Dr.Mach Mach Quintaflex User manual

Quintaflex Ceiling Model
DIRECTIONS FOR USE
Dr. Mach
Lamps and Engineering
Single Ceiling Lamp_____________________________ Order No. 5501 3031 27 Order No. 5501 3031 30 Ceiling Lamps - Combinations
Dr. Mach
59250002 Edition 02 18.06.2002 / Bak Page 1/20
GmbH u. Co., Flossmannstrasse 28, D-85560 Ebersberg
Tel.: +49 (0)8092 2093 0, Fax +49 (0)8092 2093 50
Internet: www.dr-mach.com, E-mail: info@dr-mach.de
Quintaflex Ceiling Model
Dr. Mach
Lamps and Engineering
List of contents
1. Safety instructions ..................................................................................Page 3
2. Operating the lamp Quintaflex................................................................Page 4
2.1 Checking the lamp Quintaflex with light intensity control ..................Page 4
2.2 ON/OFF switch.................................................................................Page 6
2.3 Positioning........................................................................................Page 6
2.4 Light field adjustment (merging of light fields) ..................................Page 6
3. Cleaning .................................................................................................Page 7
3.1 Sterilizable handle ............................................................................Page 7
3.2 Lamp head, splinter protection disk ..................................................Page 7
4. Maintenance...........................................................................................Page 8
4.1 Adjustments at the ceiling / wall attachment.....................................Page 8
4.2 Adjustments at the lamp head ..........................................................Page 10
4.3 Changing of spare parts ...................................................................Page 11
4.3.1 Changing the halogen bulbs....................................................Page 11
4.3.2 Changing the fuses .................................................................Page 11
4.3.3 Changing the filter disk............................................................Page 12
4.3.4 Changing the splinter protection disk ......................................Page 12
4.3.5 Changing the handle coupling.................................................Page 13
5. Data........................................................................................................Page 13
5.1 Technical data ..................................................................................Page 13
5.2 Environmental conditions .................................................................Page 13
6. Marking...................................................................................................Page 14
6.1 Specification of bulb .........................................................................Page 14
6.2 CE-mark ...........................................................................................Page 14
7. Disposal..................................................................................................Page 14
8. Spare parts.............................................................................................Page 15
9. Spare part list .........................................................................................Page 16
59250002 Edition 02 18.06.2002 / Bak Page 2/20
Quintaflex Ceiling Model
Dear customer!
Please read the safety instructions and product description carefully before working with
these lights for the first time.
Please follow the separate mounting instructions for ceiling- and wall-mounted lamps, and for fitting the stand foot and pipe.
Dr. Mach
Lamps and Engineering
1. Safety instructions
Please pay attention to the directions for use when handling the lamp.
Attention:
This device is not suitable for use in hazardous locations.
The lamp is classified as a Group 1 device according to the regulations for EEMP.
Repairs to the lamp and special installation work on the reflector or plug-in socket should only be carried out by ourselves or a company expressly authorised by our­selves.
The manufacturer is only responsible for the safety of the lamp if repairs and altera­tions have been carried out by themselves or a company who can guarantee that the safety regulations have been observed.
The manufacturer is not liable for personal or material damages if the lamp is misap­propriately or incorrectly operated or misused.
The lamp body may only be dismantled from the swivel-arm (in reverse order to its as­sembly) after the assembly locking device has been inserted and tightened since the arm is under spring tension.
Make sure that the lamp is in perfect working order before use.
59250002 Edition 02 18.06.2002 / Bak Page 3/20
Quintaflex Ceiling Model
Dr. Mach
Lamps and Engineering
2. Operating the lamp Quintaflex
2.1 Checking the lamp Quintaflex with light intensity control
The lamp Quintaflex is optionally equipped with the function of electronic light intensity control. You can adjust the light intensity at the key pad on the cardan bow.
Voltage measurement and setting
Before using the OT-lamps, a voltage measurement and, if necessary, a voltage setting has to be done. Only in this way a perfect functioning can be ensured. If the location of the lamp is changed, the voltage measurement and setting must be repeated.
Voltage measurement at the halogen bulbs
Screw
To measure the voltage proceed as follows: Mount the lamp to the ceiling or wall attachment. Switch on the lamp. In order to make the measurement, remove the cover. The cover is fixed with two screws, as shown in the figure. Loosen the two screws, remove them and take off the cover.
Screw
Measuring points
Measuring points
Voltage setting at the halogen bulbs
The measuring points for checking the voltage are in the same position at all Quintaflex-lamps. The measuring points are shown in the figure.
At measuring points 1 + 2 measure the voltage applied at the halogen bulbs. The lamp is set to the maximum light intensity.
The AC or DC voltage depends on the external power supply. Rated voltage 22,8 – 23,8V.
If the voltage measured is too high or too low, the value must be adjusted at the power supply. The setting at the transformer is made by recon­necting the wires on the secondary side.
Measuring points 1+2
59250002 Edition 02 18.06.2002 / Bak Page 4/20
Quintaflex Ceiling Model
Dr. Mach
Lamps and Engineering
Voltage measurement at the control board
Please check, whether the power supply is supplying AC or DC.

Voltage measurement at DC

Pay attention to the polarisation during the installation. If the light intensity control does not function as desired, the PLUS- and MINUS-pole at the power supply have to be changed.

Voltage measurement at AC

For installation at AC proceed as described below:
Preparation
The lamp is already switched on. Set the lamp to the lowest light intensity
Measurement
Measure the voltage at measuring points 3 + 4. The voltage must be less than 40V DC. If this value is exceeded, reset it as described be­low:
.
Measuring points
Voltage setting at the control board
Control board
J4 J5
Switch the lamp off. The position of the control board is shown in the figure. Adjust the standard setting by changing the jump­ers (see arrow). Three settings enable you to ideally set your lamp system (J4 and J5).
J5 J5 J5
J4 J4 J4
J5
J4
Take off one jumper and put it into position as shown in figure II. or III.
Switch on the lamp. Set the lamp to the lowest light intensity. Repeat the voltage measurement at measuring points 3 + 4. If the voltage is less than 40V DC, keep the chosen setting.
I. II. III.
If you measure a voltage in excess of 40V DC, switch off the lamp. Replace the first jumper to the initial position. Take off the second jumper and place it back on one of the two pins.
Switch on the lamp again. Finally repeat the voltage measurement. The voltage is now adjusted. Close the lamp by replacing the cover.
59250002 Edition 02 18.06.2002 / Bak Page 5/20
Quintaflex Ceiling Model
Dr. Mach
Lamps and Engineering
2.2 ON/OFF switch
There are no ON/OFF switches on the lamp head.
The customer must provide a two-pole ON/OFF
switch.
2.3 Positioning
For positioning use sterilizable handle 1. Handle rail 2 is used for adjusting the lights from the outside.
The cardan bow 3 is used for convenient, fully car­danic adjustment of the lamp.
2
3
In case of low room heights the lamp is delivered with a central spring arm. The spring arm has only one joint.
The coupling for the bow of the lamp body is fitted with a needle bearing. The inner ring 4 of this nee­dle bearing has already been pushed on the axis of the lamp bow.
Remark:
The instructions mentioned above refer to ceiling and wall models.
4 1
2.4 Light field adjustment (merging of light fields)
The lamps Quintaflex are equipped with the func-
1
tion of light field realignment. That means, you can merge the separated light fields to one spot as shown in the figure.
To activate the light field realignment turn steriliz­able handle 1.
59250002 Edition 02 18.06.2002 / Bak Page 6/20