Emerson 499A OZ User Manual

Model 499A OZ
Ozone Sensor
Instruction Sheet
PN 51A-499AOZ/rev.H September 2010
SPECIFICATIONS - SENSOR
Pressure: 0 to 65 psig (101 to 549 kPa abs) Temperature: 32 to 122°F (0 to 50°C) Process Connection: 1 inch MNPT Wetted Parts: Polysulfone, Viton
1
, Teflon2TFE, silicone
Cathode: gold
Type PN Wetted Materials Process Connection Maximum Temperature Maximum Pressure
1-1/2 in. tee 23567-00 CPVC, Buna N 1-1/2 in. socket
122°F (50°C)
65 psig (549 kPa abs)
915240-03 3/4 in. NFPT
2 in. tee 915240-04 PVC, Buna N 1 in. NFPT 120°F (49°C) 60 psig (515 kPa abs)
915240-05 1-1/2 in. NFPT
Low Flow* 24091-00 Polyester/polycarbonate, Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
316SS, silicone 1/4 in. OD tubing
1
Viton is a registered trademark of E.I. duPont de Nemours & Co.
2
Teflon is a registered trademark of E.I. duPont de Nemours & Co.
SPECIFICATIONS - FLOW CELLS
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
*
Pressure and temperature specifications for the low flow cell exceed the temperature and pressure specifications for the sensor.
CAUTION
Do not exceed pressure and temperature specifications. Pressure: 65 psig max (549 kPa abs max) Temperature: 32 to 122°F (0 to 50°C)
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible
with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user.
2
MODEL 499A OZ INSTALLATION
INSTALLATION
FIGURE 1. Sensor Orientation
FIGURE 2. Flow-Through 2 in. Tee
FIGURE 3. Flow-Through 1-1/2 in. Tee
Flow through 1 to 5 gpm (3.8 to 19 L/min) Open channel 1 ft/sec (0.3 m/sec) Low flow cell 2 to 5 gph (7.6 to 19 L/hr)
The Model 499AOZ sensor is NOT recommended for submersion into tanks, basins, or ponds.
FIGURE 4. Low Flow Cell (PN 24091-00)
Install the sensor in a flowing sample. Keep the sample flow as constant as possible at a value within the limits below.
MODEL 499A OZ WIRING
WIRING
3
FIGURE 5. Wiring Sensor with Standard Cable to Model 54eA Analyzer
FIGURE 6. Wiring Sensor with OptimumEMI/RFI or Variopol Cable to Model 54eA Analyzer
FIGURE 7. Wiring Sensor with Standard Cable to Model 1056 and Model 56 Analyzer
FIGURE 8. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Model 1056 and Model 56 Analyzer