diagnostics to
interpret the specific alarm.
Alarm indications are available via the Xi or the 375/475 Field Communicator
and Rosemount Analytical's Asset Management software. When the error is
corrected and/or power is cycled, the diagnostic alarms will clear or the next
error on the priority list will appear.
There are two types of alarms; recoverable and non recoverable. If an
existing alarm is recoverable, the alarm-active indication will disappear when
the alarm condition no longer exists. If an alarm is not recoverable, the alarm
indication will continue to be displayed after the cause of the alarm condition
is corrected. AC power to the Xi must be cycled to clear a non-recoverable
alarm.
Alarm messages are displayed on the Xi display window when the alarm
status display is accessed via the Xi menu. A listing of the alarm/fault
messages and the related fault status descriptions are shown in Table 5-1.
Fault conditions that give no fault indication and that allow the probe to pass
calibration are listed and discussed after Table 5-1.
NV Mem FailA checksum error was detected in the nonvolatile memory configuration data when the unit was
turned on.
Factory ModeOn Xi box I/O board, SW4, position 1 is set on "ON". On transmitter electronics board, SW1,
position 1 is set to "ON". This setting should only be used in the factory.
Board Temp HiThe transmitter electronic board temperature reading is above 126°C or the
Xi unit I/O board temperature reading is above 86°C.
O2 Htr Ramp RateThe O2 sensor heater ramp rate is greater than max allowed ramp rate indicating a run away
heater condition.
O2 Sensor OpenThe cell impedance voltage is reading less than -1.1 VDC indicating the O2 sensor wires may be
disconnected or the O2 sensor junction may be open.
O2 TC OpenThe O2 cell heater thermocouple voltage is reading more than the hardware configured threshold
voltage. This indicates the thermocouple wires may be disconnected or the thermocouple junction
may be open.
O2 Temp LowThe heater temperature is below the minimum temperature. The predefined low temperature
threshold is the 726°C.
O2 Temp HighThe heater temperature is above the defined temperature threshold. The high temperature
threshold is the defined by the "High Temp Alm SP" parameter. The default value is 750°C.
O2 T/C ShortedThe O2 sensor heater temperature thermocouple voltage is shorted.Yes
O2 T/C ReversedThe O2 sensor heater temperature thermocouple voltage is reading a negative voltage indicating
the thermocouple wire connections may be reversed.
O2 Htr FailureThe O2 sensor heater temperature is not responding to the controller and can't reach final
temperature set by the device, indicating the O2 heater may have failed.
No FlameThe Flame Status Relay Input is set to the OFF state indicating it is not safe to operate the O2
heater and the heater should be turned off.
Xmtr DisconnectCommunication failures detected between transmitter and the I/O board indicating the transmitter
has been disconnected from the I/O board.
Cal RecommendedProbe calibration is recommended. The cell impedance is above 100 ohms and has shifted 50
ohms since the last calibration; the accuracy of the O2 reading may be compromised.
Cal FailedA calibration error occurred during the last calibration. The measured slope or constant is outside
the acceptable range. (Slope: 34.5 to 57.5 mv/decade) (Constant: ±20 mv)
Cell Imp HighThe O2 sensor impedance/cell resistance value measurement is greater than 2000 Ohms
indicates the cell may be beyond its useful life.
CPU Voltage LowThe CPU voltage is less than 2.7 V indicating the CPU voltage is too low.Yes
CPU Voltage HighThe CPU voltage is more than 3.3 V indicating the CPU voltage is too high.Yes
Htr Voltage LowThe heater voltage for the O2 cell heater is below 35 volts.Yes
Htr Voltage HighThe heater voltage for the O2 cell heater is above 264 volts.Yes
Low O2The O2 reading is below the "Low O2 Alm SP".Yes
November 2010
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
CALIBRATION PASSES,
BUT STILL READS
INCORRECTLY
5-4
There are a few fault conditions where no alarm indication is present and the
probe passes calibration, but the O2 reading may still be incorrect:
An incorrect flow rate of calibration gases can cause a shifted calibration. If
the flow rate of calibration gases is too low, process gases can mix in with the
calibration gases causing a mixture at the cell that is different than what is
noted on the calibration gas bottles. Always set the calibration flow rate when
a new diffuser is installed, and never readjust this flow rate until another new
diffuser is installed. For applications with heavy particulate loading, see
"Probe Passes Calibration, But O2 Still Appears To Read Low".
No or improper reference air supplied.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Probe Passes
Calibration, O2 Still
Reads High
Figure 5-2. Probe Leakage
Paths
External Reference Air Leak - There may be a leak that is permitting
ambient air to mix with the process gases. Since many combustion processes
are slightly negative in pressure, ambient air can be drawn into the cell area,
biasing the O2 reading upward.
1. Make sure that the calibration gas line is capped tightly between
calibrations.
2. If autocal is used, make sure the check valve is seating properly.
Internal Reference Air Leak - See Figure 5-2. There may be a leak inside
the O2 Probe itself, permitting the reference air (20.95% O2) to mix with the
process gases at the cell. To confirm this leak condition, pressurize the inside
(reference side) of the probe by plugging the reference air exhaust port with
your finger for 1 minute. (The conduit ports where the signal and power wires
pass may also need to be sealed.) The O2 reading should decrease slightly. If
the O2 reading increases during this test there is a leak inside the probe.
1. Acid condensation inside the probe can degrade the hose that carries
the cal gas to the cell. Inspect this hose. Dislodging or improper
installation of the cal gas or reference air hose can cause a leakage
path.
2. The sensing cell is fastened to the end of the probe tube and uses a
corrugated washer to separate the process gases from the ambient
reference air. The corrugated washer may be damaged by corrosion.
Discard used washer.
NOTE
You should aways install a new corrugated washer whenever you remove the
sensing cell from the probe.
Bad Reference Side Cell Electrode - A bad reference side cell electrode can
cause an elevated O2 reading.This fault is usually indicated by a frequent
"Calibration Recommended" alarm and increasing cell impedance readings. A
high cell impedance can be calibrated out, but if the impedance continues to
increase rapidly, the sensing cell must be replaced.
5-5
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Probe Passes
Calibration, O
Still
2
Reads Low
How do I detect a
plugged diffuser?
Can I calibrate a badly
plugged diffuser?
The diffusion element at the end of the probe is a passive filter. It plugs very
slowly, since there is no active flow being drawn across it. In applications that
have a heavy particulate loading (coal or wood fired boilers, cement and lime
kilns, catalyst regeneration, recovery boilers, etc.), this diffusion element will
eventually plug.
NOTE
It is important not to pressurize the sensing cell during calibrations by flowing
excessive cal gas against a plugged diffuser. Always use a two-stage
regulator for setting calibration gas pressure. Calibration flow rates should be
set only when a new diffuser is installed. As the diffuser plugs, do not adjust
the flow rates upward.
The O2 cell's speed of response will degrade. The O2 trend in the control
room will become smoother.
When calibrating, the calibration gas flow rate will be noted to be lower. Never
readjust this flow upwards to correct for a plugged diffuser. Adjust this flow
only when a new diffuser is installed.
Always note the time it takes for the cell to recover to the normal process
value after the cal gas is removed. As the diffuser plugs, this recovery time will
get longer and longer. Use the Calibration Record provided to record and
track Calibration Response times.
It may not be possible to immediately replace a plugged diffuser while the
process is on line.
One can calibrate the probe without pressurizing the cell by adjusting the
calibration gas flow rate downward before calibration. For instance, say the
process is at 3%, and the first calibration gas is 8%. Adjust the flow of cal gas
downward until the reading begins to migrate from 8% to lower values,
indicating that process gases are now mixing in with the calibration gases.
5-6
Adjust the flow rate back up until this mixing is just eliminated. Calibrate at this
flow rate. Replace the diffuser at the first opportunity.
Install all protective equipment covers and safety ground leads after troubleshooting. Failure
to install covers and ground leads could result in serious injury or death.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Calibration Record
For
Rosemount Analytical In Situ O2 Probe
Probe Serial Number: ______________________________________________________________________
Probe Tag Number: _______________________________________________________________________
Probe Location: __________________________________________________________________________
Date Placed Into Service: __________________________________________________________________
DateSlopeConstantImpedanceResponse
initial
Response
final
Notes: Response
Response
When the second calibration gas is turned off, note the number of seconds required for the O2 value to begin migrating
initial
back to the process value.
When the second calibration gas is turned off, note the number of seconds required for the O2 value to settle out at the
This section identifies the calibration methods available and provides the
procedures to maintain and service the Xi.
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
The maintenance interval required is quite variable, depending on the type of
service the analyzer is placed into. The zirconium oxide sensing cell is
non-depleting, and has no specific shelf life or a defined life in flue gas
operation. The cell of a probe that is mounted inside a boiler that is burning
natural gas may shift very little over several years. Acidic compounds are the
main aggressors to the sensing cell, typically SO2 resulting from sulfur
contained in coal and heavy oil fuels, and also HCl from the combustion of
plastics in municipal incinerators and in industrial thermal oxidizers. Sensing
cells may experience significant degradation and signal shift in this type of
service, particularly if the operating levels of O2 are very low (below 1% O2).
A calibration check is generally recommended on a quarterly basis (every 3
months) by flowing bottled gas to the probe. (Make sure that the operations
personnel are notified when doing this, and also make sure that the O2 control
loop is placed in manual mode). If the probe readings vary significantly from
the bottle values, then a formal calibration should be conducted as noted in
Section 4: Startup and Operation.
http://www.raihome.com
The Xi offers a "calibration recommended" diagnostic that will indicate when
the probe needs to be calibrated.
Xi Advanced Electronics
Combustion processes that have a high level of ash or other particulate
content will cause the diffusion element on the end of the probe to plug off. A
badly plugged diffuser will cause a slower speed of response to changing O
levels in the process. This can usually be seen on the recorded trends in the
control room.
When performing a calibration check or actual calibration, the calibration flow
meter may read lower if the diffuser is badly plugged. (Never increase the flow
rate back up, however, as this can cause a shifted calibration. Adjust the
calibration flow rate only when a new diffuser is installed). Always record the
response time back to the process after the calibration gases are removed, as
noted on the calibration record at the end of this section. Diffuser pluggage
can be tracked through this record.
A visual inspection of the probe should be conducted during plant outages,
paying particular attention to condensed components. Condensation can be
reduced or eliminated by insulating the probe installation, including the probe
mount, flange, and dual blue housings.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
2
CALIBRATIONThe Xi can calibrate an O
via a handheld 375/475 Field Communicator, or automatically through the
SPS 4001B Single Probe Autocalibration Sequencer or the IMPS 4000
Intelligent Multiprobe Test Gas Sequencer.
Probe manually through the front panel display or
2
Automatic CalibrationThe Xi can be used with the SPS 4001B or IMPS 4000 in order to perform an
auto / semi-auto calibration.
Manual for further details on how to configure and perform an automatic
calibration.
Refer to the SPS 4001B or IMPS 4000 Instruction
Manual CalibrationRefer to the "Calibration - General" in Section 4: Startup and Operation to
perform a manual calibration.
REPLACEMENT PARTSRefer to Section 7: Replacement Parts for individual replacement parts and
part replacement kits. Part replacement kits are available for each of the
components discussed in this section.
6-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Xi COMPONENTS
REPLACEMENT
Figure 6-1. Xi Components
Each of the following procedures details how to replace a specific component
of the Xi. Most of these procedures include component setup instructions that
must be performed before returning the related O2 Probe to service. Refer to
Figure 6-1 and Figure 6-2 for illustrations of the Xi components.
Component replacement kits are available for each of the components
replaced in this repair section. Refer to Section 7: Replacement Parts for kit
part numbers.
6-3
Xi Advanced Electronics
Figure 6-2. Xi Front Panel
Components
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
I/O Board Replacement
Use the procedure that follows to replace and set up the I/O board in the Xi.
Disconnect and lock out power before working on any electrical components.
The I/O Board is shipped from the factory without any of the enhanced software features
activated. These features must be activated once the new board has been installed and
before the Remote Interface is put into service.
If the existing I/O Board has been operated with the Stoichiometric enhanced software
feature, this feature must be activated in the new board before the Xi Electronics is put back
into service. Failure to do so will cause a false analog output signal to the DCS.
6-4
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Replacing an Existing I/O Board
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Disconnect the 10-pin ribbon cable from the I/O Board. A new cable is
supplied in the replacement kit and should be used if the old one is
damaged.
4. Tag and disconnect wiring for Alarm Outputs, Flame Status Input and/or
SPS/IMPS as applicable.
5. Slide the I/O Board part way out of the Xi enclosure.
6. Disconnect the 4-position plug for the transmitter probe and output
wiring. Remove the I/O Board completely from the Xi enclosure.
7. See Figure 6-3. Set jumpers JP1, JP2, JP5, JP7 and JP8 to their proper
positions using the old I/O Board as a guide.
8. Set switch SW4 to their proper positions using the old I/O Board as a
guide.
9. Partially slide the new I/O Board into the Xi enclosure. Ensure the
board is correctly aligned within the slots in the enclosure.
10. Connect the 4-position plug for the transmitter probe and output wiring.
Slide the I/O Board completely into the Xi enclosure.
Figure 6-3. I/O Board Jumper & Switch Settings
6-5
Xi Advanced Electronics
11. Reinstall wiring for Alarm Outputs, Flame Status Input and/or SPS/IMPS
12. Connect the ribbon cable to the I/O Board. A new cable is supplied and
13. Swing the Xi cover up in place and tighten the four screws.
14. Prior to operating the O2 Probe and the Xi, all optional software
Part NumberSoftware Option
6A00269G01Enhanced Software Option Upgrade, Stoichiometric Function
6A00269G02Enhanced Software Option Upgrade, Programmable Reference Function
6A00269G03Enhanced Software Option Upgrade, 850°C Process Function
NOTE
For enhanced software upgrades or to enable optional software features
previously used in your Xi configuration, contact Rosemount Analytical Inc. at
1-800-433-6076.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
as applicable. See Figure 6-4 and Figure 6-5 for wiring diagrams. See
Figure 6-6 for I/O board positions in the Xi enclosure.
should be used if the old one is damaged.
enhancements (previously enabled or not) must be enabled. Notify
Rosemount Analytical Inc. and reference the following part numbers to
enable the related software options:
15. Recalibrate the O2 Probe according to the applicable calibration
Figure 6-6. I/O Board Positions
in the Xi Enclosure
6-7
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
AC Relay Board
Replacement
Use the procedure that follows to replace and set up the AC Relay board in
the Xi.
Disconnect and lock out power before working on any electrical components.
Replacing an Existing AC ReIay Board
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Tag and disconnect wiring for the Relay In.
4. Slide the AC Relay Board part way out of the Xi enclosure.
5. Tag and disconnect the two 3-position plugs for the AC input and
transmitter probe heater wiring. Remove the AC Relay Board
completely from the Xi enclosure.
6. Partially slide the new AC Relay Board into the Xi enclosure. Ensure the
board is correctly aligned within the slots in the enclosure.
7. Connect the two 3-position plugs for the AC input and transmitter probe
heater wiring. Slide the AC Relay Board fully into the Xi enclosure.
8. Reinstall wiring for Relay In. See Figure 6-7 for wiring diagram.
9. Swing the cover up in place and tighten the four screws.
Installing and configuring an AC Relay Board for the Flame Status Interlock function will
dedicate alarm output 2 for this function. Once configured for Flame Status Interlock, the
software will override any previous alarm assignments for alarm output 2 and not allow it to
be used for any other function.
Adding an AC Relay Board to the Xi
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Slide the I/O Board part way out of the Xi enclosure.
4. Make the following jumper and switch settings:
a. Set SW4, Position 4 "ON" and all other positions "OFF".
b. Move the jumpers for both JP1 and JP2 to pins 1 and 2.
5. The I/O Board must be relocated to the right-most slot inside the Xi
enclosure. If there is enough service loop on the existing wiring, slide
the I/O Board completely out of the enclosure and re-install it in the
right-most slot. Ensure the board is correctly aligned within the slots.
See Figure 6-8 for the correct location of the I/O Board.
NOTES
An AC Relay Board can only be added to single channel Xi, i.e. only one I/O
Board controlling one O2 Probe.
The Flame Status Interlock function requires a contact closure on the digital
input of the I/O Board to indicate a flame is present. When properly connected
and configured, AC power will be applied to the transmitter only when the
flame is present.
6. If there is not sufficient wiring, tag and disconnect all wiring, and
relocate the board. Then reinstall the wiring.
7. Partially slide the AC Relay Board into the left-most slot of the Xi
enclosure. The component side of the AC Relay board will be to the
right with the fuse holder at the top. Ensure the board is correctly
aligned within the slots in the enclosure. See Figure 6-8 for the correct
location of the AC Relay Board.
6-9
Xi Advanced Electronics
8. Connect two wires approximately 6" long each between the "HTR COM"
9. Connect the flame status indicator contact to the "DI+" and "DI-" on the
10. Connect the AC input and output wiring to the Transmitter. See
11. Slide the AC Relay Board completely into the Xi enclosure.
12. Swing the cover up in place and tighten the four screws.
Figure 6-8. I/O and AC Relay
Board Position in Xi Enclosure
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
and the "HTR NC" connections on the I/O Board and the "RELAY IN"
connections on the AC Relay Board; observe polarity. See Figure 6-7
for wiring details.
AC Relay Board. See Figure 6-7 for wiring details.
Figure 6-7 for wiring details.
6-10
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Power Supply Board
Replacement
Use the procedure that follows to replace the Power Supply board in the Xi.
Use this procedure to replace an original Linear Power Supply board or the
current configuration Switching Power Supply board.
Disconnect and lock out power before working on any electrical components.
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Refer to the wiring diagram in Figure 6-9. Unplug the AC input wiring
plug from the Power Supply board. A new plug is supplied in the
replacement kit and should be used if the existing plug is damaged.
4. Disconnect the 14-pin ribbon cable from the Power Supply board.
5. Remove the two long screws that secure the bracket (9, Figure 6-1) to
the Xi enclosure.
6. Hold the AC input wiring to the right and slide the Power Supply board
out of the Xi enclosure.
7. Install the new mounting bracket (9, Figure 6-1) on the new Power
Supply board (10, Figure 6-1). A new bracket and mounting screws are
provided in the replacement kit.
8. Slide the Power Supply board into the mating slots in the Xi Enclosure.
Make sure the board is correctly aligned in the slots.
9. Install and tighten the bracket mounting screws. Two new screws are
provided in the replacement kit and should be used if the existing
screws are damaged.
Figure 6-9. Power Supply
Board Wiring
6-11
Xi Advanced Electronics
10. Connect the ribbon cable to the Power Supply board. A new ribbon
11. Connect the AC power plug to the Power Supply board.
12. Swing the Xi cover up and tighten the four mounting screws.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
cable (7, Figure 6-1) is provided in the replacement kit and should be
used if the existing cable is damaged.
Xi Front Panel
Replacement
Use the procedure that follows to replace the front panel on the Xi.
Replacement kits with and without the CPU board are available. Use the
instructions that apply to the replacement kit you have.
Disconnect and lock out power before working on any electrical components.
Replacing Front Panel Assembly without CPU Board
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Disconnect the 14-pin ribbon cable going to the Power Supply board. A
new cable is supplied in the replacement kit and should be used if the
old one is damaged.
4. Disconnect the 10-pin ribbon cable(s) going to the I/O Board(s). One
new cable is supplied in the replacement kit and should be used if either
of the ribbon cables are damaged.
5. Remove the wire hinge from the right side of the Xi cover. A paper clip or
similar device can be inserted into the hole on the left side of the cover
to push the hinge out of the cover. A new hinge pin is supplied in the
replacement kit and should be used if the old one is damaged.
6. Position the new front panel assembly in place and reinstall the hinge
pin. Ensure the hinge pin is fully seated into the Xi cover.
6-12
The new front panel assembly may be supplied with protective clear membranes over the
interior and exterior of the window. Failure to remove the exterior protective membrane may
cause the display to appear distorted. The membrane may be difficult or impossible to
remove after extended use at elevated temperatures.
7. The keypad window on the new front panel may be supplied with interior
and exterior protective membranes. Remove the protective membranes
prior to final assembly and use of the Xi enclosure.
8. Reconnect the 10-pin ribbon cable from the I/O Board.
9. Reconnect the 14-pin ribbon cable from the Power Supply Board.
10. Swing the Xi cover up in place and tighten the four screws.
11. If necessary, reconfigure the appearance of the main display. Refer to
Section 3: Configuration of Xi Electronics.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Replacing Front Panel Assembly without CPU Board
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Disconnect the 14-pin ribbon cable going to the Power Supply board. A
new cable is supplied in the replacement kit and should be used if the
old one is damaged.
4. Disconnect the 10-pin ribbon cable(s) going to the I/O Board(s). One
new cable is supplied in the replacement kit and should be used if either
of the ribbon cables are damaged.
5. Remove the wire hinge from the right side of the Xi cover. A paper clip or
similar device can be inserted into the hole on the left side of the cover
to push the hinge out of the cover. A new hinge pin is supplied in the
replacement kit and should be used if the old one is damaged.
6. Place the front panel assembly on the bench. Remove the 4 screws
securing the CPU board to the front cover.
NOTE
Prior to disassembly, observe the position of the plastic bumper with rubber
insert as it is installed over the CPU board. The two longer screws are used
on the top edge of the CPU board passing through the bumper.
7. Lift the CPU board off of the front cover using care not to damage the
board.
8. Disconnect the keypad overlay ribbon cable from the CPU board.
Discard the used front cover with the keypad overlay attached.
The new front panel assembly may be supplied with protective clear membranes over the
interior and exterior of the window. Failure to remove the exterior protective membrane may
cause the display to appear distorted. The membrane may be difficult or impossible to
remove after extended use at elevated temperatures.
9. The keypad window on the new front panel may be supplied with interior
and exterior protective membranes. Remove the protective membranes
prior to final assembly and use of the Xi enclosure.
NOTE
Prior to assembly, it is recommended that the inside of the keypad overlay
window and LCD display window be gently cleaned to remove fingerprints
and accumulated dust.
10. Clean the inside of the keypad overlay window and LCD display window
to remove fingerprints and accumulated dust. Use only a soft cloth; do
not use any detergents or chemicals.
11. Connect the keypad overlay ribbon cable on the new front panel
assembly to the CPU board. Use care not to damage the board.
12. Position the CPU board in place in the front panel assembly.
6-13
Xi Advanced Electronics
13. Reinstall the bumper and 4 screws to secure the CPU board in place.
14. Position the front panel assembly and reinstall the hinge pin. Ensure the
15. Reconnect the 10-pin ribbon cable from the I/O Board.
16. Reconnect the 14-pin ribbon cable from the Power Supply Board.
17. Swing the Xi cover up in place and tighten the four screws.
18. If necessary, reconfigure the appearance of the main display. Refer to
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
The two longer screws are used with the plastic bumper on the top edge
of the CPU board.
hinge pin is fully seated into the Xi cover.
Section 3: Configuration of Xi Electronics.
DR Board Replacement
Use the procedure that follows to replace the DR board in the Xi that is
connected to a Direct Replacement probe.
Disconnect and lock out power before working on any electrical components.
1. Loosen the four screws securing the Xi cover. The screws are captive
and do not need to be completely removed.
2. Swing the Xi cover down to expose the inner components.
3. Slide the DR board part way out of the Xi enclosure.
4. Unplug the wiring harness plug from the connector J8 located on the
bottom, left-hand side of the board. Refer to Figure 6-10.
5. Unplug the wiring harness plugs from the connectors J1, J2, and J3
located on the top, right-hand side of the board. Refer to Figure 6-11.
6. Tag and disconnect the wiring to connector J4.
7. Remove the DR board from the Xi enclosure.
8. Partially slide the new DR board into the mating slots in the Xi
Enclosure. Make sure the board is correctly aligned in the slots.
9. Refer to the wiring diagram in Figure 6-10. Reconnect the wiring to
connector J4.
10. Connect the wiring harness plugs to connectors J1, J2, J3, and J8.
11. Slide the DR board completely into the Xi enclosure.
12. Swing the Xi cover up and tighten the four mounting screws.
6A00265G01Kit, Power Supply Board (Obsolete - Replaced by 6A00329G01)
6A00329G01Kit, Switching Power Supply Board
6A00266G01Kit, I/O Board*
6A00267G01Kit, AC Relay Board
6A00328G01Kit, DR Board
6A00268G01Kit, Front Panel with Overlay & CPU Board
6A00268G02Kit, Front Panel with Overlay
6A00269G01Enhanced Software Option Upgrade, Stoichiometric Function
6A00269G02Enhanced Software Option Upgrade, Programmable Reference Function
6A00269G03Enhanced Software Option Upgrade, 800°C Process Function
6A00237H24Kit, Pipe & Wall Mount
6A00237H33Kit, Panel Mount
6A00243G01Ribbon Cable, 10 Pin (CPU Board to I/O Board)
6A00242G01Ribbon Cable, 14 Pin (CPU Board to Power Supply Board)
6A00291H01Insulator
6A00381G01Kit, Hole Plug & Gland
6A00285H01Gasket, Panel
6A00287H01Gasket, Cover
*
Note:
If the existing I/O Board has been operated with the Stoichiometric enhanced software feature, this feature
must be activated in the new board before the Xi is put back into service. Failure to do so will cause a
false analog output signal to the DCS.
http://www.raihome.com
Xi Advanced Electronics
Calibration Components
Table 7-2. Replacement Parts
for Calibration Components
Part NumberDescription
1A99119G01Calibration Gas Bottles - 0.4% and 8% O2, balance
1A99119G02Two Flow Regulators (for calibration gas bottles)
1A99119G03Bottle rack
*Note:
Calibration gas bottles cannot be shipped via airfreight
OxyBalance Display and Averaging System . . . . . . . . . . page 8-6
HART HANDHELD
375/475 FIELD
COMMUNICATOR
Figure 8-1. 375/475 Field
Communicator
http://www.raihome.com
The 375/475 Field Communicator is an interface device that provides a
common communication link to HART-compatible instruments, such as the Xi.
HART Communications Protocol permits all the information available from the
Xi's electronics to be transmitted over standard 4-20 mA signal wires. By
attaching the 375/475 Field Communicator at a termination point along the
4-20 mA signal line, a technician can diagnose problems and configure and
calibrate the Xi as if he or she were standing in front of the instrument.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
Xi Advanced Electronics
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
ASSET MANAGEMENT
SOLUTIONS (AMS)
BY-PASS PACKAGES
Asset Management Solutions (AMS) software works in conjunction with the
HART Communication Protocol and offers the capability to communicate with
all HART plant devices from a single computer terminal.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
The specially designed Rosemount Analytical By-Pass Package for oxygen
analyzers has proven to withstand the high temperatures in process heaters
while providing the same advantages offered by the in situ sensor. Inconel
steel tubes provide effective resistance to corrosion, and the package uses no
moving parts, air pumps, or other components common to other sampling
systems.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
8-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
SPS 4001B
SINGLE PROBE
AUTOCALIBRATION
SEQUENCER
Figure 8-2. SPS 4001B
Xi Advanced Electronics
Rosemount Analytical Inc. specifically designed the SPS 4001B Single Probe
Autocalibration Sequencer to provide the capability to perform automatic or
on-demand calibrations. The SPS 4001B is fully enclosed in a NEMA cabinet
suited for wall-mounting. This cabinet provides added protection against dust
and minor impacts.
The SPS 4001B works in conjunction with the Xi, eliminating out-of-calibration
occurrences and the need to send a technician to the installation site.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
8-3
Xi Advanced Electronics
38890096
IMPS 4000
INTELLIGENT
MULTIPROBE
TEST GAS SEQUENCER
Figure 8-3. IMPS 4000
The IMPS 4000 Intelligent Multiprobe Test Gas Sequencer is housed within
an IP56 (NEMA 4X) enclosure and has the intelligence to provide calibration
gas sequencing of up to four Xi’s to accommodate automatic and
semi-automatic calibration routines.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
This sequencer works in conjunction with the CALIBRATION
RECOMMENDED feature, eliminating out-of-calibration occurrences and the
need to send a technician to the installation site. In addition, the IMPS 4000
provides a remote contact input to initiate a calibration from a remote location
and relay outputs to alert when a calibration is in progress, an O2 Probe is out
of calibration, calibration gases are on, and calibration gas pressure is low.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
8-4
Instruction Manual
38890098
IM-106-910Xi, Original Issue
November 2010
O2 CALIBRATION GAS
Figure 8-4. Calibration Gas
Bottles
Xi Advanced Electronics
Rosemount Analytical's O2 Calibration Gas and Service Kits have been
carefully designed to provide a more convenient and fully portable means of
testing, calibrating, and servicing.
Rosemount Analytical's oxygen analyzers. These lightweight, disposable gas
cylinders eliminate the need to rent gas bottles.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
8-5
Xi Advanced Electronics
38890099
Analytical
OXYBALANCE DISPLAY
AND AVERAGING
SYSTEM
Figure 8-5. OxyBalance
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Optional OxyBalance Display and Averaging System. Reviews up to eight
4-20 mA signals from individual probes. Trends individual outputs, calculates
four programmable averages as additional 4-20 mA outputs.
For more information, call Rosemount Analytical Inc. at 1-800-433-6076.
The XPS is primarily an interfacing electronics box installed between an
Probe and the Xi electronics. There are three variants of the XPS
O
2
electronics box, each with specific functions.
1. Remote XPS - 6A00358G01 for 44V Probes
2. Remote XPS - 6A00358G03 for 115V Probes
3. Integral XPS - 6A00365G01 for 44V Probes
This version of the XPS is for use with Westinghouse/Rosemount Analytical
World Class probes that use a 44 volt heater, and when the O
120/240V, 50/60Hz power source. The 6A00358G01 version contains
electronics to receive and process multiple raw signals from the probe.
Signal wires between the O
following:
1. Thermocouple - 2 conductors (type K lead wire)
2. Sensing Cell - 2 conductors
3. Heater - 3 conductors
4. Shield
This version of the XPS provides for probe heater control and signal
conditioning. The signal conditioning electronics result in a linear 4-20 mA
signal.
%O
2
No operator interface is included. Thereby, the 4-20 mA signal typically goes
to a Xi Advanced electronics unit. Alternately, a handheld HART 375/475
Field Communicator can be used as the operator interface. However, the Xi
or the field communicator must be the dedicated communication device. The
Xi and the field communicator cannot be used at the same time.
Probe and the G01 version XPS include the
2
Probe uses a
2
http://www.raihome.com
For heater control the G01 version of the XPS includes a transformer. The
transformer converts 120/240V AC input power to the 44V AC heater voltage
used in the Westinghouse/Rosemount Analytical World Class probe.
Xi Advanced Electronics
Figure A-1. Remote XPS
6A00358G01 and 6A00358G03
Mounting
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Remote XPS for 115V
Probes - 6A00358G03
A-2
The 6A00358G03 version of the XPS is similar to the 6A00358G01 described
above except that it does not include the transformer. The 6A00358G03
version controls any Westinghouse/Rosemount Analytical O
a 115V heater. Applicable probes include Models 218 and 218A, Oxymitter,
X-STREAM, or Model 6888 probes.
probe that uses
2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
Integral XPS
6A00365G01 with Xi
(for 44V Probes)
Figure A-2. Integral XPS with Xi
and 6A00365G01 Mounting
This XPS equipment configuration includes the XPS and a Xi Advanced
Electronics unit installed on a common mounting plate. This configuration is
intended to replace a World Class Intelligent Field Transmitter (IFT) where the
required input line voltage is 120/240V AC, particularly for 240 VAC
applications.
Other XPS Uses
Two XPS units can be wired to two O2 Probes and back to one dual-channel
Xi. This avoids having to purchase a Xi for each probe and eliminates one of
two signal cables that would otherwise be required to communicate between
the Xi and XPS units.
The following safety instructions apply specifically to all EU member
states. They should be strictly adhered to in order to assure compliance
with the Low Voltage Directive. Non-EU states should also comply with
the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points,
internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds
must be replaced. The integrity of all earth terminals must be maintained
at all times.
3. Mains supply cords should comply with the requirements of IEC227 or
IEC245.
4. All wiring shall be suitable for use in an ambient temperature of greater
than 75°C.
5. All cable glands used should be of such internal dimensions as to
provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains
supply should only be made through a circuit breaker which will
disconnect all circuits carrying conductors during a fault situation. The
circuit breaker may also include a mechanically operated isolating
switch. If not, then another means of disconnecting the equipment from
the supply must be provided and clearly marked as such. Circuit
breakers or switches must comply with a recognized standard such as
IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol
to the right, hazardous voltages are likely to be present
beneath. These covers should only be removed when
power is removed from the equipment - and then only
by trained service personnel.
8. Where equipment or covers are marked with the symbol
to the right, there is a danger from hot surfaces beneath.
These covers should only be removed by trained
service personnel when power is removed from the
equipment. Certain surfaces may remain hot to the
touch.
9. Where equipment or covers are marked with the symbol
to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one
or more of the following standards: EN61010-1, IEC417,
and ISO3864.
11. Where equipment or labels are marked "Do Not Open While Energized"
or similar, there is a danger of ignition in areas where an explosive
atmosphere is present. This equipment should only be opened when
power is removed and adequate time as specified on the label or in the
instruction manual has been allowed for the equipment to cool down and then only by trained service personnel.
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
B-2
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
Xi Advanced Electronics
DŮLEŽITÉ
Bezpečnostní pokyny pro zapojení a instalaci zařízení
Následující bezpečnostní pokyny se speciálně vztahují na všechny
členské státy EU. Pokyny by měly být přísně dodržovány, aby se
zajistilo splnění Směrnice o nízkém napětí. Pokud nejsou pokyny
nahrazeny místními či národními normami, měly by je dodržovat i
nečlenské státy EU.
1. U všech zemnicích bodů, interních a externích, by mělo být vytvořeno
odpovídající uzemnění.
2. Po instalaci nebo odstranění problémů musí být vyměněny všechny
bezpečnostní kryty a uzemnění. Vždy musí být zajištěna integrita všech
zemnicích svorek.
3. Sí˙ové kabely by měly odpovídat požadavkům normy IEC227 nebo
IEC245.
4. Všechna zapojení by měla být vhodná pro použití při vnějších teplotách
nad 75 °C.
5. Všechna použitá kabelová hrdla by měla mít takové vnitřní rozměry, aby
zajistila odpovídající zakotvení kabelu.
6. Správnou činnost zařízení zajistíte, vytvoříte-li připojení k napájecímu
zdroji pouze přes jistič, který v případě poruchy odpojí všechny obvody
s konduktory. Jistič může také obsahovat mechanický odpojovač.
Pokud ho neobsahuje, musí být zajištěn a jasně označen jiný způsob
odpojení zařízení od zdroje. Jističe nebo přepínače musí odpovídat
uznávaným normám, např. IEC947. Všechna zapojení musí odpovídat
místním normám.
7. Je-li zařízení nebo kryt označen symbolem na pravé
straně, pravděpodobně se uvnitř nachází nebezpečné
napětí. Tyto kryty by měly být sejmuty pouze po
odpojení zařízení od zdroje - a to pouze kvalifikovaným
zaměstnancem.
8. Je-li zařízení nebo kryt označen symbolem na pravé
straně, povrch zařízení může být velmi horký. Tyto kryty
by měly být sejmuty pouze kvalifikovaným
zaměstnancem po odpojení zařízení od zdroje. Některé
povrchy mohou být stále horké.
9. Je-li zařízení nebo kryt označen symbolem na pravé
straně, přečtěte si nejprve instrukce v návodu k obsluze.
10. Všechny grafické symboly používané u výrobku
pocházejí z následujících norem: EN61010-1, IEC417 a
ISO3864.
11. Pokud je zařízení nebo štítky označeno varováním „Je-li zařízení pod
napětím, neotvírejte jej“ či podobným, může dojít ve výbušném prostředí
ke vznícení. Zařízení lze otevřít pouze po jeho odpojení od zdroje a
ponechání dostatečného času na vychladnutí, jak je uvedeno na štítku
nebo v návodu k obsluze - a to pouze kvalifikovaným zaměstnancem.
B-3
Xi Advanced Electronics
Sikkerhedsinstruktion for tilslutning og installering af dette udstyr.
Følgende sikkerhedsinstruktioner gælder specifikt i alle
EU-medlemslande. Instruktionerne skal nøje følges for overholdelse af
Lavsspændingsdirektivet og bør også følges i ikke EU-lande medmindre
andet er specificeret af lokale eller nationale standarder.
1. Passende jordforbindelser skal tilsluttes alle jordklemmer, interne og
2. Efter installation eller fejlfinding skal alle sikkerhedsdæksler og
3. Forsyningskabler skal opfylde krav specificeret i IEC227 eller IEC245.
4. Alle ledningstilslutninger skal være konstrueret til omgivelsestemperatur
5. Alle benyttede kabelforskruninger skal have en intern dimension, så
6. For opnåelse af sikker drift og betjening skal der skabes beskyttelse
7. Hvor udstyr eller dæksler er mærket med dette symbol,
8. Hvor udstyr eller dæksler er mærket med dette symbol,
9. Hvor udstyr eller dæksler er mærket med dette symbol,
10. Alle benyttede grafiske symboler i dette udstyr findes i
11. Når udstyr eller etiketter er mærket "Må ikke åbnes, mens udstyret
Instruction Manual
IM-106-910Xi, Original Issue
November 2010
VIGTIGT
eksterne, hvor disse forefindes.
jordforbindelser reetableres.
højere end 75°C.
passende kabelaflastning kan etableres.
mod indirekte berøring gennem afbryder (min. 10A), som vil afbryde alle
kredsløb med elektriske ledere i fejlsitua-tion. Afbryderen skal indholde
en mekanisk betjent kontakt. Hvis ikke skal anden form for afbryder
mellem forsyning og udstyr benyttes og mærkes som sådan. Afbrydere
eller kontakter skal overholde en kendt standard som IEC947.
er farlige spændinger normalt forekom-mende bagved.
Disse dæksler bør kun afmonteres, når
forsyningsspændingen er frakoblet - og da kun af
instrueret servicepersonale.
forefindes meget varme overflader bagved. Disse
dæksler bør kun afmonteres af instrueret
servicepersonale, når forsyningsspænding er frakoblet.
Visse overflader vil stadig være for varme at berøre i op
til 45 minutter efter frakobling.
se da i betjeningsmanual for instruktion.
én eller flere af følgende standarder:- EN61010-1,
IEC417 & ISO3864.
tilføres strøm” eller lignende, er der fare for antændelse i områder, hvor
der er en eksplosiv atmosfære. Dette udstyr må kun åbnes, når
strømkilden er fjernet, og der er gået tilstrækkelig tid til, at udstyret er
kølet ned. Den nødvendige tid hertil er angivet på etiketten eller i
brugervejledningen. Udstyret må kun åbnes af en faglært person.
B-4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.