Casella CEL CEL-440 User Manual

CEL-440 & CEL-480
SOUND LEVEL METERS
Operators Handbook
060279/HB-07
August 2002
CASELLA CEL Regent House Wolseley Road Kempston Bedford MK42 7JY, U.K.
Phone: +44 (0) 1234 844 100 Fax: +44 (0) 1234 841 490 E-mail: info@casellacel.com Web: www.casellacel.com
CASELLA USA Tech Park 17 Old Nashua Road Box 15 Amherst NH 03031, U.S.A.
Toll Free: +1 800 366 2966 Fax: +1 603 672 8053 E-mail: info@casellausa.com Web: www.casellausa.com
Contents
Page 2 - CEL-440/480 Operators Handbook
Contents
Warnings !
LOAD BATTERIES following the instructions given in Section 1.3. Make sure they are inserted in the orientations shown inside the battery compartment .
A single cell installed with the wrong polarity may still allow the instrument to function, but will cause overheating severe enough to rupture a cell, with consequent risk to the operator and damage to the instrument.
DO NOT REMOVE the protective grid from Type 1 microphone capsules as this will expose the diaphragm, which is extremely vulnerable to damage.
UNDER NO CIRCUMSTANCES should these instruments be cleaned using a solvent based cleaner.
Repairs of damage caused by a failure to observe these warnings will NOT be covered by the normal warranty conditions.
Notes !
The CEL-440 and CEL-480 are supplied complete with Type 1 or Type 2 Electret Microphones and have no need of a 200 V polarizing supply. Therefore no such supply is available and it is safe to ignore the 200 V polarising supply warnings shown on CEL-250 or MK 250 microphone packaging.
Throughout this book, display screens that are available only on a CEL-480 will be shown with a dashed - - - - outline.
CEL-440/480 Operators Handbook - Page 3
Contents
Improvements Introduced with V 2.0 Firmware
New Measurement Features:
1. The ability to perform “broadband” logging measurements in a selected octave or third octave band. This allows users of narrow band instruments to restrict broad­band measurement to any available frequency band, and log data for all selected broadband parameters.
New Timer Features - CEL-480 only
2. A seventh set of Delay Timers. This allows a different set of timers to be selected for each day of the week.
3. A Repeat function for the Delay Timer. This allows a week’s settings to be used up to 999 times.
Improved Display Features:
4. The number of tests the instrument has to perform is shown during the self test routine so the user knows when it has got to the end with no problems.
5. An indication of the last used setup during the startup procedure.
6. An additional line on menus showing the selection number and the total number of choices available. During a run, this line is inhibited to indicate that the items are not changeable.
7. The option to store data or re-start the measurement when is pressed at the end of each run.
8. Storage and print out of the last TWO sets of calibration data on the default hard copy. This allows the pre- and post-measurement calibrations to be recorded.
9. The ability for scrolling to wrap-round when inspecting period data stored on a CEL- 480.
10. The spectrum is left on screen when a narrow band frequency scan finishes so the user can inspect it without having to recall it from memory.
11. The ability to store Histogram data in the meter from the SETUP/ FUNCTIONS menu.
Page 4 - CEL-440/480 Operators Handbook
Contents
Contents
Chapter ................... Page
1. INTRODUCTION . . . . . . . . . . . . . . . . . . 7
1.1 Display & Keys .................... 7
1.2 Quick Edit ....................... 9
1.3 Install Microphone, Preamplifier & Batteries .... 10
1.4 Switch Instrument ON/OFF ............. 11
1.5 Description ..................... 12
1.6 CEL-440 Sound Level Meters ............ 14
1.7 CEL-480 Sound Level Meter ............ 16
1.8 Instrument Power Supplies ............. 16
1.9 dB21 & dB22 SoundTrack Software ........ 17
2. PRELIMINARY OPERATIONS . . . . . . . . . . . . 19
2.1 Select Instrument Configuration (Language,
Microphone Response Etc.) ............. 19
2.2 Perform Acoustic Calibration Check ........ 21
3. SELECT MEASUREMENT MODE & SETUP . . . . . 25
3.1 Select Measurement Mode
(Bandwidth & CEL-480 Run Timing) ........ 25
3.2 Select Measurement Setup
(Factory- or User-Prepared Setups) ......... 28
3.2.1 Select Broadband Measurement Setup ....... 28
3.2.2 Select Narrow Band Measurement Setup ..... 33
4. OPERATION . . . . . . . . . . . . . . . . . . . . 37
4.1 Measurement .................... 37
4.1.1 Starting Measurement ............... 37
4.1.2 Broadband Measurement .............. 38
4.1.3 Narrow Band Measurement ............ 41
4.2 Recall Stored Data & Print Hard Copy ........ 44
4.3 Delete Stored Data ................. 48
4.4. Format Memory ................... 49
4.5 Use With Tape and DAT Recorders
(Including Calibration For Line Input) ........ 50
4.5.1 Recording ...................... 51
CEL-440/480 Operators Handbook - Page 5
Contents
Chapter ................... Page
4.5.2 Replay ........................ 52
5. SPECIFICATION . . . . . . . . . . . . . . . . . . 55
5.1 CEL-440 ....................... 55
5.2 CEL-480 ....................... 61
5.3 Filter Scanning ................... 62
5.4 Compliance ................... 66
6. PARTS & WARRANTY . . . . . . . . . . . . . . 67
6.1 Schedule of Parts .................. 67
6.2 Instrument Servicing & Warranty .......... 68
Getting Started
The following steps are suggested to get your CEL-440 or CEL-480 started.
1. Refer to Sections 1.1 and 1.2 to learn what is shown on the display and how to use the Quick Edit function.
2. Install the microphone, preamplifier and batteries according to Section 1.3.
3. Switch the instrument ON and follow the preliminary messages as described in Section 1.4. The messages indicate the instrument type and version.
4. Check “Hints for using Menus” between Chapters 1 and 2.
5. Select the display language and microphone response according to Section 2.1.
6. Perform an acoustic calibration check as detailed in Section 2.2.
7. Select a bandwidth, run timing and pre-set measurement set-up according to Chapter 3.
8. Start measurement and data logging as described in Section 4.1.
9. Recall stored data (and print) it according to Section 4.2.
10. Record data on a tape or DAT recorder and replay it as described in Section 4.5.
Page 6 - CEL-440/480 Operators Handbook
Introduction
1. INTRODUCTION
Please read Sections 1.1 to 1.4 before commencing measurement. For a quick introduction to instrument operations, Chapter 4 may be used as a tutorial.
1.1 Display & Keys
Stop /Run /Pause indicator
(Instrument is in Run mode)
ØØØØ Ø:1Ø
L
AS
73.8dB
L
ASmx
LL
65.4dB 69.9dB
ASmn Zpk
30 100
L
Aeq
dB
69.4dB
Period number Battery condition Principal parameter Principal level Run-Time elapsed Bandwidth Icons showing active keys Measurement/
Instrument information
Icon showing active (cursor) key Analog display of SPL Icon showing active (cursor) key
980014
MENU
DATA
Figure 1: Display & Key Identities
(Further icons may be shown during operation)
Switches instrument On/Off Changes menu (=operating mode) Enables Quick Edit function Operates display back light Move cursor (=select options) Stops a measurement Accesses stored data Starts/Pauses/Restarts a
Measurement run
CEL-440/480 Operators Handbook - Page 7
Introduction
The broadband screen in Figure 1 shows one principal and four subordinate parameters, where the principal parameter is a level as identified below.
Frequency Weighting
Level
L
AS
The following frequency weightings may be shown:
A-, C- and Z- (Linear) weighted, where Linear (Z) weighting 20 Hz to 20 kHz meets the tolerances of IEC 60651.
The following time weightings and other identities may be shown.
S Slow. F Fast.
Time Weighting
I Impulse.
pk Peak. mx Maximum. mn Minimum. eq Equivalent continuous level. AV Average level. Tm3 Taktmaximal 3 sek. Tm5 Taktmaximal 5 sek. EP,v L
TWAv Time Weighted Average is the normalised time
N Percentile sound level. AE Sound exposure level.
based noise dose normalised to a user
eq
specified period of hours and minutes. When the period is specified as 8 hours, this measurement unit will be shown as EP,d.
averaged sound pressure level with the selected frequency and time weighting that represents the total average of a persons workplace noise exposure averaged over a user specified period of hours and minutes. This unit of measurement is specified in the USA: OSHA standard 1910-95 published in 1983. When the period is specified as 8 hours, this measurement unit will be shown as TWA.
980018
Page 8 - CEL-440/480 Operators Handbook
A
L
AS
Introduction
Stop
dB
100
Quick Edit
dB
100
Stop
dB
100
Run
dB
64.8dB
100
Press
Press
Press
980015
Change
Setting
30
L
AS
30
Highlight Parameter To Select It
L
CF
L
Cpk
L
CFmx
30
Other Parameters Will Have Changed
ØØØØ Ø:1Ø
L
ASmx
73.8dB
L
Cpk
L
CFmx
All Parameter Levels Are Shown In Run Mode
Figure 2: Quick edit functions
30
L
CFmn
L
Ceq
L
72.1dB 68.6dB
CFmn
L
Ceq
1.2 Quick Edit
All control functions for the CEL-440 and CEL-480 are accessed via menu displays. In addition a powerful Quick Edit facility can be invoked by
the key to allow measurement parameters and settings to be changed quickly and easily on screen. Figure 2 shows how the quick edit function enables the cursor keys to edit the current settings.
1.3 Install Microphone, Preamplifier & Batteries
With the instrument switched OFF, insert the connector on the microphone/preamplifier unit into the socket in the cone at the top of the
CEL-440/480 Operators Handbook - Page 9
Introduction
instrument body. Face the red dot on the preamplifier to the front of the instrument so that the key engages in a keyway in the socket to ensure correct pin connection.
Serial Number Label
980016
Tripod Bush (1/4" Whit.)
Figure 3: Battery orientation
One cell installed with wrong polarity may still allow the instrument to function, but can cause overheating severe enough to rupture a cell and damage the instrument.
preamplifier unit from the instrument, pull on the knurled sleeve.)
batteries (AA or equivalent) into the battery compartment in the rear of the instrument (Figure 3).
inserted in the orientations shown inside the compartment.
(To disconnect the
Load four new 1.5 V
Make sure they are
1.4 Switch Instrument ON/OFF
1. Press to switch the instrument ON.
Early firmware may briefly show the last used screen if it has only recently been switched OFF.
Inst. Type 440/480
CEL48ØA.1 vNN.NN Preamp I/O Test ID:
p-01
Page 10 - CEL-440/480 Operators Handbook
Type 1 None
16/1Ø
NNNNNN
Inst. Version A = Broadband
Firmware version Interface status Self test progress number Inst. identification number
B = Octave band C = Third octaves 1 = Type 1 2 = Type 2
Introduction
The instrument starts a series of self tests during which it indicates the instrument type, firmware version, preamplifier type and interface status.
13Jan98 1Ø:54:42 512Kb 1% 2 Format
p-02
At the end of the self test sequence, the instrument displays memory information and battery voltage,
Memory Full Runs
Current date and time
Setup User Setup 1 Press MENU Key to Start
p-03
followed by the identity of the last used setup.
Last used setup (Not shown before V 2.0 firm-
L
ZF
dB
Date and time of last calibration
Last Cal. Ø9Jan98 12.22
p-04
Finally it enters calibration check mode (described in Chapter 2).
2. Press to switch the instrument OFF when all measurement, setup and data recall operations are finished.
1.5 Description
Both the CEL-440 and CEL-480 Sound Level Meters are available as Type 1 or Type 2 instruments to give an ability to make comprehensive
CEL-440/480 Operators Handbook - Page 11
Introduction
sound level measurements. The main difference between the CEL-440 and the CEL-480 is the additional timing facility built into the CEL-480. This allows the instrument to operate with delayed run start and stop times and for narrow band models to make repeated frequency scans.
Versions of both instruments are available for broadband measurement, broadband plus octave band measurement and broadband, octave band and third-octave band measurement. Frequency bands are scanned sequentially, with scanning controlled manually or automatically.
To simplify operation, frequently used measurement setups can be stored for re-use. The setup memory can accommodate one factory setup and up to four user specified setups for each bandwidth.
Simple procedures allow a measurement setup to be selected and the instrument to make the required measurements and save them automatically in separate data memories for each bandwidth.
Data stored in the memory can be recalled to the display for inspection, so that the operator can confirm that the results are valid before leaving the test site. When the instrument is connected to a Centronics printer, results can be printed out in a preset report format together with the setup data.
The instruments can be operated and deliver adequate results without the need for other equipment, beyond an acoustic calibrator. All operations can be controlled via the instrument keypad and simple menu options. Figure 4 shows the basic arrangement of the menu structure.
However, the instruments become even more versatile when their measurement and setup data is downloaded to a PC using the
Type 1 and Type 2 Measurement Broadband, Octave Band and Third
Octave Band versions Simultaneous measurement of up to 13
parameters in broadband mode Simultaneous measurement of up to 3
parameters in narrow band mode All broadband parameters can be
measured in a selected narrow band
1
Quick Edit function for immediate parameter change
One factory setup and up to four user specified setups stored for each bandwidth
Up to two profiles can be attached to each measurement (CEL-480 only)
Note 1: Not available before V 2.0 firmware.
Page 12 - CEL-440/480 Operators Handbook
Main Menu
80043
g
Introduction
Measure
Mode
Setup
Calibration
Memory
Configure
Status
9
Set
Bandwidth
Select
Setup
Select
Input
View/
Delete
Date & Time Memory Use Battery Volta
Set
Timers
Set
Parameters
Calibrate
Select
Directory
PC Communications Language Backlight Time & Date Microphone Response
e
Run Timer Delay Timers
SLM Parameters Measurement Period Functions Dose Parameters Profiles
Select
Data
Figure 4: Simplified menu structure
(Some of these options may not be available on all versions)
CEL-6726 dB21 or CEL-6727 dB22 Windowsbased software. This software has the facilities expected of fully featured Windows packages offering post processing, cut and paste between applications and comprehensive word processing capabilities. In addition, dB22 has extensive on screen graphing facilities
These instruments are constructed to withstand some of the toughest industrial conditions with cases formed from a polyester/polycarbonate material used, giving them a high resistance to damage. Data integrity is further protected by use of robust electret microphones.
1.6 CEL-440 Sound Level Meters
The CEL-440 is ideal for on-site noise surveys and can also monitor personal noise exposure in accordance with European ISO or USA OSHA and DOD standards. The following weightings can be set:
RMS: A, C, Z (see section 1.1) weightings,
Peak: C, Z weightings,
CEL-440/480 Operators Handbook - Page 13
Introduction
Time: F (fast), S (slow), I (impulse),
Q: 3, 4, 5, 6 energy conversion factor. Broadband models can measure all of the following parameters simul-taneously.
L
AF
L
AFmx
L
AFmn
L
Aeq
L
AIeq
L
Zpk
L
AE
L
EP,v
L
Tm3
L
Tm5
L
AFNN.N
TWA
L
Avg
v
Sound level, with current weightings (A and F are shown), Maximum level, with current weightings, Minimum level, with current weightings, Equivalent continuous level, with current frequency weighting, Equivalent continuous level, with impulse weighting, Linear peak, Sound exposure level (sometimes known as SEL), Noise dose normalised to a user selected (variable) period of hours and minutes, When the period is set to 8 hours, this will be shown as L
EP,d
. Cumulative average of fast weighted maximum values taken over 3 s periods (Taktmaximal 3), Cumulative average of fast weighted maximum values taken over 5 s periods (Taktmaximal 5), Five percentile values using current weightings with values between 0.0% and 99.9% that can be set by the user, Time Weighted Average is the normalised time averaged sound pressure level with the selected frequency and time weighting that represents the total L
of a persons workplace noise
Avg
exposure averaged over a user selected (variable) period of hours and minutes. This unit is specified in the USA: OSHA standard 1910-95 published in 1983. When the period is set to 8 hours, the unit will be shown as TWA. Average level over the measurement period.
Narrow band models can measure and scan the following parameters simultaneously.
Page 14 - CEL-440/480 Operators Handbook
Introduction
L
AF
L
AFmx
L
Aeq
All sound level measurement parameters can be set from the keypad and there are seven overlapping measurement ranges at 10 dB steps between 10-80 dB and 70-140 dB.
Sound level, with current weightings (A and F are shown), Maximum level, with current weightings, Equivalent continuous level, with current frequency weighting.
1.7 CEL-480 Sound Level Meter
The CEL-480 is recommended for detailed measurements as, in addition to the CEL-440 capabilities described above, it features automatic run timing and scanning facilities controlled by the built-in clock. This allows it to operate with user preset run durations or start and stop times.
There are time-history recording capabilities that allow up to two profiles to be saved with each result set in a separate profile store. Sampling periods can be specified between 1 second and 1 hour, while profile storage is in excess of 200 000 data points.
1.8 Instrument Power Supplies
All of these instruments are powered by a set of of four standard AA size batteries located in a compartment in the back of the instrument case, and it is recommended that alkaline IEC Type LR6 be used.
Rechargeable nickel cadmium batteries may also be employed, but will give a shorter operating life. Zinc carbon batteries are NOT recommended. If there is doubt whether the batteries will be able to power the instrument for the start and stop times set, or when measurements are needed over long periods, the instrument can be powered directly from an external 12 V DC supply without the need to remove the internal batteries.
DAMAGE to the instrument will occur if an
external power supply exceeds 14 V DC. The
nominal external supply is 12 V DC.
CEL-440/480 Operators Handbook - Page 15
Introduction
External power for the instrument is supplied via a 2.1 mm co-axial 2-line D.C. connector in the bottom of the instrument case. Terminal polarity is +12 V DC on the tip and 0 V DC ground on the sleeve.
A Casella CEL Universal Power Supply (-PC18) may be used to power the instruments from nominal 110-240 V 50/60 Hz mains supplies. To prevent loss of data should an external power supply be interrupted for any reason during measurement, it is recommended that a usable set of batteries be kept in the instrument.
In general, download all important data and remove the batteries from the instruments when they are to be out of use for some time.
Note that when an instrument is returned to battery operation after being powered from an external supply, either: switch the instrument OFF then ON again, or: disconnect ALL devices from the DC input socket, in order to avoid additional discharging of the internal batteries.
The instrument contains a built in Lithium battery to maintain stored data and setups for periods while the main batteries are absent.
1.9 dB21 & dB22 SoundTrack Software
The CEL-6726 dB21 and CEL-6727 dB22 Software enable data collected by the CEL-440 and CEL-480 to be downloaded to a PC for storage, manipulation and the production of reports. Data can be exported in ASCII format to proprietary word processing software, while profile and statistical data can be exported as tab-delimited text files suitable for use with spreadsheets such as Lotus 123™ (Release 2 or later) and MS Excel.
In general, dB22 offers similar manipulation and reporting capabilities to dB21, with the addition of on-screen graphing and reporting facilities.
dB21/22 software offer NO ON-LINE CONTROL.
dB21 software offers NO GRAPHICAL
capabilities; any graphing that may be required
must be accomplished within a spreadsheet.
®
The dB21 and dB22 Software runs under MS Windows advantage of the control facilities offered by the environment, so that once installed, users with a modest knowledge of Windows will find
Page 16 - CEL-440/480 Operators Handbook
taking
Introduction
many operations are performed almost instinctively with a minimum of learning time.
Downloading is performed under the control of the PC. Data from each of the measurement runs in the analyser is transferred as a series of associated datafiles together with a note file in which comments and information can be written. Once downloaded, the user can display the following information:
¤ Run summary for each run,
¤ Data file,
¤ Up to two profiles for each run,
¤ Notepad for user-entered notes.
CEL-440/480 Operators Handbook - Page 17
Introduction
1. and displayed on a
2. In general, and keys move the
3. In general, and keys change the
4. at the bottom of a menu list indicates
5. An indication such as 3/5 at the bottom
6. Settings left highlighted on a menu
Hints For Using Menus
menu show which option keys are active.
reverse video cursor from field to field.
entry in the cursor field.
there may be further options.
2
of a menu selected from 5 available.
screen become the active options.
shows that option 3 has been
Note 2: Not available before V 2.0 firmware.
Page 18 - CEL-440/480 Operators Handbook
The illustrations in this book are for full featured instruments, other instruments may offer fewer or no options.
Setup
Preliminary Operations
2. PRELIMINARY OPERATIONS
2.1 Select Instrument Configuration
(Language, Microphone Response Etc.)
Select the language in which the instrument will present menu options. This should be the first operation so that the displays and options will be understood.
Similarly, to get the most accurate results, these instruments must be operated with a known microphone directional response and be calibrated for this response.
When the language and microphone response are known to be correct for the proposed measurement task, go directly to Section 6 Calibration Check.
European IEC standards require measurements to be performed with microphones that have a Free Field response, while the U.S.A. ANSI standards require Random Incidence microphones. The microphone response is set via the same Configuration menu.
Normally, these more permanent characteristics will need to be set only once, as the instrument will store the settings and use them the next time it is switched on.
Proceed as follows.
Operation Press Display Shows Comments
1. Switch the instrument ON
2. Select the Main menu
MENU
2 times
L
ZF
Last Cal. Ø9Jan98 12.22
k-01
Measure Mode
Setup
Calibration Memory
k-02
2/6
Start up messages are displayed that include: Instrument Type,
dB
Firmware Version, Etc. Finally, the Configuration screen is displayed with Battery Voltage, followed by the Calibration check screen showing the current sound level and details of the last calibration.
This is the Main menu. It gives access to further menus that select: Measurement Mode, Measurement Setup, Calibration Sequence, Data Recall from Memory, Instrument Configuration, Status.
CEL-440/480 Operators Handbook - Page 19
Français
Preliminary Operations
Operation Press Display Shows Comments
3. Select the Configure option
or
Setup
Move the reverse video cursor to the required line.
Calibration Memory
Configure
4. Move to the Configure menu
5. Select the Language option
k-03
RS232 Comms
Language Backlight Time & Date
k-04
or
RS232 Comms
5/6
1/5
This menu allows the following items to be set: RS232 Communication Baud rates from 1200 to 19200, Menu Language, Backlighting Durations of 1, 5, 10, 15, 30 s or OFF, Instrument Time & Date, Microphone Response Free Field or Random.
Move the reverse video cursor to the required line.
Language
Backlight Time & Date
6. Move to the Language menu
7. Select the required language
k-05
English
Français Deutsch Italiano
k-06
or
English
Français
Deutsch
2/5
1/5
This menu allows the following languages to be selected for menu presentation: English, Français, Deutsch, Italiano, Español.
The highlighted language will be used for menu presentation. This setting will also be stored and used next time the instrument is switched on.
Italiano
k-07
2/5
Press to return to the Configuration menu.
Page 20 - CEL-440/480 Operators Handbook
Preliminary Operations
Operation Press Display Shows Comments
8. Select the required microphone response (and other configuration items) in the same way.
Date and time are set using the quick edit function as follows.
9. Select the date and time option by moving the reverse video cursor to the required line on the Configure menu.
10. Move to Date and Time menu
12
Jan
Mon 1998
The highlighted entries will be used. These settings will also be stored and used next time the instrument is switched on.
This menu allows the time and date to be set.
Ø9:51:27
11. When the date and time are to be changed
12. Return to the Main menu
k-09
Mon 1998
k-10
MENU
2 times
1/2
Jan
12
Ø9:51:27
1/2
This enables the quick edit function.
Use and to select a field, and and to
change the entry. The day will be determined automatically from the date and month entries.
Press again to save the date and time.
All of the configuration settings that have been left highlighted on the menu screens, plus any changes to the time and date, will be stored for use next time the instrument is switched on.
2.2 Perform Acoustic Calibration Check
The calibration level indicated by the instrument will depend on the microphone response set. Therefore, an acoustic calibration check should be performed only when the microphone response is correct for the required task. When the calibration is known to be acceptable, press
MENU
once to obtain a Measurement screen or twice to obtain the Main
menu that gives access to all settings and stored data.
CEL-440/480 Operators Handbook - Page 21
Loading...
+ 49 hidden pages