If you want to get on and use the instruments without any background
information, skip Section 3 and go straight to Section 4.
USERS GUIDE
1.Display
ata is Being Recorded
Battery Condition
Overload has
Occurred
easurement
dentity or Date
urrent Operating
ode, Repeatedly
ressing the
ENU key will
elect each of these in turn
SLM
REC
C-A
DOSESET UP OPTION
Measurement is Paused
Frequency Weighting
Indicates Active
dB%
960081e
Current Data or Time
Option Keys
Current Units
2.Keys
Display
Switches dosimeter ON/OFF
Changes operating mode)
Accesses stored data
Select active options from the current
menu
DATA
MENU
Starts/Pauses/Restarts a measurement
record
Confirms actions of other keys
Stops a measurement when
followed by
SET
STOP
ENTER
960080
CEL-420/460GettingStarted-Page53
Page 2
3.Introduction to the Dosimeters
The CEL-420 and CEL-460 Noise Dosimeters measure the frequency
weighted noise exposures and peak sound levels simultaneously. The dosimeter microphone and lead on both models can be replaced by a
sound level meter microphone to offer comprehensive sound level measurement. The dosimeters have the following features.
¤ Four built-in standard dose measurement setups:
OSHA, DOD, ISO85* and ISO90*,
¤ Up to three user specified dose measurement setups,
¤ A single sound level meter setup,
¤ Sixteen dose result stores,
¤ User selection of up to two profiles attached to each
CEL-460 dose measurement,
¤ Self configured storage.
The instruments are built to withstand rough industrial conditions with
cases formed from a polyester/polycarbonate material that gives them a
high resistance to damage. Data integrity is further protected by a dose
microphone lead which is designed to resist knocks and abrasions without affecting the signal passing from microphone to instrument.
Both instruments can be operated and deliver dose and SLM
results without the need for other equipment, beyond an acoustic calibrator. However, they become even more versatile when their measurement and setup data is downloaded to a PC using the CEL-6702 dB10 or
CEL-6704 dB12 Windows™ based software.
These programs have the facilities expected of fully featured Windows™ packages offering post processing, cut and paste between applications and comprehensive word processing capabilities. In addition,
dB12 has extensive on screen graphing facilities.
3.1Sound Level Meter and Intrinsically Safe Models
Both dosimeters can be converted to a miniature sound level meter by fitting a CEL-425 SLM Microphone
Adaptor. The CEL-420 then becomes
a CEL-424 SLM, while the CEL-460
becomes a CEL-464. For sound level
meter operation refer to Chapter 8.
Intrinsically safe versions are
available of both dosimeter and sound
level meter versions, with the full
model line up shown in Table 1.
Intrinsically safe versions are rated to EEx ia IIC T4 for use in
zones 0, 1 and 2 with IS Certificate Number - Ex97E2110X tested to
EN50020 (this corresponds to CENELEC apparatus group IIC, North
American grouping Class1).
3.2CEL-420 Noise Dosimeter (& CEL-422/424/426)
The following features also apply equally to the CEL-420/422/424/426.
The CEL-420 series instruments are ideal for quick on-site surveys and
for monitoring personal noise exposure in accordance with European ISO
or USA OSHA and DOD standards. For dose measurement,
they are passive instruments
with settings that may be reviewed by key strokes, but
which can be selected only by
means of the dB10 (or dB12)
software. However, all sound
level measurement parameters can be set from the keypad.
There are overlapping
measurement ranges: 50-120
& 70-140 dB, A & C RMS
weightings, Linear & C Peak
weightings, Fast, Slow &
Impulse time weightings, and energy exchange rates (Q) of 3, 4, 5 & 6.
In dose measurement mode, they display % dose, % dose projected for
an 8 hour period, peak level and measurement duration.
For those countries which are subject to European Union regulations or the equivelent ISO standards, the instruments measure the daily
sound exposure level (L
cal with the L
required by ISO 1999), while for USA OSHA regula-
EX,8h
Aeq,8hr
Pa2h according to IEC 1252, that is identi-
tions they measure the Time Weighted Average Level (TWA).
When used as a sound level meter they display sound level,
maximum sound level, peak and time-averaged (L
Aeq
or L
Avg
) sound
levels.
3.3CEL-460 Logging Noise Dosimeter
(& CEL-462/464/466)
The following features also apply equally to the CEL-460/462/464/466.
The CEL-460 series instruments are recommended for detailed measurements as they feature extended processing and memory functions.
There is an additional 30-100 dB measurement range that can be used
for environmental surveys, and automatic run timing facilities controlled
CEL-420/460GettingStarted-Page55
Page 4
by the built-in clock. There are time-history recording capabilities that allow up to two profiles to be saved with each dose result set in a separate profile store. Sampling times can be specified between 1 s and 1
hour, while profile storage is in excess of 50 000 data points.
The instruments are able to measure and save up to five user
specified L
values (exceedance level percentiles). Run durations speci-
n
fied via instrument keys, by dB10 or dB12, and start and stop times preset via dB10 or dB12 can be switched on or off from the instrument
keys. Other settings can be changed only by dB10 and dB12 Software.
When data saved by a CEL-460 instrument is downloaded to a
PC, the user can specify a period over which to re-calculate projected %
dose values. The software can also calculate six exceedance time percentiles with levels preset to values specified by measurement standards,
plus one user specified time percentile.
4.Install Battery & Microphone
Connect a new 6LF22 or equivalent 9 V battery to the terminals in the battery
compartment in the rear of the instrument.
With the instrument switched OFF,
insert the lead connector of the CEL-6681
Dosimeter Microphone (or the connector
on the CEL-425 SLM Adaptor stalk) into
960079
the socket on the top end of the case with
the red dot facing the front of the
dosimeter.
CEL-425
SLM Adaptor
Remove the
connector ONLY by pull-
Red "Dot"
ing on the knurled
sleeve.
DATA
MENU
SET
STOP
ENTER
CEL-6681
Dosimeter Microphone
990022
For IS versions, use ONLY Duracel® MN1604
batteries. Install and remove the battery and microphone
ONLY in a non-hazardous area.
Page56-CEL-420/460GettingStarted
Page 5
5.Perform Acoustic Calibration
'
Operations are performed by a sequence of key strokes.
Perform a field accuracy check
each time the instrument is switched ON
using a CEL-282 (Class 2) or CEL-284/2
(Class 1) Calibrator as follows.
Fit the microphone into the CEL4725 Coupler supplied with the calibrator,
making sure it is pushed firmly into contact
with the shoulder in the coupler cavity.
Then fit the coupler complete with microphone into a CEL-282 (or CEL-284/2) Calibrator, again ensuring that it makes contact
with the shoulder in the calibrator cavity.
(To aid removal, the coupler flange does
not fit close against the calibrator housing).
Switch the calibrator ON, then follow the instruction after the Start Up sequence of displays has finished.
Measurement data is stored in the DOSE operating mode and is acces-
sed using thekey. Current data can be inspected while the instrument is recording. When the instrument is stopped, with no data being
saved, stored data can be retrieved and displayed.
8.1Review Data From Current Run
OperationPressDisplayShowsComments
1.InspectData
FromCurrentRun
2.Inspect
FurtherDataFromCurrentRun
repeat
DatafromCurrentRunDisplaycyclesthroughdata
8.2Review Stored Data
Stored data is identified by date and measurement start time and up to
16 runs can be stored simultaneously before the memory is overwritten
by a "first-in/first-out" process.
With the dosimeter switched OFF, use both the C6671 and CEL-6672 Cables to connect the instrument to a Centronics printer as shown. This allows the dosimeter to recognise the printer cable during its Start Up
sequence.
Only
stored data
can be printed,
as a printer
cannot be connected to the
dosi-meter at
the same time
SLM settings can be changed via the instrument keypad. Connect the
CEL-425 SLM Adaptor (as described in Section 4), perform an acoustic
calibration as described in Section 5, then proceed as follows.
The OPTION operating mode contains the part of the program which
allows the following items to be checked or adjusted.
Acoustic calibration (see Section 5),
Adjustment of the calendar clock and time,
Battery voltage,
Communication status when the dosimeter is connected to a
PC for use with dB10 or dB12 Soundtrack Software,
Activating the run timer commands set by dB10 and dB12
All models of the CEL-420 and CEL-460 Dosimeters comply with the
EMC Directive 89/336/EEC of the European Union. They have been
tested according to the standard delivery schedule and comply with the
following standards.
EN 50081-1 : 1992, EN 50081-2 : 1993: Generic emission standards for residential, commercial, light industry and industrial environments.
EN 50082-1 : 1992, EN 50082-2 : 1995: Generic immunity standards (for both RF fields and electrostatic discharge) for residential, commercial, light industry and industrial environments.
CEL-420/460GettingStarted-Page65
Page 14
13.Intrinsically Safe Models
Four models in the CEL-420/460 series are available with Intrinsic Safety
approval for use in hazardous areas. Their instrument numbers are
shown in the table.
The operation of each model
is identical to its non-IS counterpart.
The Noise Dosimeter versions use
the standard CEL-6681 microphone
on a 0.75 m cable for personal use
whereas the sound level meter versions use the standard CEL-425 microphone adaptor.
The Intrinsically Safe versions are fitted with a different firmware version but are functionally identical to the standard products. They may be connected to a personal computer for setup and download of results using the CEL-6702 or 6704
Windows software packages.
Products rating - to EEx ia IIC T4 for use in zones 0, 1 and 2. Intrinsically Safe Certificate Number - Ex97E2110X tested to EN50020 (this
corresponds to CENELEC apparatus group IIC, North American grouping
Class1).
Approved Power Supply
Only the Duracel® MN1604 alkaline type battery may be used in these
meters (9 V radio battery). The use of any other battery will invalidate the
intrinsically safe certification.
Battery Life
Approximately 30 hours under typical conditions.
Special instructions
The battery must be replaced only in a non-hazardous area.
The instruments must be calibrated in a non-hazardous area following the same instructions as for the standard models.
Any repairs needed for the Intrinsically Safe models must be carried out only by an approved CASELLA CEL service centre following approved written procedures.
To ensure its conformity with the specification, this instrument is
thoroughly inspected and it's accuracy verified prior to dispatch. All technical information is filed under the instrument serial number, which
should, therefore, be quoted in any correspondence.
The manufacturers undertake to rectify any defect in the instru-
ment that is directly attributable to faulty design or assembly, and which
becomes apparent during the warranty period. In order to take advantage
of this warranty, the instrument must be returned, carriage paid, to the
manufacturer's factory or accredited agent, where necessary repairs will
be carried out.
The warranty period runs for 12 months from the date of receipt
of goods, with exceptions on certain specialised components supplied by
other manufacturers which may be warranted for shorter or longer periods by their actual manufacturers. In all such cases, the benefit of these
undertakings will be passed on to the user.
CASELLA CEL's liability is limited to items of their own manufac-
ture, and they do not accept liability for any loss resulting from the operation or interpretation of the results from this equipment.
To obtain repair under warranty, the instrument should be
packed and returned in it's original packing or an equivalent either to
CASELLA CEL's local agent, or in the case of domestic sales, to the
CASELLA CEL Service Department at Bedford. Please include the following information:
Instrument Type(s), Serial Number(s) and Firmware Version
Number(s),
Customer name and address,
Contact name and phone number,
Details of any PC and Software involved, including Version
Number(s),
Reason for returning the equipment with a detailed description
of the fault,
List of any error messages that may have been displayed.
The necessary adjustments or repairs will be carried out, and the instrument returned as soon as possible.
A comprehensive Instrument Calibration Maintenance Agree-
ment (ICMA) scheme is available to extend the initial warranty period of
this instrument. At the end of the first warranty period, it is recom-
CEL-420/460GettingStarted-Page67
Page 16
mended that the equipment be returned to the Service and Verification
Department at Bedford, where it will be inspected and entered into the
ICMA scheme as required. The warranty will then be extended for the period stated on the individual schedule. Please contact your local
CASELLA CEL agent for full details of this service.
After the warranty has expired (except on approved accounts)
service work is undertaken against quotations, and all packing and transit
costs are charged extra.
Page68-CEL-420/460GettingStarted
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.