The following safety precautions apply to both operating and maintenance personnel and
must be observed during all phases of operation, service, and repair of this instrument.
Before applying power, follow the installation instructions and become familiar with the
operating instructions for this instrument.
GROUND THE INSTRUMENT
To minimize shock hazard, the instrument chassis and cabinet must be
connected to an electrical ground. This instrument is grounded through the
ground conductor of the supplied, three-conductor ac power cable. The power
cable must be plugged into an approved three-conductor electrical outlet. Do not
alter the ground connection. Without the protective ground connection, all
accessible conductive parts (including control knobs) can render an electric
shock. The power jack and mating plug of the power cable meet IEC safety
standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
KEEP AWAY FROM LIVE CIRCUITS
Instrument covers must not be removed by operating personnel. Component
replacement and internal adjustments must be made by qualified maintenance
personnel. Disconnect the power cord before removing the instrument covers
and replacing components. Under certain conditions, even with the power cable
removed, dangerous voltages may exist. To avoid injuries, always disconnect
power and discharge circuits before touching them.
DO NOT SERVICE OR ADJUST ALONE
Do not attempt any internal service or adjustment unless another person,
capable of rendering first aid and resuscitation, is present.
DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT
Do not install substitute parts or perform any unauthorized modifications to this
instrument. Return the instrument to B&K Precision for service and repair to
ensure that safety features are maintained.
1
WARNINGS AND CAUTIONS
WARNING and CAUTION statements, such as the following examples, denote a
hazard and appear throughout this manual. Follow all instructions contained in
these statements.
A WARNING statement calls attention to an operating procedure, practice, or
condition, which, if not followed correctly, could result in injury or death to
personnel.
A CAUTION statement calls attention to an operating procedure, practice, or
condition, which, if not followed correctly, could result in damage to or
destruction of parts or the entire product.
WARNING:
WARNING:
CAUTION:
CAUTION:
Do not alter the ground connection. Without the protective
ground connection, all accessible conductive parts (including
control knobs) can render an electric shock. The power jack and
mating plug of the power cable meet IEC safety standards.
To avoid electrical shock hazard, disconnect power cord before
removing covers. Refer servicing to qualified personnel.
Before connecting the line cord to the AC mains, check the rear
panel AC line voltage indicator. Applying a line voltage other
than the indicated voltage can destroy the AC line fuses. For
continued fire protection, replace fuses only with those of the
specified voltage and current ratings.
This product uses components which can be damaged by electrostatic discharge (ESD). To avoid damage, be sure to follow
proper procedures for handling, storing and transporting parts
and subassemblies which contain ESD-sensitive components.
SAFETY SYMBOLS
Connect to safety earth ground using the wire recommended in the user
manual.
This symbol on an instrument indicates that the user should refer to the
operating instructions located in the manual.
Certification
We certify that this product met its published specifications at time of shipment from the
factory.
2
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and
other European countries with separate collection systems)
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on waste
electrical and electronic equipment (WEEE), and in jurisdictions
adopting that Directive, is marked as being put on the market
after August 13, 2005, and should not be disposed of as
unsorted municipal waste. Please utilize your local WEEE
collection facilities in the disposition of this product and
The B&K Precision model 1737 is a general purpose dual range DC power source.
This power supply can output more voltage at a lower current or more current at
a lower voltage. The 1737 provides 0-60 V DC output, adjustable with both
coarse and fine voltage controls for precise setting. The current output for the
1737 is 0-3 A for the 0-30 V range, and 0-2 A for the 30-60 V range, adjustable
with both coarse and fine current controls.
The 1737 exhibits excellent regulation and low ripple characteristics. Its
mechanical configuration conserves bench space and allows for easy portability.
Additionally, the power supply can be operated locally from the front panel or
remotely through the RS-232 interface.
This power supply is well suited for a wide variety of electrical and electronics
applications in service shops, engineering labs, production facilities, school
laboratories, and home use.
Features
• Low ripple and noise
• Excellent regulation
• Constant voltage (CV) and constant current (CC) operation
• Two 4-digit LED displays provide good visibility in bright or low light
• LED indication for CV and CC modes
• Automatic recall of last settings on power up
• RS-232 interface
• Isolated output
• Overload protection
• Reverse polarity protection
5
3 Quick Reference
7 8 6 5 9 1 2
3 4 10
11
12
3.1 Front Panel
Figure 1 - Front Panel
INDICATORS
Either the CV or CC LED indicators will be lit whenever the unit is operating. The unit
automatically changes from CV to CC operation when the preset current limit is
reached.
1) CV (Constant Voltage) Indicator. The green LED is lit in constant voltage mode.
Unit regulates output voltage at the value set by voltage controls.
2) CC (Constant Current) Indicator. The red LED is lit in constant current mode. Unit
regulates output current at the value set by current controls.
3) Green LED Display. 4-digit display continuously monitors voltage.
4) Red LED Display. 4-digit display continuously monitors current.
VOLTAGE CONTROLS
5) Voltage Coarse Control. Coarse adjustment of output voltage. Read value on the
green LED display.
6) Voltage Fine Control. Fine adjustment of output voltage. Read value on the
green LED display.
6
CURRENT CONTROL
15
14
13
7) Current Coarse Control. Coarse adjustment of current limit. Read value on the
red LED display.
8) Current Fine Control. Fine adjustment of current limit. Read value on the red LED
Use only a polarized 3-wire AC outlet. This assures that the power supply chassis,
case, and ground terminal are connected to a good earth ground and reduces
danger from electrical shock.
There is little danger of electrical shock from the power supply output, which
produces a maximum of 60 volts dc. However, there may be great danger of
electrical shock if the power supply output is connected to an external high
voltage. Some equipment being powered may contain high voltage and present a
shock hazard. Observe caution. If the power supply output is floated (referenced
to a voltage rather than earth ground) turn off the power supply and the
equipment under test when making connections. Never float the power supply to
a potential greater than 100 volts peak with respect to earth ground.
Equipment Precautions
Avoid using the power supply in ambient temperatures above +40° C. Always
allow sufficient air space around the heat sink at the rear of the power supply for
effective radiation to prevent internal heat build-up.
Although the power supply is protected against reverse polarity damage, the
circuit being powered may not include such protection. Always carefully observe
polarity; incorrect polarity may damage the equipment under test.
Do not exceed the voltage rating of the circuit being powered. Many transistors
and integrated circuits will not withstand voltage of 60 volts.
There is no need to worry about voltage spikes or overshoot damaging the
equipment under test. The voltage between the output terminals of the power
supply never exceeds the preset value as the POWER switch is turned on or off.
8
4.1 Instrument Hook-Up
1. Turn off the power supply and the equipment to be powered during hook-
up.
2. Connect the positive polarity of the device being powered to the red (+)
terminal of the power supply.
3. Connect the negative polarity of the device being powered to the black (-)
terminal of the power supply.
Figure 3 illustrates the grounding possibilities.
4.
a. If the negative polarity of the equipment or circuit being powered is also
the chassis or common, it may be grounded to earth by strapping the black
(-) terminal to the green (
b. Similarly, the positive polarity can be grounded by strapping the red (+)
terminal to the green (
c. If an earth ground reference is not required, the configuration of
may be used. The scheme in
known whether the chassis is common with either the positive or negative
polarity.
d. If the chassis or common of the equipment being powered is separate
from both the positive and negative polarity power inputs, use the
connection shown in
6. Observe proper polarity. If the circuit being powered is not equipped with
reverse polarity protection, damage to the circuit can result from reverse
polarity. Use color coded hook-up leads, for convenience in identifying
polarity, red for (+) and black for (-).
7. Make sure that the hook-up leads offer sufficient current capability and low
resistance between the power supply and the circuits being powered.
) terminal as shown in Figure 3A.
) terminal as shown in Figure 3B.
Figure 3C
Figure 3C should also be used where it is not
Figure 3D.
9
Figure 3 (A and B) - Grounding Possibilities
Figure 3 (C and D) - Grounding Possibilities
10
4.2 Typical Constant Voltage Operation
1. Before connecting the device to be powered to the power supply, determine
the maximum safe load current for the device to be powered and set the
current limit value (see Setting Current Limit procedure in this section).
2. The output voltage can be set using the voltage coarse knob (with a 1 V
resolution) and voltage fine knob (with a 10 mV resolution). After the last
adjustment you must wait 3 seconds until the state is saved.
3. Turn off power supply and connect it to the device to be powered (see
Instrument Hook-Up procedure in this section).
4. Turn on the power switch. The CV indicator should light.
5. Increase the voltage setting until the voltage LED display reads the desired
value. The fine control permits easier setting to a specific value.
6. Note the load current on the current LED display.
7. If the load current exceeds the preset current limit, the CV indicator will go
off and the CC indicator will light. In this case, the power supply
automatically switches to the constant current mode, and further rotation of
the voltage control will not increase the output voltage.
Figure 4 - Typical Constant Voltage Operation
11
4.3 Setting Current Limit
1. Determine the maximum safe current for the device to be powered.
2. Temporarily short the (+) and (-) terminals of the power supply together
with a test lead.
3. Adjust the coarse and fine current control for the desired current limit. Read
the current value on the current LED display.
4. The current limit (overload protection) has now been preset. Do not change
the current controls settings after this step.
5. Remove the short between the (+) and (-) terminals and hook up for constant
voltage operation.
Figure 5 - Setting Current Limit
12
4.4 Typical Constant Current Operation
1. Before connecting the device to be powered to the power supply, determine the
maximum safe voltage to be applied, and set the voltage controls to obtain that
voltage reading on the voltage LED display.
2. Determine the desired constant current value.
3. Set the coarse and fine current control to minimum (1 mA). Wait 3 seconds.
4. Turn off the power supply and connect it to the device to be powered.
5. Turn on the power supply. The CC indicator should light.
6. Increase the coarse and fine current control setting until the desired constant
current value is read on the display, or set the current limit in advance (before
connecting the load) as described earlier in the Setting Current Limit procedure.
7. If the load current drops below the constant current value, the CC indicator will go
off and the CV indicator will light. In this case, the power supply automatically
switches to the constant voltage mode, and further rotation of the current
controls will not increase the output current.
Figure 6 - Typical Constant Current Operation
13
4.5 Constant Voltage/Constant Current Characteristic
The working characteristic of this power supply is called a constant voltage/constant
current automatic crossover type. This permits continuous transition from constant
current to constant voltage modes in response to the load change. The intersection
of constant voltage and constant current modes is called the crossover point.
7
shows the relationship between this crossover point and the load.
For example, if the load is such that the power supply is operating in the constant
voltage mode, a regulated output voltage is provided. The output voltage remains
constant as the load decreases, up until the point where the preset current limit is
reached. At that point, the output current becomes constant and the output voltage
drops in proportion to further decreases in load. The crossover point is indicated by
the front panel LED indicators. The crossover point is reached when the CV indicator
goes off and the CC indicator comes on.
Similarly, crossover from the constant current to the constant voltage mode
automatically occurs from an increase in load. A good example of this would be seen
when charging a 12-volt battery. Initially, the open circuit voltage of the power
supply may be preset for 13.8 volts. A low battery will have a low load on the supply
and it will operate in the constant current mode, which may be adjusted for a 1 amp
charging rate. As the battery becomes charged, and its voltage approaches 13.8 volts,
its load increases to the point where it no longer demands the full 1 amp charging
rate. This is the crossover point where the power supply goes into the constant
voltage mode.
Figure
Figure 7 - Constant Voltage/Constant Current Characteristic
14
4.6 Saving the Power Supply’s Current State
The current state of the power supply is saved after 2 seconds from the last
adjustment or after the “SAVE” command is received through the RS232 interface.
4.7 Connecting Two Power Supplies in Series
Two power supplies may be connected in series to provide a variable 0-120 V output.
In this configuration, the two power supplies can supply up to 60 V-3 A or 120 V-2 A.
Figure 8 for the connection scheme.
See
When connected in series, the voltage controls of each power supply exercise control
over a 0-60 V range. Add the LED display readings together or connect an external
voltmeter across the load to determine the total output voltage.
Load current may be monitored from either supply; the readings will be identical
since they are connected in series. Also, since the supplies are connected in series, it
is only necessary to set the current limit on one of the supplies; the other may be set
for maximum.
Figure 8 - Connecting Two Power Supplies in Series
15
4.8 Connecting Two Power Supplies in Parallel
Two power supplies may be connected in parallel to double the maximum load
current. In this configuration, the two power supplies will provide 0-60 V output
at up to 4 A or 0-30 V at 6 A (heavier gauge hook-up leads are advisable). Current
equalizing resistors must be used as shown in
current limiting feature will prevent damage if current is temporarily unbalanced
during set-up.
When connected in parallel and operating in the constant voltage mode,
determine the total load current limit and preset the current limiting for each
power supply to half the total load current value. Then when the load is
connected, set the voltage controls on the two power supplies for equal voltage
readings. This should also provide approximately equal current from each supply.
Add the two current meter readings together for total load current, or connect
an external ammeter in series with the load.
Figure 9. However, the protective
Figure 9 - Connecting Two Power Supplies in Parallel
16
If the current equalizing resistors are not well matched, it is preferable that the
2
3
5
RX
TX
GND
DB-9
connector
Instrument
2
3
5
RX
TX
GND
PC
DB-9
connector
voltages be slightly unbalanced to achieve current balance. Be sure that the
supplies are adequately balanced so that both remain in the CV mode.
When connected in parallel and operating in the constant current mode, the
voltage controls of both supplies should be preset to the same value. Then when
the load is connected, the current controls of the two supplies should be
adjusted for approximately equal current from each unit. Be sure that both
supplies remain in the CC mode.
5 Remote Operation
The power supply can be connected to the RS232 interface using the 9-pin (DB-9)
serial connector on the rear panel. The 1737 uses a null modem RS232 cable. The
cable pin diagram for the DB-9 connector is shown in the picture below:
Figure 10 - RS232 Cable Pin Diagram
For all communications over the RS232 interface, the power supply uses Xon/Xoff
protocol.
NOTE: If using HyperTerminal, make sure you properly configure the “ASCII Setup”. In
“ASCII Setup”, do not append line feeds to “ASCII Sending”. “Echo typed characters
locally” can be checked, but “Send line ends with line feeds” should be left
unchecked. In “ASCII Receiving”, “Append line feeds to incoming line ends” can also
be checked.
17
5.1 RS232 Configuration Overview
Baud Rate 9600 bps
Parity Bits none
Data Bits 8
Start Bits 1
Stop Bits 1
Flow Control Xon / Xoff
Termination Character \r
Table 1 - RS232 Configuration
5.2 RS232 Commands
1.VOLT xx.xx<CR>
This command sets the output voltage value. The output voltage takes the
programmed value only if the power supply is in Constant Voltage mode.
Example: The output voltage is set at 5 V.
VOLT 05.00<CR>
2. CURR x.xxx<CR>
This command sets the output current value. The output current takes the
programmed value only if the power supply is in Constant Current mode.
Example: The output current is set at 300 mA.
CURR 0.300<CR>
18
3. SAVE<CR>
The programmed values of the output current and voltage are saved. If the
power supply is restarted, the values of the output current and voltage will
be those previously saved.
4. VOLT?<CR>
This command returns the voltage measured to the output terminals of the
power supply: xx.xxV<CR>
The voltage value returned is the same as the one displayed on the voltage
display.
5. CURR?<CR>
This command returns the current measured to the output terminals of the
power supply: x.xxxA<CR>
The current value returned is the same as the one displayed on the current
display.
6. STAT?<CR>
The power supply responds to this command with one of the following
messages:
CV<CR> if the power supply is in Constant Voltage mode
CC<CR> if the power supply is in Constant Current mode
5.3 Application Software
Remotely control your 1737 using B&K Precision’s application software, available for
download at
a simple data logging function to store and log data to a text or CSV file. RS232
commands can also be sent through the software.
www.bkprecision.com. This software provides a virtual front panel and
19
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.