The procedures in this section shall be performed in their entirety each time a component is
removed, replaced, calibrated, adjusted and during all scheduled Periodic Manufacturer’s
Certification (PMC) visits. A PMC Checklist form, P/N S010211, available from the Draeger
Medical, Inc. Technical Service Department, shall be completed by the TSR each time a PMC
is performed. Steps in the procedure marked with ( ) require a response at the corresponding
line on the checklist form.
Space is also provided on the PMC checklist form to record the results of a vapor concentration
test.
NOTE:Test equipment listed below with an asterisk (*) requires calibration at a maximum
interval of one year. Verify the dates on test equipment calibration labels. DO NOT
USE any test equipment having an expired calibration date. Notify your supervisor
immediately if any equipment is found to be out of calibration.
In the space provided at the bottom of the PMC checklist form, record the Model and
ID number of all calibrated test equipment used. Also record the calibration due
dates.
Test Equipment Required
:
*-- Multi-Meter (Fluke or equivalent)
*-- Electrical Safety Analyzer (Biotek 501 Pro or equivalent)
*-- Regulator Test Pressure Gauge, P/N S000063 or equivalent
-- Fresh Gas Outlet Volume Test Device, P/N S010158 or equivalent
-- Fresh Gas Leak Test Device, P/N 4113119 or equivalent
-- Adapter Assembly, Test Terminal, P/N 4104389 or equivalent: two are required
*-- Flowmeter Test Stand (Capnomed), P/N S000081 or equivalent
-- Breathing System Leak Test Device, P/N S010159 or equivalent
-- Baromed Pressure Test Fixture or equivalent
*-- Test Minute Volume Meter, P/N 2212300 or equivalent
*-- Digital Pressure Manometer (SenSym PDM 200CD or Equivalent)
*-- Riken Gas Indicator, Model 18H, or 1802D or equivalent
-- Stop Watch
Materials Required
:
-- Dow Corning High Vacuum Grease, P/N S4105908
-- Spiromed Sensor Lubrication Kit, P/N 2218180
-- Breathing Bag, 3 liter, P/N 9995330 or equivalent
-- Tube, Corrugated, 22 mm x 12 in. long, P/N 9995112
Test equipment illustrations are shown on the following pages.
Rev. Y
6-1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
4 1 1 3 1 1 9
F R E S H G A S L E A K T E S T D E V I C E
C O N N E C T T O
T E S T P R E S S U R E G A U G E
S V O O O 2 2
6-1ARev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
4 1 1 5 1 2 8
S I E M E N S T E S T L U N G
S V O O O 2 5
Rev. Y6-1B
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
T E S T T E R M I N A L
S 0 1 0 1 5 9
B R E A T H I N G S Y S T E M L E A K T E S T D E V I C E
S V O O O 2 3
T E S T T E R M I N A L
S 0 1 0 1 5 8
F R E S H G A S O U T L E T V O L U M E T E S T D E V I C E
6-1CRev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
S V O O O 2 7
01 0 0
S 0 0 0 0 8 1
F L O W M E T E R
T E S T S T A N D
Rev. Y6-1D
S 0 0 0 0 6 3
R E G U L A T O R T E S T
P R E S S U R E G A U G E
L
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
1 5
9 0 0
L i t e r
1 0 0
S V O O O 2 4
8 0 0
7 0 0
v o l / m i n
6 0 0
1 0
5
2 0 0
2 2 1 2 3 0 0
m l
M I N U T E
V O L U M E T E R
,
3 0 0
4 0 0
5 0 0
4 1 0 4 3 8 9
T E S T T E R M I N A
A D A P T E R
T E S T T E R M I N A L
6-1ERev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
Periodic Manufacturer’s Certification General Instructions
The purpose of these procedures is to provide detailed instructions for performing a Periodic Manufacturer’s
Certification (PMC) inspection on the Narkomed 2B anesthesia machine.
A PMC consists of a complete Periodic Manufacturer’s Service procedure and a certification level inspection
based on DMI Recommendations and equipment performance. Additional inspections are also performed to
insure proper product labeling.
Several additional documents have been created to ensure the success of this new program. Following is a brief
description of the purpose of each document.
Field Service Procedure:
Periodic Manufacturer’s Certification Forms - Part Number SP00175.
This procedure illustrates sample checklists with typical periodic maintenance items filled in, including vapor
concentrations verification tests, parts replaced, general comments and certification levels. Also included are
sample PMC labels marked to show several levels of certifications. An excerpt from DMI’s Anesthesia SystemRisk Analysis and Risk Reduction is included, and also a sample of an Executive Summary to be furnished
to the hospital’s Risk Manager or Chief of Anesthesia.
Field Service Procedure:
DMI Recommendations Guidelines Index Anesthesia Systems - Part Number S010250.
This Guideline was created to provide an assessment of each machine’s certification. It contains various
comprehensive overviews of possible equipment conditions and their associated certification levels.
The first list in the Recommendation Guidelines is a reference chart for machine certification based on
equipment status. The second is an abbreviated summary of all DMI Recommendations and Failure Codes
including the Condition Number, Equipment Condition, Recommended Corrections, Certification Code, and
Tests Affected when applicable.
There is also a matrix classified as "Failure Codes" which identifies the correct manner in which to document
equipment tests that fail, or were unable to be performed due to circumstances beyond the control of the
service technician performing the inspection. (Ex: Air cylinder supply is unavailable to perform Air High
Pressure Leak test.) The Failure Codes section also indicates suggested resolution of the situation. Failure
Code numbers begin at 34 and use the same certification levels strategy, and carry the same weight as DMI
Recommendation equipment condition codes.
Thefinal matrix is themost comprehensive index sortedby machine model and includes Equipment Condition,
Certification Code, and DMI Recommendations. It also specifies any suggested upgrade path including
ordering information that should be taken such as installing a Bellows with Pressure Limit Control
4109664-S01 Kit, after market modification kit to a machine not equipped with pressure limit control.
Rev. Y6-1F
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
The letters A, B, C, D and the Roman Numerals I, II are used as codes in the individual matrix for each model
of anesthesia machine. The letters A, B, C, and D are used in descending order to indicate the certification
level of the equipment. They are as follows:
A = Certified
B = Certified with Recommendations
C = Conditionally Certified
D = No Certification
Roman Numerals I and II do not affect the certification level but rather are provided to give further
instructions to the end user as follows:
I = The system in its present configuration shall only be used with a CO2 monitor incorporating an
apnea warning. The operator of the system is advised to frequently scan the CO2 readings and
alarm thresholds.
II = The present configuration of equipment requires that the unit operate at all times with an oxygen
analyzer that includes a low oxygen warning. The operator of the system is advised to frequently
scan the oxygen readings and alarm limits.
Following is an explanation of machine certification levels:
Certified- No recommendations apply to machine being inspected. (Only item number 33 - "No
Recommendations" shall apply for this certification level.)
Certified with Recommendations- A numbered recommendation with a code of B applies to the machine
being examined.
Conditionally Certified- A numbered recommendation with a code of BCI or BCII applies to the machine
being examined.
No Certification- A numbered recommendation with a code of D applies to the machine being examined.
When multiple recommendations apply, "No Certification" would take precedence over "Conditionally Certified"
and "Certified with Recommendations". "Conditionally Certified" would take precedence over "Certified with
Recommendations".
6-1GRev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
For example:
A Narkomed 3 could have recommendation number 21 and failure code 61.1 apply.
21 - No ventilator pressure limit control. Code is B.
61.1 - Enflurane agent is unavailable to test. Code is BC.
Correct certification for this machine is BC, which means CONDITIONALLY CERTIFIED WITH
RECOMMENDATIONS.
A Narkomed 4 could have recommendation numbers 14 and 21 apply.
14 - CO2/Agent monitor exhaust port is not properly connected to the waste gas scavenger.
Code B.
21 - No ventilator pressure limit control. Code B.
The correct certification for this machine is B, which means CERTIFIED WITH RECOMMENDATIONS.
A Narkomed 2B, 2CorGS could have recommendation 30 apply.
30 - Anesthesia machine is equipped with inhalation anesthesia vaporizers without an agent analyzer in the
breathing system. Code B.
The correct certification for this machine is B, which means CERTIFIED WITH RECOMMENDATIONS.
A Narkomed 6000 could have no DMI recommendations or failure codes apply. The correct certificaiton level
for this machine is Code A, CERTIFIED.
Code D, which means NO CERTIFICATION, also means the machine shall not receive a Periodic
Manufacturer’s Certification label. The machine shall also receive a ’WARNING - This System is
Not Certified’ label, P/N 4114857. This label shall be placed at a prominent location on the right
side of the machine after all other previous PM and ’Vigilance Audit® Validation’ labels have been
removed.
Rev.6-1H
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
PM Certification Procedure for Narkomed 2B Anesthesia System
1.Use the PM Certification form for Narkomed 2B/ 2C/ GS (P/N S010211) Anesthesia Systems (P/N S010211).
2.Completely fill in the header information.
3.Determine if the ventilator has an MJV-2 square Clippard valve. If ventilator has an MJV-2, perform the lubrication
procedure in accordance with SP00062. Write in the date of lubrication next to the "MJV-2 LUBRICATION" line
on the Periodic Manufacturer’s Certification form. If the ventilator has a Humphrey valve, indicate so with a (H)
next to the "MJV-2 LUBRICATION" line on the Periodic Manufacturer’s Certification form.
4.Replace the VENTILATOR RELIEF VALVE DIAPHRAGM in accordance with SP00075. Place a check mark and
thereplacementdateat"VENTRELIEF VALVE REPLACEMENT" lineon the Periodic Manufacturer’sCertification
form.
5.If a TEC 6 DESFLURANE vaporizer is mounted to the machine, a label stating: "WARNING THE
ADMINISTRATION OF DESFLURANE ANESTHESIA MAY REQUIRE FRESH GAS CONCENTRATIONS
HIGHER THAN COMMONLY USED WITH OTHER VOLATILE ANESTHETIC AGENTS. O2 FRESH GAS
CONCENTRATION OF LESS THAN 21% MAY BE OBTAINED WITH HIGH VAPORIZER SETTINGS.
CONTINUOUS MONITORING OF THE O2 CONCENTRATION IN THE BREATHING SYSTEM IS THEREFORE
REQUIRED. NORTH AMERICAN DRÄGER RECOMMENDS THE CONTINUOUSMONITORING OF THE
CONCENTRATION OFANESTHETIC VAPORSIN THEBREATHING SYSTEMDURING THE
ADMINISTRATION OF INHALATION ANESTHESIA."(part # 4112737-001) Shall be attached to the vapor box
immediately above the TEC 6 vaporizer.
6.Check all vapor 19 and 19.1 vaporizers for correct labeling. All vaporizers must have a label stating "THE
CONCENTRATION OUTPUT OF THIS VAPORIZER SHALL BE VERIFIED AFTER IT HAS BEEN ATTACHED
TO THE ANESTHESIA MACHINE" (part # S010015). This label shall be attached to the rear of the vaporizer
directly below the mount.
7.All Key Index Safety Systems vaporizers, (K.I.S.S.) must have a label stating "CAUTION:
BEEN COMPLETED, REINSERT PLUG INTO UPPER FILLER PORT AND TIGHTEN LOCKING SCREW" (part
# 4112520-001). This label shall be attached to the vaporizer directly above the keyed filler. Place a check mark at
"K.I.S.S. LABEL" on the PM Certification form.
8.If machine is equipped with a HALOTHANE Dräger Vapor 19 or 19.1 vaporizer, determine if vaporizer must be
inspected for soil condition one. Check the serial number plate located on the rear of the vaporizer for a plus (+)
preceding the serial number. A HALOTHANE vaporizer serial number not preceded with a (+) must be tested for
soil in accordance with SP00073. If vaporizer does not need to be inspected, indicate so with a plus (+) next to the
"Vapor Inspection (H)" line on the Vigilance Audit form. If vaporizer is soil condition 0, indicate so with "SOIL 0"
written next to the "Vapor Inspection (H)" line on the Vigilance Audit form. If vaporizer is soil one, indicate so with
"SOIL 1" written next to the "Vapor Inspection (H)" line on the Vigilance Audit form. Place a "CAUTION DO NOT
USE" label (part # 4114327) on the vaporizer, and issue a departmental alert. The TSR shall also seek permission
from the equipment operator to remove the failed vaporizer from the machine and apply a replacement vaporizer
or an adapter block onto the mount. All "SOIL 1" vaporizers must be removed from service for machine to receive
certification.
9.Perform the vapor concentration test on all Dräger vapor vaporizers in accordance with SP00073 at a six month
maximum interval. Perform the vaporizer concentration test on all Desflurane vaporizers in accordance with
SP00091 for fixed mount vaporizers and SP00189 for user removable D-tec vaporizers at a six month maximum
interval.For every vaporizertested,fill out a"VAPORVAPORIZER CALIBRATION CHECK"label(part # S010016).
Information on this label shall include your signature, type of agent, date tested, test results @ 1%, 2.5%, 4% for
H, E, I, or S vaporizers, or @ 4%, 10%, 12%, 16% for Desflurane vaporizers, and a PASS or FAIL indication. This
label shall be attached to the upper right side of the vaporizer. If vaporizer fails the concentration verification,
internal leak, or exclusion system tests, check "NO" in the "RECOMMENDED FOR USE" section on the PM
Certification form.
AFTER FILLING HAS
6-1JRev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
PM Certification Procedure for Narkomed 2B Anesthesia System
Place a "CAUTION DO NOT USE" label (part # 4114327) on the vaporizer, and issue a departmental
alert. The TSR shall also seek permission from the equipment operator to remove the failed vaporizer
from the machine and install a replacement vaporizer or an adapter block onto the mount. All
nonfunctional Dräger vapor vaporizers must be removed from service for the machine to receive
certification.
10.Proceed with PM Certification procedure. If any tests fail refer to the "Failure Codes" listing in DMI
Recommendations Guidelines Index (P/N S010250) to determine correct certification level starting point.
Failure codes shall be documented on the "RECOMMENDATIONS / GENERAL COMMENTS" section
of the PM Certification form and on the Executive Summary. If a test fails that has not been identified
by the "Failure Codes" list, consult with DMI Inc. to assess the proper certification level.
11.Based on the "EQUIPMENT CONDITION" inspect the machine for any "DMI RECOMMENDATIONS"
that would apply. Use the Narkomed 2B section of the "RECOMMENDATION GUIDELINES INDEX"
(P/N S010250). Note all applicable DMI recommendations on the Executive Summary. NOTE: If using
a carbon form, indicate the Equipment Condition number and to see reverse side under the
"RECOMMENDATIONS / GENERAL COMMENTS" section of the form.
12.Determine the correct certification level of the machine based on the combined lowest common
denominator of "Equipment Conditions" and "Failure Codes". If the machine is at least conditionally
certified fill out the "PM CERTIFICATION" label. Check the box(s) on the validation label where
appropriate. Write the month and year, (three months from date of PM Certification) next to "NEXT
VISIT DUE:" If certification level is "D", machine shall not receive a "PM CERTIFICATION" label. Any
machine not receiving a PM Certification label shall receive a "WARNING NOT CERTIFIED" label. This
label shall be placed at a prominent location on the left side of the machine after all other previous PMS
and Vigilance Audit Validation labels have been removed.
13.In the "CERTIFICATION LEVEL" section of the PM Certification form, record the last visit certification
level, the current certification level and the next visit due month and year, (three months from date of
PM Certification) in the spaces provided.
14.If applicable, remove the previous PM CERTIFICATION VALIDATION label and attach the new label
(P/N S010006 w/phone #, or P/N S010007 w/o phone #) in a prominent location on the rear of the
anesthesia machine.
15.Check the appropriate boxes on the "PM CERTIFICATION NOTICE" label, (part # S010011). If the
machine is not certified, the last box of this notice label shall be marked. Attach this notice to the flow
shield of the anesthesia machine.
16.Have the customer sign each PM Certification form or the Executive Summary, and review the
equipment conditions and the recommendations with the customer.
17.Return top copy to Draeger Medical, Inc. Service Department, keep middle copy for service organization
records, give bottom copy to customer.
Rev. Y6-1K
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.1ELECTRICAL SAFETY
( ) 6.1.1Ground Continuity
6.1.1.1Turn the System Power switch to STANDBY.
NOTE:Do not plug the safety analyzer power cord into a line isolation monitor,
as inaccurate readings may occur.
Plug the unit into the safety analyzer, and plug the power cord of the
safety analyzer into an AC receptacle.
NOTE:The BIOTECH 501 PRO will automatically test the source outlet for open
ground ( or ground resistance of 31 Ω or higher), reverse polarity, open
neutral and open line. (The latter two conditions will prevent the analyzer
from powering up.)
6.1.1.2Set the safety analyzer function switch to the GROUND WIRE
RESISTANCE position. Attach the test lead to the SINGLE LEAD
connector of the analyzer. Connect the other end of the red test lead to
the ground socket of the front panel outlet on the safety analyzer. Verify
a displayed resistance of 0.000 Ω or, if necessary, press the CALIBRATE
key on the front panel of the analyzer to zero the device.
6.1.1.3Set the safety analyzerGROUND switch to NORMAL. Setthe POLARITY
switch to OFF.
6.1.1.4The safety analyzer shall indicate 0.1 Ω or less with its test lead applied
to the following points:
( )6.1.2Circuit Isolation
6.1.2.1Disconnect the Spiromed sensor cord from the interface panel.
6.1.2.2With a multimeter set to its highest resistance range, check for continuity
between the power supply chassis and the circuit common at the
Spiromed sensor interface connector keyway pin (at the 12 o’clock
position). There shall be no continuity between these points.
--Cylinder yoke
--Power supply assembly chassis
--Each convenience outlet ground socket
--Each Auxiliary Outlet Strip ground socket
6-2
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( )6.1.3Convenience Outlet and Auxiliary Outlet Strip
NOTE:This test will check the convenience outlets and the auxiliary strip outlets
for fault conditions such as open ground (>31 Ω), reverse polarity, open
line and open neutral. This is done each time the BIOTECH 501 PRO is
powered up and allowed to cycle through its self test.
6.1.3.1Shut off and unplug the safety analyzer. Remove the anesthesia machine
plug from the analyzer, and plug it into the same outlet that was being
used by the analyzer.
6.1.3.2Plug the safety analyzer into the first convenience outlet to be tested, and
turn the analyzer power switch ON. Allow the analyzer to cycle through
its Auto Test sequence. If no wiring fault is indicated, shut off the
analyzer and move its plug to the next convenience outlet. Test this outlet
in the same manner, and continue until all convenience outlets and
auxiliary strip outlets are tested.
6.1.4Chassis Leakage Current
6.1.4.1Turn the anesthesia machine System Power switch to ON and set the
safety analyzer to the CHASSIS LEAKAGE CURRENT position.
6.1.4.2Attach the safety analyzer test lead to a cylinder yoke.
( )6.1.4.3Record the total leakage current with the Polarity and Ground switches
Verify that the leakage current is 75* microamps or less in each of the
switch positions (110 microamps or less for the 220/240 volt power supply
option).
*100 microamps if external monitors are plugged into convenience
outlets.
Rev. Y
6-3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
( ) 6.2SELF-DIAGNOSTICS
6.2.1Connect the pipeline supply or open the cylinders.
6.2.2Turn the System Power switch to ON.
6.2.3Verify that the following is displayed on the alarm CRT:
VIDEO TESTPASSNARKOMED 2B
FIRMWARE TESTPASSVERSION x.xx DIAGNOSTICS
MEMORY TESTPASSCOPYRIGHT, NAD INC. 1987-94
TIMERS TESTPASS
ANALOG TESTPASS
AUDIO TEST - PRIMARYPASS
- BACKUPPASS
SERIAL I/O TESTPASS
CLOCK TESTPASS
BACKUP MEMORY TESTPASS
AC POWER TESTPASS
RESERVE POWER TESTPASS
FUNCTIONAL
( )6.2.4Record the machine software version on the header of the checklist form.
6-4
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( ) 6.3BATTERY CIRCUIT
6.3.1Is "ON" LED lighted? __ (Y)
6.3.2With the System Power switch ON, unplug the AC power cord.
6.3.3Isyellow "AC PWR FAIL" LED lighted as long as the power cord is unplugged?
__ (Y)
6.3.4Does the "AC PWR FAIL" message appear in the Advisory display? __ (Y)
6.3.5Press and hold the "BATTERY TEST" button.
6.3.6Is green Battery Test LED lighted as long as "BATTERY TEST" button is
depressed? __ (Y)
6.3.7Release the "BATTERY TEST" button.
6.3.8Restore AC power to the machine.
6.3.9Does the "AC PWR FAIL" message disappear and the LED extinguish?
( ) 6.4CONFIGURATION
6.4.1Press the CONFIG key on the monitor controls keypad.
6.4.2The alarm CRT should display the Configuration Screen.
6.4.3Verify the correct Time and Date.
6.4.4Adjust the Volume to the highest number.
Rev. Y
6-5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.5SERVICE DATA
6.5.1Access the Configure screen, then press the following keys on the monitor
controlskeypanel: 21%keyandAPNEAALARMDISABLEkey
(simultaneously), and then the CONFIG key (while still holding the previous
two keys). The alarm CRT should then display the Service Menu Screen.
6.5.2Press the DIAGNOSTICS key. The alarm CRT should display the Diagnostics
Menu Screen.
( )6.5.5Record any Error Log codes.
( )6.5.3Record the Last Service Date.
( )6.5.4Press the RESET DATE key.
6.5.6Key Panel
6.5.6.1Press the KEY TEST key.
6.5.6.2The alarm CRT should display the outline of all keys on the display
panel.
6.5.6.3Press each key on the display panel, one at a time.
( )6.5.6.4Aseach key on the display panel is pressed, do all the corresponding keys
on the alarm CRT illuminate? ___(Y)
NOTE:The TREND key should be pressed last, because it also exits the Key
Panel Test Screen.
6-6
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.6CALIBRATIONS
6.6.1Press the CALIBRATION key on the Service Menu Screen to bring up the
Calibration Menu Screen.
6.6.2Remove the oxygen sensor from the valve dome adapter, and remove the
oxygen sensor capsule from the oxygen sensor housing.
( )6.6.3When the "OFFSET O2CELL A" and "OFFSET O2CELL B" readings
(displayed in left column) are as close as possible to each other with a
difference not greater than 8, press the O
MED key and verify that the new
2
offset values are stored.
NOTE:The higher the offset, the higher the calculated oxygen concentration at
higher oxygen concentrations.
6.6.4Put the oxygen sensor capsule into the oxygen sensor housing.
6.6.5Disconnect the Baromed breathing pressure sensor line from the absorber.
6.6.6This step intentionally left blank.
6.6.7Connect a test fixture and digital pressure manometer to the breathing
pressure sensor line.
6.6.8Pressurize the circuit to 60 cm H
6.6.9The"PRESSURE SPAN (60 cm H
O.
2
O)" reading should display between 465 and
2
519.
( )6.6.10Press the BAROMED key to store the reading. The following message should
be displayed: "PRESSURE SPAN READING STORED".
6.6.11Releasethe pressure, disconnect the manometer andtest fixture, andreconnect
the breathing pressure sensor line to the absorber.
6.6.12Press the EXIT key to exit the Calibration Screen.
6.6.13Press the EXIT key to exit the Service Menu Screen.
Rev. Y
6-7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
( ) 6.7ASCAVENGER, A/C - if applicable
6.7A.1Scavenger, A/C Cleaning
6.7A.1.1Remove all scavenger hoses one at a time, and drain all accumulated
moisture. Inspect all scavenger hoses for deterioration and replace any
worn hoses.
6.7A.1.2Remove the safety relief valve housing by unscrewing it in a counterclockwise direction.
6.7A.1.3Inspect the rubber O-ring and replace if worn.
6.7A.1.4Remove the safety relief valve from its housing by twisting it out in a
counter-clockwise direction. The tips of needle-nose pliers can be used to
turn the valve. Be careful not to damage the valve disk.
6.7A.1.5Remove any accumulated lint or dust from the valve with a soft brush.
The valve may be further cleaned with a low flow of clean air or oxygen.
The scavenger body can be cleaned with a moist cloth.
6.7A.1.6Reinstall the valve into the housing, making sure that it is threaded all
the way into the housing and that the plastic washer is properly seated
on its upper surface.
6.7A.1.7Make sure that the interior of the valve body is completely dry. Reinstall
the valve housing onto the scavenger body, making sure that the O-ring
is properly seated.
6.7A.2Scavenger, A/C Positive Pressure Relief
6.7A.2.1Connect a 19 mm scavenger hose between the bottom of the absorber pole
and the right-hand port on the scavenger. Connect a short 19 mm
scavenger hose between the APL valve and the port on the rear of the
absorber pole. Connect a 19 mm scavenger hose between the ventilator
relief valve and the left-hand port on the scavenger.
6.7A.2.2Connect a short 22 mm breathing hose from the inspiratory valve to the
expiratory valve on the absorber.
6.7A.2.3Set the Man/Auto valve to the AUTO position.
6-8
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.7A.2.4If the absorber system or ventilator bellows are equipped with a PEEP
valve, turn the PEEP valve control knob fully counter-clockwise.
6.7A.2.5Set the oxygen flow to 10 l/min. and occlude the 19 mm scavenger
terminal labeled EXHAUST.
6.7A.2.6After the ventilator bellows inflates, the flow of oxygen will exit the
system through the positive pressure safety relief valve. At this point, the
absorber system breathing pressure gauge shall indicate a pressure of
10.0 cm H
O or less.
2
( ) 6.7B SCAVENGER, OPEN RESERVOIR - if applicable
6.7B.1Scavenger, Open Reservoir Cleaning
6.7B.1.1Remove all scavenger hoses one at a time and drain all accumulated
moisture. Inspect all scavenger hoses for deterioration and replace any
worn hoses.
6.7B.1.2Disconnect the hospital vacuum source from the scavenger.
6.7B.1.3If applicable, remove the scavenger flow control needle valve assembly.
Inspect the needle valve and seat for lint or dust accumulation. Clean
with compressed air if necessary, and reinstall the needle valve assembly.
6.7B.1.4Remove the scavenger mounting screws.
6.7B.1.5Remove the two screws securing the access panel at the bottom of the
scavenger canister.
6.7B.1.6Remove and inspect the silencer; replace if needed.
6.7B.1.7Remove the reservoir canister from the scavenger body by unscrewing the
four socket head cap screws located at the top of the canister.
6.7B.1.8Remove the flowmeter from its housing by turning it counter-clockwise.
Inspect the tube and clean with compressed air if needed.
6.7B.1.9Reassemble the scavenger assembly, and reactivate the vacuum source.
Rev. Y
6-9
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.7B.2Scavenger, Open Reservoir Negative Pressure Relief
6.7B.2.1Connect a 19 mm scavenger hose between the bottom of the absorber pole
and the right-hand port on the scavenger. Connect a 19 mm scavenger
hose between the APL valve and the rear port on the absorber pole. The
left-hand scavenger port may be capped for this test, or may be connected
to the ventilator relief valve. Connect a DISS vacuum hose to the
threaded terminal on the left side of the scavenger. Alternatively, an
adapter can be used to attach a wall suction hose to the hose barb fitting
on the adapter.
6.7B.2.2Connect a short 22 mm breathing hose from the inspiratory valve to the
expiratory valve on the absorber. Set the Man/Auto selector valve to the
BAG position. Turn the APL valve control knob fully counter-clockwise.
6.7B.2.3Verify that the suction waste gas disposal system is active.
6.7B.2.4Adjust the scavenger needle valve until the flowmeter indicates between
the white lines. Close all flow control valves on the anesthesia machine.
Occlude the absorber breathing bag terminal.
6.7B.2.5Install a scavenger adapter with a hose barb (P/N 4108114) between the
19 mm hose terminal of the scavenger, and the scavenger hose. Connect
a calibrated test pressure monitor to the hose barb on the adapter and
observe the pressure reading on the test gauge. The gauge shall indicate
a pressure of 0 to -0.5 cm H
O.
2
6.7B.3Scavenger, Open Reservoir Positive Pressure Relief
6.7B.3.1Turn the scavenger needle valve fully clockwise (closed).
6.7B.3.2Open the oxygen flow control valve on the anesthesia machine to a flow
of 10 l/min. and occlude the absorber breathing bag terminal.
6.7B.3.3The flow of oxygen shall now exit the system through the relief ports
around the top of the canister. The test pressure gauge shall indicate a
pressure less than 1.0 cm H
O.
2
6.7B.3.4After the test, adjust the scavenger needle valve for a flowmeter
indication halfway between the two white lines.
6-10
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( ) 6.7CSCAVENGER INTERFACE - if applicable
6.7C.1Scavenger Inspection
6.7C.1.1Check all scavenger hoses and reservoir bag for deterioration. Replace all
worn components.
6.7C.2Primary Negative Relief Valve Cleaning:
6.7C.2.1Remove the plastic valve cover on the front surface of the scavenger body
by turning it in a counter-clockwise direction.
6.7C.2.2Remove the valve and washer from the scavenger body by turning it
counter-clockwise. A needle-nose pliers may be used to turn the valve, but
use care not to damage the valve’s fragile disk.
6.7C.2.3Brush any accumulated lint or dust off the valve with a soft brush. The
valve may be further cleaned with a low flow of clean air or oxygen.
6.7C.2.4Reinstall the plastic washer and valve into the scavenger body. Replace
the valve cover.
6.7C.3Secondary Negative Relief Valve Cleaning:
6.7C.3.1Unscrew the chrome plated valve housing on the left side of the scavenger
body by turning its fitting counter-clockwise with a wrench.
6.7C.3.2Unscrew the valve from the housing by turning it in a counter-clockwise
direction.
6.7C.3.3Brush any accumulated lint or dust off the valve with a soft brush. The
valve may be further cleaned with a low flow of clean air or oxygen.
6.7C.3.4Reinstall the valve in the housing, and then reinstall the housing into the
scavenger body. Note: This valve does not require washers or O-rings.
6.7C.4Positive Relief Valve Cleaning:
6.7C.4.1Remove the valve housing on the right side of the scavenger body by
turning it counter-clockwise.
6.7C.4.2Inspect the rubber O-ring and replace if worn.
Rev. Y
6-11
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.7C.4.3Remove the valve from the housing by turning it counter-clockwise. A
needle-nose pliers may be used to turn the valve, but use care not to
damage the valve’s fragile disk.
6.7C.4.4Brush any accumulated lint or dust off the valve with a soft brush. The
valve may be further cleaned with a low flow of clean air or oxygen.
6.7C.4.5Reinstall the valve and plastic washer into the housing.
6.7C.4.6Reinstallthe valve housing onto the scavenger body, making sure that the
O-ring is properly seated.
6.7C.5Negative Pressure:
6.7C.5.1Connect a 22mm breathing hose between the absorber’s inspiratory and
expiratory valves. Set the Man/Auto valve to the BAG position. Turn the
APL valve fully counter-clockwise. Occlude the bag mount connector.
6.7C.5.2Verify that the suction waste gas disposal system is active.
6.7C.5.3Close all flow control valves on the machine. Adjust the scavenger needle
valve to allow typical suction through the scavenger.
6.7C.5.4Install a scavenger adapter (P/N 4108114) with a hose barb between the
19mm hose terminal of the scavenger and the scavenger hose. Connect a
test monitor to the hose barb adapter and observe the pressure reading
on the test gauge. The gauge shall indicate a pressure of <-1.0 cmH
6.7C.6Positive Pressure:
6.7C.6.1Close the scavenger needle valve by turning it fully clockwise.
6.7C.6.2Push the O
Flush button to inflate the scavenger reservoir bag. Open the
2
oxygen flow control valve to 10 l/min.
6.7C.6.3Observe the pressure reading on the test gauge. The gauge shall indicate
a pressure of <10.0 cmH
O.
2
6.7C.6.4Remove the test equipment. Readjust the scavenger needle valve to allow
typical suction through the scavenger.
O.
2
6-12
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.7DSUCTION REGULATOR
6.7D.1Verify that the suction bottle is attached to the suction regulator.
6.7D.2Verify that vacuum is attached to the ¾ in. DISS vacuum connection.
6.7D.3Set the vacuum on/off valve to the OFF (vertical) position.
6.7D.4Connect a digital pressure meter to the collecting inlet stem of the suction
bottle.
6.7D.5Set the digital pressure meter to the mmHg scale.
6.7D.6Turn the vacuum control knob fully counter-clockwise.
( )6.7D.7What is the vacuum indicated on the digital pressure meter? (0)
6.7D.8Turn the vacuum control knob fully clockwise and verify that the vacuum
control knob stops.
6.7D.9Set the vacuum on/off valve to the ON position.
6.7D.10Set the regulator to indicate 250 mmHg.
( )6.7D.11What is the vacuum indicated on the digital pressure meter? (200-300 mmHg)
6.7D.12Return all controls to their original positions.
Rev. Y
6-13
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.8BREATHING SYSTEM
( )6.8.1Absorber System Inspection
6.8.1.1Remove the inspiratory and the expiratory valve domes.
6.8.1.2Is there a broken or bent pin on the valve assembly?
Inspiratory ___ (N)Expiratory ___ (N)
6.8.1.3Is there a broken pin on the valve domes?
Inspiratory ___ (N)Expiratory ___ (N)
6.8.1.4Is the valve disc in good condition?
Inspiratory ___ (Y)Expiratory ___ (Y)
6.8.1.5Are the valve dome washers in good condition? ___ (Y)
6.8.1.6Reinstall the inspiratory and expiratory valve domes.
6.8.1.7Remove the inspiratory and expiratory valve assemblies.
6.8.1.8Remove the Spiromed sensor.
6.8.1.9Are all three washers in good condition? ___ (Y)
6.8.1.10Reinstall the inspiratory valve.
6.8.1.11Are the two (2) spring clips on the absorber rods?
___ (Y)
6.8.1.12Inspect the following: canisters and gaskets, dust cup and O-ring,
condition of soda lime.
6.8.1.13Are the absorber canisters and dust cup in good condition? ___ (Y)
6.8.1.14Is the cm H
6.8.1.15Remove the O
O gauge at zero (0)? ___ (Y)
2
Med sensor plug from the inspiratory valve dome adapter
2
and examine the two O-rings at the bottom of the plug.
6-14
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.8.1.16Reinstall the O
Med sensor plug into the inspiratory valve dome adapter.
2
6.8.1.17Examine the two O-rings at the bottom of the sensor.
( )6.8.2Repack MAN/AUTO Selector Valve
[Does not apply to machines with later design selector valve.]
6.8.2.1Remove the four screws securing the stick shift block to the selector valve
body and remove the block.
6.8.2.2Remove the spring and valve channel from the valve body.
6.8.2.3Remove all residual lubricant from the valve channel.
6.8.2.4Remove all residual lubricant from the valve body.
6.8.2.5Apply a minimal amount of "stop cock" lubricant (Dow Corning High
Vacuum Grease, P/N S4105908) to the tapered surface of the valve
channel, and ensure complete coverage of lubricant.
6.8.2.6Insert the valve channel into the valve body.
6.8.2.7Insert the spring into the stick shift block.
6.8.2.8Align the index pins on the stick shift block to the holes in the valve
channel.
6.8.2.9Secure the stick shift block to the selector valve body with the four screws
that were previously removed.
6.8.2.10Operate the selector valve handle and verify smooth movement.
Rev. Y
6-15
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
( )6.8.3Lubrication, Spiromed Sensor
6.8.3.1Locate the four lateral holes at the sides of the Spiromed sensor marked
by printed arrows.
CAUTION:Use only Sensor Lubrication Kit P/N 2218180 for the following
procedure.
6.8.3.2Remove the protective cover from the pipette and open the lubricant
bottle.
6.8.3.3Dip the tip of the pipette into the lubricant and draw one drop of
lubricant into the pipette by pulling the pin upwards.
6.8.3.4Insert the pipette into one of the four holes as far as it will go. Push the
pin forward to its stop to inject lubricant into the hole.
6.8.3.5Repeat the procedure on the three remaining holes.
6.8.3.6Wipe any lubricant residue from the exterior of the sensor.
6.8.3.7Reattach the sensor to the absorber top dome.
6.8.3.8Reattach the expiratory valve to the sensor.
6-16
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.8.4Freshgas Leak
6.8.4.1Turn the System Power switch to STANDBY.
6.8.4.2Remove the 15 mm connector from the FRESHGAS OUTLET.
6.8.4.3Is the common gas outlet assembly in good condition? ___ (Y)
6.8.4.4Connect a digital pressure manometer and Fresh Gas Leak Test Device
to the freshgas outlet.
6.8.4.5Apply 50 cm H
O of pressure to the system.
2
( )6.8.4.6After thirty (30) seconds, what is the pressure on the manometer? ___
(>40 cm H
2
O)
6.8.4.7Turn on the left mounted vaporizer to the first graduated marking.
6.8.4.8Apply 50 cm H
O of pressure to the system.
2
( )6.8.4.9After thirty (30) seconds, what is the pressure on the manometer? ___
(>40 cm H
2
O)
6.8.4.10Turn off the vaporizer.
6.8.4.11Remove the test equipment from the Fresh Gas Outlet.
6.8.4.12Turn the System Power switch to ON.
6.8.4.13Open the O
then close the O
flow control valve to 5 l/min., purge the system for 5 seconds,
2
flow control valve.
2
6.8.4.14Turn the System Power switch to STANDBY.
Rev. Y
( )6.8.4.15Turn on the center mounted vaporizer to the first graduated marking,
repeat Steps 6.8.4.4 and 6.8.4.8 thru 6.8.4.14. ___ (>40 cm H
O)
2
( )6.8.4.16Turn on the right mounted vaporizer to the first graduated marking,
repeat Steps 6.8.4.4 and 6.8.4.8 thru 6.8.4.14. ___ (>40 cm H
O)
2
6.8.4.17Reconnect the 15 mm connector from the absorber system to the
FRESHGAS OUTLET.
6.8.4.18Is the FRESHGAS OUTLET label on the freshgas outlet? ___ (Y)
6-17
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.8.5Breathing System Leak
6.8.5.1Verify the System Power switch is at STANDBY.
6.8.5.2Close all flow control valves.
6.8.5.3Interconnect the inspiratory and expiratory valves with a 12-inch hose.
6.8.5.4Attach a test terminal with a cuff inflation bulb (P/N S010159) to the bag
mount.
6.8.5.5Set the Man/Auto selector valve to BAG.
6.8.5.6Close the APL valve.
6.8.5.7Apply 50 cm H
O pressure to the absorber system.
2
( )6.8.5.8After 30 seconds, what is the pressure in the absorber system? ___ cm
H
O(>30)
2
6.8.6APL Valve
6.8.6.1Open the APL valve to its stop.
6.8.6.2Turn the SYSTEM POWER switch to ON.
6.8.6.3Set the oxygen flow to 8 l/min.
( )6.8.6.4What is the pressure on the absorber pressure gauge? ___ cm H2O(<3)
6.8.6.5Close the oxygen flow control valve, turn the System Power switch to
STANDBY, and remove the test terminal from the bag mount.
6.8.7Absorber Flow Direction and Leak
6.8.7.1Expiration Valve Leak
6.8.7.1.1 Close the APL valve.
6.8.7.1.2 Connect a 22mm hose between the inspiration valve and the bag
mount.
6.8.7.1.3 Connect a test terminal to the expiration valve.
6.8.7.1.4 Connect a Capnomed flowmeter to the test terminal.
6-18
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.8.7.1.5 Turn the System Power switch to ON, turn up the oxygen flow until
the system pressurizes to 30 cmH
O.
2
( )6.8.7.1.6 What is the value indicated on the flowmeter? ___ (≤60ml/min).
6.8.7.1.7 Remove all test equipment, and turn the System Power switch to
STANDBY.
6.8.7.2Inspiration Valve Leak
6.8.7.2.1 Connect a test terminal to the inspiratory valve.
6.8.7.2.2 Connect a tee fitting (P/N 4109292) and calibrated pressure meter
to the test terminal.
6.8.7.2.3 Connect a pressure bulb to the open port of the tee adapter.
6.8.7.2.4 Connect another test terminal to the bag connector.
6.8.7.2.5 Connect a Capnomed flowmeter to the test terminal on the bag
mount.
6.8.7.2.6 Pressurize the system to 30 cmH
O.
2
( )6.8.7.2.7 What is the value indicated on the flow meter? ___ (≤60 ml/min).
6.8.7.2.8 Remove all test equipment.
6.8.7.3Flow Direction
6.8.7.3.1 Attach a breathing circuit with a 3-liter bag at the Y-piece to the
inspiration and expiration valves.
6.8.7.3.2 Attach a 3-liter bag to the swivel bag mount.
6.8.7.3.3 Turn the System Power switch to ON.
6.8.7.3.4 Set the O
6.8.7.3.5 Inflate the simulated lung by briefly using the O
flow to 4 l/min.
2
Flush.
2
Rev. Y
6.8.7.3.6 Partially open the APL valve.
6-19
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.8.7.3.7 Squeeze the breathing bag attached to the bag mount at a rate of
approximately 10 BPM. Readjust the APL valve if required to
properly ventilate the simulated lung.
( )6.8.7.3.8 Observe the operation of each unidirectional valve disc at eye level
and make sure the inspiratory valve disc raises only during the
inspiration phase, and the expiratory valve raises only during the
exhalation phase. Watch the valves until satisfied that both valves
operate correctly, and move freely without sticking.
6.8.7.3.9 Remove the breathing circuit.
6.8.7.3.10Open the APL valve.
( )6.8.8Absorber PEEP Valve w/By-Pass - if applicable
6.8.8.1Connect a 12 in. hose between the inspiratory and expiratory valves, and
set the O
flow to 5 l/min.
2
*6.8.8.2Place the PEEP bypass in the ON position.
6.8.8.3Adjust the absorber PEEP valve clockwise to the maximum position.
6.8.8.4Does the PEEP valve adjust smoothly? ___ (Y)
6.8.8.5What is the maximum PEEP? ___ cm H
O (15-22)
2
*6.8.8.6Place the PEEP bypass in the OFF position.
6.8.8.7Does the PEEP return to ≤3cmH
O? ___(Y)
2
6.8.8.8Adjust the absorber PEEP valve counterclockwise to its minimum
position.
6.8.8.9Close the O
flow control valve.
2
* These items apply only to machines with a PEEP by-pass.
6-20
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.8.9Bain Circuit Adapter - if applicable
6.8.9.1Close the APL valve by turning the knob fully clockwise.
6.8.9.2Insert the O
sensor plug into the O2sensor inlet on the Bain Circuit.
2
6.8.9.3Attach a test terminal with a cuff inflation bulb (P/N S010159) to the
Breathing Bag port on the Bain Circuit.
6.8.9.4Attach a cmH
O digital pressure meter to the female quick connect
2
connection on the Bain Circuit.
6.8.9.5Occlude the expiration port on the Bain Circuit.
6.8.9.6Apply 50cmH
O to the Bain Circuit via test terminal and inflation bulb.
2
( )6.8.9.7After 30 seconds, what is the pressure on the cmH2O digital pressure
meter? (45 to 50 cmH
6.8.9.8Verify that the pressure indicated on the cmH
2
O)
O gauge is within 3 cmH2O
2
of the digital pressure meter reading.
6.8.9.9Open the APL valve by turning the knob fully counter-clockwise.
6.8.9.10Connect a test hose from the fresh gas outlet to the Expiration port of the
Bain Circuit.
6.8.9.11Set the O
flow to 10 L/min.
2
( )6.8.9.12What is the pressure on the cmH2O digital pressure meter? (≤ 3 cmH2O)
6.8.9.13Verify that the pressure indicated on the cmH
O gauge is within 3 cmH2O
2
of the digital pressure meter reading.
6.8.9.14Remove the test terminal and inflation bulb from the Breathing Bag port.
6.8.9.15Return all controls to their original positions.
Rev. Y
6-21
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
( ) 6.9MANUAL SPHYGMOMANOMETER - if applicable
6.9.1Insert the male Luer fitting of the Sphygmomanometer squeeze bulb-hose
assembly into the female Luer fitting labeled BP BULB on the front of the
machine.
6.9.2Wrap the blood pressure cuff around an "E" cylinder.
6.9.3Hand-pump the squeeze bulb until pressure of 200 mm Hg is indicated on the
Sphygmomanometer gauge on the machine. Pinch the hose adjacent to the
Luer fitting to assure that the hose-bulb is not the source of any leak.
6.9.4After thirty (30) seconds, what is the pressure on the Sphygmomanometer
gauge? ___ mm Hg (190-200)
6.9.5Bleed the pressure.
6.9.6The Sphygmomanometer should indicate within the band.
6.9.7Attach a test gauge in series with the sphygmomanometer gauge being tested.
6.9.8Hand-pump the squeeze bulb until the machine gauge indicates 100 mm Hg.
6.9.9What does the test gauge indicate? ___ mm Hg (90-110).
6.9.10Hand-pump the squeeze bulb until the machine gauge indicates 200 mm Hg.
6.9.11What does the test gauge indicate? ___ mm Hg (180-220).
6.9.12Hand-pump the squeeze bulb until the machine gauge indicates 300 mm Hg.
6.9.13What does the test gauge indicate? ___ mm Hg (210-390).
6.9.14Remove the test gauge.
6.9.15Remove the blood pressure cuff from the "E" cylinder.
6-22
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( ) 6.10 VAPOR EXCLUSION SYSTEM
6.10.1Set all vapors to zero (0).
6.10.2Adjust the handwheel on the left vapor (viewed from the front of the machine)
to any concentration above zero (0).
6.10.3Is it possible to adjust the center vapor? ___ (N)
6.10.4Is it possible to adjust the right (viewed from the front of the machine) vapor?
___(N)
6.10.5Adjust the handwheel on the left vapor to zero (0).
6.10.6Adjust the handwheel on the center vapor to any concentration above zero 0).
6.10.7Is it possible to adjust the left vapor? ___ (N)
6.10.8Is it possible to adjust the right vapor? ___ (N).
6.10.9Adjust the handwheel on the center vapor to zero (0).
6.10.10Adjust the handwheel on the right vapor to any concentration above zero (0).
6.10.11Is it possible to adjust the left vapor? ___ (N).
6.10.12Is it possible to adjust the center vapor? ___ (N).
6.10.13Adjust the handwheel on the right vapor to zero (0).
Rev. Y
6-23
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.11 YOKES & GAUGES
( )6.11.1Yokes & Check Valves
6.11.1.1Turn the System Power switch to STANDBY.
6.11.1.2Disconnect all pipeline hoses and close all cylinder valves.
6.11.1.3Remove cylinder or yoke plug from each yoke assembly.
6.11.1.4Do all the yoke handles adjust smoothly? ___ (Y)
6.11.1.5Are the two (2) yoke pins installed securely in each yoke? ___ (Y)
6.11.1.6Is there only one (1) cylinder washer on each yoke assembly? ___ (Y)
6.11.1.7Is there a yoke plug attached to each yoke assembly? ___ (Y)
6.11.1.8Is the proper gas I.D. label affixed to each yoke assembly? ___ (Y)
6.11.1.9Attach a cylinder to each yoke assembly, open the cylinder valve, let the
pressure stabilize, close the cylinder valve, and remove the cylinder from
the yoke assembly.
6.11.1.10 Does the yoke check valve assembly prevent the escape of excessive
pressure? ___ (Y)
6.11.1.11 Attach the cylinders to the yokes.
6-24
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( )6.11.2Cylinder Gauges
6.11.2.1Are the pressure gauges correct for the gases indicated by the flowmeters?
___ (Y)
6.11.2.2Is the gauge closest to the table top for cylinder supply pressure? ___ (Y)
6.11.2.3Bleed all pressure from the cylinder circuit.
6.11.2.4Are the cylinder gauges at zero (0) PSI? ___ (Y)
6.11.2.5Open the cylinder valves.
6.11.2.6Do the cylinder pressure gauges respond properly? ___ (Y)
( )6.11.3Pipeline Gauges
6.11.3.1Are the gauges right below the flowmeters for pipeline supply pressure?
___ (Y)
6.11.3.2Are the pipeline pressure gauges at zero (0) PSI? ___ (Y)
6.11.3.3Connect the pipeline supply.
6.11.3.4Do the pipeline pressure gauges respond properly? ___ (Y)
6.11.3.5Are the correct gas identification labels affixed at each of the pipeline
inlets? ___ (Y)
6.11.3.6Does the back panel identify each of the pipeline inlets properly? ___ (Y)
6.11.3.7Disconnect the pipeline hoses.
Rev. Y
6-25
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.12 HIGH PRESSURE REGULATOR
Minimum cylinder pressures for this test are: N2O & CO2: 600 psi
O2, Air, He, He/O2, N2: 1000 psi
6.12.1Verify the System Power switch is ON.
6.12.2Verify all cylinder valves are closed except the O
valve.
2
6.12.3Set the oxygen flow to 4 liters per min.
6.12.4Open the other gas flow control valves to drain pressure from the system.
6.12.5Close the O
cylinder valve, and press the O2Flush valve to drain oxygen
2
pressure from the system.
6.12.6Turn the System Power switch to STANDBY.
6.12.7Remove the table top from the machine and remove the top cabinet drawer.
Refer to replacement procedure 4.2.
6.12.8Locate the TEE fitting in the ¼ in. diameter O
regulator output line, and
2
remove the plug from the TEE fitting.
6.12.9Connect a dedicated O
NOTE:Two test gauges are required to avoid contamination of the O
other gases: a dedicated test gauge for O
test gauge to the TEE fitting.
2
, and a second test gauge for the
2
circuit from the
2
other gases.
6.12.10Open the O
cylinder valve. Turn the System Power switch to ON, and turn the
2
gas selector switch to ALL GASES - if applicable.
6.12.11Verify the oxygen flow is 4 liters per min.
( )6.12.12On the test gauge, what is the regulator output press.? ___PSI (43-49)
NOTE:Leave the dedicated O
test gauge connected for later use in the Oxygen Supply
2
Pressure Alarm Test.
6.12.13Turn the System Power switch to STANDBY.
6.12.14For the other pressure regulators, locate their corresponding TEE fittings in
the ¼ in. diameter regulator output line, and remove the plug from the TEE
fitting (one at a time).
6.12.15Connect the second test gauge to the TEE fitting.
6-26
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.12.16Open the corresponding cylinder valve and set the System Power switch to ON.
6.12.17Set the corresponding flow to 4 l/min.
( )6.12.18On the second test gauge, what is the regulator output pressure? ___ PSI (40 -
49 for N2O), (43 - 49 for Air)
6.12.19Depressurize the gas circuit.
6.12.20Remove the test gauge and replace the plug in the TEE fitting.
( )6.12.21Repeat Steps 6.12.13 thru 6.12.20 for all other gases except CO2circuits.
Rev. Y
6-27
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.13 HIGH PRESSURE LEAK
( )6.13.1Oxygen High Pressure Leak
6.13.1.1Verify the System Power switch is at STANDBY.
6.13.1.2Open one (1) oxygen cylinder valve.
6.13.1.3Let the pressure stabilize.
6.13.1.4Close the oxygen cylinder valve and remove the cylinder.
6.13.1.5Observe the oxygen cylinder pressure gauge.
6.13.1.6After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.1.7Attach the cylinder.
( )6.13.2Nitrous Oxide High Pressure Leak
6.13.2.1Turn the System Power switch to ON.
6.13.2.2Open one (1) oxygen cylinder valve and one (1) nitrous oxide cylinder
valve.
6.13.2.3Adjust the oxygen flow to 4 l/min.
6.13.2.4Let the pressure stabilize.
6.13.2.5Close the nitrous oxide cylinder valve and remove the cylinder.
6.13.2.6Observe the nitrous oxide cylinder pressure gauge.
6.13.2.7After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.2.8Attach the cylinder.
6.13.2.9Close the oxygen flow control valve.
6-28
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( )6.13.3Air High Pressure Leak - if applicable
*6.13.3.1Set the gas selector switch to ALL GASES.
6.13.3.2Open the air cylinder valve.
6.13.3.3Let the pressure stabilize.
6.13.3.4Close the air cylinder valve and remove the cylinder.
6.13.3.5Observe the air cylinder pressure gauge.
6.13.3.6After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.3.7Attach the air cylinder.
( )6.13.4Oxygen-Helium High Pressure Leak - if applicable
*6.13.4.1Set the gas selector switch to ALL GASES.
6.13.4.2Open the oxygen-helium cylinder valve.
6.13.4.3Let the pressure stabilize.
6.13.4.4Close the oxygen-helium cylinder valve and remove the cylinder.
6.13.4.5Observe the oxygen-helium cylinder pressure gauge.
6.13.4.6After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.4.7Attach the oxygen-helium cylinder.
( )6.13.5Carbon Dioxide High Pressure Leak - if applicable
*6.13.5.1Set the gas selector switch to ALL GASES.
6.13.5.2Open the carbon dioxide cylinder valve.
Rev. Y
6.13.5.3Let the pressure stabilize.
6.13.5.4Close the carbon dioxide cylinder valve and remove the cylinder.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-29
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.13.5.5Observe the carbon dioxide cylinder pressure gauge.
6.13.5.6After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.5.7Attach the carbon dioxide cylinder.
( )6.13.6Nitrogen High Pressure Leak - if applicable
6.13.6.1Set the gas selector switch to ALL GASES.
6.13.6.2Open the nitrogen cylinder valve.
6.13.6.3Let the pressure stabilize.
6.13.6.4Close the nitrogen cylinder valve and remove the cylinder.
6.13.6.5Observe the nitrogen cylinder pressure gauge.
6.13.6.6After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.6.7Attach the nitrogen cylinder.
( )6.13.7Helium High Pressure Leak - if applicable
6.13.7.1Set the gas selector switch to ALL GASES.
6.13.7.2Open the helium cylinder valve.
6.13.7.3Let the pressure stabilize.
6.13.7.4Close the helium cyl. valve and remove the cylinder.
6.13.7.5Observe the helium cylinder pressure gauge.
6.13.7.6After two (2) minutes, what is the pressure loss? ___ PSI (<50).
6.13.7.7Attach the helium cylinder.
6-30
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.14 OXYGEN SUPPLY FAILURE PROTECTION
6.14.1Nitrous Oxide O.F.P.D.
6.14.1.1Open and close the oxygen cylinder valve.
6.14.1.2Open the nitrous oxide cylinder valve.
6.14.1.3Set the O
and N2O flows to 4 l/min.
2
( )6.14.1.4Does the flow of nitrous oxide cease when the oxygen pressure is
depleted? ___ (Y)
6.14.1.5Connect the O2pipeline supply.
6.14.1.6Close the nitrous oxide cylinder valve and bleed the pressure from the
circuit.
6.14.1.7Connect the N
6.14.1.8Disconnect the O
O pipeline supply.
2
pipeline supply.
2
( )6.14.1.9Does the flow of nitrous oxide cease when the oxygen pressure is
depleted? ___ (Y)
6.14.1.10 Close the nitrous oxide flow control valve.
6.14.1.11 Disconnect the N
O pipeline supply.
2
Rev. Y
6-31
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.14.2Air O.F.P.D. - if applicable
*6.14.2.1Set the gas selector switch to ALL GASES.
6.14.2.2Open the air cylinder valve - if applicable.
6.14.2.3Connect the air pipeline hose - if applicable.
6.14.2.4Open one (1) oxygen cylinder valve.
6.14.2.5Set the air flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.2.6Close the oxygen cylinder valve.
( )6.14.2.7Does the flow of air cease when the oxygen pressure is depleted? ___ (Y)
6.14.2.8Close the air flow control valve.
6.14.2.9Disconnect the air pipeline hose - if applicable.
6.14.3Oxygen-Helium O.F.P.D. - if applicable
*6.14.3.1Set the gas selector switch to ALL GASES.
6.14.3.2Open the oxygen-helium cylinder valve.
6.14.3.3Open one (1) oxygen cylinder valve.
6.14.3.4Set the oxygen-helium flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.3.5Close the oxygen cylinder valve.
( )6.14.3.6Does the flow of oxygen-helium cease when the oxygen pressure is
depleted? ___ (Y)
6.14.3.7Close the oxygen-helium flow control valve.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-32
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.14.4Carbon Dioxide O.F.P.D. - if applicable
*6.14.4.1Set the gas selector switch to ALL GASES.
6.14.4.2Open the cylinder of carbon dioxide.
6.14.4.3Open one (1) oxygen cylinder valve.
6.14.4.4Set the carbon dioxide flow to 500 ml; set the oxygen flow to 4 l/min.
6.14.4.5Close the oxygen cylinder valve.
( )6.14.4.6Does the flow of carbon dioxide cease when the oxygen pressure is
depleted? ___ (Y)
6.14.4.7Close the carbon dioxide flow control valve.
*6.14.4.8Set the gas selector switch to O
2+N2
O.
6.14.5Nitrogen O.F.P.D. - if applicable
6.14.5.1Set the gas selector switch to ALL GASES.
6.14.5.2Open the nitrogen cylinder valve.
6.14.5.3Open one (1) oxygen cylinder valve.
6.14.5.4Set the nitrogen flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.5.5Close the oxygen cylinder valve.
( )6.14.5.6Does the flow of nitrogen cease when the oxygen pressure is depleted? ___
(Y)
6.14.5.7Close the nitrogen flow control valve.
6.14.5.8Set the gas selector switch to O
2+N2
O.
6.14.5.9Close the oxygen flow control valve.
Rev. Y
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-33
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.14.6Helium O.F.P.D. - if applicable
6.14.6.1Set the gas selector switch to ALL GASES.
6.14.6.2Open the helium cylinder valve.
6.14.6.3Open one (1) oxygen cylinder valve.
6.14.6.4Set the helium flow to 4 l/min; set the oxygen flow to 4 l/min.
6.14.6.5Close the oxygen cylinder valve.
( )6.14.6.6Does the flow of helium cease when the oxygen pressure is depleted? ___
(Y)
6.14.6.7Close the helium flow control valve.
6.14.6.8Set the gas selector switch to O
+N2O.
2
6.14.6.9Close the oxygen flow control valve.
6.14.7Oxygen Supply Pressure Alarm
6.14.7.1If not already connected, connect a dedicated O
fitting in the O
regulator output line.
2
test gauge to the TEE
2
6.14.7.2Open and close an oxygen cylinder.
6.14.7.3Set the oxygen flow to 2 l/min.
( )6.14.7.4What is the pressure on the dedicated O2test gauge when the "O2
SUPPLY PRESSURE" LED turns on? ___ PSI (34 - 40)
6.14.7.5Does the O2 SUPPLY LOW message appear in the Caution display? ___
(Y)
6.14.7.6Bleed the remaining O
pressure from the system, then close the flow
2
control valve.
6.14.7.7Remove the test gauge from the TEE fitting in the O
line and replace the plug.
6-34
regulator output
2
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.15 FLOWMETERS
( )6.15.1Oxygen Flowmeter
6.15.1.1Open the O
cylinder valve.
2
6.15.1.2Is it possible to adjust the flow of oxygen over the full range of the
flowmeters? ___ (Y)
6.15.1.3Close the O
6.15.1.4Connect the O
cylinder valve and bleed the pressure.
2
pipeline supply, and verify the operation of the oxygen
2
flowmeter.
6.15.1.5Is the correct flow control knob and label attached to the oxygen flow
control valve? ___ (Y)
6.15.1.6Close the oxygen flow control valve.
*6.15.1.7Set the Gas Selector switch to the O
+N2O position.
2
( )6.15.1.8What is the minimum flow of oxygen? ___ ml (100-200 ml/min);
(0 ml for 4107615A (Minimum O
Flow Elimination))
2
( )6.15.2Nitrous Oxide Flowmeter
6.15.2.1Set the oxygen flow to 4 l/min.
6.15.2.2Open the nitrous oxide cylinder valve.
6.15.2.3Is it possible to adjust the flow of nitrous oxide over the full range of the
flowmeter? ___ (Y)
6.15.2.4Close the nitrous oxide cylinder valve and bleed the pressure.
6.15.2.5Connect the N
O flowmeter.
N
2
6.15.2.6Is the correct flow control knob and label attached to the N
O pipeline supply, and verify the proper operation of the
2
O flow control
2
valve? ___ (Y)
6.15.2.7Close the oxygen and nitrous oxide flow control valves.
*These items do not apply to 2-gas machines or later design (P/N 4113329-001) ORC.
Rev. Y
6-35
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
( )6.15.3Air Flowmeter - if applicable
*6.15.3.1Set the gas selector switch to ALL GASES.
**6.15.3.2Open the air cylinder valve.
**6.15.3.3Is it possible to adjust the flow of the air over the full range of the
flowmeter? ___ (Y)
**6.15.3.4Close the air cylinder valve and bleed the pressure.
6.15.3.5Close the air flow control valve.
6.15.3.6Connect the Air pipeline supply and verify the proper operation of the air
flowmeter. Close the air flow control valve.
6.15.3.7Is the correct flow control knob and label attached to the air flow control
valve? ___ (Y)
*6.15.3.8What is the minimum flow of oxygen? ___ ml (0)
( )6.15.4Oxygen-Helium Flowmeter - if applicable
*6.15.4.1Set the gas selector switch to ALL GASES.
6.15.4.2Open the oxygen-helium gas cylinder valve.
6.15.4.3Is it possible to adjust the flow of the oxygen-helium over the full range
of the flowmeter? ___ (Y)
6.15.4.4Close the oxygen-helium flow control valve.
6.15.4.5Is the correct flow control knob and label attached to the oxygen-helium
flow control valve? ___ (Y)
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
**these items do not apply to 4-gas machines.
6-36
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( )6.15.5Carbon Dioxide Flowmeter - if applicable
*6.15.5.1Set the gas selector switch to ALL GASES.
6.15.5.2Open the carbon dioxide gas cylinder valve.
6.15.5.3Is it possible to adjust the flow of the carbon dioxide over its range of 600
ml/min.? ___ (Y)
6.15.5.4Close the carbon dioxide flow control valve.
6.15.5.5Is the correct flow control knob and label attached to the carbon dioxide
flow control valve? ___ (Y)
( )6.15.6Nitrogen Flowmeter - if applicable
6.15.6.1Set the gas selector switch to ALL GASES.
6.15.6.2Open the nitrogen gas cylinder valve.
6.15.6.3Is it possible to adjust the flow of the nitrogen over the full range of the
flowmeter? ___ (Y)
6.15.6.4Close the nitrogen flow control valve.
6.15.6.5Is the correct flow control knob and label attached to the nitrogen flow
control valve? ___ (Y)
( )6.15.7Helium Flowmeter - if applicable
6.15.7.1Set the gas selector switch to ALL GASES.
6.15.7.2Open the helium gas cylinder valve.
6.15.7.3Is it possible to adjust the flow of helium over the full range of the
flowmeter? ___ (Y)
6.15.7.4Close the helium flow control valve.
Rev. Y
6.15.7.5Is the correct flow control knob and label attached to the helium flow
control valve? ___ (Y)
6.15.7.6Set the gas selector switch to O
+N2O.
2
*These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-37
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.16 O2 MED
6.16.1Disconnect the O2sensor cable from the oxygen sensor interface.
6.16.2The following messages shall appear on the display: "O
CAL", "O
6.16.3Reconnect the O
6.16.4The "O
6.16.5Press the "21%" key to calibrate the O
ALRM OFF".
2
Med sensor.
2
SENS DISC" message shall disappear.
2
Med.
2
SENS DISC", "O2NOT
2
NOTE:Make sure the sensor is stabilized in ambient air for several minutes.
( )6.16.6After the calibration is completed, what is the oxygen concentration? ___ % (21)
6.16.7This step intentionally left blank.
6.16.8The alarm CRT shall display "% OXYGEN LOW" under the warning heading.
There should be a continuous audible alarm.
6.16.9What is the low oxygen alarm default? ___ % (30)
6.16.10This step intentionally left blank.
6.16.11Press the LO oxygen alarm key. Does a box appear around the low alarm
limit? ___ (Y)
6.16.12Verify that the low alarm limit has a range from 21 to 99% (18 to 99% with
firmware version 2.03 and later).
6.16.13Place the sensor into the inspiratory valve dome adapter. Set the Man/Auto
selector to BAG, close the APL valve, attach a 12-inch hose to the inspiratory
valve, and occlude the bag mount.
6.16.14Set the oxygen flow to 4 l/min.
6.16.15Set the low limit to the lowest setting possible, and verify the "% OXYGEN
LOW" message has cleared.
6.16.16Press the HI oxygen alarm key. Does a box appear around the high alarm
limit? ___(Y)
6-38
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.16.17What is the high oxygen alarm default? ___ % (100)
6.16.18Verify that the high alarm limit has a range from 100 to 22%.
6.16.19Set the high alarm to 95.
6.16.20the alarm CRT shall display "% OXYGEN HI" under the ADVISORY column.
6.16.21Return the high alarm to 100.
6.16.22The "% OXYGEN HI" message will now disappear.
( )6.16.23Within three minutes, what is the oxygen concentration? ___% (97-100)
Rev. Y
6-39
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.17 OXYGEN CONCENTRATIONS
6.17.1Oxygen + Nitrous Oxide Concentration Test
6.17.1.1Verify the oxygen flow is at 4 l/min.
6.17.1.2Depress the O
6.17.1.3Does the O
Flush button for 5 seconds.
2
Med read 97-100% after the value stabilizes?__(Y)
2
6.17.1.4Set the nitrous oxide flow to 2 l/min.
( )6.17.1.5After the value stabilizes, what is the oxygen concentration? ___ % (64-70)
6.17.1.6Close the nitrous oxide flow control valve.
6.17.2Oxygen + Air Concentration Test - if applicable
6.17.2.1Depress the O
6.17.2.2Does the O
FLUSH button for 5 seconds.
2
Med read 97-100% after the value stabilizes?__(Y)
2
*6.17.2.3Set the gas selector switch to ALL GASES.
6.17.2.4Verify the O
flow is at 4 l/min; set the air flow to 2 l/min.
2
( )6.17.2.5After the value stabilizes, what is the O2concentration? ___%(71-77)
6.17.2.6Close the air cylinder valve and bleed the pressure from the circuit - if
applicable.
6.17.2.7Close the air flow control valve.
*6.17.2.8Set the gas selector switch to O
2+N2
O.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-40
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.17.3Oxygen + Helium and Oxygen Concentration - if applicable
6.17.3.1Depress the "O
6.17.3.2Does the O
FLUSH" for 15 seconds.
2
Med read 97-100% within 3 minutes? ___ (Y)
2
*6.17.3.3Set the gas selector switch to ALL GASES.
6.17.3.4Verify the oxygen flow is at 4 l/min.
6.17.3.5Set the oxygen-helium flow to 2 l/min.
( )6.17.3.6After the value stabilizes, what is the O2concentration? ___%(72-78)
6.17.3.7Close the oxygen-helium cylinder valve and bleed the pressure from the
circuit.
6.17.3.8Close the oxygen-helium flow control valve.
*6.17.3.9Set the gas selector switch to O
2+N2
O.
6.17.4Oxygen + Carbon Dioxide Concentration - if applicable
6.17.4.1Depress the O
FLUSH button for 15 seconds.
2
6.17.4.2Set the oxygen flow to 1000 ml.
6.17.4.3Does the O
MED read 97-100% within 3 min?___(Y)
2
*6.17.4.4Set the gas selector switch to ALL GASES.
6.17.4.5Set the carbon dioxide flow to 500 ml.
( )6.17.4.6What is the O2concentration after the value stabilizes? ___%(64-70)
6.17.4.7Close the carbon dioxide cylinder valve and bleed the pressure from the
circuit.
6.17.4.8Close the carbon dioxide flow control valve.
*6.17.4.9Set the gas selector switch to O
2+N2
O.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
6-41
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.17.5Oxygen + Nitrogen Concentration - if applicable
6.17.5.1Depress the O2FLUSH button for 5 seconds.
6.17.5.2Does the O
MED read 97-100% within 3 minutes?___ (Y)
2
6.17.5.3Set the gas selector switch to ALL GASES.
6.17.5.4Set O
flow to 4 l/min; nitrogen flow to 2 l/min.
2
( )6.17.5.5What is the oxygen concentration after the value stabilizes? ___ % (64-70)
6.17.5.6Close the nitrogen cylinder valve and bleed the pressure from the circuit.
6.17.5.7Close the nitrogen flow control valve.
6.17.5.8Set the gas selector switch to O
2+N2
O.
6.17.5.9Close the oxygen flow control valve.
6.17.6Oxygen + Helium Concentration - if applicable
6.17.6.1Depress the O
6.17.6.2Does the O
FLUSH button for 15 seconds.
2
MED read 97-100% within 3 minutes? ___ (Y)
2
6.17.6.3Set the gas selector switch to ALL GASES.
6.17.6.4Set O
flow to 4 l/min; helium flow to 2 l/min.
2
( )6.17.6.5What is the oxygen concentration after the value stabilizes? ___% (64-70)
6.17.6.6Close the helium cylinder valve and bleed the pressure from the circuit.
6.17.6.7Close the helium flow control valve.
6.17.6.8Set the gas selector switch to O
2+N2
O.
6.17.6.9Close the oxygen flow control valve.
6-42
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( ) 6.18 AUXILIARY OXYGEN FLOWMETER - if applicable
6.18.1Close the auxiliary oxygen flowmeter flow control valve.
6.18.2Connect a cm H
O pressure manometer to the outlet.
2
6.18.3Is there an increase in pressure? ___ (N)
6.18.4Apply a pressure of 50 cm H
6.18.5After 30 seconds, what is the pressure on the manometer? ___(>40 cm H
O to the manometer.
2
2
O)
6.18.6Remove the gauge and test fixture.
6.18.7Is it possible to adjust the flow over the full range of the auxiliary oxygen
flowmeter? ___ (Y)
6.18.8Set the flow rate to 5 l/min.
6.18.9Hold the sensor from a calibrated O
Med at the auxiliary oxygen flowmeter
2
outlet.
6.18.10After the value stabilizes, what is the oxygen concentration? ___ % (97-100)
6.18.11Remove the O
Med sensor from the auxiliary oxygen flowmeter, and insert it
2
into the inspiratory valve dome adapter.
6.18.12Close the flowmeter flow control valve.
Rev. Y
6-43
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.19AOXYGEN RATIO CONTROLLER - if applicable
*6.19A.1Set the Gas Selector switch to O2+N2O.
6.19A.2Depress the O
FLUSH for 5 seconds.
2
6.19A.3Set the oxygen flow to 1000 ml/min.
6.19A.4Open the nitrous oxide flow control valve to the stop position.
( )6.19A.5What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19A.6Adjust the oxygen flow to 1.5 l/min.
( )6.19A.7What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19A.8Adjust the oxygen flow to 2 l/min.
( )6.19A.9What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19A.10 Adjust the oxygen flow to 4 l/min.
( )6.19A.11 What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19A.12 Reduce the O
flow to 500 ml/min. Verify that the N2O flow is greater than or
2
equal to 600 ml/min.
6.19A.13 Close the oxygen flow control valve.
( )6.19A.14 What is the flow of nitrous oxide? ___ ml/min. (0) or 375-750 ml/min on
machines with later design ORC (P/N 4113329-001)
6.19A.15 What is the oxygen concentration with the O2flow control valve closed?___%
(>21)
6.19A.16 Close the nitrous oxide flow control valve.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
6-44
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.19BOXYGEN RATIO MONITOR CONTROLLER
6.19B.1Depress the O
6.19B.2Turn the N
FLUSH for 15 seconds.
2
O flow control valve counter-clockwise to its stop position.
2
6.19B.3Slowly increase the oxygen flow until nitrous oxide begins to flow.
6.19B.4What is the oxygen flow rate? ___(200-400 ml/min.)
( )6.19B.5Slowly increase the oxygen flow until the "O2/N2O FLOW RATIO" LED on the
alarm panel is lighted.
6.19B.6What is the nitrous oxide flow rate? ___(150-300 ml/min.)
6.19B.7Set the oxygen flow to 1000 ml/min.
6.19B.8Adjust the nitrous oxide flow until the ORMc controls the nitrous oxide.
( )6.19B.9What is the oxygen concentration after the value stabilizes? ___% (21-29)
6.19B.10 Does the "O
O FLOW RATIO" LED turn "ON" on the ALARM PANEL? ___
2/N2
(Y)
6.19B.11 Does the "O
O LOW" message appear in the ADVISORY column? ___ (Y)
2/N2
6.19B.12 Open the nitrous oxide flow control valve to the stop position.
6.19B.13 Adjust the oxygen flow to 1.5 l/min.
( )6.19B.14 What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19B.15 Adjust the oxygen flow to 2 l/min.
( )6.19B.16 What is the oxygen concentration after the value stabilizes? ___ % (21-29)
6.19B.17 Adjust the oxygen flow to 4 l/min.
( )6.19B.18 What is the oxygen concentration after the value stabilizes? ___ % (21-29)
*6.19B.19 Set the Gas Selector switch to ALL GASES.
* Does not apply to 2-gas machines.
Rev. Y
6-45
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.19B.20 Is the "O
*6.19B.21 Set the Gas Selector switch to "O
O FLOW RATIO" alarm activated? ___ (N)
2/N2
O".
2+N2
6.19B.22 Close the oxygen flow control valve.
( )6.19B.23 What is the flow of nitrous oxide? ___ ml/min. (0)
6.19B.24 Is the "O
O FLOW RATIO" alarm activated? ___ (N)
2/N2
6.19B.25 Close the nitrous oxide flow control valve.
* Does not apply to 2-gas machines.
6-46
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.20 BAROMED
6.20.1Disconnect the breathing pressure sensor line from the absorber.
6.20.2Connect a test pressure gauge and syringe to the breathing pressure sensor
line.
6.20.3Press the THRESHOLD key on the main keypad. A box shall appear around
the Threshold Alarm Limit.
6.20.4What is the threshold alarm default? ___ cm H
6.20.5Verify that the threshold alarm limit has a range from 5 to 30 cm H
O (12)
2
O.
2
6.20.6Adjust the threshold to 10 cm H2O.
6.20.7Press the HIGH key in the Baromed section of the main keypad. A box shall
appear around the High Pressure Alarm Limit.
6.20.8What is the high alarm limit default? ___ cm H
O (50)
2
6.20.9Verify that the high alarm limit has an adjustable range from 30 to 120 cm
O. (Later software versions allow range up to 125 cm H2O).
H
2
6.20.10Set the high alarm limit to 65 cm H
6.20.11Increase the pressure to 25 cm H
O.
2
O; decrease the pressure to 20 cm H2O.
2
6.20.12Does the THRESHOLD LOW message appear in the Advisory column? ___ (Y)
6.20.13Set the Man/Auto valve to AUTO, and turn the ventilator ON.
6.20.14Bleed the pressure and start a stopwatch.
( )6.20.15What is the time when APNEA-PRESSURE appears in the Caution column?
___ sec (13-17)
( )6.20.16What is the time when the APNEA-PRESSURE appears in the Warning
column? ___ sec (26-34)
6.20.17After the APNEA-PRESSURE alarm is displayed in the Warning column on
remote display, slowly increase the test pressure.
( )6.20.18At what pressure does the APNEA-PRESSURE alarm deactivate? ___ cm H2O
(7-13)
Rev. Y
6-47
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.20.19Adjust the threshold to 18 cm H
6.20.20Increase the pressure to 20 cm H
O.
2
O, maintain the pressure, and start a
2
stopwatch.
( )6.20.21What is the time when CONTINUOUS PRES appears in the Caution column?
___ sec (12-18)
( )6.20.22Decreasing the pressure slowly, what is the pressure at which the
CONTINUOUS PRES alarm deactivates? ___ cm H
O (15-21)
2
6.20.23Slowly increase the pressure.
( )6.20.24At what pressure does the VENT PRESSURE HI alarm activate? ___ cm H2O
(62-68)
6.20.25Bleed the pressure.
6.20.26Slowly create a sub-atmospheric pressure.
( )6.20.27At what pressure does the SUB ATM PRESSURE alarm activate? ___ cm H2O
(-7 to -13)
6.20.28Disconnect the test gaugeand syringe;reconnect the breathing pressure sensor
line to the absorber.
6.20.29Does the SUB ATM PRESSURE alarm deactivate? ___(Y)
6.20.30Enable the APNEA ALARM by pressing the APNEA ALARMS DISABLE key
in the main keypad.
6.20.31Verify that APNEA ALARM DISABLE key will not disable the APNEA Alarm
when the ventilator switch is ON.
6.20.32Remove the test equipment and reconnect the breathing pressure sensor line
to the absorber.
6-48
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.21 VENTILATOR
NOTE:Set the Pressure Limit control to MAX, and the PEEP valve to the minimum
position, if applicable.
**6.21.1Set the Man/Auto selector to BAG.
6.21.2Set the FREQUENCY to 10 BPM.
6.21.3Set the I:E RATIO to 1:2.
6.21.4Set the Tidal Volume to 1000 ml.
6.21.5Attach a patient circuit to the absorber system.
6.21.6Adjust the O
flow to 3 l/min.
2
6.21.7Is the APNEA-P ALARM OFF message displayed in the Advisory column? (Y)
(if no, press the APNEA ALARMS DISABLE key in the Pressure Alarms
section of the main keypad.
6.21.8Is the VOL-ALARMS OFF message displayed in the Advisory column? (Y) (if
no, press the VOLUME ALARMS DISABLE key in the Minute Volume Alarms
section of the main keypad.
6.21.9Turn the ventilator on.
**6.21.10Verify the FAULT indicator turns on (Y)
6.21.11Set the Man/Auto selector switch to AUTO.
**6.21.12Verify the FAULT indicator turns off (Y)
6.21.13Do the APNEA-P ALARM OFF and VOL-ALARMS OFF messages disappear
from the Advisory column? (Y)
6.21.14Adjust the INSPIRATORY FLOW to the maximum of the LOW zone.
6.21.15Occlude the Y-piece with your thumb.
6.21.16Inflate the bellows by momentarily pressing the O
Flush.
2
( )6.21.17What is the peak inspiratory pressure? ___ cm H2O (>30 cm H2O)
6.21.18Attach a 3-liter bag to the Y-piece.
** These items only apply to machines with an AV2+ ventilator.
Rev. Y
6-49
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.21.19Using a stopwatch, time the inspiratory phase.
( )6.21.20What is the inspiratory time? ___ seconds (1.8-2.2)
6.21.21Using a stopwatch, time the expiratory phase.
( )6.21.22What is the expiratory time? ___ seconds (3.6-4.4)
**6.21.23Press and hold the EXTENDED RANGE switch and scroll the I:E ratio dial
counter-clockwise and verify the extended I:E values increment (2:1, 3:1 and
4:1); return the I:E ratio to 2:1.
**6.21.24Using a stopwatch, time the inspiratory phase.
( )**6.21.25What is the inspiratory time? ___ seconds (3.6 - 4.4)
**6.21.26Using a stopwatch, time the expiratory phase.
( )**6.21.27What is the expiratory time? ___ seconds (1.8 - 2.2)
( )6.21.28Adjust the FREQUENCY and I:E RATIO through the following settings and
verify that the ventilator cycles properly:
FREQ.I:E RATIOFREQ.I:E RATIOFREQ.I:E RATIO
111:1441:2.5771:4
221:1.5551:3881:4.5
331:2661:3.5991:4.5
6.22 BELLOWS DRIVE GAS LEAK: ADULT
6.22.1Remove the ventilator hose from the VENTILATOR HOSE terminal.
6.22.2Attach a test terminal to the bellows assembly ventilator hose terminal.
6.22.3Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.22.4Set the FREQUENCY to 1 BPM.
6.22.5Set the I:E RATIO to 1:1.
6.22.6Set the INSPIRATORY FLOW to the maximum.
( )6.22.7What is the flow indicated during the inspiratory phase? ___ (<50 ml)
6.22.8Remove the test terminal and flowmeter test stand.
6.22.9Turn the ventilator OFF.
6-50
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.23 SPIROMED
6.23.1Press the LOW key in the Minute Volume Alarms section of the main keypad.
A box shall appear around the Minute Volume Alarm Limit.
6.23.2What is the low minute volume alarm default? ___ (3.0)
6.23.3Verify that the minute volume has a low alarm limit range from 0.5 to 10.0 by
increments of 0.1.
6.23.4Adjust the low minute volume alarm to 2.0 liters.
6.23.5Disable the VOLUME ALARM by pressing the VOLUME ALARMS DISABLE
key in the Minute Volume Alarms section of the main keypad. Turn on the
ventilator and start a stopwatch.
( )6.23.6What is the time when APNEA-VOLUME appears in the Caution column? ___
sec (13-17)
( )6.23.7What is the time when APNEA-VOLUME appears in the Warning column? ___
sec (26-34)
( )6.23.8After one (1) minute, does the MINUTE VOLUME LOW message appear in the
Caution column? ___ Y
6.23.9Insert a test minute volumeter in between the Spiromed sensor and the
exhalation valve.
6.23.10Attach a patient circuit with a 3-liter bag to the absorber system.
6.23.11Reconnect the ventilator hose to the Ventilator Hose terminal.
6.23.12Adjust the FREQUENCY to 6 BPM.
6.23.13Adjust the I:E RATIO to 1:2.
6.23.14Adjust the flow to the maximum of the LOW zone.
6.23.15Adjust the oxygen flow to 2 l/min.
6.23.16Adjust the Tidal Volume to 200 ml.
Rev. Y
6-51
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.23.17After the first breath is detected, does the APNEA-VOLUME in the Warning
column and the MINUTE VOLUME LO in the Caution column deactivate? ___
(Y)
6.23.18Adjust the low alarm limit above the indicated minute volume.
6.23.19Does the MINUTE VOLUME LO message appear in the Caution column? ___
(Y)
6.23.20Adjust the low alarm limit below the indicated minute volume.
6.23.21Does the MINUTE VOLUME LO message in the Caution column deactivate?
___ (Y)
6.23.22Increase the tidal volume to 1000 ml and the frequency to 10 BPM.
6.23.23Press the O
Flush momentarily to inflate the bellows.
2
( )6.23.24Are the tidal and minute volumes on the Spiromed and on the test volumeter
within 20% of each other? ___ (Y)
6.23.25Create a reverse flow by loosening the expiratory valve dome.
( )6.23.26Each time a reverse flow greater than 20 ml is detected, does the REVERSE
FLOW message appear in the Advisory column? ___ (Y)
6.23.27Tighten the expiratory valve dome.
6.23.28Disconnect the Spiromed sensor from the VOLUME SENSOR interface.
6.23.29Do the VOL SENSOR DISC and VOL ALARMS OFF messages appear in the
Advisory column? ___ (Y)
6.23.30Connect the Spiromed sensor to the Volume Sensor interface and verify that
the alarms clear.
6-52
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.24ABELLOWS: ADULT - if applicable
6.24A.1Set the FREQUENCY to 10 BPM.
6.24A.2Set the I:E RATIO to 1:2.
6.24A.3Adjust the O
flow to 300 ml/min.
2
6.24A.4Adjust the INSPIRATORY FLOW to MED.
6.24A.5Adjust the Tidal Volume to 200 ml.
6.24A.6Press the O
Flush momentarily to inflate the bellows.
2
( )6.24A.7What is the Tidal Volume on the test volumeter? ___ ml (125-250)
6.24A.8Adjust the Tidal Volume to 1000 ml.
6.24A.9Press the O
Flush momentarily to inflate the bellows.
2
( )6.24A.10 What is the Tidal Volume on the test volumeter? ___ ml (900-1100)
6.24A.11 Adjust the O
flow to 5 l/min.
2
6.24A.12 Adjust the INSPIRATORY FLOW to maximum.
6.24A.13 Adjust the Tidal Volume to maximum.
( )6.24A.14 What is the Tidal Volume on the test volumeter? ___ ml (≥1400)
6.24BVENTILATOR RELIEF VALVE: ADULT
6.24B.1Adjust the O
flow to 10 l/min.
2
6.24B.2Adjust the INSPIRATORY FLOW to MED.
6.24B.3Adjust the I:E RATIO to 1:3, and the FREQUENCY to 10.
6.24B.4Adjust the Tidal Volume to 1200 ml.
( )6.24B.5What is the PEEP? ___ cm H2O(≤3)
6.24B.6Adjust the O
flow to 500 ml.
2
( )6.24B.7Does the ventilator deliver the full Tidal Volume during 5 consecutive cycles?
___ (Y)
6.24B.8Does the bellows stop adjust smoothly? ___ (Y)
Rev. Y
6-53
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.25ABELLOWS: PEDIATRIC EXTERNAL - if applicable
6.25A.1Set the ventilator frequency to 10 BPM.
6.25A.2Set the I:E Ration to 1:2.
6.25A.3Adjust the tidal volume to 300 ml/min.
6.25A.4Turn the ventilator ON.
6.25A.5Adjust the Inspiratory Flow control on the ventilator to Medium.
6.25A.6Adjust the fine flow control of the Ped Bellows attachment fully clockwise.
6.25A.7Set the O
to 3 l/min.
2
6.25A.8Occlude the 15mm patient side of the Y-piece and fill the bellows by pressing
the O
Flush button.
2
6.25A.9Observe the absorber breathing pressure gauge as the ventilator cycles.
6.25A.10 What is the pressure when the bellows completes its downward travel?
___cmH
O >30
2
6.25A.11 What is the pressure when the bellows completes its upward travel? ___cmH
≤3
6.25A.12 Verify that with the Pediatric Bellows Fine Flow Control turned fully counterclockwise the bellows does not collapse during inspiration.
6.25A.13 Installa3Lbreathing bag onto the patient circuit Y-piece.
6.25A.14 Fill the bellows by pressing the O
6.25A.15 Adjust the O
flow to 10 l/min.
2
Flush button.
2
( )6.25A.16 What is the PEEP? ___cmH2O(≤3)
O
2
6.25A.17 Adjust the O
flow to 500 ml/min.
2
( )6.25A.18 Does the ventilator deliver the full tidal volume during 5 consecutive cycles?
___(Y)
6.25A.19 Set the ventilator frequency to 20 BPM.
6-54
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.25A.20 Adjust the O
flow to 5 l/min.
2
( )6.25A.21 What is the tidal volume on the test volumeter? ___ml (>250)
6.25A.22 Remove the ventilator hose from the ventilator hose terminal.
6.25A.23 Attach a test terminal to the bellows assembly ventilator hose terminal.
6.25A.24 Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.25A.25 Set the frequency to 1 BPM.
6.25A.26 Set the I:E RATIO to 1:1.
( )6.25A.27 What is the flow indicated during the inspiratory phase? ___(<50 ml)
6.25A.28 Remove the test terminal and flowmeter test stand.
6.25A.29 Return all controls to their original position.
Rev. Y
6-55
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.25BBELLOWS: PEDIATRIC INTERNAL - if applicable
6.25B.1Set the ventilator frequency to 10 BPM.
6.25B.2Set the I:E Ration to 1:2.
6.25B.3Turn the ventilator ON.
6.25B.4Adjust the Inspiratory Flow control on the ventilator to Medium.
6.25B.5Set the O
to 3 l/min.
2
6.25B.6Occlude the 15mm patient side of the Y-piece and fill the bellows by pressing
the O
Flush button.
2
6.25B.7Observe the absorber breathing pressure gauge as the ventilator cycles.
6.25B.8What is the pressure when the bellows completes its downward travel?
___cmH
O >30
2
6.25B.9What is the pressure when the bellows completes its upward travel? ___cmH
≤3
6.25B.10 Installa3Lbreathing bag onto the patient circuit Y-piece.
6.25B.11 Fill the bellows by pressing the O
6.25B.12 Adjust the O
flow to 10 l/min.
2
Flush button.
2
( )6.25B.13 What is the PEEP? ___cmH2O(≤3)
6.25B.14 Adjust the O
flow to 500 ml/min.
2
O
2
6.25B.15 Adjust the inspiratory flow control until the bellows collapses entirely.
( )6.25B.16 Does the ventilator deliver the full tidal volume during 5 consecutive cycles?
___(Y)
6.25B.17 Set the ventilator frequency to 20 BPM.
6.25B.18 Adjust the O
flow to 5 l/min.
2
6.25B.19 Adjust the inspiratory flow control until the bellows collapses to the 100 ml
mark on the pediatric bellows assembly.
6-56
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
( )6.25B.20 What is the tidal volume on the test volumeter? ___ml (65-135)
6.25B.21 Adjust the inspiratory flow control until the bellows collapses entirely.
( )6.25B.22 What is the tidal volume on the test volumeter? ___ml (>240)
6.25B.23 Close the O
flow control valve.
2
6.25B.24 Remove the ventilator hose from the ventilator hose terminal.
6.25B.25 Attach a test terminal to the bellows assembly ventilator hose terminal.
6.25B.26 Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.25B.27 Set the frequency to 1 BPM.
6.25B.28 Set the I:E RATIO to 1:1.
( )6.25B.29 What is the flow indicated during the inspiratory phase? ___(<50 ml)
6.25B.30 Remove the test terminal and flowmeter test stand.
6.25B.31 Return all controls to their original position.
Rev. Y
6-57
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.26 TRACE & TREND
( )6.26.1Trace:Press the TRACE key and verify that the lower half of the alarm
CRT displays the Breathing Pressure waveform.
( )6.26.2Trend:Press the TREND key and verify that the lower half of the alarm
CRT can selectively display percent oxygen, breaths per minute, and
minute volume.
NOTE:Information must be available in order to be displayed.
6.26.2.1Turn the ventilator OFF.
6.27 PRESSURE LIMIT CONTROLLER (AV-2, or optional PLC, if applicable)
6.27.1Turn the ventilator ON.
6.27.2Set the Inspiratory Flow to the bottom of the low range.
6.27.3Set the oxygen flow rate to 4 l/min.
6.27.4Set the Pressure Limit Control to the MAX position.
6.27.5Occlude the Y-piece with your thumb.
( )6.27.6Slowly increase the Inspiratory Flow setting until a peak pressure of 80 cm
H
O is achieved.
2
6.27.7Set the Pressure Limit Control to 30.
( )6.27.8What is the peak pressure? ___ cm H2O (27-33)
6.27.9Set the Pressure Limit Control to the Min position.
( )6.27.10What is the peak pressure? ___ cm H2O (<15)
6.27.11Remove your thumb from the Y-piece.
6.27.12Set the Inspiratory Flow to the maximum of the LOW zone.
6.27.13Close the oxygen flow control valve.
6.27.14Turn the ventilator OFF.
6-58
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.28 BELLOWS PEEP VALVE - if applicable
6.28.1Connect a 12 in. hose between the inspiratory and expiratory valves, and set
the oxygen flow to 5 l/min.
6.28.2Adjust the PEEP valve clockwise to the maximum position.
6.28.3Turn the ventilator ON.
( )6.28.4What is the PEEP on the Baromed monitor? ___ cm H2O (18-25)
6.28.5Adjust the PEEP valve counter-clockwise to the minimum position.
6.28.6What is the PEEP on the Baromed monitor? ___ cm H
6.28.7Close the O
6.28.8Turn the ventilator OFF.
( ) 6.29 AUDIO SILENCE
6.29.1Turn the System Power switch to STANDBY, then turn it ON.
6.29.2Verify the 120-sec. delay at power-up and allow a full countdown.
6.29.3Create any warning alarm.
6.29.4Press the key labeled with a crossed-out speaker twice.
6.29.5Does the "SILENCE 120" message appear on the display? Is the audio alarm
silenced during the count down? ___ (Y)
6.29.6Does the "SILENCE" count down to 0 and then disappear? ___ (Y)
6.29.7Press the key labeled with a crossed-out speaker once.
flow control valve.
2
O(≤3)
2
Rev. Y
6.29.8Does the "SILENCE 60" message appear on the display? ___ (Y)
6.29.9Press the key labeled with a speaker producing sound.
6.29.10Does the "SILENCE" message disappear from the display? ___ (Y)
6-59
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)NM2B
6.30 OXYGEN FLUSH VALVE
Turn the System Power switch to STANDBY.
6.30.1Press and release the O
FLUSH button.
2
6.30.2Does the flow of oxygen stop immediately? __ (Y)
6.30.3Close the APL valve.
6.30.4Connect a 12-inch hose to the inspiratory valve.
6.30.5Set the Man/Auto selector to BAG.
6.30.6Occlude the bag mount.
6.30.7Insert the sensor from a calibrated O
Med into the valve dome adapter on the
2
inspiratory valve.
6.30.8Close all flow control valves.
6.30.9Press the O
FLUSH button.
2
( )6.30.10What is the O2concentration after the value stabilizes?___%O2(97-100)
6.30.11Remove the O
Med sensor and install the plug.
2
6.30.12Remove the test minute volumeter (P/N 2212300) from the absorber and
connect it to the common gas outlet, using Fresh Gas Outlet Volume Test
Device (P/N S010158).
6.30.13Press and hold the O
FLUSH button for 15 seconds; multiply the value by 4.
2
( )6.30.14What is the oxygen flush flow rate? ___l/min. (45-65)
6.30.15Remove the test minute volumeter and test fixture, and reconnect the fresh gas
hose.
6.30.16Turn the System Power switch to ON.
6-60
Rev. Y
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM2BPMC PROCEDURE (continued)
6.31 FINAL TESTS
( )6.31.1Operator’s Instruction Manual
6.31.1.1Verify that the availability/locationof the machine’s Operator’s
Instruction Manual is in close proximity of the machine.
( )6.31.2Auxillary Lamp - if applicable
6.31.2.1Verify that the lamp head (shroud) can move approximately 90°.
6.31.2.2Verify that the lamp head (shroud) can be positioned at any point within
the 90° range and holds fast.
6.31.2.3Verify that the lamp can rotate 360° and hits a stop located within 45° on
either side of the top center point.
6.31.2.4Verify that the lamp is plugged into a live AC receptacle.
6.31.2.5Turn the lamp ON.
6.31.2.6Adjust the dimmer control and verify that the light varies with the
adjustment of the knob.
6.31.2.7Turn the lamp OFF.
( )6.31.3Final Check
6.31.3.1Verify that all cylinder pressure gauges indicate zero.
6.31.3.2Verify that the pipeline hoses are connected to the hospital pipeline.
6.31.3.3Verify that the APL valve knob is turned completely counter-clockwise
(fully open).
6.31.3.4Place the Auto/Man selector in the BAG position.
6.31.3.5Verify that the O
6.31.3.6Verify that the valve dome is plugged.
Med sensor is removed from the valve dome adapter.
2
Rev. Y
6.31.3.7Verify that the machine is plugged into a live outlet.
6.31.3.8Return all machine controls and settings to their original state.
6-61
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
DrägerService
RETURN TO CD-ROM TABLE OF CONTENTS
®
DrägerService is a division of
Draeger Medical, Inc.
3122 Commerce Drive
Telford, PA 18969
Tel: (215) 721-5402