User’s manual
V300+ / V600+ / V1000+
Version N°2.0 – E
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 1 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 2 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
Contents
|
Description |
Page |
1. |
Description of the instrument |
4-5 |
2. |
Display functions |
5 |
3. |
Distintive characteristics |
6 |
4. |
Specifications |
6 |
5. |
Delivery |
6 |
6. |
Unpacking and Installation |
7 |
7. |
Getting started |
8 |
8. |
Replacing the battery pack |
9 |
9. |
Ajusting the measuring pressure |
10 |
10. |
Ajustement of the floating probe suspension |
10 |
11. |
Main functions |
11 |
12. |
Secondary functions |
12 |
13. |
How to start up |
13 |
|
13.1 Setting the probe constant |
13 |
|
13.2 Measuring heights |
13 |
|
13.3 Measuring diameters and centerlines |
14 |
|
13.4 Measuring with min. and max. mode |
15 |
|
13.5 Measuring with delta mode |
16 |
|
13.6 Measuring of squareness |
16 |
|
13.7 Measuring with two references |
17 |
14. |
Initialization of printing mode |
18 |
15. |
Printer |
19 |
16. |
AC adaptor, external contact |
20 |
17. |
RS232C connector |
20 |
18. |
Remote control of the instrument |
21 |
19. |
Optional accessories |
22-26 |
20. |
Maintenance |
27 |
21. |
Transport |
27 |
22. |
Complaints / Repairs |
27 |
23. |
Distributors worldwide |
28-30 |
TRIMOS SA |
Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 3 |
Tel ++ 41 21 |
633 01 01 Fax ++41 21 633 01 02 office@trimos.ch |
http://www.trimos.ch |
35 34 33 32 44 31
36
37
38 39 45 40 41 42 43
6
10
9
8
7
30
6
29
12
28
11
27
3
2 26
1 25
24
5 4
14 15
13 15
17
18
19
20
21
22
23
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 4 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
1.Description of the instrument
1.Handwheel allowing the displacement of the measuring probe.
2.Locking device to activate the fine ajustment.
3.Fine adjustment screw.
4.Switch to acivate the air cushion. (Version C)
5.Handle allowing the displacement of the instrument.
6.Measuring probe.
7.Screw to adjust the measuring pressure (0.7N-1.6N, ajustable) (wrench /2mm)
8.Lever for fast displacement.
9.Screw for locking the measuring carriage (wrench 2mm) (Screw, chromiumplated)
10.Control screw to ajust the floating probe suspension. (wrench 2mm)
11.Bracket for printer.
12.Printer RS 232.
13.RS 232 connector.
14.Green light = Indicates the trickle charge with AC adapter connected.
15.Red light = Indicates fast charging.
16.Connector for AC adaptor or data transmission foot pedal.
17.Green light = normal measuring mode.
18.Orange light = measuring diameters.
19.Yellow light = measuring centerlines.
20.NORMAL or DIAMETER/CENTERLINE mode key.
21.PRESET key = set the display at zero or memorizes preset values.
22.PRINT key = RS232 data output.
23.ON-OFF key = ON / OFF Switch.
24.Numerical keyboard.
25.CLEAR key = reinitialises the minimum/maximum/delta memories.
26.MIN/MAX/DELTA mode key.
27.MM/INCH key, true conversion Metric – Inch – Metric.
28.REF I / REF II selection key.
29.Resolution key. (0.010/0.001mm or .0005/.00005”)
30.Probe constant key.
2.Display functions
31.Indication of sending data.
32.Indication of DELTA mode.
33.Indication of MAX mode.
34.Indication of MIN mode.
35.Indication of REFII mode.
36.Indication of low battery level – connect the AC adaptor.
37.Indication of measuring values.
38.Indication of MM mode.
39.Indication of INCH mode.
40.Indication of measuring direction.
41.Indication of centerline mode.
42.Indication of diameter mode.
43.Indication of PRESET mode.
44.Indication of CONSTANT mode.
45.Indication of keyboard locked.
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 5 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
3. Distintive characteristics
The V300,V600 and V1000 are height measuring instruments, self-contained, controlled by a microprocessor.
o Easy to operate.
o Self-contained operation for a period of: without air cushion : max.100h.
|
: with air cushion : max. 50h. |
o |
Rechargeable battery. |
o |
Big display. |
oAcceptance of measured values by an acoustic signal.
4.Specifications
Designation |
V300 |
|
V300C |
|
V600 |
V600C |
|
V1000 |
|
V1000C |
Measuring range |
300 mm or 12’’ |
|
600 mm or 24’’ |
|
1000 mm or 40’’ |
|||||
Application range |
535 mm or 21’’ |
|
835 mm or 32.8’’ |
|
1235 mm or 48.6’’ |
|||||
Resolution |
|
|
0.01 mm / 0.001 mm or .0005’’ / .00005’’ |
|
|
|||||
Overal measuring accuracy |
3 µm + (L(mm)/300 = µm or .00012’’ + (L(inch)/300000) = inch |
|||||||||
Repeatability |
|
|
|
± 2 s = ³ 2 µm or .00008’’ |
|
|
|
|||
Max. displacement speed |
|
|
|
|
|
|
|
|
|
|
of the measuring carriage |
|
|
|
|
3m/120’’ per second |
|
|
|
||
Measuring pressure |
|
|
|
0.7 N – 1,6 N (adjustable) |
|
|
|
|||
Measuring system |
|
|
differencial capacitance Sylvac (patented) |
|
|
|||||
Power supply |
|
|
Battery powered (rechargeable) |
|
|
|||||
Operational time |
100 h. |
|
50 h. |
|
100 h. |
50 h. |
|
100 h. |
|
50 h. |
Squareness deviation overal |
|
|
|
|
|
|
|
|
|
|
(in measuring direction) |
5µm or .0002’’ |
|
8 µm or .0003’’ |
|
12 µm or .00047’’ |
|||||
Data output |
|
|
|
|
RS 232 C |
|
|
|
||
Operational temperature |
|
|
|
|
|
|
|
|
|
|
limit |
|
|
+ 10 °C to +40°C or 50° F to 104° F |
|
|
|||||
Total height |
565 mm or 22.2’’ |
|
865 mm or 34’’ |
|
1275 mm or 50’’ |
|||||
Weight |
|
10 kg |
|
12,3 kg |
|
|
15,3 kg |
5. Delivery
The TRIMOS V300, V600 and V1000 are supplied as follows : o Ruby ball probe dia. 4 mm (V-1)
oHex screwdriver wrench 2 mm (T016-INB.2), Hex screwdriver wrench 2,5 mm (T016INB.3)
|
o AC adaptor set 8.5V / 700mA sector, depending on country. |
|
|||
|
o |
Dust cover. |
|
|
|
|
o Test and guarantee certificates, user’s manual. |
|
|||
|
|
|
|
|
|
|
Vertical automatic |
Code number |
|
Code number |
|
|
|
|
without air cushion |
|
with air cushion |
|
Measuring range 300mm /12’’, 220 V |
V300/220 |
|
V300C/220 |
|
|
Measuring range 300mm /12’’, 240 V |
V300/240 |
|
V300C/240 |
|
|
Measuring range 300mm /12’’, 110 V |
V300/110 |
|
V300C/110 |
|
|
Measuring range 300mm /12’’, 100 V |
V300/100 |
|
V300C/100 |
|
|
Measuring range 600mm /24’’, 220 V |
V600/220 |
|
V600C/220 |
|
|
Measuring range 600mm /24’’, 240 V |
V600/240 |
|
V600C/240 |
|
|
Measuring range 600mm /24’’, 110 V |
V600/110 |
|
V600C/110 |
|
|
Measuring range 600mm /24’’, 100 V |
V600/100 |
|
V600C/100 |
|
|
Measuring range 1000mm /40’’, 220 V |
V1000/220 |
|
V1000C/220 |
|
|
Measuring range 1000mm /40’’, 240V |
V1000/240 |
|
V1000C/240 |
|
|
Measuring range 1000mm /40’’, 110 V |
V1000/110 |
|
V1000C/110 |
|
|
Measuring range 1000mm /40’’, 100 V |
V1000/100 |
|
V1000C/100 |
|
TRIMOS SA |
Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 6 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
Options : see optional accessories (page 22-26)
TRIMOS SA |
Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 7 |
Tel ++ 41 21 |
633 01 01 Fax ++41 21 633 01 02 office@trimos.ch |
http://www.trimos.ch |
6. Unpacking and installation
The V300, V600 and V1000 are supplied in a shock prof box.
The instrument itseff is surrounded with a protection cover.
- To unpack the instrument lay the box in a horizontal position,
Brand name visible on top, open it and take the instrument out carefully.
-Take off the protection cover.
-Insert the ruby ball (V-1).
-Unlock completely the transport safety screw (9), Screw, chromium-plated.
-Clean the instrument, especially the three point air bearings on the base.
DO NOT USE CHEMICAL PRODUCTS.
-Place the instrument on a granite surface table.
FOR FUTURE TRANSPORT KEEP THE ORIGINAL PACKING.
V.RE300/600/1000
V.REC300 (V300)
V.REC600 (V600) V.REC1000 (V1000)
V.CA300 (V300) TVA.CA600 (V600) V.CA1000 (V1000)
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 8 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
7. Getting started
The instrument stands on a granite surface table (or e.i. cast Iron table)
-Switch ON the instrument by pressing the key (ON).
-The display will show
rEF I
M M
-The reference of the column must be taken :
-Move the probe (A) slowly and cross the two arrows to take the reference. This operation will memorized this reference point improve
the accuracy of the instrument.
-The display will begin to count.
If during this operation the display doesn’t count.
restart the operation again.
-If the indicator (36) showing complete battery discharge, connect the AC adaptor. The instrument remains opérational.
Complete recharging takes 15 hours.
A
0.125
M M
B 0.125
M M
-It will not do any harm by connecting the AC adaptor continiously, because it will charge the instrument with a maintenance power.
IT IS NOT NECESSARY TO WAIT FOR THE INSTRUMENT TO BE FULLY CHARGED BEFORE USE :
INSTRUMENT CAN BE USED DURING CHARGE CYCLE.
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 9 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch
8. REPLACING THE BATTERY PACK
As sson as the autonomy of the instrument becomes unsufficient, the battery pack should be changed.(Lifespan about 4 to 6 years)
1.Purchase a battery pack from your TRIMOS distributor.
2.Switch off the instrument.
3.Place the instrument in an horizontal position onto a table. (steady)
4.Unscrew the base plate. (5 screws)
5.Disconnect the battery pack (A)
6.Connect the new battery pack and refit the base plate. Take care not to damage Any connecting cables.
7.Charge the new battery pack ~15 heures.
Battery pack
Code number 151-331.009
TRIMOS SA Av. de Longemalle 5 CH 1020 Renens Switzerland |
Page 10 |
Tel ++ 41 21 633 01 01 Fax ++41 21 633 01 02 office@trimos.ch http://www.trimos.ch