AZN 0
This option is used to select the auto zero
tracking.
Options : 0.25d, 0.5d,
0-AUTO
This option is used to select the Initial zero
setting range (mechanism) when turn the
indicator.
Options : 0%, 2%, 5%, 10%, 20%
P1 REF
0- RANGE
This option is used to select the (SAZSM)
manual zero range when press the ZERO key.
Options: 2%, 4%, 10%, 20%,
C AP
This display will show xxxxxx for setting the
capacity.
CAL
Calibrate, see detail in section 10.
P 2 CAL
CNT
This display will show xxxxxx for indicating the
internal counts.
P-AUTO
Automatic print setting, when the weight is
stable, the data will be printed automatic
P-PRT
Manual print setting, when the weight is stable,
press the PRINT key, the data will be printed.
P3 P RT
P cont
Connect PC, the data will be send to PC
AUDIT TRAIL COUNTER :
CALIBRATION:
When the scale self-check press the SMPL/ key, it will show ”CAL X”, it is an Audit
trail counter for each Successful CALIBRATION.
PARAMETER (OPERATIONAL) COUNTER :
When the scale self-check press the N/G key, it will show “opt X”, it is Audit trail
counter for any change in parameter setting.
SECTIOIN 12 TROUBLE SELF CHECKING
Problem Possible cause Common solutions
Display is
blank, No
turn on test
Mains power is turned off
Power supply not plugged in
Power supply faulty
Internal battery not charged
Display turned off
Check power is going to the
scale and switch is on.
Verify the voltage going to the
scale matches the power supply
labels on the power module or
scale.