3. Setup and Calibration Requirements .............................................................................................. 4
TO RUN THE SELF TEST............................................................................................................... 4
4. Matching EZ scale to another scale ................................................................................................. 5
SCALE MATCHING EXAMPLE ..................................................................................................... 5
SCALE INFORMATION SHEET .................................................................................................... 5
SCALE MATCHING WORK SHEET .............................................................................................. 5
CONNECTING EZ INDICATOR TO OTHER LOAD CELLS ....................................................... 6
TO CHANGE THE SETUP & CALIBRATION NUMBERS .......................................................... 6
TO RETURN TO WEIGH MODE .................................................................................................... 7
5. Getting started .................................................................................................................................. 8
6. System operation............................................................................................................................... 9
TURNING ON THE SCALE ............................................................................................................9
TURNING OFF THE SCALE .......................................................................................................... 9
TO ZERO BALANCE THE SCALE ................................................................................................. 9
USING THE HELP KEY ................................................................................................................ 10
TO SELECT GROSS MODE .......................................................................................................... 10
TO SELECT NET MODE ............................................................................................................... 10
TO SELECT HOLD MODE............................................................................................................ 11
TO EXIT HOLD MODE.................................................................................................................. 11
TO CANCEL HOLD MODE ........................................................................................................... 11
TO ENTER A PRESET ................................................................................................................... 12
TO CLEAR THE PRESET ALARM ............................................................................................... 13
TO CLEAR THE PRESET ALARM USING THE TR OPTION................................................... 13
TO PRELOAD A TARE VALUE .................................................................................................... 14
TO USE THE PRE-ALARM ........................................................................................................... 15
TO CHANGE THE PRE-ALARM VALUE .................................................................................... 15
TO START THE MIX TIMER ........................................................................................................ 16
TO CLEAR THE MIX TIMER........................................................................................................ 16
TO RESTART THE MIX TIMER ................................................................................................... 16
TO START THE ROTATION COUNTER ..................................................................................... 17
TO CLEAR THE ROTATION COUNTER..................................................................................... 17
TO RESTART THE ROTATION COUNTER ................................................................................ 17
All rights reserved. Reproduction of any part of this manual in any form whatsoever without Digi-Star’s express written permission is
forbidden. The contents of this manual are subject to change without notice. All efforts have been made to assure the accuracy of the
contents of this manual. However, should any errors be detected, Digi-Star would greatly appreciate being informed of them. The above
notwithstanding, Digi-Star can assume no responsibility for any errors in this manual or their consequence.
Copyright ! 2001 Digi-Star.
Fort Atkinson, Wisconsin (USA)
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
1. About the product
1. About the product
FEATURES OF MODELS EZ3200, EZ3200V AND EZ3200V RC
"
Three recipe programming modes
- Ingredient percentage of total
- Amount per animal
- Amount per ingredient
"
Accumulation for tracking ingredient usage
"
200 Recipe memory locations
"
99 Programmable ingredient names to simplify loading & unloading
"
Ingredients can be entered in any order
"
Scrolling Help messages for easy recipe programming & operation
"
Front panel calibration without simulator or weights
"
[Select] and [Function] keys to simplify appearance and allow for future expansion
"
A Hold feature to hold the weight stable while moving the scale system
"
Large 1.7" display for greater readability (1" for Model 3200)
Display EZ3200:6-digit alpha numeric LCD, fiberoptic back lighting
Display height EZ3200 = 1" (25.4 mm)
Display height EZ3200V and 3200V RC = 1.7" (43.2 mm)
"
Environmental enclosure: IP65, IEC529
"
Connectors:AMP, gold plated contacts
copyright - 10/24/2001
!
1
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
2. Mounting and connection
2. Mounting and connection
INDICATOR MOUNTING
The indicator is easily attached to the Indicator Mounting Bracket by hooking the top over the
plate and securing the bottom with two screws and nuts (size# 10 - 24 × 5/8" or M5 × 16mm).
POWER CONNECTION
Warning!
Always disconnect the indicator power cord before “jump starting” or fast charging a battery.
Disconnect all indicator leads before welding on equipment. Failure to do so can cause surges
which will damage the scale.
The power cable should be connected directly to a vehicle battery or regulated power supply. The
scale end of the power cable is attached to the J901 connector located on the bottom panel of the
scale.
Connect the RED wire from the power cable to +12VDC and the BLACK wire to GROUND. The
indicator is fused internally at 4 amps.
If a remote 12 Vdc alarm is to be used, connect
the +12Vdc side of the alarm to the power cable
ORANGE wire and the GROUND side of the
alarm to the frame (= ground).
The alarm output is fused for a maximum drain of 10 amps. The remote alarm connection
!
(Optional)
Serial
Port
Cable
Remote
Display
Cable
(Optional)
Junction
Box
Cable
Power
Cable
may also be used for motor control purposes when used with a relay.
REMOTE INPUT CONNECTION
If the remote input is to be used, connect one side of the normally open momentary switch or
relay contact to the power cable BLUE wire, and the other side to the frame or other GROUND
connection. If your power cable does not contain a blue wire and you desire to use this feature,
contact your dealer for a special cable. A process control box is available for motor control and
remote enter preset capability.
2
copyright - 10/24/2001
!
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
LOAD CELL CONNECTION
The indicator operates with strain gage load cells.
The system is normally supplied with a J-BOX cable
going between the scale indicator and the load cell
junction box. Extension Kits are available from your
dealer in various lengths.
To connect the load cells, attach the junction box
cable to the J902 connector on the bottom panel of
the scale. Connect the load cell cables to the junction box as shown left.
Follow color key on circuit board to insure proper
connection of load cell wires.
to J902 connector of indicator
LIGHTNING PROTECTION
Additional protection can be achieved with the proper installation of grounding rods. Please call
(920) 563-1400 to request Digi-Star Form F3050.
TECHNICAL MANUAL
Technical Manual available upon request. Please call (920) 563-1400 to request Digi-Star Manual
for New EZ3200.
copyright - 10/24/2001
!
3
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
3. Setup and Calibration Requirements
3. Setup and Calibration Requirements
Warning!
This indicator was calibrated at the factory to weigh accurately with your system.
Additional calibration is not necessary under normal conditions.
The Short Form Setup & Calibration procedure allows you to change the SETUP and CAL
numbers of the indicator. You may want to perform this procedure if:
1. the indicator is being connected to different load cells, or
2. you want to adjust the calibration to match another scale system (chapter 4).
Before continuing, first write down the current SETUP and CAL numbers of your EZ indicator.
These numbers are displayed during the Self Test.
TO RUN THE SELF TEST
1. With the indicator already ON, press
2. Press
3. Press
to "pause" the Self Test while numbers are displayed.
again to "resume".
to start the Self Test.
SETUP #CAL #
Do not attempt to calibrate the scale if the indicator is not reading stable weights. The
!
calibration procedure will not fix instability, inconsistencies, or flashing "RANGE"
messages.
copyright - 10/24/2001
!
4
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
4. Matching EZ scale to another scale
4. Matching EZ scale to another scale
Sometimes two different scales are used to weigh the same load. When this is done, the weight
measured by each scale may not be the same. This can be caused by one or both of the two scales
being slightly out of calibration. This indicator has the ability to match any other scale, even if
that scale is not calibrated.
To match your EZ scale (Scale A) to another scale (Scale B) you must determine the Calibration
Multiplier. To do this, place a load on Scale A (feed wagon, etc...) and write down the weight
displayed. Repeat several times to determine the average weight. Next, place the same load on
Scale B and again write down the weight displayed.
Repeat several times to determine the average weight. Use the following formula to determine
the Calibration Multiplier for the EZ’s "CAL" number:
!
It is important to use an average of several weights before calibrating the scale.
You will need the number and type of load cells
used in the new scale system. You will also need
the current "SETUP" and "CAL" as described
above. Once you have written down this
information, contact your nearest Scale Service
Center for new "SETUP" and "CAL" numbers.
Follow the instructions “To Change the Setup /
Calibration Numbers” shown below.
Press and hold the
key, to enter Short Form Setup & Calibration.
The first message displayed is SETUP.
Next, the actual SETUP number is displayed.
!
Press the key for additional help
information during Setup and Calibration.
If the correct SETUP number is displayed, press
the
key to advance to the CAL number.
key, then press the
127062
CAL
1. Press the
to count upward.
2. Press the
flashing.
1
When the correct SETUP number is displayed,
2
press the
This displays the CAL message,
followed by the CAL number.
!
The CAL number is not a weight. It is a
reference value the indicator uses to determine the
weight. This number directly affects the accuracy
of the scale system.
Change the CAL number using the same method
described in Steps 1& 2. When the display shows
the correct number, press the
key to cause the “flashing” digit
key to select which digit is
key to advance to the CAL number.
key.
copyright - 10/24/2001
!
This causes the number to be stored permanently
in the indicator and returns the indicator to the
weighing mode.
6
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
4. Matching EZ scale to another scale
TO RETURN TO WEIGH MODE
1
To exit setup without changing any values, press
and hold the
key, then press the
key.
and
2
copyright - 10/24/2001
!
7
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
5. Getting started
To operate the scale, first attach the scale's power cord to connector J901 and the loadcell cable to
connector J902 on the bottom panel of the scale.
5. Getting started
Press the
key. A brief message is displayed (such as
HELLO
The scale enters the GROSS mode.
HELLO
).
A warm up period of ten to fifteen minutes provides the most accurate readings. If the scale is
holding a load for a long period of time (ex. overnight), the weight displayed may vary because of
zero shift created by changes in temperature. This does not affect the accuracy of the scale.
For example, if the system was loaded with 1000lbs, it might read 1200lbs the following day. The
change in temperature "zero shifted" the ZERO/BALANCE from 0 to 200lbs. When unloading the
scale, the display will count from 1200 to 200lbs for a total of 1000lbs.
After this warm up period, press and release
The word
is displayed to show completion of the ZERO/BALANCE step.
ZERO
. Then, within three seconds, press
.
Now the scale is ready to weigh!
copyright - 10/24/2001
!
8
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
6. System operation
EZ3200 V shown - EZ3200 and ES3200V RC operations are the same.
!
LB and KG annunciators are located along right hand edge of the 3200V display.
!
6. System operation
TURNING ON THE SCALE
HELLO
TURNING OFF THE SCALE
Press
A brief message will be displayed (such as
The scale then enters the GROSS weighing mode.
GROSS mode displays the weight change since the
unit was last ZERO/BALANCED.
Pressing
operation starts the self test.
Press .
.
a second time during normal system
HELLO
).
TO ZERO BALANCE THE SCALE
and
zero
If the supply power is below the low battery threshold (10.5 Volts), the message
LOW BATTERY VOLTAGE
(approx. every five seconds) to alert the operator of the low battery condition.
Loss of power does not affect the Zero/Balance or Setup/Calibration values.
copyright - 10/24/2001
!
is displayed. The message
Pressand then within three seconds, press.
The [Zero] key will "balance off" empty trailer, bin,
or platform weight.
The message
the step and the scale is placed in the GROSS
mode.
Pressing only
ZERO/BALANCE PRESS NET/ GROSS - THEN ZERO
displayed.
will be periodically shown on the display
LO BAT
9
is displayed to show completion of
ZERO
will cause the message:
TO
to be
CANNOT balance --
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
USING THE HELP KEY
provides additional information about the weighing modes, setup/calibration, and recipe
programming.
6. System operation
Pressing
TO SELECT GROSS MODE
TO SELECT NET MODE
while displaying weight will display information about the last key pressed.
9850
%
%
0
!
Gross Mode displays the weight change since
the unit was last Zero/Balanced.
1. Press
!
The scale is in Gross Mode if there is a flashing
arrow (% ) pointing toward the word Gross, next to
the display.
Net Mode displays the weight change after a
Tare has been performed. Tare creates a
temporary zero at that weight value.
1. Press
enter the Net Mode.
or
.
to set a temporary "zero" point and
2. If in Gross Mode, press
or
3. If in Load/Unload Mode, press
If the Tare Function has not been previously performed, the unit will stay in the Gross Mode
!
and the message
The scale is in Net Mode if there is a flashing arrow (% ) pointing toward the word Net, just
above the [Hold] and [Net/Gross] keys.
2/31
FOR NET MODE PRESS TARE
two times to place the scale in Net Mode.
will scroll across the display.
or
key is an alternating action key. If the scale is
in the Gross Mode, pressing the [Net/Gross]
key will place it in the Net Mode. If the scale
is in the Net Mode, pressing the [Net/Gross]
key will place it in the Gross Mode.
.The [Net/Gross]
copyright - 10/24/2001
!
10
Digi-Star Model EZ3200/EZ3200V/EZ3200V RC
TO SELECT HOLD MODEHold Mode prevents the displayed weight
from changing due to “zero shift” while
moving the scale.
6. System operation
hold
TO EXIT HOLD MODE
9850
%
TO CANCEL HOLD MODE
!
Press
the Hold Mode.
!
The scale is in Hold Mode if the word
flashing on the display and the flashing
is only displayed for a brief time.
Press .
At this time the scale reactivates and adjusts the
Zero/Balance to maintain the gross weight
displayed. Small changes in weight can occur while
moving the scale system to new locations for
loading or unloading. This change is called "zero
shift" and is due to several factors including terrain
changes and mechanical stresses.
Press .
to "hold" the displayed weight and enter
is
HOLD
hold weight
9850
%
!
Cancelling the Hold Mode prevents the scale from
adjusting the Zero/Balance and returns the system
to the normal weighing mode.
Use this if weight is accidentally added while scale
is still in Hold Mode.
copyright - 10/24/2001
!
11
Loading...
+ 31 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.