Elenco 3 1-2 Digit with Temperature User Manual

Page 1
OPERATOR’S
INSTRUCTION MANUAL
M-2625
AUTO RANGING
DIGITAL MULTIMETER
with Temperature Probe
Copyright © 2007 Elenco®Electronics, Inc.
Page 2
Contents
1. Safety Information 3,4
2. Safety Symbols 5
3. Front Plate of the Meter 5 Input Terminal 6 Function Rotary Switch 7 Function Press Button 8,9 LCD Display 10
4. Power On Option 11
5. Auto Power Off 11
6. Measurement Range 11
7. Making Measurements 12 Preparation 12 Measuring Voltage 12
Measuring DC Voltage 13 Measuring AC Voltage 14,15
Measuring Current 16
Measuring DC Current 17-19
Measuring AC Current 19-21 Measuring Resistance 21,22 Testing Diodes 23 Testing for Continuity 24 Measuring for Capacitance 25,26 Measuring Frequency 27 Measuring Duty Cycle 28,29 Measuring Temperature 29,30
8. Maintenance 31 Battery Replacement 31 Fuse Replacement 32
9. General Specifications 33-35
-2-
Page 3
Introduction
This series LCD auto ranging and auto power-off digital multimeter is a portable, compact, 3¾ digit multimeter. It is ideally suited for field, lab, shop, car, and home.
CAUTION: Read, understand and follow all Safety Rules and Operating Instructions in this manual before using this product.
1. Safety Information
Use the meter only as specified in this manual. Otherwise, the protection provided by the meter may be impaired. Refer to safety information in Table 1.
Table 1. Safety Information
Warning
!
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use the meter if it is damaged. Before you use the meter,
inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal.
Check the test leads for continuity. Replace damaged test leads before you use the meter.
• The RESPONSIBLE BODY shall be made aware that, if the
equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
• Do not use the meter if it operates abnormally. Protection may be
impaired. When in doubt, have the meter serviced.
• Do not operate the meter around explosive gas, vapor, or dust.
• Do not apply more than the rated voltage, as marked on the meter,
between terminals or between any terminal and earth ground.
• Before use, verify the meter’s operation by measuring a known
voltage.
• When measuring current, turn off circuit power before connecting
the meter in the circuit. Remember to place the meter in series with the circuit.
-3-
Page 4
Table 1. Safety Information (continued)
Warning
• When servicing the meter, use only specified replacement parts.
• Use caution when working above 30VAC rms, 42V peak, or 60VDC.
Such voltages pose a shock hazard.
• Avoid working alone.
• When using the probes, keep your fingers behind the finger guards
on the probes.
• Connect the common test lead before you connect the live test lead.
When you disconnect test leads, disconnect the live test lead first.
• Remove test leads from the meter with the battery door or portions
of the cover removed or loosened.
• To avoid false readings which could lead to possible electric shock
or personal injury, replace the batteries as soon as the low battery indicator ( ) appears.
• Use only type “AA” batteries or equivalent, properly installed in the
meter case, to power the meter.
• To avoid the potential for fire or electrical shock, do not connect the
thermocouples to electrically live circuits.
+
!
Caution
To avoid possible electric shock or personal injury, follow these guidelines:
• Disconnect circuit power and discharge all high-voltage capacitors
before testing resistance, continuity, diodes, or capacitance.
• Use the proper terminals, function, and range for your
measurements.
• Before measuring current, check the meter’s fuses and turn power
OFF to the circuit before connecting the meter to the circuit.
-4-
Page 5
2. Safety Symbols
Symbols used on the meter and in this manual are explained in Table 2.
Table 2. International Electrical and Relative Symbols
AC (Alternating Current) Important Information
~
DC and AC
~
DC (Direct Current) Earth Ground
Equipment Safety Standard
Double Insulated
3. Front Plate of Meter
LCD Display
Function Press Button
Function Rotary Switch
!
Caution, risk of electric shock
Low Battery Indicator
+
Input Terminal
Figure 1. Front plate of the meter
-5-
Page 6
Input Terminal
About terminal function refer to Table 3 and Figure 2.
Figure 2. Input Terminal
Table 3. Input Terminal
Terminal Function Description
/ /)/OF
V /Ω/ Hz
mA
COM
10A
Volts, ohm, diode, frequency, DUTY, temperature and capacitance measurement, and testing for continuity terminal
Milliampere current, microampere current measurement terminal
Common terminal for all measurements
Ampere current measurement terminal
Function Rotary Switch
About function rotary switch refer to Table 4 and Figure 3.
-6-
Page 7
Table 4. Function Rotary Switch Symbols
Position of Switch
V
~
)
Hz
Figure 3. Function Rotary Switch
Function
DC/AC voltage measurement
Resistance measurement, diode test, testing for continuity
Capacitance measurement
Frequency measurement / Duty cycle measurement
-7-
Page 8
Table 4. Function Rotary Switch Symbols (continued)
Position of Switch
O
F
µµ
A
~
mA
~
A
~
OFF
Temperature measurement
DC/AC microampere current measurement
DC/AC milliampere current measurement
DC/AC ampere current measurement
Power off position
Function
Function Press Button
About press function button refer to Figure 4.
Figure 4. Function Press Button
1. SELECT (state choice)
There are several functions at one rotary switch position; press the
SELECT button to select the desired function: ~V, A~ range for AC or DC voltage, AC or DC current, resistance,
diode, or audible continuity measurement.
2. RANGE (range selection)
With the power on and the meter on the auto range mode, press the RANGE button to select the desired measurement range manually.
The range goes from minimum to maximum by pressing the RANGE button multiple times, and returns to minimum after reaching the maximum range. Pressing and holding the RANGE button for over two seconds returns the meter to auto range. This button does not affect the frequency, capacitance, diode or continuity range.
, µµA~, mA~,
-8-
Page 9
3. REL(relative measurement) Press the REL a “REL” symbol appears on the LCD. The result of the relative measurement is the difference between the measuring value and the reference value. The reference value is produced the same as the momentary value by press the REL button again to exit this function and the “REL” symbol disappears from the display. This button does not affect the frequency, duty cycle, diode and continuity test functions.
4. HOLD (data hold) Pressing the HOLD button enables the data hold function and an “ H ” symbol appears on the display. The momentary value appears on the display. Press the HOLD button again to exit this function and the “ H ” symbol disappears from the display. This button does not affect the frequency measurement function.
button to use the relative measuring function and
button. Press the REL
5. Hz/DUTY When making a frequency, DC or AC voltage measurement, press this button to switch between the Hz or duty cycle test mode.
6. RESET (reset button) Press this button to clear all values and to restart the meter.
(frequency and duty cycle measurement switch button)
LCD Display
About function rotary switch refer to Table 5 and Figure 5.
-9-
Page 10
Figure 5. LCD Display
Table 5. LCD Display Symbols
Sequ. No. Symbol Description
1 8.8.8.8
2
3 AC
4
+
5 RS232
6 Auto
7 REL
8
H
9
10
11 hFE
12
/ kΩ/ M
Hz / kHz / MHz
The display digit group
Negative symbol
AC (alternating current) indicator
Battery power is weakening
Serial transmission (not used on this meter)
Auto range indicator
Relative measurement
Data hold
Diode symbol
Audible continuity function indicator
Transistor gain indicator (not used)
Resistance units
Frequency units
-10-
Page 11
Table 5. LCD Display Symbols (continued)
Sequ. No. Symbol Description
13
O
C OF
14 %
mV / V
15
nF / µµF
mA / A
16
Temperature units
O
C not used on this meter)
(
Percent symbol
Voltage units Capacitance units Current units
Overload
4. Power On Option
To release the Auto Power Off function, press and hold the
SELECT button when the meter is on.
5. Auto Power Off
When the meter has been on for 30 minutes without any action from the user, the meter will automatically switch to the standby mode. To bring back the display, turn the rotary switch to a different position, push any button, or restart the meter.
6. Measurement Range
A measurement range determines the maximum measurement value. Most of the functions of the meter have more than one range.
1. It is important to select the correct range
• If the range is too small, “OL” shows on the display for overload.
• If the range is too large, the value of the reading will be very small with less resolution.
2. Auto range and manual range options
• In the auto range mode, the meter automatically selects the best range for the input detected.
• In the auto range mode, press the RANGE button to enter the manual range and select the desired range.
-11-
Page 12
• When you change the measurement function, the meter will regard its auto range as current range.
• In the auto range mode, the “AUTO” symbol will appear on display and the manual range disappears.
3. Enter or exit manual range mode
• Press the RANGE button to select the manual range mode and the “AUTO” symbol disappears. Each press of the RANGE button increases the range. When the maximum range is reached, the meter restarts at the lowest range.
• Press and hold the RANGE button for more than two seconds to return to the autorange mode and the “AUTO” symbol will appear on the display.
7. Making Measurements
Preparation
Checking the Battery Voltage
Rotate the Function Rotary Switch to any position away from the OFF position. The battery voltage is adequate when the indication is clear and the symbol is not on the display.
+
If the symbol appears on the display, or nothing shows on the display at all, follow the battery replacement procedure shown in related items in this manual and replace with new batteries specified.
Measuring Voltage
Voltage is the difference in electrical potential between two points. The polarity of AC (alternating current) voltage varies over time, while the polarity of DC (direct current) voltage is constant.
+
Danger
!
To avoid the danger of electric shock, never make measurements on a circuit over 600V AC/DC (electrical potential to ground 300V DC/AC). Do not operate the Function Rotary Switch during measurement. Do not make a measurement when opening the battery cover and the meter case.
-12-
Page 13
Measuring DC Voltage
1. Turn on the meter, then set up the meter to measure DC volts as shown in Figure 6.
2. Insert the black test lead into the COM terminal and the red test lead into the terminal.
3. Set the Function Rotary Switch to the ~V position (then, “AUTO” and “mV” symbols are indicated on the display).
4. Connect the black test lead to the negative side of the circuit under test and the red test lead to the positive side of the circuit. The measured value is indicated on the display. If you connect the test leads to the opposite polarity, the “–” symbol appears on the display.
Figure 6. DC Voltage Measurement
-13-
Page 14
Measuring AC Voltage
1. Turn on the meter, then set up the meter to measure AC volts as shown in Figure 7.
2. Insert the black test lead into the COM terminal and the red test lead into the terminal.
3. Set the Function Rotary Switch to the ~V position and press SELECT to select the AC voltage measuring mode (then, “AUTO”, “AC” and “V” symbols are indicated on the display).
4. Connect the test leads to the circuit under test, then the measured value is indicated on the display.
5. When measuring voltage less than 400mV, press the RANGE button to switch to manual range mode and select the AC 400mV range. Doing so will provide a better resolution for your measurement.
6. You can press the Hz/DUTY button to read the signal frequency or duty cycle under measurement from the display.
-14-
Page 15
Figure 7. AC Voltage Measurement
Note: When taking a measurement of less than 20mV at the 400mV
AC range, the measurement value cannot be indicated correctly. Even if shorted, the input line at the 4V AC range, 1 ~ 3 digits may remain indicated. In that case, by pressing the REL
be indicated.
button, “0” will
-15-
Page 16
Measuring Current
Warning
!
Never attempt an in-circuit current measurement where the open-circuit potential to earth is greater than 300V. You may damage the meter or be injured if the fuse blows during such a measurement.
Caution
To avoid possible damage to the meter or to the equipment under test, check the meter’s fuses before measuring current. Use the proper terminals, function, and range for your measurement. Never place the probes across (in parallel with) any circuit or component when the leads are plugged into the current terminals. Maximum measurement time allowed at the 10A current range function is 15 seconds. If you carry on making a measurement continuously over 15 seconds, it can cause an erroneous measurement and damage the meter.
Danger
!
Do not operate the Function Rotary Switch during measurement. Do not make a measurement when opening the battery cover or the meter case.
Current is the flow of electrons through a conductor. To measure current, you must open the circuit under test, then place the meter in series with the circuit.
To measure AC or DC current, proceed as follows:
-16-
Page 17
Measuring DC Current
1. Turn off power to the circuit. Discharge all high-voltage capacitors.
2. Turn on the meter, then set up the meter to measure DC current as shown in Figure 8.
3. Insert the black test lead into the COM terminal and the red lead into an input appropriate for the measurement range as shown in Table 6.
4. Connect the black test lead to the negative side of the circuit under test and the red test lead to the positive side of the circuit so the meter is in series with the circuit.
5. Turn the on the power for the circuit under test.
6. The measured value is indicated on the display.
If you connect the test leads to the reverse polarities, a “–” symbol will be shown on the display.
Note: To avoid blowing the meter’s 500mA fuse, use the mA terminal only if you are sure the current is less than 400mA.
Table 6. Current Measurement
Rotary Switch Input Ranges
µµ
A
mA
A
~
~
~
mA
mA
10A
40µA / 400µA
40mA / 400mA
4A / 10A
-17-
Page 18
Figure 8. DC Current Measurement
-18-
Page 19
Figure 8. DC Current Measurement (continued)
Measuring AC Current
1. Turn off power to the circuit. Discharge all high-voltage capacitors.
2. Turn on the meter, then set up the meter to measure AC current as shown in Figure 9.
3. Insert the black test lead into the COM terminal and the red lead in an input appropriate for the measurement range as shown in Table 6.
4. Connect the test leads to the circuit under test so the meter is in series with the circuit.
5. Set the meter to AC mode by pressing the SELECT button (then, “AUTO” “AC” and “µA”, “mA”, or “A” symbols are indicated on the display).
6. Turn on the power to the circuit under test.
7. The measured value is shown on the display.
-19-
Page 20
Note: To avoid blowing the meter’s 500mA fuse, use the mA terminal only if you are sure the current is less than 400mA.
Figure 9. AC Current Measurement
-20-
Page 21
Figure 9. AC Current Measurement (continued)
Measuring Resistance
Caution
To avoid possible damage to the meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance.
Resistance is an opposition to current flow. The unit of resistance is the ohm (). The meter measures resistance by sending a small current through the circuit.
-21-
Page 22
1. To measure resistance, turn on the meter, then set up the meter as shown in Figure 10.
2. Insert the black test lead into the COM terminal and the red test lead into the terminal.
3. Set the Function Rotary Switch to the position (then, “AUTO” “OL” and “M” symbols are indicated on the display).
4. Connect the test leads to both ends of the resistance under test, then the measured value is shown on the display.
Keep the following in mind when measuring resistance:
• Because the meter’s test current flows through all possible paths between the probe tips, the measured value of a resistor in a circuit is often different from the resistor’s rated value.
• The test leads can add 0.1to 0.2of error to resistance measurements. To test the leads, touch the probe tips together and read the resistance of the leads. If necessary, you can press the REL
button to automatically subtract its value.
Figure 10. Resistance Measurement
-22-
Page 23
Testing Diodes
Caution
To avoid possible damage to the meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, silicon controlled rectifiers (SCRs), and other semiconductor devices. The test sends a current through a semiconductor junction, then measures the junction’s voltage drop. A typical junction drops 0.5V to 0.8V.
To test a diode out of a circuit, set up the meter as shown in Figure
11. In a circuit, a similar diode should still indicate a forward-bias reading of 0.5V to 0.8V; however, the reverse-bias reading can vary depending on the resistance of other pathways between the probe tips.
Figure 11. Diode Test
-23-
Page 24
Note: Open-circuit voltage between measuring terminals is approximately 1.5V (measuring current approximately 1.5mA).
Testing for Continuity
Caution
To avoid possible damage to the meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing for continuity.
Continuity is the presence of a complete path for current flow. The continuity test features a beeper that sounds if a circuit is complete (the resistance less than 120approx.). The beeper allows you to perform quick continuity tests without having to watch the display.
The continuity function detects intermittent opens and shorts lasting as little as 1 millisecond (0.001 second). These brief contacts cause the meter to emit a short beep. Refer to Figure 12 for continuity testing setup instructions.
Figure 12. Continuity Test
-24-
Page 25
Keep the following in mind while doing a continuity test:
• Even with the test leads shorted, the indicated value may not be “0”. This is because of the resistance of the test leads and not a fault. If necessary, you can press the REL automatically subtract its value, then “0” will be indicated.
button to
Measuring Capacitance
Caution
To avoid possible damage to the meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC voltage function to confirm that the capacitor is indeed discharged.
Capacitance is the ability of a component to store an electrical charge. The unit of capacitance is the farad (F). Most capacitors are in the nanofarad (nF) to microfarad (µF) range. The meter measures capacitance by charging the capacitor with a known current for a known period of time according to the measuring capacitance:
Measuring capacitance <4µF Measuring time is about 2 seconds Measuring capacitance <40µF Measuring time is about 7 seconds Measuring capacitance <100µF Measuring time is about 15 seconds
To measure capacitance, proceed as follows:
1. Turn on the meter, then set up the meter as shown in Figure 13.
2. Insert the black test lead into the COM terminal and the red test lead into the terminal.
3. Set the Function Rotary Switch to the ) position (then, “AUTO” and “nF” symbols are indicated on the display).
4. Connect the test leads to both ends of the capacitor under test, then the measured value is shown on the display. If the capacitor is polarized, connect the red test lead to the positive lead and the black test lead to the negative lead.
The following are some tips for measuring capacitance:
• To improve the measurement accuracy of small value capacitors, press the REL to subtract the residual capacitance of the meter and leads.
button with the test leads open
-25-
Page 26
Figure 13. Capacitance Measurement
-26-
Page 27
Measuring Frequency
Frequency is the number of cycles a signal completes each second.
The meter measures the frequency of a voltage or current signal by counting the number of times the signal crosses a threshold level each second.
To measure frequency, proceed as follows:
1. Turn on the meter, then set up the meter as shown in Figure 14.
2. Insert the black test lead into the COM terminal and the red test
lead into the terminal.
3. Set the Function Rotary Switch to the Hz position (then, the “Hz”
is shown on the display).
4. Connect the test leads to the circuit under test. The meter auto
ranges to one of the frequency ranges, then the measured value is shown on the display.
Figure 14. Frequency Measurement
-27-
Page 28
Measuring Duty Cycle
Duty cycle (or duty factor) is the percentage of time a signal is above or below a trigger level during one cycle (Figure 16).
To measure duty cycle, proceed as follows:
1. Turn on the meter, then set up the meter as shown in Figure 15.
2. Insert the black test lead into the COM terminal and the red test
lead into the terminal.
3. Set the Function Rotary Switch to the Hz position. Then press
Hz/DUTY once to select DUTY Cycle measurement (then, a “%” symbol is indicated on the display).
4. Connect the test leads to the circuit under test. The measured
value is shown on the display.
Figure 15. Duty Cycle Measurement
-28-
Page 29
Figure 16. Duty Cycle Measurement
Measuring Temperature
Warning
!
To avoid the potential of fire or electrical shock, do not connect the thermocouples to electrically live circuits.
To measure temperature, proceed as follows:
1. Turn on the meter, then set up the meter as shown in Figure 17.
O
2. Set the Function Rotary Switch to the
“AUTO” and “OF” symbols are indicated on the display).
3. Plug the positive leg (+) of the type K thermocouples into the
and read the value on the display. The range of the TP-03 thermocouples is –58
4. Opening the input will display the temperature at the meter
terminals.
terminal and the negative leg (–) into the COM terminal
O
F~+752OF.
F position (then, the
-29-
Page 30
Figure 17. Temperature Measurement
-30-
Page 31
8. Maintenance
General Maintenance
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture in the terminals can affect readings. Clean the terminals as follows:
1. Turn the meter off and remove all test leads.
2. Shake out any dirt that may be in the terminals.
3. Soak a new swab with alcohol. Work the swab around each
terminal.
Replacing the Batteries
Replace the batteries with two “AA” batteries (R6P) or equivalent.
Warning
!
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the low battery indicator ( ) appears.
Replace the batteries as follows (refer to Figure 18):
1. Turn the Function Rotary Switch to the OFF position and remove
the test leads from the terminals.
2. Remove the battery door by using a screwdriver to turn the
battery door screw counterclockwise.
3. Replace the batteries. Secure the door by turning the screws
clockwise.
+
Screw
Battery: Two R6P (AA) or equivalent
Figure 18. Battery Replacement
-31-
Page 32
Replacing the Fuses
Warning
!
To avoid electrical shock or damage to the meter, only use replacement fuses 10A/250V FAST.
Referring to Figure 19, examine or replace the meter’s fuses as follows:
1. Turn the Function Rotary Switch to the OFF position and remove
the test leads from the terminals.
2. Remove the holster from the meter. Then, using a screwdriver,
unscrew the two screws on the back case and remove it.
3. Remove either fuse by gently prying one end loose, then sliding
the fuse out of its bracket.
4. Install ONLY the specified replacement fuses with the amperage,
voltage, and speed ratings: 10A/250V FAST.
5. Reinstall the back case. Secure the back case by turning the
screws clockwise.
Screw
5x20 0.5A/250V FAST or ∅6x25
5x20 0.5A/250V FAST or ∅6x25
6x25 10A/250V FAST
®
Elenco
part # 531101
5x20 0.5A/250V FAST
®
Elenco
part # 533004
Figure 19. Fuse Replacement
-32-
Page 33
9. General Specifications
Display: 3¾ digit LCD with a maximum reading of 3999. Range Control: Auto range or manual range control. Polarity: Automatic negative polarity indication. Zero Adjustment: Automatic Overrange Indication: Only the “OL” display. Low Battery: The symbol “ ” is displayed when the battery
voltage is below approximately 2.4V. Auto Power Off: 30 minutes after stopping the switch or no key­input, the meter automatically enters the power off mode. Safety Standards: EMC/LVD. The meter is up to the standards of IEC1010 pollution degree 2; overvoltage category II (double insulation). Operating Environment: Temperature 32 humidity <80% RH. Storage Environment: Temperature –4 humidity <85% RH.
Power: Two 1.5V “AA” batteries. Dimensions: 6 3/32”/155mm (H) x 2 15/16”/75mm (W) x 1 5/16”/
33mm (D)
Weight: Approx. 9.2oz./260g. (including battery and holster)
Detailed Accuracy Specifications
Accuracies are ± (% of reading + number in last digit) at 73OF ±41OF/23OC ±5OC, <75% RH.
The detailed accuracy specifications are shown in Tables 7 to 18.
Table 7. DCV Detailed Accuracy Specifications
Range Resolution Accuracy Note
400mV 0.1mV
4V 1mV
40V 10mV
400V 100mV
600V 1V ±(0.8% rdg + 5)
+
±(0.5% rdg + 5)
O
- 104OF (0Oto 40OC),
O
- 140OF (–20Oto 60OC),
Input Resistance: 10M Overload Protect: 400mV range 250V rms. The rest 600VDC or 600VAC rms.
-33-
Page 34
Table 8. ACV Detailed Accuracy Specifications
Range Resolution Accuracy Note
400mV 0.1mV ±(1.5% rdg + 8) Frequency Response:
4V 1mV
40V 10mV
400V 100mV
600V 1V ±(1.2% rdg + 5)
±(0.8% rdg + 5)
50Hz - 400Hz Input Resistance: 10M Overload Protect: 400mV range 250V rms. The rest 600VDC or 600VAC rms.
Table 9. DCA Detailed Accuracy Specifications
Range Resolution Accuracy Note
40µA 0.1µA
400µA 1µA
40mA 10µA
400mA 100µA
4A 1mA
10A 10mA
±(2.0% rdg + 5)
±(1.5% rdg + 5)
±(2.0% rdg + 5)
Overload Protect: Fast fuse
0.5A/250V and fast fuse 10A/250V. 10A for 15sec. maximum Input Voltage Drop: <0.4V
Table 10. ACA Detailed Accuracy Specifications
Range Resolution Accuracy Note
40µA 0.1µA
400µA 1µA
40mA 10µA
400mA 100µA
4A 1mA
10A 10mA
±(2.5% rdg + 3)
±(2.0% rdg + 5)
±(2.5% rdg + 5)
Overload Protect: Fast fuse
0.5A/250V and fast fuse 10A/250V. 10A for 15sec. maximum Input Voltage Drop: < Frequency Response: 50Hz - 400Hz
Table 11. Resistance Detailed Accuracy Specifications
Range Resolution Accuracy Note
400 100m
4k 1
40k 10
400k 100
4M 1k
40M 10k ±(2.0% rdg + 5)
±(1.0% rdg + 5)
Overload Protect: 250V rms
0.4V
-34-
Page 35
Table 12. Capacitance Detailed Accuracy Specifications
Range Resolution Accuracy Note
40nF 10pF ±(3.5% rdg + 10)
400nF 100pF
4µF 1nF
40µF 10nF
100µF 100nF ±(3.5% rdg + 5)
±(3.0% rdg + 5)
Overload Protect: 250V rms
Table 13. Diode Detailed Accuracy Specifications
Range Description Note
Display reads approximate forward voltage of diode
Forward DC current approx. 1.5mA; Reversed DC voltage approx. 1.5V.
Overload protection: 250V rms
Table 14. Frequency & Duty Cycle Detailed Accuracy Specifications
Range Resolution Accuracy Sensitivity Note
5.12Hz
-
10MHz
0.1% - 99.9% 0.1% ±(2.5% rdg + 5) 1.5V rms
0.001Hz
-
10kHz
±(0.1% rdg + 5)
<1MHz:
0.7V rms; >1MHz:
1.5V rms
Overload Protect: 250V rms Duty Cycle: 10Hz-1kHz
Table 15. Temperature Detailed Accuracy Specifications
Range Accuracy Note
–58OF - 752OF
±0.75% ±5OF
±1.5% ±5OF
5OF (room temperature)
-35-
Page 36
Elenco®Electronics, Inc.
150 Carpenter Avenue
Wheeling, IL 60090
(847) 541-3800
Web site: www.elenco.com
e-mail: elenco@elenco.com
Loading...