Druck UPS III Specifications

UPS-III
Druck Loop Calibrator
The UPS-III is a rugged and extremely compact loop calibrator. Measuring 77 mm x 129 mm
(3 in. x 5 in.) and weighing just 275 g (9.7 oz), it ts
comfortably in a shirt pocket. It is an essential tool for loop testing, instrument maintenance and valve set-up, with an easy to read display and simple to use time saving features.
Features
• Measure or source 0 to 24 mA
• Accuracy 0.01% of reading
Dual mA or % readout, linear or ow
• Step, span check, valve check, ramp
• 60 VDC measurement and continuity
• HART® compatible
• Adjustable display contrast
• Adjustable resolution
• Uses industry standard AA batteries
• Battery voltage and battery low indicator
• Information screen reports serial number, software version and calibration date
• Primary reading in mA or %
bhge.com
Auto power o (enable or disable)
• Pin protected closed case calibration
• Optional protective over-boot with bench stand, hanging loop and quick connect belt clip
• Optional AC line adaptor
Graphic Display
Auto Step and Ramp
The graphic display with menu-driven interface is easier to use than traditional knobs, switches, multi-function keys and dual key sequences.
Measure or Source 0 to 24 mA
An internal 24 V loop supply is available in both measure and source modes, essential during plant shutdowns.
Accuracy 0.01% of Reading
This includes 12 month stability and temperature
eects. The UPS-III can maintain a 4:1 calibration ratio
with the latest instrumentation and is typically 20 to 80 times more accurate than a Digital Multi-Meter (DMM).
Simultaneous % Readout
Displays mA and the percentage value of 4 to 20 mA or 0 to 20 mA. The UPS-III also converts mA readings to %
ow. This is not possible with many loop calibrators.
Outputs have adjustable rates (1 to 599 s) for single­handed loop testing, valve set-up and slew rate checks.
“Fast Set” Output
Allows valves to be set to within 0.001 mA using the arrow keys. Holding a key quickly ramps the output to the next valve. The system is ideal for alarm trip tests.
60 VDC Measurement
For loop diagnostics and maintenance of voltage output instruments
Continuity and Switch Test Facility
This reduces DMM dependence when fault nding,
checking loop integrity and testing switches.
HART Compatible
Internal 235 Ω loop resistor (menu selectable)
EMC Compliance to EN 61326-1
Step, Span Check and Value Check
Step quickly through linearity tests and zero/span adjustments. In valve mode, seating can be checked
and the positioner set.
Step Mode 0% 25% 50% 75% 100%
4 to 20 mA linear
0 to 20 mA linear
4 to 20 mA
ow
0 to 20 mA
ow
4 to 20 mA span check
0 to 20 mA span check
4 to 20 mA valve
4 8 12 16 20
0 5 10 15 20
4 5 8 13 20
0 1.25 5 11.25 20
4 --- --- --- 20
0 --- --- --- 20
3.8, 4, 4.2 --- 12 --- 19, 20, 21
Complies with heavy industry standard. Most loop calibrators are only suitable for light industrial use.
Loading...