Emerson designs, manufactures, and tests its products to meet many national and international
standards. Because these instruments are sophisticated technical products, you must properly install,
use, and maintain them to ensure they continue to operate within their normal specifications. You
must adhere to the following instructions and integrate them into your safety program when
installing, using, and maintaining Emerson's Rosemount™ products. Failure to follow the proper
instructions may cause any one of the following situations to occur: loss of life, personal injury,
property damage, damage to this instrument, and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
• Install equipment as specified in the installation instructions of the appropriate Quick Start Guide
and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's performance,
place the safe operation of your process at risk, and may result in fire, electrical hazards, or
improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal
injury.
Note
The information contained in this document is subject to change without notice.
WARNING
Pressure and temperature
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig
and the process temperature is lowered to a safe level.
Do not insert or retract retractable sensors when process pressures are in excess of 64 psig (542
kPa) for option 21 or 35 psig (343 kPa) for option 25.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end
users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your
system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for
all systems used within the facility.
2Emerson.com/Rosemount
November 2020Quick Start Guide
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions.
Application compatibility is entirely the operator's responsibility.
M. 1½-in. (38.1 mm) male pipe thread (MPT) close nipple
N. 1½-in. x 1-in. (38.1 x 25.4 mm) FPT reducing coupling
O. Ball valve (PN 23240-00) optional
Table 1-1: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23240-00
OptionDimension A
(in./mm)
2121.6/54914.0/35523.1/587
2536.1/91728.5/72437.6/955
Dimension B
(in./mm)
Dimension C
(in./mm)
Quick Start Guide9
Quick Start GuideNovember 2020
Figure 1-4: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23765-00
A. Dimension (see Table 1-2)
B. Dimension (see Table 1-2)
C. Housing titanium
D. Retraction stop collar
E. Electrode housing
F. pH electrode
G. 1¼-in. (38.1 mm) MPT close nipple
H. Ball valve
I. Packing adapter
J. Knurl retainer
K. Ball valve kit PN 23765-00
10Emerson.com/Rosemount
November 2020Quick Start Guide
Table 1-2: Rosemount 3400HT and 3400HTVP Sensor Dimensions with
Optional Ball Valve PN 23765-00
OptionA: in./mmB: in./mm
2121.6/54912.2/310
2536.1/91726.7/678
Rosemount 3300HT and 3300HTVP flow through and insertion
installation
1½-in. (38.1 mm) pipe tee (PN 2002011) with 1-in. (25.4 mm) threaded
connections.
Note
Mount the sensor at least 10 degrees above the horizon.
Figure 1-5: Straight Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
Quick Start Guide11
Quick Start GuideNovember 2020
Figure 1-6: Angle Flow
A. 1½-in. (38.1 mm) pipe tee with 1-in. (25.4 mm) threaded connections:
PN 2002011
B. Flow
Figure 1-7: Pipe Y Installation
A. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
B. 1½-in. (38.1 mm) pipe "Y"
12Emerson.com/Rosemount
November 2020Quick Start Guide
Figure 1-8: Rosemount 3400HT and 3400HTVP Mounting Details Retraction Version
A. Weldalet (1½-in. [38.1 mm] FPT)
B. Electrode housing tip
C. Electrode
D. Flow
E. Pipe "Y"
F. Pipe tee
G. Flow
Quick Start Guide13
Quick Start GuideNovember 2020
Figure 1-9: Rosemount 3300 and 3500VP Flow through Installation
A. 1½-in. (38.1 mm) schedule 80 CPVC tee with 1-in. (25.4 mm) female
national pipe thread (FNPT) connections (code 16) straight flow shown
B. Always mount sensor at least 10 degrees above horizontal.
C. 1½-in. (38.1 mm) pipe "Y"
D. 1½-in. x 1-in. (38.1 x 25.4 mm) reducing bushing
E. Flow
F. 1½-in. (38.1 mm) schedule 80 CPVC with view or cleaning port
G. Angle flow shown
14Emerson.com/Rosemount