13. DISPOSAL OF THE PRODUCT ........................................................... 27
Type 8025/8035 battery powered
2
English
Type 8025/8035 battery powered
About this manual
1. ABOUT THIS MANUAL
This manual describes the entire life cycle of the device. Please
keep this manual in a safe place, accessible to all users and any new
owners.
This manual contains important safety information.
Failure to comply with these instructions can lead to hazardous
situations.
• This manual must be read and understood.
1.1. Symbols used
danger
Warns against an imminent danger.
• Failure to observe this warning can result in death or in serious
injury.
Warning
Warns against a potentially dangerous situation.
• Failure to observe this warning can result in serious injury or
even death.
attention
Warns against a possible risk.
• Failure to observe this warning can result in substantial or minor
injuries.
note
Warns against material damage.
• Failure to observe this warning may result in damage to the
device or system.
Indicates additional information, advice or important
recommendations.
Refers to information contained in this manual or in other
documents.
→ Indicates a procedure to be carried out.
1.2. Definition of the word "device"
The word "device" used within this manual refers to the battery
powered flowmeter-totalizer type 8025 or 8035.
English
3
Type 8025/8035 battery powered
Intended use
2. INTENDED USE
Use of the device that does not comply with the instructions
could present risks to people, nearby installations and the
environment.
• The flowmeter-totalizer 8025 or 8035 is intended to measure
the flow rate and totalize the volume of liquid.
• This device must be protected against electromagnetic interference, ultraviolet rays and, when installed outdoors, the effects
of climatic conditions.
• This device must be used in compliance with the characteristics
and commissioning and use conditions specified in the contractual documents and in the user manual.
• Requirements for the safe and proper operation of the device
are proper transport, storage and installation, as well as careful
operation and maintenance.
• Only use the device as intended.
→ Observe any existing restraints when the device is exported.
3. BASIC SAFETY INFORMATION
This safety information does not take into account:
• any contingencies or occurences that may arise during installation,
use and maintenance of the devices.
• the local safety regulations for which the operating company
is responsible including the staff in charge of installation and
maintenance.
Danger due to high pressure in the installation.
• Stop the circulation of fluid, cut off the pressure and drain the
pipe before loosening the process connections.
Danger due to electrical voltage.
• Put the power switch to OFF before carrying out work on the
device.
• Shut down and isolate the electrical power source before carrying out work on the system.
• Observe all applicable accident protection and safety regulations for
electrical equipment.
Danger due to high temperatures of the fluid.
• Use safety gloves to handle the device.
• Stop the circulation of fluid and drain the pipe before loosening
the process connections.
Danger due to the nature of the fluid.
• Respect the prevailing regulations on accident prevention and safety
relating to the use of aggressive fluids.
4
English
Type 8025/8035 battery powered
Basic safety information
Various dangerous situations
To avoid injury take care:
• not to use the device for the measurement of gas flow rates.
• not to use the device in explosive atmospheres.
• not to use the device in an environment incompatible with the
materials it is made of.
• not to use fluid that is incompatible with the materials the device
is made of.
• not to subject the device to mechanical loads (e.g. by placing
objects on top of it or by using it as a step).
• not to make any external or internal modifications to the device.
• to prevent any unintentional power supply switch-on.
• to ensure that installation and maintenance work are
carried out by qualified, authorised personnel in possession of
the appropriate tools.
• to use the device only if in perfect working order and in compliance with the instructions provided in the operating instructions.
• to observe the general technical rules when installing and using
the device.
note
The device may be damaged by the fluid in contact with.
• Systematically check the chemical compatibility of the component materials of the device and the fluids likely to come into
contact with it (for example: alcohols, strong or concentrated
acids, aldehydes, alkaline compounds, esters, aliphatic compounds, ketones, halogenated aromatics or hydrocarbons,
oxidants and chlorinated agents).
note
Elements / Components sensitive to electrostatic discharges
• This device contains electronic components sensitive to electrostatic discharges. They may be damaged if they are touched by
an electrostatically charged person or object. In the worst case
scenario, these components are instantly destroyed or go out of
order as soon as they are activated.
• To minimise or even avoid all damage due to an electrostatic
discharge, take all the precautions described in the EN 613405-1 and 5-2 norms.
• Also ensure that you do not touch any of the live electrical
components.
English
5
1
2
Type 8025/8035 battery powered
General information
4. GENERAL INFORMATION
4.1. Manufacturer's address and
international contacts
To contact the manufacturer of the device, use following address:
Bürkert SAS
Rue du Giessen
BP 21
F-67220 TRIEMBACH-AU-VAL
You may also contact your local Bürkert sales office.
The addresses of our international sales offices are available on the
internet at: www.burkert.com
4.2. Warranty conditions
The condition governing the legal warranty is the conforming use
of the 8025 or the 8035 in observance of the operating conditions
specified in this manual.
4.3. Information on the Internet
You can find the user manuals and technical data sheets regarding
the type 8035 or 8025 at: www.burkert.com
5. DESCRIPTION
5.1. Area of application
The battery powered flowmeter-totalizer 8025 or 8035 is intended
to measure the flow rate and totalize the volume of liquid.
5.2. General description
The device type 8025 is made up of a flow sensor with paddle
wheel and an electronic module with display.
The device type 8035 is made up of an SE35 electronic module
and of an S030 fitting-sensor incorporating the paddle wheel which
measure the flow.
The device is fed with 2 batteries.
5.3. Description of the name plates
1. Type of the device
TOTAL :8025- SUPPLY: BATT9V
COIL SHORT - OUT: /
S/N 1439
Made in France
00418403 W43MA
Fig. 1: Name plate of the
8025
2. Power supply
3. Type of output: without
4. Manufacturing code
5. Conformity logo
457836
6. Specification of the flow sensor
7. Serial number
8. Order code
6
English
12
Type 8025/8035 battery powered
Technical data
1. Type of the device
TOTAL :SE35/8035 COIL 9V
IP65
S/N 1439
Made in France
00423921 W43MA
2. Specification of the flow sensor
3. Power supply
4. Manufacturing code
5. Conformity logo
457836
6. Protection rating
Fig. 2: Name plate of the
8035
7. Serial number
8. Order code
5.4. Available versions
• The flowmeter-totalizer type 8035 is made up of an SE35 elec-
tronic module with order code 423921 and of an S030 fittingsensor which has to be ordered separately.
• The flowmeter-totalizer type 8025 exists in 2 versions:
Type of flow sensor Seal materialOrder code
short, with coilFKM
long, with coilFKM
418403
418405
6. TECHNICAL DATA
6.1. Conditions of use
Ambient temperature0 to +60 °C
Air humidity< 80%, without condensation
Height above see
level
Protection ratingIP65, cover lid screwed tight and screwed
6.2. Conformity to standards and
directives
The device conforms to the EC directives through the following
standards:
• EMC: EN 61000-6-2, EN 61000-6-3
• LVD: EN 61 010-1
• Vibration: EN 60068-2-6
• Shock: EN 60068-2-27
• Pressure: article 3§3 of the Pressure Directive 97/23/CE. Acc. to
the Pressure Directive 97/23/CE: the device can only be used in
the following cases (depending on the max. pressure, the DN of
the pipe and the fluid)
max. 2000 m
plugs.
English
7
203
Type 8025/8035 battery powered
Technical data
Type of fluidConditions
8025 8035
Fluid group 1 §
1.3.a
Fluid group 2 §
1.3.a
Fluid group 1 §
1.3.b
Fluid group 2 §
1.3.b
1)
For the type 8035 : S030 fitting-sensor, DN6 to DN65, in PP,
only DN25DN ≤ 25
DN ≤ 32
or DN > 32 and
PNxDN ≤1000
DN ≤ 32
or DN > 32 and
PNxDN ≤ 1000
DN ≤ 25
or DN > 25 and
PNxDN ≤ 2000
PNxDN≤ 2000
DN ≤ 400DN ≤ 200
PVC, PVDF, brass or stainless steel.
6.3. General technical data
6.3.1. Mechanical data
PartMaterial
Housing / sealPC / NBR
Cover with lid / sealPC / silicone
Front foilPolyester
1)
ScrewsStainless steel
NutPC
PartMaterial
Flow sensor (exposed to the
PVDF / FKM
fluid) / seal (only with 8025)
Axis and bearings of the paddle
ceramic
wheel
R=90
91
21
85.5
102
164.5
49.5
30
88
136
88
Fig. 3: Dimensions in mm of the 8025
M20x1.5 screw plugs / sealPA / neoprene
8
English
Type 8025/8035 battery powered
Technical data
Table 1: Height H in mm of flowmeter-totalizer 8025 associated to
an S020 fitting
H
H
T-fitting
Saddle
Spigot, in
plastic
Welding tab
with radius,
in stainless
steel
DN20185
DN25185
DN32188
DN40192188
DN50198223193
DN65198221206199
DN80226212204
DN100231219214
DN110227
DN125234254225
DN150244261236
T-fitting
Saddle
Spigot, in
plastic
Welding tab
with radius,
in stainless
steel
DN180268
DN200280282257
DN250300317
DN300312336
DN350325348
DN400340
English
9
Type 8025/8035 battery powered
Technical data
180
91
21
88
88
95
105
75
104
114
Fig. 4: Dimensions in mm of the SE35 electronic module
Table 2: Height H in mm of SE35 electronic module associated to
an S030 fitting-sensor
DN (mm)060815202532405065
H (mm)134 134 139 137 137 140 144 151 151
1)
H
The height H does not
depend on the type of connection or on the material.
6.3.2. General data
Type80258035
Pipe diameter
Type of fittingS020S030
Fluid temperature
Fluid pressure
Flow rate measurement
• Measurement range
• Measurement error
with a teach-in
procedure (teach-in
function)
• Measurement error
with K factor of fitting
used
• Linearity
• Repeatability
Power supply2 batteries of 9 V (6LR61)
1)
Determined in the following reference conditions: fluid = water,
water and ambient temperatures = 20 °C, upstream and downstream distances respected, appropriate pipe dimensions.
DN20 (except for DN specified
p.13) to DN400
DN6 to
DN65
The fluid temperature may be restricted by the
fluid pressure and by the material of the fitting
used. See Fig. 5 et Fig. 6
The fluid pressure may be restricted by the fluid
temperature and by the material of the fitting
used. See Fig. 5 et Fig. 6
• 0,2 to 10 m/s
• ±0,5 % of the full scale
• ± (0,5% of the full scale + 2,5 % of the
measured value) 1)
• ±0,5 % of the full scale 1)
• ±0,4 % of the measured value
Minimum autonomy: 1 year at 20° C
10
English
11
Type 8025/8035 battery powered
Installation
PVDF
PVC + PP
PVC (PN10)
application range
Metal
PVDF (PN10)
PP (PN10)
+80
P fluid (bar)
10
9
8
7
6
5
4
3
2
1
0
-150+20 +40+60
Fig. 5: Type 8025, temperature-pressure dependency curves,
depending on the S020 fitting material
application range
PVC + PP
PVC (PN10)
A
Metal
PVDF (PN10)
PP (PN10)
P fluid (bar)
16
15
14
13
12
11
10
9
PVDF
8
7
6
5
4
3
2
1
0
-10 +10 +30 +50+70 +90 +110
-30
Fig. 6: Type 8035, temperature-pressure dependency curves,
depending on the S030 sensor-fitting material
T fluid (°C)
T fluid (°C)
7. INSTALLATION
7.1. Safety instructions
danger
Risk of injury due to high pressure in the installation.
• Stop the circulation of fluid, cut off the pressure and drain the
pipe before loosening the process connections.
Risk of injury due to electrical voltage.
• Shut down and isolate the electrical power source before carrying out work on the system.
• Put the power switch to OFF before carrying out work on the
device.
• Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high fluid temperatures.
• Use safety gloves to handle the device.
• Stop the circulation of fluid and drain the pipe before loosening
the process connections.
Risk of injury due to the nature of the fluid.
• Respect the prevailing regulations on accident prevention and
safety relating to the use of aggressive fluids.
English
11
Warning
Risk of injury due to non-conforming installation.
• The fluid installation can only be carried out by qualified and
skilled staff with the appropriate tools.
• Respect the assembly instructions for the fitting used.
Risk of injury due to unintentional switch on of power supply
or uncontrolled restarting of the installation.
• Take appropriate measures to avoid unintentional activation of
the installation.
• Ensure that the restart of the installation is controlled after any
interventions.
Protect this device against electromagnetic interference, ultraviolet rays and, when installed outdoors, the
effects of the climatic conditions.
7.2. Installation onto the pipe
danger
Type 8025/8035 battery powered
Installation
The 8025 flowmeter-totalizer has to be inserted into an S020 fitting
mounted on a pipe.
The SE35 electronic module has to be inserted into an S030 fittingsensor mounted on a pipe.
7.2.1. Installation of the fitting onto the
pipe
→ Choose an S020 or S030 fitting appropriate to the velocity of
the fluid inside the pipe: refer to the graphs at right.
→ Install the fitting on the pipe as described in the operating
instructions of the fitting used.
Risk of injury due to high pressure in the installation.
• Stop the circulation of fluid, cut off the pressure and drain the
pipe before loosening the process connections.
Risk of injury due to the nature of the fluid.
• Respect the prevailing regulations on accident prevention and
safety relating to the use of aggressive fluids.
12
English
gpm
Type 8025/8035 battery powered
Installation
Table 3: Diagram flow/ fluid velocity/ DN of S020 fitting (on left) and S030 fitting-sensor (on right)
Example: for a nominal flow = 10 m3/h and an optimal flow velocity between 2 and 3 m/s.
→ use a DN40 fitting (or DN50 for the asterisked fittings).
US gpm
20000
10000
5000
2000
1000
500
200
100
50
20
10
5
2
1
0.5
0.2
0.1
0.05
Flow rate
l/min
100000
50000
30000
20000
10000
5000
3000
2000
1000
500
200
100
50
20
10
5
2
1
0.5
0.2
3
/h
m
5000
2000
1000
500
200
100
50
20
10
5
2
1
0.5
0.2
0.1
0.05
0.02
0.01
0.10.3 0.5135 10
0.30.513510
Fluid velocity
DN 400
DN 350
DN 300
DN 250
DN 200
DN 150
DN 125
DN 100
DN 80
DN 65
DN 50 (DN65)*
DN 40 (DN50)*
DN 32 (DN40)*
DN 25 (DN32)*
DN 20 (DN25)*
Example
m/s
30
fps
Flow rate
m3/h
l/min
2000
1000
500
200
100
50
20
10
1)
0.5
0.2
0.1
0.05
500
5000
200
2000
100
1000
50
500
20
200
10
100
5
50
2
20
5
2
1
1
10
0.5
5
3
0.2
2
0.1
1
0.05
0.5
0.02
0.3
0.2
0.01
0.1 0.2 0.3 0.5 1235 10
0.3 0.5 1235 10 20 30
Fluid velocity
* For the fittings:
• with external thread connections acc. to SMS 1145,
• with welding end connections acc. to SMS 3008, BS 4825 / ASME BPE or DIN 11850 Rg 2,
• Clamp acc. to SMS 3017 / ISO 2852, BS 4825 / ASME BPE or DIN 32676.
1)
The device cannot be installed on the DN20 fittings listed above
DN 65
(DN65)*
DN 50
DN 40
(DN50)*
DN 32
(DN40)*
DN 25
(DN32)*
DN 20
(DN25)*
DN 15
(DN15 / DN20)*
DN 08
DN 06
m/s
fps
English
13
Type 8025/8035 battery powered
Installation
7.2.2. Installation of 8025 on the S020
fitting
→ Insert nut 3 on the fitting 5.→ Insert the snap ring 2 into the groove 4.→ Check that seal 6 is correctly inserted on the
flow sensor.
→ Slowly insert the device 1 into the fitting.
If the mounting is correctly done the device cannot
turn around anymore.
6
1
2
→ Hand lock the assembly with the nut 3.
Fig. 7: Installation of 8025 on the S020 fitting
7.2.3. Installation of the SE35 electronic
module on the S030 fitting-sensor
→ Insert the electronic module 2 in the
S030 fitting-sensor.
→ Secure with a 30° clockwise rotation.→ Tighten the lateral screw 3 to lock
the electronic module with the
fitting-sensor.
3
7.3. Description of the electronic
board connections
Terminal block:
+ : positive supply (red wire)
- : negative supply (black wire)
3
4
5
2
1
Connector : branchement du
capteur de débit
Terminal block
OFF ON
Connector
Fig. 9: Electronic board
connections
Fig. 8: Installation of the SE35 electronic module on the S030
fitting-sensor
14
English
Type 8025/8035 battery powered
Operating and commissioning
8. OPERATING AND
COMMISSIONING
8.1. Safety instructions
Warning
Risk of injury due to non-conforming operating.
Non-conforming operating could lead to injuries and damage the
device and its surroundings.
• The operators in charge of operating must have read and understood the contents of this manual.
• In particular, observe the safety recommendations and intended
use.
• The device/installation must only be operated by suitably trained
staff.
Warning
Danger due to non-conforming commissioning.
Non-conforming commissioning could lead to injuries and damage
the device and its surroundings.
• Before commissioning, make sure that the staff in charge have
read and fully understood the contents of the manual.
• In particular, observe the safety recommendations and intended
use.
• The device / the installation must only be commissioned by
suitably trained staff.
8.2. Switch on the device and lock/
unlock the ENTER key
→ Unfasten the screw and flip
the lid.
→ Unfasten the 4 screws and
take off the cover of the
housing.
Switch on the device:
→ Put the switch 1 to ON
ENTER key unlocked
Fig. 11: Lock and unlock the ENTER key
Switch 1
OFF ON
Switch 2: see
Fig. 11
Fig. 10: Electronic board
ENTER key locked
English
15
ENTER
Type 8025/8035 battery powered
Operating and commissioning
8.3. Operating levels of the device
The device has two operating levels: the Process level and the Configuration level.
The Process level makes it possible:
• to read the flow rate measured by the device and the value of the
totalizers, the main and the daily one.
• to reset the daily totalizer.
• to access the Configuration level.
The Configuration level makes it possible to set the device
parameters.
Table 4: Default settings of the device
FunctionDefault value
LANGUAGEEnglish
UNIT of the flow ratel/s
UNIT of the totalizerslitre
Number of decimals2
K FACTOR46,60
FILTERFilter 2
Process level
Configuration level
12.6 l/s
87654 l
+
> 5 s
2)
231 l.
+
0......9
> 2 s
To reset the daily
totalizer (identified by
a dot after the volume
units).
2)
Possible if the ENTER key is unlocked (see chap. 8.2)
Fig. 12: Diagram of the levels of the device
Once the settings of a parameter is done if you do not want
to adjust another parameter, go to the "END" parameter of
the Configuation level and press
ENTER
to save the settings
and go back to the Process level.
lANGUAGE
UNit
K-FACtOR
FiltER
tOtAl
END
Process level
16
English
Type 8025/8035 battery powered
Operating and commissioning
8.4. Description of the navigation
keys
Fig. 13: Foil of the device
You want to...Press...
Move in the parameters of
a level.
Access the Configuration
level.
To go to the next parameter.
ENTER
+ simultaneously for 5 s,
in the Process level
You want to...Press...
Reset the daily totalizer.
0......9
+ simultaneously for 2 s,
when the daily totalizer is displayed in
the Process level
Select the displayed
ENTER
parameter.
Confirm the displayed
ENTER
value.
Modify a numerical value.
•
0......9
•
the blinking digit.
0......9
•
point.
to increase the blinking digit.
to select the digit at the left of
+ to move the decimal
English
17
Type 8025/8035 battery powered
Operating and commissioning
8.5. Details of the Process level
This level is active by default when the device is energized.
Value of the measured flow rate, displayed
in the unit chosen in the "UNIT" parameter of
the Configuration level.
Value of the main totalizer, total volume of
fluid counted by the device since the last
reset.
Value of the daily totalizer (identified by a
dot after the volume units), volume of fluid
counted by the device since the last reset.
0......9
+
> 2 s
Resetting the daily
totalizer.
0......9
12.6 l/s
87654 l
231 l.
8.6. Details of the Configuration level
To access the Configuration level, simultaneously press keys
ENTER
for at least 5 s.
This level comprises the following configurable parameters:
lANGUAGE
UNit
K-FACtOR
0......9
FiltER
tOtAl
Choosing the display language
Choosing the flow rate unit, the
number of decimals and the unit
the totalizers are displayed in.
Entering the K factor of the fitting
used or have it defined through a
teach-in procedure.
Choosing the degree of attenuation
of the measured flow rate variations, with effect on the displayed
flow rate.
Resetting both totalizers.
Going back to the Process level
END
and saving the settings.
Fig. 14: Details of the Process level
18
English
Fig. 15: Diagram of the Configuration level
Type 8025/8035 battery powered
Operating and commissioning
8.6.1. Choosing the display language
When the device is energized for the first time, the display language
is English.
lANGUAGEENGlish
DEUtsCh
FRANçAis
itAliANO
→ Confirm the dis-
played language:
The selected language is immediately
active.
UNit
Fig. 16: Diagram of the "LANGUAGE" parameter of the
Configuration level
8.6.2. Choosing the flow rate units, the
number of decimals and the units of
the totalizers
The max. flow rate that can be displayed depends on the
number of decimals chosen:
• 9999 if the number of decimals = 0,
• 999,9 if the number of decimals = 1,
• 99,99 if the number of decimals = 2,
• 9,999 if the number of decimals = 3.
The maximal volume that can be displayed by the totalizers
depends on the unit of the chosen volume:
• 9 999 999 if the chosen volume unit is "litre",
3
• 999 999 if the chosen volume unit is "m
The "UNIT" parameter makes it possible to choose:
• the flow rate units.
• a fixed number of decimals (choose 0, 1, 2 or 3) to display the
flow rate in the Process level.
• the volume units of the totalizers if the unit previously chosen is in
litres or in m
3
.
", or "gallon".
English
19
Type 8025/8035 battery powered
Operating and commissioning
UNit
FlOw RAtE
lit/sEC
lit/miN
lit/h
m3/miN
m3/h
Us GAl/s
Us GAl/m
Us GAl/h
imp GA/s
imp GA/m
GA/h
tOtAl
K-FACtOR
Fig. 17: Diagram of the "UNIT" parameter of the Configuration level
→ Choose the flow rate unit.→ Confirm
DEC pt 0
DEC pt 1
DEC pt 2
DEC pt 3
If the chosen unit is in litres or m3.
litRE
→ Choose the totalizer unit.→ Confirm
m3
→ Choose the number
of decimal positions.
→ Confirm
If the chosen unit
is in gallons.
20
English
Type 8025/8035 battery powered
Operating and commissioning
8.6.3. Entering or determining the K factor of the fitting used
The device determines the flow rate in the pipe using the fitting K factor.
The K factor of the fitting used can either be entered (see Fig. 18) or determined using a teach-in procedure (See Fig. 19).
K-FACtOR
ENTER
tEACh NOK=000.0
ENTER
The display shows the last K factor of the fitting, may it have been
entered or determined by a teach-in procedure.
→ Edit the parameter.
0......9
ENTER
Fig. 18: Entering the K factor of the fitting used
→ Enter the K factor (value between 0,0001 and 9999,9) of the
K=48.30FiltER
fitting used.
→ Confirm the displayed value.
English
21
The teach-in procedure is done depending on a known volume.
→ Prepare a tank with a known volume.→ Stop the fluid circulation.→ confirm "TEACH Y.": "FILL END." is displayed.
Type 8025/8035 battery powered
Operating and commissioning
K-FACtORtEACh NO
ENTER
tEACh YFill END
The device calculates
the K factor of the
fitting and displays it.
ENTER
Fig. 19: Teach-in procedure depending on a volume
22
English
K=48.30FiltER
ENTER
ENTER
ENTER
0000.0 l
0......9
0100.0 l
→ Charge the pipe to fill the tank.→ When the tank is full, confirm "FILL END":
Do not enter a volume value of 0,0
because the calculated K-factor would
be of 0,0 and the device would not be
operating.
→ Enter the volume (value > 0,0 and ≤ 9999,9)
of fluid that passed in the circuit.Confirm.
Type 8025/8035 battery powered
Maintenance and troubleshooting
8.6.4. Configuring the filter
This parameter makes it possible to dampen the fluctuations ot the
display.
Ten filters are available.
FiltERFiltER 0
→ Choose the filter.
0......9
FiltER 9
tOtAl
Fig. 20: Diagram of the "FILTER" parameter of the Configuration level
8.6.5. Resetting both totalizers
tOtAlREs NO
→ Confirm
→ Choose to reset the
two totalizers or not.
REs YEs
END
Fig. 21: Diagram of the "TOTAL" parameter of the Configuration
level
The daily totalizer can be reset from the Process level (see
chap. 8.5).
→ Confirm
9. MAINTENANCE AND
TROUBLESHOOTING
9.1. Safety instructions
danger
Risk of injury due to high pressure in the installation.
• Stop the circulation of fluid, cut off the pressure and drain the
pipe before loosening the process connections.
Risk of injury due to electrical voltage.
• Shut down and isolate the electrical power source before carrying out work on the system.
• Put the power switch to OFF before carrying out work on the
device.
• Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high fluid temperatures.
• Use safety gloves to handle the device.
• Stop the circulation of fluid and drain the pipe before loosening
the process connections.
Risk of injury due to the nature of the fluid.
• Respect the prevailing regulations on accident prevention and
safety relating to the use of aggressive fluids.
English
23
Type 8025/8035 battery powered
Maintenance and troubleshooting
Warning
Risk of injury due to non-conforming maintenance.
• Maintenance must only be carried out by qualified and skilled
staff with the appropriate tools.
• Ensure that the restart of the installation is controlled after any
interventions.
9.2. Cleaning the device
The device can be cleaned with a cloth dampened with water or a
detergent compatible with the materials the device is made of.
Please feel free to contact your Bürkert supplier for any additional
information.
9.3. Cleaning the flow sensor of the
8025
note
• Use a cleaning product that is compatible with the materials the
flow sensor is made of.
• Do not use any abrasive acting materials.
note
After cleaning of the flow sensor:
• Rinse the flow sensor.
• Check the seal and replace it if necessary.
9.4. Replacing the batteries
Dispose of the used batteries in compliance with the prevailing legislation.
→ Open the housing as indicated in Fig. 10 chap. 8.2.→ Disconnect the flow sensor to make easier the access to the
batteries.
→ Take off the batteries connectors.→ Replace the batteries one after the other to keep the data regis-
tered in the device.
→ Plug the connectors onto the batteries.→ Reconnect the flow sensor.→ Close the cover.→ Tighten the screws.→ Close the lid.→ Tighten the screw.
24
English
Type 8025/8035 battery powered
Spare parts and accessories
9.5. If you encounter problems
DisplayPossible causeRecommended action
"ERROR"The user parameters
are lost.
→ Parameterise the
device again.
→ If the message appears
repeatedly send the
device back to your
supplier.
"LOW BATT" The battery power
is too low to ensure
the operating of the
device.
Value of the
flow rate or of
the totalizer
fashes
The battery power
is getting low but
the device is still
operating.
→ Replace the batteries
(see chap. 9.4).
→ Plan to replace the
batteries
10. SPARE PARTS AND
ACCESSORIES
attention
Risk of injury and/or damage caused by the use of unsuitable
parts.
Incorrect accessories and unsuitable replacement parts may
cause injuries and damage the device and the surrounding area.
• Use only original accessories and original replacement parts
from Bürkert.
10.1. SE35 electronic module
Spare partOrder code
Cover in PC, with lid, incl. window, screws,
sticked foil and electronic board
Set with 8 FLOW foils553191
Housing incl. coil as a measuring element425247
Set including:
• 2 M20x1.5 screw plugs
• 2 flat seals
425433
444705
English
25
Loading...
+ 61 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.