Store the Dose Badge in Stop Mode and ONLY
in a fully charged state. To ensure optimum
operation, the Dose Badge should be left on
trickle charge when not in use.
To avoid accelerated power loss DO NOT store
the Dose Badge in an area with a high ambient
lighting level.
DAMAGE WILL OCCUR if discharged batteries
are left in the Dose Badge or in the Reader
Unit.
Avoid damage by checking the battery charge
state every 2 weeks (14 days).
Caution !
Recharge the battery in the Dose Badge only
with the battery charger supplied.
Return the Dose Badge to Casella CEL or an
authorised service centre for replacement of
the internal battery after 12 months from the
date of supply.
It is recommended that the instructions and
information contained in this handbook be
read before operating the equipment.
The CEL-310 Dose Badge is a unique solution to problems associated
with the measurement of personal noise exposure. It is controlled and
downloads data via infrared signals and in consequence needs no
external controls or cables. This allows it to be used in situations where
the wearer is working in confined spaces, or where there is a risk of
cables becoming caught in machinery.
The Dose Badge can be configured to make measurements
according to ISO or OSHA standard, where it provides the following
information.
Run timeDuration of the measurement,
L
AEq/LAVG
L
/ TWADaily personal noise exposure,
EP,d
% DoseNoise exposure as a percentage of a preset
ExposureSound exposure in Pa
Peak Exceeded Whether 140 dB(C) Peak has been exceeded.
Average sound pressure level during the
measurement period,
total,
2
h (Pascal squared hours),
Figure 1: Dose Badges and the Reader Unit
CEL-310DoseBadgeUsersHandbook-Page5
Introduction
The system is delivered with the Dose Badge configured for
either ISO or OSHA measurement. The user can change the configuration
by means of the Reader Unit, allowing the Dose Badge to be altered for
different Standards and Guidelines. A system may be founded with a
single Dose Badge and Reader, then expanded by adding further Dose
Badges and an appropriate number of chargers.
This manual describes the operation of the Dose Badge, Reader
Unit and Windshield.
1.1The System
A Dose Badge system consists of a number of individual Dose Badges
controlled by a single Reader Unit. The Reader provides the method to
calibrate, start, stop and download measurements from a Dose Badge,
then download this information to a PC.
The system can be configured to measure according to ISO or
OSHA standards.
Those authorities who follow European Union measurement
procedures should use the ISO configuration. This satisfies the
requirements of IEC 61252 and employs the following settings.
Q = 3Energy exchange rate,
80 dBThreshold level,
85 dBCriterion level,
8 hoursCriterion time,
Freq. weighting A for all sound level measurements,
C for peak sound pressures,
Time weighting None.
Those authorities who follow U.S. measurement procedures
should use the OSHA configuration. This satisfies the requirements of
ANSI S1.25-1991 and employs the following settings.
Q = 5Energy exchange rate,
80 dBThreshold level,
90 dBCriterion level,
8 hoursCriterion time,
Freq. weighting A for all sound level measurements
C for peak sound pressures,
Time weighting Slow weighting.
Page6-CEL-310DoseBadgeUsersHandbook
M
Introduction
When a measurement is downloaded from the Reader to the
Dose Badge Software, the configuration of the Dose Badge is shown.
PLEASE NOTE that the Microphone Response shown as “F”
does not indicate Fast time weighting, but rather Free Field microphone
response.
The Reader controls the Dose Badge via an infrared link with
information sent backwards and forwards in a similar way to a television
remote control.
PLEASE NOTE that the range of communication is very much
shorter than a television remote control.
Each Dose Badge contains a rechargeable battery, where a
single charge of the battery is intended to supply sufficient power for a
measurement lasting up to 16 hours. However, the Dose Badge can run
for much longer when the battery is fully charged.
1.2Layout & Controls
The layout of the Dose Badge is shown in Figure 2.
The microphone is mounted at the top of the case in a shock
mount that reduces the noise generated from movement or handling.
A window allows access for
the infrared control signals from the
Reader Unit. Ensure that this window
is clean and free from dust and dirt.
The charging connection for the
internal battery is at the centre of the
bottom panel. This threaded insert is
also used to attach the Dose Badge.
The Reader Unit, shown in
Figure 3, has an integral acoustic
calibrator that has been specifically
designed to suit the unique shape and
performance of the Dose Badge. The
calibrator cavity is located in the top
panel of the Reader.
icrophone
Infrared
window
0000
SerialNumber
CEL-310
ANSI S1.25-1991Class: see manual
or IEC 61252: 1993
Charging/
mounting
stud
www.casellacel.com
03002
Figure 2: The Dose Badge
CEL-310DoseBadgeUsersHandbook-Page7
Introduction
Acoustic calibrator
in centre of top panel
Reset Dose Badge
Calibrate Dose Badge
Start Dose Badge
measurement
Accept changes &
select options
Display menu
RS232 connector to
side of bottom panel
03001
Dose Badge Reader
Reset
Cal
OK
Menu
Read
Stop
Run
Cancel
Navigator keys to select/change/enter data/
step through information shown on the display
Figure 3: The Reader
Infrared window to
side of top panel
Display
Read measurement
from Dose Badge
Stop Dose Badge
measurement
Cancel & exit from
menu
Power On/Off
An infrared port beside the acoustic calibrator in the top panel, is
used to communicate with the Dose Badge. Ensure that this window is
clean and free from dust and dirt.
The RS232 connector used to download data to a PC is located in
the bottom panel of the Reader, while the battery compartment is located
in the rear panel.
Page8-CEL-310DoseBadgeUsersHandbook
Preparation
2.PREPARATION
It is recommended that the instructions and information contained in this
handbook be read before operating the equipment.
2.1Unpack the Equipment
Carefully remove all components of the Dose Badge system from the
shipping container or carrying case and inspect them for possible damage
or missing items. If there appears to be damage or something missing,
please contact Casella CEL or your local representative immediately. The
following components should be included.
CEL-310Dose Badge,
CEL-315Dose Reader Unit,
C6801/2RS232 Cable (2 m),
-PC18Mains Power Supply (with UK, EU or USA plug
as ordered,
CEL-6803Hard Hat Mounting Kit,
CEL-6804dBLink Dose Badge Software on a CD,
CEL-6805dBDatabase Software (ordered separately),
CEL-6806Carrying Case for Dose Badge System with
up to 10 Badges,
060349/HBOperating Manual.
At least one of the following items will also be supplied as ordered
CEL-6807/2Charger for 2 Dose Badges,
CEL-6807/5Charger for 5 Dose Badges,
CEL-6807/10Charger for 10 Dose Badges.
Please check that any other components that were ordered have been
included with the equipment.
2.2Quick Start
This is the recommended procedure for performing personal sound
exposure measurements using a single Dose Badge and Reader.
1.Charge the Dose Badge battery (Section 2.3).
CEL-310DoseBadgeUsersHandbook-Page9
Preparation
2.Insert usable batteries into the Reader Unit (Section 2.4).
3.Use the Reader menu to check that the configuration set for
the Dose Badge is suitable for the proposed measurements
(Section 2.5).
4.Insert the Dose Badge into the calibration cavity on the Reader
(Section 4.1).
5.Pressto reset the Dose Badge (Section 4.1).
6.Pressto calibrate the Dose Badge automatically (Section 4.1).
7.Use the Reader to reset the Dose Badge (Section 4.2).
8.Remove the Dose Badge from the calibration cavity and attach
9.Pressto start the measurement (Section 4.4).
10.At the end of the measurement shift, pressto stop the run
11.Remove the Dose Badge from the wearer and push it into the
12.Pressto perform a second calibration check (Section 4.1).
13.Pressto transfer the Dose Badge session data to the Reader
14.View the results (Section 4.8) and print a hard copy (Section 4.7).
15.Download measurements to the PC software (Section 4.9).
Reset
Cal
it to the wearer (Section 4.3).
Run
Stop
(Section 4.5).
calibration cavity on the Reader (Section 4.1).
Cal
Read
via the infrared link (Section 4.6).
2.3Charge the Dose Badge Battery
The Dose Badge contains a rechargeable battery pack which is NOT
USER REPLACEABLE.
The indicators on the Dose Badge charger show the following
states:
Red FlashingDose Badge not connected,
Red ConstantLow battery,
Green FlashingCharging cycle,
Green ConstantTrickle charging.
Page10-CEL-310DoseBadgeUsersHandbook
Preparation
Carefully screw
the Dose Badge
on to the stud
Charging studs
Power socket
Figure 4: Fitting a Dose Badge on to a Charger
1.Make sure the Dose Badge is in Stop Mode.
2.Carefully screw it on to the charging stud of a CEL-6807 Charger.
The five badge CEL-6807/5 charger is shown in Figure 4.
3.Connect the Charger unit to the mains supply and switch the
power ON.
Charging indicators
03004
When the charging process begins, the battery will first be fully
discharged to ensure that it does not develop a “memory”.
During this period, which may take up to 2 hours, the green
charge indicator should give “short” flashes (i.e. off more than
lit).
Once the battery has been discharged, the green charge indicator
will give “long” flashes (i.e. lit more than off) to indicate that the
battery is receiving a charge. Charging may take up to 12 hours.
When charging is finished, the green indicator will remain on
continuously to show that the Dose Badge is receiving a trickle
charge to keep it topped up.
CEL-310DoseBadgeUsersHandbook-Page11
Preparation
4.To ensure optimum operation, leave the Dose Badge on trickle
charge when not in use.
2.4Insert Batteries into the Reader Unit
The Reader Unit requires two AA type batteries for operation. These are
located under the plastic cover on the back of the instrument.
1.Before changing the main batteries, make sure the Reader is OFF.
The memory and clock in the Reader are protected for up to
10 minutes when the main batteries are removed (and it is OFF).
2.Locate the battery compartment cover and remove it from the
instrument.
3.Remove the two batteries from the compartment and replace
them with two batteries of equivalent type and quality.
The recommended battery should be alkaline type AA (LR6 /
AM3 / MN1500 / Mignon).
4.Make sure that the polarity of the batteries is correct, as indicated
by the diagram in the compartment.
5.Replace the battery cover.
2.5Switch the Reader Unit ON & Check
Configuration
1.Presson the Reader to switch it ON.
(To switch OFF, pressagain.)
It will display a start screen that is followed by a screen showing
Reader status information.
001 of 010
Serial No.
Meas. time
Meas. Date
Run Time
17:05:28 20/06/02
NNNNNN
16:55
06/02/02
07:45:12
M-01
The Reader will automatically switch itself OFF after four minutes
of inactivity.
Page12-CEL-310DoseBadgeUsersHandbook
Battery condition
Current time and date
shown in message line
Preparation
The display backlight will be lit when any key is pressed and will
automatically switch off after 10 seconds to preserve battery
power.
2.If the display contrast is not satisfactory, pressto display the
Main Menu.
The menu structure for the Reader is shown on a fold out sheet
at the back of this handbook.
Reader Main Menu
Clear Memory
Set Clock
Print Current Data
Print All Data
10 Measurements
M-02
Options
Message line
Active key option
The Main Menu offers a series of options that scroll up or down
the screen when the reverse video cursor is moved.
3.Useorto move the cursor and select Adjust LCDContrast.
4.Pressto display the Adjust LCD Contrast Menu.
OK
Two options are offered: Lighten or Darken.
Menu
5.Move the cursor to select the required option.
6.Pressagain to save the setting and return to the Main Menu.
(Alternatively, pressto save the setting and return to the
Menu
OK
status screen.)
7.To check the configuration, pressto display the Main Menu.
Menu
8.Check the clock setting by usingto move the cursor and
select Set Clock.
The time and date are indicated in the message line at the bottom
of the display.
9.Check the time history (profile) status by usingto select
Time History Control.
The current setting is indicated in the message line.
CEL-310DoseBadgeUsersHandbook-Page13
Preparation
10.Check the setting for criterion time (i.e. duration of a nominal
working day) by usingto select Criterion Time.
The criterion time is indicated in the message line.
11.Check the setting for criterion level (i.e. level to give 100% dose
in a “normalised” 8 hour day) by usingto select Criterion
Level.
The criterion level is indicated in the message line.
12.Check the dose badge settings by usingto select DoseBadge Settings.
The current settings are indicated in the message line.
13.When all of the settings are correct for the proposed job, proceed
directly to Chapter 4: Operation.
However, if any of the settings checked in this section are
incorrect, change them according to Chapter 3: Configuration.
Page14-CEL-310DoseBadgeUsersHandbook
Loading...
+ 31 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.