Elenco Digital Bench Multimeter User Manual

DVM645BI
BENCH MULTIMETER
TAFELMULTIMETER
MULTIMETRE DE TABLE
BANCO MULTÍMETRO
TISCHMULTIMETER
User Manual
Gebruikershandleiding
Manuel d'utilisation
Gebrauchsanleitung
DVM645BI – BENCH MULTIMETER
Introduction
This manual contains instructions and warnings that must be observed to ensure safe operation and to keep the meter in a safe condition.
WARNING
READ "SAFETY INFORMATION"
BEFORE USING THE BENCH MULTIMETER
This multimeter is a portable 4000-count instrument that is designed for use in the laboratory, the field, at home, and in other environments. This multimeter features a compact design with rounded corners for easy handling, with a rugged shock­resistant and fire-retardant case. It also has electronic overload protection for all functions and ranges.
Unpacking and Inspection
Upon removing your new Bench Multimeter from its packaging, you should have the following items:
1. Bench multimeter
2. Test lead set (one black, one red)
3. Carrying strap
4. Power lead
5. Instruction manual If any of the above items are missing or are received in a damaged condition, please contact the distributor from whom you purchased the unit.
Safety Precautions
Injury or death can occur even with low voltage and low current. It is extremely important that you read this safety information before using your multimeter. Follow all safety practices and proper operating procedures for the equipment under test.
1. Exercise extreme caution when: measuring voltages above 20 volts,
measuring currents greater than 10mA, measuring AC power lines with inductive loads, measuring AC power lines during electrical storms.
2. Always inspect your multimeter, test leads and accessories for signs of
damage or abnormality before every use. If there are any abnormalities
DVM645BI GB 1
(i.e. broken or damaged test leads, cracked case, display not reading, etc.) do not attempt to do any measurements.
3. Never earth yourself when making electrical measurements. Do not touch
exposed metal pipes, outlets, fixtures, etc., which might be at earth potential. Keep your body insulated from earth by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
4. Never touch exposed wiring, connections, test probe tips, or any live
circuit conductors when making measurements.
5. Never replace the protective fuse inside the multimeter with a fuse other
than the specified or approved type. Replace only with the same type of fuse. To avoid electric shock, disconnect the test leads and any input signals before replacing the fuse.
6. Replace only with the same type of battery. To avoid electric shock,
disconnect the power lead from the mains, and the test leads and any input signal before replacing the battery.
7. Do not operate this multimeter in an explosive atmosphere (ie. in the
presence of inflammable gases or fumes, vapour or dust).
8. Measuring voltages that exceed the limits of the multimeter may damage
the meter and expose the operator to a shock hazard. Always respect the meter voltage limits as stated on the front of the meter.
9. Never apply more than 500V DC between the COM connector and earth.
10. When testing for the presence of voltage or current, make sure the
voltage or current ranges are functioning correctly. Take a reading of a known voltage or current before assuming a zero reading indicates no current or voltage.
11. Do not attempt to calibrate or service the meter unless you are trained to
do so and another person capable of rendering first aid and resuscitation is present.
12. Remember: Think Safety, Act Safely
Safety Information
The multimeter complies with Protection Class II, Overvoltage CAT II of IEC1010­1 (EN61010-1). Pollution degree 2 in accordance with IEC-664 indoor use. If the equipment is used in a non-compliant manner, the protection provided by the equipment may be insufficient.
This product complies with the requirements of the following European Community Directives: 89/336/EEC (EMC) and 73/23/EEC (LVD) as amended by 93/68/EEC (CE marking).
DVM645BI GB 2
Chapter 1 A quick tour
Explanation of Symbols
Attention ! Refer to the operating instructions
Dangerous voltage may be present at terminals
Ground (Earth Terminal)
AC - Alternating Current
DC - Direct Current
Audible Continuity
Equipment protected throughout by Double Insulation (Protection Class II)
DVM645BI GB 3
Instrument Layout
Figure 1-1. Bench Multimeter (Front)
Figure 1-2. Bench Multimeter (Back)
DVM645BI GB 4
1. V
ΩΩΩΩ
Hz Volts, Ohms, Diode, and Frequency Input Terminal
This is the positive input terminal for all functions except current measurement. The red test lead is connected to this.
2. COM Common Terminal
This is the negative (ground) input terminal for all measurement modes. The black test lead is connected to this.
3. mA Milliamp Input Terminal
This is the positive input terminal for current measurements (AC or DC) up to 400mA. The red test lead is connected to this.
4. A 10 Amp Input Terminal
This is the positive input terminal for current measurements (AC or DC) up to 10A. The red test lead is connected to this.
5. Function / Range Selector Rotary Switch
This rotary switch selects the function, and selects the desired range.
6. Function / Range Selector Buttons
This button selects the function, and selects the desired range.
7. The LCD Display
The LCD display indicates the measured value of a signal, the function mode, and indicators.
8. The Battery Cover
9. Power Supply Input with Fuse
WARNING
To avoid user injury and multimeter damage, the multimeter must be set to
the same AC voltage value as the mains before connecting the power lead
to the mains and switching on.
This switch is used to turn the AC Power on and off. When the AC Power is on, the DC Power is automatically switched off.
10. The Power Switch
DVM645BI GB 5
Using the Rotary Switch
Turn the multimeter on by turning the rotary switch to any function.
Using the Buttons
The buttons are push type switches. The functions are as follows:
1. You can turn the back light on or off by pressing the
BACK LIGHT Button
button. When powered by the battery, the back light will automatically switch off again after a short period of time.
2. In the resistance and continuity range, each press of the
FUNCTION SELECTOR Button
button switches alternately between resistance and continuity. In the frequency and adaptive test range, each press of the
button switches alternately between frequency and adaptive test. In the DC and AC amps range, each press of the
button switches alternately between DC amps or AC amps. In the DC and AC milliamps range, each press of the
button switches
alternately between DC milliamps or AC milliamps.
3. Press the
Data Hold: HOLD Button
button to toggle in and out of the Data Hold mode, except if the multimeter is already in the Min/Max Recording Hold mode. When in the Data Hold mode, the "
" indicator is displayed and the last reading is held on the display, the beeper emits a tone. Pressing the
button when the multimeter is in the Data Hold mode will cause it to exit Data Hold and enter the Min/Max Recording Hold mode. In the Min/Max Recording Hold mode, press the readings, press
again to resume recording.
button to stop recording the
However, the multimeter is still operating in Data Hold mode, therefore, the buzzer will still sound if the over range or continuity mode occurs. And the range will also change if it is in auto range.
4.
Delay Data Hold:
Press the
button to toggle in and out of the Data Hold mode, and the " "
DELAY HOLD Button
indicator turns on or off.
DVM645BI GB 6
5. Min/Max Recording Hold: Press the
button to enter the Min/Max Recording Hold mode. The minimum
MIN/MAX Button
and maximum values are then reset to the current input, the readings are stored in memory, and the "
" indicator turns on. Push the button to cycle through the minimum (MIN), maximum (MAX), and current readings. The "MIN" or "MAX" indicator turns on to indicate what value is being displayed. In the Min/Max Recording Hold mode, press the
button to stop recording the readings, press again to restart recording. If recording is stopped, the minimum, maximum, or current values and the analogue display are frozen. In the Min/Max Recording Hold mode, if a new minimum value exceeds the actual minimum readings, or a new maximum value is an overload value, the minimum or maximum value will be held on the display, but the analogue display will continue to be active. If in auto range, the mode is held just before the range.
6.
Relative Display:
Press the
button to enter the Relative Display mode, the "
REL Button
REL
" indicator turns on, the displayed reading is stored as a reference value, and the display is then set to zero. In the Relative Display mode, the value shown on the LCD is always the difference between the stored reference value and the current reading. If the new reading is the same as the reference value, the display will indicate zero. The bar graph displays absolute value, not relative value. And the over range also occurs according to the absolute value. With auto-range, the range is held just before range. Press and hold down the
button for more than one second to exit the relative
mode.
RANGE Button
7. Press the
button to select the Manual Range mode and turn off the "
AUTO
" indicator. In the Manual Range mode, each time the
button is pressed, the range (and the input range indicator) increments and a new value is displayed. To exit the Manual Range mode and return to auto range, press and hold down the
button for more than one second. The "
AUTO
" indicator turns back on. The DC V, AC V, resistance and capacitance measurements are always first set to the auto range by default. The frequency measurement is always set to the auto range. The continuity, diode, ADP, DC A and AC A measurements are always set to the manual range. When the power is turned on, the default range is selected.
DVM645BI GB 7
8. Data Memory: Press the
button to enter the Data Memory mode and turn on the "MEM"
MEM STO Button
indicator. All the display data is then stored in the memory. The stored data can be read out and displayed on the LCD, which then switches to the Data Recall Hold mode.
9. Press the "
Data Recall Hold:
button to enter the Data Recall mode and turn on the "
" indicator. The stored data can be read out and displayed on the LCD, which
MEM RCL Button
MEM
" and
then switches to the Data Recall Hold mode. While storing data in the memory display, the " Recall Hold mode is exited by pressing the
MEM
" indicator flashes. The Data
button or the button.
Buzzer Output
A 2kHz buzzer sounds in the following cases:
1. When the key is operated.
2. When the over-range occurs, except for resistance (ΩΩΩ), frequency (Hz), continuity (
) and diode ( ) measurements.
3. When the continuity measurement results in less than 40Ω.
4. When the relative mode is exited.
5. When the mode is changed from manual to auto range by operating the button. Please refer to the timing chart for the timing.
Low Battery Indication
The "
" indicator is displayed when the battery voltage falls below the reliable
operating level.
DVM645BI GB 8
Chapter 2 Measurements
Introduction
WARNING To avoid user injury and multimeter damage, the multimeter must be set to the correct value of mains voltage before connecting the power lead to the mains and switching on.
Before doing any measurements always examine the multimeter and accessories for any damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation and make sure the lead plugs fit snugly into the multimeter sockets. If there are any abnormalities, do not attempt to do any measurements.
DVM645BI GB 9
Measuring DC Volts
Circuit
Red
Figure 2-1. Measuring DC Volts
WARNING To avoid possible electric shock, multimeter damage and/or equipment damage, do not attempt to do any voltage measurements if the voltage is above 1000V DC / 750V AC RMS. 1000V DC / 750V AC RMS are the maximum voltages that this multimeter is designed to measure.
c Turn the Selector to Volts DC d Connect the leads as shown e
is not available in Volts DC, all other buttons can be used.
1. Insert the black and red test leads into the COM and VΩΩΩ
Hz input terminals
respectively.
2. Select the desired DC voltage range, or set to auto range.
3. When the magnitude of the voltage to be measured is unknown, always start with the highest range.
4. Connect the test leads in parallel with the circuit to be measured. Be careful not to touch any live conductors with your hands. Note the reading.
5. For DC voltage readings, the red lead should be connected to the positive side of the circuit, the black lead to the negative side. A minus sign on the left-hand side of the LCD will appear if the leads are connected the other way around.
6. When all measurements have been completed, disconnect the test leads from the circuit under test. Remove the test leads from the multimeter.
DVM645BI GB 10
Measuring AC Volts
Circuit
Red
Figure 2-2. Measuring AC Volts
WARNING To avoid possible electric shock, multimeter damage and/or equipment damage, do not attempt to do any voltage measurements if the voltage is above 1000V DC / 750V AC RMS. 1000V DC / 750V AC RMS are the maximum voltages that this multimeter is designed to measure.
c Turn the Selector to Volts AC d Connect the leads as shown e
is not available in Volts AC, all other buttons can be used.
1. Insert the black and red test leads into the COM and VΩΩΩ
Hz input terminals
respectively.
2. Select the desired AC voltage range, or set to auto range.
3. When the magnitude of the voltage to be measured is unknown, always start with the highest range.
4. Connect the test leads in parallel with the circuit to be measured. Be careful not to touch any live conductors with your hands. Note the reading.
5. When all measurements have been completed, disconnect the test leads from the circuit under test. Remove the test leads from the multimeter.
DVM645BI GB 11
Measuring DC and AC Amps
Circuit
Figure 2-3. Measuring Amps
Red
WARNING Do not attempt to measure high currents. Use a current clamp for measurements >10A. The 10A input terminal is protected by a F15A/250V fast blow ceramic fuse.
c Turn the selector to DC and AC Amps. d Connect the leads as shown e
is not available in DC and AC Amps, all other buttons can be used.
Each press of the button
switches alternately between DC and AC Amps.
1. Insert the black and red test leads into the COM and A input terminals respectively.
2. Turn off or disconnect the circuit to be measured from all power sources, connect the multimeter in series with the conductor where the current is to be measured.
3. Switch on the power to the circuit under test. Note the reading.
4. After completing the measurement, turn off the power to the circuit under test, disconnect the test leads from the multimeter.
5. The DC and AC Amps measurements are always fixed range.
DVM645BI GB 12
Measuring DC and AC Milliamps
Circuit
Figure 2-4. Measuring Milliamps
Red
WARNING Do not attempt to measure high currents. The mA input terminal is protected by a F1A/250V fast blow ceramic fuse.
c Turn the selector to DC and AC Milliamps. d Connect the leads as shown. e
is not available in DC and AC Milliamps, all other buttons can be used.
Each press of the button
switches alternately between DC or AC Milliamps.
1. Insert the black and red test leads into the COM and mA input terminals respectively.
2. Select the desired DC current range or AC current range.
3. Turn off or disconnect the circuit to be measured from all power sources, connect the multimeter in series with the conductor where the current is to be measured.
4. Switch on the power to the circuit under test. Note the reading.
5. After completing the measurement, turn off the power to the circuit under test, disconnect the test leads from the multimeter.
6. The DC and AC Milliamps measurements are always fixed range.
DVM645BI GB 13
Measuring Frequency and Adaptive Measurement
Circuit
Figure 2-5. Measuring Frequency and Adaptive Measurement (ADP)
Red
c Turn the selector to frequency and adaptive measurement. d Connect the leads as shown e
is not available in frequency and adaptive measurement, all other buttons
can be used.
Each press the button
switches alternately between frequency and adaptive
measurement.
1. Insert the black and red test leads into the
COM
and VΩΩΩΩ
Hz
input terminals
respectively.
2. In the frequency test, it is not necessary to know the range, as the multimeter will automatically range up or down to display the best resolution. Manual range is not available.
3. Ensure that the amplitude level of the signal to be measured is not greater than the input voltage limit (250V DC/AC rms). The signal amplitude must also be greater than the sensitivity level.
4. Attach the probe tips to the points across which the frequency is to be measured, and read the result directly from the display.
5. The frequency range is always set to auto range.
The Adaptive (ADP) Measurement is as follows:
1. Connect the ADP signal terminal (-) and (+) to the COM and VΩΩΩ
Hz input
terminals respectively.
2. The ADP voltage is supplied to the multimeter directly. The result is displayed on a scale of 10 units per 1mV.
3. The ADP measurement is always fixed range. The full range is 400mV DC.
DVM645BI GB 14
Measuring Capacitance
Red
Figure 2-6. Measuring Capacitance
WARNING Turn off the power to the device under test and discharge all capacitors.
c Turn the selector to capacitance. d Connect the leads as shown: insert the black and red test leads into the
Hz
and VΩΩΩΩ e
is not available with capacitance, all other buttons can be used.
input terminals respectively.
COM
1. Turn off the power to the device under test and discharge all capacitors.
2. Completely discharge the capacitor before measuring its capacitance value.
3. Set to the capacitance range that gives the most accurate reading or select auto range.
4. Connect the crocodile clips to the capacitor leads or insert the capacitor leads into the multimeter measuring socket. Always observe the correct polarity when measuring electrolytic capacitors.
5. Read the capacitance value directly from the display. The accuracy of the capacitance measurement can be improved by first pressing the
button (Relative mode) in order to zero the display and automatically subtract the residual multimeter and test lead capacitance. The Relative mode also selects manual range.
6. Residual charge in the capacitor, or capacitors with poor insulation resistance or poor dielectric absorption may cause measurement errors.
NOTE:
A safe way to discharge a capacitor is to connect a 100kΩ resistor across
the two capacitor leads.
DVM645BI GB 15
Testing Diodes
Red
Figure 2-7. Diode Test
WARNING
Measurements must always be done with the circuit power OFF.
c Turn the selector to Diode. d Connect the leads as shown: insert the black and red test leads into the
and VΩΩΩΩ e
Hz
input terminals respectively.
and are not available in diode test, all other buttons can be used.
COM
1. Select the (
2. Insert the black and red test leads into the COM and VΩΩΩ
) position by turning the rotary switch.
Hz input terminals
respectively.
3. The red lead should be connected to the anode and the black lead to the cathode of the diode.
4. The typical voltage drop should be about 0.6V for a silicon diode or 0.3V for a germanium diode.
5. If the diode is reverse biased or there is an open circuit and the reading will be between 3.000V and 3.400V.
6. The Diode test is always fixed range.
DVM645BI GB 16
Measuring Resistance and Continuity
Red
Figure 2-8. Measuring Resistance and Continuity
c Turn the selector to Resistance and Continuity measurement. d Connect the leads as shown. e
is not available in continuity test, all other buttons can be used.
Each press of the button
switches alternately between resistance and
continuity measurement.
Resistance measurement is as follows:
1. Select resistance by pressing the
2. Insert the black and red test leads into the
button.
COM
and
VΩΩΩ
Hz
input terminals
respectively.
3. Select the desired ohms () range or set to auto range. Connect the black and red test probes to the circuit or device under test. First make sure that the circuit or device is dead.
4. Turn off any power to the resistor to be measured. Discharge any capacitors. Any voltage present during a resistance measurement will cause inaccurate readings and could damage the meter if the overload protection of 250V DC or AC rms is exceeded.
5. Open circuits will be displayed as an overload condition and the MSD (Most Significant Digit) will blink.
DVM645BI GB 17
6. The resistance of the test leads can diminish accuracy in the lowest (400Ω) range. The error is usually 0.1 to 0.2 for a standard pair of test leads. To determine the error, short the test leads together and the use the
Relative mode to automatically subtract the lead resistance from the resistance measurement.
7. After all measurements have been completed, disconnect the test leads from the circuit and multimeter input terminals.
The Continuity (
1. Select the continuity measurement by pressing the
2. Insert the black and red test leads into the
) measurement is as follows:
COM
and VΩΩΩΩ
button.
Hz
input terminals
respectively.
3. Connect the black and red test probes to the circuit or device under test. First make sure that the circuit or device is dead.
4. An audible tone will sound when the resistance is less than approximately 40.
5. After the continuity measurement has been completed, disconnect the test leads from the circuit and multimeter input terminals.
6. The continuity measurement is always fixed range. Open circuit voltage is approximately 0.45V.
DVM645BI GB 18
Chapter 4 Specifications
General Specifications
Maximum voltage between terminals and earth:
rms (sine)
Display:
Measuring rate:
Over range indications:
3 3/4 digit (4000 count) digital indication. Frequency range: 9999 counts
max. 42 segment analogue bar graph Full indicators, automatic polarity indication
Digital 2 times per second Analogue bar graph 20 times per second Capacitance 1 time per second
MSD (Most Significant Digit) blinks
CAT. II 1000V DC or 750V AC
Low Battery indication: The "
voltage drops below the reliable operating level.
Temperature coefficient: 0.15 x specified accuracy per °C
< 18°C to >28°C
Temperature:
Storage: -10°C to 60°C
Relative humidity:
Back light Safety: Power requirements Size Weight
Accuracy is given as ± (% of reading + number of least significant digit) at 18°C to 28°C, with relative humidity up to 75%. All specifications assume less than 1 year since calibration.
Operating: 5°C to 35°C
20% to 75% RH (5°C to 35°C)
" indicator is displayed when the battery
DVM645BI GB 19
Resolution and Accuracy
DC Volts (mV DC)
Range resolution Accuracy (% reading + digits)
400mV 0.1mV ± (0.3% rdg + 5d)
4V 1mV ± (0.3% rdg + 2d)
40V 10mV ± (0.3% rdg + 2d)
400V 100mV ± (0.3% rdg + 2d)
1000V 1V ± (0.3% rdg + 2d)
Input impedance: 10M, < 100pF Overload Protection: 1000V DC or 750V AC RMS
AC Volts (true rms, ac-coupled)
Range resolution Accuracy (% reading + digits)
4V 1mV
40V 10mV
400V 100mV
± (0.8% rdg + 5d) 50 to 60Hz ± (1.2% rdg + 5d) 45 to 1kHz
750V 1V
Input impedance: 10M, < 100pF Overload Protection: 1000V DC or 750V AC RMS AC coupled true RMS
DC Current
Range resolution Accuracy (% reading + digits)
4mA 1µA ± (0.8% rdg + 5d)
40mA 10µA ± (0.8% rdg + 5d)
400mA 100µA ± (0.8% rdg + 5d)
10A
10mA ± (1.5% rdg + 10d)
(20A for 30
seconds) Input protection: 1A/250V fuse for mA input, 15A / 250V fuse for A input Load voltage: 600mV max. for mA input, 900mV max. for A input
DVM645BI GB 20
AC Current
Range resolution Accuracy (% reading + digits)
4mA 1µA ± (1.5% rdg + 5d) 45 to 400Hz
40mA 10µA ± (1.5% rdg + 5d) 45 to 400Hz
400mA 100µA ± (1.5% rdg + 5d) 45 to 400Hz
10A
10mA ± (2% rdg + 10d) 45 to 400Hz
(20A for 30
seconds) Input protection: 1A/250V fuse for mA input, 15A / 250V fuse for A input Load voltage: 600mV max. for mA input, 900mV max. for A input AC coupled true RMS
Resistance
Range resolution Accuracy (% reading + digits)
400 0.1
4k 1Ω
40kΩ 10Ω
400kΩ 100Ω
4M 1k
40M 10k
± (0.5% rdg + 5d) ± (0.5% rdg + 3d) ± (0.5% rdg + 3d) ± (0.5% rdg + 3d)
± (1% rdg + 5d)
± (1.5% rdg + 10d)
Open circuit Voltage: 0.45V Input protection: 250V RMS
Continuity Test
Continuity threshold: approx. 40 Continuity threshold: 2kHz buzzer tone Input protection: 250V RMS
Diode Test
Test current: 0.6mA Open circuit voltage: approx. 3.0V Input protection: 250V RMS
DVM645BI GB 21
Capacitance
Range resolution Accuracy (% reading + digits)
4nF 1pF ± (2% rdg + 40d) in relative mode
40nF 10pF ± (2% rdg + 5d) in relative mode
400nF 100pF ± (2% rdg + 5d) in relative mode
4µF 1nF ± (0.5% rdg + 5d)
40µF 10nF ± (2% rdg + 5d) at <= 20µF
± (5% rdg + 5d) at > 20µF
Input protection: 250V RMS
Frequency Counter
Ranges: 100Hz, 1kHz, 10kHz, 100kHz, 1MHz Resolution: 0.01Hz in the 100Hz range Accuracy: ± (0.1% + 4d) Sensitivity: 100mV rms for 1Hz to 20kHz, 500mV rms for 20kHz to 1MHz Input protection: 250V RMS
Adaptive Range
Display: 10 units per 1mV DC Accuracy: ± (0.3% + 5d) Input protection: 250V RMS
DVM645BI GB 22
Chapter 4 Maintenance
Introduction
Any repairs or servicing not covered by this manual should only be done by qualified personnel.
Battery Replacement
WARNING To avoid electric shock, disconnect the power leads from the mains, and disconnect the test leads and any input signals before replacing the battery. Only replace with same type of battery.
When the multimeter displays the " maintain proper operation. Use the following procedure to replace the battery:
1. Disconnect the test leads from any live source, turn the rotary switch to off, and remove the test leads from the input terminals.
2. Remove the screws on the battery cover and open it.
3. Remove the dead battery and replace with a new equivalent 9V battery.
4. Two types of battery are used in the bench multimeter: one is a NEDA 1604 6F22 006P type x 1 or equivalent 9V battery, the other is IEC LR6 AM3 AA 1.5V x 6.
5. Never use the bench multimeter unless the battery cover is in place and properly fastened.
" indicator, the battery must be replaced to
DVM645BI GB 23
Fuse Replacement
WARNING To avoid electric shock, disconnect the test leads and any input signals before replacing the fuses. Replace only with the same type of fuse. The A input terminal is protected by a F 15A/250V fast blow ceramic fuse. The mA input terminal is protected by a F 1A/250V fast blow ceramic fuse.
Use the following procedure to examine or replace the multimeter fuses:
1. Turn the power switch to off, and disconnect the power cable from the mains.
2. Disconnect the test leads from any live source, turn the rotary switch to off, and remove the test leads from the input terminals.
3. Open the tools cover on the top, and open the fuse cover in the tool case.
4. Remove the blown fuse, replace with a fuse of the same size and rating. Make sure the new fuse is centred in the fuse holder.
5. Replace the blown fuse with one of the same rating.
6. The A input terminal is protected by a F 15A/250V fast blow ceramic fuse, Ø 6 x 30mm. The ceramic fuse, Ø 5 x 20mm.
7. Fuses rarely need replacement and almost always blow as a result of operator error.
8. Never use the bench multimeter unless the fuse cover is in place and fully fastened.
mA
input terminal is protected by a F 1A/250V fast blow
Power Fuse Replacement
WARNING To avoid electric shock, disconnect the power cable from the mains, and disconnect the test leads and any input signals before replacing the power fuse. Only replace with same type of fuse
Use the following procedure to examine or replace the power fuse:
1. Turn the power switch to off, and disconnect the power cable from mains.
2. Disconnect the test leads from any live source, turn the rotary switch to off, and remove the test leads from the input terminals.
3. Replace the power fuse in the power supply input. Replace the blown fuse with one of the same rating.
4. The power fuse is 80mA/250V, Fast, Ø 5 x 20mm
DVM645BI GB 24
Other notes
1. Do not use abrasives or solvents on the bench multimeter, use a damp cloth with mild detergent only.
2. If any faults or abnormalities are observed, the bench instrument should not be used and needs to be inspected.
DVM645BI GB 25
Loading...