INSPIRED INNOVATION
u
LCC and LCD Stackable
Clean Process Ovens
Save valuable fl oor space
Stainless steel interior and exterior
e Despatch LCC/LCD stackable clean process ovens are designed to save
valuable oor space and provide a variety of tailored options for your speci c
needs. Typical applications for these ovens include die-bond curing and other
semiconductor packaging processes, depyrogenation, sterilization and drying
for life sciences. e oven is con gured for 220/240 volts and 50/60 hz.
For production environment processes where minimal contamination is
essential, the LCC and LCD Series ovens o er the highest standards in HEPA
ltration. Re-circulated air ow is 100% HEPA (High E ciency Particulate
Air) ltered for operation at ISO Class 5 (Class 100) or better within the oven
chamber. is oven can also be ordered without HEPA ltration.
FEATURES AT A GLANCE
u
Available in air or nitrogen
atmosphere confi gurations (100ppm
oxygen capability). Forced exhaust
is standard on all models for rapid
cooling. Programmable control of the
nitrogen and cooling water is standard
on the nitrogen atmosphere units.
u
Lockable disconnect switch on the
control panel for easy servicing.
u
Protocol 3
display, integrated data logging
capabilities and USB port for simple
oven set-up and data export.
u
Stainless steel interior and exterior
with all interior seams continuously
welded on the insulation side to
protect the work chamber from
contamination.
u
Programmable door lock with
electronic release prevents operators
from opening oven door when cycle
is in process.
TM
control with large LCD
u
Stackable design allows for
independent batch processing.
u
HEPA fi ltration– Recirculated airfl ow
is 100% HEPA fi ltered for operation
at ISO Class 5 (Class 100) throughout
the cycle.
Magnehelic™ gauge monitors the
HEPA fi lter pressure drop so you know
when it is time to replace the fi lter.
u
Modbus communications connection
(RS485) for remote monitoring and
recording.
u
End of cycle and high limit audible
and visual alarm, that can be seen
and heard from a distance.
u
UL & C-UL listed open control panel
MODEL OPTIONS
u
Two sizes: 1.6 or 5.1 cu ft
(45 or 144 liter)
u
Maximum temperature of
260˚C (500˚F) with LCC
350˚C (662˚F) with LCD
u
With HEPA fi ltration
(Can be ordered without)
u
Air or nitrogen atmospheres
MinneaPolis • shanGhai • berlin • sinGaPore • hsinchu

LCC/LCD 1-16 1-51
Physical sPecifications
Chamber size (width x depth x height)
* Clear opening width is reduced by 1.5” (3.8 cm)
due to 3/4” (1.9 cm) shelf supports on each side.
Capacity in cubic feet (liters) 1.6 (45) 5.1 (144)
Overall size (width x depth x height) 32.5 x 35.5 x 20.75 in.
Electrical single phase 50/60 HZ
* Heater capacity based on 240 volt supply. Operating
at 208V will result in 25% reduction in heater output.
Number of shelves provided
Maximum number of shelves 5 on 2” (5 cm) centers 8 on 2” (5 cm) centers
Approximate net weight 250 lbs (114 kg) 380 lbs (172 kg)
Approximate shipping weight 350 lbs (159 kg) 525 lbs (238 kg)
FUNCTIONAL SPECIFICATIONS
Time to temperature with no load (40˚C to 100˚C) 7 minutes 5 minutes
Time to temperature with no load (40˚C to 200˚C) 30 minutes 27 minutes
Time to temperature with no load (40˚C to 260˚C) 45 minutes 35 minutes
Time to temperature with no load (40˚C to 350˚C)
*For LCD and LLD only, LCC and LLC max temp is 260˚C
Cooling time to teperature with no load (100˚C to 55˚C)
*Based on cooling water supplied at 2 GPM (7.6lpm), 16˚C
Cooling time to teperature with no load (200˚C to 55˚C)
*Based on cooling water supplied at 2 GPM (7.6lpm), 16˚C
Cooling time to teperature with no load (260˚C to 55˚C)
*Based on cooling water supplied at 2 GPM (7.6lpm), 16˚C
Cooling time to temperature with no load (350˚C to 55˚C)
(For LCD only, LCC max temp is 260˚C)
*Based on cooling water supplied at 2 GPM (7.6 lpm), 16˚C
Temperature uniformity at 100˚C +/- 1˚C +/- 1˚C
Temperature uniformity at 200˚C +/- 2˚C +/- 2˚C
Temperature uniformity at 260˚C +/- 3˚C +/- 3˚C
Temperature uniformity at 350˚C (For LCD only) +/- 4˚C +/- 4˚C
Control stability +/- 0.5˚C +/- 0.5˚C
LCC operating range with 20˚C ambient air atmosphere (nitrogen atmosphere).
LCD operating range with 20˚C ambient -
air atmosphere (nitrogen atmosphere)
Maximum load capacity 200 lbs (90 kg) 200 lbs (90 kg)
Maximum shelf capacity 50 lbs (23 kg) 25 lbs (11 kg)
1/4 HP recirculating fan 240 CFM (113 lps) 435 CFM (205 lps)
15* x 14 x 14 inches
38* x 36 x 36 cm
83 x 90 x 53 cm
220-240 volts*
Heater: 3 kW
2 2
60 minutes 50 minutes
35 minutes (air)
30 minutes (nitrogen)*
65 minutes (air)
55 minutes (nitrogen)*
75 minutes (air)
60 minutes (nitrogen)*
90 minutes (air)
80 minutes (nitrogen)*
40˚C-260˚C (35˚C-260˚C) 45˚C-260˚C (35˚C-260˚C)
40˚C-350˚C (35˚C-350˚C) 45˚C-350˚C (35˚C-350˚C)
23* x 20 x 20 inches
58* x 51 x 51 cm
40.5 x 42.5 x 27 in.
103 x 108 x 68 cm
220-240 volts*
Heater: 6 kW
40 minutes (air)
25 minutes (nitrogen)*
75 minutes (air)
40 minutes (nitrogen)*
85 minutes (air)
45 minutes (nitrogen)*
100 minutes (air)
50 minutes (nitrogen)*
OPTIONS
u
Silicone-free construction
u
Non-HEPA filtration model
u
Nitrogen atmosphere
u
Door interlock switch turns off heater and
fan when door is opened
u
Cleaning and triple bagging in clean room
prior to shipment
u
Chart recorders
u
Three color process stack light
u
Skirted stands with leveling legs
u
Extra shelves
u
Oxygen concentration monitor system
u
Data acquisition software
u
Customization available
u
CE compliance
Warning: Despatch Stackable Clean Room
Ovens are not to be used with flammable
solvents, combustible materials or enclosed
containers. If your process involves flammable
solvents, consult factory about modifications.
Notes: Uniformity figures are based on a nine-point test conducted in an empty oven after stabilization period.
Uniformity can vary slightly depending on unit and operating conditions. Class 100 HEPA filtration will limit ramp rates.
Minimum operating temperature and cooling times are based on 20°C ambient temperature measured at the fresh air inlet
Specifications are subject to change without notice. If the existing specifications differ from yours, ask about our customizing capabilities.
service and technical suPPort
service parts: 1-800-473-7373
international service/main: 1-952-469-8230
service fax: 1-952-469-8193
service@despatch.com
Global headquarters
phone: 1-888-DESPATCH (1-888-337-7282)
international/main: 1-952-469-5424
fax: 1-952-469-4513
sales@despatch.com
8860 207th Street West
Minneapolis, MN 55044 USA
www.despatch.com
© 2012 Despatch Industries. All rights reserved. Despatch is a registered trademark of Despatch Industries in the U.S. and other countries. LCC-LCD 4-12