The following safety precautions apply to both operating and maintenance personnel and must be followed
during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
•Read and understand the safety and operational information in this manual.
•Apply all the listed safety precautions.
•Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating
the instrument at an incorrect line voltage will void the warranty.
•Make all connections to the instrument before applying power.
•Do not operate the instrument in ways not specied by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety
standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no
liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard denes safety category ratings that specify the amount of electrical energy
available and the voltage impulses that may occur on electrical conductors associated with these category
ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a
maximum voltage of the circuit to be tested, which denes the voltage impulses expected and required
insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be
connected to the mains supply. The voltages in the environment are typically
derived from a limited-energy transformer or a battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected
to the mains supply at a standard wall outlet or similar sources. Example
measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be
connected to the mains installation of a building. Examples are measurements
inside a building’s circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be
connected to the primary power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specied
in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or
higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do
so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains
power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and
ensure it is appropriate for your country of use.
Ground the Instrument
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety
ground. This instrument is grounded through the ground conductor of the supplied, three-conductor AC
line power cable. The power cable must be plugged into an approved three-conductor electrical outlet.
The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible
conductive parts (including control knobs) may provide an electric shock. Failure to use a properlygrounded approved outlet and the recommended three-conductor AC line power cable may result in
injury or death.
Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference
of potential only and is not to be used as a safety ground. Do not operate in an explosive or ammable
atmosphere.
Do not operate the instrument in the presence of ammable gases or vapors, fumes, or nely-divided
particulates.
The instrument is designed to be used in oce-type indoor environments. Do not operate the instrument
•In the presence of noxious, corrosive, or ammable fumes, gases, vapors, chemicals, or nely-divided
particulates.
•In relative humidity conditions outside the instrument’s specications.
•In environments where there is a danger of any liquid being spilled on the instrument or where any
liquid can condense on the instrument.
•In air temperatures exceeding the specied operating temperatures.
•In atmospheric pressures outside the specied altitude limits or where the surrounding gas is not air.
•In environments with restricted cooling air ow, even if the air temperatures are within specications.
•In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating
temperature range is 0∘C to 40∘C and 20% to 80% relative humidity, with no condensation allowed.
Measurements made by this instrument may be outside specications if the instrument is used in nonoce-type environments. Such environments may include rapid temperature or humidity changes, sunlight,
vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric elds, or strong
magnetic elds.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on
or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label
it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of
the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants,
solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth
or as instructed in this manual. Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a
life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal
adjustments must be made by qualied service-trained maintenance personnel who are aware of the
hazards involved when the instrument’s covers and shields are removed. Under certain conditions,
even with the power cord removed, dangerous voltages may exist when the covers are removed. To
avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections
(for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are
no hazardous voltages present on any conductors by measurements with a properly-operating voltagesensing device before touching any internal parts. Verify the voltage-sensing device is working properly
before and after making the measurements by testing with known-operating voltage sources and test for
both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of
rendering rst aid and resuscitation is present.
Do not insert any object into an instrument’s ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition
in the circuit exists.
Fuse replacement must be done by qualied service-trained maintenance personnel who are aware of
the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from
the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage
ratings, and current ratings specied in this manual or on the back of the instrument. Failure to do so
may damage the instrument, lead to a safety hazard, or cause a re. Failure to use the specied fuses
will void the warranty.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the
instrument to B&K Precision for service and repair to ensure that safety and performance features are
maintained.
For continued safe use of the instrument
•Do not place heavy objects on the instrument.
•Do not obstruct cooling air ow to the instrument.
•Do not place a hot soldering iron on the instrument.
•Do not pull the instrument with the power cord, connected probe, or connected test lead.
•Do not move the instrument when a probe is connected to a circuit being tested.
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other
European countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on waste electrical
and electronic equipment (WEEE), and in jurisdictions adopting that
Directive, is marked as being put on the market after August 13, 2005,
and should not be disposed of as unsorted municipal waste. Please
utilize your local WEEE collection facilities in the disposition of this
product and otherwise observe all applicable requirements.
The 9140 Series triple output multi-range DC power supplies combine industry-leading power density
and performance with an extensive set of features in a compact 2U form factor.
Three isolated output channels each produce 100 W of clean power with exceptionally low ripple and
noise characteristics. Combine channels to output up to 300 W, distributing the power across all three
channels. Multiple outputs paired with advanced list mode programming, data logging, and protection
features make these power supplies suitable for a wide range of benchtop or test system applications.
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the
shipping carton, and check for any obvious signs of physical damage that may have occurred during
transportation. Report any damage to the shipping agent immediately. Save the original packing carton
for possible future reshipment. Every power supply is shipped with the following contents:
•1 x 9140 or 9141 Power Supply
•1 x AC Power Cord
•1 x Certicate of Calibration
•1 x Test Report
Note:
Ensure the presence of all the items above. Contact
the distributor or B&K Precision if anything is missing.
1.3 Features
•Three independent galvanically isolated, oating output channels providing up to 100 W per channel
or 300 W total when combining all three channels
•High power density, compact 2U half-rack form factor
•Multi-ranging operation delivers rated power at various voltage/current combinations
•Combine outputs in parallel or series to increase current or voltage (up to 24 A or 180 V depending
on the model)
•Advanced list mode programming with internal storage for up to 10 list mode programs and list
sequencing capabilities
•Digital I/O block oers external triggering, voltage fault and remote inhibit capabilities
•Overvoltage (OVP), overcurrent (OCP), overtemperature (OTP), and includes a key-lock function
•NISPOM-compliant sanitization procedure to securely erase and return the instrument to factory
settings
The 9140 Series is compatible with the optional 19-inch rackmount kit model RK2US. The RK2US
installation instructions can be downloaded from bkprecision.com
Before connecting and powering up the instrument, please review and go through the instructions in this
chapter.
2.1 Input Power and Fuse Requirements
The supply has a universal AC input that accepts line voltage input within:
91409141
AC Line Input100 - 240 VAC±10%, 47 to 63 Hz
AC Line PhaseSingle Phase
Maximum Rated
Input Power
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF
position and verify that the AC power cord, including the extension line, is compatible with the rated
voltage/current and that there is sucient circuit capacity for the power supply. Once veried, connect
the cable rmly.
The included AC power cord is safety certied for this instrument operating in rated range. To
change a cable or add an extension cable, be sure that it can meet the required power
ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
500 VA
SHOCK HAZARD
The power cord provides a chassis ground through a third conductor. Verify that your
power outlet is of the three-conductor type with the correct pin connected to earth ground.
Any disassembling of the case or changing the fuse not performed by an
authorized service technician will void the warranty of the instrument
SHOCK HAZARD:
Hazardous voltages may exist at the outputs and the load connections when using a power supply
with a rated output greater than 60 V. To protect personnel against accidental contact with hazardous
voltages, ensure that the load and its connections have no accessible live parts. Ensure that the load
wiring insulation rating is greater than to the maximum output voltage of the power supply.
The rear output terminals can accept wire sizes AWG 24 to AWG 12 (See Table 2.1).However, we
recommend using 12 AWG if current output is between 5 to 10 A. Refer to Table 2 below to determine
the proper wire size.
The output terminals of the 9240 Series are isolated from earth ground. An earth ground terminal is
provided on the front panel for convenience. Any output terminal may be grounded. However, output
terminals must not exceed ±200 VDC.
Complete the following steps to verify the Power supply is ready for use.
Getting Started 21
Verify AC
Input Voltage
Connect Power
& Self-Test
After Power on, during the self-test, the following should be displayed:
Verify and check to make sure proper AC voltages are available to power the
instrument. The AC voltage range must meet the acceptable specication as
explained in section “2.1 Input Power and Fuse Requirements”.
Connect AC power cord to the AC receptacle in the rear panel and press the
power button. It will run through a self-test procedure initially before booting to
the main screen.
Figure 2.2 Initializing
After the system scan is complete the LCD will display as shown in Figure 2.3
Note:
The initial power on state is based on the settings set on Power-ON
At power-on, the power supply will automatically enter the front-panel operation mode and the instrument
can be controlled via the front panel keys and knob.
3.1 Keys
3.1.1 Main Keys
There are three main keys:
Navigates the 5 available screens.
When navigating any menu
returns the key will return "Home"
main display
5 available displays:
– Main page:
Overview of the 3 outputs.
– Single Channel Display 1 to
3: Detailed channel overview.
– Live Output Monitoring:
Oscilloscope like overview
of the measured output
parameters.
More information on each display in 3.2.
Opens up a menu giving access
to various save and load options,
regardless of the current display
of the screen.
Opens the main menu bar. If
a menu bar is already open,
The unit has six soft keys, which are located beneath the screen. Each key selects the corresponding
function. Functions will vary depending on the current menu or display.
Figure 3.1 Soft Keys
3.1.3 Numeric Keys
The numeric keys allow the conguration of various parameters. Using the numeric keys provides a fast
and precise input. Thekey can be found with the numeric keys. Pressing enter will assign the
selected value to the desired parameter.
Figure 3.2 Numeric Keys
3.1.4 Output Control
Thebutton toggles the output On/O. When output is enabled, the ON button will turn green
and the supply will display the state of the channel.
represents constant current mode.represents constant voltage mode. This is indicated for
each channel independently. When output is disabled, the channel state will display.
Thekey toggles all the channels On/O simultaneously. Its function may vary depending on the
coupling state of the channels. For more information go to Coupling.
The live output monitoring display shows metered output voltage and current over time.
It is always active and will continuously plot and overwrite the oldest value when the display graph is full.
Figure 3.5Live Output Monitoring Display
The Live output display can be congured by setting the: Voltage, Current, and Time Scale.
Voltage Scale
The voltage scale of each output can be set from 10 mV/Div up to 20 V/Div.
1. While in Live Output Display use the softkeys to select Voltage Scale.
2. Select a channel to congure.
3. Use the rotary knob to navigate through the available scales.
Current Scale
The current scale of each output can be set from 10 mA/Div up to 3 A/Div.
1. While in Live Output Display use the softkeys to select Current Scale.
2. Select a channel to congure.
3. Use the rotary knob to navigate through the available scales.
Follow the steps below to set the output voltage or current. The values can be entered using the numeric
keypad or the rotary knob.
1. Using the keypad: Use the soft keys to select the channel VSET or ISET settings beneath channel
view on the display. Then use the numeric keypad to input the voltage or current value then press
to conrm.
2. Using the rotary knob: Use the soft keys to select the channel VSET or ISET settings beneath
the channel view on display. Then turn the knob clockwise to increment or counter-clockwise to
decrement the value.
a. Use thebelow the rotary knob to move the cursor. Press the knob in to conrm.
Over voltage protection is always enabled, however the user can set the OVP voltage limits.
1. Press thebutton then use the soft keys to select Output cong > Protection settings
2. Use the soft keys to select a channel to adjust.
3. Select OVP Limit.
4. Use the numeric keypad or rotary knob to set the desired OVP limit.
When OVP protection is tripped during operation, the output will be de-rated to 0 V and the icon
will appear between the wattage and Vset.
4.1.2 Over Current Protection (OCP)
Over current protection state and limit can be set in the protection settings.
1. Press thebutton then use the soft keys to select Output cong > Protection settings
2. Use the soft keys to select a channel to adjust.
3. Press OCP State to toggle the over current protection on or o.
4. Select OCP Limit and use the numeric keypad or rotary knob to set limit value.
When OCP protection is tripped during operation, the output will be de-rated to 0 V and the icon
will appear between the wattage and Vset.
4.1.3 Over Temperature Protection & Over Temperature Warning (OTP & OTW)
The state and limit of both OTP & OTW cannot be set or congured in the protection settings. OTP is set
to 85 °C and OTW is set to 75 °C. If the temperature reaches either OTW, the OT icon will blink on the
display with a low period beep sound. If temperature reaches OTP the OT icon will show on the display
and the output will turn o. The output will not be allowed to turn on until the temperature drops to 75 °C.
Congure the maximum or minimum voltage value the user can set from the protection
settings.
1. Press thebutton then use the soft keys to select Output cong >
Protection settings
2. Use the soft keys to select a channel to adjust.
3. Select Vmax or Vmin and set a maximum or minimum value using the numeric keypad or rotary knob.
4.1.5 Protection Clear
Protections are cleared from the output conguration settings as shown in the following steps.
1. Press thebutton then use the soft key to select Output cong >
Protection Clear.
2. Press the soft key to clear the protection for the corresponding channel, or select
All CH Clear to clear protections on all three channels.
4.2 Output Settings
4.2.1 Remote Sense
Remote sense can be used to compensate for voltage drops (up to 1 V) due to resistance from test leads
connected to your device under test (DUT), thus providing more accurate output voltage. When remote
sense is enabled, the positive (+S) and positive lead (+) of the DC output are connected to the positive
(+) of the load. Similarly, the negative sense (-S) and negative lead (-) of the DC output are connected
to the negative (-) end of the load shown in gure 4.1
Never connect any power source into any of the terminals at any time during operation.
When output is enabled, DO NOT use your hands to touch the terminals or the screws that are
designed to tighten wires to the terminals. Doing so may create a shock hazard under high
voltage output conditions.
DO NOT at any time disconnect the wires from the S+ and S- terminals to the DUT while output
is enabled (ON). Doing so may damage the power supply and cause unstable output.
Remote sense is disabled by default, see the following steps to enable remote sense:
1. Press thebutton then use the soft key to select Output cong > Output Settings.
2. Select a desired channel to congure.
3. Press Remote Sense using the soft key to enable remote sense.
4.2.2 Output Timer
Output timer can be set for each channel individually. When the output timer is enabled, an hourglass
icon will be displayed next to the channel number. The timer countdown will begin when the channel
output is enabled. The output is disabled when the countdown reaches 0. The timer can be set from 1
second to 99 hours 59 minutes and 59 seconds.
Follow the steps below to enable and set the output timer:
1. Press thebutton then use the soft key to select Output cong > Output Settings.
2. Select a desired channel to congure. Push Timer State to enable the timer.
3. Select Timer Setting and set a desired time for the output to be enabled using the numeric keypad or
rotary knob. The timer format is: HH:MM:SS. Use the keys to move the cursor.
Series mode changes the UI to accurately represent the possible outputs congurations.
Series mode does not internally wire the outputs. The outputs must be connected externally to increase
the available voltage range.
Press thebutton then use the soft key to select Output Cong > Op. Mode/Coupling > Operation
Mode > Series.
The image below illustrates the external wiring required to connect all channels in series. If CH1+ CH2
is chosen CH2’s positive terminal does not connect to CH3’s negative terminal.
Single CH displays more information about the output, and allows conguration of more parameters.
The voltage congurations will now range from 0 v to 64 V for model 9140 and 0 V to 120 V for model
9141.
Figure 4.7 Series CH1+2 Single Channel Display
Protection Setting menu will also adjust the congurable values.
Operation mode will be set to Series All. Coupling will not be available in Series All mode.
Press thebutton then use the soft key to select Output Cong > Op. Mode/Coupling > Operation
Mode > Series > Series All.
Figure 4.9 Series All
In Series All modewill only toggle between the Series All display and live output monitoring.
Series All CH display will adjust to voltage conguration between 0 V and 96 V for model 9140 and 0 V
to 180 V for model 9141.
Figure 4.10 Series All Main Page
Protection Setting menu will also adjust the congurable values.
Parallel Mode does not internally wire the outputs. The outputs must be connected externally to increase
the available current range.
Parallel Mode changes the UI to accurately represent the possible output congurations.
Press thebutton then use the soft key to select Output Cong > Op. Mode/Coupling > Operation
Mode > Parallel.
The image below illustrates the external wiring required to connect all channels in Parallel. If CH1+ CH2
is chosen CH2’s terminals do not connect to CH3’s terminals.
Voltage levels are tracked on either CH1+2 or All Channels.
Press thebutton then use the soft key to select Output Cong > Op. Mode/Coupling >
Operation Mode > Tracking.
Tracking mode allows real time voltage tracking across all channels or between CH1 and CH2.
When tracking mode is enabled, list mode will still operate independently on the channel that has list
enabled. Tracking channels will not track the list changes.
Enable or disable the output coupling between multiple output channels, by using the soft keys to toggle
individual channel coupling On or O.
Press thebutton then use the soft key to select Output Cong > Op. Mode/Coupling
If coupling is enabled,will follow the coupling settings. Channel output On/O buttons will also
follow the coupling settings if the correlated channel is coupled to another channel.
Example:
If CH1 and CH2 are coupled and the coupled CHs are on,will not turn on
CH3. It will turn o CH1 and CH2.
However, if all channels are OFFwill turn all CHs ON.
The 9140 and 9141 are capable of storing up to 10 programmable lists in the internal memory. Each list
can have up to 100 congurable steps.
List memory is shareable across all 3 channels, allowing the channels to run the same list or dierent list
simultaneously.
If all ten list numbers are taken, a list can also be saved to a USB. Lists can be loaded from a USB,
allowing for a larger collection of programmed lists.
5.1 List Setup
Press thebutton then use the soft key to select List Setup.
Use the softkeys to select one of the channels CH1, CH2, or CH3.
Select a list program from the memory to run on the channel.
Use the numeric keypad or rotary knob to select list number.
Pressto assign the list to a channel.
Pace:
Set the pace in which steps are executed.
a. Dwell: The next step outputs once the dwell time elapsed.
List Mode48
b. Trigger: After the dwell time elapsed, wait for a trigger before outputting the next step.
Trigger Source
Set the trigger source.
a. Manual: Select to use a manual trigger. The Trigger softkeywill be availableonce List State is on and
exit out of the menu to the main display.Press it to manually trigger the list program. If the step dwell
time has not elapsed, then Trigger will not have a white background. When dwell timeelapses, Trigger background
color will turn white to indicate the trigger is ready.
Use the rotary knob to navigate through the le paths of the USB. Use thebutton to expand folders.
Once the desired location has been found select Save by pressing the softkey furthest to the left.
Load a previously saved list from a USB to selected List Number.
Use the rotary knob to navigate through the le paths of the USB. Use thebutton to expand folders.
Once the desired location has been found select Load by pressing the softkey furthest to the left.
Figure 5.4 Load from USB
Note:
The list must be saved/overwritten at the selected list number location before it is able to be assigned.
5.2.1 List Number
1. Select a list program from the internal memory.
2. Use the numeric keypad or rotary knob to select a list (0 to 9).
Set the next list program to run after the current list elapse.
a. To make a list run continuously and indenitely, set Next to the same number as List Number.
b. To run dierent list sequences continuously, set Next to each other.
Example Set Next on List 1 to 2.
Set Next on List 2 to 1.
5.2.3 Repeat
To repeat a list, set Repeat using the numeric keypad or rotary knob and press.
5.2.4 Steps
Add, delete, clear all, or edit steps.
Add Step
Use the softkeys to add a step to the list. The step added will be a duplication of the step selected before
pressing Add Step.
To select a step to duplicate, use the rotary knob to navigate the available steps. If the list is new, only
one default step will be available. Pressing Add Step will duplicate the default step.
When setting Vset the list will not limit the voltage range based on mode.
Therefore, Vmax will be 96.960 for model 9140 and 180.800 for model 9141.
To prevent a protection trip, verify the Vset and Iset values in the list is within
the range of the congured operation mode before running the list.
Current Sets Iset value.
When setting Iset the list will not limit the current range based on mode.
Therefore, Imax will be 24.240 for model 9140 and 12.120 for the 9141.
To prevent a protection trip, verify the Vset and Iset values in the list is within
the range of the congured operation mode before running the list.
List Mode54
Step Parameters
BOST Press the knob or to enable/disable a trigger-out signal at the beginning of the step (BOST).
EOST Press the knob or to enable/disable a trigger-out signal at the beginning of the step (EOST).
List allows a sequence of outputs with up to 100 congurable steps. With sharable memory across all
three channels, a list can run simultaneously on various channels. This allows for a list to be compatible
in normal, parallel or series mode.
Before running a list its parameters must be congured.
1. Congure the parameters of each step. See 5.2
2. Set the settings of the channel which will output the list. See 5.1
a. Enable List State on desired channel.
b. Assign a List Number to the channel.
c. Select Pace {Dwell or Trigger}.
d. Select Trigger Source {Manual, Digital IO, or Remote }.
e. Select After List output {DC or Last}.
3. After conguring the list and channel parameters pressto return to the main page.
– Vset will be replaced with List Start if List State was enabled.
Figure 5.7 List State Disabled
4. Select List start to start the list.
– Once the list begins List Start will change to Abort List
– The interaction with the list will vary depending on the chosen Pace. See 5.1
5. Select List Abort to end the list before all steps elapse. If the all steps elapse the list will end and the
output will be set to chosen After List parameter.
– If After List is set to DC the output will output the Vset and Iset values.
– If After List is set to Last the output will output the values of the last step in the list.
Note:
While the list is running the user can adjust the current step output by setting Vset or Iset on the main
display. Doing so will immediately change the value of the ongoing output. However, the set value will
not be saved to the list. Once the dwell time of the step has elapsed the list will continue and output
the next step.
If Trigger source is set to manual, Vset/Iset will not be available. The rotary knob and numeric keypad
can be used to adjust the selected value. To adjust both Vset/Iset navigate to the single CH display of
the output using thekey.Vset/Iset will be available as shown in gure 5.9.
The data logger can record the output voltage, current, and error codes of all three channels. Log data
can be congured to record either Voltage only, Current only, or both.
Connect a USB ash drive to the front panel USB port. Max. Recording time will vary based on the ash
drive size and the amount of the data being logged.
Figure 6.1 Data Logger Max Time
6.1 Using the Data Logger Function
Press thebutton then use the soft key to select Data logger. The Data Logger settings can be
selected in this menu.
Before starting the Data Logger its settings must be adjusted.
1. Set the desired Sampling Interval See 6.2.1
2. Select a File Path to determine where to store the collected data. see 6.2.2
3. Enable or disable T. Stamp Filename. See 6.2.3
4. Choose what data to record. See 6.2.4
5. Enable or disable Status Code. See 6.2.5
6. Select a Trigger Source. See 6.2.6
7. After setting all desired parameters press datalog Start/Stop to begin recording. Staring data log
will vary based on chosen trigger. See6.2.7
Press thebutton then use the soft key to select Data logger > Sampling Interval.
Use the numeric keypad or rotary motor to select the sampling interval. Pressto assign the selected
value. (.2s to 300s)
6.2.2 File Path
Press thebutton then use the soft key to select Data logger > File Path.
Using the rotary knob andnavigate through the le path of the USB. Once the desired location
has been found select save by pressing the softkey furthest to the left.
Verify that the correct path was chosen in the Data Logger menu.
Figure 6.2 File Path Selected
6.2.3 T. Stamp Filename
Press thebutton then use the soft key to select Data logger > T. Stamp Filename
Toggle to enable/disable time and date on the le name.
Data order will vary depending on Log Data selected.
6.2.5 Status Code
Press thebutton then use the soft key to select Data logger > Status Code.
Toggle to enable/disable status code.
Enabled Status Code will record all codes reported. See 6.3 for an example.
In gure 6.3 the code 0x0001 is returned. This indicates OVP has been triggered.
Data Logger 61
Table 6.2 denes each code.
CodeDescription
0x0001Over Voltage Protection Triggered
0x0002Over Current Protection Triggered
0x0004
0x0008Sense not connected
Table 6.2 Status Code
Over Temperature
Protection Triggered
6.2.6 Trigger Source
Press thebutton then use the soft key to select Data logger > Trigger Source.
Datalog can be triggered in three ways: Manual, Digital IO, and Remote.
•Manual: Select to use manual trigger. The trigger softkey will be available as Datalog Start/Stop.
See section 6.2.7
•Digital IO: Select to use the digital I/O pins to receive a trigger. To assign a function to a pin see
section7.4.1.
Press thebutton then use the soft key to select Utilities > User Settings >Date.
Use the number pad to set the date.
YY/MM/DD
7.1.4 Time
Press thebutton then use the soft key to select Utilities > User Settings >Time.
Use the number pad to set the time.
HH:MM:SS
7.1.5 Screen Intensity
Press thebutton then use the soft key to select Utilities > User Settings >Next Pg. > Screen
Intensity.
Use the rotary knob or number pad to set the screen brightness.
Lowest 1
Highest 10
7.1.6 Language
Press thebutton then use the soft key to select Utilities > User Settings >Next Pg. > Language.
Set the language for the On Screen Help guide. It does not change the language for the menus. For
more information about On Screen Help see section 7.8
The gateway address is by default the IP address of the network device that connects the instrument.
If IP Mode is set to DHCP Gateway IP does not have to be set.
To set Gateway IP :
Press thebutton then use the soft key to select Utilities > I/O Cong > LAN Settings > GatewayIP.
Use the numeric keypad to enter an IP address. Thenkeys are used to separate each number in
the string. After inputting each number pressto set the address.
Primary DNS
Domain Name Service (DNS) is the system that automatically translates Internet addresses to the numeric
machine addresses. The DNS address is the IP address of the system that performs this service.
If IP Mode is set to DHCP DNS will automatically be set.
To set the Primary DNS :
Press thebutton then use the soft key to select Utilities > I/O Cong > LAN Settings > PrimaryDNS .
Use the numeric keypad to enter an IP address. Thenkeys are used to separate each number in
the string. After inputting each number pressto set the address.
Hostname
Each instrument comes with a unique hostname that can be changed. The default hostname has the
following format:
Enable a fault condition, which generates a protection fault signal on the digital port.
Dierent conditions such as over voltage, over current, or over temperature will generate a fault event.
The fault event will “disable” all outputs without turning them o.
The outputs are “disabled” by setting the outputs as low as possible.
7.4.1 Polarity
Default: Positive
Set polarity of selected pin.
Positive
A logic true signal is received as a voltage high.
Negative
A logic true signal is received as a voltage low.
7.4.1 Inhibit Mode
Receive an external input signal that controls the output state of all channels.
Default: O
To set Inhibit Mode:
1. Assign the inhibit In function to pin 3 by:
Press thebutton then use the soft key to select Utilities > I/O Cong > Digital I/O > Pin 3 >Inhibit In.
2. Select Inhibit Mode: Press thebutton then use the soft key to select Utilities > I/O Cong >Digital I/O > Inhibit Mode.
A logic-true transition signal will disable the power supply. The output will remain disabled.
Live
Power supply output follows the state of the inhibit signal. If the inhibit signal is true the output is disabled.
When it is false the output is enabled.
7.5 Test/Admin
7.5.1 Self Test
Run a Module test.
Press thebutton then use the soft key to select Utilities > Test/Admin > Self-Test > Start.
Figure 7.9 Self Test Complete
7.5.2 Security
The Security Settings are locked and can be accessed by entering the default code 77416699.
Press thebutton then use the soft key to select Utilities > Test/Admin > Security > Change
Code.
Calibrate
Enter calibration mode.
Press thebutton then use the soft key to select Utilities > Test/Admin > Security > Calibrate .
See chapter Calibration Adjustment Procedure to adjust the units calibration.
Firmware Update
Update the rmware and module. Press thebutton then use the soft key to select Utilities >
Test/Admin > Security > Firmware Update .
NISPOM
Select for complete memory wipe. Press thebutton then use the soft key to select Utilities >
Test/Admin > Security > NISPOM .
Lock
Sets security status to Lock. Press thebutton then use the soft key to select Utilities > Test/Admin
> Security > Lock .
If Done is selected without locking the security settings, the user will be prompted that the security settings
will not be locked.
7.6 Error Log
View up to 50 previously set error codes.
Press thebutton then use the soft key to select Utilities > Error Log.
Errors are placed in the order that they were encountered. 1 being the most recent.
The error log displays up to 50 error codes.
After reaching 50 error codes no more codes will be reported in the error log.
To continue ling error codes the error log list must be cleared
The error log can be saved into a USB ash drive connected to the USB port on the front panel.
Save the log by:
Pressing thebutton then use the soft key to select Utilities > Error Log > Save .
Use the rotary knob and thebutton to navigate through the le path of the USB. Once the desired
location has been found select Save by pressing the softkey furthest to the left.
Clear the data log by:
Pressing thebutton then use the soft key to select Utilities > Error Log > Clear All > Yes.
Using the rotary knob navigate through the le paths of the USB. Use thebutton to expand folders.
Once the desired location has been found press Save.
The output settings will be saved after a short delay.
Figure 9.2 Save to USB
9.2 Recall the Instrument’s Settings
Recall previously saved output settings from the internal memory. A total of 10 (0 to 9) user settings can
be saved.
To recall settings :
Press thebutton then use the soft keys to select Recall > Recall from INT.
Use the numeric keypad to enter a number from 0 to 9 to recall the output settings stored in that address.
The settings will be recalled after a short delay.
Recall from USB
Recall the instrument’s output settings from a USB drive.
Press thebutton then use the soft keys to select Recall > Recall from USB.
Use the rotary knob to navigate through the le paths of the USB. Use thebutton to expand folders.
Once the desired location has been found press Load.
The settings will be recalled after a short delay.
Performance verication ensures the instrument will meet specications listed in the datasheet.
Load regulation can be tested for both the front and rear outputs.
Note:
All specications apply to the unit after a temperature stabilization time of 15 minutes over an ambient
temperature range of 23 °C ± 5 °C. Specications are valid for single unit operation only.
CV Load Regulation Front Panel Output
1. Turn o the power supply before making any connections.
2. Connect the: DMM, electronic load, and the rear panel remote sense to the power supply’s terminal
being tested.
(Remote sense must correspond with the channel being tested.)
4. Set the power supply settings of the channel being tested to the values given in table 11.1.
5. Set the electronic load settings to the stated values in table 11.1.
6. Enable remote sense. See section Remote Sense.
7. Enable the channel output.
8. Enable the electronic load. Monitor the power supply to ensure it remains in CV mode.If the power
supply switches to CC slightly lower the current on the electronic load until the power supply return
to CV mode.
9. Give the power supply some time to settle and record the DMM reading as V load.
10.Disable the electronic load and record the DMM reading as V noload.
11.Take the dierence between V load and V noload (V load - V noload) and record the value as the
voltage regulation. The dierence should be within the specication limit.
CV Load Regulation Rear Panel Output
1. Turn o the power supply before making any connections.
2. Connect the: DMM, electronic load, and the rear panel remote sense to the power supply’s rear +/terminal being tested.Use jumpers to connect the +S to + and -S to -. See gure 11.2
(Remote sense must correspond with the channel being tested.)
3. Turn on the power supply.
4. Set the power supply settings of the channel being tested to the values given in table 11.1.
5. Set the electronic load settings to the stated values in table 11.1.
8. Enable the electronic load. Monitor the power supply to ensure it remains in CV mode.If the power
supply switches to CC slightly lower the current on the electronic load until the power supply return
to CV mode.
9. Give the power supply some time to settle and record the DMM reading as V load.
10.Disable the electronic load and record the DMM reading as V noload.
11.Take the dierence between V load and V noload (V load - V noload) and record the value as the
voltage regulation. The dierence should be within the specication limit.