BK Precision 2190D User Manual

Model: 2190D
100 MHz Digital Storage Oscilloscope
USER MANUAL
Safety Summary
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument.
Before applying power to this instrument:
Read and understand the safety and operational information in
this manual.
Apply all the listed safety precautions. Verify that the voltage selector at the line power cord input is set
to the correct line voltage. Operating the instrument at an incorrect line voltage will void the warranty.
Make all connections to the instrument before applying power. Do not operate the instrument in ways not specified by this
manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Category rating
The IEC 61010 standard defines safety category ratings that specify the amount of electrical energy available and the voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested, which defines the voltage impulses expected and required insulation clearances. These categories are:
Category I (CAT I): Measurement instruments whose measurement inputs are not intended to be connected to the mains supply. The voltages in the environment are typically derived from a limited-energy transformer or a
ii
battery.
Category II (CAT II): Measurement instruments whose measurement inputs are meant to be connected to the mains supply at a standard wall outlet or similar sources. Example measurement environments are portable tools and household appliances.
Category III (CAT III): Measurement instruments whose measurement inputs are meant to be connected to the mains installation of a building. Examples are measurements inside a building's circuit breaker panel or the wiring of permanently-installed motors.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power entering a building or other outdoor wiring.
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating. Failure to do so will lower the category rating of the measuring system.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 120 V RMS or 240 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Ground the Instrument
iii
To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded
through the ground conductor of the supplied, three-conductor AC line power cable. The power cable must be plugged into an approved three­conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.
Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved
outlet and the recommended three-conductor AC line power cable may result in injury or death.
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a reference of potential only and is not to be used as a safety ground.
Do not operate in an explosive or flammable atmosphere
Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely-divided particulates.
iv
The instrument is designed to be used in office-type indoor environments. Do not operate the instrument
In the presence of noxious, corrosive, or flammable fumes, gases,
vapors, chemicals, or finely-divided particulates.
In relative humidity conditions outside the instrument's
specifications.
In environments where there is a danger of any liquid being
spilled on the instrument or where any liquid can condense on the instrument.
In air temperatures exceeding the specified operating
temperatures.
In atmospheric pressures outside the specified altitude limits or
where the surrounding gas is not air.
In environments with restricted cooling air flow, even if the air
temperatures are within specifications.
In direct sunlight.
This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0 °C to 40 °C and the operating humidity is ≤ 85 % relative humidity at 40 °C, with no condensation allowed. Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.
Do not operate instrument if damaged
If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument's power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair.
v
Notify B&K Precision of the nature of any contamination of the instrument.
Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual.
Not for critical applications
This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.
Do not touch live circuits
Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument's covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed. To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltage­sensing device before touching any internal parts. Verify the voltage­sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test
vi
for both DC and AC voltages. Do not attempt any service or adjustment unless another person capable of rendering first aid and resuscitation is present. Do not insert any object into an instrument's ventilation openings or other openings.
Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.
Servicing
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
Cooling fans
This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm clearance around the sides of the instrument that contain air inlet and exhaust ports. If mounted in a rack, position power devices in the rack above the instrument to minimize instrument heating while rack mounted. Do not continue to operate the instrument if you cannot verify the fan is operating (note some fans may have intermittent duty cycles). Do not insert any object into the fan's inlet or outlet.
vii
For continued safe use of the instrument
Do not place heavy objects on the instrument. Do not obstruct cooling air flow to the instrument. Do not place a hot soldering iron on the instrument. Do not pull the instrument with the power cord, connected
probe, or connected test lead.
Do not move the instrument when a probe is connected to a
circuit being tested.
viii
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European
Union and other European countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
ix
CE Declaration of Conformity
This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards.
Low Voltage Directive
- EN61010-1: 2001
- EN61010-031: 2002+A1: 2008
EMC Directive
- EN 61326-1:2006
- EN 61000-3-2: 2006+A2: 2009
- EN 61000-3-3: 2008
x
Safety Symbols
Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument.
Electric Shock hazard
Alternating current (AC)
Chassis (earth ground) symbol.
Ground terminal
On (Power). This is the In position of the power switch when instrument is ON.
Off (Power). This is the Out position of the power switch when instrument is OFF.
Off (Supply). This is the AC mains connect/disconnect switch on top of the instrument.
CAUTION indicates a hazardous situation which, if not avoided, will result in minor or moderate injury
WARNING indicates a hazardous situation which, if not avoided, could result in death or serious injury
DANGER indicates a hazardous situation which, if not avoided, will result in death or serious injury.
xi
Table of Contents
Safety Summary ........................................................................................i
Compliance Statements ........................................................................ viii
Safety Symbols ......................................................................................... x
1 General Information .........................................................................1
1.1 Product Overview ......................................................................... 1
1.2 Package Contents ......................................................................... 1
1.3 Front Panel ................................................................................... 2
Front Panel Description ........................................................................ 2
1.4 Back Panel .................................................................................... 3
Back Panel Description ......................................................................... 4
1.5 Display Information ...................................................................... 4
User Interface Description ................................................................... 5
2 Getting Started .................................................................................6
2.1 Input Power Requirements .......................................................... 6
Input Power .......................................................................................... 6
2.2 Preliminary Check ......................................................................... 6
Verify AC Input Voltage ........................................................................ 7
Connect Power ..................................................................................... 7
Self Test ................................................................................................ 7
Self Cal .................................................................................................. 7
Check Model and Firmware Version .................................................... 8
Function Check ..................................................................................... 8
Probe Safety ....................................................................................... 10
Probe Attenuation .............................................................................. 11
Probe Compensation .......................................................................... 11
3 Functions and Operating Descriptions ............................................ 13
3.1 Menu and Control Button .......................................................... 14
3.2 Connectors ................................................................................. 16
3.3 Auto Setup .................................................................................. 17
xii
3.4 Default Setup .............................................................................. 19
3.5 Universal Knob............................................................................ 20
3.6 Vertical System ........................................................................... 21
Using Vertical Position Knob and Volts/div Knob ............................... 21
3.7 Channel Function Menu ............................................................. 22
Setting up Channels ............................................................................ 25
3.8 Math Functions ........................................................................... 31
FFT Spectrum Analyzer ....................................................................... 32
3.9 Using REF .................................................................................... 39
3.10 Horizontal System....................................................................... 40
Horizontal Control Knob ..................................................................... 41
Window Zone ..................................................................................... 42
3.11 Trigger System ............................................................................ 43
Signal Source ...................................................................................... 44
Trigger Type ........................................................................................ 45
Coupling .............................................................................................. 63
Position ............................................................................................... 63
Slope and Level ................................................................................... 63
Trigger Holdoff ................................................................................... 64
3.12 Signal Acquisition System ........................................................... 65
3.13 Display System ............................................................................ 70
X-Y Format .......................................................................................... 74
3.14 Measure System ......................................................................... 75
Scale Measurement ............................................................................ 75
Cursor Measurement ......................................................................... 75
3.15 Storage System ........................................................................... 88
Recalling Files ..................................................................................... 90
Creating Folders and Files .................................................................. 90
Save/Recall Setup ............................................................................... 91
Save/Recall Waveform ....................................................................... 95
3.16 Utility System............................................................................ 102
System Status ................................................................................... 106
Language .......................................................................................... 107
Self Calibration ................................................................................. 107
Self Test ............................................................................................ 108
Update Firmware.............................................................................. 111
xiii
Pass/Fail ............................................................................................ 112
Waveform Record ............................................................................ 117
Recorder (Scan Mode Only) ............................................................. 121
Help Menu ........................................................................................ 124
4 Application Examples ................................................................... 125
4.1 Taking Simple Measurements .................................................. 125
4.2 Taking Cursor Measurements................................................... 126
4.3 Capturing a Single-Shot Signal .................................................. 128
4.4 Analyzing Signal Details ............................................................ 129
4.5 Triggering on a Video Signal ..................................................... 130
4.6 Application of X-Y Function ...................................................... 131
4.7 Analyzing a Differential Communication Signal........................ 133
5 Remote Control ............................................................................ 134
6 Message Prompts and Troubleshooting ........................................ 135
6.1 Message Prompts ..................................................................... 135
6.2 Troubleshooting ....................................................................... 136
7 Specifications ............................................................................... 138
8 Calibration .................................................................................... 142
SERVICE INFORMATION ....................................................................... 143
LIMITED ONE-YEAR WARRANTY ........................................................... 144
1
1 General Information
1.1 Product Overview
The B&K Precision 2190D digital storage oscilloscope (DSO) is a portable benchtop instrument used for making measurements of signals and waveforms. The oscilloscope’s bandwidth is capable of capturing up to 100 MHz signals with a real time sampling rate of up to 1 GSa/s. With up to 40k points of deep memory, it allows for capturing more details of a signal for analysis and displayed on the large color LCD display.
Features:
2 Channels, Bandwidth: 100 MHz Single channel real-time sampling rate of up to 1GSa/s Up to 40k points of memory depth 7” Color TFT LCD display Trigger types: Edge, Pulse, Video, Slope and Alternative Unique Digital Filter function and Waveform recorder function Auto measure 32 parameters (Voltage and Time) Standard interface: USB Host, USB Device, RS-232, Pass/Fail Out
1.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every instrument is shipped with the following contents:
1 x 2190D 100 MHz Digital Storage Oscilloscope 1 x User Manual (printed) 1 x AC power cord 1 x USB type A to Type B cable
2
2 x 1:1/10:1 Passive Oscilloscope Probes
Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.
1.3 Front Panel
It is important for you to familiarize yourself with the DSO’s front panel before operating the instrument. Below is a brief introduction of the front panel function operation.
Figure 1.1 – Front Panel
Front Panel Description
Front USB (Type A) Connector
Menu function keys, Menu On/Off, Print keys
Input Channels (1 MΩ BNC)
5
10
1
2
3
3
Probe Compensator (1 kHz and Ground)
Horizontal Controls (Time Base)
Trigger Controls
Auto Function Button
Menu and Measurement Buttons
Universal Knob
Vertical Controls
1.4 Back Panel
The following images show back and side panel connection locations.
Figure 1.2 – Back Panel
4
5
6
6
10
9 4 5 7 8
4
Back Panel Description
Security Lock Receptacle
Pass/Fail Output
RS-232 Connector
Rear USB (Type B) Device Connector
Power Input Connector
AC Power Switch
1.5 Display Information
Figure 1.3 – Display Screen
13
11
10
7
3
12
1
1 2 3 4 5
6
5
User Interface Description
Trigger Level Display Marker
Vertical Display Markers (Ground Reference)
Channel Source, Coupling Type, Volts/Division, BW Limit Indicator
Timebase Setting
Horizontal Trigger Position Display
Frequency Counter
Trigger Source, Type, and Level Indicator
Rear USB Indicator
Print Key Save Function Indicator
Waveform Display Preview
Horizontal Trigger Position Marker
USB Flash Drive Indicator
Trigger Status
13
1 2 3 4 5
6
7
8
9
10
11
12
6
2 Getting Started
Before connecting and powering up the instrument, please review and go through the instructions in this chapter.
2.1 Input Power Requirements
Input Power
The supply has a universal AC input that accepts line voltage and frequency input within:
100 – 240 V (+/- 10%), 50 /60 Hz (+/- 5%)
100 – 127 V, 45 – 440 Hz
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sufficient circuit capacity for the power supply. Once verified, connect the cable firmly.
The included AC power cord is safety certified for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
2.2 Preliminary Check
Complete the following steps to verify that the oscilloscope is ready for use.
7
Verify AC Input Voltage
Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specification as explained in section 2.1.
Connect Power
Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the ON position to turn ON the instrument. The instrument will have a boot screen while loading, after which the main screen will be displayed.
Self Test
The instrument has 3 self-test options to test the screen, keys, and the LED back lights of the function, menu, and channel keys as shown below.
Figure 2.1 – Self Test Menu
To perform the self test, please refer to the Self Test section for further instructions.
Self Cal
This option runs an internal self calibration procedure that will
check and adjust the instrument. To perform the self calibration, please refer to the Self Calibration section for further instructions.
8
Check Model and Firmware Version
The model and firmware version can be verified from within the menu system.
Press Utility and select System Status option. The software/firmware version, hardware version, model, and serial number will be displayed. Press the Single key to exit.
Function Check
Follow the steps below to do a quick check of the oscilloscope’s functionality.
1. Power on the oscilloscope. Press “DEFAULT SETUP” to show the
result of the self check. The probe default attenuation is 1X.
Figure 2.2 – Scope Layout
2. Set the switch to 1X on the probe and connect the probe to
channel 1 on the oscilloscope. To do this, align the slot in the probe connector with the key on the CH 1 BNC, push to connect, and twist to the right to lock the probe in place. Connect the probe tip and reference lead to the PROBE COMP connectors.
9
Figure 2.3 – Probe Compensation
3. Press “AUTO” to show the 1 kHz frequency and about 3V peak-
peak square wave in couple seconds.
Figure 2.4 – 3 Vpp Square Wave
4. Press “CH1” two times to turn off channel 1, Press“CH2” to
change screen into channel 2, reset the channel 2 as step 2 and step 3.
PROBE
COMP
CH1
10
Probe Safety
A guard around the probe body provides a finger barrier for protection from electric shock.
Figure 2.5 – Oscilloscope Probe
Connect the probe to the oscilloscope and connect the ground terminal to ground before you take any measurements.
SHOCK HAZARD
To avoid electric shock when using the probe, keep fingers behind the guard on the probe body.
To avoid electric shock while using the probe, do not touch metallic portions of the probe head while it is connected to a voltage source. Connect the probe to the oscilloscope and connect the ground terminal to ground before you take any measurements.
11
Probe Attenuation
Probes are available with various attenuation factors which affect the vertical scale of the signal. The Probe Check function verifies that the Probe attenuation option matches the attenuation of the probe.
You can push a vertical menu button (such as the CH 1 MENU button), and select the Probe option that matches the attenuation factor of your probe.
NOTE: The default setting for the Probe option is 1X.
Be sure that the attenuation switch on the probe matches the Probe option in the oscilloscope. Switch settings are 1X and 10X.
NOTE: When the attenuation switch is set to 1X, the probe limits
the bandwidth of the oscilloscope to 6 MHz (according to Probe spec). To use the full bandwidth of the oscilloscope, be sure to set
the switch to 10X
Probe Compensation
As an alternative method to Probe Check, you can manually perform this adjustment to match your probe to the input channel.
12
Figure 2.6 – Probe Compensation Setup
1. Set the Probe option attenuation in the channel menu to 10X. Do
so by pressing CH1 button and selecting “Probe” from menu.
Select 10X. Set the switch to 10X on the probe and connect the probe to channel 1 on the oscilloscope. If you use the probe hook-tip, ensure a proper connection by firmly inserting the tip onto the probe.
2. Attach the probe tip to the PROBE COMP 3V connector and the
reference lead to the PROBE COMP Ground connector. Display the channel and then push the “AUTO” button.
3. Check the shape of the displayed waveform.
Figure 2.7 – Compensation Illustration
4. If necessary, adjust your probes compensation trimmer pot.
Repeat as necessary.
AUTO BUTTO
PROBE
COMP
CH1
Overcompensated
Compensated Correctly
Undercompensated
13
3 Functions and Operating
Descriptions
To use your oscilloscope effectively, you need to learn about the following oscilloscope functions:
Menu and control button
Connector
Auto Setup
Default Setup
Universal knob
Vertical System
Channel Function Menu
Math Functions
Using REF
Horizontal System
Trigger System
Acquiring signals System
Display System
Measuring waveforms System
Utility System
Storage System
Online Help function
14
3.1 Menu and Control Button
Figure 3.1 – Control Buttons
Channel buttons (CH1, CH2): Press a channel button to turn
that channel ON or OFF and open the channel menu for that channel. You can use the channel menu to set up a channel. When the channel is on, the channel button is lit.
MATH: Press to display the Math menu. You can use the
MATH menu to use the oscilloscopes Math functions.
REF: Press to display the Ref Wave menu. You can use this
menu to save and recall four or two reference waveforms to and from internal memory.
15
HORI MENU: Press to display the Horizontal menu. You can
use the Horizontal menu to display the waveform and zoom in a segment of a waveform.
TRIG MENU: Press to display the Trigger menu. You can use
the Trigger menu to set the trigger type (Edge. Pulse, Video, Slope, Alternative) and trigger settings.
SET TO 50%: Press to stabilize a waveform quickly. The
oscilloscope can set the trigger level to be halfway between the minimum and maximum voltage level automatically. This is useful when you connect a signal to the EXT TRIG connector and set the trigger source to Ext or Ext/5.
FORCE: Use the FORCE button to complete the current
waveform acquisition whether the oscilloscope detects a trigger or not. This is useful for Single acquisitions and Normal trigger mode.
SAVE/RECALL: Press to display the Save/Recall menu. You
can use the Save/Recall menu to save and recall up to 20 oscilloscope setups and 10 waveforms to/from internal memory or a USB memory device (limited by memory capacity of the USB flash drive). You can also use it to recall the default factory settings, to save waveform data as a comma-delimited file (.CSV), and to save the displayed waveform image.
ACQUIRE: Press to display Acquire menu. You can use the
Acquire menu to set the acquisition Sampling Mode (Sampling, Peak Detect, and Average).
MEASURE: Press to display a menu of measurement
parameters.
CURSORS: Display the Cursor Menu. Vertical Position
controls adjust cursor position while displaying the Cursor Menu and the cursors are activated. Cursors remain
displayed (unless the “Type” option is set to “Off”) after
leaving the Cursor Menu but are not adjustable.
DISPLAY: Press to open the Display menu. You can use the
Display menu to set grid and waveform display styles, and persistence.
UTILITY: Press to open the Utility menu. You can use the
Utility menu to configure oscilloscope features, such as
16
sound, language, counter, etc. You can also view system status and update software.
DEFAULT SETUP: Press to reset the oscilloscope’s settings to
the default factory configuration.
HELP: Enter the online help system. AUTO: Automatically sets the oscilloscope controls to
produce a usable display of the input signals.
RUN/STOP: Continuously acquires waveforms or stops the
acquisition.
Note: If waveform acquisition is stopped (using the
RUN/STOP or SINGLE button), the TIME/DIV control expands or compresses the waveform.
SINGLE: Acquire a single waveform and then stops waveform
acquisition.
3.2 Connectors
Figure 3.2 – Connectors
Channel Connector (CH1, CH2): Input connectors for waveform
display.
EXT TRIG: Input connector for an external trigger source. Use the
Trigger Menu to select the “Ext” or “Ext/5” trigger source.
Probe Compensation: 1 kHz voltage probe compensation output
and ground. Use to electrically match the probe to the oscilloscope input circuit.
17
3.3 Auto Setup
The 2190D Digital Storage Oscilloscope has an Auto Setup function that identifies the waveform types and automatically adjusts controls to produce a usable display of the input signal.
Press the AUTO front panel button, and then press the menu option button adjacent to the desired waveform as follows:
Figure 3.3 – Auto Setup
Table 3.1 – Autoset Menu
Option
Description
(Multi-cycle sine)
Auto set the screen and display several cycles signal.
(Single-cycle sine)
Set the screen and auto display single cycle signal.
(Rising edge)
Auto set and show the rising time.
18
(Falling edge)
Auto set and show the falling time.
(Undo Setup)
Causes the oscilloscope to recall the previous setup.
Auto Set determines the trigger source based on the following conditions:
If multiple channels have signals, channel with the lowest
frequency signal.
No signals found, the lowest-numbered channel displayed when
Auto set was invoked.
No signals found and no channels displayed, oscilloscope displays
and uses channel 1.
Table 3.2 – Auto Set Function Menu Items
Function
Setting
Acquire Mode
Adjusted to Sampling
Display Format
Y-T
Display Type
Set to Dots for a video signal, set to Vectors for an FFT spectrum; otherwise, unchanged
Vertical Coupling
Adjusted to DC or AC according to the input signal
Bandwidth Limit
Off(full)
V/div
Adjusted
VOLTS/DIV adjustability
Coarse
19
Signal inverted
Off
Horizontal position
Center
Time/div
Adjusted
Trigger type
Edge
Trigger source
Auto detect the channel which has the input signal
Trigger slope
Rising
Trigger mode
Auto
Trigger coupling
DC
Trigger holdoff
Minimum
Trigger level
Set to 50%
NOTE: The AUTO button can be disabled. Please see “Appendix C Disabling Auto Function” for details.
3.4 Default Setup
The oscilloscope is set up for normal operation when it is shipped from the factory. This is the default setup. To recall this setup, press the DEFAULT SETUP button. For the default options, buttons and controls when you press the DEFAULT SETUP button, refer to “Appendix A Default Setup”.
The DEFAULT SETUP button does not reset the following settings:
Language option
Saved reference waveform files
Saved setup files
Display contrast
20
Calibration data
3.5 Universal Knob
Figure 3.4 – Universal Knob
You can use the Universal knob with many functions, such as adjusting the hold off time, moving cursors, setting the pulse width, setting the video line, adjusting the upper and lower frequency limit, adjusting the X and Y masks when using the pass/fail function, etc. You can also turn the
“Universal” knob to adjust the storage position of setups, waveforms,
pictures when saving/recalling, and to select menu options. With some functions, the light indicator above the knob will turn on to indicate that the knob can be used to make changes or adjustments for that function. The knob can also be pushed to make a selection after changes/adjustments have been made.
21
3.6 Vertical System
The vertical control could be used for displaying waveform, rectify scale and position.
Figure 3.5 – Vertical System Controls
Using Vertical Position Knob and Volts/div Knob
Vertical “POSITION” Knob
1. Use the Vertical “POSITION” knobs to move the channel
waveforms up or down on the screen. This button’s
resolution varies as per the vertical scale.
2. When you adjust the vertical position of channels
waveforms, the vertical position information will display on the bottom left of the screen. For example “Volts Pos=24.6mV”.
3. Press the vertical “POSITION” knob to set the vertical
position to zero.
“Volts/div” Knob
1. Use the “Volts/div” knobs to control how the
oscilloscope amplifies or attenuates the source signal of
22
channel waveforms. When you turn the “volts/div”
knob, the oscilloscope increases or decreases the vertical size of the waveform on the screen with respect to the ground level.
2. When you press the “Volt/div” knob, you can switch
“Volts/div” option between “Coarse” and “Fine”. The vertical scale is set by a 1-2-5 step sequence in Coarse mode. Increase in the clockwise direction and decrease in the counterclockwise direction. In Fine mode, the knob changes the Volts/Div scale in small steps between the coarse settings. Again, increase in the clockwise direction and decrease in the counterclockwise direction.
3.7 Channel Function Menu
Table 3.3 – Channel Function Menu
Option
Setting
Introduction
Coupling
DC
AC
GND
DC passes both AC and DC components of the input signal.
AC blocks the DC component of the input signal and attenuates signals below 10 Hz.
GND disconnects the input signal.
BW limit
On
Off
Limits the bandwidth to reduce display noise; filters the signal to reduce noise and other unwanted high frequency components.
Volts/Div
Coarse
Fine
Selects the resolution of the Volts/Div knob
Coarse defines a 1-2-5 sequence.
23 Fine changes the resolution to small steps between the coarse settings.
Probe
1X, 5X 10X, 50X 100X, 500X, 1000X
Set to match the type of probe you are using to ensure correct vertical readouts.
Next Page
Page 1/3
Press this button to enter the second page menu
Table 3.4 – Channel Function Menu 2
Option
Setting
Instruction
Invert
on
off
Turn on invert function.
Turn off invert function.
Filter
Press this button to enter the “Digital Filter menu”.
Next Page
Page 2/3
Press this button to enter the third page menu.
Table 3.5 – Channel Function Menu 3
Option
Setting
Introduction
Unit
V A Set the scale unit to voltage
Set the scale unit to current.
Skew
-100 ns – 100ns
Set the skew time between two channels.
Next Page
Page 3/3
Press this button to return to the first page menu.
24
Table 3.6 – Digital Filter Menu
Option
Setting
Introduction
Digital Filter
On
Off
Turn on the digital filter.
Turn off the digital filter.
Type
Setup as LPF (Low Pass Filter).
Setup as HPF (High Pass Filter).
Setup as BPF (Band Pass Filter).
Setup as BRF (Band Reject Filter).
Upper_limit
Turn the “Universal” knob to set
upper limit.
Lower_limit
Turn the “Universal” knob to set
lower limit.
Return
Return to the second page menu.
“GND” Coupling: Use GND coupling to display a zero-volt
waveform. Internally, the channel input is connected to a zero­volt reference level.
Fine Resolution: The vertical scale readout displays the actual
Volts/Div setting while in the fine resolution setting. Changing the setting to coarse does not change the vertical scale until the VOLTS/DIV control is adjusted.
NOTE:
The oscilloscope’s vertical response rolls off slowly above its specified bandwidth. Therefore, the FFT spectrum can show valid frequency information higher than the oscilloscope’s bandwidth. However, the magnitude information near or above the
25
bandwidth will not be accurate.
If the channel is set to DC coupling, you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol.
If the channel is set to AC coupling, the DC component of the signal is blocked allowing you to use greater sensitivity to display the AC component of the signal.
Setting up Channels
Each channel has its own separate Menu. The items are set up separately according to each channel.
1. Setup Channel Coupling
Take CH1 for example; the tested signal is a sine wave signal with DC deflection:
Press “CH1”→“Coupling”→“AC”, Set to AC couple mode.
This will block the DC component of the input signal.
Figure 3.6 – AC Coupling
26
Press “CH1”→“Coupling”→“DC”, Set to DC couple mode.
Both DC and AC components of the input signal will be captured.
Figure 3.7 – DC Coupling
Press “CH1”→“Coupling”→“GND”, Set to GROUND mode.
This disconnects the input signal.
Figure 3.8 – Ground Coupling
2. Bandwidth Limiting
Take CH1 for example:
27
Press “CH1”→“BW Limit”→ “On”, and bandwidth will be
limited to 20 MHz.
Press “CH1”→“BW Limit”→ “Off”, and bandwidth limit
will be disabled.
Figure 3.9 – Bandwidth Limit
3. Volts/Div Settings
Vertical scale adjust have Coarse and Fine modes, Vertical sensitivity range of 2 mV/div – 10 V/div. Take CH1 for example:
Press “CH1”→“Volts/Div”→“Coarse”. It is the default
setting of Volts/Div, and makes the vertical scaling in a 1­2-5-step sequence from 2 mV/div, 5 mV/div, 10 mV/div to 10 V/div.
Press “CH1”→ Volts/Div”→ Fine”. This setting changes
the vertical control to small steps between the coarse settings. It will be helpful when you need to adjust the waveform vertical size in smaller steps.
28
Figure 3.10 – Coarse/Fine Control
4. Setting Probe Attenuation
In order to set the attenuation coefficient, you need to specify it in the channel operation Menu. If the attenuation coefficient is 10:1, the input coefficient should be set to 10X, so that the Volts/div information and measurement testing is correct. Take CH1 for example, when you use the 100:1 probe:
Press “CH1”→“Probe” →“100X”
Figure 3.11 – Probe Attenuation Setting
29
5. Inverting Waveforms
Take CH1 for example:
Press “CH1”→ Next Page “Page 1/3” →“Invert”→“On”:
Figure 3.12 – Invert Waveform Screen
6. Using the Digital Filter
Press “CH1”→ Next Page “Page 1/3”→ “Filter”, display
the digital filter menu. Select “Filter Type”, then select “Upper Limit” or “Lower Limit” and turn the “Universal”
knob to adjust them.
Press “CH1”→ Next Page “Page 1/3”→ “Filter” →“Off”.
Turn off the Digital Filter function.
30
Figure 3.13 Digital Filter Menu
Press “CH1”→ “Next Page “Page 1/3”→ “Filter” → “On”.
Turn on the Digital Filter function.
Figure 3.14 – Digital Filter Adjustment Screen
31
3.8 Math Functions
Math shows the results after +,-,*, / and FFT operations of the CH1 and CH2. Press the MATH button to display the waveform math operations. Press the MATH button again to remove the math waveform display.
Table 3.7 – Math Function Menu
Function
Setting
Description
Operation
+, -, *, /, FFT
Math operates between signal source CH1 and CH2.
Source A
CH1 – CH2
Select CH1 or CH2 as Source A.
Source B
CH1 – CH2
Select CH1 or CH2 as Source B.
Invert
on
off
Invert the MATH waveform.
Turn off MATH Invert function.
Next Page
Page 1/2
Enter the second page of MATH menu.
Table 3.8 – Math Function Menu 2
Function
Setting
Description
Use universal knob to adjust the vertical position of the MATH waveform.
Use universal knob to adjust the vertical scale of the MATH waveform.
Next Page
Page 2/2
Go back to first page of MATH menu.
32
Table 3.9 – Math Function Description
Operation
Setting
Description
A+B
Source A waveform adds Source B waveform.
A-B
Source B waveform is subtracted from Source A waveform.
*
A*B
Source A multiplied by Source B
/
A/B
Source A divided by Source B
FFT
Fast Fourier Transform.
Figure 3.15 – Math Waveform
FFT Spectrum Analyzer
The FFT process mathematically converts a time-domain signal into its frequency components. You can use the Math FFT mode to view the following types of signals:
Analyze the harmonic wave in the Power cable.
33
Test the harmonic content and distortion in the system
Show the Noise in the DC Power supply
Test the filter and pulse response in the system
Analyze vibration
Table 3.10 – FFT Function Menu 1
FFT Option
Setting
Description
Source
CH1, CH2
Select this channel as the FFT source.
Window
Hanning
Hamming
Rectangular
Blackman
Select FFT window types.
FFT ZOOM
1X
2X
5X
10X
Changes the horizontal magnification of the FFT display.
Next Page
Page 1/2
Enter the second page of FFT menu.
Table 3.11 – FFT Function Menu 2
FFT Option
Setting
Description
Scale
Vrms
Set Vrms to be the Vertical Scale unit.
dBVrms
Set dBVrms to be the vertical
34
Scale unit.
Display
Split
Full screen
Display FFT waveform on half screen.
Display FFT waveform on full screen.
Next Page
Page 2/2
Return to the first page of FFT menu.
To use the Math FFT mode, you need to perform the following steps:
1. Set up the source (time-domain) waveform.
2. Press the AUTO button to display an YT waveform.
3. Turn the vertical “POSITION” knob to move the YT waveform to
the center vertically (zero divisions).
4. Turn the horizontal “POSITION” knob to position the part of the
YT waveform that you want to analyze in the center eight divisions of the screen. The oscilloscope calculates the FFT spectrum using the center 1024 points of the time-domain waveform.
5. Turn the “Volts/div” knob to ensure that the entire waveform
remains on the screen.
6. Turn the “S/div” knob to provide the resolution you want in the
FFT spectrum.
7. If possible, set the oscilloscope to display many signal cycles.
To display FFT correctly, follow these steps:
1. Push the “MATH” button.
2. Set the “Operation” option to FFT.
3. Press the “Source” button to select “CH1” or “CH2according to
input signal channel.
4. Turn the “Time/div” knob to adjust the sampling rate (this
parameter is displayed behind the time base parameter), making sure it is at least double the input signal frequency. (to avoid aliasing according to Nyquist’s theorem)
35
Displaying the FFT Spectrum
Press the MATH button to display the Math Menu. Use the options to select the Source channel, Window algorithm, and FFT Zoom factor. You
can display only one FFT spectrum at a time. You can select “Full screen” or “Split” in “Display” option to display FFT waveform on full screen or
display channel waveform and its FFT waveform on half screen at a time.
Select FFT window
Windows reduce spectral leakage in the FFT spectrum. The FFT assumes that the YT waveform repeats forever. With an integral number of cycles, the YT waveform starts and ends at the me amplitude and there are no discontinuities in the signal shape A non-integral number of cycles in the YT waveform causes the signal start and end points to be at different amplitudes. The transitions between the start and end points cause discontinuities in the signal that introduce high-frequency transients.
Table 3.12 – FFT Window Description
Window
Characteristics
Applications
Rectangular
Best frequency resolution, worst magnitude resolution. This is essentially the
Symmetric transients or bursts. Equal­amplitude sine waves with fixed frequencies.
36
same as no window.
Broadband random noise with a relatively slowly varying spectrum.
Hanning Hamming
Better frequency, poorer magnitude accuracy than Rectangular. Hamming has slightly better frequency resolution than Hanning.
Sine, periodic, and narrow-band random noise. Asymmetric transients or bursts.
Blackman
Best magnitude, worst frequency resolution.
Single frequency waveforms, to find higher order harmonics.
Magnifying the FFT Spectrum
You can magnify and use cursors to take measurements on the FFT
spectrum. The oscilloscope includes an “FFT Zoom” option to magnify horizontally, press this option button to select “1X”, “2X”, “5X” or “10X”. Moreover, you also can turn the “Universal” knob to magnify FFT
waveform horizontally in a 1-2-5 step. To magnify vertically, turn the “Volts/div” knob.
Measuring an FFT Spectrum Using Cursors
You can take two measurements on FFT spectrums: magnitude (in dB) and frequency (in Hz). Magnitude is referenced to 0 dB, where 0 dB equals 1 VRMS. You can use the cursors to take measurements at any zoom factor. Use horizontal cursors to measure amplitude and vertical cursors to measure frequency.
If you input a sine signal into channel 1, follow these steps:
Measure FFT Amplitude
1. Input a sine signal to channel 1, and press the
“AUTO” button.
37
2. Press the “MATH” button to enter the “MATH”
menu.
3. Press the “Operation” option button to select “FFT”.
4. Press the “Source” option button to select “CH1”.
5. Press CH1 button to display CH1 menu.
6. Turn the “Time/div” knob to adjust the sampling
rate (at least double the frequency of input signal).
7. If FFT is displayed on full screen, press CH1 button
again to remove channel waveform display.
8. Press the “CURSOR” button to enter “Cursor” menu.
9. Press the “Cursor Mode” button to select “Manual”.
10. Press the “Type” option button to select “Voltage”.
11. Press the “Source” option button to select “MATH”.
12. Press the “CurA” option button; turn the “Universal”
knob to move Cursor A to the highest point of the FFT waveform.
13. Press the “CurB” option button, turn the “Universal”
knob to move Cursor B to the lowest point of the FFT waveform.
14. The amplitude (T) displays on the top of the left
screen.
Figure 3.16 Measuring FFT Amplitude
Measure FFT Frequency
1. Press the CURSOR button.
38
2. Press the “Cursor Mode” button to select “Manual”.
3. Press the “Type” option button to select “Time”.
4. Press the “Source” option button to select “MATH”.
5. Press the “CurA” option button, turn the “Universal”
button to move Cursor A to the highest position of the FFT waveform.
6. The value of CurA displaying on the top of the left
screen is FFT frequency. This frequency should be the same as input signal frequency.
Figure 3.17 – Measuring FFT Frequency
NOTE:
The FFT of a waveform that has a DC component or offset can cause incorrect FFT waveform magnitude values. To minimize the DC component, choose AC Coupling on the source waveform.
To display FFT waveforms with a large dynamic range, use the dBVrms scale. The dBVrms scale displays component magnitudes using a log scale.
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. This frequency is half that of the sample rate provided it is within the analog bandwidth of the oscilloscope. Frequencies above the Nyquist frequency will be undersampled, which causes aliasing.
39
3.9 Using REF
The reference control saves waveforms to a nonvolatile waveform memory. The reference function becomes available after a waveform has been saved.
Table 3.13 – REF Function Menu
Option
Setting
Description
Source
CH1,CH2,
CH1 off
CH2 off
Choose the waveform display to store.
REFA
REFB
Choose the reference location to store or recall a waveform.
Save
Stores source waveform to the chosen reference location.
REFA
REFB
on
off
Recall the reference waveform on the screen.
Turn off the reference waveform.
Press the Ref button to display the “Reference waveform menu”.
40
Figure 3.18 – Reference Waveform Menu
Operation step:
1. Press the “REF” menu button to display the “Reference waveform
menu”.
2. Press the “Source” option button to select input signal channel.
3. Turn the vertical “POSITION” knob and “Volts/div” knob to adjust
the vertical position and scale.
4. Press the third option button to select “REFA” or “REFB” as
storage position.
5. Press the “Save” option button.
6. Press the bottom option button to select “REFA On” or “REFB On”
to recall the reference waveform.
NOTE:
X-Y mode waveforms are not stored as reference waveforms.
You cannot adjust the horizontal position and scale of the reference waveform.
3.10 Horizontal System
Shown below are two knobs and one button in the HORIZONTAL area.
41
Figure 3.19 – Horizontal Controls
Table 3.14 – Horizontal System Function Menu
Option
Setting
Description
Delayed
On
Off
Turn on this function for main timebase waveform to display on the top half screen and window timebase waveform to display on the below half screen at the same time.
Turn off this function to only display main timebase waveform on the screen.
Horizontal Control Knob
You can use the horizontal controls to change the horizontal scale and position of waveforms. The horizontal position readout shows the time represented by the center of the screen, using the time of the trigger as zero. Changing the horizontal scale causes the waveform to expand or contract around the screen center.
Horizontal “POSITION” knob
1. Adjust the horizontal position of all channels and
math waveforms (the position of the trigger relative
42
to the center of the screen). The resolution of this control varies with the time base setting.
2. When you press the horizontal “POSITION” Knob,
you can set the horizontal position to zero.
“Time/div” knob
1. Used to change the horizontal time scale to magnify
or compress the waveform. If waveform acquisition is stopped (using the RUN/STOP or SINGLE button), turn the Time/div knob to expand or compress the waveform.
2. Select the horizontal Time/div (scale factor) for the
main or the window time base. When Window Zone is enabled, it changes the width of the window zone by changing the window time base.
Display scan mode
When the Time/div control is set to 100 ms/div or slower and the trigger mode is set to Auto, the oscilloscope enters the scan acquisition mode. In this mode, the waveform display updates from left to right. There is no trigger or horizontal position control of waveforms during scan mode.
Window Zone
Use the Window Zone option to define a segment of a waveform to see more detail. This function behaves like zooming into a portion of the captured waveform. The window time base setting cannot be set slower than the Main time base setting.
You can turn the Horizontal Position and Time/div controls to enlarge or minimize waveforms in the Window Zone.
If you want to see a section of the waveform in details, follow these steps:
1. Press the “HORI MENU” button to enter the “Horizontal menu”.
2. Turn the “Time/div” knob to change the main timebase scale.
3. Press the “Delayed” option button to select “On”.
43
Figure 3.20 – Horizontal Delay Menu
4. Turn the “Horizontal Position” knob (adjust window’s position) to
select the window that your need and expanded window waveform display on the below half screen at the same time.
3.11 Trigger System
The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, the oscilloscope converts unstable displays or blank screens into meaningful waveforms.
Here are three buttons and one knob in the trigger area. See below:
44
Figure 3.21 Trigger Controls
“TRIG MENU” Button: Press the “TRIG MENU” Button to
display the “Trigger Menu”.
“LEVEL” Knob: The LEVEL knob is to set the
corresponding signal voltage of trigger point in order to
sample. Press the “LEVEL” knob to set trigger level to
zero.
“SET TO 50%” Button: Use the “SET TO 50%” button to
stabilize a waveform quickly. The oscilloscope can set the Trigger Level to be about halfway between the minimum and maximum voltage levels automatically. This is useful when you connect a signal to the EXT TRIG BNC and set the trigger source to Ext or Ext/5.
“FORCE Button: Use the FORCE button to complete the
current waveform acquisition whether the oscilloscope detects a trigger or not. This is useful for SINGLE acquisitions and Normal trigger mode.
Signal Source
You can use the Trigger Source options to select the signal that the oscilloscope uses as a trigger. The source can be any signal connected to a channel BNC, to the EXT TRIG BUS, or the AC power line (available only with Edge Trigger).
45
Trigger Type
The scopes have five trigger types: Edge, Video, Pulse, Slope, and Alternate.
Edge Trigger
Use Edge triggering to trigger on the edge of the oscilloscope input signal at the trigger threshold.
Table 3.15 – Edge Trigger Function Menu
Option
Setting
Description
Type
Edge
With Edge highlighted, the rising or falling edge of the input signal is used for the trigger.
Source
CH1
CH2
Triggers on a channel whether or not the waveform is displayed.
EXT
Does not display the trigger signal; the Ext option uses the signal connected to the EXT TRIG front-panel BNC and allows a trigger level range of -1.2V to +1.2V.
EXT/5
Same as Ext option, but attenuates the signal by a factor of five, and allows a trigger level range of +6V to -6V.This extends the trigger level range.
AC Line
This selection uses a signal derived from the power line as the trigger source; trigger coupling is set to DC and the trigger level to 0 volts.
Slope
Trigger on Rising edge of the trigger signal.
Trigger on Falling edge of the trigger signal.
Trigger on Rising edge and Falling edge of the trigger signal.
46
Mode
Auto
Use this mode to let the acquisition free-run in the absence of a valid trigger; This mode allows an untriggered, scanning waveform at 100 ms/div or slower time base settings.
Normal
Use this mode when you want to see only valid triggered waveforms; when you use this mode, the oscilloscope does not display a waveform until after the first trigger.
Single
When you want the oscilloscope to acquire a single waveform, press the “SINGLE” button.
Set up
Enter the “Trigger Setup Menu” (See Table 3.16).
Table 3.16 – Trigger Setup Function Menu
Option
Setting
Explain
Coupling
DC
Passes all components of the signal
AC
Blocks DC components, attenuates signals below 50 Hz.
HF Reject
Attenuates the high-frequency components above 150 kHz.
LF Reject
Blocks the DC component, attenuates the low-frequency components below 7 kHz.
Holdoff
Using the “universal” knob to adjust
holdoff time (sec), the holdoff value is displayed.
Holdoff
Reset
Reset holdoff time to 100ns.
47 Return
Return to the first page of “Trigger main menu”.
Figure 3.22 – Trigger Menu Screen
Operating Instructions:
1. Setup Type
Press the “TRIG MENU” button to display “Trigger” menu. Press the “Type” option button to select “Edge”.
2. Set up Source
According to the input signal, press the “Source” option
button to select “CH1”, “CH2”, “EXT”, “EXT/5” or “AC Line”.
3. Set up Slope
Press the “Slope” option button to select “ ”,
or ”.
4. Set up Trigger mode
Press the “Trigger mode” option button to select “Auto”,
“Normal”, “Single”.
Auto: The waveform refreshes at a high speed whether the trigger condition is satisfied or not.
48
Normal: The waveform refreshes when the trigger condition is satisfied and waits for next trigger event occurring when the trigger condition is not satisfied.
Single: The oscilloscope acquires a waveform when the trigger condition is satisfied and then stops
5. Set up Trigger coupling
Press the “Set Up” button to enter the “Trigger Setup
Menu”.
Press the “Coupling” option button to select “DC”, “AC”,
“HF Reject” or “LF Reject”.
Pulse Trigger
Use Pulse Width triggering to trigger on aberrant pulses.
Table 3.17 – Pulse Trigger Function Menu 1
Option
Setting
Description
Type
Pulse
Select the pulse to trigger the pulse match the trigger condition.
Source
CH1 CH2 EXT EXT/5 AC Line
Select input signal source.
49
When
(Positive pulse width less
than pulse width setting)
(Positive pulse width larger
than pulse width setting)
(Positive pulse width equal
to pulse width setting)
(Negative pulse width less
than pulse width setting)
(Negative pulse width
larger than pulse width setting)
(Negative pulse width
equal to pulse width setting)
Select how to compare the trigger pulse relative to the value selected in the Set Pulse Width option.
Set Width
20.0ns10.0s
Selecting this option can turn the universal to set up the pulse width.
Next Page
Page 1/2
Press this button to enter the second page.
Figure 3.23 – Pulse Trigger Menu 1
50
Table 3.18 – Pulse Trigger Function Menu 2
Option
Setting
Description
Type
Pulse
Select the pulse to trigger the pulse match the trigger condition.
Mode
Auto Normal single
Select the type of triggering; Normal mode is best for most Pulse Width trigger applications.
Set up
Enter the “Trigger setup menu”.
Next Page
Page 2/2
Press this button to return to the first page.
Figure 3.24 – Pulse Trigger Menu 2
Operating Instructions:
1. Setup Type
Press the “TRIG MENU” button to display “Trigger” menu. Press the “Type” option button to select “Pulse”.
2. Set up condition
Press the “When” option button to select “ ”, “
”, “ ”, “ ”, “ ”or“ ”.
51
3. Set up pulse width
Turn the “Universal” knob to set up width.
Video Trigger
Trigger on fields or lines of standard video signals.
Table 3.19 – Video Trigger Function Menu 1
Option
Setting
Description
Type
Video
When you select the video type, put the couple set to the AC, then you could trigger the NTSC,PAL and SECAM video signal.
Source CH1 CH2
Select the input source to be the trigger signal.
EXT EXT/5
Ext and Ext/5 use the signal applied to the EXT TRIG connector as the source.
Polarity
(Normal)
Normal triggers on the negative edge of the sync pulse.
(Inverted)
Inverted triggers on the positive edge of the sync pulse.
Sync
Line Num All lines Odd field Even Field
Select appropriate video sync.
Next Page
Page 1/2
Enter the second page of “Video trigger menu”.
Table 3.20 - Video Trigger Function Menu 2
Option
Setting
Description
Type
Video
When you select the video type, put the couple set to the
52 AC, then you could trigger the
NTSC, PAL and SECAM video signal.
Standard
NTSC Pal/Secam
Select the video standard for sync and line number count.
Mode
Auto
Use this mode to let the acquisition free-run in the absence of a valid trigger; This mode allows an untriggered, scanning waveform at 100 ms/div or slower time base settings.
Normal
Use this mode when you want to see only valid triggered waveforms; when you use this mode, the oscilloscope does not display a waveform until after the first trigger.
Single
When you want the oscilloscope to acquire a single waveform, press the “SINGLE” button.
Set up
Enter the “Trigger setup menu”.
Next Page
Page 2/2
Return to the first page of “Video Trigger menu”.
53
+-
Figure 3.25 – Video Trigger Menu
Operating Instructions:
1. Set up Type
Press the “TRIG MENU” button to display “Trigger” menu. Press the “Type” option button to select “Video”.
2. Set up Polarity
Press the “Polarity” option button to select “ ” or “
”.
3. Set up Synchronization
Press the “Sync” option button to select “All Lines”, “Line
Num”, “Odd Field”, and “Even Field”.
If you select “Line Num”, you can turn the “Universal”
knob to set the appointed line number.
4. Set up Standard
Press the “Next Page - Page 2/2” option Press the “Standard” option button to select
“PAL/SECAM” or “NTSC”.
54
Slope Trigger
Trigger on positive slope or negative slope of waveform, according to setup time of the oscilloscope.
Table 3.21 – Slope Trigger Function Menu 1
Option
Setting
Instruction
Type
Slope
Trigger on positive slope of negative slope according to setup time of the oscilloscope.
Source
CH1
CH2
EXT
EXT/5
Select trigger source. When
Select trigger condition.
Time <Set time>
Turn the “Universal” knob to set
slope time. Time setup range is 20ns-10s.
Next Page
Page 1/2
Enter the second page of slope trigger.
55
Figure 3.26 – Slope Trigger Menu 1
Table 3.22 – Slope Trigger Function Menu 2
Option
Setting
Instruction
Type
Slope
Trigger on positive slope or negative slope according to setup time of the oscilloscope.
Vertical
Select the trigger level that can be
adjusted by “LEVEL” knob. You can adjust “LEVEL A”, “LEVEL B” or adjust
them at the same time.
Mode
Auto
Use this mode to let the acquisition free­run in the absence of a valid trigger; This mode allows an untriggered, scanning waveform at 100 ms/div or slower time base settings.
56 Normal
Use this mode when you want to see only valid triggered waveforms; when you use this mode, the oscilloscope does not display a waveform until after the first trigger.
Single
When you want the oscilloscope to acquire a single waveform, press the “SINGLE” button.
Set up
Enter “Trigger setup menu” (See Table
3.16).
Next Page
Page 2/2
Return to the first page of slope trigger.
Figure 3.27 – Slope Trigger Menu 2
Operating Instructions:
Follow the next steps after “Slope Trigger” is selected:
1. Input a signal to CH1 or CH2.
2. Press the “AUTO” button.
3. Press the “TRIG MENU” button to enter “Trigger menu”.
4. Press the “Type” option button to select “Slope”.
5. Press the “Source” option button to select “CH1” or “CH2”.
57
6. Press the “When” option button to select “ ”, “ ”,
”, “ ”, “ ” and “ ”.
7. Press the “Time” button, turn the “Universal” knob to adjust
slope time.
8. Press the “Next Page - Page 1/2” option button to enter the
second page of the “Slope trigger menu”.
9. Press the “Vertical” option button to select trigger level that can
be adjusted.
10. Turn the “LEVEL” knob.
Alternate Trigger
The trigger signal comes from two vertical channels when you use alternate trigger. In this mode, you can observe two irrelative signals at the same time. You can select different trigger types for two vertical signals, and selected types cover edge, pulse, video and slope trigger. Trigger information of two channel signals display on the bottom right of the screen.
Figure 3.28 – Alternate Trigger Menu
Table 3.23 – Alternate Trigger Edge Mode Function Menu 1
Option
Setting
Description
58
Type
Alternate
When using alternate trigger, the triggered signal comes from two vertical channels. In this mode, you can observe two irrelative signals at a time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1 CH2
Set trigger type information for CH1 signal Set trigger type information for CH2 signal
Mode
Edge
Set trigger type of vertical channel signal to edge
Next Page
Page 1/2
Go to the second page of the TRIGGER Menu.
Table 3.24 – Alternate Trigger Edge Mode Function Menu 2
Option
Setting
Description
Slope
Triggering on rising edge. Triggering on falling edge. Triggering on rising edge and falling edge.
Set up
Enter “Trigger setup menu” (See Table
3.16).
Next Page
Page 2/2
Go to the back to the first page of the TRIGGER Menu.
Table 3.25 – Alternate Trigger Pulse Mode Function Menu 1
Option
Setting
Description
Type
Alternate
The trigger signal comes from two vertical channels when you use alternative trigger. In this mode, you can observe two irrelative signals at the same time.
59
Channels
CH1-CH2
Set the trigger channels
Source
CH1 CH2
Set trigger type information for CH1 signal Set trigger type information for CH2 signal
Mode
Pulse
Set trigger type of the vertical channel signal to Pulse trigger.
Next Page
Page 1/2
Enter the second page of Alternative trigger menu.
Table 3.26 – Alternate Trigger Pulse Mode Function Menu 2
Option
Setting
Description
When
Select how to compare the trigger pulse relative to the value selected in the Set Pulse Width option.
Set Width
20.0ns-10.0s
Selecting this option can turn the universal to set up the pulse width.
Set up
Enter the “Trigger Setup Menu” (See Table 3.16).
Next Page
Page 2/2
Press this button to return to the first page.
Table 3.27 – Alternate Trigger Video Mode Function Menu 1
Option
Setting
Description
60
Type
Alternative
The trigger signal comes from two vertical channels when you use alternative trigger. In this mode, you can observe two irrelative signals at the same time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1 CH2
Set trigger type information for CH1 signal Set trigger type information for CH2 signal
Mode
Video
Set trigger type of the vertical channel signal to Video trigger.
Next Page
Page 1/2
Enter the second page of Alternative trigger menu.
Table 3.28 – Alternate Trigger Video Mode Function Menu 2
Option
Setting
Description
Polarity
(Normal) (Inverted)
Normal triggers on the negative edge of the sync pulse. Inverted triggers on the positive edge of the sync pulse.
Sync
Line Num All lines Odd field Even Field
Select appropriate video sync.
Standard
NTSC Pal/Secam
Select the video standard for sync and line number count.
Set up
Enter the “Trigger Setup Menu” (See Table 3.16).
Next Page
Page 2/2
Press this button to return to the first page.
61
Table 3.29 – Alternate Trigger Slope Mode Function Menu 1
Option
Setting
Description
Type
Alternative
The trigger signal comes from two vertical channels when you use alternative trigger. In this mode, you can observe two irrelative signals at the same time.
Channels
CH1-CH2
Set the trigger channels
Source
CH1 CH2
Set trigger type information for CH1 signal Set trigger type information for CH2 signal
Mode
Slope
Set trigger type of the vertical channel signal to slope trigger.
Next Page
Page 1/2
Enter the second page of the alternative trigger.
Table 3.30 – Alternate Trigger Slope Mode Function Menu 2
Option
Setting
Description
When
Select slope trigger condition.
Time
<Set time>
Turn the “Universal” knob to set the
slope time. Time setup range is 20ns­10s.
Vertical
Select the trigger level that can be
adjusted by “LEVEL” knob. You can adjust “LEVEL A”, “LEVEL B” or adjust
62
them at the same time.
Set up
Enter “Trigger setup menu” (See Table
3.16).
Next Page
Page 2/2
Return to the first page of “Alternative trigger menu”.
Operating Instructions:
To use alternate triggering, follow these steps:
1. Input two irrelative signals to channel 1 and channel 2.
2. Press the AUTO button.
3. Press the TRIG MENU button to enter “trigger menu”.
4. Press the “Type” option button to select “Alternative”.
5. Press the “Channels” option button to select “CH1-CH2”
6. Press the “Source” option button to select “CH1”.
7. Press the CH1 button and turn the “Time/div” knob to optimize
waveform display.
8. Press “Mode” option button to select “Edge”, “Pulse”, “Slope”
or “Video”.
9. Set the trigger according to trigger edge.
10. Press the “Source” option button to select “CH2”.
11. Press the CH2 button and turn the “Time/div” knob to optimize
waveform display.
12. Repeat steps 8 and 9.
63
Coupling
Use “Coupling” to make sure the signal passes through the trigger circuit.
It is useful for obtaining a steady waveform.
If you use trigger coupling, you should press the “TRIG MENU” button and
then select “Edge”, “Pulse”, “Video”, or “Slope” trigger. Then select the Coupling” option in the “Set Up menu”.
Position
The horizontal position control establishes the time between the trigger
position and the screen center. You can adjust the horizontal “POSITION”
knob control to view waveform data before the trigger, after the trigger, or some of each. When you change the horizontal position of a waveform, you are changing the time between the trigger and the center of the display actually. (This appears to move the waveform to the right or left on the display.)
Slope and Level
The Slope and Level controls help to define the trigger. The Slope option (Edge trigger type only) determines whether the oscilloscope finds the trigger point on the rising and/or the falling edge of a signal.
The TRIGGER LEVEL knob controls at what point on the edge the trigger occurs.
64
Figure 3.29 – Rising and Falling Edge Illustration
NOTE: Press the SINGLE button when you want the oscilloscope to
acquire a single waveform.
Trigger coupling affects only the signal passed to the trigger system. It does not affect the bandwidth or coupling of the signal displayed on the screen.
Normal Polarity Sync triggers always occur on negative-going horizontal sync pulses. If the video waveform has positive-going horizontal sync pulses, use the Inverted Polarity selection.
Trigger Holdoff
You can use the Trigger Holdoff function to produce a stable display of complex waveforms. Holdoff is time between when the oscilloscope detects one trigger and when it is ready to detect another. The oscilloscope will not trigger during the holdoff time. For a pulse train, you can adjust the holdoff time so the oscilloscope triggers only on the first pulse in the train.
Falling edge
Rising edge
T
r
i
g g
e r
l
e v
e l c
a n
b
e a
d j
u
s
t
e d
v
e
r
t
i
c
a l
l
y
65
Holdoff time
Trigger position
Trigger level
Figure 3.30 – Trigger Holdoff Illustration
If you want to change holdoff time, please follow the steps below:
1. Press the “TRIG MENU” button to show the “TRIG
Menu”.
2. Press the “Type” option button to select trigger type.
3. Press the “Set Up” option button to enter the “Trigger
setup menu”.
4. Press the “Holdoff” option button and turn the
“Universal” knob to change the holdoff time until the
waveform triggers steadily.
NOTE: Use trigger holdoff to help stabilize the display of aperiodic waveforms.
3.12 Signal Acquisition System
Shown below is the “ACQUIRE” button for entering the menu for “Acquiring Signals”.
Table 3.31 – Acquire Function Menu
Option
Setting
Description
Acquisition
Sampling
Use for sampling and accurately display most of the waveform.
66
Peak Detect
Detect the noise and decrease the possibility of aliasing.
Average
Use to reduce random or uncorrelated noise in the signal display.
Averages (4, 16, 32, 64, 128, 256)
Select number of averages.
Sinx/x
Sinx x
Use sin interpolation Use linear interpolation
Mode
Equ time Real time
Set the Sampling mode to Equivalent time. Set the Sampling mode to Real time.
Sa Rate
Displays system sampling rate.
When you acquire a signal, the oscilloscope converts it into digital form and displays a waveform. The acquisition mode defines how the signal is digitized and the time base setting affects the time span and level of detail in the acquisition.
Sampling: In this acquisition mode, the oscilloscope samples
the signal in evenly spaced intervals to construct the waveform. This mode accurately represents signals most of the time.
Advantage: You can use this mode to reduce random noise. Disadvantage: This mode does not acquire rapid variations in the
signal that may occur between samples. This can result in aliasing may cause narrow pulses to be missed. In these cases, you should use the Peak Detect mode to acquire data.
67
Figure 3.31 – Acquire Menu
Peak Detect: Peak Detect mode capture the maximum and
minimum values of a signal Finds highest and lowest record points over many acquisitions. Advantage: In this way, the oscilloscope can acquire and display narrow pulses, which may have otherwise been missed in Sample mode.
Disadvantage: Noise will appear to be higher in this mode.
Figure 3.32 – Peak Detect
68
Average: The oscilloscope acquires several waveforms, averages
them, and displays the resulting waveform.
Advantage: You can use this mode to reduce random noise.
Figure 3.33 – Average Acquisition
Equivalent Time Sampling: The equivalent time sampling mode
can achieve up to 20 ps of horizontal resolution (equivalent to 50GSa/s). This mode is good for observing repetitive waveforms.
Real Time Sampling: The scope has a maximum Real-time
sampling rate of 1GSa/s.
“RUN/STOP” Button: Press the RUN/STOP button when you want
the oscilloscope to acquire waveforms continuously. Press the button again to stop the acquisition.
“SINGLE” Button: Press the SINGLE button to acquire a single
waveform. Each time you push the SINGLE button, the oscilloscope begins to acquire another waveform. After the oscilloscope detects a trigger it completes the acquisition and stops. When you push the RUN/STOP or SINGLE buttons to start an acquisition, the oscilloscope goes through the following steps:
69
1. Acquire enough data to fill the portion of the waveform
record to the left of the trigger point. This is also called the pre-trigger.
2. Continue to acquire data while waiting for the trigger
condition to occur.
3. Detect the trigger condition.
4. Continue to acquire data until the waveform record is
full.
5. Display the newly-acquired waveform.
Time Base: The oscilloscope digitizes waveforms by acquiring the
value of an input signal at discrete points. The time base allows you to control how often the values are digitized. To adjust the time base to a horizontal scale that suits your purpose, use the Time/div knob.
Time Domain Aliasing: Aliasing occurs when the oscilloscope
does not sample the signal fast enough to construct an accurate waveform record. When this happens, the oscilloscope displays a waveform with a frequency lower than the actual input waveform, or triggers and displays an unstable waveform.
Sampled points
Apparent low-frequency waveform due to aliasing
Actual high-frequency waveform
Figure 3.34 – Time Domain Aliasing Illustration
Operating Instructions:
Set up Sampling Format
You can press the “Acquisition” option button or turn the “Universal” knob to select “Sampling” mode, “Peak Detect” mode or “Average” mode.
70
Set up Averages
When you select “Average” format, you can press the “Averages” option button to select “4”, “16”, “32”, “64”, “128”or “256”.
Set up function interpolation
You can select Sinx interpolation or linear interpolation.
Set up Sampling Mode
Press the “Mode” option button to select “Real Time” or “Equ Time”.
Set up Sampling Rate
The sampling rate is based on the time division scaling on the screen. Adjust the sampling rate by turning the Time/div front panel knob. The sampling rate is shown under the “Sa Rate” display.
3.13 Display System
The display function can be setup by pressing the “DISPLAY” button.
Table 3.32 – Display System Menu 1
Option
Setting
Description
Type
Vectors
Dots
Vectors fill the space between adjacent sample points in the display. There is no link between adjacent sample points.
Persist
Off 1 sec 2 sec 5 sec Infinite
Sets the length of time each displayed sample point remains displayed.
Intensity <Intensity>
Set waveforms’ intensity.
Brightness <Brightness>
Set grid brightness. Next Page
Page 1/3
Press this button to enter second
71
page.
Figure 3.35 – Display Menu 1
Table 3.33 – Display System Menu 2
Option
Setting
Description
Format
YT XY
YT format displays the vertical voltage in relation to time (horizontal scale). XY format displays a dot each time a sample is acquired on channel 1 and channel 2
Screen
Normal Inverted
Set to normal mode. Set to invert color display mode.
Grid
Display grids and axes on the screen. Turn off the grids. Turn off the grids and axes.
Menu Display
2sec 5sec 10sec
Set menu display time on screen.
72
20sec Infinite
Next Page
Page 2/3
Press this button to enter the second page of “Display menu”.
Figure 3.36 – Display Menu 2
Table 3.34 – Display System Menu 3
Option
Setting
Description
Skin
Classical Modern Tradition Succinct
Set up screen style.
Next Page
Page 3/3
Press this button to return to the first page.
Operating Instructions:
Set up waveform display type
1. Press the “DISPLAY” button to enter the “Display” menu.
2. Press the “Type” option button to select “Vectors” or “Dots”.
73
Set up Persist
Press “Persist” option button to select “Off”, “1 Sec”, “2 Sec”, “5 Sec” or “Infinite”. You can use this option to observe some special waveforms.
Figure 3.37 – Persist Screen
Set up Intensity
Press the “Intensity” option button and turn the “Universal” knob to adjust waveforms’ intensity.
Set up Brightness
Press the “Brightness” option button and turn the “Universal” knob
to adjust grid brightness.
Set up display format
1. Press the “Next Page” option button to enter second
display menu.
2. Press the “Format” option button to select “YT” or “XY”.
Set up Screen
Press the “Screen” option button to select “Normal” or “Inverted” to
set the screen display color.
Set up Grid
Press the “Grid” option button to select “ ”, “ ”or“
”to set the screen whether display grid or not.
Set up Menu Display
74
Press the “Menu Display” option button to select “2 sec”, “5sec”, “10sec”, “20sec” or “Infinite” to set menu display time on screen.
Set Skin
Press the “skin” option button or turn the “Universal” knob to select “Classical”, “Modern”, “Traditional” or “Succinct”.
X-Y Format
Use the XY format to analyze phase differences, such as those represented by Lissajous patterns. The format plots the voltage on channel 1 against the voltage on channel 2, where channel 1 is the horizontal axis and channel 2 is the vertical axis. The oscilloscope uses the untriggered Sample acquisition mode and displays data as dots.
NOTE: The oscilloscope can capture a waveform in normal YT mode at any sampling rate. You can view the same waveform in XY mode.
To do so, stop the acquisition and change the display format to XY.
Operation steps:
Channel 1 “Volt/div” and vertical “POSITION” set up the
horizontal scale and position.
Channel 2 “Volt/div” and vertical “POSITION” set up the
horizontal scale and position.
Turn the “Time/div” knob to adjust the sampling rate The following functions are forbidden in XY display form:
o Benchmark wave form and wave mathematic o Cursor o Auto (resets display format to YT) o Trigger Control o Horizontal Position Knob o Vector Display Type o Scan Display
75
3.14 Measure System
The oscilloscope displays the voltage in relation to time and tests the waveform displayed. Different measurement techniques such as scale, Cursor and auto measure modes are used.
Scale Measurement
This method allows you to make a quick, visual estimate. For example, you might look at waveform amplitude and determine that it is a little more than 100 mV. You can take simple measurements by counting the major and minor graticule divisions involved and multiplying by the scale factor. For example, if you counted five major vertical graticule divisions between the minimum and maximum values of a waveform and knew you had a scale factor of 100 mV/div, then you could easily calculate your peak-to­peak voltage as follows:
5 𝑑𝑖𝑣𝑖𝑠𝑖𝑜𝑛𝑠 ×
100 𝑚𝑉
𝑑𝑖𝑣𝑖𝑠𝑖𝑜𝑛
= 500 𝑚𝑉
Cursor Measurement
Press the “CURSORS” button to display the “Cursor” menu. The cursor measurement has three modes: Manual, Track, and Auto Measure.
Manual Mode
Table 3.35 – Manual Cursor Menu
Option
Setting
Description
Cursor Mode
Manual
In this menu, set the manual cursor measure.
Type
Voltage
Time
Use cursor to measure voltage parameters. Use cursor to measure time
76
parameters.
Source
CH1 CH2 MATH REFA REFB
Select input signal channel. Cur A
Select this option, use “Universal” knob to adjust cursor A.
Cur B
Select this option, use “Universal” knob to adjust cursor B.
In this mode, the screen displays two horizontal parallel cursors or vertical parallel cursors to measure voltage or time. You can move the cursor by
turning the “Universal” knob. Before using cursors, you should make sure
that you have set the signal source as the channel for measuring.
Voltage Cursor: Voltage cursors appear as horizontal lines on the
display and measure the vertical parameters.
Time Cursor: Time cursors appear as vertical lines on the display
and measure the horizontal parameters.
Cursor Moving: Use the “universal” knob to move cursor 1 and
cursor 2. They could be moved when the corresponding cursor option are selected, and cursor value will display on the bottom left and top left of the screen when you move the cursor.
To do manual cursor measurements, follow these steps:
1. Press CURSOR button to enter the cursor function menu.
2. Press the “Cursor Mode” option button to select “Manual”.
3. Press the “Type” option button to select “Voltage” or “Time”.
4. Press the “Source” option button to select “CH1”, “CH2”,
“MATH”, “REFA”, “REFB” according to input signal channel.
5. Select “Cur A”, turn the “Universal” knob to adjust Cursor A.
6. Select “Cur B”, turn the “Universal” knob to adjust Cursor B.
7. The measurement values are displayed on the top of the left
corner.
77
If the measurement type is set to “Voltage”, the values are as followed:
The voltage increment between Cursor A and Cursor
B: ΔV
The value of Cur A: CurA The value of Cur B: Cur B
If the measurement type is set to “Time”, the values are as followed:
The time increment between Cursor A and Cursor B:
ΔT
The reciprocal of time increment between Cursor A
and Cursor B: 1/ΔT
The value of Cur A: CurA The value of Cur B: Cur B
Figure 3.38 – Cursor Menu (Manual)
Track Mode
Table 3.36 – Track Mode Menu
Option
Setting
Description
Cursor Mode
Track
In this mode, set track cursor measure.
78
Cursor A
CH1 CH2 NONE
Set the input signal channel that the Cursor A will measure.
Cursor B
CH1 CH2 NONE
Set the input signal channel that the Cursor B will measure.
Cur A
Select this option, turn the
“Universal” knob to adjust
horizontal coordinate of Cursor A.
Cur B
Select this option, turn the
“Universal” knob to adjust
horizontal coordinate of Cursor B.
In this mode, the screen displays two cross cursors. The cross cursor sets
the position on the waveform automatically. You could adjust cursor’s
horizontal position on the waveform by turning the “Universal” knob. The oscilloscope displays the values on the top left of the screen.
To do track cursor measurement, follow these steps:
1. Press CURSOR button to enter the cursor measure function
menu.
2. Press the “Cursor Mode” option button to select “Track”.
3. Press the “Cursor A” option button to select the input signal
channel.
4. Press the “Cursor B” option button to select the input signal
channel.
5. Select “Cur A”, turn the “Universal” knob to move Cursor A
horizontally.
6. Select “Cur B”, turn the “Universal” knob to move Cursor B
horizontally.
7. The measurement values are displayed on the top left of the
screen:
AT: The horizontal position of Cursor A (Time cursor centered around the midpoint of screen).
79
A→V: The Vertical position of Cursor A (Voltage cursor centered around channel ground level).
B→T: The horizontal position of Cursor B (Time cursor
centered around the midpoint of screen).
B→V: The Vertical position of Cursor B (Voltage cursor
centered around channel ground level).
ΔT: Horizontal space between Cursor A and Cursor B (Time value between two cursors).
1/ΔT: The reciprocal of horizontal space between cursor A and cursor B.
ΔV: Vertical space between Cursor A and Cursor B (Voltage value between two cursors).
Figure 3.39 – Cursor Menu (Track)
Auto Mode
This mode will take effect with automatic measurements. The instruments will display cursors while measuring parameters automatically. These cursors demonstrate the physical meanings of these measurements.
80
To do auto cursor measurements, follow these steps:
1. Press the CURSOR button to enter “Cursor measure menu”.
2. Press the “Cursor Mode” option button to select “Auto”.
3. Press the “MEASURE” button to enter “Auto cursor measure
mode menu” to select the parameter that you want to
measure.
Figure 3.40 – Auto Mode
Auto Measurement
When you take automatic measurements, the oscilloscope does all the calculation for you. The measurements use all the recorded points in the memory, which are more accurate than measurements made using the graticule lines or cursor measurements because these measurements are confined to be made by only using points on the display and not all the data points recorded by the oscilloscope.
Press the ‘MEASURE’ button for Automatic Test.
There are three auto measurement types: Voltage Measure, Time Measure, and Delay Measure. There are a total of 32 measurement parameters.
81
Table 3.37 – Auto Measurement Menu
Option
Description
Voltage
Press this button to enter the Voltage measure menu.
Time
Press this button to enter the Time measure menu.
Delay
Press this button to enter the Delay measure menu.
All Mea
Press this button to enter the All Measurement menu.
Return
Press this option button to return to the home page of the auto measure menu.
Figure 3.41 – Auto Measure Menu
Table 3.38 – Auto Voltage Measurement Menu
Option
Setting
Description
Source
CH1, CH2
Select input signal source for Voltage measure.
Type
Vpp, Vmax, Vmin, Vpp, Vamp, Vtop, Vbase, Vavg, Mean, Vrms, Cycle Vrms, FOVShoot, FPREShoot, ROVShoot, RPREShoot
Press the “Type”
button or turn the
“Universal” knob to
select Voltage measure parameter.
82
Display the corresponding icon and measure value of your selected Voltage measure parameter.
Return
Return to the first page of auto measurement menu.
Table 3.39 – Auto Time Measurement Menu
Option
Setting
Description
Source
CH1, CH2
Select input signal source for Time measure.
Type
Period, Freq, +Width, -Width, Rise Time, Fall Time, BWidth, +Duty, ­Duty
Press the “Type”
button or turn the
“Universal” knob to
select Time measure parameter.
Display the corresponding icon and measure value of your selected time measure parameter.
Return
Return to the first page of auto measurement menu.
83
Table 3.40 – Auto Delay Measurement Menu
Option
Setting
Description
Source
CH1, CH2
Select any two input signal source for Delay measure.
Type
Phase, FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF
Press the “Type” button or turn the “Universal” knob
to select Delay measure parameter.
Display the corresponding icon and measure value of your selected Delay measure parameter.
Return
Return to the first page of auto measurement menu.
Table 3.41 – All Measurement Menu
Option
Setting
Description
Source
CH1 CH2
Select input signal channel.
Voltage
On Off
Turn on the all measurement function to measure voltage parameters. Turn of the all measurement function to measure voltage parameters.
Time
On Off
Turn on the all measurement function to measure Time parameters. Turn of the all measurement function to measure Time parameters.
Delay
On Off
Turn on the all measurement function to measure Delay parameters. Turn of the all measurement function to measure Delay parameters.
Return
Return to the “All Measure main
84
menu”.
Table 3.42 Types of Measurements
Measure Type
Description
Vmax
The most positive peak voltage measured over the entire waveform.
Vmin
The most negative peak voltage measured over the entire waveform.
Vpp
Measures the absolute difference between the maximum and minimum peaks of the entire waveform.
Vtop
Measures the highest voltage over the entire waveform.
Vbase
Measures the lowest voltage over the entire waveform.
Vamp
Voltage between Vhig and Vlow of a waveform.
Vavg
The arithmetic mean over the first cycle in the waveform.
Mean
The arithmetic mean over the entire waveform.
Crms
The true Root Mean Square voltage over the first cycle in the waveform.
Vrms
The true Root Mean Square voltage over the entire waveform.
ROVShoot
Defined as (Vmax-Vhig)/Vamp after the waveform rising.
FOVShoot
Defined as (Vmin-Vlow)/Vamp after the waveform falling.
RPREshoot
Defined as (Vmin-Vlow)/Vamp before the waveform rising.
85
FPREshoot
Defined as (Vmax-Vhig)/Vamp before the waveform falling.
Rise Time
Measures the time between 10% and 90% of the first rising edge of the waveform.
Fall Time
Measures the time between 90% and 10% of the first falling edge of the waveform.
BWid
The duration of a burst. Measured over the entire waveform.
+ Wid
+ Width Measures the time between the first rising edge and the next falling edge at the waveform 50% level.
- Wid
-Width Measures the time between the first falling edge and the next rising edge at the waveform 50% level.
+ Duty
Measures the first cycle waveform. Positive Duty Cycle is the ratio between positive pulse width and period.
Duty
Measures the first cycle waveform. Negative Duty Cycle is the ratio between negative pulse width and period.
Phase
The amount one waveform leads or lags another in time. Expressed in degrees, where 360 degrees comprise one waveform cycle.
FRR
The time between the first rising edge of source X and the first rising edge of source Y.
FRF
The time between the first rising edge of source X and the first falling edge of source Y.
FFR
The time between the first falling edge of source X and the first rising edge of source Y.
FFF
The time between the first falling edge of source X and the first falling edge of source Y.
86
LRR
The time between the first rising edge of source 1 and the last rising edge of source 2.
LRF
The time between the first rising edge of source X and the last falling edge of source Y.
LFR
The time between the first falling edge of source X and the last rising edge of source Y.
LFF
The time between the first falling edge of source X and the last falling edge of source Y.
If you want to measure voltage parameters, please follow the steps below:
1. Press the MEASURE button to enter the Auto measurement
menu.
2. Press the first option button to enter the second measurement
menu.
3. Select measure type. If you press the “Voltage” option button,
“Voltage measurement” menu will display on the screen.
4. Press the Source option button to select CH1, CH2
according to input signal channel.
5. Press the Type option button to select the parameter type that
you want to measure. The corresponding icon and value will display below the measure parameter.
Loading...