Condec DLR334 Operating Manual

DLR3110
Rack-mountable Pneumatic
Pressure Calibration Console
• Operation
• Calibration
• Maintenance
• Service
CONDEC Sales Phone No.: (888) 295-8475
CONDEC Web Site: WWW.4CONDEC.COM
74843
Contents
About This Manual................................................................................................................................... 1
1.0 Introduction.................................................................................................................................. 1
2.0 Installation and Wiring................................................................................................................ 3
2.1 Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Repacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Rear Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.4 Hardware Configuration and Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.5 Display/Keyboard Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.6 Freeze Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.6.1 Freeze Mode Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.7 Min and Max Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.7.1 Resetting Min and Max Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.8 Remote Display Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.0 Operation...................................................................................................................................... 8
3.1 Initial Setup Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Pressure Measurement Sequence (Gage) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Pressure Measurement Sequence (Absolute). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.0 Re-Calibration Only.................................................................................................................... 10
4.1 Pneumatic Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Instrument Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.3 Re-Calibration Procedure (Zero, Linearity and Hysteresis) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.0 Configuration and Calibration.................................................................................................... 12
5.1 Pneumatic Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.2 Instrument Calibration Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.3 Mode Selection Switch (S1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.4 Main Set-Up Mode Menu Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.5 Auto Set-Up Configuration Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.5.1 Auto Set-Up Menu (General). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.5.2 Auto Set-Up Menu (Gage). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.5.3 Auto Set-Up Menu (Absolute) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.6 Calibration Set-Up Configuration Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.6.1 Calibration Mode Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.7 Modifying Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.8 Span Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.9 Resetting Span Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.10 Overwriting Span Calibration Points and Hysteresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.11 Configuration Set-Up Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.12 Conversion Set-Up Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.13 I/O Set-Up Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.13.1 Port 1 Set-Up Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.13.2 Port 1 Printer Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
5.13.3 Port 1 Duplex Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5.13.4 Port 1 RS485 Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.13.5 Port 1 Simplex Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.13.6 Port 2 Set-Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.13.7 Time and Date Set-Up Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.13.8 Input Set-Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.13.9 Print and Zero Wiring Configration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Copyright © 2002 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
December 2002
ii
6.0 Serial Input/Output..................................................................................................................... 32
6.1 ASCII - Data Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.2 Printer (Port One or Port Two) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.3 Duplex (Port One Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.3.1 Full Duplex Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.3.2 Command Code <cmd> Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.3.3 Basic RS485 Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.4 Continuous Simplex Output (Compressed Format) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.5 Serial Output Wiring and Hardware Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.0 Maintenance and Service.......................................................................................................... 39
7.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2 Maintenance and Service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2.1 Panel/Chassis Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2.2 ORION-2C (PN 55283) or ORION-3A (PN 55287) Manifold Removal. . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.3 ORION Manifold - Valve Seat Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.4 ORION Manifold - Vernier Control Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.2.5 ORION Manifold - Vernier Control Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.6 ORION-2C Manifold - Valve Seat Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.7 ORION Manifold - Panel Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.2.8 ORION-2C Manifold - Valve Adjustment Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.2.9 ORION-3A Manifold - Valve Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2.10 Pressure Limit Control (Standard Pneumatic), Regulator Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.2.11 Pressure Limit Control (Standard Pneumatic), Regulator Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.2.12 Pressure Limit Control (Tescom), Regulator Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.13 Pressure Limit Control (Tescom), Regulator Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2.14 Panel Gauge Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.2.15 Panel Gauge Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7.2.16 Test Port Quick-Connect Fitting (PN 59762) Removal and Installation - 3 and 100 Full Scale PSI . . . . 48
7.2.17 Test Port Filter (PN 54188) Removal and Installation - 250 and 2500 Full Scale PSI . . . . . . . . . . . . . . 49
7.2.18 Test Port Quick-Connect Fitting (PN 59004) and Filter (PN 54188) Removal and Installation - 5000 and
7.2.19 Test Port (Output) Hose Quick-Connect Fitting and Filter Removal and Installation (PN 56991)- 5000 and
7.2.20 Input Port Filter (PN 54188) Removal and Installation - 2500 Full Scale PSI. . . . . . . . . . . . . . . . . . . . . 50
7.2.21 AC Fuse (PN 58076) - Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2.22 Input Port Filter (PN 54188) Removal and Installation - 5000 and 10000 Full Scale PSI. . . . . . . . . . . . 51
7.2.23 AC Fuse (PN 58076) - Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2.24 AC Power/EMI Line Filter (PN 58870) - Removal and Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2.25 Power Switch (PN 58878) - Removal and Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
10000 Full Scale PSI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
10000 Full Scale PSI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
7.3 ORION-2C Valve Assembly Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.4 ORION-3A Valve Assembly Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.0 Model Numbering System......................................................................................................... 62
9.0 Auto Selection Default Tables ................................................................................................... 63
10.0 Conversion Factors.................................................................................................................... 66
11.0 Options, Replacement Kits ........................................................................................................ 67
12.0 Specifications............................................................................................................................ 68
DLR3110 Warranty and Return Policy.................................................................................................... 69
DLR3110 Return Material Authorization Form....................................................................................... 70
DLR3110 Rack Mount Digital Pressure Indicator Manual

About This Manual

This manual is intended for use by service technicians responsible for installing and servicing the DLR3110 Digital Pressure Calibrator.
The DLR3110 series rack-mount combines the DLR Digital Pressure Calibrator with the Orion vernier for calibration of a wide variety of pressure sensing and measuring devices.
The DLR3110 utilizes a repeatable sensor coupled to microprocessor-based electronic circuitry and a selectable units conversion display system to provide digital representation of measured pressure. The unit has one test port. Dependent on model, a front panel mounted gauge tells the operator system pressure and includes a pressure regulator so the operator can not over-pressure a unit under test. Test hose is supplied standard for the customer. Front panel buttons provide
Warning
equipment. This is a high-pressure system. Only personnel formally trained in the use of pneumatic pressure equipment should be permitted to operate it. Potentially dangerous conditions can be produced through negligent handling or operation of the console due to the high pressure used within the unit.
These units are strictly for use with pneumatic pressures. Erroneous readings and potential damage could result from the introduction of hydraulic fluids into the internal tubing lines.
Before attempting to use either style pressure calibrator, the following instructions must be read and understood by personnel using the
Authorized distributors and their employees can view or download this manual from the Condec distributor site at
www.4condec.com
.
selection of the desired functions, including zeroing (GAGE models only).

1.0 Introduction

Using microprocessor technology, the rack-mountable DLR3110 instrument offers a combination of features, performance, versatility and reliability not previously available in a single, rack-mount pressure calibration instrument. Some of the features of the DLR3110 are listed below:
Streamlined, multi-function keyboard.
Selectable configurations, automatic set-up mode and multiple conversion units.
•Based on model, accuracy equal to or better than ±0.05%, ±0.1% or ±0.25% of indicated value from 20% of full scale.
•Alpha-numeric display prompting provides the operator with functional status information during both operation and calibration.
•A manually adjustable regulator (when applicable) allows the maximum system input pressure to be adjusted to any desired value higher (typically 20 – 50%) than the full scale range of the device being tested. This regulator helps protect the unit under test from being inadvertently over-pressurized.
•No manual alignment or potentiometer adjustments are required for system calibration. (50,000 displayed graduations; ten-point linearity compensation; automatic gain adjustment).
•Calibration integrity: Tamper-proof design. Once calibrated, numerous safeguards guarantee the integrity of pressure readings obtained.
•Operation: All controls, indicators and test port are accessible from the front panel. A version with test port on rear of unit is available by special order. See Section 3.0 on page 8 for instructions on system operation.
•For safe, clean operation, all pressure components are made of brass, copper, aluminum, or stainless steel and proof-tested to at least 150% of maximum operating pressure. In addition, the system contains a high-pressure burst disk and relief valves to protect both the operator and system components from harm in the event of inadvertent over-pressurization.
The heart of this calibration system is a stable and repeatable pressure transducer. These sensors produce an electrical output signal that is linearly proportional to the applied pressure.
Introduction
1
2
By combining these sensors with microprocessor-based circuitry, an even higher degree of operational accuracy and precision has been accomplished. For example, computer-generated correction curves for both the non-linearity and the hysteresis of the sensors improve these characteristics by an order of magnitude or more.
Two micro-metering valves and vernier are provided to control the external nitrogen source while the digital display indicates the magnitude of the applied test pressure. Also, a push-button switch provides the “zeroing” of the pressure display (on GAGE models). Over-pressure protection, based on model, is provided via a fully adjustable pressure regulator which is manually set for each new device being tested.
Figure 1-1 provides an overview of the DLR3110’s function.
VERNIER
ADJUST
PRESSURE LIMIT
MONITOR
FILTER
20u
INPUT PORT
NOTE: MAX INPUT BASED\P ON MODEL.
PRESSURE LIMIT
CONTROL
BURST DISK
3000 PSI
±50 PSI
RELIEF
VALVE
NOT REQUIRED ON 5K OR 10K MODELS
Figure 1-1. DLR3110 Flow Diagram
PRESSURE
INCREASE
MICROPROCESSOR-BASED
DIGITAL INDICATOR
DLR3110 FLOW DIAGRAM
FILTER
TEST PORT20u
(QUICK DISCONNECT)
VENT
VENT/VACUUM
PORT
TRANSDUCER
DLR3110 Rack Mount Digital Pressure Calibrator Manual

2.0 Installation and Wiring

2.1 Unpacking and Inspection

Carefully remove the instrument from its shipping container. A visual inspection of the instrument's external surfaces should be performed immediately after unpacking. If obvious damage occurred during transit, notify the shipping agency and distributor as soon as possible to receive instructions on how to proceed after an assessment of the damage is completed.
If the instrument shows no signs of damage, check to be sure all the required equipment and accessories were included. Keep original shipping carton for repacking in case repair is necessary.

2.2 Repacking

If the DLR3110 must be returned for repair, recalibration, or modification, be sure that it is properly cushioned and packed and that a description of the work to be performed is included. The original shipping carton should be kept for this purpose.
2.3 Rear Panel Configuration
The rear panel of the rack mount DLR3110 Series, contains the following items:
AC power cord (1) and input receptacle.
7/16-20, 37
NOTE: The maximum input pressure, supplied by user, is noted below input pressure port.
The unit's identification plate (4) and J6, 25 pin D connector, (5) for serial I/O communication interface.
•Optional if required items: Connector J5, 37 pin D connector, (6) for the Parallel BCD output. Connector J2 (7) for the Freeze mode cable. Test port (8) in place of front mounted.
°
-4 AN male fittings, vent/vacuum port (2) and input pressure port (3).
VACUUM
PUMP
2 5
TEST PORT
J2
FREEZE MODE
VENT/VACUUM
PORT
4
78
J3
3
J1
INPUT MAX. XXX
REAR VIEW
Figure 2-1. DLR3110 Rear View
6
AC INPUT
J5
J6
FUSE
1
Installation and Wiring
3
4
2.4 Hardware Configuration and Wiring
The CPU board in the DLR series has been designed for easy access to wiring and configuration of all available inputs and outputs. In addition to easy access, the wire and configuration mapping provide clear labeling of all input and output signals and the associated configurations of their device drivers.
J2
GND
LO
HI
S9
S8
S7
S6
S5
S4
C38
E11
E12
REMOTE SENSING
7
1
J3
+EXC
TB3
OUT
TB2
U18
10
RS485
AB
RS422
BA
1
3
1
S3
S2
20mA LOOP
P2
-OUT
+OUT
-IN P1
+IN
-OUT
+OUT
SIG COM
IN
OUT
OUTOUT
+5V
I/O 2
I/O 1
RETURN
S10
P1 IN
RS232
OFF
20mA RS485 RS422
6
-EXC
-SENSE
-SIG
+SIG
+SENSE
+EXC
1
CPU
TB1
TB1
ASSEMBLY KKR8924-[] DATE [] REV []
PORT 1 CONFIGURED FOR 20ma CURRENT LOOP
RS232
P1
PORT 1 CONFIGURED FOR RS232
P2
SHOWN CONFIGURED AS INPUTS
SERIAL INPUT/OUTPUT SELECTION
JUMPER POSITION
P1
20mA
RS232
S9
S8
S7
S6
RS485
RS422
JUMPER POSITION
BA
AB
P1
20mA
RS232
RS485
S9
AB
PORT 1 CONFIGURED FOR RS485 OR RS422 (TRANSMITTER)
S8
RS422
S7
PORT 1 CONFIGURED FOR RS485 OR
BA
RS422 (RECEIVER)
S6
INPUT/OUTPUT CONFIGURA TION
JUMPER POSITION
INS5OUT
S4
JUMPER POSITION
INS5OUT
S4
SHOWN CONFIGURED AS OUTPUTS
PORT 1 SERIAL INPUT SELECTION
THE PORT 1 INPUT SWITCH SELECTS FROM WHICH DEVICE THE SERIAL DATA IS RECEIVED
F1
C2
C1
3
20mA
P1
RS232
IN
E10
14
8
E8
S13
-EXC
Figure 2-2. CPU Configuration
NOTE: Unit is shipped previously set up and configured. For information on changing hardware configuration see Section 5.11 on page 21.

2.5 Display/Keyboard Functions

See Figure 2-3 on page 5 for information on display and keyboard functions.
Six-digit LED display of the measured pressure (1).
Three annuciators: Two of the LEDs are used for indicating whether the unit is operating as a GAGE pressure unit or as an absolute (ABS) pressure unit. The LED marked delta P display mode (2).
Eight character LED display for indicating the conversion units that the pressure data is currently being displayed in. This display also scrolls prompting messages if corrective action is required or a problem occurred in performing an operation (3).
For example, if attempting to perform a push to ZERO while the pressure data is in motion, the display shows
Unit is in MOTION
In GAGE units, the
.
function key (4) allows the indicator to be zeroed if it is within the limits
ZERO
defined by the ZERO configuration.
DISPLAY
key (5). In the normal operating mode this key is used to select and vie w all acti v e displays that were enabled in the Configuration mode (i.e., Pressure data, MIN/MAX data, Freeze mode data, and delta P display mode data).
indicates that the unit is in the
P
DLR3110 Rack Mount Digital Pressure Calibrator Manual
TARE
function key (6). The AUTO TARE (Auto.tr) feature is enabled in the Configuration mode under
the ZERO SET selection. When AUTO TARE is enabled, pressing the
TARE
key automatically tares out the value of the current pressure display reading and shifts the display to the P (delta P) mode. The resultant ZERO reference reading is useful in cases where it is desirable to subtract the atmospheric pressure from an absolute reading to obtain a quick gauge reference.
NOTE: This pseudo-gage reading is only valid for the current atmospheric pressure that was tared out. Any subsequent changes in atmospheric pressure result in an error.
PRINT
key (7) issues a demand serial printout of the display data from the indicator.
UP
(8) and
DOWN
(9) arrow keys step through and display the data in the conversion units that were
selected in the Configuration mode.
1
P Gage Abs
PSI
2
54
6
7
8 9
3
ZERO
DISPLAY
TARE
CE ENTER
PRINT
YES
NO
Figure 2-3. Display/Keyboard Functions

2.6 Freeze Mode

Freeze mode, used for testing pressure switches, is an edge triggered input signal that freezes the last display update for the number of seconds (1 – 10 seconds) selected in the Configuration mode. The Freeze mode is enabled in the CONFIG ? Freeze mode).
menu under freeze time selections (Refer to Section 5.11 on page 21 for enabling the
JUMPER PLUG SHOWN ON S5 CONFIGURED AS AN INPUT
PRESSURE SWITCH
(+5 VDC)
S5
I/O 2
IN
S4
OUT
TB2
I/O 1
RETURN
(DIGITAL COMMON)
Figure 2-4. Freeze Mode CPU Connections
Either opening or closing of a switch contact between an active input (I/O 1 or I/O 2) and RETURN “freezes”
the display (for the pre-selected amount of time in seconds) as shown above. However, while the display is in the freeze condition, it does not respond to another contact transition
.
Installation and Wiring
5
6
2.6.1 Freeze Mode Specifications
The following bullet points outline general specifications for the DLR3110’s Freeze mode.
The freeze input “freezes”
the display for approximately five seconds on the rise and fall time of the
input (edge not level triggered) and can not be re-triggered.
Input pulse > 100 ms required to guarantee detection.
Freeze mode “freezes”
the 6 digit numeric and LED bar displays in whatever mode it is in (for example,
Normal, Net, Min and Max mode).
•Alphanumeric display prompts
•During the “freeze”
the front panel keys are inactive.
Inputs are a +5 VDC logic level. Sinking current approximately
FREEZE
.
0.5 mA.

2.7 Min and Max Mode

The Min and Max mode is designed to capture and store the highest and the lowest pressure readings. These values may be recalled at anytime in the normal operating mode by repeatedly pressing the
DISPLAY
the desired value is displayed. When either the minimum or maximum value is selected, the alphanumeric display flashes between the mode (Min and Max) and its units.
NOTE: The Min and Max mode pressure data is automatically converted to any of the active conversion units by stepping through the conversion units using the UP or DOWN arrow keys.
Both the Min and Max mode are independent of each other and are enabled (turned ON) in the Configuration mode (see Section 5.11 on page 21).
key until

2.7.1 Resetting Min and Max Values

To reset the stored minimum or maximum value:
1. Select the Min or Max display to be reset using the key. One of the following messages appears on the alphanumeric
2. Press the
YES OR NO; RESET MIN? YES OR NO.
CE
3. At the display prompt shown above, press the
resetting the value.
NOTE: Min and Max data work with the standard pressure, not delta pressure data.
DISPLAY
key to reset the value or press the
YES
key.
display:
NO
RESET MAX?
key to abort

2.8 Remote Display Mode

The following describes the setup for connecting the optional six-digit remote display (PN 74683). The display duplicates the DLR3110 six-digit front panel display for all ranges and units selected (see Figure 2-5 on page 7).
NOTE: Remote display does not show current unit or mode designation.
See Table 5-10 on page 24 for setting the following parameters:
Parameter
Port 1 SIMPLEX
Baud 9600
Parity EVEN
Port 1 Modify
delay 1 AUTO
t –d 1 OFF data 1 COMPRESS EOL 1 CR
SOt 1 NONE
EOt 1 NONE
Configuration
Table 2-1. DLR3110 Parameters
DLR3110 Rack Mount Digital Pressure Calibrator Manual
JUMPER PLUGS SHOWN ON S6 & S7 CONFIGURED AS RS232
WIRE\PCOLOR
BLACK
P1
RS232
S7
S6
RS422
TB3
SIG COM
IN
BA
1
OUT
OUT
P1
P2
RS232
RED
Figure 2-5. DLR3110 CPU to Remote Display PN 74683 Wiring
J6
7
2
P2-4 ISO GND
P2-5 RS232 RXD
REMOTE DISPLAY
Installation and Wiring
7
8

3.0 Operation

The following sections provide instructions on various operation procedures for the DLR3110.

3.1 Initial Setup Procedure

To prepare for actual calibration usage, see Figure 3-1 on page 9 and proceed as follows:
1. Check that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
2. Plug in the power cord (see Section 2.3 on page 3) and power up the unit by pressing the power switch (3). Allow at least 10 minutes warm-up time.
3. If applicable, use the PRESSURE LIMIT CONTROL REGULATOR (4) to adjust the maximum system input pressure, as read by the PRESSURE LIMIT MONITOR (5), to any desired value higher (typically 20–50% higher) than the full-scale range of the Device-Under-Test (DUT). This procedure fully protects the DUT from being accidentally over-pressurized.
4. Connect the male end of the test hose to the TEST PORT (6) fitting.
NOTE: DLR3110 may be ordered with Test port on rear of unit. See Section 2.3 on page 3.
5. Connect the swivel fitting end (7/16-20) of the test (output) hose to the DUT using adapters if required. Tighten all connections properly.

3.2 Pressure Measurement Sequence (Gage)

Use the following steps to measure pressure with the GAGE model DLR3110:
1. With reference to Figure 3-1 on page 9, check that the indicator on the right end of the display indicates
Gage
(8). If not, see Section 5.11 on page 21.
switch (7). DLR3110’s LEDs momentarily display the word
2. Press and release the the display returns to a (2) is open, by momentarily depressing the
3. To apply pressure, close the VENT the INPUT valve (1) approximately 1/2 turn counter-clockwise until the numerical display begins to move. The pressure may change rapidly until reaching approximately 90% of the desired final value.
4. Use either the INPUT (1) or VENT valve (2) to obtain a specific pressure reading. Both provide precise control. As the pressure approaches the desired value, the valve being used for control should be rotated slowly clockwise to its closed position. With a little experience, pressure values very close to the desired final value can be quickly achieved.
5. To obtain exact pressure readings, slowly rotate the VERNIER control knob (10) in the direction required (clockwise to increase pressure) as indicated by the electronic numerical display.
ZERO
0
reading. The instrument may be zeroed at anytime, as long as the VENT valve
switch (7) for less than 5 seconds.
ZERO
valve (2) approximately two turns clockwise until it stops, then open
Zero

3.3 Pressure Measurement Sequence (Absolute)

Use the following steps to measure pressure with the ABSOLUTE model DLR3110:
1. With reference to Figure 3-1 on page 9, check that the indicator on the right end of the display indicates
Abs
(9). If not, see Section 5.11 on page 21.
2. If only pressure measurements greater than barometric are required, continue to step 2.1. If pressure measurements above and below atmospheric pressure are required go to step 3.
a) To apply pressure, close the VENT valve (2) by turning clockwise approximately two times until it
stops and open the INPUT valve (1) by turning counter-clockwise approximately 1/2 turn until the numerical display begins to move. In general, the pressure may be changed rapidly until reaching approximately 90% of the desired final value.
b) Use either the INPUT (1) or VENT valve (2) to obtain a specific pressure reading. Both provide
precise control. As the pressure approaches the desired value, the valve being used for control should be rotated slowly clockwise to its closed position. With a little experience, pressure values very close to the desired final value may be quickly achieved.
c) To obtain exact pressure readings, slowly rotate the VERNIER control knob (10) in the direction
required (clockwise to increase pressure) as indicated by the electronic numerical display.
and then
DLR3110 Rack Mount Digital Pressure Calibrator Manual
3. If pressure measurements above and below atmospheric pressure are required, connect a VACUUM PUMP to the VACUUM/VENT port (2) as shown in Figure 2-1 on page 3.
4. Open the VENT valve (2), close the INPUT valve (1) and apply power to the vacuum pump and allow it to evacuate the system for several minutes or until the digital display reading reaches equilibrium near 0 PSIA.
5. With the vacuum pump still running, close the VENT valve (2) and check for system leaks. If there are none, continue to step 5.1.
a) To apply pressure, close the VENT valve (2) by turning clockwise approximately two turns to its
stop and open the INPUT valve (1) by turning counter-clockwise approximately 1/2 turn until the numerical display begins to move. In general, the pressure may be changed rapidly until reaching approximately 90% of it desired final value.
b) Use either the INPUT (1) or VENT valve (2) to obtain a specific pressure reading. Both provide
precise control. As the pressure approaches the desired value, the valve being used for control should be rotated slowly clockwise to its closed position. With a little experience, pressure values very close to the desired final value may be quickly achieved.
c) To obtain exact pressure readings, slowly rotate the VERNIER control (10) knob in the direction
required (clockwise to increase pressure) as indicated by the electronic numerical display.
7
DISPLAY
ZERO
POWER
89
P Gage Abs
PSI
PRINT
TARE
ENTER
CE
YES
NO
4
INCREASE
DECREASE
CONTROL
INPUT
1
56
PRESSURE LIMIT
MONITOR
VERNIER
DUT
TEST PORT
VENT
23 10
Figure 3-1. Pressure Measurement Sequence
Operation
9

4.0 Re-Calibration Only

NOTE: This section is for the technician that is familiar with the front keyboard menus and is doing nothing more than a re-calibration. The re-calibration must be done in the same base units as originally calibrated. For a more in-depth configuration and calibration explanation, see Section 5.0 on page 12.
The step-by-step calibration sequence provided on the following pages permits a qualified technician to calibrate an entire DLR3110 instrument in approximately 45 minutes.
The instrument has been designed and programmed to provide the operator with various prompting symbols and legends during each phase of the calibration sequence. Also, to prevent unauthorized tampering or calibrations, numerous features have been incorporated to minimize this potential danger.
However, it must be emphasized that when performing these tests, the computer within the DLR3110 is actually being re-programmed and as such, it is imperative that the pressure standard being used is in satisfactory operating condition and that the technician fully understands its operating characteristics and methods of usage. In addition, the DLR3110 itself must be warmed up for approximately 30 minutes and electrically stabilized prior to performing a calibration cycle.

4.1 Pneumatic Calibration Set-Up

Figure 4-1 on page 11 defines a typical calibration set-up using a floating piston type dead weight tester. However, any type of precision standard is acceptable as long as its basic accuracy is twice that of the DLR3110’s or better.
To permit proper calibration at least an on/off and a vent valve must be provided (connected as shown in Figure 4-1 on page 11).
10

4.2 Instrument Calibration Set-Up

Check that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
The DLR3110 is placed into its Calibrate mode by closing rocker switch the DLR3110's LED and alpha-numeric display is used to provide operator prompting symbols as well as displaying the various data formats employed.
S1 position 2. In the Calibrate mode,

4.3 Re-Calibration Procedure (Zero, Linearity and Hysteresis)

Use the following steps for the re-calibration procedure of the DLR3110.
1. The alpha-numeric display reads
2. Press the
3. Press the
4. Press the
5. Press the
models and ABS for ABSOLUTE models. For GAGE models continue to step 6. For ABSOLUTE models go to step 7.
6. Press the INPUT valve is closed (rotate clockwise until it stops) and that the VENT valve is open (two turns counter-clockwise from its stop).
Press the Close the VENT valve (rotate clockwise until it stops). Go to step 7.
7. The alpha-numeric display reads Set the input pressure from the test standard to the target value on the display. W ait for system to stabilize and press the ENTER key on the display panel.
8. The alpha-numeric display reads Set the input pressure from the test standard to the target value on the display. W ait for system to stabilize and press the target values.
DOWN arrow key on the display panel. The alpha-numeric display reads CAPACITY. DOWN arrow key on the display panel. The display reads XXXX x .x. DOWN arrow key on the display panel. The LED display reads base.
DOWN arrow key on the display panel. The alpha-numeric display reads GAGE for GAGE
DOWN arrow key on the display panel. The alpha-numeric display reads ZERO CAL. V erify the
ENTER key on the display panel. The LED display reads 0.00 (number of zeros arbitrary).
ENTER key on the display panel. Repeat this step for the next eight predetermined SPAN
FILTER X, where “X” represents a numerical variable.
SPAN 1 CAL TARGET=XXX, where “X” represents a numerical variable.
SPAN 2 CAL TARGET=XXX, where “X” represents a numerical variable.
DLR3110 Rack Mount Digital Pressure Calibrator Manual
9. The alpha-numeric display reads HYST. CAL TARGET=XXX (X represents the preset target value on
display). Set the input pressure from the test standard to the target value on the display. Wait for system to stabilize and press the ENTER key on the display panel.
NOTE: For ABS models, use a vacuum pump to reach the applicable SPAN or HYST target values that are below local barometric pressure.
10. The alpha-numeric display reads FILTER 5. Save the calibration data by opening rocker switch S1
position 2. DLR3110 returns to normal mode.
11. Vent system and remove test standard.
"ON-OFF"
2-WAY VALVE
PNEUMATIC PRESSURE
TESTER
7
89
4
INCREASE
5
VENT
6
P Gage Abs
PSI
ZERO
POWER
DISPLAY
TARE
PRINT
YES
CE
ENTER
NO
3
DECREASE
CONTROL
INPUT
1
Figure 4-1. Instrument Calibration Set-Up
NOTE: AC input and input port are on back side rack-mountable calibrator.
PRESSURE LIMIT
10
MONITOR
VERNIER
TEST PORT
VENT
2
Re-Calibration Only 11
5.0 Configuration and Calibration
The following sections cover configuration and calibration of the DLR3110. For information on calibrating the unit without configuration changes, see Section 4.0 on page 10.
During the following procedures, the computer within the DLR3110 is actually being re-programmed. The pressure standard being used during calibration should be in satisfactory condition and fully understood by the service technician. In addition, the DLR3110 must be warmed up for approximately 30 minutes and electrically stabilized prior to performing a calibration cycle.

5.1 Pneumatic Calibration Set-Up

Figure 5-1 defines a typical calibration set-up using a floating piston type dead weight tester. Any precision standard is acceptable to use providing its basic accuracy is at least twice that of the DLR3110’s.

5.2 Instrument Calibration Set-Up

Check that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop).
"ON-OFF"
2-WAY VALVE
PNEUMATIC PRESSURE
TESTER
7
DISPLAY
ZERO
POWER
3
8
TARE
PRINT
CE
ENTER
9
P Gage Abs
PSI
YES
NO
4
INCREASE
DECREASE
CONTROL
INPUT
1
5
PRESSURE LIMIT
10
Figure 5-1. Instrument Calibration Set-up
NOTE: AC input and input port are on the back side of rack-mountable calibrator.
MONITOR
VERNIER
VENT
6
TEST PORT
VENT
2

5.3 Mode Selection Switch (S1)

Main set-up mode - Allows access to the Configuration and Calibration modes.
Calibration mode - Accesses the Calibration mode directly.
I/O mode - Allows direct access to the set up of the Serial I/O ports (ports 1 and 2), BCD, Analog output,
discrete inputs/outputs.
Diagnostic mode - Accesses a diagnostic mode for troubleshooting and evaluation (f actory access only).
12 DLR3110 Rack Mount Digital Pressure Calibrator Manual
Switch Position
Mode
1 2 3 4
Normal Closed Open Open Open
Main set-up Open Closed Open Open
Calibration Open Open Closed Open Diagnostic Open Open Open Closed
Table 5-1. Mode Selection Switch Modes
1
+
+
C4
R4
R3
CR7
R2
C6
1
U1
C8
U2
C5
E15
C31
R20
C10
+
C22
C7
U14
+
11
1
J5
EXPANDER
C14
C20
U8
20
U7
10
C17
U6
C18
C37
U5
R6
DIAG
I/O CAL
MAIN
R7
5
10
RP1
J1
U20
1
S1
MODE
6
J4
DISP
KEYBD
C35
DIAG
I/O CAL
MAIN
OPEN
12 43
1
S1
MODE
FRONT OF DLR3110
Figure 5-2. Rocker Switch S1 Location

5.4 Main Set-Up Mode Menu Selections

Enter the main setup mode by closing switch 1 of S1 located on the main board (see Section 5.3 on page 12). All other switch positions on Press
YES key to accept selections, NO key to reject.
LED Display Alpha-Numeric Display Key Pressed Remarks
SET AUTO ? YES key See Section 5.5 on page 14
SET-UP CONFIG ? YES key See Section 5.11 on page 21
---- CAL ? YES key See Section 5.6.1 on page 17
SET-UP I/O ? YES key See Table 5-10 on page 24
SET-UP CONV. ? YES key See Section 5.12 on page 22
S1 must be open. To save the changes and leave MAIN set-up, open switch 1 of S1.
NO key Continue to CONFIG?
NO key Continue to CAL?
NO key Continue to I/O?
NO key Continue to CONV?
NO key Return to AUTO?
Table 5-2. Main Set-Up Mode Selection Parameters
Configuration and Calibration 13
AUTO ?
ZERO
DISPLAY
TARE
CE
PRINT
ENTER
YES
NO
Figure 5-3. Auto Set-Up Mode
5.5 Auto Set-Up Configuration Mode Selection
The auto set-up automatically sets the following conditions in the DLR3110:
SET-UP - CONFIG configuration parameters
SET-UP - CONV conversion parameters
SET-UP - I/O configuration of inputs/outputs
---- CAL sets capacity, count by, abs/gage, and cal units; resets all calibration values; sets all cal target
values for span and hysteresis.

5.5.1 Auto Set-Up Menu (General)

LED Display Alpha-Numeric Display Key Pressed* Remarks
SET AUTO ? YES key See Section 5.4 on page 13
---- GAGE ? YES key See Section 5.5.2 on page 15 NO key Continue to ---/ABS?
---- ABS ? YES key See Section 5.5.3 on page 16 NO key Continue to ---/SPECIAL?
---- SPECIAL. ? NO key Allocated for future use, continue to
----/CUSTOM ?
---- CUSTOM ? NO key Allocated for future use, continue to
----/EXIT ?
---- EXIT ? YES key Return to main set-up mode selections
(see Section 5.4 on page 13).
NO key Returns to ----/GAGE ?
Table 5-3. Auto Set-Up Mode Selection Parameters
* Press YES key to accept selection; NO key to reject.
Upon completion and accepting an AUTO set-up configuration, the program automatically advances the indicator to the calibration mode (see Section 5.6.1 on page 17).
14 DLR3110 Rack Mount Digital Pressure Calibrator Manual

5.5.2 Auto Set-Up Menu (Gage)

See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under the AUTO set-up mode.
LED Display Alpha-Numeric Display Key Pressed Remarks
SET AUTO ? YES key See Section 5.5 on page 14
---- GAGE ? YES key See Section 5.5 on page 14
SET-UP 3 PSIG ? NO key Continue to press NO key until the
alpha-numeric display shows the DLR3110’s full scale range.
YES key Press YES key when the alpha-numeric
display shows the DLR3110’s full scale range.
SET-UP ACCEPT ?
YES or NO.
XXXX CAPACITY YES key “XXX” shown on LED display represents
EXIT ? YES key Return to SET-UP/ I/O?. (see Section 5.4
NO key Continue to press NO key until the
alpha-numeric display shows the DLR3110’s full scale range. See note below.
YES key Accept data
the full scale range and is based on previous selection.
on page 13).
NO key Returns to XXX/CAPACITY
Table 5-4. Auto Set-Up Mode (Gage) Parameters Selections
NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
Configuration and Calibration 15

5.5.3 Auto Set-Up Menu (Absolute)

See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under the AUTO set-up mode.
LED Display Alpha-Numeric Display Key Pressed Remarks
SET AUTO ? YES key See Section 5.5 on page 14
---- GAGE ? NO key See Section 5.5 on page 14
ABS ? YES key See Section 5.5 on page 14
SET-UP 15 PSIA NO key Continue to press NO key until the
alpha-numberic display shows the DLR3110’s full scale range.
YES key Press YES key when the alpha-numeric
display shows the DLR3110’s full scale range.
SET-UP ACCEPT ?
YES or NO
——YES key Accept data
XXXX CAPACITY YES key “XXX” shown on LED display represents
EXIT ? YES key Return to SET-UP/ I/O?. (see Section 5.4
NO key Continue to press NO key until the
alpha-numberic display shows the DLR3110’s full scale range. See note below.
the full scale range and is based on previous selection.
on page 13).
NO key Return to XXX/CAPACITY
Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections
NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit menu (see Table 5-3 on page 14).
5.6 Calibration Set-Up Configuration Mode Selection
Use one of the following methods for configuring the calibration set-up mode:
Method 1
CAL (SELECT). See Section 5.4 on page 13 for information on selecting main mode settings.
Method 2
CAL (DIRECT). Closing only rocker switch position CALIBRATION mode. See Section 5.3 on page 12.
NOTE: If the AUTO SET-UP selection is accepted and completed, the indicator automatically advances to the calibration mode (see Section 5.5 on page 14).
2 on switch S1 places the DLR3110 directly into
16 DLR3110 Rack Mount Digital Pressure Calibrator Manual
CAL ?
ZERO
DISPLAY
TARE PRINT
CE ENTER
YES
NO
Figure 5-4. Calibration Mode

5.6.1 Calibration Mode Menu

Press LEFT or RIGHT arrow keys to change values. The various menus may be exited or data selection aborted at any time by pressing the either the
YES or NO key is pressed.
LED Display Alpha-Numeric Display Key Pressed Remarks
**XXX CAPACITY DOWN arrow To change the capacity value, see
**XXX X .5 To change the resolution value, press LEFT
BASE PSI Press LEFT or RIGHT arrow key to select
---- GAGE ?
*XX ZERO CAL See step 1 below Does not exist on ABS type units. Disregard
*XX SPAN 1 TARGET VALUE=XX See step 3 below To change the target value, see Section 5.7
CE key. If the CE key is pressed, the prompt Exit ? YES or NO scrolls continuously until
Section 5.7 on page 19.
or RIGHT arrow key.
the base unit used to calibrate indicator. If for any reason this selection is changed, the indicator must be calibrated to the new base unit selection.
Press LEFT or RIGHT arrow key to select
or ABS ?
choice.
steps 1 and 2 below on ABS units.
on page 19.
Table 5-6. Calibration Mode Parameter Selections
*XX represents existing data. **XXX represents the present capacity of the DLR3110.
Use the following steps to set calibration data in the DLR3110 using the Calibration mode menu.
1. Verify that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is open (two turns counter-clockwise from its stop). See Figure 5-1 on page 12.
2. Press the
ENTER key on the display panel keypad. The LED display reads 0.00. Close the VENT valve
(rotate clockwise until it stops).
NOTE: If ZERO is not accepted, display shows either SIGNAL INPUT UNDER MINIMUM ZERO RANGE or INPUT ABOVE CAL INPUT RANGE (NO CAL messages).
3. Set the input pressure from the test standard to the target value on the display. W ait for system to stabilize and press the ENTER key on the display panel.
NOTE: For pneumatic setup, see Section 5.1 on page 12. For absolute models, use a vacuum pump to reach the applicable SPAN target values that are below local barometric pressure.
Configuration and Calibration 17
LED Display Alpha-Numeric Display Key Pressed Remarks
CAL SPAN OK N/A Span has been accepted. Proceed to step
4 below. Span may be accepted, but based on span value, alpha-numeric display may return with scrolling message “FULL SCALE WILL NOT BE OBTAINED”.
NO CAL SPAN CAL N/A Span has not been accepted. Repeat step
3 above. If span is not accepted, display returns with one of the following scrolling NO CAL messages: “SIGNAL INPUT UNDER THE MINIMUM ZERO RANGE” or “INPUT ABOVE CAL INPUT RANGE”.
Table 5-7. Calibration Mode Span Messages
4. The alpha-numeric display reads SPAN 2 CAL TARGET=X (X represents the existing target value on
display). Change target value if required. Set the input pressure from the test standard to the target value on the display. Wait for system to stabilize and press the ENTER key on the display panel. See Table 5-7 as required.
5. The alpha-numeric display reads
SPAN 3 CAL TARGET=X (X represents the existing target value on
display). Change target value if required. Set the input pressure from the test standard to the target value on the alpha-numeric display . Wait for system to stabilize and press the ENTER key on the display panel. Repeat this step for SPAN target values 4 – 10.
LED Display Alpha-Numeric Display Key Pressed Remarks
*XX HYST.CAL TARGET=XXX See step 6 below To change the target value, see Section 5.7
on page 19.
*XX FILTR 5 DOWN arrow To change target value 1 to 10 by 1, 10 to
20 by 2, 20 to 50 by 5, press UP/DOWN arrow keys.
---- EXIT ? YES key Return to main menu display SET-UP/IO ?. See Section 5.4 on page 13.
NO key Return to XXX/CAPACITY. See
Section 5.6.1 on page 17.
Table 5-8. Calibration Mode Hysteresis and Filter Target Parameters
6. The alpha-numeric display reads
HYST.CAL TARGET=XXX (X represents the preset target value on
display. See Table 5-8.). Set the input pressure from the test standard to the target value on the alpha-numeric display. Wait for system to stabilize and press the ENTER key on the display panel.
NOTE: For absolute models, use a vacuum pump to reach the applicable hysteresis target value if below local barometric pressure.
7. Save the calibration data by opening rocker switch S1 position 2. The DLR3110 returns to normal mode (see Section 5.3 on page 12).
8. Vent system and remove test standard and/or vacuum pump.
18 DLR3110 Rack Mount Digital Pressure Calibrator Manual

5.7 Modifying Data

Data values previously set are easily altered by pressing either of the horizontal arrow keys to enter the modification mode. Upon entering the mode, the least significant digit indicates the current position in the data that can be changed. Press the desired position indicated by the flashing digit, then use the
UP or DOWN arrow key to increment or decrement
the value of the selected digit. Once the new data value has been set, press the abort. If an abort is issued, the data reverts back to its former value. The display prompts with
(LSD) begins to flash. A flashing digit
LEFT or RIGHT arrow key to select a
ENTER key to accept or the CE to
ENTER or ABORT
accordingly.

5.8 Span Calibration Points

There are a total of 10 calibration points (SPAN 1 - 10 CAL TARGET) which are sequentially entered. It is not required that all calibration points be entered. If calibration is performed with less than 10 points, the curve is extrapolated to full capacity based upon the slope correction between the last two entered SPAN CAL points. In the example below, if only four span calibration points were entered (SPAN 1 to SPAN 4), the slope correction from SPAN 4 to full capacity would be defined by the computed slope between SPAN 3 and SPAN 4.
NOTE: If fewer than 10 SPAN CAL points are desired during calibration, simply step past the remaining calibration points using the DOWN arrow key, until the HYSTERESIS calibration point is reached.
HYSTERESIS\PERROR
DESIRED LINEARITY FROM ZERO TO RATED OUTPUT
SPAN 4
mV/V OUTPUT
SPAN 1
0
LOAD
Figure 5-5. Span Calibration Points
SPAN 3
SPAN 2
NON-LINEARITY\PERROR

5.9 Resetting Span Calibration Points

To reset the span calibration points, use the DOWN arrow key to step to the desired SPAN CAL point and then press the
CE key. The correction factor of the selected SPAN CAL point and all others beyond will be reset to
unity (see Section 5.6 on page 16).
NOTE: X’s shown on LED display and alpha-numeric display columns represent existing data.
After a reset has been performed, entry of new SPAN CAL points proceeds as described in Section 5.6.1 on page 17, starting with step 3.
NOTE: If all 10 SPAN CAL points are not used in the calibration, the last span calibration point to appear (SPAN x CAL TARGET=xx.xx) is not an entered correction point. It only indicates the next span calibration that can be entered.

5.10 Overwriting Span Calibration Points and Hysteresis

Any SPAN CAL point can be overwritten with a new value provided that it remains within the range bounded by the span calibration points above and below the span point to be overwritten. Figure 5-6 on page 20 shows SPAN 2 CAL point being redefined within the boundaries set by SPAN 1 and SPAN 3. HYSTERESIS can also be overwritten as long as the new entry is below 80% of the full capacity value. For information on how to reach SPAN CAL point and HYSTERESIS, see Section 5.6 on page 16.
Configuration and Calibration 19
Loading...
+ 51 hidden pages