Casella CEL CEL-593 User Manual

Page 1
Getting Started With
CEL-553, CEL-573 & CEL-593
Sound Level Analysers
“Series 3”
1. Introduction
Issue: 6
Series 3 of the CEL-500 analysers introduces a new instrument: the CEL-553 which is available ONLY as a real-time third-octave sound level meter, with user initiated storage of screen data.
Instruments in Series 3 can measure noise levels, calculate the commonly required time averaged parameters and save the results.
The standard memory is capable of storing about 200 000 data points, which is equivalent to 6 000 data screens while the optional extended memory can store approximately 1 000 000 data points, equivalent to 30 000 data screens.
Depending on model they can also analyse and store noise data in broadband and octave bands.
With the exception of the CEL-553 which has only SLM mode, the standard versions of these instruments can be operated in three different ways (measurement modes) as follows.
¤
Real time sound level meter with user
initiated storage of screen data.
¤
Event sound level meter with manual,
remote or threshold triggering
(CEL-593 only).
¤
Environmental sound level meter with
user selected sampling periods
(CEL-573 and CEL-593). The measurement parameters available on standard versions vary from model to
model and from mode to mode as shown in Table 1 on page 73.
To simplify measurement procedures, pre-programmed setups are stored by the instrument. There is a separate directory for each application, mode and bandwidth, each of which contains one initial setup provided by the factory, and up to six more defined by the user. Similarly, stored data can be accessed via separate directories for each bandwidth, or via an additional directory that includes ALL data.
The following additional measurement applications can be installed.
¤
Fastore capability that samples and
saves data at very short time intervals.
¤
Building Acoustics.
¤
Loudness Determination. Each instrument in the series can be up- (or down-) graded to any other model, to include more (or fewer) features. Consult your local agency or CEL for details.
Series 3 instruments with Version 7 and later firmware feature a revised startup sequence where Application, Acquisition Mode and Bandwidth are selected via separate menu screens, plus the option of measuring sound exposure level (SEL) in broadband SLM and a revised keypad appearance.
For a description of all CEL-500 series instruments, also refer to the Reference Handbook.
Getting Started - CEL-553/573/593 - Page 59
Page 2
2. Menus & Controls
Menus and results are shown on a high definition screen in both alphanumeric and graphical formats, while operator commands are exercised via the keypad in association with menu options shown on the instrument display. The earlier and later keypads are shown here for easy series identification. The keys have exactly the same functions despite a different appearance.
Figure 1 (opposite) shows a typical screen display with the currently selected option indicated by the reverse video cursor field. Options in the four lowest boxes on the screen are mirrored by the four areas on the lowest part of the keypad which allow the respective option to be implemented.
On most screens, the shallow boxes immediately above the four option boxes indicate the current instrument bandwidth, measurement or recall mode, or real time measurement status.
ON/OFF
980007e
Series 3 Keypad
Earlier Series 1 and 2 Keypad
3. Installing Microphone, Preamplifier, Interface Module & Batteries
Warnings !
To prevent DAMAGING the microphone, the 200 V polarizing supply must be OFF when a CEL-250 or MK 250 Electret Microphone is used. DO NOT remove the protective grid from Type 1 microphones as this will expose the diaphragm to possible damage. DO NOT fit microphones or preamplifiers to the instrument with the power ON. Screw microphones and adaptors only finger tight. Do not use pliers or other tools. Repairs of damage caused by a failure to observe the warnings in this book will not be covered by the normal warranty conditions.
If a CEL-250 or MK 250 microphone is supplied and you dont know whether the polarizing voltage is OFF,
anything else
Chapter 4.
When the polarizing voltage is
to be OFF
supplied, install microphone, preamplifier, interface and batteries as follows.
1. With the power OFF, locate the preamplifier in the guides (Figure 2) and push firmly into place.
2. Screw the CEL-192 Type 1 Microphone on to a CEL-527 or CEL-525 Type 1 Preamplifier.
Or:
already OFF, screw the CEL-250 (or MK 250) Type 1 Microphone on to a CEL-527 or CEL-525 Preamplifier.
Or:
Microphone on to a CEL-530 Type 2 Preamplifier.
, check it as described in
, or some other microphone is
ONLY when polarizing voltage is
Screw the CEL-292 Type 2
before doing
Esc
EXIT
known
Page 60 - Getting Started - CEL-553/573/593
Page 3
Indicates set up loaded by
C a i
default when switched ON
Non-deletable set up provided by factory
Reverse video cursor shows selected item
Current retrieve mode/ status/record number will be shown here
Current status/screen title/narrow band cursor status/message icons
Battery condition, icon empties as power is drained. When empty at 6.4 V, icon blinks
Overload has occurred: is removed from SLM screens after 2 s, remains on EVT, ENV and FST screens
Pause has occurred during measurement or PAUSE key pressed
Event capture is following an event
Fastore is following an event
SETUP 1:
INITIALSETUP
2:
22-JUN-98
3:
25-JUN-98
4:
26-JUN-98
5:
EMPTY
6:
EMPTY
7:
EMPTY
LOAD
ursor keys to
ccess next screen
n same level or run
STORE DELETE REVIEW
Cursor keys to access next level or run
-
/
SLM
-
Lights display backlight for 30 s
STOP
Esc
DisplayIconsandScreenOptions
Timer is active
Current application, bandwidth, or mode:
-/­1/1 1/3 SLM EVT ENV FST RTC RT TMS RCS LOU
Current measurement status: STOP WAIT SYNC RUN PAUSE RECALL
Current box options
Escape from current screen
Keys that implement current box options
Broadband Octave band Third octave Real time sound level meter Event (CEL-593 only) Environmental (Not CEL-553) Fastore operations Rev. time- continuous Rev. time- Impulse
I
Transmit side measurement Receive side measurement Loudness determination
Measurement but no storage Waiting for timer Synchronising with clock time Measurement plus storage Run paused Recalling stored data
Figure1: Series 3
Thisscreenisobtained (after checking
themicrophonetype) bypressing
,USEMODE& SETUPxxxx
980006e
Cursor keys to select options within screen
Getting Started - CEL-553/573/593 - Page 61
Page 4
Or:
Screw the MK 301 Microphone on to the A67 Adaptor supplied with it, then screw the adaptor on to a CEL-527 or CEL-525 Preamplifier. Although Types 1 and 2 preamplifiers have a similar microphone thread, other physical limitations prevent operation with an incorrect pairing.
3. Locate the interface module in the guides (Figure 2) and push firmly into place.
4. When an active interface is to be used, install it before switching ON otherwise it will not be recognised by the firmware.
Warning !
LOAD BATTERIES in the correct orientations shown inside the battery compartment.
940028
One or two cells installed with the wrong polarity can still allow the instrument to function, but will provoke overheating severe enough to rupture a cell, with consequent risk to the operator and damage to the instrument.
5. Insert 8 standard AA size batteries in the orientations shown inside the battery compartment (and Figure 2). It is recommened that alkaline IEC Type LR6 be used.
6. Proceed to Chapter 5 to try simple measurements, or to Chapter 6 for acoustic calibration.
Figure 2: Installing batteries,
preamplifier and interface
+
-
-
+
-
+
+
+
-
-
­+
4. Checking Polarizing Voltage & Microphone Type
When a CEL-250* (or MK 250) Electret Microphone is fitted and you do not know whether the polarizing voltage is OFF,
before doing anything else
as follows.
Note*:Microphones are identified by a name
engraved on the side of the sleeve.
1. If the microphone is a CEL-250 (or an MK 250) carefully unscrew it from the preamplifier.
2. Locate the preamplifier in the guides (Figure 2) and push firmly into place.
3. Observing the battery warning in Chapter 3, insert eight standard AA batteries in the orientations shown
Page 62 - Getting Started - CEL-553/573/593
, check it
980008e
PreviousMode
Previous
Screen
Backlight
"Series3 & 2"
NextMode
Setup/Review - ScreenOptions
inside the battery compartment (and in Figure 2). It is recommened that alkaline IEC Type LR6 be used.
Select Screen Entries
Next Screen
Next/ Previous Value
Page 5
4. Press the power key to switch the instrument ON.
5. Wait for the self test screens to clear and show the Configuration screen.
CONFIGURATION
CEL-593.C1M
I
Type 1
Time 16-Jun-98 Ø9:33:31 INSTRUMENT ID (DPB) = NNNNNN CURRENT MODE = SLM SELECT CHOICE FROM BELOW SET
CLOCK
APPLI­CATION
IEC 651 IEC 8Ø4Type 1
USE MODE
:1979 :1985
-
/
-
CAL
GS-01
M suffix indicates that Fastore,
Building Acoustics, Loudness
Determination and Standard
Application (SLM) are available.
R suffix indicates that Fastore,
Building Acoustics and Standard
Application (SLM) are available.
L suffix indicates that Fastore,
Loudness Determination and
Standard Application (SLM) are
available.
T suffix indicates that Fastore
and Standard Application (SLM) are
available.
No suffix indicates that only
Standard Application (SLM) is
available.
and bandwidth, which indicates the current battery voltage.
7. If the voltage shown for a NEW SET of batteries is only 9-10 V, check battery orientation.
8. Press OPTION to obtain the Options screen.
OPTIONS DELETE MEMORY CALIBRATE SET CLOCK MIC. RESPONSE CONFIGURATION
SET PRINTER LCD CONTRAST SET UP RS 232 POL. VOLTS SET RUN I. D.
-
/
STOPSLM
-
SELECT
GS-03
POL. VOLTS option is not available
while a Type 2 preamplifier is fitted
9. Use or to move the cursor to the POL.VOLTS option.
10. Press SELECT key to display the Select Polarizing Voltage screen which shows the current status.
SELECT POLARIZING VOLTAGE
200 VOLT POL. VOLTAGE FOR
AIR CONDENSER MICROPHONES
IS
ON
6. Press the USE MODE key to display the Status screen for the (last used) mode
STATUS : STOP 16-JUN-98 BATTERY MEMORY RUN START TIME RUN ELAPSE TIME APP: SOUND ANALYSER MODE: BROADBAND SLM
-
/
SLM
-
RECALL
OPTION SETUP REAL
Ø9 : 35 : 15
1Ø.8Ø VOLTS
Ø% FULLØRUNS
-
/
SLM
-
TIME
GS-02
CHANGE
GS-04
11. If the polarizing voltage is OFF, go to step 12, if it is ON, press the CHANGE key to switch it OFF.
Battery Voltage
Ø Runs with N% Full indicates
memory in use by another
Application
Getting Started - CEL-553/573/593 - Page 63
Page 6
Also use this screen to switch the polarizing voltage ON for condenser microphones.
12. Press power to switch the instrument OFF before re-installing the microphone.
Note: The FORMAT option on the last System
Test screen should be treated with great respect ! It purges all stored data and set ups, except for the initial set ups provided by the factory.
5. Trying Simple Octave Band Measurements
980009e
Previous
Screen
(Same Setup)
Backlight
"Series3 & 2"
PreviousData
StoredinThisRun
(WhenAvailable)
NextData
StoredinThisRun
(WhenAvailable)
Change Frequency
Next Screen (Same Setup)
Commands
Measurement- ScreenOptions
Observe the Warnings in Chapter 3.
With microphone type identified and polarizing voltage checked (Chapter 4), with batteries, preamplifier, microphone and interface installed (Chapter 3), try some simple measurements with a B or C version instrument.
1. Press the power key to switch ON.
2. Wait for the Start up screens to clear
and display the Configuration screen.
3. When the Current Mode is shown as
1
⁄1SLM, press the USE MODE key to
display the Status screen for this bandwidth and proceed to step 8.
4. When the Current Mode is not shown as
1
⁄1SLM, press the APPLICATION key
to display the Select Application screen. (The L suffix example also shows Fastore and Loudness.)
SELECT APPLICATION
1 SOUND ANALYSER 23FASTORE
BUILDING ACOUSTICS
4
LOUDNESS
Use to select
CEL-593.C1L
STOP
LOAD
5. Use and to move the reverse video cursor to SOUND ANALYSER, then press the LOAD key to load this Application and display a similar Select Acquisition Mode screen.
SELECT ACQUISITION MODE
1
SOUND LEVEL METER
2
ENVIRONMENTAL
3
EVENT
Use to select
CEL-593.C1L
STOP
LOAD
6. Use and to move the reverse video cursor to SOUND LEVEL METER, then press the LOAD key to
GS-05
SLM ENV EVT
GS-06
Page 64 - Getting Started - CEL-553/573/593
Page 7
SELECT BANDWIDTH
CEL-553.C1T
-
1BROADBAND 23OCTAVE BAND
THIRD OCTAVE BAND
/
SLM
-
1
/
SLM
1
1
/
SLM
3
Use to select
STOP
LOAD
GS-07
load this Application and display a similar Select Bandwidth screen. This identifies bandwidths as
1
/1octave band and1/3third
band,
-
/-broad-
octave band.
7. Use and to move the reverse video cursor to OCTAVE BAND
1
⁄1SLM, and press the LOAD key to
load the indicated bandwidth for use and display the Status screen while using the Initial Set Up that will have been loaded automatically.
8. If the voltage shown for a NEW SET of batteries is only 9-10 V, check battery orientation.
9. Press the REAL TIME key to display the octave band measurement screen.
L
250Hz
1
/
SLM
1
56.4
STOP
RUN
GS-08
AL
F
RANGE
SPLF
RANGE
Within an Application,
Acquisition Mode and Bandwidth
can also be changed by using
& to change the identity
shown in the Application Box
Frequency and level of marked
band shown here
Measurement commences with the settings Lin plus A-weighted and SPL Fast for narrowband. Data storage is not possible without further action.
10. If necessary, use RANGE keys to give the correct dynamic range for the measurement.
11. Place the instrument near a noise source and observe the movement of the real time octave band columns.
12. Use and keys to obtain a readout of the octave band levels.
13. Press RUN key to enable data storage.
250Hz
AL
F
11:16:31
STOP
SPLF
1
RECORD
1
/
SLM
1
DATA
L
56.4
RUN
RESET
GS-09
14. Press RECORD key to store a snapshot of the spectrum.
15. Observe that a DATA box has been added to indicate that stored data is available.
16. Change the noise or the instrument position and store further spectra.
17. Press DATA key to see the first stored spectrum.
18. Press to see the next stored spectrum.
19. Press STOP key to leave Run status.
20. Press ESC to return to the Stop Status screen.
STATUS : STOP 16 - APR - 98 BATTERY MEMORY
1Ø : 14 : 13
1Ø.8Ø VOLTS
1% FULL1RUNS RUN START TIME RUN ELAPSE TIME APP: SOUND ANALYSER MODE: BROADBAND SLM
RECALL
-
/
SLM
-
OPTION SETUP REAL
-
/
SLM
-
TIME
GS-10
21. Press to move to the recall side of the display. With Version 7 or later firmware, the cursor will already be on the recall side of the display.
Getting Started - CEL-553/573/593 - Page 65
Page 8
22. Press until1⁄1SLM is indicated in the left hand cursor field.
23. Press RECALL key to display the
1
⁄1SLM store Directory.
24. Use to select the date and time of the record just stored.
25. Press to display the Header screen for the run.
26. Press to display the Recall data screen 1 for the first spectrum.
27. Press to display the Recall data screen for the second spectrum.
28. Press ESC to return to the Status screen.
29. Press power to switch the analyser OFF.
You have now operated the instrument, using all of the steps required to store and recall data.
These are the steps required to
perform a simple noise survey.
Easy wasnt it ?
6. Field Accuracy Check (Acoustic Calibration)
Setting the clock and performing a calibration can be carried out in any Application and will be valid for all Applications.
With batteries, preamplifier, microphone and interface installed according to Chapter 3 above, proceed as follows.
1. Press the power key to switch the
instrument ON.
2. Wait for the self test screens to clear
and show the Configuration screen.
3. When the date and time are correct, go
to step 8.
4. When the date and time are incorrect
use the SET CLOCK option to display
the Set Clock screen.
5. Set the date and time using cursor and
value keys.
6. Press the SET key.
7. Press ESC key to return to the
Configuration screen.
8. Use the CAL option to display the first
Calibrate screen.
Page 66 - Getting Started - CEL-553/573/593
The Calibration screen can also be reached via the CAL option on the Options screen, see Chapter 10.
9. When the previous calibration is acceptable, use the ABORT option and go directly to step 14. When the previous calibration is not acceptable, use the CAL option to produce the second Calibrate screen.
10. Fit a CEL-284/2 Calibrator over a Type 1 microphone and either a CEL-282 or a CEL-284/2 Calibrator over a Type 2 microphone, making sure the microphone is correctly seated against the shoulder in the cavity (Figure 3).
11. Switch the calibrator ON. These calibrators generate a nominal level of 114.0 dB at 1 kHz.
12. For the CEL-192, CEL-250 and MK 250, if necessary use the LEVEL keys to adjust the sensitivity until 114.0 dB is indicated. For the MK 301, adjust the indicated level to 94.0 dB.
Page 9
e
CALIBRATION
LIN LEVEL FIT CALIBRATOR AND USE
PRESS SAVE-CAL TO SAVE AND
USE NEW SETTING OR PRESS
ABORT TO USE LAST SETTING
STANDARD CALIBRATION
ABORT
SAVE
more
CAL
DIRECTION
65.Ø dB
TO SET LEVEL
LEVEL LEVEL
GS-11
For the CEL-292 (due to differences in cavity volume), adjust the indicated level to 113.7 dB.
13. Once the correct level is indicated, use the SAVE CAL option to store the calibration for future use.
14. Press ESC key to display the current Status screen to start measurement and logging, then follow the instructions in Chapter 7.
Calibrator
Shoulder
'O' Ring
Microphone & Preamplifier
900049
Figure 3: Fitting acoustic calibrator
7. Practical Measurement, Logging & Printout
980009e
Previous
Screen
(Same Setup)
Backlight
"Series3 & 2"
Measurement- ScreenOptions
Observe the Warnings in Chapter 3.
With batteries, preamplifier, microphone and interface installed as detailed in
PreviousData
StoredinThisRun
(WhenAvailable)
NextData
StoredinThisRun
(WhenAvailable)
Change Frequency
Next Screen (Same Setup)
Commands
Chapter 3, a field accuracy check performed according to Chapter 6, and the Configuration screen displayed, proceed as follows.
1. When the Application, mode and bandwidth shown in the status field are suitable for the proposed measurement task, press the USE MODE key to display the Status screen for the mode and bandwidth, then proceed to step 6.
2. When the Application and bandwidth shown in the status field, are not those required for the proposed SLM measurement task, press the APPLICATION key to display the Select Application screen.
Getting Started - CEL-553/573/593 - Page 67
Page 10
3. Use and to move the reverse video cursor to SOUND ANALYSER, then press the LOAD key to load this Application and display a similar Select Acquisition Mode screen (shown in Chapter 5).
SELECT APPLICATION
1 SOUND ANALYSER 23FASTORE
BUILDING ACOUSTICS
4
LOUDNESS
Use to select
CEL-593.C1L
STOP
LOAD
GS-05
4. Use and to move the reverse video cursor to highlight the required Mode. then press the LOAD key to load the Mode and display a similar Select Bandwidth screen (see Chapter 5).
5. Use and to move the reverse video cursor to highlight the required bandwidth, then press LOAD to load the default setup for the indicated bandwidth and display the relevant Status screen.
6. If the voltage shown for a NEW SET of batteries is only 9-10 V, check battery orientation.
7. When the instrument is switched ON, once the self test is completed, the default setup ( in Figure 1) for each mode and bandwidth is loaded automatically for immediate use.
8. When the default setup is known to be suitable for the current task, go directly to step14.
9. When it is not known if the default setup is suitable, use the SETUP option to display the current Set Up screen and directory (see Figure 1).
10. When none of the listed setups are suitable, use the REVIEW option on the Set Up screen, edit an existing setup as described in Chapter 11 before using it, then go to step 13.
-
/
SLM SETTINGS
-
LIN
'A'
BARGRAPH(S) FAST RECORD TIMER= OFF
Q=3 Q=
_
PEAK= LINPEAK
more
-
REVIEW
/
-
PREV VALUE
FAST SLOW
STOPSLM 1
NEXT VALUE
GS-12
Indicates further parameters on
other screens accessed by
and
11. When the default setup which has been loaded for the mode (indicated by
), is found to be suitable, go directly
to step 13.
12. When one of the other listed setups for the bandwidth is suitable, use or keys to select it followed by the LOAD option to load it for use.
13. Press ESC key to display the current Status screen
14. Press the REAL TIME key to display the real time Stop screen.
L
250Hz
1
/
SLM
1
56.4
STOP
RUN
GS-08
AL
F
RANGE
SPLF
RANGE
15. If necessary, press RANGE keys one or more times to obtain the correct range for the task.
16. Press the RUN key to start calculation and logging. All data from the run is stored in a file that is identified by its start date and time. This file is held in the directory of results for the current mode and bandwidth.
Page 68 - Getting Started - CEL-553/573/593
Page 11
Sound Level Meter
modes offer the option of storing a record with preset duration or an instantaneous snapshot of the values by means of a RECORD key, and will also store a snapshot automatically at the end of a run if the RECORD key has not been used.
250Hz
F
11:16:31
STOP
In
Event
SPLF
AL
1
RECORD
modes, data is stored only
1
/
SLM
1
DATA
L
56.4
RUN
RESET
GS-09
after the threshold conditions have been met, or event storage has been initiated manually, or by remote signal via a CEL-503.
Environmental
modes store data automatically at the end of each set measurement period and run. Any snapshot or other timed data taken during a run will be stored with the run data file and identified 1, 2, 3, etc in sequence.
17. In SLM mode, press the RESET key to reset the parameters on display so that new accumulated totals or maxima can be determined.
18. In SLM mode, press the RECORD key to store a snapshot or timed record. (The key may be used again to store further snapshots or timed records.)
19. In any mode, inspect stored data from the current run by pressing the DATA key which is displayed ONLY when stored data is available.
20. In any mode, when a CEL-501 is fitted and connected to a Centronics printer
using a C6623/2 Cable (with the printer on-line), print the current screen by pressing the PRINT key.
250Hz
AL
F
11:16:31
STOP
SPLF
1
RECORD
1
/
SLM
1
REAL TIME
L
56.4
RUN
PRINT
GS-13
21. After inspecting and printing data as it is collected, press REAL TIME to return to the Real Time Run screen.
22. When the measurement run is complete or the required logging period has elapsed, press the STOP key to terminate the run.
23. When a further run with the same setup is required, start it by pressing the RUN key again and following the instructions from step 17.
24. When operation with the same mode and bandwidth but another setup is required, use SETUP and REVIEW options (step 7 and Chapter 11) to obtain a different setup, then follow the instructions from step 13.
25. When operation with a different mode or bandwidth is required, press ESC key to return to the Status screen, use
or to change the mode and bandwidth, use SETUP and REVIEW options (step 3 and Chapter 11) to obtain the required setup, then follow the instructions from step 13.
26. When data retrieval operations are required, press Esc key and go to step 4 in Chapter 8.
27. When no further operations are required, press power key to switch the instrument OFF.
Getting Started - CEL-553/573/593 - Page 69
Page 12
8. Retrieving Stored Data
980010e
Previous
Screen
(Same
Record)
Backlight
"Series3 & 2"
PreviousRecord
(SameScreen)
NextRecord
(SameScreen)
Change Frequency
Next Screen (Same Record)
Next/Previous Run
Recall-ScreenOptions
To retrieve stored data, it is not necessary to have the preamplifier, microphone and interface installed.
However, when data is to be downloaded, the relevant interface must be attached (see Chapter 3).
1. Switch the instrument ON.
2. Wait for the self test screens to clear
and the relevant Configuration screen
to be displayed.
3. When the cursor is on the Measure
(right side) Application Box, use to
move it to the Recall (left side)
Application Box.
4. When the required data was stored
using a different Application from that
currently loaded, change the Appli-
cation as described in Chapters 5 or 7.
5. When the required data was stored
using the current Application, press
the USE MODE key to obtain the
Status screen for the (last used) mode
and bandwidth for this Application.
Version 7 and later firmware will
highlight the recall field.
6. Press or keys sufficient times to
display the required recall mode and
bandwidth in the cursor field.
7. Use the RECALL option to display the
Directory screen for the bandwidth.
8. Press or to move the cursor to the
required entry, identified by date and
time.
9. Press to display the Header screen
for the identified run showing a
summary of data stored.
HEADER
START END LENGTH TIMED RECS
MIC=FF 2ØØV=ON TYPE=1 F, S L, A Q 3
PRINT
LAST CAL 16-APR-98 16-APR-98 1Ø:13:23 16-APR-98 1Ø:44:38
Ø DAYS Ø:31:16
4
-
/
RECALL
SLM
-
PREV NEXT RUN RUN
GS-14
10. Press PRINT to obtain a printout of the Header screen on a Centronics printer connected via a CEL-501 and C6623/2 Cable.
Note: The PRINT option becomes available
only when the CEL-501 was fitted before the instrument was switched ON and the printer is on-line.
11. Press (or ) to display the first (or last) data screen from the identified run, and press PRINT to obtain a printout of the screen.
12. Continue to press (or ) to display the remaining data screens from the run, pressing PRINT to obtain printouts of the screens if required.
13. Press (or ) to display the equivalent screens from the next (or previous) timed record in the same run, pressing PRINT to obtain printouts of the screens when required.
250Hz
L
1
/
SLM
1
PREV NEXT RUN RUN
56.4
RECALL
GS-15
F
PRINT
SPLF
AL
111:16:31
14. When all data from this run has been displayed and printed, use the PREV RUN or NEXT RUN options to obtain the header for the previous or next run
Page 70 - Getting Started - CEL-553/573/593
Page 13
in this store directory, then resume the instructions from step 8.
15. When all required data from this directory has been displayed and printed, press ESC key to return to the Status screen.
16. When data from another bandwidth is to be displayed, use or keys
sufficient times to display the required store directory and follow the instructions from step 8.
17. When further measurement and logging are required, refer to Chapter 7.
18. When no further operations are required, press power to switch the instrument OFF.
9. Performing Full Data Download
A full download of all data stored in the instrument is possible only when a CEL-502 RS232 Interface is fitted to the instrument and connected via C6621/2 or C6622/2 Cables to a PC running application software.
The following Windows programs are
available for Series 3.
dB1 (CEL-6694) Download and Control Software for ALL Applications, including Loudness. This program has no graphing capabilites but can export data to other standard software packages for graphing and printing if required.
dB2 (CEL-6695) Software for Sound Analyser Application with comprehensive graphing and data manipulation fqcilities.
dB3 (CEL-6696) Software for Fastore Application with comprehensive graphing and data manipulation fqcilities.
dB4 (CEL-6697) Software for Building Acoustics Application with comprehensive graphing and data manipulation fqcilities.
Downloading takes place under the control of the relevant software, and the
10. Setting Instrument Options
SETUP RS 232
BAUD RATE 192ØØ
PREV VALUE
only operations to be performed on the analyser are to switch it ON and let it pass the Self test and Calibration screens.
The baud rate is set via the SET UP RS232 screen accessed from the SET UP RS232 option on the Options screen. This screen becomes available only when the CEL-502 was fitted before the instrument was switched ON.
Make sure the baud rate settings on the instrument and the communication port options in the software are the same to achieve reliable transfer of data.
NEXT VALUE
GS-16
The following instrument options can be set from the Options screen, which is obtained by using the OPTION key available on any Status screen.
Most of these options can be selected ONLY while in STOP status to avoid corrupting any measurement in progress.
DELETE MEMORY Allows blocks of stored data to be deleted. It may take a considerable time to delete large data blocks.
CALIBRATE Allows a field accuracy check (acoustic calibration) to be performed, see Chapter 6.
Getting Started - CEL-553/573/593 - Page 71
Page 14
SET CLOCK
Allows time and date to be set.
MIC. RESPONSE Allows microphone response to be changed between free field to meet IEC standards and random incidence that meets ANSI standards.
CONFIGURATION Displays the instrument configuration.
LCD CONTRAST Allows the display contrast to be changed.
SET PRINTER Allows the printer output format to be specified as Epson dot matrix or Laserjet compatible so data displayed on the screen can be printed via a CEL-501 Parallel (Centronics) Interface (Chapter 8).
SETUP RS232 Allows the RS232 baud rate to be speci- fied so that stored data can be transferred over a CEL-502 RS232 Interface.
11. Editing Stored Set Ups
OPTIONS DELETE MEMORY CALIBRATE SET CLOCK MIC. RESPONSE CONFIGURATION
SET PRINTER LCD CONTRAST SET UP RS 232 POL. VOLTS SET RUN I. D.
-
/
STOPSLM
-
SELECT
GS-03
POL. VOLTS Allows the 200 V polarising voltage required for condenser microphones to be switched OFF to permit electret microphones to be used (Chapter 4).
RUN I.D. Allows an identity code consisting of a single letter and three numbers to be assigned to each run by means of NEXT VALUE and PREV. VALUE keys.
When switched ON, the set ups marked
will be loaded for each measurement
mode and bandwidth, for immediate use.
Up to seven set ups can be stored for each available bandwidth and mode, giving the following maximum numbers for an instrument without additional applications.
CEL-553: 7 x 3 x 1 = 21,
CEL-573: 7 x 3 x 2 = 42,
CEL-593: 7 x 3 x 3 = 63. Table 1 on page 15 lists the bandwidths, operating modes and measurement parameters available for all instruments in the CEL-500 series.
Each bandwidth has its own directory, where the first entry is an INITIAL SET UP provided by the factory (Figure 1) which can neither be deleted nor overwritten.
Any stored set up can be edited and the changed version used with or without saving. However, if an edited version of the initial set up is to be saved, it must be given a new identity. Set ups other than the initial set up may be overwritten.
Page 72 - Getting Started - CEL-553/573/593
980008e
Previous
Screen
Backlight
"Series3 & 2"
PreviousMode
NextMode
Select Screen Entries
Next Screen
Next/ Previous Value
Setup/Review - ScreenOptions
When a set up has been modified for use with no intent to save it, do not press the LOAD key as this will reload the un-modified version.
With the instrument in Stop mode (i.e. not logging data), starting from a Status screen, edit and store a set up as follows.
1. Press or keys sufficient times to display the required mode and bandwidth in the Application Box.
Page 15
Table 1: Bandwidths
Operating Modes &
Measurement Parameters
CEL-5X3.C1/C2 (
CEL-5X3.A1/A2 (
Broad Band Octave & Third Octave
−
−
⁄
)
−
Real Time Sound Level Meter
Freq Weightings
Time Weightings Exch Rate Q Parameters
2 (from Lin, A & C)
2 (from Slow, Fast & Impulse) 2 (from Q=3, 4, 5 & 6) SPL Max & Min
& IL
L
eq avg
(with Q=3)
epd
Tm3<m5
eq
L LinPeak/C Peak L TWA (with Q=5 & Slow wtg) SEL L
Environmental Mode (Not CEL-553)
Freq Weightings
Time Weightings Exch Rate Q Parameters
Meas Period Durations
Auto Start/Stop Times Ln values Profiles
1
2 (from Lin, A & C)
1 (from Slow, Fast & Impulse) 1 (from Q=3, 4, 5 & 6) SPL not stored Max & Min SEL of period
& IL
L
eq avg
Tm3<m5
eq
2
or 1 sec L
eq
L LinPeak/C Peak L
0.5, 1, 5, 10, 15, 30 sec 1, 5, 10, 15, 20, 30, 60 min
24 sets 5 1 SecMax
A cumulative result set and set of histograms are ALWAYS stored
Event Mode (CEL-593 Only)
Freq Weightings Time Weightings Exch Rate Q Parameters
Profiles Triggering
Threshold Triggering Delays
1 (from Lin, A & C) 1 (from Slow, Fast & Impulse) Fixed at Q=3 Max Level Time of Max SEL Duration Event L
eq
LinPeak/C Peak 1 SecMax
Manual or Threshold + Delay 40 to 139 dB 1 to 30 sec
1
1
⁄
⁄
⁄
) B1/B2 (
−
1
3
1 x Broad & Narrow Band 1 x Optional Broad Band (from Lin, A & C) 1 (from Slow, Fast & Impulse) Fixed at Q=3 SPL Max L
eq
1 x Broad & Narrow Band 1 x Optional Broad Band (from Lin, A & C) 1 (from Slow, Fast & Impulse) Fixed at Q=3 SPL not stored Max L
eq
0.5, 1, 5, 10, 15, 30 sec 1, 5, 10, 15, 20, 30, 60 min
24 sets 5
1 (from Lin, A & C) 1 (from Slow, Fast & Impulse) Fixed at Q=3 Max Level Time of Max SEL Duration Event L
eq
LinPeak/C Peak 1 SecMax Spectrum Profiles Manual or Threshold + Delay 40 to 139 dB 1 to 30 sec
1
−
⁄
⁄
−
1
)
Note 1: With Series 3 & 2" instruments, period Ln values are calculated when a run is≥1 minute
in length. (Series 1" instruments need at least 4 minutes for broadband and 8 minutes for narrow band.) Cumulative Ln values will be available for display as soon as a broad band run≥5 seconds duration or a narrow band run≥10 seconds duration is complete.
Note 2: Available only with periods of 5 minutes or longer.
Getting Started - CEL-553/573/593 - Page 73
Page 16
2. Press the SETUP key to display the relevant Setup screen (Figure 1) that offers the following options. LOAD: Loads the highlighted set up for use and sets against the entry for default loading next time the instrument is switched ON. STORE: Stores the current set up in the highlighted space and sets . DELETE: Deletes the highlighted entry. If it is in current use, it remains active until another is selected. REVIEW: Displays the current set up.
3. Use or to select an entry.
4. Press REVIEW to access a setup for editing. Table 2 suggests settings for some typical measurement tasks.
5. Use cursor keys to select entries and VALUE keys to edit them.
6. When the more legend is shown, access further relevant screens by pressing or .
Table 2: Typical Measurement Tasks
Task Suggested Settings
Determine Noise Exposure of Employees Working Near an Impact Press
Measurements to Aid Selection of Hearing Protection Suitable for the Employees Above
Reduce Noise Levels in Location Adjacent to a Noise Source
Perform Annual Octave Band Noise Survey in Factory at 102 Locations
Investigate Intermittent Industrial Noise Disturbing a Quiet Neighbour­hood
Police Cumulative Noise Limits Set for a Quarry
Determine the Effect of Running New Trains Through Quiet Country­side
Use broad band ENV with 1 second period and EVT enabled, Measure Leq and LinPeak, Inspect period profiles, Check events.
Use Store SPLmax spectra followed by L Apply data to protector selection chart.
Use Compare stored SPL spectra from source with SPL spectra from location, Check for noise leaks/improvements.
Use Store L
Use enabled, Measure L background noise, Check for events.
Use broad band ENV with 60 minute period, Measure L L
95
Use enabled, Measure L as L Check event frequency profiles, Use data to find projected daily L
1
/
SLM
3
LIN FAST
(2nd BARGRAPH ONLY)
'A'
RECORD TIMER OFF
REVIEW
Q=3
=
more
1
/
3
PREV VALUE
SETTINGS
7. Select and edit entries as in 5 and 6 above.
8. When all entries have been made, press ESC key to recall the Set Up Directory.
9. Select a store location (This must be done).
10. Press STORE key to save the edited version.
1
⁄
octave band real time SLM,
1
1
⁄
octave band real time SLM,
3
1
⁄
octave band real time SLM,
1
spectra snapshots at each location.
eq
1
⁄
octave band ENV with 10 minute period and EVT
3
and determine Ln spectra values such as L
eq
, SPLmax and determine Ln values such as
eq
background noise.
1
⁄
octave band ENV with 10 minute period and EVT
3
, SEL and determine Ln spectra values such
eq
background noise,
95
spectra,
eq
from train frequency.
eq
STOPSLM 1
NEXT VALUE
GS-17
95
Page 74 - Getting Started - CEL-553/573/593
Loading...