Casella CEL CEL-633 User Manual

1
3
2
4
Figure 9. Memory screens
Taking Measurements (Refer to Figure 10)
The measurement options and screen details will vary, depending on the instrument type you are using.
The instrument must be in the Measurement Stop mode (with red bars at the top and bottom of the screen) before you can start a measurement.
Press the Run/Stop key ( ) to start making measurements. The screen shows green bars at the top and bottom.
D
Figure 10. Example Stop and Run screens
Run Screens
C
4
;
A
Figure 11. Graph and Main screens
B
B
max eq
B
D
The Play symbol is displayed at the top left-hand corner of the screen, and the Pause symbol is displayed at the bottom left-hand corner of the screen. The left­hand soft key ( ) is the Pause/Run control and toggles between these two modes when operated.
The screen displays PAUSED! when measurements are paused. The incremental run number and its duration are displayed at the top right-hand corner of the screen. If the Run Duration Timer is set, an icon that shows a stopwatch is displayed next to the countdown timer. To enable or disabled the Run Duration Timer, use the Settings screens (refer to Figure 6). If the Run Duration Timer is not set, the digital timer clock counts up until you stop the measurement run manually (by pressing the Run/Stop
D
key ).
For instruments that include this option, the broadband values can be viewed as a bargraph or as a graphical time history on the Graph screen. Press the View key ( ) to toggle between the Graph and the Main screens.
Certain instrument types provide two additional Octave (or 1/3 Octave) screens (refer to Figure 12). Use the View key ( ) to toggle the screens. Use the left/right navigation keys ( ) to move across the frequency bands in the Graph view. Use the up/down navigation keys to toggle between showing L and L functions
Each bar in the graph represents an octave band measured in Hz or kHz. To view a table of the octave values, press the View key ( ).
In 1/3 Octave mode, use the up/down navigation keys to change the displayed frequencies on the table.
Some instruments in the CEL-63x range can display all the screens shown in Figure 11 and Figure12.
The Marker screen allows you to tag noise data, for example noise from a road vehicle, with a specific identity using one of four markers.
When the measurement run has finished, press the Run/Stop key ( ). A screen is displayed asking you to confirm the action - press Yes to end the run.
Figure 12. graph and list views1/3 Octave
Storage
Switch the CEL-63x instrument off when you are not using it to record measurements, and remove the batteries if the instrument will not be used for extended periods of time.
Always protect the instrument from physical damage and from water.
To protect the microphone, you should always keep the windshield fitted.
More information
For more information about any of the topics summarised in this Quick Start guide, refer to the User Manual.
Instrument types
Table 1 below lists the instrument types and shows the capabilities of each instrument in the range. Please remember that the features that are available to you and the screens that you can see depend on the type of instrument you are using.
Store cumulative data
Application
Occupational
Environmental
Occupational (with logging)
Environmental
(with logging)
Model number
CEL-630
CEL-631
CEL-632
CEL-633
Data markers
Table 1. CEL-63x instrument types, applications and capabilities
Regent House, Wolseley Road, Kempston, Bedford, MK42 7JY, U.K.
E-mail for technical support: techsupport@casellacel.com
For online Technical Support: http://supportdesk.casellacel.com
Document number FG40-03 © 2010 Casella Measurement
CASELLA MEASUREMENT
Phone: +44 (0) 1234 844 100
Fax: +44 (0) 1234 841 490
Logging of Time History
Run duration timers
Automatic On/Off timers
Statistical parameters (Ln%)
Field Guide
CEL-63x Series Sound Level Meters
1
2
3
4
B
5
6
1. Windshield (covering the removable microphone)
2. Fixed amplifier
3. On/Off key
4. Display
Figure 1. The CEL-63x series
Introduction
The guide assumes this is the first time you have used the instrument. It therefore includes instructions to install the batteries, switch ON and set up the instrument to be ready for use, calibrate it, take measurements, and store it safely.
We want you to get the best possible performance from the CEL-63x instrument, and we recommend you refer to the Operator’s Manual for detailed information.
A
5. Soft keys and
6. Navigation keys
7. Run/Stop key
A B
C
D
7
Fit the batteries
The CEL-63x instrument requires three AA batteries.
Remove the battery compartment cover and fit the batteries using the correct polarities. Refit the battery compartment cover.
Note: The batteries can be zinc carbon, alkaline or rechargeable. DO NOT mix battery types.
Switch the instrument ON
Press the On/Off key. Confirm the battery condition symbol shows there is a good level of charge in the batteries.
The initialisation screen is displayed for approximately 10 seconds, and then the measurement screen is displayed in the STOP mode (with red bars at the top and the bottom of the screen). Press the Menu key to access the settings, memory and instrument status. The screens displayed depend on the model of instrument in use.
Settings screen
MENU
1 2
3
SETTINGS
5
A BCA B
C
C
1. Settings 2. Memory results 3. Instrument status A. Exit / Back B. Select / Edit
C. Navigation keys
Figure 2. Settings screen
LANGUAGE
4
A B
C
C
Set language
Select Settings and press to continue. Use navigation keys to select the Language symbol . Press to continue. Use navigation keys to select the correct language. Press Select ( ). Press Exit ( ).
Set frequency modes
Some instruments in the CEL-63x range allow you to select either Octave or 1/3
Octave with which to take measurements. Select Settings and press to
continue. Use navigation keys to select the Octave symbol . Press to continue. Use navigation keys to select Octave mode. Press Select ( ). Press Exit ( ).
Set the time and date
Select Settings and press to continue. Use navigation keys to select the Set Clock symbol . Press to continue. Use navigation keys to select the time or date.
B C B
Press to edit. Use navigation keys to adjust time/date. Press Save ( ). Press Exit
A
( ).
(Refer to Figure 2)
1 B C
4 B C
B A
(Refer to Figure 2)
1 B
C 5 B
C B A
(Refer to Figure 3)
1 B C
2 B C
2
1
Calibration (refer to Figure 4)
Figure 3. Setting the time and date
Calibration is auto-set to the reference level as set within the Settings menu. Refer to the manual to set the calibration reference level.
1
BA A
Figure 4. Calibration
Select the Measurement Stop screen. Fit the acoustic calibrator fully over the removable microphone. Press the calibrator’s On/Off key . The CEL-63x instrument detects the signal and activates the calibration screen (with yellow bars at the top and bottom). Press to start the calibration. The instrument calibrates and displays the word PASSED when finished. Press Exit ( ) to return to the Stop screen. Press and hold the calibrator’s On/Off key to switch the calibrator off. Remove the acoustic calibrator.
Setup
(Refer to Figure 5)
1 B C
Select Settings and press to continue. Use navigation keys to select the Setup
2
symbol . Press to continue. Use navigation keys to select the required measurement view for your local legislation. Refer to the User Manual for detailed information about the measurement views.
Note: This instrument simultaneously records all noise functions for every measurement it completes. The Measurement View screens simply define which functions are displayed on the measurement screens.
1
B
B C
2
Figure 5. Settings options
1
A
Measurement control(Refer to Figure 6)
Select Settings and use the navigation keys to select the Measurement control
1 B
symbol . Press to continue. Use the navigation keys to select the Measure ­Start/Stop symbol . Press Edit ( ). Use the navigation keys to select the Mode (Key
Press / Fixed Duration / Timers). Press Save ( ). Use the navigation keys to select additional Fixed duration or Timers options. Press Edit ( ), and use the navigation keys to set the run duration. Press Save ( ) to continue (if required).
2 B
B
B
B
2
1
Measurement data sets (Refer to Figure 7)
Cumulative measurements made by the instrument apply for the entire measurement duration. They do not show how the sound level changed during the period of measurement. The instrument can also make Periodic measurements, which store measurements acquired during a repeating interval of time. When one interval ends, the next one begins until the periodic measurements stop.
The instrument’s Profile recording function provides an additional time-history recording channel, which may operate in parallel with either Cumulative or Periodic data recording. The Profile time history channel can provide faster recording resolution but for a reduced set of noise parameters.
Select Settings and use the navigation keys to select the Measurement control symbol . Press to continue. Use the navigation keys to select the Data Sets
1 B
symbol . Set the Periodic and Profile measurement intervals and press Exit.
2
Figure 6. Run duration
2
1
Figure 7. Measurement data sets
1
Figure 8. Memory results
Memory results
Press the Menu key to access the settings, memory and instrument status. Use the navigation keys to select the Memory Results symbol and press Select ( ) to continue. Use the navigation keys to select a result in the list, and press Select ( ) to select the measurement.
Use the navigation keys to select View Results (see Figure 9) and press Select ( ) to display the selected memory results. For additional data press View ( ). The screen displays detailed, read-only data for the selected measurement. Press View
B
( ) to see the data in a graphical format, using the Octave format saved in the Memory Results. Use the navigation keys to move the vertical line across the graph. Each graph bar represents an octave level measured in Hz or kHz. Press View ( ) again to view tabular octave results.
Use the navigation keys to toggle between functions such as L and L . Press View
B
( ) to view the results in tabular form. The results stored for the octave band graph display functions L and L .
Press Exit ( ) to leave the Memory Results, or press View ( ) to cycle the Memory Results screens again. Note that the display screens will vary according to the model of CEL-63x instrument you are using.
(Refer to Figure 8)
1 B
B
1 B
B
max Aeq
max eq
A B
Delete Memory Results
Refer to the User Manual for details and instructions to delete the results stored in the instrument’s memory.
B