Druck UNIK 5000 Specifications

Page 1
UNIK 5000
Pressure Sensing Platform
The new UNIK 5000 is a high performance configurable solution to pressure measurement. The use of micromachined silicon technology and analogue circuitry enables best in class performance for stability, low power and frequency response. The new platform enables you to easily build up your own sensor to match your own precise needs. This high performance, configurable solution to pressure measurement employs modular design and lean manufacturing techniques.
High Quality
The combination of a high technology sensor, together with advanced signal conditioning and packaging techniques, provides an ideal long term solution for reliable, accurate and economical measurements.
Bespoke as Standard
Custom-built from standard components, manufacturing sensors to your requirement is fast and simple; each UNIK 5000 is a “bespoke” pressure sensing solution, but with the short lead times and competitive pricing you would expect from standard products.
Expertise
We have the people and the knowledge to support your needs for accurate and reliable product performance; our team of experts can help you make the right sensor selection, guiding you and providing the help and tools you need. It is important to ensure that the sensor material and performance selected are suitable for your application.
Features
Ranges from 70 mbar (1 psi) to 700 bar (10000 psi)
Accuracy to ±0.04% Full Scale (FS) Best Straight Line (BSL)
Stainless Steel construction
Frequency response to 3.5 kHz
High over pressure capability
Hazardous Area certifications
mV, mA, voltage and configurable voltage outputs
Multiple electrical and pressure connector options
Operating temperature ranges from –55 to +125°C (-67 to +257°F)
Online configuration tool
Druck.com
Page 2
UNIK 5000 Specifications
Supply and Outputs
Measurement Operating Pressure Ranges:
Gauge Ranges
Any zero based range 70 mbar to 70 bar (1 to 1000 psi)
Note: All psi values are approximate.
Sealed Gauge Ranges
Any zero based range 10 to 700 bar (145 to 10000 psi)
Absolute Ranges
Any zero based range 100 mbar to 700 bar (1.5 to 10000 psi)
Differential Ranges
Wet/Dry
Uni-directional or bi-directional 70 mbar to 35 bar (1 to 500 psi) Wet/Wet Uni-directional or bi-directional 350 mbar to 35 bar (5 to 500 psi) Line pressure: 70 bar (1000 psi) maximum
Barometric Ranges
Barometric ranges are available with a minimum span of 350 mbar (5.1 psi)
Non-Zero Based Ranges
Non-zero based ranges are available. For non-zero based gauge ranges, please contact Druck to discuss your requirements.
Over Pressure
10 × FS for ranges up to 150 mbar (2 psi)
6 × FS for ranges up to 700 mbar (10 psi)
2 × FS for barometric ranges
4 × FS for all other ranges (up to 200 bar for ranges
≤70 bar and up to 1200 bar for ranges >70 bar)*
For differential versions the negative side must not exceed the positive side by more than:
6 × FS for ranges up to 150 mbar (2 psi)
4 × FS for ranges up to 700 mbar (10 psi)
2 × FS for all other ranges up to a maximum of 15
bar (200 psi)
Containment Pressure
Ranges up to 150 mbar (2 psi) gauge 10 x FS
Ranges up to 70 bar (1000 psi) gauge 6 x FS (200 bar (2900 psi) max)
Ranges up to 70 bar (1000 psi) absolute
200 bar (2900 psi)
Ranges above 70 bar (1000 psi)
1200 bar (17400 psi)*
Differential (-ve port) must not exceed positive port by more than 6 × FS (15 bar (200 psi) maximum.
* 600 bar (8700 psi) for pressure connector options PX, RA and RF
Electronics
Option
0 mV Passive 2.5 to 12 10 mV/V^ <2 at 10 V
1 mV Linearised 7 to 12 10 mV/V^ <3 2 mA 7 to 28** 4-20 mA <30 3 0 to 5 V 4-wire 7 to 16** 0 to 5 V <3 4 0 to 5 V 3-wire 7 to 16** 0 to 5 V* <3 5 Basic
6 0 to 10 V 4-wire 12 to 16** 0 to 10 V <3 7 0.5 V to 4.5 V
8 Configurable
9 Configurable
^ with a 10 V supply mV output sensors give 100 mV over the full scale pressure.
Output is ratiometric to the supply voltage
Output reduces pro-rata for pressure ranges below 350 mbar (5 psi)
*0 to 5 V 3-wire output is non true zero. At pressures below 1% of span the output will be fixed at approximately 50 mV
**32 V in non-hazardous area operation ~ Supply voltage is between [Maximum Output + 1 V] (7 V
minimum) to 16 V (32 V in non-hazardous area operation)
Description
Configurable (3-wire)
Ratiometric
(4-wire)
(3-wire)
Supply
voltage
(V)
See
below
5.0 ± 0.5 0.5 to 4.5 V <3
7 to 36 See
7 to 36 See
Output
See
below
below
below
Current
Consumption
(mA)
<3
See
below
See
below
Basic Configurable (Option 5), Configurable 4-Wire (Option 8), Configurable 3-Wire (Option 9)
Any pressure signal output configurations will be available, subject to the following limitations:
Output Specification
Minimum span: 4 V 2 V Maximum span: 10 V 20 V Maximum output
limit: Maximum zero
offset: Current
consumption:
Reverse output response:
Maximum operating temperature:
Output voltage range can be specified to a resolution of
0.1 V. The output will continue to respond to 110% FS. i.e. if a 0
to 10 V output is specified, the output will continue to increase proportionally to applied pressure until at least 11 V.
Option 5: Not true zero, the output will saturate at < 50 mV.
Options 8, 9: On startup <100 mA drawn for 10 ms typically.
Options 8, 9: Shunt calibration: not available with reverse output.
Basic Configurable (Option 5)
11 V ±10 V
Span / 2 ±Span
< 3 mA < 20 mA @ 7 Vdc
No Yes
+125˚C +80˚C
Configurable (Options 8, 9)
decreasing
to < 5 mA @ 32 Vdc
Page 3
Examples
Configuration Allowed Not Allowed
Basic Configurable (Option 5)
Configurable (Options 8, 9)
0 to 5 V 1 to 4 V (span too small)
0.5 to 4.5 V 4 to 11 V (offset too big)
1 to 6 V
1 to 11 V
-10 to 0 V 0 to 12 V (outside ±10 V limits)
0 to 5 V 6 to 10 V (offset too big)
-5 to 5 V 0 to 0.5 V (span too small)
-2 to 10 V
1 to 6 V
10 to 0 V
Power-Up Time
mV, Voltage and current versions: 10 ms
Configurable 3-wire and 4-wire versions: 500 ms
Insulation
500 Vdc: 100 MΩ
500 Vac: < 5 mA leakage current (mV and mA versions only)
Shunt Calibration
Shunt Calibration provides a customer accessible connection which, when applied, causes a shift in output of 80% FS in order to simulate applied pressure. It is fitted to the mV, Configurable 4-wire and Configurable 3-wire versions as standard. It is not available with DIN, M12 x 1 or M20 x 1.5 electrical connectors (options 7, D, G and R). Shunt calibration is activated in different ways depending on the electrical connector and version:
mV versions: connect Shunt Cal to -ve Supply or where available, connect both Shunt Cal connections together.
Configurable 4-wire and Configurable 3-wire versions: connect Shunt Cal to -ve Output or, where available, connect both Shunt Cal connections together.
Note: Not available with reverse output.
Performance Specifications
There are three grades of performance specification: Industrial, Improved and Premium.
Accuracy
Voltage, Current and mV Linearised
Combined effects of non-linearity, hysteresis and repeatability:
Industrial: ±0.2% FS BSL
Improved: ±0.1% FS BSL
Premium: ±0.04% FS BSL
mV Passive
70 bar Industrial/Improved: ±0.25% FS BSL Premium not available > 70 bar
Industrial/Improved: ±0.5% FS BSL Premium not available
Note 1: For the barometric pressure range, accuracy is of span, not full scale.
Note 2: For bi-directional ranges, accuracy is specified for each direction separately.
Zero Offset and Span Setting
Demountable electrical connector options allow access to potentiometers that give at least ±5% FS adjustment (see Electrical Connector section).
Factory Set To:
Product Description Industrial Improved
Current and Voltage Versions (Demountable Electrical Connections and Cable Gland)
Current and Voltage Versions (All Other Electrical Connections)
mV Versions ±3.0 mV ±3.0 mV
±0.5% FS ±0.2% FS
±1.0% FS ±1.0% FS
and Premium
Long Term Stability
±0.05% FS typical (±0.1% FS maximum) per year increasing pro-rata for pressure ranges below 350 mbar
Temperature Effects
Four compensated temperature ranges can be chosen. Industrial Accuracy performance:
-10 to +50°C (+14 to +122°F): ±0.75% FS TEB*
-20 to +80°C (-4 to +176°F): ±1.5% FS TEB
-40 to +80°C (-40 to +176°F): ±2.25% FS TEB
-40 to +125°C (-40 to +257°F):±2.25% FS TEB
*TEB = Temperature Error Band
Improved and Premium Accuracy performance:
-10 to +50°C (+14 to +122°F): ±0.5% FS TEB
-20 to +80°C (-4 to +176°F): ±1.0% FS TEB
-40 to +80°C (-40 to +176°F): ±1.5% FS TEB
-40 to +125°C (-40 to +257°F):±1.5% FS TEB
Temperature effects increase pro-rata for pressure ranges below 350 mbar (5 psi) and are doubled for barometric ranges.
Line Pressure Effects (Differential Version Only)
Zero shift: <±0.03% span/bar of line pressure Span shift: <±0.03% span/bar of line pressure Effects increase pro-rata for differential pressure ranges
below 700 mbar (10 psi).
Physical Specifications
Environmental Protection
See Electrical Connector section
Hyperbaric Pressure: 20 bar (300 psi) maximum
Operating Temperature Range
See Electrical Connector section
Pressure Media
Fluids compatible with stainless steel 316L and Hastelloy C276.
Page 4
For the wet/dry differential version, negative pressure port: Fluids compatible with stainless steel 316L, stainless steel 304, Pyrex, silicon and structural adhesive.
Enclosure Materials
Stainless steel (body), nitrile- or silicone-rubber (o-rings, gaskets), EPDM (gaskets), PVDF (depth cone), PTFE (vent filter), Nickel plated brass (lock rings), glass filled nylon (electrical connector assemblies), delrin (depth cone). Cable sheaths as specified (see Electrical Connector).
Pressure Connector
Available options are:
G1/4 Female*
G1/4 Male Flat
G1/4 Male 60° Internal Cone
G1/4 Male Flat Long
G1/4 Male Flat with Snubber
G1/4 Male Flat with Cross Bore Protection
G1/4 Male with Nipple
G1/4 Quick Connect
G1/8 Male 60° Internal Cone
G1/2 Male via Adaptor*
1/4 NPT Female*
1/4 NPT Male
1/8 NPT Male
1/8-27 NPT Female
1/2 NPT Male via Adaptor
7/16-20 UNJF Female*
7/16-20 UNF Male Short Flat*
7/16 UNF Long 37° Flare Tip*
7/16-20 UNJF Male 74° External Cone
3/8-24 UNJF
1/4 Swagelok Bulkhead
M10 X 1 80° Internal Cone
M12 X 1 60° Internal Cone
M14 X 1.5 60° Internal Cone
M20 X 1.5 Male (3 mm bore)
M20 X 1.5 Male (8 mm bore)
Depth Cone (G1/4 Female Open Face)
M12 x 1.0 74° External Cone
Quick Release Male
VCR Female*
VCR Male*
NW16 Flange
R3/8 Male
R1/4 Male
Choose connectors marked * for pressure ranges over 70 bar. Other pressure connectors may be available, contact Druck to discuss your requirement.
General Certifications
CRN Certified 0F13650.517890YTN ADD1/REV1, 0F13828.2 (sensor types K and O) and CSA 0F13650.56 ADD1 for pressure ranges up to and including 350 bar (5000 psi).
Electrical Connector
Various electrical connector options are available offering different features:
Code
Description Maximum Operating
No.
0 No Connector -55 to +125 -67 to +257 - Y 1 Cable Gland -40 to +80 -40 to +176 65 N 2 Raychem Cable -55 to +125 -67 to +257 65 N 3 Polyurethane
Depth 4 Hytrel Depth -40 to +80 -40 to +176 68 N 6/E Bayonet
MIL-C-26482 7 DIN 43650 Form
A Demountable A/F Bayonet
MIL-C-26482
Demountable C 1/2 NPT Conduit -40 to +80 -40 to +176 65 N D Micro DIN (9.4
mm pitch) G M12x1 4pin -55 to +125 -67 to +257 67 N K Zero Halogen
Cable
Demountable N Kynar Depth -40 to +80 -40 to +176 68 N R M20 x 1.5 Inline -40 to +80 -40 to +176 65 Y
Note 1: Electronics output options 8 and 9 are restricted to a maximum operating temperature of 80°C (176°F).
Note 2: Hazardous area approved versions are restricted to a maximum operating temperature range of -40°C to +80°C (-40°F to +176°F).
Note 3: Electrical connector option R IP65 rating only with suitable conduit/cable fitting.
Temperature Range °C °F
-40 to +80 -40 to +176 68 N
-55 to +125 -67 to +257 67 N
-40 to +80 -40 to +176 65 Y
-55 to +125 -67 to +257 65 Y
-40 to +80 -40 to +176 65 N
-40 to +80 -40 to +176 65 Y
IP Rating
Zero/ Span Adjust
CE Conformity
RoHS 2011/65/EU
Pressure Equipment Directive 2014/68/EU Sound Engineering Practice
ATEX 2014/34/EU (Optional)
EMC Directive 2014/30/EU
BS EN 61000-6-1: 2007: Susceptibility - Light Industrial
BS EN 61000-6-2: 2005: Susceptibility - Heavy Industrial*
BS EN 61000-6-3: 2007+A1:2011: Emissions - Light Industrial
BS EN 61000-6-4: 2007+A1:2011: Emissions - Heavy Industrial
BS EN 61326-1: 2013: Electrical Equipment for Measurement, Control and Laboratory Use
BS EN 61326-2-3: 2013: Particular Requirements for Pressure Transducers
*PMP/PTX models with electrical connector options 6, A, C, E, F, G, N and R with cable screen connected to case.
Hazardous Area Approvals (optional)
General applications:
IECEx/ATEX Intrinsically Safe ‘ia’ Group IIC
INMETRO Intrinsically Safe ‘ia’ Group IIC
NEPSI Intrinsically Safe ‘ia’ Group IIC
FM Approved (Canada & US) Intrinsically Safe Exia Class I, Division 1, Groups A, B, C & D and Class I, Zone 0 AEx/Ex ia Group IIC; Single Seal
Mining applications:
IECEx/ATEX Intrinsically Safe ‘ia’ Group I
INMETRO Intrinsically Safe ‘ia’ Group I
For full certification details, refer to the type-examination certificates (or approval listings) and supplied hazardous area installation instructions.
Page 5
Electrical Connector
Connector Type Option
code
Molex 0 1 Red +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
2 Yellow - +ve Output +ve Output +ve Output +ve Output +ve Output
3 Green - - -ve Output -ve Output 0V Common -ve Output
4 Blue -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
5 Orange - - - Shunt Cal Shunt Cal Shunt Cal
6 Black Case Case Case Case Case -
Cable (Not Raychem)
Raychem Cable 2 Red +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
Bayonet 6, A A +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
DIN A Micro DIN
Bayonet Alternative Wiring Options
M12 X 1 4-Pin
Zero Halogen Cable (Demountable)
M20 x 1.5 Female Demountable
1, 3, 4, C, N Red +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
Yellow - +ve Output +ve Output +ve Output +ve Output +ve Output
Blue - - -ve Output -ve Output 0V Common -ve Output
White -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
Orange - - - Shunt Cal Shunt Cal Shunt Cal
Black - - - - - -
Screen - - - - - -
White - +ve Output +ve Output +ve Output +ve Output +ve Output
Green - - -ve Output -ve Output 0V Common -ve Output
Blue -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
Black - - - Shunt Cal Shunt Cal Shunt Cal
Screen - - - - - -
B -ve Supply +ve Output +ve Output +ve Output +ve Output +ve Output
C - - -ve Output -ve Output 0V Common -ve Output
D - 0V Common -ve Supply -ve Supply 0V Common -ve Supply
E - - - Shunt Cal Shunt Cal Shunt Cal
F - - - - - Shunt Cal
7 D
E, F A +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
G 1 +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
K Pink +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
R +ve +ve Supply - - - - -
1 +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply +ve Supply
2 -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
3 - +ve Output +ve Output +ve Output +ve Output +ve Output
E Case Case -ve Output -ve Output 0V Common -ve Output
B - 0V Common -ve Supply -ve Supply 0V Common -ve Supply
C - +ve Output +ve Output +ve Output +ve Output +ve Output
D -ve Supply - -ve Output -ve Output 0V Common -ve Output
E - - - Shunt Cal Shunt Cal Shunt Cal
F - - - Shunt Cal Shunt Cal -
2 +ve Output +ve Output +ve Output +ve Output +ve Output
3 -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
4 Case Case -ve Output -ve Output 0V Common -ve Output
White - +ve Output +ve Output +ve Output +ve Output +ve Output
Green - - -ve Output -ve Output 0V Common -ve Output
Blue -ve Supply 0V Common -ve Supply -ve Supply 0V Common -ve Supply
Grey - - - Shunt Cal Shunt Cal Shunt Cal
Brown - - - - - -
Yellow - - - - - -
Screen - - - - - -
-ve -ve Supply - - - - -
Electronics Option
4 to 20 mA Voltage (3-
wire) and Basic Configurable
Voltage (4-wire)
Configurable Voltage (4-Wire)
Configurable Voltage (3-Wire)
mV
Page 6
Ordering Information (see the online configuration tool at Druck.com)
(1) Select model number
Main Product Variant PMP Amplified Pressure Transducer PDCR mV Pressure Transducer PTX 4-20 mA Pressure Transmitter Product Series 5 UNIK 5000 Diameter and Material 0 25mm Stainless Steel Electrical Connector (Note 6)
0 No Electrical Connector (Note 7) 1 Cable Gland (Polyurethane Cable) 2 Raychem Cable 3 Polyurethane Cable (Depth) 4 Hytrel Cable (Depth) 6 MIL-C-26482 (6-pin Shell Size 10) (Mating connector not supplied) 7 DIN 43650 Form A Demountable (Mating connector supplied) A Demountable MIL-C-26482 (6-pin Shell Size 10) (Mating connector not supplied) C 1/2” NPT Conduit (Polyurethane cable) D Micro DIN (9.4 mm Pitch) (Mating connector supplied) E MIL-C-26482 (6 pin Shell Size 10) Alternative Wiring (Mating connector not supplied) F Demountable MIL-C-26482 (6 pin Shell Size 10) Alternative Wiring (Mating connector not supplied) G M12 x 1 4-pin male (Mating connector not supplied) K Zero Halogen Cable Demountable N Kynar Cable (Depth) R M20 x 1.5 Inline Female Conduit Demountable (Note 8) Electronics Option 0 mV Passive 4-wire (PDCR) (Note 1) 1 mV Linearised 4-wire (PDCR) 2 4 to 20 mA 2-wire (PTX) 3 0 to 5 V 4-wire (PMP) 4 0 to 5 V 3-wire (PMP) 5 Basic Configurable 3-wire (PMP) 6 0 to 10 V 4-wire (PMP) 7 0.5 to 4.5 V Ratiometric 3-wire (PMP) (Note 5)
8 Configurable 4-wire (PMP) (Note 4, 5) 9 Configurable 3-wire (PMP) (Note 4, 5)
Compensated Temperature Range TA -10 to +50 °C (14 to +122 °F) TB -20 to +80 °C (-4 to +176 °F)
TC -40 to +80 °C (-40 to +176 °F) TD -40 to +125 °C (-40 to +257 °F) (Note 2, 5)
Accuracy A1 Industrial
A2 Improved A3 Premium
Calibration CA Zero/Span Data
CB Room Temperature CC Full Thermal
CR Room Temperature/Measured Reading Hazardous Area Approval (Note 6) H0 None H1 IECEx/ATEX Intrinsically Safe ‘ia’ Group IIC H2 IECEx/ATEX Intrinsically Safe ‘ia’ Group I H6 FM (C & US) Intrinsically Safe ‘ia’ Group IIC/ABCD HA IECEx/ATEX Intrinsically Safe ‘ia’ Groups I/IIC [H1 + H2] HS IECEx/ATEX/FM (C & US) Intrinsically Safe ‘ia’ Groups IIC/ABCD [H1 + H6] J1 IECEx/ATEX/NEPSI Intrinsically Safe ‘ia’ Group IIC JA INMETRO Intrinsically Safe ‘ia’ Group IIC JB INMETRO Intrinsically Safe ‘ia’ Group I JF INMETRO Intrinsically Safe ‘ia’ Group I/IIC [JA + JB] Pressure Connector PA G1/4 Female (Note 3) PB G1/4 Male Flat PC G1/4 Male 60° Internal Cone PD G1/8 Male 60° Internal Cone PE 1/4 NPT Female (Note 3) PF 1/4 NPT Male PG 1/8 NPT Male PH M20x1.5 Male (3 mm bore) PJ M14x1.5 60° Internal Cone PK M12x1 Internal Cone PL 7/16-20 UNJF Male 74° External Cone PN G1/2 Male via Adaptor (Note 3) PQ G1/4 Quick Connect PR 1/2 NPT Male via Adaptor (Note 3) PS 1/4 Swagelok Bulkhead PT G1/4 Male Flat Long PU 7/16-20 UNF Long 37° Flare Tip (Note 3) PV 7/16-20 UNJF Female (Note 3) PW Depth Cone (G1/4 Female Open Face) PX 7/16-20 UNF Male Short Flat (Note 3, 9) PY 3/8-24 UNJF PZ M10 x 1 80° Internal Cone RA VCR Female (Note 3, 9) RB G1/4 Male Flat with Snubber RC G1/4 Male Flat with Cross Bore Protection RD M12 x 1.0 74° External Cone RE Quick Release Mount RF VCR Male (Note 3, 9) RJ M20 x 1.5 Male (8 mm bore) RQ NW16 Flange RT 1/8-27 NPT Female RU R3/8 Male RV R1/4 Male RW G1/4 Male with Nipple
PTX 5 0 7 2 - TA - A2 - CB - H0 - PA (Example Model Number)
Page 7
Ordering Notes
Note 1: Premium Accuracy is not available on this version Note 2: Please ensure that the electrical connector
selected is option 0, 2, 6, A, E, F or G. Note 3: Select one of these pressure connectors for
pressure ranges over 70 bar
Note 4: Maximum operating temperature is 80°C (176°F) Note 5: Hazardous area certifications not available Note 6: Hazardous area certifications are restricted by
electrical connector options in line with the following table:
Connector
Approval 0 1 2 3 4 6/E 7 A/F C D G K N R
H0 Y Y Y Y Y Y Y Y Y Y Y Y Y Y
H1 Y Y Y Y Y Y Y Y Y Y Y - - Y H2 Y - Y Y Y Y - - Y - Y - - ­H6 Y Y Y Y Y Y Y Y Y Y Y - - Y HA Y - Y Y Y Y - - Y - Y - - ­HS Y Y Y Y Y Y Y Y Y Y Y - - Y
J1 Y Y Y Y Y Y Y Y Y Y Y - - Y JA Y Y Y Y Y Y Y Y Y Y Y - - Y JB Y - Y Y Y Y - - Y - Y - - -
JF Y - Y Y Y Y - - Y - Y - - -
Note 7: Available with component certification, use of which requires incorporation into certified apparatus with an IP rated enclosure appropriate to the certification type supplied
Note 8: Electronics option 2 only Note 9: Pressure ranges less than 500 bar
2) State pressure range and units
(e.g., 0 to 10 bar, -5 to +5 bar)
Pressure unit options are:
Symbol Description
bar bar mbar millibar psi pounds/sq. inch Pa Pascal hPa hectoPascal kPa kiloPascal MPa MegaPascal mmH2O mm water cmH2O cm water mH2O metres water inH2O inches water ftH2O feet water mmHg mm mercury inHg inches mercury kgf/cm2 kg force/sq. cm atm atmosphere Torr torr
5) Output options 5, 8 and 9
State voltage output at minimum and maximum pressure (e.g., output –1 to 9 V)
Typical order examples
PTX5012-TB-A2-CA-H0-PA: 0 to 10 bar, gauge, 3 m cable PMP5028-TC-A3-CC-H0-PE: -15 to 75 psi, gauge, 15 ft cable, output voltage -1 to 5 volts PDCR5071-TB-A1-CB-H0-PA: 0 to 100 bar, sealed gauge
Accessories (order as separate line items)
P/N 202-034-05: Sensor termination enclosure P/N 600-914: Desiccant gel pack for STE P/N 222-117-01: Short sink weight (to fit G1/4 female) P/N 222-179-01: Depth cone injection moulded for PW connection P/N 163-009: Mating connector for MIL-C-26482 (Electrical connector options 6, A, E, and F) (Note 10)
Cable Assemblies (Note 10)
A made up electrical connector with a length of cable.
(1) Select Part Number
Main Product (Note 10)
UNIKCABLE Cable assembly Electrical connector
6 MIL-C-26482 7 DIN 43650 D MICRO DIN (9.4 mm Pitch)
Cable 1 Polyurethane cable (Note 11)
2 Raychem cable
UNIKCABLE - 6 - 2 (Example Part Number)
(2) State cable length and units (Integer value only)
Minimum length 1 m (3 ft) Maximum length 200 m (600 ft)
Example: UNIKCABLE-6-2 5 m
Note 10: Not considered suitable for use in hazardous areas.
Note 11: This cable is the same as that used in the electrical connector cable gland (Option 1 in the main product).
3) State pressure reference (e.g., gauge)
Reference options are:
gauge
absolute
barometric
sealed gauge
wet/dry differential
wet/wet differential
4) State cable lengths and units
Integer values only, e.g. 1 m cable, 8 ft. Minimum length 1 m (3 ft) cable (only required with certain electrical connectors). Maximum cable length 100 m (300 ft) for approval options not H0; 200 m (600 ft) for approval option H0.
Page 8
CONDUIT DEMOUNTABLE
Mechanical Drawings
72
HIGH PRESSURE CONSTRUCTION
65 (2.56)
MEDIUM PRESSURE
CONSTRUCTION
LOW/MEDIUM PRESSURE CONSTRUCTION
Ø
25
(0.98)
(2.83)
S/N *******
*** TO *** ba
Supply: *** T
Output: *** T
Temp. Range: DRUCK LTD. GR
DRUCK LTD. GR
#### 50##-T*-
********** AC
S/N *******
*** TO *** ba
Supply: *** T
Output: *** T
Temp. Range:
69.5 (2.73)
S/N *******
*** TO *** ba
Supply: *** T
Output: *** T
Temp. Range: DRUCK LTD. GR
15.5 BOTH ENDS
(0.61)
DRUCK LTD. GR
Supply: *** T
Output: *** T
Temp. Range:
Ø
WET/WET & WET/DRY DIFFERENTIAL CONSTRUCTION
80
(3.15)
7 (0.27)
15 (0.53)
MALE PRESSURE CONNECTION
( A/F)
22
(0.86)
16
(0.62)
DEPTH CONE
PRESSURE ADAPTOR
TYP25
(0.98)
OPTIONAL
PRESSURE ADAPTOR
NOTES:
[1] DIMENSIONS SHOWN ARE FOR STANDARD LENGTH PRODUCTS WITH THE FOLLOWING ELECTRICAL OUTPUT OPTIONS: mV LINEARISED (PDCR) 4 TO 20 mA (PTX) STANDARD VOLTAGE OPTIONS (PMP)
BASIC CONFIGURABLE (PMP) FOR mV PASSIVE (PDCR) - SUBTRACT 10 mm (0.39 in)
FOR CONFIGURABLE 4-WIRE AND CONFIGURABLE 3-WIRE (PMP) - ADD 15 mm (0.59 in)
[2] REFER TO PAGE 4 FOR LIST OF PRESSURE CONNECTION OPTIONS (ORIENTATION NOT CRITICAL)
15.5 (0.61)
15.5 (0.61)
72
(2.83)
[3] ALL DIMENSIONS ARE IN MILLIMETRES (INCHES IN PARENTHESES)
[4] HIGH PRESSURE IS >70 BAR MEDIUM PRESSURE INDUSTRIAL ACCURACY >1 BAR TO < 50 BAR LOW/MEDIUM PRESSURE INDUSTRIAL ACCURACY < 1 BAR, > 50 BAR TO < 70 BAR IMPROVED/PREMIUM ACCURACY < 70 BAR
UNIK 5
PRESSU #### 50##
UNIK 5
PRESSU
UNIK 5
PRESSU #### 50##
********** AC
S/N *******
*** TO *** ba
15.5 (0.61)
Ø25 (0.98)
#### 50##-T*-
UNIK 5
PRESSU
M12x1 4-PIN
Ø
25 (0.98)
Ø
Ø
25
25 (0.98)
(0.98)
Ø
25 (0.98)
25 TYP (0.98)
4
(0.98)
Ø
25
CABLE GLAND WITH POLYURETHANE CABLE
7 (0.27)
Ø
25
(0.98)
Ø
25
(0.98)
35 (1.37)
DIN 43650 DEMOUNTABLE
42
(1.65)
32
(1.25)
Ø
25
Ø
25
(0.98)
(0.98)
CONDUIT WITH POLYURETHANE CABLE
A/F)(22
4 (0.15)
(0.98)
Ø
25
BAYONET MIL-C-26482
NON-DEMOUNTABLE
(1.92)
TYP (0.82)21
RAYCHEM CABLE
60 (2.36)
17 (0.66)
49
Ø
DEPTH CABLE
30
(1.18)
BAYONET MIL-C-26482
A/F
22
(0.86)
49
28 (1.10)
29 (1.14)
16 (0.62)
DEMOUNTABLE
(1.32)
39 (1.5)
25 (0.98)
Ø25 (0.98)
MICRO DIN (9.4 mm)
15.5 (0.6)
33.45 (1.3)
Ø
Druck.com
Copyright 2019. Baker Hughes Company. This material contains one or more registered trademark of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and
company names are trademarks of their respective holders.
920-483L
47 (1.85)
M20 x 1.5 INLINE FEMALE
Ø
44 (1.73)
16.25 (0.64)
Ø30 (1.18)
ZERO HALOGEN CABLE DEMOUNTABLE
Loading...