LOAD BATTERIES following the instructions given in
Section 1.3. Make sure they are inserted in the orientations
shown inside the battery compartment .
A single cell installed with the wrong polarity may still allow
the instrument to function, but will cause overheating
severe enough to rupture a cell, with consequent risk to the
operator and damage to the instrument.
DO NOT REMOVE the protective grid from Type 1
microphone capsules as this will expose the diaphragm,
which is extremely vulnerable to damage.
UNDER NO CIRCUMSTANCES should these instruments be
cleaned using a solvent based cleaner.
Repairs of damage caused by a failure to observe these
warnings will NOT be covered by the normal warranty
conditions.
Notes !
The CEL-440 and CEL-480 are supplied complete with Type 1 or Type 2
Electret Microphones and have no need of a 200 V polarizing supply.
Therefore no such supply is available and it is safe to ignore the 200 V
polarising supply warnings shown on CEL-250 or MK 250 microphone
packaging.
Throughout this book, display screens that are available only on a
CEL-480 will be shown with a dashed - - - - outline.
CEL-440/480OperatorsHandbook-Page3
Contents
Improvements Introduced with V 2.0 Firmware
New Measurement Features:
1.The ability to perform “broadband” logging measurements in a
selected octave or third octave band.
This allows users of narrow band instruments to restrict broadband measurement to any available frequency band, and log data
for all selected broadband parameters.
New Timer Features - CEL-480 only
2.A seventh set of Delay Timers.
This allows a different set of timers to be selected for each day
of the week.
3.A Repeat function for the Delay Timer.
This allows a week’s settings to be used up to 999 times.
Improved Display Features:
4.The number of tests the instrument has to perform is shown
during the self test routine so the user knows when it has got to
the end with no problems.
5.An indication of the last used setup during the startup procedure.
6.An additional line on menus showing the selection number and
the total number of choices available.
During a run, this line is inhibited to indicate that the items are not
changeable.
7.The option to store data or re-start the measurement when
is pressed at the end of each run.
8.Storage and print out of the last TWO sets of calibration data on
the default hard copy.
This allows the pre- and post-measurement calibrations to be
recorded.
9.The ability for scrolling to wrap-round when inspecting period data
stored on a CEL- 480.
10.The spectrum is left on screen when a narrow band frequency
scan finishes so the user can inspect it without having to recall it
from memory.
11.The ability to store Histogram data in the meter from the SETUP/
FUNCTIONS menu.
The following steps are suggested to get your CEL-440 or CEL-480
started.
1.Refer to Sections 1.1 and 1.2 to learn what is shown on the
display and how to use the Quick Edit function.
2.Install the microphone, preamplifier and batteries according to
Section 1.3.
3.Switch the instrument ON and follow the preliminary messages
as described in Section 1.4.
The messages indicate the instrument type and version.
4.Check “Hints for using Menus” between Chapters 1 and 2.
5.Select the display language and microphone response according
to Section 2.1.
6.Perform an acoustic calibration check as detailed in Section 2.2.
7.Select a bandwidth, run timing and pre-set measurement set-up
according to Chapter 3.
8.Start measurement and data logging as described in Section 4.1.
9.Recall stored data (and print) it according to Section 4.2.
10.Record data on a tape or DAT recorder and replay it as described
in Section 4.5.
Page6-CEL-440/480OperatorsHandbook
Introduction
1.INTRODUCTION
Please read Sections 1.1 to 1.4 before commencing measurement. For a
quick introduction to instrument operations, Chapter 4 may be used as a
tutorial.
1.1Display & Keys
Stop /Run /Pauseindicator
(Instrument is in Run mode)
ØØØØØ:1Ø
L
AS
73.8dB
L
ASmx
LL
65.4dB69.9dB
ASmnZpk
30100
L
Aeq
dB
69.4dB
Period number
Battery condition
Principal parameter
Principal level
Run-Time elapsed
Bandwidth
Icons showing active keys
Measurement/
Instrument information
Icon showing active (cursor) key
Analog display of SPL
Icon showing active (cursor) key
980014
MENU
DATA
Figure 1: Display & Key Identities
(Further icons may be shown during operation)
Switches instrument On/Off
Changes menu (=operating mode)
Enables Quick Edit function
Operates display back light
Move cursor (=select options)
Stops a measurement
Accesses stored data
Starts/Pauses/Restarts a
Measurement run
CEL-440/480OperatorsHandbook-Page7
Introduction
The broadband screen in Figure 1 shows one principal and four
subordinate parameters, where the principal parameter is a level as
identified below.
Frequency Weighting
Level
L
AS
The following frequency weightings may be shown:
A-, C- and Z- (Linear) weighted, where Linear (Z) weighting 20 Hz
to 20 kHz meets the tolerances of IEC 60651.
The following time weightings and other identities may be shown.
specified period of hours and minutes.
When the period is specified as 8 hours, this
measurement unit will be shown as EP,d.
averaged sound pressure level with the selected
frequency and time weighting that represents
the total average of a persons workplace noise
exposure averaged over a user specified period
of hours and minutes.
This unit of measurement is specified in the
USA: OSHA standard 1910-95 published in 1983.
When the period is specified as 8 hours, this
measurement unit will be shown as TWA.
980018
Page8-CEL-440/480OperatorsHandbook
A
L
AS
Introduction
Stop
dB
100
Quick Edit
dB
100
Stop
dB
100
Run
dB
64.8dB
100
Press
Press
Press
980015
Change
Setting
30
L
AS
30
Highlight Parameter To Select It
L
CF
L
Cpk
L
CFmx
30
Other Parameters Will Have Changed
ØØØØØ:1Ø
L
ASmx
73.8dB
L
Cpk
L
CFmx
All Parameter Levels Are Shown In Run Mode
Figure 2: Quick edit functions
30
L
CFmn
L
Ceq
L
72.1dB68.6dB
CFmn
L
Ceq
1.2Quick Edit
All control functions for the CEL-440 and CEL-480 are accessed via menu
displays. In addition a powerful Quick Edit facility can be invoked by
thekey to allow measurement parameters and settings to be
changed quickly and easily on screen. Figure 2 shows how the quick edit
function enables the cursor keys to edit the current settings.
1.3Install Microphone, Preamplifier
& Batteries
With the instrument switched OFF, insert the connector on the
microphone/preamplifier unit into the socket in the cone at the top of the
CEL-440/480OperatorsHandbook-Page9
Introduction
instrument body. Face the red dot on the preamplifier to the front of the
instrument so that the key engages in a keyway in the socket to ensure
correct pin connection.
Serial Number Label
980016
Tripod Bush (1/4" Whit.)
Figure 3: Battery orientation
One cell installed with wrong polarity may still
allow the instrument to function, but can
cause overheating severe enough to rupture a
cell and damage the instrument.
preamplifier unit from the
instrument, pull on the
knurled sleeve.)
batteries (AA or equivalent)
into the battery
compartment in the rear of
the instrument (Figure 3).
inserted in the orientations
shown inside the
compartment.
(To disconnect the
Load four new 1.5 V
Make sure they are
1.4Switch Instrument ON/OFF
1.Pressto switch the instrument ON.
Early firmware may briefly show the last used screen if it has only
recently been switched OFF.
The instrument starts a series of self tests during which it
indicates the instrument type, firmware version, preamplifier
type and interface status.
13Jan98 1Ø:54:42
512Kb
1%
2
Format
p-02
At the end of the self test sequence, the instrument displays
memory information and battery voltage,
Memory
Full
Runs
Currentdateandtime
Setup
User Setup 1
Press MENU
Key to Start
p-03
followed by the identity of the last used setup.
Lastusedsetup(NotshownbeforeV2.0firm-
L
ZF
dB
Dateandtimeoflastcalibration
Last Cal.
Ø9Jan98 12.22
p-04
Finally it enters calibration check mode (described in Chapter 2).
2.Pressto switch the instrument OFF when all measurement,
setup and data recall operations are finished.
1.5Description
Both the CEL-440 and CEL-480 Sound Level Meters are available as
Type 1 or Type 2 instruments to give an ability to make comprehensive
CEL-440/480OperatorsHandbook-Page11
Introduction
sound level measurements. The main difference between the CEL-440
and the CEL-480 is the additional timing facility built into the CEL-480.
This allows the instrument to operate with delayed run start and stop
times and for narrow band models to make repeated frequency scans.
Versions of both
instruments are available for
broadband measurement,
broadband plus octave band
measurement and
broadband, octave band and
third-octave band
measurement. Frequency
bands are scanned
sequentially, with scanning
controlled manually or
automatically.
To simplify
operation, frequently used
measurement setups can be
stored for re-use. The setup
memory can accommodate
one factory setup and up to
four user specified setups for each bandwidth.
Simple procedures allow a measurement setup to be selected
and the instrument to make the required measurements and save them
automatically in separate data memories for each bandwidth.
Data stored in the memory can be recalled to the display for
inspection, so that the operator can confirm that the results are valid
before leaving the test site. When the instrument is connected to a
Centronics printer, results can be printed out in a preset report format
together with the setup data.
The instruments can be operated and deliver adequate results
without the need for other equipment, beyond an acoustic calibrator. All
operations can be controlled via the instrument keypad and simple menu
options. Figure 4 shows the basic arrangement of the menu structure.
However, the instruments become even more versatile when
their measurement and setup data is downloaded to a PC using the
Type 1 and Type 2 Measurement
Broadband, Octave Band and Third
Octave Band versions
Simultaneous measurement of up to 13
parameters in broadband mode
Simultaneous measurement of up to 3
parameters in narrow band mode
All broadband parameters can be
measured in a selected narrow band
1
Quick Edit function for immediate
parameter change
One factory setup and up to four user
specified setups stored for each
bandwidth
Up to two profiles can be attached to
each measurement (CEL-480 only)
Note1:NotavailablebeforeV2.0firmware.
Page12-CEL-440/480OperatorsHandbook
Main Menu
80043
g
Introduction
Measure
Mode
Setup
Calibration
Memory
Configure
Status
9
Set
Bandwidth
Select
Setup
Select
Input
View/
Delete
Date & Time
Memory Use
Battery Volta
Set
Timers
Set
Parameters
Calibrate
Select
Directory
PC Communications
Language
Backlight
Time & Date
Microphone Response
e
Run Timer
Delay Timers
SLM Parameters
Measurement Period
Functions
Dose Parameters
Profiles
Select
Data
Figure 4: Simplified menu structure
(Some of these options may not be available on all versions)
CEL-6726 dB21 or CEL-6727 dB22 Windows™ based software. This
software has the facilities expected of fully featured Windows™
packages offering post processing, cut and paste between applications
and comprehensive word processing capabilities. In addition, dB22 has
extensive on screen graphing facilities
These instruments are constructed to withstand some of the
toughest industrial conditions with cases formed from a
polyester/polycarbonate material used, giving them a high resistance to
damage. Data integrity is further protected by use of robust electret
microphones.
1.6CEL-440 Sound Level Meters
The CEL-440 is ideal for on-site noise surveys and can also monitor
personal noise exposure in accordance with European ISO or USA OSHA
and DOD standards. The following weightings can be set:
RMS:A, C, Z (see section 1.1) weightings,
Peak:C, Z weightings,
CEL-440/480OperatorsHandbook-Page13
Introduction
Time:F (fast), S (slow), I (impulse),
Q:3, 4, 5, 6 energy conversion factor.
Broadband models can measure all of the following parameters
simul-taneously.
L
AF
L
AFmx
L
AFmn
L
Aeq
L
AIeq
L
Zpk
L
AE
L
EP,v
L
Tm3
L
Tm5
L
AFNN.N
TWA
L
Avg
v
Sound level, with current weightings (A and F
are shown),
Maximum level, with current weightings,
Minimum level, with current weightings,
Equivalent continuous level, with current
frequency weighting,
Equivalent continuous level, with impulse
weighting,
Linear peak,
Sound exposure level (sometimes known as SEL),
Noise dose normalised to a user selected
(variable) period of hours and minutes,
When the period is set to 8 hours, this will be
shown as L
EP,d
.
Cumulative average of fast weighted maximum
values taken over 3 s periods (Taktmaximal 3),
Cumulative average of fast weighted maximum
values taken over 5 s periods (Taktmaximal 5),
Five percentile values using current weightings
with values between 0.0% and 99.9% that can
be set by the user,
Time Weighted Average is the normalised time
averaged sound pressure level with the selected
frequency and time weighting that represents
the total L
of a person’s workplace noise
Avg
exposure averaged over a user selected
(variable) period of hours and minutes.
This unit is specified in the USA: OSHA standard
1910-95 published in 1983. When the period is
set to 8 hours, the unit will be shown as TWA.
Average level over the measurement period.
Narrow band models can measure and scan the following parameters
simultaneously.
Page14-CEL-440/480OperatorsHandbook
Introduction
L
AF
L
AFmx
L
Aeq
All sound level measurement parameters can be set from the keypad
and there are seven overlapping measurement ranges at 10 dB steps
between 10-80 dB and 70-140 dB.
Sound level, with current weightings (A and F
are shown),
Maximum level, with current weightings,
Equivalent continuous level, with current
frequency weighting.
1.7CEL-480 Sound Level Meter
The CEL-480 is recommended for detailed measurements as, in addition
to the CEL-440 capabilities described above, it features automatic run
timing and scanning facilities controlled by the built-in clock. This allows
it to operate with user preset run durations or start and stop times.
There are time-history recording capabilities that allow up to two
profiles to be saved with each result set in a separate profile store.
Sampling periods can be specified between 1 second and 1 hour, while
profile storage is in excess of 200 000 data points.
1.8Instrument Power Supplies
All of these instruments are powered by a set of of four standard AA size
batteries located in a compartment in the back of the instrument case,
and it is recommended that alkaline IEC Type LR6 be used.
Rechargeable nickel cadmium batteries may also be employed,
but will give a shorter operating life. Zinc carbon batteries are NOT
recommended. If there is doubt whether the batteries will be able to
power the instrument for the start and stop times set, or when
measurements are needed over long periods, the instrument can be
powered directly from an external 12 V DC supply without the need to
remove the internal batteries.
DAMAGE to the instrument will occur if an
external power supply exceeds 14 V DC. The
nominal external supply is 12 V DC.
CEL-440/480OperatorsHandbook-Page15
Introduction
External power for the instrument is supplied via a 2.1 mm co-axial 2-line
D.C. connector in the bottom of the instrument case. Terminal polarity is
+12 V DC on the tip and 0 V DC ground on the sleeve.
A Casella CEL Universal Power Supply (-PC18) may be used to
power the instruments from nominal 110-240 V 50/60 Hz mains supplies.
To prevent loss of data should an external power supply be interrupted
for any reason during measurement, it is recommended that a usable set
of batteries be kept in the instrument.
In general, download all important data and remove the batteries
from the instruments when they are to be out of use for some time.
Note that when an instrument is returned to battery operation
after being powered from an external supply, either: switch the
instrument OFF then ON again, or: disconnect ALL devices from the DC
input socket, in order to avoid additional discharging of the internal
batteries.
The instrument contains a built in Lithium battery to maintain
stored data and setups for periods while the main batteries are absent.
1.9dB21 & dB22 SoundTrack Software
The CEL-6726 dB21 and CEL-6727 dB22 Software enable data collected
by the CEL-440 and CEL-480 to be downloaded to a PC for storage,
manipulation and the production of reports. Data can be exported in
ASCII format to proprietary word processing software, while profile and
statistical data can be exported as tab-delimited text files suitable for use
with spreadsheets such as Lotus 123™ (Release 2 or later) and MS
Excel™.
In general, dB22 offers similar manipulation and reporting
capabilities to dB21, with the addition of on-screen graphing and
reporting facilities.
dB21/22 software offer NO ON-LINE CONTROL.
dB21 software offers NO GRAPHICAL
capabilities; any graphing that may be required
must be accomplished within a spreadsheet.
®
The dB21 and dB22 Software runs under MS Windows
advantage of the control facilities offered by the environment, so that
once installed, users with a modest knowledge of Windows will find
Page16-CEL-440/480OperatorsHandbook
taking
Introduction
many operations are performed almost instinctively with a minimum of
learning time.
Downloading is performed under the control of the PC. Data
from each of the measurement runs in the analyser is transferred as a
series of associated datafiles together with a note file in which
comments and information can be written. Once downloaded, the user
can display the following information:
Select the language in which the instrument will present menu options.
This should be the first operation so that the displays and options will be
understood.
Similarly, to get the most accurate results, these instruments
must be operated with a known microphone directional response and be
calibrated for this response.
When the language and microphone response are known to be
correct for the proposed measurement task, go directly to Section 6
Calibration Check.
European IEC standards require measurements to be performed
with microphones that have a Free Field response, while the U.S.A.
ANSI standards require Random Incidence microphones. The
microphone response is set via the same Configuration menu.
Normally, these more permanent characteristics will need to be
set only once, as the instrument will store the settings and use them the
next time it is switched on.
All of the configuration settings that have been left highlighted on the
menu screens, plus any changes to the time and date, will be stored for
use next time the instrument is switched on.
2.2Perform Acoustic Calibration Check
The calibration level indicated by the instrument will depend on the
microphone response set. Therefore, an acoustic calibration check
should be performed only when the microphone response is correct for
the required task. When the calibration is known to be acceptable, press
MENU
once to obtain a Measurement screen or twice to obtain the Main
menu that gives access to all settings and stored data.
CEL-440/480OperatorsHandbook-Page21
9
1
M
Preliminary Operations
It is recommended that an acoustic calibration check
be performed both before and after a measurement run.
Perform the calibration check on the sound level meter using a
CEL-284/2 (Class 1) Calibrator for instruments with Type 1 accuracy and
a CEL-282 (Class 2) for Type 2 instruments. These calibrators provide a
nominal level of 114.0 dB at 1 kHz. However, the exact value to which
the instrument must be calibrated will depend on the microphone type in
use, the microphone response and local atmospheric conditions.
DO NOT remove the protective metal grid from
Type 1 microphones.
With a Type 1
instrument (i.e. 1/2"
microphone), fit the
CEL-284/2 Calibrator
directly on to the
microphone, making sure it
is pushed firmly into
contact with the shoulder in
the calibrator cavity
(Figure 5).
With a Type 2
instrument (i.e. 1/4"
microphone), fit the
CEL-4725 Coupler, supplied
with the calibrator, into the
cavity of the CEL-282 (or CEL-284/2) Calibrator making sure it is pushed
firmly into contact with the shoulder in the calibrator cavity (Figure 5).
Then fit the calibrator complete with coupler on to the instrument
microphone, again ensuring that it is pushed firmly into contact with the
shoulder in the coupler cavity.
DO NOT lay the sound level meter and calibrator on a
horizontal surface during calibration, as the combined
weight will cause the microphone to move inside the
calibrator cavity and produce an incorrect level.
Support the sound level meter and calibrator in an upright
position. To aid removal, the coupler flange does not fit tightly against
the calibrator housing.
Calibrator
Boss
/2"
icrophone
Shoulder
'O'Ring
1/4"Microphone&
PreamplifierUnit
80017
Microphone
Coupler
Figure 5: Fitting the acoustic calibrator
Page22-CEL-440/480OperatorsHandbook
Preliminary Operations
A Calibration screen is displayed at the end of the start up
sequence and this will be the normal entry to the calibration check.
However, after making changes to the instrument configuration as
described above, or after performing some other operation, the
Calibration screen can be obtained via the Calibration option on the Main
menu followed by the Microphone Option on the Calibration Menu.
When the Calibration screen shows a “Scale ... ” line instead of
“Last Cal”, the last calibration was performed via line input. Use the
Calibration and Microphone options as described above.
When the calibration check is complete, pressonce to obtain a
Measurement screen or twice to obtain the Main menu that gives
access to all settings and stored data.
3.1Select Measurement Mode
(Bandwidth & CEL-480 Run Timing)
The measurement mode specifies the bandwidth that is to be used for
measurement. Note that broadband versions of the CEL-440 have no
measurement mode option, see step 2.
With a CEL-480, measurement mode also specifies how
measurements will be timed as follows.
Timers OffManually timed. This allows the user to start and
stop a measurement run whenever they want.
DurationRun for a predetermined duration after pressing
the run key. This can be used to time a work
shift, or to measure some particularly noisy
operation with a known work cycle.
Sync timerRun for a predetermined duration that is
synchronized to start at the same time as the
next measurement period, which is set via the
Setup menu. This method is suggested when
precisely timed measurements are required.
Delay TimerRun with preset start and stop times. This allows
the instrument to start and stop up to seven
accurately timed measurement runs while
unattended. In this way, different measuring
times can be set for each day of the week with
the option of repeating them over a period of
several weeks.
Each time is set as Day:Hour:Minute, so that
delays can be set up to a month ahead. Set the
Day to 00, to use this as a 24 hour timer.
Once set, timer settings become valid for all measurement modes (i.e.
for all bandwidths and all setups).
Display screens that are available only for a CEL-480 are shown
When the required measurement mode and timer settings have been
selected, pressonce to obtain a Measurement screen or twice to
MENU
obtain the Main menu that gives access to all settings and stored data.
3.2Select Measurement Setup
(Factory- or User-Prepared Setups)
The Setup specifies the particular parameters that are to be measured.
Selection procedures for broadband and narrow band measurements
differ, as a wider choice of parameters is available for broadband, while
several frequency band scanning options are available for narrow band. A
factory configured setup and up to four user setups can be stored for
each available bandwidth.
Firmware version 2.0 introduced the facility for all broadband
measurement parameters to be obtained from a single narrowband,
provided octave or third octave band are available.
In addition, the CEL-480 can make measurements at specified
time intervals and store period noise data and exceedance (L
and profiles. Display screens that are available only for a CEL-480 are
shown with a dashed ----line.
) values
n
3.2.1 Select Broadband Measurement Setup
The factory setup for single band (broadband8) measurement contains a
set of standard parameters selected for general purpose use.
User setups contain parameters that have been selected to
perform some particular task and are saved under a user setup identity.
Changes to user setups are saved for re-use, however any changes to
the factory setup will be lost and the standard settings offered each time
the instrument is re-started. Select and save single band setups as follows.
When the required single band setup has been selected, press
once to obtain the Broadband Measurement screen or twice to obtain
the Main menu that gives access to all settings and stored data.
3.2.2 Select Narrow Band Measurement Setup
The factory setup for narrow band measurement contains a standard set
of parameters selected for general purpose use. Octave and third octave
setups have the same parameter selections that are selected in a similar
way.
User setups contain parameters that have been selected to
perform some particular task and are saved under a user setup identity.
Changes to user setups are saved for re-use. However any changes to
the factory setup will be lost and the standard settings offered each time
the instrument is re-started.
The following frequency band scanning modes are available.
ManualBands are scanned manually with the user
judging when the level is stable in a band.
(This is the factory setting.)
AutoBands are scanned automatically with the level
in a band judged stable according to the criteria
included under Specification.
TimedBands are scanned automatically with the level
in a band allowed to stabilize over user selected
preset periods between 5 and 60 seconds.
The timed setting shows how many bands are to
be scanned.
selected preset intervals between 0 minutes and
1 hour.
When using timed scanning, make sure the repeat time is greater than
the number of channels to be scanned multiplied by the stabilization
time, otherwise some bands will not have been scanned before the scan
re-starts for the next repeat period.
For example, the maximum of 30 bands x 10 s would require a
When the required narrow band setup has been selected, press
once to obtain the Narrow Band Measurement screen or twice to obtain
the Main menu that gives access to all settings and stored data.
With the instrument configured, calibrated, measurement parameters
specified and timers set as described in the preceding chapters perform
measurements as described in this chapter.
Alternatively, this chapter may also be used as a tutorial to give
an idea of how the instrument functions by accepting the currently
elected measurement parameters and the last saved calibration.
Data stored in the instrument can be recalled to the display for
inspection. This allows the operator to check the quality of results before
leaving the test site. In addition, a data set from any stored run can be
printed on a Centronics compatible printer.
Connect the printer as shown in Figure 6 and set it On-Line.
can be used to display a Header screen for the most recently saved run.
Page44-CEL-440/480OperatorsHandbook
DATA
Measure Mode
View Current
Operation
C6672 (0.5 m)
Active Printer
Interface
Centronics Printer
73.8dB
LLL
69.4dB
ASmxASAeq
LL
65.4dB 69.9dB
ASmn Zpk
CEL-440
CEL-480
C6724/2
Output
Cable
(2m)
Figure 6: Printer connection
Continue from Step 6 below. To display data from any stored run,
proceed as follows.
Formatting the memory (strictly re-formatting) is a global operation that
deletes all stored data (runs) and all user setups from every available
bandwidth.
CEL-440/480OperatorsHandbook-Page49
Operation
Use this option with great caution !
The option to format the instrument memory is displayed on the
4.5Use With Tape and DAT Recorders
(Including Calibration For Line Input)
The sound level meters can be connected to a DAT or tape recorder and
function as an accurately calibrated input system.
This enables the measured sound (or vibration) levels to be
recorded for further calculation and analysis.
When recording, connect the recorder to the OUT Phono
jack-socket in the bottom of the sound level meter (Figure 6). This socket
supplies an unconditioned AC signal with a maximum level of 0.5 V RMS.
When replaying, insert the recorded signal via the IN Phono
jack-socket shown in Figure 7. Suitable signals from other sources may
also be inserted for measurement or analysis in this way.
If an electrical calibration of the instrument using the microphone
substitution method is required, insert the calibration signal via a
CEL-516 Line Input Adaptor (or the earlier CEL-216) screwed in to a
CEL-495 Preamplifier in place of the Type 1 microphone (which may also
Page50-CEL-440/480OperatorsHandbook
Operation
2.1 mm Coaxial
DC Power
Phono Jack
Socket
AC line Out
OUTIN
Miniature 8-way DIN
Interface Socket
Phono Jack
Socket
AC line In
980037
Figure 7: Bottom panel of the instrument
be used with a Type 2 instrument), then perform an acoustic calibration
as described in Section 2.2.
4.5.1 Recording
The following guide lines are suggested for recording sound (or vibration
signals).
1.Use a CEL-284/2 or CEL-282 Calibrator to perform an acoustic
calibration check of the sound level meter.
The sound level meter should indicate the correct level for the
microphone type and response, as described in Section 3.2.
2.Connect the input of the recorder to the instrument OUT terminal.
3.Select broadband L
measure mode on the sound level meter.
ZF
4.Use the 50 - 120 dB measurement range.
5.Switch the recorder ON and start it operating in record mode.
6.For accurately repeatable recordings, adjust the Record Level
control on the recorder until the calibrator signal gives a reading
of -6 dB on the tape recorder VU meter.
Using this level means that the top of the recording range on the
tape (= 0 VU) occurs in the same place as overload (120 dB) on
CEL-440/480OperatorsHandbook-Page51
Operation
this range of the sound level meter and ensures the maximum
dynamic range for recording.
7.Record about 30 seconds of the calibration signal.
A calibration signal with a known level is inserted via the sound
level meter and recorded, so that when replayed, it gives an
indication of the record/replay characteristics of the tape.
When the recorded material is to be used for frequency analysis,
it may also be useful to calibrate each tape individually by
recording signals produced by a suitable generator at the
analyser filter band centre frequencies and then replaying them
through the relevant bands.
8.Switch the calibrator OFF and remove it from the microphone.
When the VU reading is estimated (which it usually must be at
these levels), repeatability from one series of recordings to
another may suffer. Therefore, either keep the Record Level
control in the same position for ALL recordings, or perform a
new calibration for EACH series of measurements.
9.Without touching the Record Level on the recorder, if required,
change the measurement range of the instrument to
accommodate the noise (or vibration) signal.
10.Make a note of the range used for measurement.
11.Proceed with recording the noise (or vibration) signal.
4.5.2 Replay
The following guide lines are suggested for replaying sound or vibration
signals.
1.Start from the sound level meter Line Input Calibration screen
(see Section 2.2).
Broadband L
the sound level meter.
2.Connect the recorder output to the instrument IN terminal.
3.Switch the recorder to replay mode.
Page52-CEL-440/480OperatorsHandbook
measure mode will be selected automatically by
ZF
Operation
4.Replay the recorded calibration signal.
5.Adjust the Replay Level control on the recorder so that the sound
level meter display indicates the calibration level for the
microphone and response used.
6.Stop the recorder.
The system is now calibrated to replay at accurately determined
levels with the calibration signal near the top of the input range,
which gives the widest possible replay dynamic range. However,
the instrument measurement range may need to be adjusted so
that the display scale gives a correct indication of the recorded
values.
L
ZF
Scale5Ø-12Ø
dB
dB
Dateandtimeoflastcalibration
Ø9Jan98 12.22
k-20
7.Pressto enter quick edit mode.
8.Useandto select the measurement range that was noted
for the measurement.
9.Pressagain to confirm the calibration and any scale change.
The instrument replay scale now matches the scale used for
the recording so that the levels indicated by the display will be
correct.
10.Pressto display the measurement screen, then replay the
MENU
recorded material and perform any measurements and analysis
required.
CEL-440/480OperatorsHandbook-Page53
Operation
Page54-CEL-440/480OperatorsHandbook
5.SPECIFICATION
5.1CEL-440
Instrument Standards:
Sound Level Meter to IEC 60651: 1979 Type 1
relevant.
Integrating Averaging Sound Level Meter to IEC 60804:
2000 Type 1 & Type 2 as relevant.
Sound Level Meter to ANSI S1.4: 1997 Type S(1) & Type S(2)
as relevant.
IEC 60651 and IEC 60804 Standards:
The sound level at which the sound level meter conforms to
specifications of clause 12 of IEC 60651 is 74 dB.
The tests as specified in 12.5.2.6 and 12.5.4.4 of IEC 60651
were carried out with the following cables and accessories:
The configuration for the normal mode of operation is specified
with the microphone connected via the standard preamplifier and
the following instrument parameters selected.
Frequency weighting:A-weighted sound pressure level,
Time Constant:Fast,
Measurement Range:70-140 dB.
Electromagnetic and Electrostatic compatibility:
There is no degradation in performance or loss of functionality
following the application of electrostatic discharges.
The approved configuration for reference orientation as viewed
by the operator is with the instrument in a horizontal position
facing the noise source where the receiver of the
electromagnetic field replaces the noise source.
Specification
I & Type 2I as
CEL-440/480OperatorsHandbook-Page55
Specification
All settings and orientations of the instrument have similar
radio-frequency emissions. These radio-frequency emissions are
within the specified limits of this standard.
All modes of operation and connecting device configurations
have similar immunity to power and radio frequency fields.
Time Weightings:
S, F and
RMS Frequency Weightings:
A, C and Z (LIN, 0 dB) - select only one at a time.
A and C weightings according to ANSI S1.4 and IEC 60651.
Z (LIN) weighting 20 Hz to 20 kHz to the tolerances of |EC 60651
weightings for the appropriate Type of instrument (1 or 2).
Peak Frequency Weightings:
C and Z (LIN as above) - select only one at a time.
Order 6, Type 3-D, optional range per ANSI S1.11 1986,
Class 2 to IEC 1260: 1995,
9 bands with centres from 31.5 Hz to 8 kHz (B and C models).
Third-octave Filter:
Order 6, Type 3-D, optional range per ANSI S1.11 1986,
Class 2 to IEC 1260: 1995,
28 bands with centres from 25 Hz to 12.5 kHz (C models).
Calibration:
Direct for Type 1 microphone (CEL-250).
Direct for Type 2 microphone/preamplifier unit (CEL-485).
Calibration Reference Conditions:
60 - 130 dB measurement range,
o
20
C air temperature,
65% Relative Humidity,
101.325 kPa atmospheric pressure.
114.0 dB at 1 kHz Nominal reference level.
Reference direction of incidence for all microphones is perpendicular to the front face (diaphragm surface) of the microphone.
No cable correction is required for microphone cables up to 10 m
when calibration is performed with the C6716 or C6717 Cables
fitted.
Alternative Calibration:
Suitable for direct line input from tape recorder with direct
scaling to suit recorded range.
Calendar & Clock:
Date and time accuracy better than 5 sec per day.
Data Storage:
Cumulative data set of overall values for all runs.
Broad Band Data Storage:
Manual storage of 999 complete sets of results.
Page58-CEL-440/480OperatorsHandbook
Narrow Band Data Storage:
Manual storage of 999 octave or 1/3 oct sets of results with
choice of:
Auto scanning of frequency bands with measurement and step
on each time the current band has stabilised,
Timed scanning with a predetermined time for each band that
can be set between 5 and 60 s,
Manual scanning of bands with saving of band data at each step.
Only 1 spectrum can be captured per run as the CEL-440 has
no auto repeat mode.
The memory can save a mixture of full octave and 1/3 octave
measurements. (L, L
mx,Leq
The number of bands scanned can be set between all available
and a single band.
Data Recall:
Stored results can be recalled to the display (even while a
measurement is in progress).
Data sets can be downloaded to a Centronics printer to give a
preformatted hard copy.
Data can be downloaded according to RS 232 standards to a PC
for further manipulation under the control of dB 21 or dB 22.
Outputs and Inputs:
AC output signal for tape recording or headphone.
0.5 V RMS Maximum with minimum load impedance 22 kΩ.
AC input signal for analysis inserted via a line input terminal in
the bottom panel.
0.5 V RMS Maximum with maximum source impedance 22 kΩ,
or AC input signal inserted via a CEL-516 Input Adaptor, 10 V
RMS Maximum with maximum source impedance 100 kΩ.
Effect of Temperature:
Accuracy better than ±0.5 dB over the range -10 to +50
when corrected as indicated here.
Specification
result during measurement).
o
C
CEL-440/480OperatorsHandbook-Page59
Specification
The temperature correction to be applied for CEL-250 and
MK 250 Type 1 microphones:
Corrected level = Actual level - 0.02 (t - 20)
Where t is temperature
The temperature correction to be applied for CEL-485 Type 2
microphone/preamplifier unit is applied automatically by the
preamplifier.
Storage range: -20 to +60
Effects of Humidity:
Less than 0.5 dB over the range 30 to 90% relative humidity
(provided there is no condensation), relative to the value at
65% RH and 40
Storage range: 0 to 90%RH in the absence of condensation.
Effects of Magnetic Fields:
Less than 3 dB(A) above the minimum valid reading when tested
according to IEC 60651 clause 8.4 at 50 Hz.
Effects of Vibration:
Less than 6 dB when tested according to IEC 60651 clause 8.3.
Effect of Operator:
Negligible on free field measurements when standing more than
2 m behind the SLM.
Tripod mounting:
Socket to mount on standard 1/4" camera tripod thread.
Battery Life:
At least 14 hours in broad band mode when using LR6 battery.
External Power:
9 to 14 V DC at typical 150 mA.
The batteries may safely be left in the instrument while it is
operating from an external supply. Remove the batteries when
the instrument is to be out of service for some time. A built in
lithium battery will maintain stored data and set ups.
o
C.
o
C.
o
C.
Page60-CEL-440/480OperatorsHandbook
Specification
Note that when an instrument is returned to battery operation
after being powered from an external supply, either switch the
instrument OFF then ON again, or disconnect ALL devices from
the DC input socket, in order to avoid additional discharging of
the internal batteries.
Dimensions:
340 x 100 x 40 mm (13.5 x 4 x 1.5 in),
500 gm (17.5 oz) excluding batteries.
5.2CEL-480
In addition to those listed above for the CEL-440, the CEL-480 has the
following features.
7 user selectable by date and time to nearest minute.
(Before V 2.0 firmware: 6 start and stop times.)
Broad Band Storage Capability:
999 runs, each with up to 9999 periods plus profiles up to a
maximum of 245 000 total results.
CEL-440/480OperatorsHandbook-Page61
Specification
Narrow Band Storage Capability:
999 runs, each with up to 9999 octaves or third octaves band
scans in each run up to a maximum of 16 000 octave results and
7 200 third octave results with all frequency bands measured.
No profiles are stored for narrow band.
Stored Parameters:
L, L
mx,Leq
storage of up to all 3 spectra if selected by user.
Profile times must be less than or equal to the period times.
L
Tm3
of 3 seconds.
Digital Control:
From PC to change setup, perform in-house testing for service
and recall, and to control operation of the meter’s measurements.
per band simultaneously in narrow band mode with
mx,Leq
for two broadband weightings Z and
mx,Lmn,Leq,LI
.
not available for period times that are not integer multiples
eq,LAV,Lpk,LTm3
,
5.3Filter Scanning
The operation of the sequential measurement of filter bands in the
CEL-440 and CEL-480 may be effected in two ways.
For both methods, the starting band in the scan may be preset to
any available band in the instrument, similarly the final band to be
measured can also be preset. The only restriction is that the start band
must be lower in frequency than the final band.
The first method enables the user to select a fixed measurement
interval for all measured bands ranging from 5 seconds to 1 minute in 1
second increments.
Page62-CEL-440/480OperatorsHandbook
Specification
Once scanning has been started, each band will be measured in
turn for the preset duration, with the resultant L, L
mx,Leq
the period saved in the store. The instrument then steps on to the next
band automatically and, after a suitable wait for the filter to stabilise,
starts the process of measurement again until all bands have been
measured. After measurement is completed, the results may be
examined on screen or downloaded as required.
The second method performs a self-timed autoscan. This
operates differently in that the length of time during which each filter
band is measured will be determined by how steady the readings in the
band are. The measuring process is the same as for the fixed interval
method, but the meter looks for a steady L
value, and as soon as the
eq
criteria defined for “steady” are detected, it stores this level and
immediately steps on to the next band.
Depending on the nature of the signal being analysed, the auto
scanned method may make for rapid stepping through all bands, possibly
noticeably quicker than the fixed interval method, however if the signal is
fluctuating considerably, it may take longer. This method is likely to
produce the most reliable reading of band levels when having equal time
period measurements is not of importance.
The criteria for concluding a band measurement and stepping on
to the next band are threefold.
at the end of
1.The last 4 readings of L
must all be within 0.1 dB of each other.
eq
2.The last 4 readings of Leq must all be within 0.5 dB of each other
and have successively increased and decreased, or vice-versa,
within ±1 dB (e.g. the second reading is higher than the first,
but the third reading is lower than the second and the fourth is
higher than the third).
3.The measurement has lasted for the maximum period set for the
Criteria 1 and 2 ensure that a steady or near steady reading has been
reached and provided the signal continues as in the immediate past, the
reading is the best L
value that will be obtained.
eq
CEL-440/480OperatorsHandbook-Page63
Specification
Criteria 3 ensures that even when the meter is unable to reach a
steady L
as the source permits, without measuring for extremely lengthy periods.
seconds for all octaves installed, and about 60 seconds for all third
octaves installed. Scanning times can be shortened by restricting the
number of bands to be measured.
adequate reading in the band are factory set in the meter and are listed
below. The maximum measurement time per band for Criterion 3 is
always fixed at 4 times the minimum measurement time.
, it will have measured for long enough to have as good an L
eq
Using a typical pink noise source, autoscans take around 20
A settling time for each band plus a minimum time to obtain an
Table 2: Octave Band Settling and Minimum Measurement Times
The CEL-440 and CEL-480 Sound Level Meters comply with the EMC
Directive 89/336/EEC of the European Union. They have been tested
according to the standard delivery schedule and comply with the
following standards.
EN 50081-1 : 1992, EN 50081-2 : 1993: Generic emission
standards for residential, commercial, light industry and industrial
environments.
EN 50082-1 : 1992, EN 50082-2 : 1995: Generic immunity
standards (for both RF fields and electrostatic discharge) for residential,
commercial, light industry and industrial environments.
Page66-CEL-440/480OperatorsHandbook
Parts & Warranty
6.PARTS & WARRANTY
6.1Schedule of Parts
A complete CEL-440 Sound Level Meter consists of:
CEL-440Sound Level Meter,
Plus as applicable:
CEL-250Type 1
CEL-495Type 1 Preamplifier,
or CEL-485Type 2
Assembly.
The following additional items are also included:
016004 (4 off) Alkaline-Manganese Battery 1.5 V (IEC LR6),
060279CEL-440/480 Operator’s Handbook.
A complete CEL-480 Sound Level Meter consists of:
CEL-480Sound Level Meter,
Plus as applicable:
CEL-250Type 1
CEL-495Type 1 Preamplifier,
or CEL-485Type 2
Assembly.
The following additional items are also included:
CEL-6726dB21 Download Software on a single 3
When the instrument is delivered, check that all of these items have
been supplied.
The following standard accessories may be ordered separately.
CEL-2962Windshield for Type 1 Microphones,
CEL-4672Windshield for Type 2 Microphones,
CEL-282Acoustic Calibrator Class 2L (Includes CEL-4725
Microphone Adaptor),
CEL-284/2Acoustic Calibrator Class 1L (Includes CEL-4725
Microphone Adaptor),
C6724Communication Cable to PC,
1
/2“ Electret Microphone (or MK 250),
1
/4“ Electret Microphone and Preamplifier
1
/2“ Electret Microphone (or MK 250),
1
/4“ Electret Microphone and Preamplifier
1
⁄2"
CEL-440/480OperatorsHandbook-Page67
Parts & Warranty
C6672Active Printer Cable to Centronics printer
(Requires C6724 Cable).
6.2Instrument Servicing & Warranty
To ensure its conformity with the specification, this instrument is
thoroughly inspected and it’s accuracy verified prior to dispatch. All
technical information is filed under the instrument serial number, which
should, therefore, be quoted in any correspondence.
The manufacturers undertake to rectify any defect in the
instrument that is directly attributable to faulty design or assembly, and
which becomes apparent during the warranty period. In order to take
advantage of this warranty, the instrument must be returned, carriage
paid, to the manufacturer’s factory or accredited agent, where necessary
repairs will be carried out.
The warranty period runs for 12 months from the date of receipt
of goods, with exceptions on certain specialised components supplied by
other manufacturers which may be warranted for shorter or longer
periods by their actual manufacturers. In all such cases, the benefit of
these undertakings will be passed on to the user.
CASELLA CEL liability is limited to items of their own
manufacture, and they do not accept liability for any loss resulting from
the operation or interpretation of the results from this equipment.
To obtain repair under warranty, the instrument should be
packed and returned in it’s original packing or an equivalent either to
CASELLA CEL’s local agent, or in the case of domestic sales, to the
CASELLA CEL Service Department at Bedford. Please include the
following information:
Instrument Type(s), Serial Number(s) and Firmware Version
Number(s),
Customer name and address,
Contact name and phone number,
Details of any PC and Software involved, including Version
Number(s),
Reason for returning the equipment with a detailed description
of the fault,
List of any error messages that may have been displayed.
Page68-CEL-440/480OperatorsHandbook
Parts & Warranty
The necessary adjustments or repairs will be carried out, and the
instrument returned as soon as possible.
A comprehensive Instrument Calibration Maintenance
Agreement (ICMA) scheme is available to extend the initial warranty
period of this instrument. At the end of the first warranty period, it is
recommended that the equipment be returned to the Service and
Verification Department at Bedford, where it will be inspected and
entered into the ICMA scheme as required. The warranty will then be
extended for the period stated on the individual schedule. Please contact
your local CASELLA CEL agent for full details of this service.
After the warranty has expired (except on approved accounts)
service work is undertaken against quotations, and all packing and transit
costs are charged extra.
CEL instrumentation and software is designed, manufactured, and serviced by:
CASELLA CEL LIMITED
and
CASELLA CEL INC.
CASELLA Group Companies
CEL-440/480OperatorsHandbook-Page69
Parts & Warranty
Page70-CEL-440/480OperatorsHandbook
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.