Druck UNIK 5000 Specifications

UNIK 5000
Pressure Sensing Platform
The new UNIK 5000 is a high performance configurable solution to pressure measurement. The use of micromachined silicon technology and analogue circuitry enables best in class performance for stability, low power and frequency response. The new platform enables you to easily build up your own sensor to match your own precise needs. This high performance, configurable solution to pressure measurement employs modular design and lean manufacturing techniques.
High Quality
The combination of a high technology sensor, together with advanced signal conditioning and packaging techniques, provides an ideal long term solution for reliable, accurate and economical measurements.
Bespoke as Standard
Custom-built from standard components, manufacturing sensors to your requirement is fast and simple; each UNIK 5000 is a “bespoke” pressure sensing solution, but with the short lead times and competitive pricing you would expect from standard products.
Expertise
We have the people and the knowledge to support your needs for accurate and reliable product performance; our team of experts can help you make the right sensor selection, guiding you and providing the help and tools you need. It is important to ensure that the sensor material and performance selected are suitable for your application.
Features
Ranges from 70 mbar (1 psi) to 700 bar (10000 psi)
Accuracy to ±0.04% Full Scale (FS) Best Straight Line (BSL)
Stainless Steel construction
Frequency response to 3.5 kHz
High over pressure capability
Hazardous Area certifications
mV, mA, voltage and configurable voltage outputs
Multiple electrical and pressure connector options
Operating temperature ranges from –55 to +125°C (-67 to +257°F)
Online configuration tool
Druck.com
UNIK 5000 Specifications
Supply and Outputs
Measurement Operating Pressure Ranges:
Gauge Ranges
Any zero based range 70 mbar to 70 bar (1 to 1000 psi)
Note: All psi values are approximate.
Sealed Gauge Ranges
Any zero based range 10 to 700 bar (145 to 10000 psi)
Absolute Ranges
Any zero based range 100 mbar to 700 bar (1.5 to 10000 psi)
Differential Ranges
Wet/Dry
Uni-directional or bi-directional 70 mbar to 35 bar (1 to 500 psi) Wet/Wet Uni-directional or bi-directional 350 mbar to 35 bar (5 to 500 psi) Line pressure: 70 bar (1000 psi) maximum
Barometric Ranges
Barometric ranges are available with a minimum span of 350 mbar (5.1 psi)
Non-Zero Based Ranges
Non-zero based ranges are available. For non-zero based gauge ranges, please contact Druck to discuss your requirements.
Over Pressure
10 × FS for ranges up to 150 mbar (2 psi)
6 × FS for ranges up to 700 mbar (10 psi)
2 × FS for barometric ranges
4 × FS for all other ranges (up to 200 bar for ranges
≤70 bar and up to 1200 bar for ranges >70 bar)*
For differential versions the negative side must not exceed the positive side by more than:
6 × FS for ranges up to 150 mbar (2 psi)
4 × FS for ranges up to 700 mbar (10 psi)
2 × FS for all other ranges up to a maximum of 15
bar (200 psi)
Containment Pressure
Ranges up to 150 mbar (2 psi) gauge 10 x FS
Ranges up to 70 bar (1000 psi) gauge 6 x FS (200 bar (2900 psi) max)
Ranges up to 70 bar (1000 psi) absolute
200 bar (2900 psi)
Ranges above 70 bar (1000 psi)
1200 bar (17400 psi)*
Differential (-ve port) must not exceed positive port by more than 6 × FS (15 bar (200 psi) maximum.
* 600 bar (8700 psi) for pressure connector options PX, RA and RF
Electronics
Option
0 mV Passive 2.5 to 12 10 mV/V^ <2 at 10 V
1 mV Linearised 7 to 12 10 mV/V^ <3 2 mA 7 to 28** 4-20 mA <30 3 0 to 5 V 4-wire 7 to 16** 0 to 5 V <3 4 0 to 5 V 3-wire 7 to 16** 0 to 5 V* <3 5 Basic
6 0 to 10 V 4-wire 12 to 16** 0 to 10 V <3 7 0.5 V to 4.5 V
8 Configurable
9 Configurable
^ with a 10 V supply mV output sensors give 100 mV over the full scale pressure.
Output is ratiometric to the supply voltage
Output reduces pro-rata for pressure ranges below 350 mbar (5 psi)
*0 to 5 V 3-wire output is non true zero. At pressures below 1% of span the output will be fixed at approximately 50 mV
**32 V in non-hazardous area operation ~ Supply voltage is between [Maximum Output + 1 V] (7 V
minimum) to 16 V (32 V in non-hazardous area operation)
Description
Configurable (3-wire)
Ratiometric
(4-wire)
(3-wire)
Supply
voltage
(V)
See
below
5.0 ± 0.5 0.5 to 4.5 V <3
7 to 36 See
7 to 36 See
Output
See
below
below
below
Current
Consumption
(mA)
<3
See
below
See
below
Basic Configurable (Option 5), Configurable 4-Wire (Option 8), Configurable 3-Wire (Option 9)
Any pressure signal output configurations will be available, subject to the following limitations:
Output Specification
Minimum span: 4 V 2 V Maximum span: 10 V 20 V Maximum output
limit: Maximum zero
offset: Current
consumption:
Reverse output response:
Maximum operating temperature:
Output voltage range can be specified to a resolution of
0.1 V. The output will continue to respond to 110% FS. i.e. if a 0
to 10 V output is specified, the output will continue to increase proportionally to applied pressure until at least 11 V.
Option 5: Not true zero, the output will saturate at < 50 mV.
Options 8, 9: On startup <100 mA drawn for 10 ms typically.
Options 8, 9: Shunt calibration: not available with reverse output.
Basic Configurable (Option 5)
11 V ±10 V
Span / 2 ±Span
< 3 mA < 20 mA @ 7 Vdc
No Yes
+125˚C +80˚C
Configurable (Options 8, 9)
decreasing
to < 5 mA @ 32 Vdc
Examples
Configuration Allowed Not Allowed
Basic Configurable (Option 5)
Configurable (Options 8, 9)
0 to 5 V 1 to 4 V (span too small)
0.5 to 4.5 V 4 to 11 V (offset too big)
1 to 6 V
1 to 11 V
-10 to 0 V 0 to 12 V (outside ±10 V limits)
0 to 5 V 6 to 10 V (offset too big)
-5 to 5 V 0 to 0.5 V (span too small)
-2 to 10 V
1 to 6 V
10 to 0 V
Power-Up Time
mV, Voltage and current versions: 10 ms
Configurable 3-wire and 4-wire versions: 500 ms
Insulation
500 Vdc: 100 MΩ
500 Vac: < 5 mA leakage current (mV and mA versions only)
Shunt Calibration
Shunt Calibration provides a customer accessible connection which, when applied, causes a shift in output of 80% FS in order to simulate applied pressure. It is fitted to the mV, Configurable 4-wire and Configurable 3-wire versions as standard. It is not available with DIN, M12 x 1 or M20 x 1.5 electrical connectors (options 7, D, G and R). Shunt calibration is activated in different ways depending on the electrical connector and version:
mV versions: connect Shunt Cal to -ve Supply or where available, connect both Shunt Cal connections together.
Configurable 4-wire and Configurable 3-wire versions: connect Shunt Cal to -ve Output or, where available, connect both Shunt Cal connections together.
Note: Not available with reverse output.
Performance Specifications
There are three grades of performance specification: Industrial, Improved and Premium.
Accuracy
Voltage, Current and mV Linearised
Combined effects of non-linearity, hysteresis and repeatability:
Industrial: ±0.2% FS BSL
Improved: ±0.1% FS BSL
Premium: ±0.04% FS BSL
mV Passive
70 bar Industrial/Improved: ±0.25% FS BSL Premium not available > 70 bar
Industrial/Improved: ±0.5% FS BSL Premium not available
Note 1: For the barometric pressure range, accuracy is of span, not full scale.
Note 2: For bi-directional ranges, accuracy is specified for each direction separately.
Zero Offset and Span Setting
Demountable electrical connector options allow access to potentiometers that give at least ±5% FS adjustment (see Electrical Connector section).
Factory Set To:
Product Description Industrial Improved
Current and Voltage Versions (Demountable Electrical Connections and Cable Gland)
Current and Voltage Versions (All Other Electrical Connections)
mV Versions ±3.0 mV ±3.0 mV
±0.5% FS ±0.2% FS
±1.0% FS ±1.0% FS
and Premium
Long Term Stability
±0.05% FS typical (±0.1% FS maximum) per year increasing pro-rata for pressure ranges below 350 mbar
Temperature Effects
Four compensated temperature ranges can be chosen. Industrial Accuracy performance:
-10 to +50°C (+14 to +122°F): ±0.75% FS TEB*
-20 to +80°C (-4 to +176°F): ±1.5% FS TEB
-40 to +80°C (-40 to +176°F): ±2.25% FS TEB
-40 to +125°C (-40 to +257°F):±2.25% FS TEB
*TEB = Temperature Error Band
Improved and Premium Accuracy performance:
-10 to +50°C (+14 to +122°F): ±0.5% FS TEB
-20 to +80°C (-4 to +176°F): ±1.0% FS TEB
-40 to +80°C (-40 to +176°F): ±1.5% FS TEB
-40 to +125°C (-40 to +257°F):±1.5% FS TEB
Temperature effects increase pro-rata for pressure ranges below 350 mbar (5 psi) and are doubled for barometric ranges.
Line Pressure Effects (Differential Version Only)
Zero shift: <±0.03% span/bar of line pressure Span shift: <±0.03% span/bar of line pressure Effects increase pro-rata for differential pressure ranges
below 700 mbar (10 psi).
Physical Specifications
Environmental Protection
See Electrical Connector section
Hyperbaric Pressure: 20 bar (300 psi) maximum
Operating Temperature Range
See Electrical Connector section
Pressure Media
Fluids compatible with stainless steel 316L and Hastelloy C276.