Emerson LIQ-MAN-ABR-226 User Manual

Instruction Manual 226
LIQ_MAN_ABR_226 April 2014
Submersion/Insertion Toroidal Sensor
For additional information, please visit our website at www.rosemountanalytical.com
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user.
CAUTION
Specifications
Wetted Material: PEEK (glass-lled), EPDM gasket Temperature: 248 °F (120 °C) maximum Pressure: 295 psig (2135 kPa abs)
Installation
Keep at least 2.4 in. (60 mm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely submerged in the liquid. Mounting the sensor in a vertical pipe run with the ow from bottom to top is best. If the sensor must be installed in a horizontal pipe run, mount the sensor in the 3 o’clock or 9 o’clock position.
Figure 1. Submersion Installation Showing Mounting Hardware
Screw the rear facing thread of the PEEK adapter into the pipe. Do not let the back end of the sensor get wet.
226 Instruction Manual
April 2014 LIQ_MAN_ABR_226
Figure 2. Insertion Installation Showing Mounting Hardware
Wiring
Keep sensor wiring away from ac conductors and high current demanding equipment. Do not cut cable. Cutting the cable may void the warranty.
Note
For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product. 1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemountanalytical/Liquid/Sensors/ Pages/Wiring_Diagram.aspx 1055: http://www3.emersonprocess.com/raihome/sp/liquid/wiring/1055/
Figure 3. Wire Functions
226-54
226-56
clear
2
Instruction Manual 226
LIQ_MAN_ABR_226 April 2014
Figure 4. Wiring for 226-54 and 226-56 sensors to 1056 and 56 analyzers
Figure 5. Wiring for 226-54 and 226-56 sensors to 54eC analyzer
3
226 Instruction Manual
April 2014 LIQ_MAN_ABR_226
Figure 6. Wiring for 226-54 and 226-56 sensors to Xmt-T panel mount transmitter
Figure 7. Wiring for 226-54 and 226-56 sensors to Xmt-T pipe/wall mount transmitter
4