The following safety precautions apply to both operating and maintenance
personnel and must be followed during all phases of operation, service,
and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in
this manual.
Apply all the listed safety precautions.
Verify that the voltage selector at the line power cord input is set
to the correct line voltage. Operating the instrument at an
incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power.
Do not operate the instrument in ways not specified by this
manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in
this manual violates the safety standards of design, manufacture, and
intended use of the instrument. B&K Precision assumes no liability for a
customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the
amount of electrical energy available and the voltage impulses that may
occur on electrical conductors associated with these category ratings. The
category rating is a Roman numeral of I, II, III, or IV. This rating is also
accompanied by a maximum voltage of the circuit to be tested, which
defines the voltage impulses expected and required insulation clearances.
These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs
are not intended to be connected to the mains supply. The voltages in the
environment are typically derived from a limited-energy transformer or a
Category II (CAT II): Measurement instruments whose measurement inputs
are meant to be connected to the mains supply at a standard wall outlet or
similar sources. Example measurement environments are portable tools
and household appliances.
Category III (CAT III): Measurement instruments whose measurement
inputs are meant to be connected to the mains installation of a building.
Examples are measurements inside a building's circuit breaker panel or the
wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement
inputs are meant to be connected to the primary power entering a
building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher
category rating than what is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a
category rating equal to or higher than the instrument's category rating to
maintain the instrument's category rating. Failure to do so will lower the
category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains
power environment. The mains power should be 120 V RMS or 240 V RMS.
Use only the power cord supplied with the instrument and ensure it is
appropriate for your country of use.
To minimize shock hazard, the instrument chassis and cabinet must be
connected to an electrical safety ground. This instrument is grounded
through the ground conductor of the supplied, three-conductor AC line
power cable. The power cable must be plugged into an approved threeconductor electrical outlet. The power jack and mating plug of the power
cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground
connection, all accessible conductive parts (including control knobs) may
provide an electric shock. Failure to use a properly-grounded approved
outlet and the recommended three-conductor AC line power cable may
result in injury or death.
Unless otherwise stated, a ground connection on the instrument's front or
rear panel is for a reference of potential only and is not to be used as a
safety ground.
Do not operate in an explosive or flammable atmosphere
Do not operate the instrument in the presence of flammable gases or
vapors, fumes, or finely-divided particulates.
The instrument is designed to be used in office-type indoor environments.
Do not operate the instrument
In the presence of noxious, corrosive, or flammable fumes, gases,
vapors, chemicals, or finely-divided particulates.
In relative humidity conditions outside the instrument's
specifications.
In environments where there is a danger of any liquid being
spilled on the instrument or where any liquid can condense on
the instrument.
In air temperatures exceeding the specified operating
temperatures.
In atmospheric pressures outside the specified altitude limits or
where the surrounding gas is not air.
In environments with restricted cooling air flow, even if the air
temperatures are within specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2
environment. The operating temperature range is 0 °C to 40 °C and the
operating humidity is ≤ 85 % relative humidity at 40 °C, with no
condensation allowed. Measurements made by this instrument may be
outside specifications if the instrument is used in non-office-type
environments. Such environments may include rapid temperature or
humidity changes, sunlight, vibration and/or mechanical shocks, acoustic
noise, electrical noise, strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid,
chemical, or other material gets on or inside the instrument, remove the
instrument's power cord, remove the instrument from service, label it as
not to be operated, and return the instrument to B&K Precision for repair.
Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact
cleaners, abrasives, lubricants, solvents, acids/bases, or other such
chemicals. Clean the instrument only with a clean dry lint-free cloth or as
instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body
or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel.
Component replacement and internal adjustments must be made by
qualified service-trained maintenance personnel who are aware of the
hazards involved when the instrument's covers and shields are removed.
Under certain conditions, even with the power cord removed, dangerous
voltages may exist when the covers are removed. To avoid injuries, always
disconnect the power cord from the instrument, disconnect all other
connections (for example, test leads, computer interface cables, etc.),
discharge all circuits, and verify there are no hazardous voltages present
on any conductors by measurements with a properly-operating voltagesensing device before touching any internal parts. Verify the voltagesensing device is working properly before and after making the
measurements by testing with known-operating voltage sources and test
for both DC and AC voltages. Do not attempt any service or adjustment
unless another person capable of rendering first aid and resuscitation is
present.
Do not insert any object into an instrument's ventilation openings or other
openings.
Hazardous voltages may be present in unexpected locations in circuitry
being tested when a fault condition in the circuit exists.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify
this instrument. Return the instrument to B&K Precision for service and
repair to ensure that safety and performance features are maintained.
Cooling fans
This instrument contains one or more cooling fans. For continued safe
operation of the instrument, the air inlet and exhaust openings for these
fans must not be blocked nor must accumulated dust or other debris be
allowed to reduce air flow. Maintain at least 25 mm clearance around the
sides of the instrument that contain air inlet and exhaust ports. If mounted
in a rack, position power devices in the rack above the instrument to
minimize instrument heating while rack mounted. Do not continue to
operate the instrument if you cannot verify the fan is operating (note some
fans may have intermittent duty cycles). Do not insert any object into the
fan's inlet or outlet.
Do not place heavy objects on the instrument.
Do not obstruct cooling air flow to the instrument.
Do not place a hot soldering iron on the instrument.
Do not pull the instrument with the power cord, connected
probe, or connected test lead.
Do not move the instrument when a probe is connected to a
This product is subject to Directive
2002/96/EC of the European
Parliament and the Council of the
European Union on waste electrical
and electronic equipment (WEEE),
and in jurisdictions adopting that
Directive, is marked as being put on
the market after August 13, 2005,
and should not be disposed of as
unsorted municipal waste. Please
utilize your local WEEE collection
facilities in the disposition of this
product and otherwise observe all
applicable requirements.
Disposal of Old Electrical & Electronic Equipment (Applicable in the
European Union and other European countries with separate collection
systems)
This instrument meets the requirements of 2006/95/EC Low Voltage
Directive and 2004/108/EC Electromagnetic Compatibility Directive with
the following standards.
The B&K Precision 2190E digital storage oscilloscope (DSO) is a portable
benchtop instrument used for making measurements of signals and
waveforms. The oscilloscope’s bandwidth is capable of capturing up to 100
MHz signals with a real time sampling rate of up to 1 GSa/s. With up to 40k
points of deep memory, more details of a signal can be captured and
displayed on a large color LCD display for analysis.
Features:
2 channels with bandwidth up to 100 MHz
Single channel real-time sampling rate of up to 1 GSa/s
Up to 40k points of memory depth
7” Color TFT LCD display
Trigger types: Edge, Pulse, Video, Slope, and Alternative
Digital filter and waveform recorder functions
Automatic measurement of 32 parameters (Voltage and Time)
Standard USB host, USBTMC device, and LAN ports
1.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving
it. Unpack all items from the shipping carton, and check for any obvious
signs of physical damage that may have occurred during transportation.
Report any damage to the shipping agent immediately. Save the original
packing carton for possible future reshipment. Every instrument is shipped
with the following contents:
1 x Quick Start guide (printed)
1 x AC power cord
1 x USB type A to Type B cable
2 x 1:1/10:1 Passive Oscilloscope Probes
Verify that all items above are included in the shipping container. If
anything is missing, please contact B&K Precision.
1.3 Front Panel
It is important for you to familiarize yourself with the DSO’s front panel
before operating the instrument. Below is a brief introduction of the front
panel function operation.
The included AC power cord is safety certified for this
instrument operating in rated range. To change a cable or add
an extension cable, be sure that it can meet the required
power ratings for this instrument. Any misuse with wrong or
unsafe cables will void the warranty.
Before connecting and powering up the instrument, please review and go
through the instructions in this chapter.
2.1 Input Power Requirements
Input Power
The supply has a universal AC input that accepts line voltage and
frequency input within:
100 – 240 V (+/- 10%), 50/60 Hz (+/- 5%)
100 – 127 V, 45 – 440 Hz
Before connecting to an AC outlet or external power source, be sure that
the power switch is in the OFF position and verify that the AC power cord,
including the extension line, is compatible with the rated voltage/current
and that there is sufficient circuit capacity for the power supply. Once
verified, connect the cable firmly.
2.2 Preliminary Check
Complete the following steps to verify that the oscilloscope is ready for
use.
Verify and check to make sure proper AC voltages are available to
power the instrument. The AC voltage range must meet the
acceptable specification as explained in section 2.1.
Connect Power
Connect AC power cord to the AC receptacle in the rear panel and
press the power switch to the ON position to turn ON the
instrument. The instrument will have a boot screen while
loading, after which the main screen will be displayed.
Self-Test
The instrument has 3 self-test options to test the screen, keys,
and the LED back lights of the function, menu, and channel keys
as shown below.
Figure 2.1 – Self Test Menu
To perform the self-test, please refer to the Self Test section for
further instructions.
Self-Cal
This option runs an internal self-calibration procedure that will
check and adjust the instrument. To perform the self-calibration,
please refer to the Self Calibration section for further
instructions.
The model and firmware version can be verified from within the menu
system.
Press Utility and select System Status option. The software/firmware
version, hardware version, model, and serial number will be displayed.
Press the Single key to exit.
Function Check
Follow the steps below to do a quick check of the oscilloscope’s
functionality.
1. Power on the oscilloscope. Press “DEFAULT SETUP” to show the
result of the self-check. The probe default attenuation is 1X.
Figure 2.2 – Scope Layout
2. Set the switch to 1X on the probe and connect the probe to
channel 1 on the oscilloscope. To do this, align the slot in the
probe connector with the key on the CH 1 BNC, push to connect,
and twist to the right to lock the probe in place. Connect the
probe tip and reference lead to the PROBE COMP connectors.
To avoid electric shock when using the probe, keep fingers
behind the guard on the probe body.
To avoid electric shock while using the probe, do not touch
metallic portions of the probe head while it is connected to a
voltage source. Connect the probe to the oscilloscope and
connect the ground terminal to ground before you take any
measurements.
A guard around the probe body provides a finger barrier for protection
from electric shock.
Figure 2.5 – Oscilloscope Probe
Connect the probe to the oscilloscope and connect the ground terminal to
ground before you take any measurements.
NOTE: The default setting for the Probe option is 1X.
NOTE: When the attenuation switch is set to 1X, the probe limits
the bandwidth of the oscilloscope to 6 MHz (according to Probe
spec). To use the full bandwidth of the oscilloscope, be sure to set
the switch to 10X
Probes are available with various attenuation factors which affect the
vertical scale of the signal. The Probe Check function verifies that the
Probe attenuation option matches the attenuation of the probe.
You can push a vertical menu button (such as the CH 1 MENU button), and
select the Probe option that matches the attenuation factor of your probe.
Be sure that the attenuation switch on the probe matches the Probe
option in the oscilloscope. Switch settings are 1X and 10X.
Probe Compensation
As an alternative method to Probe Check, you can manually perform this
adjustment to match your probe to the input channel.
1. Set the Probe option attenuation in the channel menu to 10X. Do
so by pressing CH1 button and selecting “Probe” from menu.
Select 10X. Set the switch to 10X on the probe and connect the
probe to channel 1 on the oscilloscope. If you use the probe
hook-tip, ensure a proper connection by firmly inserting the tip
onto the probe.
2. Attach the probe tip to the PROBE COMP 3V connector and the
reference lead to the PROBE COMP Ground connector. Display
the channel and then push the “AUTO” button.
3. Check the shape of the displayed waveform.
Figure 2.7 – Compensation Illustration
4. If necessary, adjust your probe’s compensation trimmer pot.
Channel buttons (CH1, CH2): Press a channel button to turn
that channel ON or OFF and open the channel menu for that
channel. You can use the channel menu to set up a channel.
When the channel is on, the channel button is lit.
MATH: Press to display the Math menu. You can use the
MATH menu to use the oscilloscope’s Math functions.
REF: Press to display the Ref Wave menu. You can use this
menu to save and recall four or two reference waveforms to
and from internal memory.
HORI MENU: Press to display the Horizontal menu. You can
use the Horizontal menu to display the waveform and zoom
in a segment of a waveform.
TRIG MENU: Press to display the Trigger menu. You can use
the Trigger menu to set the trigger type (Edge. Pulse, Video,
Slope, Alternative) and trigger settings.
SET TO 50%: Press to stabilize a waveform quickly. The
oscilloscope can set the trigger level to be halfway between
the minimum and maximum voltage level automatically. This
is useful when you connect a signal to the EXT TRIG
connector and set the trigger source to Ext or Ext/5.
FORCE: Use the FORCE button to complete the current
waveform acquisition whether the oscilloscope detects a
trigger or not. This is useful for Single acquisitions and
Normal trigger mode.
SAVE/RECALL: Press to display the Save/Recall menu. You
can use the Save/Recall menu to save and recall up to 20
oscilloscope setups and 10 waveforms to/from internal
memory or a USB memory device (limited by memory
capacity of the USB flash drive). You can also use it to recall
the default factory settings, to save waveform data as a
comma-delimited file (.CSV), and to save the displayed
waveform image.
ACQUIRE: Press to display Acquire menu. You can use the
Acquire menu to set the acquisition Sampling Mode
(Sampling, Peak Detect, and Average).
MEASURE: Press to display a menu of measurement
parameters.
CURSORS: Display the Cursor Menu. Vertical Position
controls adjust cursor position while displaying the Cursor
Menu and the cursors are activated. Cursors remain
displayed (unless the “Type” option is set to “Off”) after
leaving the Cursor Menu but are not adjustable.
DISPLAY: Press to open the Display menu. You can use the
Display menu to set grid and waveform display styles, and
persistence.
UTILITY: Press to open the Utility menu. You can use the
Utility menu to configure oscilloscope features, such as
Auto set the screen and display
several cycles signal.
(Single-cycle sine)
Set the screen and auto display
single cycle signal.
The 2190E Digital Storage Oscilloscope has an Auto Setup function that
identifies the waveform types and automatically adjusts controls to
produce a usable display of the input signal.
Press the AUTO front panel button, and then press the menu option
button adjacent to the desired waveform as follows:
Auto detect the channel which has the
input signal
Trigger slope
Rising
Trigger mode
Auto
Trigger coupling
DC
Trigger holdoff
Minimum
Trigger level
Set to 50%
NOTE:The AUTO button can be disabled. Please see “Education
Mode” for details.
3.4 Default Setup
The oscilloscope is set up for normal operation when it is shipped from the
factory. This is the default setup. To recall this setup, press the DEFAULT SETUP button. For the default options, buttons and controls when you
press the DEFAULT SETUPbutton, refer to “Table 3.3 – Default Setup
Table” below.
The DEFAULT SETUP button does not reset the following settings:
You can use the Universal knob with many functions, such as adjusting the
hold off time, moving cursors, setting the pulse width, setting the video
line, adjusting the upper and lower frequency limit, adjusting the X and Y
masks when using the pass/fail function, etc. You can also turn the
“Universal” knob to adjust the storage position of setups, waveforms,
pictures when saving/recalling, and to select menu options. With some
functions, the light indicator above the knob will turn on to indicate that
the knob can be used to make changes or adjustments for that function.
The knob can also be pushed to make a selection after
changes/adjustments have been made.
3.6 Vertical System
The vertical control could be used for displaying waveform, rectify scale
and position.
1.Use the Vertical “POSITION” knobs to move the channel
waveforms up or down on the screen. This button’s
resolution varies as per the vertical scale.
2. When you adjust the vertical position of channels
waveforms, the vertical position information will display
on the bottom left of the screen. For example “Volts Pos=24.6mV”.
3. Press the vertical “POSITION” knob to set the vertical
position to zero.
•“Volts/div” Knob
1. Use the “Volts/div” knobs to control how the
oscilloscope amplifies or attenuates the source signal of
channel waveforms. When you turn the “volts/div”
knob, the oscilloscope increases or decreases the vertical
size of the waveform on the screen with respect to the
ground level.
2. When you press the “Volt/div” knob, you can switch
“Volts/div” option between “Coarse” and “Fine”. The
vertical scale is set by a 1-2-5 step sequence in Coarse
Option
Setting
Introduction
Coupling
DC
AC
GND
DC passes both AC and DC
components of the input signal.
AC blocks the DC component of the
input signal and attenuates signals
below 10 Hz.
GND disconnects the input signal.
BW limit
On
Off
Limits the bandwidth to reduce
display noise; filters the signal to
reduce noise and other unwanted
high frequency components.
Volts/Div
Coarse
Fine
Selects the resolution of the
Volts/Div knob
Coarse defines a 1-2-5 sequence.
Fine changes the resolution to small
steps between the coarse settings.
mode. Increase in the clockwise direction and decrease
in the counterclockwise direction. In Fine mode, the
knob changes the Volts/Div scale in small steps between
the coarse settings. Again, increase in the clockwise
direction and decrease in the counterclockwise
direction.
The oscilloscope’s vertical response rolls off slowly above its
specified bandwidth. Therefore, the FFT spectrum can show valid
frequency information higher than the oscilloscope’s bandwidth.
However, the magnitude information near or above the
bandwidth will not be accurate.
•“GND” Coupling: Use GND coupling to display a zero-volt
waveform. Internally, the channel input is connected to a zerovolt reference level.
•Fine Resolution: The vertical scale readout displays the actual
Volts/Div setting while in the fine resolution setting. Changing the
setting to coarse does not change the vertical scale until the
VOLTS/DIV control is adjusted.
If the channel is set to DC coupling, you can quickly measure the
DC component of the signal by simply noting its distance from the
ground symbol.
If the channel is set to AC coupling, the DC component of the
signal is blocked allowing you to use greater sensitivity to display
the AC component of the signal.
Setting up Channels
Each channel has its own separate Menu. The items are set up separately
according to each channel.
1. Setup Channel Coupling
Take CH1 for example; the tested signal is a sine wave signal with
DC deflection:
• Press “CH1”→“Coupling”→“AC”, Set to AC couple mode.
This will block the DC component of the input signal.
• Press “CH1”→“Coupling”→“DC”, Set to DC couple mode.
Both DC and AC components of the input signal will be
captured.
Vertical scale adjust have Coarse and Fine modes, Vertical
sensitivity range of 2 mV/div – 10 V/div.
Take CH1 for example:
• Press “CH1”→“Volts/Div”→“Coarse”. It is the default
setting of Volts/Div, and makes the vertical scaling in a 12-5-step sequence from 2 mV/div, 5 mV/div, and 10
mV/div to 10 V/div.
• Press “CH1”→ Volts/Div”→ Fine”. This setting changes
the vertical control to small steps between the coarse
settings. It will be helpful when you need to adjust the
waveform vertical size in smaller steps.
In order to set the attenuation coefficient, you need to specify it
in the channel operation Menu. If the attenuation coefficient is
10:1, the input coefficient should be set to 10X, so that the
Volts/div information and measurement testing is correct.
Take CH1 for example, when you use the 100:1 probe:
Math shows the results after +,-,*, / and FFT operations of the CH1 and
CH2. Press the MATH button to display the waveform math operations.
Press the MATH button again to remove the math waveform display.
The FFT process mathematically converts a time-domain signal into its
frequency components. You can use the Math FFT mode to view the
following types of signals:
• Analyze the harmonic wave in the Power cable.
• Test the harmonic content and distortion in the system
• Show the Noise in the DC Power supply
• Test the filter and pulse response in the system
To use the Math FFT mode, you need to perform the following steps:
1. Set up the source (time-domain) waveform.
2. Press the AUTO button to display an YT waveform.
3. Turn the vertical “POSITION” knob to move the YT waveform to
the center vertically (zero divisions).
4. Turn the horizontal “POSITION” knob to position the part of the
YT waveform that you want to analyze in the center eight
divisions of the screen. The oscilloscope calculates the FFT
spectrum using the center 1024 points of the time-domain
waveform.
5. Turn the “Volts/div” knob to ensure that the entire waveform
remains on the screen.
6. Turn the “S/div” knob to provide the resolution you want in the
FFT spectrum.
7. If possible, set the oscilloscope to display many signal cycles.
To display FFT correctly, follow these steps:
1. Push the “MATH” button.
2. Set the “Operation” option to FFT.
3. Press the “Source” button to select “CH1” or “CH2” according to
input signal channel.
4. Turn the “Time/div” knob to adjust the sampling rate (this
parameter is displayed behind the time base parameter), making
sure it is at least double the input signal frequency. (to avoid
aliasing according to Nyquist’s theorem)
Displaying the FFT Spectrum
Press the MATH button to display the Math Menu. Use the options to
select the Source channel, Window algorithm, and FFT Zoom factor. You
can display only one FFT spectrum at a time. You can select “Full screen”
or “Split” in “Display” option to display FFT waveform on full screen or
display channel waveform and its FFT waveform on half screen at a time.
Best frequency
resolution, worst
magnitude resolution.
This is essentially the
same as no window.
Symmetric transients
or bursts. Equalamplitude sine waves
with fixed frequencies.
Broadband random
noise with a relatively
slowly varying
spectrum.
Select FFT window
Windows reduce spectral leakage in the FFT spectrum. The FFT assumes
that the YT waveform repeats forever. With an integral number of cycles,
the YT waveform starts and ends at the me amplitude and there are no
discontinuities in the signal shape A non-integral number of cycles in the
YT waveform causes the signal start and end points to be at different
amplitudes. The transitions between the start and end points cause
discontinuities in the signal that introduce high-frequency transients.
Better frequency,
poorer magnitude
accuracy than
Rectangular. Hamming
has slightly better
frequency resolution
than Hanning.
Sine, periodic, and
narrow-band random
noise. Asymmetric
transients or bursts.
Blackman
Best magnitude, worst
frequency resolution.
Single frequency
waveforms, to find
higher order
harmonics.
Magnifying the FFT Spectrum
You can magnify and use cursors to take measurements on the FFT
spectrum. The oscilloscope includes an “FFT Zoom” option to magnify
horizontally, press this option button to select “1X”, “2X”, “5X” or “10X”.
Moreover, you also can turn the “Universal” knob to magnify FFT
waveform horizontally in a 1-2-5 step. To magnify vertically, turn the
“Volts/div” knob.
Measuring an FFT Spectrum Using Cursors
You can take two measurements on FFT spectrums: magnitude (in dB) and
frequency (in Hz). Magnitude is referenced to 0 dB, where 0 dB equals 1
VRMS. You can use the cursors to take measurements at any zoom factor.
Use horizontal cursors to measure amplitude and vertical cursors to
measure frequency.
If you input a sine signal into channel 1, follow these steps:
• Measure FFT Amplitude
1. Input a sine signal to channel 1, and press the
“AUTO” button.
2. Press the “MATH” button to enter the “MATH”
menu.
3. Press the “Operation” option button to select “FFT”.
4. Press the “Source” option button to select “CH1”.
4. Press the “Source” option button to select “MATH”.
NOTE:
The FFT of a waveform that has a DC component or offset can
cause incorrect FFT waveform magnitude values. To minimize the
DC component, choose AC Coupling on the source waveform.
To display FFT waveforms with a large dynamic range, use the
dBVrms scale. The dBVrms scale displays component magnitudes
using a log scale.
The Nyquist frequency is the highest frequency that any real-time
digitizing oscilloscope can acquire without aliasing. This frequency
is half that of the sample rate provided it is within the analog
5. Press the “CurA” option button, turn the “Universal”
button to move Cursor A to the highest position of
the FFT waveform.
6. The value of CurA displaying on the top of the left
screen is FFT frequency. This frequency should be
the same as input signal frequency.
Turn on this function for main time base
waveform to display on the top half screen
and window time base waveform to display
on the below half screen at the same time.
Turn off this function to only display main
time base waveform on the screen.
Shown below are two knobs and one button in the HORIZONTAL area.
Figure 3.18 – Horizontal Controls
Table 3.15 – Horizontal System Function Menu
Horizontal Control Knob
You can use the horizontal controls to change the horizontal scale and
position of waveforms. The horizontal position readout shows the time
represented by the center of the screen, using the time of the trigger as
zero. Changing the horizontal scale causes the waveform to expand or
contract around the screen center.
• Horizontal “POSITION” knob
1. Adjust the horizontal position of all channels and
math waveforms (the position of the trigger relative
to the center of the screen). The resolution of this
control varies with the time base setting.
2. When you press the horizontal “POSITION” Knob,
you can set the horizontal position to zero.
•“Time/div” knob
1. Used to change the horizontal time scale to magnify
or compress the waveform. If waveform acquisition
is stopped (using the RUN/STOP or SINGLE button),
turn the Time/div knob to expand or compress the
waveform.
2. Select the horizontal Time/div (scale factor) for the
main or the window time base. When Window Zone
is enabled, it changes the width of the window zone
by changing the window time base.
•Display scan mode
When the Time/div control is set to 100 ms/div or slower and the
trigger mode is set to Auto, the oscilloscope enters the scan
acquisition mode. In this mode, the waveform display updates
from left to right. There is no trigger or horizontal position control
of waveforms during scan mode.
Window Zone
Use the Window Zone option to define a segment of a waveform to see
more detail. This function behaves like zooming into a portion of the
captured waveform. The window time base setting cannot be set slower
than the Main time base setting.
You can turn the Horizontal Position and Time/div controls to enlarge or
minimize waveforms in the Window Zone.
If you want to see a section of the waveform in details, follow these steps:
1. Press the “HORI MENU” button to enter the “Horizontal menu”.
2. Turn the “Time/div” knob to change the main time base scale.
3. Press the “Delayed” option button to select “On”.
Figure 3.19 – Horizontal Delay Menu
4. Turn the “Horizontal Position” knob (adjust window’s position) to
select the window that your need and expanded window
waveform display on the below half screen at the same time.
3.11 Trigger System
The trigger determines when the oscilloscope starts to acquire data and
display a waveform. When a trigger is set up properly, the oscilloscope
converts unstable displays or blank screens into meaningful waveforms.
Here are three buttons and one knob in the trigger area. See below:
• “TRIG MENU” Button: Press the “TRIG MENU” Button to
display the “Trigger Menu”.
•“LEVEL” Knob: The LEVEL knob is to set the
corresponding signal voltage of trigger point in order to
sample. Press the “LEVEL” knob to set trigger level to
zero.
• “SET TO 50%” Button: Use the “SET TO 50%” button to
stabilize a waveform quickly. The oscilloscope can set
the Trigger Level to be about halfway between the
minimum and maximum voltage levels automatically.
This is useful when you connect a signal to the EXT TRIG
BNC and set the trigger source to Ext or Ext/5.
•“FORCE Button: Use the FORCE button to complete the
current waveform acquisition whether the oscilloscope
detects a trigger or not. This is useful for SINGLE
acquisitions and Normal trigger mode.
Signal Source
You can use the Trigger Source options to select the signal that the
oscilloscope uses as a trigger. The source can be any signal connected to a
channel BNC, to the EXT TRIG BUS, or the AC power line (available only
with Edge Trigger).
With Edge highlighted, the rising or falling edge
of the input signal is used for the trigger.
Source
CH1
CH2
Triggers on a channel whether or not the
waveform is displayed.
EXT
Does not display the trigger signal; the Ext
option uses the signal connected to the EXT TRIG
front-panel BNC and allows a trigger level range
of -1.2V to +1.2V.
EXT/5
Same as Ext option, but attenuates the signal by
a factor of five, and allows a trigger level range
of +6V to -6V.This extends the trigger level
range.
AC Line
This selection uses a signal derived from the
power line as the trigger source; trigger coupling
is set to DC and the trigger level to 0 volts.
Slope
Trigger on Rising edge of the trigger signal.
Trigger on Falling edge of the trigger signal.
Trigger Type
The scopes have five trigger types: Edge, Video, Pulse, Slope, and
Alternate.
Edge Trigger
Use Edge triggering to trigger on the edge of the oscilloscope input signal
at the trigger threshold.
Trigger on Rising edge and Falling edge of the
trigger signal.
Mode
Auto
Use this mode to let the acquisition free-run in
the absence of a valid trigger; This mode allows
an untriggered, scanning waveform at 100
ms/div or slower time base settings.
Normal
Use this mode when you want to see only valid
triggered waveforms; when you use this mode,
the oscilloscope does not display a waveform
until after the first trigger.
Single
When you want the oscilloscope to acquire a
single waveform, press the “SINGLE” button.
Set up
Enter the “Trigger Setup Menu” (See Table 3.17).
Option
Setting
Explain
Coupling
DC
Passes all components of the signal
AC
Blocks DC components, attenuates
signals below 50 Hz.
HF Reject
Attenuates the high-frequency
components above 150 kHz.
LF Reject
Blocks the DC component, attenuates the
low-frequency components below 7 kHz.
Holdoff
Using the “universal” knob to adjust
holdoff time (sec), the holdoff value is
displayed.
Select the pulse to
trigger the pulse match
the trigger condition.
Source
CH1
CH2
EXT
EXT/5
AC Line
Select input signal
source.
Press the “Trigger mode” option button to select “Auto”,
“Normal”, or “Single”.
Auto: The waveform refreshes at a high speed whether
the trigger condition is satisfied or not.
Normal: The waveform refreshes when the trigger
condition is satisfied and waits for next trigger event
occurring when the trigger condition is not satisfied.
Single: The oscilloscope acquires a waveform when the
trigger condition is satisfied and then stops
5. Set up Trigger coupling
Press the “Set Up” button to enter the “Trigger Setup
Menu”.
Press the “Coupling” option button to select “DC”, “AC”,
“HF Reject” or “LF Reject”.
Pulse Trigger
Use Pulse Width triggering to trigger on aberrant pulses.
When you select the video
type, put the couple set to the
AC, then you could trigger the
NTSC, PAL and SECAM video
signal.
Standard
NTSC
Pal/Secam
Select the video standard for
sync and line number count.
Mode
Auto
Use this mode to let the
acquisition free-run in the
absence of a valid trigger; This
mode allows an untriggered,
scanning waveform at 100
ms/div or slower time base
settings.
Normal
Use this mode when you want
to see only valid triggered
waveforms; when you use this
mode, the oscilloscope does
not display a waveform until
after the first trigger.
Single
When you want the
oscilloscope to acquire a single
waveform, press the “SINGLE”
button.
Trigger on positive slope or negative
slope according to setup time of the
oscilloscope.
Vertical
Select the trigger level that can be
adjusted by “LEVEL” knob. You can
adjust “LEVEL A”, “LEVEL B” or adjust
them at the same time.
Mode
Auto
Use this mode to let the acquisition freerun in the absence of a valid trigger; This
mode allows an untriggered, scanning
waveform at 100 ms/div or slower time
base settings.
Use this mode when you want to see
only valid triggered waveforms; when
you use this mode, the oscilloscope does
not display a waveform until after the
first trigger.
Single
When you want the oscilloscope to
acquire a single waveform, press the
“SINGLE” button.
Set up
Enter “Trigger setup menu” (See Table
3.17).
Next Page
Page 2/2
Return to the first page of slope trigger.
Figure 3.26 – Slope Trigger Menu 2
Operating Instructions:
Follow the next steps after “Slope Trigger” is selected:
1. Input a signal to CH1 or CH2.
2. Press the “AUTO” button.
3. Press the “TRIG MENU” button to enter “Trigger menu”.
4. Press the “Type” option button to select “Slope”.
5. Press the “Source” option button to select “CH1” or “CH2”.
6. Press the “When” option button to select “”, “”,
“”, “”, “” and “”.
7. Press the “Time” button, turn the “Universal” knob to adjust
slope time.
8. Press the “Next Page - Page 1/2” option button to enter the
second page of the “Slope trigger menu”.
9. Press the “Vertical” option button to select trigger level that can
be adjusted.
10. Turn the “LEVEL” knob.
Alternate Trigger
The trigger signal comes from two vertical channels when you use
alternate trigger. In this mode, you can observe two irrelative signals at
the same time. You can select different trigger types for two vertical
signals, and selected types cover edge, pulse, video and slope trigger.
Trigger information of two channel signals display on the bottom right of
the screen.
Table 3.25 – Alternate Trigger Pulse Mode Function Menu 1
Option
Setting
Description
Type
Alternate
The trigger signal comes from two
vertical channels when you use
alternative trigger. In this mode, you
can observe two irrelative signals at the
same time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1
CH2
Set trigger type information for CH1
signal
Set trigger type information for CH2
signal
Mode
Pulse
Set trigger type of the vertical channel
signal to Pulse trigger.
Next Page
Page 1/2
Enter the second page of Alternative
trigger menu.
Option
Setting
Description
When
Select how to compare the trigger
pulse relative to the value selected in
the Set Pulse Width option.
Set Width
20.0ns-10.0s
Selecting this option can turn the
universal to set up the pulse width.
Table 3.26 – Alternate Trigger Pulse Mode Function Menu 2
The trigger signal comes from two
vertical channels when you use
alternative trigger. In this mode, you
can observe two irrelative signals at
the same time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1
CH2
Set trigger type information for CH1
signal
Set trigger type information for CH2
signal
Mode
Video
Set trigger type of the vertical
channel signal to Video trigger.
Next Page
Page 1/2
Enter the second page of Alternative
trigger menu.
Option
Setting
Description
Polarity
(Normal)
(Inverted)
Normal triggers on the negative edge
of the sync pulse.
Inverted triggers on the positive edge
of the sync pulse.
Table 3.27 – Alternate Trigger Video Mode Function Menu 1
Table 3.28 – Alternate Trigger Video Mode Function Menu 2
Select the video standard for sync and
line number count.
Set up
Enter the “Trigger Setup Menu” (See
Table 3.17).
Next Page
Page 2/2
Press this button to return to the first
page.
Option
Setting
Description
Type
Alternative
The trigger signal comes from two
vertical channels when you use
alternative trigger. In this mode, you
can observe two irrelative signals at the
same time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1
CH2
Set trigger type information for CH1
signal
Set trigger type information for CH2
signal
Mode
Slope
Set trigger type of the vertical channel
signal to slope trigger.
Next Page
Page 1/2
Enter the second page of the
alternative trigger.
Table 3.29 – Alternate Trigger Slope Mode Function Menu 1
6. Press the “Source” option button to select “CH1”.
7. Press the CH1 button and turn the “Time/div” knob to optimize
waveform display.
8. Press “Mode” option button to select “Edge”, “Pulse”, “Slope”
or “Video”.
9. Set the trigger according to trigger edge.
10. Press the “Source” option button to select “CH2”.
11. Press the CH2 button and turn the “Time/div” knob to optimize
waveform display.
12. Repeat steps 8 and 9.
Coupling
Use “Coupling” to make sure the signal passes through the trigger circuit.
It is useful for obtaining a steady waveform.
If you use trigger coupling, you should press the “TRIG MENU” button and
then select “Edge”, “Pulse”, “Video”, or “Slope” trigger. Then select the
“Coupling” option in the “Set Up menu”.
Position
The horizontal position control establishes the time between the trigger
position and the screen center. You can adjust the horizontal “POSITION”
knob control to view waveform data before the trigger, after the trigger,
or some of each. When you change the horizontal position of a waveform,
you are changing the time between the trigger and the center of the
display actually. (This appears to move the waveform to the right or left on
the display.)
NOTE: Press the SINGLE button when you want the oscilloscope to
acquire a single waveform.
Trigger coupling affects only the signal passed to the trigger system.
It does not affect the bandwidth or coupling of the signal displayed
on the screen.
Normal Polarity Sync triggers always occur on negative-going
horizontal sync pulses. If the video waveform has positive-going
horizontal sync pulses, use the Inverted Polarity selection.
Falling edge
Rising edge
T
r
i
g
g
e
r
l
e
v
e
l
c
a
n
b
e
a
d
j
u
s
t
e
d
v
e
r
t
i
c
a
l
l
y
The Slope and Level controls help to define the trigger. The Slope option
(Edge trigger type only) determines whether the oscilloscope finds the
trigger point on the rising and/or the falling edge of a signal.
The TRIGGER LEVEL knob controls at what point on the edge the trigger
occurs.
Figure 3.28 – Rising and Falling Edge Illustration
NOTE: Use trigger holdoff to help stabilize the display of aperiodic
waveforms.
You can use the Trigger Holdoff function to produce a stable display of
complex waveforms. Holdoff is time between when the oscilloscope
detects one trigger and when it is ready to detect another. The
oscilloscope will not trigger during the holdoff time. For a pulse train, you
can adjust the holdoff time so the oscilloscope triggers only on the first
pulse in the train.
Figure 3.29 – Trigger Holdoff Illustration
If you want to change holdoff time, please follow the steps below:
1. Press the “TRIG MENU” button to show the “TRIG
Menu”.
2. Press the “Type” option button to select trigger type.
3. Press the “Set Up” option button to enter the “Trigger
setup menu”.
4. Press the “Holdoff” option button and turn the
“Universal” knob to change the holdoff time until the
Use for sampling and accurately
display most of the waveform.
Peak Detect
Detect the noise and decrease the
possibility of aliasing.
Average
Use to reduce random or
uncorrelated noise in the signal
display.
Averages
(4, 16, 32, 64,
128, 256)
Select number of averages.
Sinx/x
Sinx
x
Use sin interpolation
Use linear interpolation
Mode
Equ time
Real time
Set the Sampling mode to Equivalent
time.
Set the Sampling mode to Real time.
Sa Rate
Displays system sampling rate.
Shown below is the “ACQUIRE” button for entering the menu for
“Acquiring Signals”.
Table 3.31 – Acquire Function Menu
When you acquire a signal, the oscilloscope converts it into digital form
and displays a waveform. The acquisition mode defines how the signal is
digitized and the time base setting affects the time span and level of detail
in the acquisition.
Sampling: In this acquisition mode, the oscilloscope samples
the signal in evenly spaced intervals to construct the waveform.
This mode accurately represents signals most of the time.
Advantage: You can use this mode to reduce random noise.
Disadvantage: This mode does not acquire rapid variations in the
signal that may occur between samples. This can result in aliasing
may cause narrow pulses to be missed. In these cases, you should
use the Peak Detect mode to acquire data.
Figure 3.30 – Acquire Menu
Peak Detect: Peak Detect mode capture the maximum and
minimum values of a signal Finds highest and lowest record
points over many acquisitions.
Advantage: In this way, the oscilloscope can acquire and display
narrow pulses, which may have otherwise been missed in Sample
mode.
Disadvantage: Noise will appear to be higher in this mode.
Equivalent Time Sampling: The equivalent time sampling mode
can achieve up to 20 ps of horizontal resolution (equivalent to
50GSa/s). This mode is good for observing repetitive waveforms.
Real Time Sampling: The scope has a maximum Real-time
sampling rate of 1GSa/s.
“RUN/STOP” Button: Press the RUN/STOP button when you want
the oscilloscope to acquire waveforms continuously. Press the
button again to stop the acquisition.
“SINGLE” Button: Press the SINGLE button to acquire a single
waveform. Each time you push the SINGLE button, the
oscilloscope begins to acquire another waveform. After the
oscilloscope detects a trigger it completes the acquisition and
stops.
When you push the RUN/STOP or SINGLE buttons to start an
acquisition, the oscilloscope goes through the following steps:
1. Acquire enough data to fill the portion of the waveform
record to the left of the trigger point. This is also called
the pre-trigger.
2. Continue to acquire data while waiting for the trigger
Sampled points
Apparent low-frequency
waveform due to aliasing
Actual high-frequency waveform
condition to occur.
3. Detect the trigger condition.
4. Continue to acquire data until the waveform record is
full.
5. Display the newly-acquired waveform.
Time Base: The oscilloscope digitizes waveforms by acquiring the
value of an input signal at discrete points. The time base allows
you to control how often the values are digitized. To adjust the
time base to a horizontal scale that suits your purpose, use the
Time/div knob.
Time Domain Aliasing: Aliasing occurs when the oscilloscope
does not sample the signal fast enough to construct an accurate
waveform record. When this happens, the oscilloscope displays a
waveform with a frequency lower than the actual input
waveform, or triggers and displays an unstable waveform.
Figure 3.33 – Time Domain Aliasing Illustration
Operating Instructions:
Set up Sampling Format
You can press the “Acquisition” option button or turn the “Universal” knob
to select “Sampling” mode, “Peak Detect” mode or “Average” mode.
Set up Averages
When you select “Average” format, you can press the “Averages” option
button to select “4”, “16”, “32”, “64”, “128”or “256”.
Vectors fill the space between
adjacent sample points in the
display.
There is no link between
adjacent sample points.
Persist
Off
1 sec
2 sec
5 sec
Infinite
Sets the length of time each
displayed sample point remains
displayed.
Intensity
<Intensity>
Set waveforms’ intensity.
Brightness
<Brightness>
Set grid brightness.
Next Page
Page 1/3
Press this button to enter second
page.
You can select Sinx interpolation or linear interpolation.
Set up Sampling Mode
Press the “Mode” option button to select “Real Time” or “Equ Time”.
Set up Sampling Rate
The sampling rate is based on the time division scaling on the screen.
Adjust the sampling rate by turning the Time/div front panel knob. The
sampling rate is shown under the “Sa Rate” display.
3.13 Display System
The display function can be setup by pressing the “DISPLAY” button.
YT format displays the vertical
voltage in relation to time
(horizontal scale). XY format
displays a dot each time a sample
is acquired on channel 1 and
channel 2
Screen
Normal
Inverted
Set to normal mode.
Set to invert color display mode.
Grid
Display grids and axes on the
screen.
Turn off the grids.
Turn off the grids and axes.
NOTE: The oscilloscope can capture a waveform in normal YT mode
at any sampling rate. You can view the same waveform in XY mode.
To do so, stop the acquisition and change the display format to XY.
Press the “Grid” option button to select “”, “”or“
”to set the screen whether display grid or not.
Set up Menu Display
Press the “Menu Display” option button to select “2 sec”, “5sec”,
“10sec”, “20sec” or “Infinite” to set menu display time on screen.
Set Skin
Press the “skin” option button or turn the “Universal” knob to select
“Classical”, “Modern”, “Traditional” or “Succinct”.
X-Y Format
Use the XY format to analyze phase differences, such as those represented
by Lissajous patterns. The format plots the voltage on channel 1 against
the voltage on channel 2, where channel 1 is the horizontal axis and
channel 2 is the vertical axis. The oscilloscope uses the untriggered Sample
acquisition mode and displays data as dots.
Operation steps:
Channel 1 “Volt/div” and vertical “POSITION” set up the
horizontal scale and position.
Channel 2 “Volt/div” and vertical “POSITION” set up the
horizontal scale and position.
Turn the “Time/div” knob to adjust the sampling rate
The following functions are forbidden in XY display form:
o Benchmark wave form and wave mathematic
o Cursor
o Auto (resets display format to YT)
o Trigger Control
NOTE: The CURSORS and MEASURE buttons can be disabled. Please
see “Education Mode” for details.
o Vector Display Type
o Scan Display
3.14 Measure System
The oscilloscope displays the voltage in relation to time and tests the
waveform displayed. Different measurement techniques such as scale,
Cursor and auto measure modes are used.
Scale Measurement
This method allows you to make a quick, visual estimate. For example, you
might look at waveform amplitude and determine that it is a little more
than 100 mV. You can take simple measurements by counting the major
and minor graticule divisions involved and multiplying by the scale factor.
For example, if you counted five major vertical graticule divisions between
the minimum and maximum values of a waveform and knew you had a
scale factor of 100 mV/div, then you could easily calculate your peak-topeak voltage as follows:
5 𝑑𝑖𝑣𝑖𝑠𝑖𝑜𝑛𝑠 ×
100 𝑚𝑉
𝑑𝑖𝑣𝑖𝑠𝑖𝑜𝑛
= 500 𝑚𝑉
Cursor Measurement
Press the “CURSORS” button to display the “Cursor” menu.
The cursor measurement has three modes: Manual, Track, and Auto
Measure.
Use cursor to measure voltage
parameters.
Use cursor to measure time
parameters.
Source
CH1
CH2
MATH
REFA
REFB
Select input signal channel.
Cur A
Select this option, use “Universal”
knob to adjust cursor A.
Cur B
Select this option, use “Universal”
knob to adjust cursor B.
Table 3.35 – Manual Cursor Menu
In this mode, the screen displays two horizontal parallel cursors or vertical
parallel cursors to measure voltage or time. You can move the cursor by
turning the “Universal” knob. Before using cursors, you should make sure
that you have set the signal source as the channel for measuring.
Voltage Cursor: Voltage cursors appear as horizontal lines on the
display and measure the vertical parameters.
Time Cursor: Time cursors appear as vertical lines on the display
and measure the horizontal parameters.
Cursor Moving: Use the “universal” knob to move cursor 1 and
cursor 2. They could be moved when the corresponding cursor
option are selected, and cursor value will display on the bottom
left and top left of the screen when you move the cursor.
ΔV: Vertical space between Cursor A and Cursor B (Voltage
value between two cursors).
Figure 3.38 – Cursor Menu (Track)
Auto Mode
This mode will take effect with automatic measurements. The instruments
will display cursors while measuring parameters automatically. These
cursors demonstrate the physical meanings of these measurements.
To do auto cursor measurements, follow these steps:
1. Press the CURSOR button to enter “Cursor measure menu”.
2. Press the “Cursor Mode” option button to select “Auto”.
3. Press the “MEASURE” button to enter “Auto cursor measure
mode menu” to select the parameter that you want to
Press this button to enter the Voltage measure menu.
Auto Measurement
When you take automatic measurements, the oscilloscope does all the
calculation for you. The measurements use all the recorded points in the
memory, which are more accurate than measurements made using the
graticule lines or cursor measurements because these measurements are
confined to be made by only using points on the display and not all the
data points recorded by the oscilloscope.
Press the ‘MEASURE’ button for Automatic Test.
There are three auto measurement types: Voltage Measure, Time
Measure, and Delay Measure. There are a total of 32 measurement
parameters.