DLR334/DLR3110
Digital Pressure Indicator
Applications
• Test stands
• On-site manufacturing pressure monitoring
• Transfer standard
• Low-cost digital data logging of readings
DLR334 - Table top model
DLR334 -Table top/panel mount model
DLR3110 precision patented valve assembly and vernier, test port
and test hose, pressure limit control and monitor.***
Test port may be mounted on front or rear of unit
19 inch rack-mount version of DLR334 plus
Please refer to price schedule for part numbers and
part numbering guide.
Wi
th high-accuracy features and a full range of options,
the DLR334 digital pressure indicator and the DLR3110
19 inch rack-mount are the obvious choice for a wide
range of pressure applications.
From jet engine to automotive, machine tool to general
manufacturing, the DLR334 and DLR3110 are perfect for
use in test stand monitoring. With large red LEDs, selectable resolution, up to 20 selectable engineering units, and
up to 0.05% of reading accuracy, these unique models
provide reliable, repeatable performance up to 10,000 psi.
Flexible features and options also allow the DLR334 and
DLR3110 to serve as a calibrators. Other advantages,
such as a variety of network communications capabilities, peak hold and freeze modes, and optional relays and
setpoints, only enhance the DLR334 and DLR3110’s performance power. The DLR334 panel mount or table top
and the 19 inch rack-mounted DLR3110 lend themselves
to custom use in calibrating a variety of devices, including pressure gauges, pressure transducers, pressure
switches, relief valves, and burst discs.
The DLR334 and DLR3110...providing the ultimate in
accuracy and affordability for high-pressure applications.
Standard Features
• Three models available:
DLR334 table top model
DLR334 table top/panel mount model
DLR3110 19 inch rack-mounted version
• 115 or 230 VAC operation
• Ranges from vacuum to 10,000 psi
• 20 selectable engineering units
• Freeze mode for pressure switch testing
• 0.05% of reading, 0.1% of reading, or the economical 0.25% of
reading model
• Selectable resolution
• NIST traceable
• Digital outputs: RS-232, RS-422, and RS-485
• Min/Max pressure readings
• Accuracy is percent of reading from 20% of full scale
Options/Accessories
23113** ....... TMU295 ticket printer
81076** ....... TM-U220D tape printer
15547 ......... 25-pin printer interface cable
19297 ......... DLR334 panel mount adapter kit (included in panel
mount models, not available for black extruded model)
63255 ......... Additional operating and maintenance
manual for DLR334
DLR3110 only
58478 ......... Pressure trap (5000 psi and above)
Literature sheet PN 58596
58483 ......... Pressure trap (250-2500 psi)
Literature sheet PN 58609
58487 ......... Pressure trap (3-100 psi)
Literature sheet PN 58621
74843 ......... Additional operating and maintenance
manual for DLR3110
55279 ......... Additional test port (output) 5 ft. long hose with
quick-disconnect plug (SST) on one end and 1/4 in.
37º female AN swivel tube coupling on the other end
(DLR3110 250-2500 psi operation)
55280 ......... Additional test port (output) 4-1/2 ft. long hose with
quick-disconnect (SST) plug on one end and 1/4 in.
37º female AN swivel pressure fitting on opposite end.
(DLR3110 5000 psi and above operation)
55281 ......... Additional test port (output) 5 ft. long hose with
quick-disconnect plug (brass) on one end and 1/4 in.
37º female AN swivel tube coupling on the other end.
Additional options available*
Standard unit plus two relays and setpoints
Standard unit plus 0-10 VDC and 4-20 mA outputs
Standard unit plus parallel BCD output
* Please refer to price pages for part numbers
and part numbering guide.
** Consult factory for printer specifications.
*** Pressure limit control and monitor are not required for 5000 psi models and above.
(DLR3110 3-100 psi operation)
(Specifications on back)
CP-5
DLR334/DLR3110
Specications
OVERALL ACCURACY:
0.05% of reading, 0.1% of reading, 0.25% of reading all from
20% of full scale and above. Below 20% of full scale units
remain 0.05%, 0.1%, or 0.25% of 20% of full scale range.
Includes all effects of linearity, hysteresis, repeatability, and
ambient temperature. No temperature effect from
40°F to 140°F. NIST traceable.
TEMPERATURE RANGE:
Operating: 40°F to 140°F (4°C to 60°C)
Storage: 0°F to 185°F (-17°C to 85°C)
Warm up: 3 minutes to rated accuracy, 15 minutes to
complete stability
CALIBRATION:
Full NIST traceability. Calibration points can be adjustable for
operator convenience. Calibration via front panel and/or
digital interface.
RESOLUTION:
Adjustable, 50,000 display counts (maximum) of range
DISPLAY SIzE:
Two-line, six-digit, 0.6 in. high numeric display, eight alpha
characters per line, three annunciator light bars, bright red Light
Emitting Diode (LED)
TABLE TOP
PRESSURE HOSE (DLR3110 ONLY):
Quantity supplied: One output
Length: 4 ft. nominal
DISPLAY UPDATE:
Adjustable, 100 ms maximum
MEDIA COMPATIBILITY:
Any media compatible with 17-4PH or 15-5PH stainless steel
OVERPRESSURE:
200% of range (minimum)
POWER REqUIREMENTS:
115 or 230 VAC
CASE MATERIAL:
Single piece extruded aluminum or panel mount sheet metal
WEIGHT:
DLR334 Table top: 7.5 lb (3.4 kg)
DLR334 Panel mount or table top: 8.7 lb (3.9 kg)
DLR3110: 19 lb (8.6 kg)
PANEL MOUNT
OR TABLE TOP
DIMENSIONS
DLR334 - Table Top Model
A = 8.46 in. (214 mm) C = 10.5 in. (266.70 mm)
B = 2.94 in. (74.67 mm)
DLR334 - Panel Mount or Table Top Model
A = 7.56 in. (192.02 mm) D = 10.35 in. (262.89 mm)
B = 9.14 in. (232.15 mm) F = 7.38 in. ±0.03 (187.45 mm)
C = 3.78 in. (96.01 mm) G = 3.6 in. ±0.03 (91.44 mm)
DLR3110 - 19 inch Rack-mount (not shown)
19 in. W x 7 in. H x 10 in. deep (482.6 mm x 177.8 mm x 254 mm)
Excluding front handles
Rack-mount dimensions are for reference only. Consult factory for
cutout dimensions.
Specifications subject to change without notice
© Condec 2012 Catalog PN 64749 1/12