Emerson 5081T-FF, 5081T-FI User Manual

Model 5081T-FF/FI
Instruction Sheet
PN 51A-5081T-FF/FI/rev.A November 2005
Your purchase from Rosemount Analytical, Inc. has result­ed in one of the finest instruments available for your par­ticular application. These instruments have been designed, and tested to meet many national and interna­tional standards. Experience indicates that its perform­ance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, commis­sioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against haz­ards may be impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; per­sonal injury; property damage; damage to this instru­ment; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manu­al covers your model and options. If not, call 1-800­854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, oper­ation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pres­sure sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protective cov­ers must be in place unless qualified personnel are per­forming maintenance.
• If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
ESSENTIAL INSTRUCTIONS
READ THIS P
AGE BEFORE PROCEEDING!
WARNING
To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or understand and adhere to the manufacturer's live maintenance procedures.
W
ARNING
Substitution of components may impair Intrinsic Safety or suitability for Division 2.
WARNING
Do not remove or replace while circuit is live unless area is known to be non-hazardous.
WARNING
Explosion Hazard - Do not disconnect equipment unless area is known to be non-hazardous.
For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.
FOUNDATION Fieldbus®Two-Wire Conductivity Transmitter
PHYSICAL SPECIFICATIONS
HOUSING: Epoxy-polyester painted over low-copper alu-
minum. Neoprene O-rings on cover. 160.5 mm x 175.3 mm x 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.)
DIAMETER: 155.4 mm (6.1 in.) ELECTRICAL CONDUIT OPENINGS: 3/4 in. FNPT POWER SUPPLY AND LOAD: A power supply voltage of
9 Vdc to 32 Vdc at 22 mA is required; Intrinsically Safe installations may be limited to a maximum of 2-3 transmitters per node, depending on the barrier used.
LOCAL READOUT:
Main Display is 4 digits, 20 mm tall (0.8 in.) Message Display is ten digits, 7 mm tall (0.3 in.)
AUTOMATIC TEMPERATURE COMPENSATION:
3-wire Pt100 RTD Conductivity: 0 to 200 °C (32 to 392 °F) % concentration: 0 to 100 °C (32 to 212°F)
AMBIENT TEMPERATURE: -20 to 65° C (-4 to 149° F) RELATIVE HUMIDITY: 0-95% with enclosure sealed. CE: EMI/RFI certified: EN61326-1
HAZARDOUS AREA CLASSIFICATION: Intrinsic Safety:
Class I, II, III, Div. 1 Groups A-G T4 Tamb = 70°C Exia Entity Class I, Groups A-D Class II, Groups E-G Class III T4 Tamb = 70°C
ATEX 0600
II 1 G Baseefa03ATEX0399 EEx ia IIC T4 Tamb = -20°C to +65°C
Non-Incendive:
Class I, Div. 2, Groups A-D Dust Ignition Proof Class II & III, Div. 1, Groups E-G NEMA 4X Enclosure Class I, Div. 2, Groups A-D Suitable for Class II, Div. 2, Groups E-G T4 Tamb = 70°C
Explosion-Proof:
Class I, Div. 1, Groups B-D Class II, Div. 1, Groups E-G Class III, Div. 1 Class I, Groups B-D Class II, Groups E-G Class III Tamb = 65°C max
TRANSMITTER SPECIFICATIONS @ 25°C
MEASURED RANGE*: 50 to 2,000,000 µS/cm (see chart) ACCURACY: ± 1.0% of reading REPEATABILITY: ± 0.25% of reading STABILITY: 0.25% of output range/month, non-cumulative AMBIENT TEMPERATURE COEFFICIENT: ± 0.2% of FS/°C COMPATIBLE RTD: 100with Automatic Recognition TEMPERATURE SLOPE ADJUSTMENT: 0-5%/° C % CONCENTRATION RANGES:
Sodium Hydroxide: 0 to 15% Hydrochloric Acid: 0 to 16% Sulfuric Acid: 0 to 25% and 96 to 99.7%
LOOP SPECIFICATIONS
LOOP ACCURACY: With a standard Model 228 or 225
sensor with 20' cable, laboratory accuracy at 25°C can be as good as ±2% of reading and ±50 µS/cm.
To achieve optimum performance, standardize the sen-
sor in the process at the conductivity and temperature of interest.
Results under real process conditions, at different tem-
peratures, or using other sensors may differ from above.
RTD ACCURACY: Utilizing a perfect 100 Ohm RTD after
1 point temperature standardiztion, temperature read­ing can be as good as ±0.5°C.
RECOMMENDED SENSORS:
Model 222 Flow-Through Model 225 Clean-In-Place (CIP) Model 226 Submersion/Insertion Model 228 Submersion/Insertion/Retractable Model 242 Flow-Through*
* Model 242-06 or 242-08 with 5081T do not have Intrinsically Safe
approvals.
MODEL 5081T-FF/FI SPECIFICATIONS
2
NOTE: Values shown are for 25°C conductivity with a temperature slope of 2% per degree C. The maximum range
value will be lower for solutions with a higher temperature slope. Minimum conductivity depends on sensor.
RECOMMENDED RANGES FOR TOROIDAL SENSORS
Conductivity Sensor
Model Number 226 228 225 222 (1in.) 222 (2 in.) 242
Nominal Cell Constant 1.0 3.0 3.0 6.0 4.0 *
Minimum Conductivity (µµS/cm) 50 200 200 500 500 100*
Maximum Conductivity (µµS/cm) 1,000,000 2,000,000 2,000,000 2,000,000 2,000,000 2,000,000*
* Model 242 values depend on sensor configuration and wiring.
33
MODEL 5081T-FF/FI SPECIFICATIONS
TRANSMITTER DISPLAY DURING CALIBRATION AND PROGRAMMING (FIGURE 1)
1. Continuous display of conductivity, % concentration, or custom reading.
2. Units: µS/cm, mS/cm, ppm, %, or none.
3. Current menu appears here.
4. Submenus, prompts, and diagnostic readings appear hear.
5. Commands available in each submenu or at each prompt appear here.
6. Hold appears when the transmitter is in hold.
7. Fault appears when the transmitter detects a sensor or instrument fault.
INFRARED REMOTE CONTROLLER (FIGURE 2)
1. Pressing a menu key allows the user access to calibrate, program, or diagnostic menus.
2. Press ENTER to store data and settings. Press NEXT to move from one submenu to the next. Press EXIT to leave without storing changes.
3. Use the editing keys to scroll through lists of allowed settings or to change a numerical setting to the desired value.
4. Pressing HOLD puts the transmitter in hold. Pressing RESET causes the transmitter to abandon the present operation and return to the main display.
FIGURE 2. INFRARED REMOTE CONTROLLER
1.
4.
3.
2.
CALIBRATE PROGRAM DIAGNOSE
/-[5ES-U1
EXIT NEXT ENTER
mS
F A U L T
H O L D
6
7
3
4
1
2
5
FIGURE 1. TRANSMITTER DISPLAY DURING
CALIBRATION AND PROGRAMMING
The program display screen allows access to
calibration and programming menus.
#'"
REMOTE CONTROL
INTRINSICALLY SAFE EQUIPMENT HAZARDOUS AREA LOCATIONS: CLASS I, DIV 1, GP A, B, C, D CLASS I, DIV 2, GP A, B, C, D T3C Tamb = 40°C T3 Tamb = 80°C
1.5Vdc AAA BATTERIES EVEREADY E92/1212 DURACELL MN2400/PC2400
IS/I/1/A,B,C & D NI/I/2/A,B,C & D T4 Tamb = 40°C T3A Tamb = 80°C
Baseefa02ATEX0198 II 1G EExia IIC T4 1180
1.5Vdc AAA BATTERIES EVEREADY E92/1212 DURACELL MN2400/PC2400 ROSEMOUNT ANALYTICAL 92606 USA
WARNING: TO PREVENT IGNITION CHANGE BATTERIES IN A NONHAZARDOUS AREA
ONLY
IRC - INFRARED REMOTE CONTROL
LR 34186 Exia
SUBSTITUTION OF
COMPONENTS MAY
IMPAIR INTRINSIC SAFETY
PN 23572-00
YEAR
4
U17:!<[9:1
-E< /!
922<U
'"*#\=\22
/1[[!/9P<U
<Q2U
4-SG
U<[9:1
2D>[=T
'"""
25.0C 12.00mA
CAL key PROG key DIAG key HOLD key
Model 5081T-FF/FI
Process Display Screen
µS/cm
/-[5ES-U1
/1[[!/9P<U
<1P<9S!"
U17:!-G6
U17:
G5<:[-A
<1U>:!/V<=
G12->[=
PROGRAM
Process Display
FIGURE 1. Menu Tree
MODEL 5081T-FF/FI MENU TREE
PROGRAM MENU MNEMONICS
4Q[G
Transmitter on hold
2->[U
Fault condition current output setting
U1
7
:
Temperature menu header
U->U9
Automatic temperature compensation
U
7
-P
Manual temperature compensation input
G5<:[-A
Display menu header
UA:
Conductivity measurement type
U1
7
:!
°C / °F toggle selection
/9G1
Security code
922<U
Conductance Offset value
CALIBRATION
DIAGNOSTICS
MODEL 5081T-FF/FI DEFAULT SETTINGS
VARIABLE NAME MNEMONIC FACTORY SETTINGS CUSTOMER SETTINGS
Program Menu
Temperature
UHOR
Auto temperature compensation UDVUQ on ___________
Manual temperature UODP 25.0°C (overridden by auto) ___________
Temperature compensation algorithm /97: ([LPHDS or P9P1) LInEAr ___________
Display
GLTRNDY
Measurement type UYR (/QPG>/ or P-94 or 4/[ CondUC ___________
or 4$<9&[ or 4$<9&4 or /VTU)
Temperature (°C or °F) UHOR C ___________
Output (mA or %) QVURVU Cur ___________
Security Code FQGH 000 ___________
Custom Curve
<1U>:!!/V<U
___________
Reference temperature U!SHI 25.0°C ___________
Calibrate Menu
Cell constant /1[[!/QPTU 3.00 ___________
Temperature slope U17:!TN9RH 2.000 ___________
Diagnose Menu
Diagnose SAMPLE READINGS
(Each segment displays the current value in the transmitter.)
Absolute conductivity -ET 1000 µS ___________
Off Set 9II<U 0.0 µS ___________
Cell constant /1[[!/9P<U 3.00/cm ___________
Temperature slope UTNQRH 2.000 ___________
Software version TQIU A02.09 ___________
Hardware version 4-SG 01 ___________
Show fault warnings 2D>[=< none ___________
5
6
MODEL 5081T-FF/FI WIRING
WIRING
Wire sensor as shown below in Figure 2. Keep sensor wiring separate from power wiring. For best EMI/RFI protection, use shielded output signal cable in an earth-grounded metal conduit. Refer to the sensor instruction manual for more details.
FIGURE 2. Wiring Model 5081T-FF/FI
77
MODEL 5081T-FF/FI WIRING
WIRING THROUGH A JUNCTION BOX
The sensor can be wired to the analyzer through a remote junction box (PN 23550-00). Wire the extension cable and sensor cable point-to-point. Refer to the sensor instruction manual for more details.
Factory-terminated (PN 23747-00) and unterminated (PN 9200275) connecting cable are available. The use of fac­tory-terminated cable is strongly recommended. To prepare unterminated cable for use, follow the instructions in the sensor instruction manual.
For maximum EMI/RFI protection, the outer braid of the sensor cable should be connected to the outer braided shield of the extension cable. At the instrument, connect the outer braid of the extension cable to earth ground.
POWER WIRING
For general purpose areas, wire power as shown in Figure 3.
FIGURE 3. Power Supply/Current Loop Wiring for Model 5081T-FF/FI
FOUNDATION FIELDBUS COMMUNICATIONS AND POWER SUPPLY
9 - 32 VDC
8
MODEL 5081T-FF/FI INSTALLATION
INSTALLATION
UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analytical immediately.
ROTATING THE DISPLAY
The Model 5081T-FF/F1 display can be rotated 90° left or right. Disengage the cover lock and remove the front cover. Remove the three screws holding the PCB stack and gen­tly lift the display board. Do not disengage the ribbon cable between the display board and the CPU board. Rotate the display. The black infrared sensor will be at the top of the display.
INSTALLATION
See Figure 4.
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temper­atures.
2. Install the analyzer in an area where vibrations and elec­tromagnetic and radio frequency interference are mini­mized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.
4. The conduit connections on the sides of the 5081 hous­ing should be sealed to prevent moisture from entering the enclosure.
5. The transmitter must not be mounted with both conduit openings at the top
.
FIGURE 4. Mounting Model 5081T-FF/FI
INCH
MILLIMETER
Loading...
+ 16 hidden pages