BK Precision 2165A, 2630, 2625, 2620A User Manual

INSTRUCTION
MANUAL
SPECTRUM ANALYZERS
MODELS
2165A 2620A
TEST INSTRUMENT SAFETY
Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must sometimes be performed where exposed voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will stop most human heartbeats. Voltage as low as 35 volts dc or ac rms should be considered dangerous and hazardous since it can produce a lethal current under certain conditions. Higher voltages pose an even greater threat because such voltage can more easily produce a lethal current. Your normal work habits should include all accepted practices to prevent contact with exposed high voltage, and to steer current away from your heart in case of accidental contact with a high voltage. You will significantly reduce the risk factor if you know and observe the following safety precautions:
5. Don’t expose high voltage needlessly. Remove
housings and covers only when necessary. Turn off equipment while making test connections in high­voltage circuits. Discharge high-voltage capacitors after removing power.
6. If possible, familiarize yourself with the equipment
being tested and the location of its high voltage points. However, remember that high voltage may appear at unexpected points in defective equipment.
7. Use an insulated floor material or a large, insulated
floor mat to stand on, and an insulated work surface on which to place equipment; and make certain such surfaces are not damp or wet.
8. Use the time proven “one hand in the pocket”
technique while handling an instrument probe. Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path.
9. When testing ac powered equipment, remember that
ac line voltage is usually present on some power input circuits such as the on-off switch, fuses, power transformer, etc. any time the equipment is connected to an ac outlet, even if the equipment is turned off.
1. Some equipment with a two-wire ac power cord,
including some with polarized power plugs, is the “hot chassis” type. This includes most recent television receivers and audio equipment. A plastic or wooden cabinet insulates the chassis to protect the customer. When the cabinet is removed for servicing, a serious shock hazard exists if the chassis is touched. Not only does this present a dangerous shock hazard, but damage to test instruments or the equipment under test may result from connecting the ground lead of most test instruments to a “hot chassis”. To test “hot chassis” equipment, always connect an isolation transformer between the ac outlet and the equipment under test. The B+K Precision Model TR-110 Isolation Transformer, or Model 1653A or 1655A AC Power Supply is suitable for most applications. To be on the safe side, treat all two-wire ac equipment as “hot-chassis” unless you are sure it has an isolated chassis or an earth ground chassis.
2. On test instruments or any equipment with a 3-wire ac
power plug, use only a 3-wire outlet. This is a safety feature to keep the housing or other exposed elements at earth ground.
3. B+K Precision products are not authorized for use in
any application involving direct contact between our product and the human body, or for use as a critical component in a life support device or system. Here, “direct contact” refers to any connection from or to our equipment via any cabling or switching means. A “critical component” is any component of a life support device or system whose failure to perform can be reasonably expected to cause failure of that device or system, or to affect its safety or effectiveness.
4. Never work alone. Someone should be nearby to
render aid if necessary. Training in CPR (cardio­pulmonary resuscitation) first aid is highly recommended.
2
Instruction Manual
for
Models
2615A, 2620A, 2625, 2630
SPECTRUM ANALYZERS
3
TABLE OF CONTENTS
TEST INSTRUMENT SAFETY .....Inside front cover
SPECIFICATIONS ....................................................5
OPTIONAL ACCESSORIES ....................................6
GENERAL INFORMATION.....................................7
Symbols ..................................................................7
Tilt handle................................................................7
Safety.......................................................................7
Operating Conditions................................................8
Maintenance.............................................................8
Selecting the Line Voltage........................................8
INTRODUCTION ......................................................9
General ....................................................................9
Operating Considerations .........................................9
CONTROLS AND INDICATORS...........................10
CALIBRATION.......................................................15
Vertical Calibration................................................15
Horizontal Calibration............................................15
Page
Page
INTRODUCTION TO SPECTRUM ANALYSIS....17
General .................................................................. 17
Types of Spectrum Analyzers................................. 17
Spectrum Analyzer Requirements........................... 18
Frequency Measurements ....................................... 18
Resolution.............................................................. 18
Sensitivity.............................................................. 19
Video Filtering.......................................................19
Spectrum Analyzer Sensitivity................................ 19
Frequency Response ............................................... 20
Tracking Generators............................................... 20
APPENDIX–dBm CONVERSION ..........................22
CUSTOMER SUPPORT..........................................23
INSTRUMENT REPAIR SERVICE ....................... 23
WARRANTY SERVICE INSTRUCTIONS............ 24
LIMITED ONE-YEAR WARRANTY..................... 25
4
Frequency
Input
SPECIFICATIONS
Frequency range:
0.15 MHz to 1050 MHz (–3 dB) (Models 2625 and 2630)
0.15 MHz to 500 MHz (–3 dB) (Models 2615A and 2620A)
Center frequency display accuracy: ±100 kHz Marker accuracy: ±(0.1% span + 100 kHz) Frequency display resolution: 100 kHz
(4½ digit LED for Models 2625 and 2630) (4 digit LED for Models 2615A and 2620A)
Frequency scanwidth:
100 kHz/div. to 100 MHz/div. (Models 2625 and 2630)
in 1-2-5 steps and 0 Hz/div. (Zero Scan)
50 kHz/div. to 50 MHz/div. (Models 2615A and 2620A)
in 1-2-5 steps and 0 Hz/div. (Zero Scan)
Frequency scanwidth accuracy: ±10%
Frequency stability: Drift: <150 kHz / hour IF-Bandwidth (–3 dB):
Resolution: 400 kHz and 20 kHz
(Models 2625 and 2630)
Resolution: 250 kHz and 20 kHz
(Models 2615A and 2620A)
Video-Filter on: 4 kHz
Sweep rate: 43 Hz
Amplitude
Amplitude range: –100 dBm to +13 dBm Screen display range: 80 dB (10 dB / div.) Reference level: –27 dBm to +13 dBm
(in 10 dB steps)
Reference level accuracy: ±2 dB Average noise level:
–99 dBm (20 kHz BW) (Models 2625 and 2630) –99 dBm (20 kHz BW) (Models 2615A and 2620A)
Distortion:
<–75 dBc; 2nd and 3rd harmonic
3rd order intermod.: –70 dBc (two signals >3 MHz apart)
Sensitivity: dB above average noise level Log scale fidelity: ±2 dB (without attn.) Ref.: 250 MHz IF gain: 10 dB adjustment range
NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.com for the most current product
information.
Input impedance: 50? Input connector: BNC Input attenuator: 0 to 40 dB (4 x 10 dB steps) Input attenuator accuracy: ±1 dB/10 dB step Max. input level: +10 dBm, ±25VDC (0 dB attenuation)
+20 dBm (40 dB attenuation)
Tracking Generator (Models 2620A and 2630 only)
Output level range: –50 dBm to +1 dBm
(in 10 dB steps and variable)
Output attenuator: 0 to 40 dB (4 x 10 dB steps) Output attenuator accuracy: ±l dB Output impedance: 50? (BNC) Frequency range:
0.15 MHz to 1050 MHz (Model 2630)
0.1 MHz to 500 MHz (Model 2620A)
Frequency response: ±1.5 dB Radio Frequency Interference (RFI): <20 dBc
General
Display: CRT, 6 inch, 8 x 10 div. internal graticule Trace rotation: Adjustable on front panel Output Probe Power: 6V Line voltage: 115 / 230V ±10%, 50-60Hz Power consumption: approx. 27W Operating ambient temperature:
+10°C to +40°C
Protective system: Safety Class I (IEC 1010-1) Weight:
Approx. 15.4 lb. (6 kg) (Models 2625 and 2630) Approx. 13.2 lb. (5 kg) (Models 2615A and 2620A)
Dimensions:
4.9 in. (125 mm) H x 11.2 in. (285mm) W x 15 in. (380 mm) D.
Accessories Supplied
Power Cord Instruction Manual
5
Near Field “Sniffer” Probe Set; Model PR
-
261
Antenna Kit; Model AN
-18
OPTIONAL ACCESSORIES
The PR-261 is the ideal tool kit for the investigation of RF electromagnetic fields. It is indispensable for EMI pre-compliance testing during product development, prior to third party testing. The set includes three hand-held probes with a built-in pre-amplifier covering the frequency range from 10 kHz to 1000 MHz. The set includes one magnetic field probe, one electric field probe and one high impedance probe. All have high sensitivity and are matched to the 50? inputs of spectrum analyzers. The power can be supplied either from the batteries or through a power cord directly connected to a B+K Precision Models 2615A, 2620A, 2625 and 2630. Signal feed is via a 1.5 meter BNC-cable. When used in conjunc­tion with a spectrum analyzer, the probes can be used to locate and qualify EMI sources. They are especially suited to locate emission “hot spots” on PCBs and cables, as well as evaluate EMC problems at the breadboard and prototype level. They enable the user to evaluate radiated fields and perform shield effectivity comparisons. Mechanical screening performance and immunity tests on cables and components are easily performed. Faulty components and poor bonding locations can be isolated.
The magnetic probe incorporates a high degree of rejection of both stray and direct electric fields, and provides far greater repeatability than with conventional field probes. Measurements can be made on the very near field area that is close to components or radiation sources. The electric field (monopole) probe has the highest sensitivity of all three probes. It can be used to check screening and perform pre-compliance testing on a comparative basis. The high impedance probe is used to measure directly on the components under test or at the conductive trace of a PC board. It has an input capacitance of only 2 pF and supplies virtually no electrical charge to the device under test.
Probe Set Specifications
All probes are electrically shielded and are supplied in a carrying case.
Frequency range: 100 kHz–1,2GHz
Power supply: 6V from Spectrum Analyzers or Batteries
Operating current: 10-15 mA
Probe dimensions: 40 x 19 x 195 mm (approx.)
Broad band antenna is useful for radiated signal measurement.
Deluxe Carrying Case; Model LC-210
Rugged cordura carrying case is foam padded for instrument protection, has zipped pockets for manual and accessories, and includes web hand strap and shoulder strap.
Viewing Hood; Model VH-26
Shades CRT to block ambient light and improve definition of the display.
50-ohm to 75-ohm Matching Network; Model ZTF-1
Most RF networks (except cable TV) have an impedance of 50 ohms. The spectrum analyzers also have a 50 ohm input impedance, which allows direct connection. Cable TV networks have an impedance of 75 ohms. To use the spectrum analyzer with 75 ohm networks, Model ZTF-1 will match the 75 ohm network to the 50 ohm input impedance of the spectrum analyzer.
50-ohm Feedthru Termination; Model TE-26
The output levels of the tracking generator of the Models 2620 and 2630 are correct only when terminated into 50 ohms. The Model TE-26 provides a 50 ohm termination and a BNC feedthru connection so that the tracking generator output may be fed into high impedance circuit at a calibrated level.
Rack Mount Adapter; Model RM-26
Model RM-26 mounts the spectrum analyzer to standard 19-inch racks.
6
GENERAL INFORMATION
The Models 2615A, 2620A, 2625, and 2630 spectrum ana­lyzers are easy to operate. The logical arrangement of the controls allows anyone to quickly become familiar with the operation of the instrument, however, experienced users are also advised to read through these instructions so that all functions are understood. Immediately after unpacking, the instrument should be checked for mechanical damage and loose parts in the interior. If there is transport damage, the supplier must be informed immediately. The instrument must then not be put into operation.
Symbols
ATTENTION - refer to manual
Danger - High voltage
Protective ground (earth) terminal
Tilt handle
To view the screen from the best angle, there are three different positions (C, D, E) for setting up the instrument (see Figure 1). If the instrument is set down on the floor after being carried, the handle automatically remains in the upright carrying position (A). In order to place the instrument onto a horizontal surface, the handle should be turned to the upper side of the Spectrum Analyzer (C). For the D position (10° inclination), the handle should he turned to the opposite direction of the carrying position until it locks in place automatically underneath the instrument. For the E position (20° inclination), the handle should be pulled to release it from the D position and swing backwards until it locks once more. The handle may also be set to a position for horizontal carrying by turning it to the upper side to lock in the B position. At the same time, the instrument must be lifted, because otherwise the handle will jump back.
Safety
This instrument has been designed and tested in accordance with IEC Publication 1010-1, Safety requirements for elec-
trical equipment for measurement, control, and laboratory use. The CENELEC regulations EN 61010-1 correspond to
this standard. It has left the factory in a safe condition. This instruction manual contains important information and warn­ings which have to be followed by the user to ensure safe operation and to retain the Spectrum Analyzer in a safe condition.
Fig. 1. Tilt Handle Operation
The case, chassis and all measuring terminals are connected to the protective earth contact of the appliance inlet. The instrument operates according to Safety Class I (three-- conductor power cord with protective earthing conductor and a plug with earthing contact). The mains/line plug shall only be inserted in a socket outlet provided with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor.
The mains/line plug should be inserted before connections are made to measuring circuits. The grounded accessible metal parts (case, sockets, jacks) and the mains/line supply contacts (line/live, neutral) of the instrument have been tested against insulation breakdown with 2200V DC. Under certain conditions, 50 Hz or 60 Hz hum voltages can occur in the measuring circuit due to the interconnection with other mains/line powered equipment or instruments. This can be avoided by using an isolation transformer (Safety Class II) between the mains/line outlet and the power plug of the device being investigated. Most cathode-ray tubes develop X-rays. However, the dose equivalent rate falls far below the maximum permissible value of 36pA/kg (0.5mR/h). Whenever it is likely that protection has been impaired, the instrument shall be made inoperative and be secured against any unintended operation. The protection is likely to be impaired if, for example:
shows visible damage
fails to perform the intended measurements
7
Selecting the Line Voltage
GENERAL INFORMATION
has been subjected to prolonged storage under
unfavorable conditions (e.g. in the open or in moist environments)
has been subjected to severe transport stress (e.g. in
poor packaging).
Operating Conditions
The instrument has been designed for indoor use. The permissible ambient temperature range during operation is +10°C (+50°F) to +40°C (+104°F). It may occasionally be subjected to temperatures between +10°C (+50°F) and – 10°C (+14°F) without degrading its safety. The permissible ambient temperature range for storage or transportation is – 40°C (+14°F) to +70°C (+158°F).
The maximum operating altitude is up to 2200m. The maximum relative humidity is up to 80%.
If condensed water exists in the instrument it should be acclimatized before switching on. In some cases (e.g. instrument extremely cold) two hours should be allowed before the instrument is put into operation. The instrument should be kept in a clean and dry room and must not be operated in explosive, corrosive, dusty, or moist environments. The spectrum analyzer can be operated in any position, but the convection cooling must not be impaired. For continuous operation the instrument should be used in the horizontal position, preferably tilted upwards, resting on the tilt handle.
The specifications stating tolerances are only valid if the instrument has warmed up for 60 minutes at an ambient temperature between +15°C (+59°F) and +30°C (+86°F). Values without tolerances are typical for an average instrument.
Maintenance
Various important properties of the spectrum analyzer should be carefully checked at certain intervals. Only in this way it is certain that all signals are displayed with the accuracy on which the technical data are based. The exterior of the instrument should be cleaned regularly with a dusting brush. Dirt which is difficult to remove on the casing and handle, the plastic and aluminum parts, can be removed with a moistened cloth (99% water + 1% mild detergent). Spirit or washing benzene (petroleum ether) can be used to remove greasy dirt. The screen may be cleaned with water or washing benzene (but not with spirit (alcohol) or solvents), it must then be wiped with a dry clean lint-free cloth. Under no circumstances may the cleaning fluid get into the instrument. The use of other cleaning agents can attack the plastic and paint surfaces.
The spectrum analyzer operates at mains/line voltages of 115V AC and 230V AC. The voltage selection switch is located on the rear of the instrument and displays the selected voltage. The correct voltage can be selected using a small screwdriver.
Remove the power cable from the power connector prior to making any changes to the voltage setting. The fuses must also be replaced with the appropriate value (see Fuse Type ) prior to connecting the power cable. Both fuses are externally accessible by removing the fuse cover located above the 3-pole power connector.
The fuseholder can be released by pressing its plastic retainers with the aid of a small screwdriver (see Figure 2). The retainers are located on the right and left side of the holder and must be pressed towards the center. The fuse(s) can then be replaced and pressed in until locked on both sides.
Use of patched fuses or short-circuiting of the fuseholder is not permissible; B+K Precision assumes no liability whatsoever for any damage caused as a result, and all warranty claims become null and void.
Fuse type:
Size 5 x 20 mm; 250-Volt AC; must meet IEC specification 127, Sheet III (or DIN 41 662 or DIN 41 571, sheet 3).
Time characteristic: time-lag
Line voltage 115V~ ±10%: Fuse rating: T 315mA Line voltage 230V~ ±10%: Fuse rating: T 160mA
Fig. 2. Fuse Replacement
8
General
INTRODUCTION
The spectrum analyzer permits the detection of spectrum components of electrical signals in the frequency range of 0.15 to 1050 MHz for Models 2625 and 2630 and 0.15 to 500 MHz for Models 2615 and 2620. The detected signal and its content have to be repetitive. In contrast to an oscilloscope operated in Yt mode, where the amplitude is displayed on the time domain, the spectrum analyzer displays amplitude on the frequency domain (Yf). The individual spectrum components of “a signal” become visible on a spectrum analyzer. The oscilloscope would display the same signal as one resulting waveform.
Models 2625 and 2630
The spectrum analyzer works according to the triple superhet receiver principle. The signal to be measured (fin = 0.15 MHz to 1050 MHz) is applied to the 1st mixer where it is mixed with the signal of a variable voltage controlled oscillator (fLO 1350 MHz – 2350 MHz). This oscillator is called the lst LO (local oscillator). The difference between the oscillator and the input frequency (fLO – fin = 1st IF) is the first intermediate frequency, which passes through a waveband filter tuned to a center fre­quency of 1350 MHz. It then enters an amplifier, and this is followed by two additional mixing stages, oscillators and amplifiers. The second IF is 29.875 MHz and the third is 2.75 MHz. In the third IF stage, the signal can be selectively transferred through a filter with 400 kHz or 20 kHz bandwidth before arriving at an AM demodulator. The logarithmic output (video signal) is transferred directly, or via a low pass filter to another amplifier. This amplifier output is connected to the Y deflection plates of the CRT.
Models 2615A and 2620A
The spectrum analyzer works according to the triple superhet receiver principle. The signal to be measured (fin = 0.5 MHz to 500 MHz) is applied to the 1st mixer where it is mixed with the signal of a variable voltage controlled oscillator (fLO 610 MHz – 1110 MHz). This oscillator is called the 1st LO (local oscillator). The difference between the oscillator and the input frequency (fLO – fin = 1st IF) is the first intermediate frequency, which passes through a waveband filter tuned to a center frequency of 609.5 MHz. It then enters an amplifier, and this is followed by two additional mixing stages, oscillators and amplifiers. The second IF is 29.5 MHz and the third is 2.9 MHz. In the third IF stage, the signal can be selectively transferred through a filter with 250 kHz or 20 kHz bandwidth before arriving at an AM demodulator. The logarithmic output (video signal) is transferred directly, or via a low pass filter to another amplifier. This amplifier output is connected to the Y deflection plates of the CRT.
The X deflection is performed with a ramp generator voltage. This voltage can also be superimposed on a dc voltage which allows for the control of 1st LO. The spectrum analyzer scans a frequency range depending on the ramp height. This span is determined by the scanwidth setting. In ZERO SCAN mode only the direct voltage controls the 1st LO.
Models 2620A and 2630 each include a tracking generator. This generator provides sine wave voltages within the frequency range of 0.1 to 1050 MHz for Model 2630 and 0.1 to 500 MHz for Model 2620A. The tracking generator frequency is determined by the first oscillator (1st LO) of the spectrum analyzer section. Spectrum analyzer and tracking generator are frequency synchronized.
Operating Considerations
It is very important to read Safety in the GENERAL INFORMATION Section including the instructions prior to operating the spectrum analyzer. No special knowledge is necessary for the operation of the spectrum analyzer. The straightforward front panel layout and the limitation to basic functions guarantee efficient operation immediately. To ensure optimum operation of the instrument, some basic instructions need to be followed.
CAUTION
The most sensitive component of the spectrum analyzer is the input section. It consists of the signal attenuator and the first mixer. Without input attenuation, the voltage at the input must not exceed +10 dBm (0.7Vrms) AC or ±25 volt DC. With a maximum input attenuation of 40 dB the AC voltage must not exceed +20 dBm.
Exceeding these limits will damage the input attenuator and/or the first mixer.
Prior to examining unidentified signals, the presence of unacceptable high voltages has to be checked. It is also recommended to start measurements with the highest possible attenuation and a maximum frequency range. The user should also consider the possibility of excessively high signal amplitudes outside the covered frequency range, although not displayed (e.g. 1200 MHz). The frequency range of 0 Hz to 150 kHz is not specified. Spectral lines within this range would be displayed with incorrect amplitude.
A particularly high intensity setting shall be avoided. The way signals are displayed on the spectrum analyzer typically allows for any signal to be recognized easily, even with low intensity.
Due to the frequency conversion principle, a spectral line is visible at 0 Hz. It is called IF-feedthrough. The line appears when the 1st LO frequency passes the IF amplifiers and filters. The level of this spectral line is different in each instrument. A deviation from the full screen does not indicate a malfunctioning instrument.
9
Loading...
+ 19 hidden pages