Rosemount™ 3300HT/3300HTVP/ 3400HT/3400HTVP
PERpH-X High Performance pH/ORP Sensors
Quick Start Guide November 2020
Essential instructions
Read this page before proceeding!
Emerson designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. You must adhere to the following instructions and integrate them into your safety program when installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for
clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
Install equipment as specified in the installation instructions of the appropriate Quick Start Guide
and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Pressure and temperature
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level.
Do not insert or retract retractable sensors when process pressures are in excess of 64 psig (542 kPa) for option 21 or 35 psig (343 kPa) for option 25.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
2 Emerson.com/Rosemount
November 2020 Quick Start Guide
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the operator's responsibility.
Contents
Install........................................................................................................................................... 5
Start up and calibrate................................................................................................................. 25
Maintenance.............................................................................................................................. 27
Rosemount pH/ORP sensor(s) product certifications..................................................................29
Declaration of Conformity..........................................................................................................36
China RoHS tables...................................................................................................................... 38
Intrinsically safe sensor installation drawing - FM........................................................................39
Quick Start Guide 3
Quick Start Guide November 2020
4 Emerson.com/Rosemount
November 2020 Quick Start Guide

1 Install

1.1 Storage

1. Emerson recommends storing electrodes in their original shipping containers until needed.
2. Do not store at temperatures below 23 °F (-5 °C).
3. Store electrodes with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode has a limited shelf life of one year.

1.2 Installation

For sensor dimensions, see Figure 1-1 through Figure 1-4.
For sensor orientation and installation, see Figure 1-5 through Figure 1-11.
Quick Start Guide 5
Quick Start Guide November 2020
Figure 1-1: Rosemount 3300HT Sensor Dimensional Drawing
A. Sensor body B. 2-in. (50.8 mm) Swagelok® fitting process connector assembly PN
23166-00/01
C. 1-in. (25.4 mm) national pipe thread (NPT)
6 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-2: Rosemount 3300HTVP Sensor Dimensional Drawing
A. Cable with Variopol receptacle (female) connector B. Variopol plug (male) connector C. Sensor body
D. Retraction stop collar
Quick Start Guide 7
Quick Start Guide November 2020
Figure 1-3: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23240-00
A. Dimension (see Table 1-1) B. Dimension (see Table 1-1) C. Dimension (see Table 1-1)
D. Cable
E. Cable bushing polypropylene F. 1-in. by 1-in. (25.4 x 25.4 mm) Swagelok fitting kit (PN 23166-00 or
23166-01) required to connect sensor directly to process or to ball valve
G.
WARNING
Residual pressure and process may remain trapped between ball valve and male connector.
Maximum pressure at retraction:
Option 21: 64 psig (441.2 kPa) Option 25: 35 psig (241.3 kPa)
8 Emerson.com/Rosemount
November 2020 Quick Start Guide
H. O.D. titanium housing
I. O.D. retraction stop collar J. Electode housing Ryton
K. pH electrode
L. 1½-in. (38.1 mm) female pipe thread (FPT) ball valve PN 9340065
M. 1½-in. (38.1 mm) male pipe thread (MPT) close nipple
N. 1½-in. x 1-in. (38.1 x 25.4 mm) FPT reducing coupling O. Ball valve (PN 23240-00) optional
Table 1-1: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23240-00
Option Dimension A
(in./mm)
21 21.6/549 14.0/355 23.1/587
25 36.1/917 28.5/724 37.6/955
Dimension B (in./mm)
Dimension C (in./mm)
Quick Start Guide 9
Quick Start Guide November 2020
Figure 1-4: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23765-00
A. Dimension (see Table 1-2) B. Dimension (see Table 1-2) C. Housing titanium
D. Retraction stop collar
E. Electrode housing
F. pH electrode G. 1¼-in. (38.1 mm) MPT close nipple H. Ball valve
I. Packing adapter J. Knurl retainer
K. Ball valve kit PN 23765-00
10 Emerson.com/Rosemount
November 2020 Quick Start Guide
Table 1-2: Rosemount 3400HT and 3400HTVP Sensor Dimensions with Optional Ball Valve PN 23765-00
Option A: in./mm B: in./mm
21 21.6/549 12.2/310
25 36.1/917 26.7/678
Rosemount 3300HT and 3300HTVP flow through and insertion installation
1½-in. (38.1 mm) pipe tee (PN 2002011) with 1-in. (25.4 mm) threaded connections.
Note
Mount the sensor at least 10 degrees above the horizon.
Figure 1-5: Straight Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
Quick Start Guide 11
Quick Start Guide November 2020
Figure 1-6: Angle Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
B. Flow
Figure 1-7: Pipe Y Installation
A. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing B. 1½-in. (38.1 mm) pipe "Y"
12 Emerson.com/Rosemount
November 2020 Quick Start Guide
Figure 1-8: Rosemount 3400HT and 3400HTVP Mounting Details ­Retraction Version
A. Weldalet (1½-in. [38.1 mm] FPT) B. Electrode housing tip C. Electrode
D. Flow
E. Pipe "Y" F. Pipe tee
G. Flow
Quick Start Guide 13
Quick Start Guide November 2020
Figure 1-9: Rosemount 3300 and 3500VP Flow through Installation
A. 1½-in. (38.1 mm) schedule 80 CPVC tee with 1-in. (25.4 mm) female
national pipe thread (FNPT) connections (code 16) straight flow shown B. Always mount sensor at least 10 degrees above horizontal. C. 1½-in. (38.1 mm) pipe "Y"
D. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
E. Flow F. 1½-in. (38.1 mm) schedule 80 CPVC with view or cleaning port
G. Angle flow shown