BK Precision 1744A Instruction Manual

0-35 V, 0-10 A
DC POWER SUPPLY
With Dual
INSTRUCTION
MANUAL
Model 1744A & 1745A
2
TEST INSTRUMENT SAFETY
WARNING
Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must sometimes be performed where exposed high voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will stop most human heartbeats. Voltage as low as 35 volts dc or ac rms should be considered dangerous and hazardous since it can prod uce a lethal current under certain conditions. Higher voltages are even more dangerous. Your normal work habits should include all a ccepted practices to prevent contact with exposed high voltage, and to steer current away from your heart in case of accidental con tact with a high voltage. Observe the following safety precautions:
1. There is little danger of electrical shock from the dc output of this power supply. However, there are several other possible test
conditions using this power supply that can create a high voltage shock hazard: a. If the equipment under test is the “hot chassis” type, a serious shock hazard exists unless the equipment is unplugged (just turning
off the equipment does not remove the hazard), or an isolation transformer is used.
b. If the equipment under test is “powered up” (and that equipment uses high voltage in any of its circuits), the power supply outputs
may be floated to the potential at the point of connection. Remember that high voltage may appear at unexpected points in defective equipment. Do not float the power supply output to more than 100 volts peak with respect to chassis or earth ground.
c. If the equipment under test is “off” (and that equipment uses high voltage in any of its circuits under normal operation), di scharge
high-voltage capacitors before making connections or tests. Some circuits retain high voltage long after the equipment is turned off.
2. Use only a polarized 3-wire ac outlet. This assures that the power supply chassis, case, and ground terminal are connected to a good
earth ground and reduces danger from electrical shock.
3. Don’t expose high voltage needlessly. Remove housings and covers only when necessary. Turn off equipment while making test
connections in high-voltage circuits. Discharge high-voltage capacitors after removing power.
(continued on inside back cover)
3
TABLE OF CONTENTS
Page
TEST INSTRUMENT SAFETY -------------- inside front cover
INTRODUCTION -------------------------------- ----------------- 4
FEATURES -------------------------------------------------------- 5
SPECIFICATIONS ------------------------------------------------ 6
CONTROLS AND INDICATORS ------------------------------- 8
OPERATING INSTRUCTIONS -------------------------------- 10
Safety Precautions --------------------------------------------- 10 Equipment Precautions ---------------------------------------- 10 Hook-Up -------------------------------------------------------- 10 Typical Constant Voltage Operation ------------------------- 13 Setting Current Limit ------------------------------------------ 14 Typical Constant Current Operation ------------------------- 15 Constant Voltage/Constant Current Characteristic --------- 16 Connecting Two Power Supplies in Series ------------------ 16 Connecting Two Power Supplies in Parallel ---------------- 19
Page
APPLICATIONS ------------------------------------------------ 20
General -------------------------------------------------------- 20 Electronics Servicing ---------------------------------------- 20 Electronics Manufacturing ---------------------------------- 20 Electronics Design Lab -------------------------------------- 20 Electronics Education ---------------------------------------- 21 Battery Charging --------------------------------------------- 21
MAINTENANCE ----------------------------------------------- 22
Fuse Replacement -------------------------------------------- 22 Line Voltage Conversion ------------------------------------ 22 Adjustments --------------------------------------------------- 22 1745A/1744A Calibration ----------------------------------- 23 Instrument Repair Service ----------------------------------- 23
WARRANTY SERVICE INSTRUCTIONS ----------------- 25
LIMITED TWO-YEAR WARRANTY ----------------------- 26
SPANISH MANUAL ------------------------------------------- 28
INTRODUCTION
The B+K Precision Models 1744A & 1745A DC Power
Supplies are high quality, general purpose dc power sources. They provide 0-35 volts dc output, adjustable with both coarse and fine voltage controls for precise settability. The current output is 0-10 amps, adjustable with both coarse and fine current controls (model 1745A). Two large LED meters continuously monitor the output voltage and current (model 1745A). The VOLTAGE meter is green, while the CURRENT meter is red (model 1745A).
These models exhibit excellent regulation and low ripple characteristics. The circuit design incorporates a pre-regulator, which greatly reduces internal power dissipation at low output voltages. The styling is both attractive and functional.
These instruments may be used in constant voltage or constant current applications. The crossover from constant voltage to
constant current modes is smooth and automatic. LED’s indicate the “CV” (constant voltage) or “CC” (constant current) mode of
operation. In constant voltage applications, a current limit may be preset. When load variations cause the current to reach the preset limit, the unit then regulates output current rather than output voltage.
Current limits are adjustable from 5% to 100% of maximum. In constant current applications, the maximum voltage may be preset. When load variations cause current to drop below the regulated value, the unit reverts to regulated voltage operation at the preset value.
Reverse polarity protection prevents accidental damage to the power supply from improper connection to an external voltage, and current limiting protects the equipment being powered, as well as the power supply.
The output is isolated from chassis and earth ground, which permits full flexibility of connections. When needed, the (+) or (-) polarity may be strapped to ground, or either polarity may be floated to an external voltage. Two supplies may be connected in series as a 0-to-70 volt power source, or two supplies may be connected in parallel, with suitable balancing resistors, for up to twice the output current.
These power supplies are well suited for a wide variety of electrical and electronics applications, including service shops, engineering labs, production testing, school laboratories, and home use by hobbyists.
4
5
FEATURES
0-35 VOLTS
Continuously variable over 0-to-35 volts with coarse & fine controls.
0-10 AMPS
0-to-10 amps current rated for continuous duty at full output current.
Coarse and fine current adjust controls.
LABORATORY QUALITY
Excellent regulation, low ripple.
CONSTANT VOLTAGE OR CONSTANT CURRENT
Provides regulated dc voltage output or regulated dc current output.
Crossover is smooth and automatic.
LED DISPLAYS (model 1745A)
Two large, easy-to-read LED digit displays monitors output voltage
and output current, and provides good visibility in bright or low
light.
Two Current Ranges (model 1744A)
High-Low switch selects full output (0-10A), or on quarter output (0-2.2A). Low range improves current settability and meter resolution.
LED INDICATORS
Act as pilot light and identify mode of operation and metering.
PRE-REGULATOR
Limits internal dissipation for higher reliability.
ISOLATED OUTPUT
Either polarity may be floated or grounded.
OVERLOAD PROTECTION
Fully adjustable current limiting (from 5% to 100% of
maximum output current) protects circuit under test and the
power supply.
REVERSE POLARITY PROTECTION
Prevents damage to power supply from external voltage of
reverse polarity.
STYLING
Modern functional styling. Logical, convenient layout of
controls.
SPECIFICATIONS
OUTPUT VOLTAGE: 0 to 35 VDC, coarse and fine adjustment.
OUTPUT CURRENT: 0 to 10 A, coarse and fine adjustment.
CONSTANT VOLTAGE OPERATION
Voltage Regulation Line (108-132 V): 0.2% +2 mV Load (no load to full load): ≤0.04% + ≤0.04% + 2 mV
Recovery Time: ≤100 μs typical.
Ripple Voltage RMS: 1 mV.
Temperature Coefficient
(0° to +40 °C): ≤300 PPM/°C
CONSTANT CURRENT OPERATION
Adjustable Current Limits: 5% to 100% Current Regulation Line (108-132 V): ≤0.4% + 5 mA Load ≤0.4% + 5 mA Current Ripple: ≤3 mA typical.
METERING Model 1745A Voltmeter 4 digit Green LED Display Range: 0 to 99.99 V. Accuracy: (0.7% + 9 digits)
Ammeter 4 digit Red LED Display Hi Range: 0 to 9.999 A. Accuracy: (0.7% + 9 digits).
Model 1744A Voltmeter Range: 0 to 40 V Accuracy: +2.5%
Ammeter Hi Range: 0 to 11 A Lo Range: 0 to 2.2 A. Accuracy: +2.5%
AC INPUT: Domestic: 120 VAC ±10%, 60 Hz International: 120/220/230/240 VAC - 10%, 50/60 Hz
POWER CONSUMPTION: Approximately 420 W or less at
full load.
PROTECTION: Reverse polarity protection, Current limiting.
6
7
SPECIFICATIONS
NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.com for the most current
product information.
Note 1: Important: Even with noticeable Thermal Drift, this high resolution power supply will be considerably more accurate than any standard three digit display bench power supply. Thermal Drift: Since this power supply has greater resolution than standard bench power supplies they are more susceptible to Thermal Drift. Thermal Drift occurs on almost every type of power supply but is more apparent on high resolution types. Thermal Drift results in the metering of the power supply to either slowly increase or decrease with the change in the power supply’s internal temperature. As the power supply outputs more power its internal temperature will increase causing the metering (primarily the current) to slowly increase. As the power demand is deceased the power supply will cool causing the metering (primarily the current) to slowly decrease. If the power supply remains with a constant output of power for more than fifteen minutes the power supply metering will remain constant and should not continue to drift.
TEMPERATURE RANGE
Operation: 0° to +40° C, 75% R.H. Storage: -15° to +70° C, 85% R.H.
DIMENSIONS (HxWxD): 5.7” x 10.5” x 15” (145 x 267 x 38 mm)
WEIGHT: 27 lb. (12.25 kg)
ACCESSORIES SUPPLIED: Instruction Manual, Spare Fuse
8
CONTROLS AND INDICATORS
INDICATORS
Either the “CC” or “CV” and the LED indicators will be lit
whenever the unit is operating, thus serving as a pilot light. The unit automatically changes from CV to CC operation when the preset current limit is reached.
1. C.C. (Constant Current) Indicator. Red LED lights in constant
current mode. Unit regulates output current at value set by CUR­RENT controls.
2. C.V. (Constant Voltage) Indicator. Green LED lights in constant
voltage mode. Unit regulates output voltage at value set by VOLTAGE controls.
3. GREEN LED Display (1745A). 4 digit display continuously
monitors voltage.
4. RED LED Display (1745A). 4 digit display continuously
monitors current.
VOLTAGE CONTROLS
5. Coarse Control. Coarse adjustment of output voltage. Read value
on GREEN LED display.
6. Fine Control. Fine adjustment of output voltage. Read value on
GREEN LED display.
CURRENT CONTROL
7. Coarse CURRENT Control. Adjusts current limit in
constant voltage mode. Adjusts constant current value in constant current mode. Current can be read from RED LED display.
8. Fine Control (1745A). Fine adjustment of current limit
or current value. Current can be read from RED LED display.
OUTPUT TERMINALS
9. “+” Terminal (Red). Positive polarity output terminal.
10. GND Terminal. Earth and chassis ground.
11. “-” Terminal (Black). Negative polarity output
terminal.
REAR PANEL CONTROLS
12. Fuse
13. Power Cord
9
CONTROLS AND INDICATORS
Figure 1. Front Panel Controls and Indicators.
Figure 2. Rear Panel
REAR PANEL CONTROLS
7. Fuse.
8. Power Cord.
9. 110/220 Line Voltage Conversion Switch.