Casella CEL DryCal DC-Lite flow calibrator User Manual

Bios International Corporation
10 Park Place
Butler, New Jersey, USA 07405
Phone (973) 492 8400
Toll Free (800) 663 4977
Fax (973) 492 8270
Email sales
Web www.biosint.com
© 2003 Bios International Corporation
MK01-1 Rev. C 1.09
DryCal®DC-Lite Manual
Table of Contents
1.0 DC-Lite Features 2
2.0 Unpacking Checklist 3
3.0 General Description 3
4.0 Theory of Operation 3–4
5.0 Operating Instructions 4
5.1 DC-Lite Button Panel 4
5.2 Turning the Power On 5
5.3 Disabling & Re-Enabling the 5-Minute Auto-Shutoff Feature 5
5.4 Taking Readings 5–6
5.5 Resetting the Averaging Sequence 6
5.6 Printing 6–7
5.7 Stop & Reset 7
5.8 Resetting a Printed Sequence 7
5.9 Printing to a PC 8
5.10 Hard Reset Button 8
6.0 Battery System 8
6.1 Charging the Battery 9
6.2 Battery Maintenance & Storage 9
7.0 Isolating the DryCal from Other Instruments 9–10
7.1 Use with Instruments that Contain Internal
Mass Flow Controllers (MFCs) 10
7.2 Use with Personal Air Samplers 10
7.3 Calibrating Rotameters 10
7.4 Use with Magnehelic Manometers 10–11
8.0 Maintenance, Quality Assurance 11
8.1 Maintenance 11
8.2 Leak-Test Procedure 11–12
8.3 Air Containing Particulates 12
8.4 Return Authorization 12
8.5 Shipment 12
8.6 Long-Term Storage 12–13
8.7 Calibration 13
9.0 Calibration Statement 13–14
10.0 Limited Warranty 14
DryCal®DC-Lite Specifications
Size 5" x 5" x 2.75" 127 mm x 127 mm x 70 mm
Weight 42 oz 1200 g
Flow Ranges | Air Flow Accuracy Specifications based on averaged readings.
Lower limit is based on self-tested maximum leakage.
Model Optimum Flow Range (±1%) Extended Flow Range
L 10–500 ml/min 1 ml/min–500 ml/min
ML 50 ml/min–2 L/min 5 ml/min–5 L/min
M 100 ml/min–7 L/min 10 ml/min–12 L/min
MH 200 ml/min–20 L/min 20 ml/min –20 L/min
H 500 ml/min–30 L/min 50 ml/min–30 L/min
Contact Bios for extended flow range specifications, or visit our website
at www.biosint.com/products/dclite_models.htm
Battery System 6V rechargeable, sealed lead-acid, 6-8 hours typical operation
AC Battery Charger | Power Adapter Wall-mounted, single-station charge,
input: 100 to 120 VAC, 60 Hz., output: 12 VDC. Optional input: 200 to 240 VAC,
50 Hz., output 12 VDC.
Operating Modes Single reading, 10 readings, or auto-mode.
Temperature Range 0–55 °C
Humidity Range 0–70% non-condensing
Printer Port Standard parallel (Not compatible with printers that require
Microsoft
®
Windows™)
Warranty Product, 1 year; battery, 6 months
The annual recalibration program offered by Bios is elective and is not included
as a warranty item. All specifications are subject to change.
Please contact Bios or visit our web site at www.biosint.com for the most
current information.
2.0 Unpacking Checklist
Your DryCal DC-Lite has been packaged with care and includes all components
necessary for operation. Please take a moment to check that you have received
the following items. If you believe you have not received a full shipment or have
any other questions, please contact Bios immediately.
Your DryCal DC-Lite Includes
Single-Station Battery Charger
Tubing Kit
Additional High Flow Tubing (with DCL-MH and DCL-H only)
Certificate of Calibration
Instruction Manual
Registration Card
3.0 General Description
The DryCal DC-Lite is a field-portable primary flow calibrator used for industrial
hygiene, environmental and laboratory flow measurement applications.
The DC-Lite uses patented dry piston technology and infrared sensors to obtain
volumetric flow rates quickly and accurately.
Housed in a small, sturdy case, each unit employs a variety of popular user con-
veniences such as push-button read and auto-read functions, a large alphanumeric
display, battery level indicator, 5-minute automatic shut-off and a parallel printer
port for data-logging.
4.0 Theory of Operation
The DryCal DC-Lite can be used to measure air flow rates for either a vacuum flow
source (connected to the outlet port) or a pressure flow source (connected to the
inlet port). Before a reading is initiated, or between readings, a computer-directed
valve performs a bypass function. This allows the air to pass through the DryCal
valve, bypassing the flow-measuring cell which is then able to reset.
As a reading is initiated (by pressing the Read button) the internal valve closes
and the flow source evacuates or pressurizes the air in the flow-measuring cell. The
piston rises at the rate of evacuation or pressurization. A precision encoder system
provides two finely collimated light beams with a known distance between the
beams.
After a suitable acceleration interval the piston breaks the first infrared light beams
as it passes. The flow reading is completed when the second infrared beam
3
2
1.0 DC-Lite Features
Front and Side
Back and Side
Cell Viewing Window
Charging Jack
Inlet Port
Connect tubing here for pressure applications
Outlet Port
Connect tubing here for suction applications
Parallel Printer Port
Reset Button