LOAD BATTERIES following the instructions given in
Section 1.3. Make sure they are inserted in the orientations
shown inside the battery compartment .
A single cell installed with the wrong polarity may still allow
the instrument to function, but will cause overheating
severe enough to rupture a cell, with consequent risk to the
operator and damage to the instrument.
DO NOT REMOVE the protective grid from Type 1
microphone capsules as this will expose the diaphragm,
which is extremely vulnerable to damage.
UNDER NO CIRCUMSTANCES should these instruments be
cleaned using a solvent based cleaner.
Repairs of damage caused by a failure to observe these
warnings will NOT be covered by the normal warranty
conditions.
Notes !
The CEL-440 and CEL-480 are supplied complete with Type 1 or Type 2
Electret Microphones and have no need of a 200 V polarizing supply.
Therefore no such supply is available and it is safe to ignore the 200 V
polarising supply warnings shown on CEL-250 or MK 250 microphone
packaging.
Throughout this book, display screens that are available only on a
CEL-480 will be shown with a dashed - - - - outline.
CEL-440/480OperatorsHandbook-Page3
Contents
Improvements Introduced with V 2.0 Firmware
New Measurement Features:
1.The ability to perform “broadband” logging measurements in a
selected octave or third octave band.
This allows users of narrow band instruments to restrict broadband measurement to any available frequency band, and log data
for all selected broadband parameters.
New Timer Features - CEL-480 only
2.A seventh set of Delay Timers.
This allows a different set of timers to be selected for each day
of the week.
3.A Repeat function for the Delay Timer.
This allows a week’s settings to be used up to 999 times.
Improved Display Features:
4.The number of tests the instrument has to perform is shown
during the self test routine so the user knows when it has got to
the end with no problems.
5.An indication of the last used setup during the startup procedure.
6.An additional line on menus showing the selection number and
the total number of choices available.
During a run, this line is inhibited to indicate that the items are not
changeable.
7.The option to store data or re-start the measurement when
is pressed at the end of each run.
8.Storage and print out of the last TWO sets of calibration data on
the default hard copy.
This allows the pre- and post-measurement calibrations to be
recorded.
9.The ability for scrolling to wrap-round when inspecting period data
stored on a CEL- 480.
10.The spectrum is left on screen when a narrow band frequency
scan finishes so the user can inspect it without having to recall it
from memory.
11.The ability to store Histogram data in the meter from the SETUP/
FUNCTIONS menu.
The following steps are suggested to get your CEL-440 or CEL-480
started.
1.Refer to Sections 1.1 and 1.2 to learn what is shown on the
display and how to use the Quick Edit function.
2.Install the microphone, preamplifier and batteries according to
Section 1.3.
3.Switch the instrument ON and follow the preliminary messages
as described in Section 1.4.
The messages indicate the instrument type and version.
4.Check “Hints for using Menus” between Chapters 1 and 2.
5.Select the display language and microphone response according
to Section 2.1.
6.Perform an acoustic calibration check as detailed in Section 2.2.
7.Select a bandwidth, run timing and pre-set measurement set-up
according to Chapter 3.
8.Start measurement and data logging as described in Section 4.1.
9.Recall stored data (and print) it according to Section 4.2.
10.Record data on a tape or DAT recorder and replay it as described
in Section 4.5.
Page6-CEL-440/480OperatorsHandbook
Introduction
1.INTRODUCTION
Please read Sections 1.1 to 1.4 before commencing measurement. For a
quick introduction to instrument operations, Chapter 4 may be used as a
tutorial.
1.1Display & Keys
Stop /Run /Pauseindicator
(Instrument is in Run mode)
ØØØØØ:1Ø
L
AS
73.8dB
L
ASmx
LL
65.4dB69.9dB
ASmnZpk
30100
L
Aeq
dB
69.4dB
Period number
Battery condition
Principal parameter
Principal level
Run-Time elapsed
Bandwidth
Icons showing active keys
Measurement/
Instrument information
Icon showing active (cursor) key
Analog display of SPL
Icon showing active (cursor) key
980014
MENU
DATA
Figure 1: Display & Key Identities
(Further icons may be shown during operation)
Switches instrument On/Off
Changes menu (=operating mode)
Enables Quick Edit function
Operates display back light
Move cursor (=select options)
Stops a measurement
Accesses stored data
Starts/Pauses/Restarts a
Measurement run
CEL-440/480OperatorsHandbook-Page7
Introduction
The broadband screen in Figure 1 shows one principal and four
subordinate parameters, where the principal parameter is a level as
identified below.
Frequency Weighting
Level
L
AS
The following frequency weightings may be shown:
A-, C- and Z- (Linear) weighted, where Linear (Z) weighting 20 Hz
to 20 kHz meets the tolerances of IEC 60651.
The following time weightings and other identities may be shown.
specified period of hours and minutes.
When the period is specified as 8 hours, this
measurement unit will be shown as EP,d.
averaged sound pressure level with the selected
frequency and time weighting that represents
the total average of a persons workplace noise
exposure averaged over a user specified period
of hours and minutes.
This unit of measurement is specified in the
USA: OSHA standard 1910-95 published in 1983.
When the period is specified as 8 hours, this
measurement unit will be shown as TWA.
980018
Page8-CEL-440/480OperatorsHandbook
A
L
AS
Introduction
Stop
dB
100
Quick Edit
dB
100
Stop
dB
100
Run
dB
64.8dB
100
Press
Press
Press
980015
Change
Setting
30
L
AS
30
Highlight Parameter To Select It
L
CF
L
Cpk
L
CFmx
30
Other Parameters Will Have Changed
ØØØØØ:1Ø
L
ASmx
73.8dB
L
Cpk
L
CFmx
All Parameter Levels Are Shown In Run Mode
Figure 2: Quick edit functions
30
L
CFmn
L
Ceq
L
72.1dB68.6dB
CFmn
L
Ceq
1.2Quick Edit
All control functions for the CEL-440 and CEL-480 are accessed via menu
displays. In addition a powerful Quick Edit facility can be invoked by
thekey to allow measurement parameters and settings to be
changed quickly and easily on screen. Figure 2 shows how the quick edit
function enables the cursor keys to edit the current settings.
1.3Install Microphone, Preamplifier
& Batteries
With the instrument switched OFF, insert the connector on the
microphone/preamplifier unit into the socket in the cone at the top of the
CEL-440/480OperatorsHandbook-Page9
Introduction
instrument body. Face the red dot on the preamplifier to the front of the
instrument so that the key engages in a keyway in the socket to ensure
correct pin connection.
Serial Number Label
980016
Tripod Bush (1/4" Whit.)
Figure 3: Battery orientation
One cell installed with wrong polarity may still
allow the instrument to function, but can
cause overheating severe enough to rupture a
cell and damage the instrument.
preamplifier unit from the
instrument, pull on the
knurled sleeve.)
batteries (AA or equivalent)
into the battery
compartment in the rear of
the instrument (Figure 3).
inserted in the orientations
shown inside the
compartment.
(To disconnect the
Load four new 1.5 V
Make sure they are
1.4Switch Instrument ON/OFF
1.Pressto switch the instrument ON.
Early firmware may briefly show the last used screen if it has only
recently been switched OFF.
The instrument starts a series of self tests during which it
indicates the instrument type, firmware version, preamplifier
type and interface status.
13Jan98 1Ø:54:42
512Kb
1%
2
Format
p-02
At the end of the self test sequence, the instrument displays
memory information and battery voltage,
Memory
Full
Runs
Currentdateandtime
Setup
User Setup 1
Press MENU
Key to Start
p-03
followed by the identity of the last used setup.
Lastusedsetup(NotshownbeforeV2.0firm-
L
ZF
dB
Dateandtimeoflastcalibration
Last Cal.
Ø9Jan98 12.22
p-04
Finally it enters calibration check mode (described in Chapter 2).
2.Pressto switch the instrument OFF when all measurement,
setup and data recall operations are finished.
1.5Description
Both the CEL-440 and CEL-480 Sound Level Meters are available as
Type 1 or Type 2 instruments to give an ability to make comprehensive
CEL-440/480OperatorsHandbook-Page11
Introduction
sound level measurements. The main difference between the CEL-440
and the CEL-480 is the additional timing facility built into the CEL-480.
This allows the instrument to operate with delayed run start and stop
times and for narrow band models to make repeated frequency scans.
Versions of both
instruments are available for
broadband measurement,
broadband plus octave band
measurement and
broadband, octave band and
third-octave band
measurement. Frequency
bands are scanned
sequentially, with scanning
controlled manually or
automatically.
To simplify
operation, frequently used
measurement setups can be
stored for re-use. The setup
memory can accommodate
one factory setup and up to
four user specified setups for each bandwidth.
Simple procedures allow a measurement setup to be selected
and the instrument to make the required measurements and save them
automatically in separate data memories for each bandwidth.
Data stored in the memory can be recalled to the display for
inspection, so that the operator can confirm that the results are valid
before leaving the test site. When the instrument is connected to a
Centronics printer, results can be printed out in a preset report format
together with the setup data.
The instruments can be operated and deliver adequate results
without the need for other equipment, beyond an acoustic calibrator. All
operations can be controlled via the instrument keypad and simple menu
options. Figure 4 shows the basic arrangement of the menu structure.
However, the instruments become even more versatile when
their measurement and setup data is downloaded to a PC using the
Type 1 and Type 2 Measurement
Broadband, Octave Band and Third
Octave Band versions
Simultaneous measurement of up to 13
parameters in broadband mode
Simultaneous measurement of up to 3
parameters in narrow band mode
All broadband parameters can be
measured in a selected narrow band
1
Quick Edit function for immediate
parameter change
One factory setup and up to four user
specified setups stored for each
bandwidth
Up to two profiles can be attached to
each measurement (CEL-480 only)
Note1:NotavailablebeforeV2.0firmware.
Page12-CEL-440/480OperatorsHandbook
Main Menu
80043
g
Introduction
Measure
Mode
Setup
Calibration
Memory
Configure
Status
9
Set
Bandwidth
Select
Setup
Select
Input
View/
Delete
Date & Time
Memory Use
Battery Volta
Set
Timers
Set
Parameters
Calibrate
Select
Directory
PC Communications
Language
Backlight
Time & Date
Microphone Response
e
Run Timer
Delay Timers
SLM Parameters
Measurement Period
Functions
Dose Parameters
Profiles
Select
Data
Figure 4: Simplified menu structure
(Some of these options may not be available on all versions)
CEL-6726 dB21 or CEL-6727 dB22 Windows™ based software. This
software has the facilities expected of fully featured Windows™
packages offering post processing, cut and paste between applications
and comprehensive word processing capabilities. In addition, dB22 has
extensive on screen graphing facilities
These instruments are constructed to withstand some of the
toughest industrial conditions with cases formed from a
polyester/polycarbonate material used, giving them a high resistance to
damage. Data integrity is further protected by use of robust electret
microphones.
1.6CEL-440 Sound Level Meters
The CEL-440 is ideal for on-site noise surveys and can also monitor
personal noise exposure in accordance with European ISO or USA OSHA
and DOD standards. The following weightings can be set:
RMS:A, C, Z (see section 1.1) weightings,
Peak:C, Z weightings,
CEL-440/480OperatorsHandbook-Page13
Introduction
Time:F (fast), S (slow), I (impulse),
Q:3, 4, 5, 6 energy conversion factor.
Broadband models can measure all of the following parameters
simul-taneously.
L
AF
L
AFmx
L
AFmn
L
Aeq
L
AIeq
L
Zpk
L
AE
L
EP,v
L
Tm3
L
Tm5
L
AFNN.N
TWA
L
Avg
v
Sound level, with current weightings (A and F
are shown),
Maximum level, with current weightings,
Minimum level, with current weightings,
Equivalent continuous level, with current
frequency weighting,
Equivalent continuous level, with impulse
weighting,
Linear peak,
Sound exposure level (sometimes known as SEL),
Noise dose normalised to a user selected
(variable) period of hours and minutes,
When the period is set to 8 hours, this will be
shown as L
EP,d
.
Cumulative average of fast weighted maximum
values taken over 3 s periods (Taktmaximal 3),
Cumulative average of fast weighted maximum
values taken over 5 s periods (Taktmaximal 5),
Five percentile values using current weightings
with values between 0.0% and 99.9% that can
be set by the user,
Time Weighted Average is the normalised time
averaged sound pressure level with the selected
frequency and time weighting that represents
the total L
of a person’s workplace noise
Avg
exposure averaged over a user selected
(variable) period of hours and minutes.
This unit is specified in the USA: OSHA standard
1910-95 published in 1983. When the period is
set to 8 hours, the unit will be shown as TWA.
Average level over the measurement period.
Narrow band models can measure and scan the following parameters
simultaneously.
Page14-CEL-440/480OperatorsHandbook
Introduction
L
AF
L
AFmx
L
Aeq
All sound level measurement parameters can be set from the keypad
and there are seven overlapping measurement ranges at 10 dB steps
between 10-80 dB and 70-140 dB.
Sound level, with current weightings (A and F
are shown),
Maximum level, with current weightings,
Equivalent continuous level, with current
frequency weighting.
1.7CEL-480 Sound Level Meter
The CEL-480 is recommended for detailed measurements as, in addition
to the CEL-440 capabilities described above, it features automatic run
timing and scanning facilities controlled by the built-in clock. This allows
it to operate with user preset run durations or start and stop times.
There are time-history recording capabilities that allow up to two
profiles to be saved with each result set in a separate profile store.
Sampling periods can be specified between 1 second and 1 hour, while
profile storage is in excess of 200 000 data points.
1.8Instrument Power Supplies
All of these instruments are powered by a set of of four standard AA size
batteries located in a compartment in the back of the instrument case,
and it is recommended that alkaline IEC Type LR6 be used.
Rechargeable nickel cadmium batteries may also be employed,
but will give a shorter operating life. Zinc carbon batteries are NOT
recommended. If there is doubt whether the batteries will be able to
power the instrument for the start and stop times set, or when
measurements are needed over long periods, the instrument can be
powered directly from an external 12 V DC supply without the need to
remove the internal batteries.
DAMAGE to the instrument will occur if an
external power supply exceeds 14 V DC. The
nominal external supply is 12 V DC.
CEL-440/480OperatorsHandbook-Page15
Introduction
External power for the instrument is supplied via a 2.1 mm co-axial 2-line
D.C. connector in the bottom of the instrument case. Terminal polarity is
+12 V DC on the tip and 0 V DC ground on the sleeve.
A Casella CEL Universal Power Supply (-PC18) may be used to
power the instruments from nominal 110-240 V 50/60 Hz mains supplies.
To prevent loss of data should an external power supply be interrupted
for any reason during measurement, it is recommended that a usable set
of batteries be kept in the instrument.
In general, download all important data and remove the batteries
from the instruments when they are to be out of use for some time.
Note that when an instrument is returned to battery operation
after being powered from an external supply, either: switch the
instrument OFF then ON again, or: disconnect ALL devices from the DC
input socket, in order to avoid additional discharging of the internal
batteries.
The instrument contains a built in Lithium battery to maintain
stored data and setups for periods while the main batteries are absent.
1.9dB21 & dB22 SoundTrack Software
The CEL-6726 dB21 and CEL-6727 dB22 Software enable data collected
by the CEL-440 and CEL-480 to be downloaded to a PC for storage,
manipulation and the production of reports. Data can be exported in
ASCII format to proprietary word processing software, while profile and
statistical data can be exported as tab-delimited text files suitable for use
with spreadsheets such as Lotus 123™ (Release 2 or later) and MS
Excel™.
In general, dB22 offers similar manipulation and reporting
capabilities to dB21, with the addition of on-screen graphing and
reporting facilities.
dB21/22 software offer NO ON-LINE CONTROL.
dB21 software offers NO GRAPHICAL
capabilities; any graphing that may be required
must be accomplished within a spreadsheet.
®
The dB21 and dB22 Software runs under MS Windows
advantage of the control facilities offered by the environment, so that
once installed, users with a modest knowledge of Windows will find
Page16-CEL-440/480OperatorsHandbook
taking
Introduction
many operations are performed almost instinctively with a minimum of
learning time.
Downloading is performed under the control of the PC. Data
from each of the measurement runs in the analyser is transferred as a
series of associated datafiles together with a note file in which
comments and information can be written. Once downloaded, the user
can display the following information:
Select the language in which the instrument will present menu options.
This should be the first operation so that the displays and options will be
understood.
Similarly, to get the most accurate results, these instruments
must be operated with a known microphone directional response and be
calibrated for this response.
When the language and microphone response are known to be
correct for the proposed measurement task, go directly to Section 6
Calibration Check.
European IEC standards require measurements to be performed
with microphones that have a Free Field response, while the U.S.A.
ANSI standards require Random Incidence microphones. The
microphone response is set via the same Configuration menu.
Normally, these more permanent characteristics will need to be
set only once, as the instrument will store the settings and use them the
next time it is switched on.
All of the configuration settings that have been left highlighted on the
menu screens, plus any changes to the time and date, will be stored for
use next time the instrument is switched on.
2.2Perform Acoustic Calibration Check
The calibration level indicated by the instrument will depend on the
microphone response set. Therefore, an acoustic calibration check
should be performed only when the microphone response is correct for
the required task. When the calibration is known to be acceptable, press
MENU
once to obtain a Measurement screen or twice to obtain the Main
menu that gives access to all settings and stored data.
CEL-440/480OperatorsHandbook-Page21
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.