Bistos BT-500 User Manual

4.5 (11)

BT-500

Infant Incubator Operation Manual

Keep this manual for future reference

P/N : 500-ENG-OPM-EUR-R12

Information and descriptions contained in this manual are the property of Bistos Co., Ltd. and may not be copied, reproduced, disseminated, or distributed without express written permission from Bistos Co., Ltd

Information furnished by Bistos Co., Ltd is believed to be accurate and reliable. However, no responsibility is assumed by Bistos for its use, or any infringements of patents or other rights of third parties that may result from its use. No license is granted by implication or otherwise under any patent or patent rights of Bistos.

Revision 12

July, 2019

Copyright © Bistos Co., Ltd. 2018. All rights reserved.

BT-500 Operation Manual

1

Table of Contents

1Safety

1.1Instructions for the safe operation and use of the device

1.2Warnings

1.3Shock hazard

1.4General precautions on environment

1.5Symbols

2Introduction

2.1General

2.2Brief device description

2.3Intended use

2.4Operating principles

2.5Essential performance

2.6Accessaries and options

2.7Appearance of device

2.7.1Front view

2.7.2Front view detail

2.7.3Rear view

2.7.4Side view

2.8Description of each part

2.8.1Control shell

2.8.2Hood

2.8.3Mattress tray

2.8.4Stand

3Install and connection

3.1IV pole assembly

3.1.1IV external monitor

3.1.2IV plate

3.1.3IV ringer pole

3.2Air filter assembly

3.3Connection of power and cable

3.3.1Power connection

3.3.2Cable connection

3.3.3Sensor module connection

3.4Placement of infant

3.5Movement and installation

4Operation

4.1System start-up

4.2LED

4.3Key and knob operation

4.3.1Key

4.3.2Knob

4.4Displays

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

2

4.5Temperature measurement and control

4.6Humidity Measurement and control

4.7O2 measurement

4.8Weighing scale measurement

4.9SpO2 and Pulse rate measurement

4.10Menu function

4.11Pulse oximeter

4.12External monitor

4.13Shut down

4.14Calibration of O2 module

5Alarms

5.1System alarms

5.2Temperature alarms

5.3Humidity alarms

5.4Oxygen alarms

5.5Weighing scale alarms

5.6Alarm self-test

6Cleaning & Maintenance

6.1General cleaning method and precautions

6.2Hood

6.3Shell, sensor module, scale module and basket

6.4Water tank

6.5Skin temperature sensor and SpO2 sensor

6.6Drain of residual water

6.7Regular inspection

6.8Battery replacement and disposal

6.9Disposal of the device

7Specification

8Troubleshooting

8.1General checking

8.2Alarm message checking

9Declaration on EMC

9.1Electromagnetic emissions

9.2Recomended separation distances between portable and mobile RF communications equipment and the device

9.3Electromagnetic immunity

Warranty

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

3

1 Safety

1.1Instructions for the Safe Operation and Use of the device

Examine the incubator and any accessories periodically to ensure that the cables, line cords and instruments do not have visible evidence of damage that may affect patient safety or performance. The recommended inspection interval is once per week or less. Do not use the incubator if there is any visible sign of damage.

Only the AC line cord supplied with the BT-500 is approved for use with the Unit.

Do not attempt to service the BT-500 incubator. Only qualified service personnel by Bistos Co., Ltd. should attempt any needed internal servicing.

The BT-500 is not specified or intended for operation during the use of defibrillators or during defibrillator discharge.

The BT-500 is not specified or intended for operation in the presence of electrosurgical equipment.

The BT-500 is not specified or intended for operation in conjunction with any other type of equipment except the specific devices that have been identified for use in this Operator’s Manual.

Perform periodic safety testing to insure proper patient safety. This should include leakage current measurement and insulation testing. The recommended testing interval is once per year.

Do not operate the BT-500 incubator if it fails to pass the power on self-test procedure.

WARNING

CAUTION

SHOCK HAZARD

Be informed that it may cause serious injury or death to the patient, property damage, material losses against the “WARNING” sign.

Be informed that it may cause no harm in life but lead to injury against the “CAUTION” sign.

Be informed that it may cause serious electrical shock to the patient or operator, property damage, material losses against the “SHOCK HAZARD” sign.

1.2Warnings

WARNING Thoroughly read and understand the manual prior to use of the incubator. Failure to do so could result in personal injury or equipment damage.)

Incubator misuse may result in harm to an infant. Only properly trained personnel should use the incubator as directed by an appropriately qualified attending physician aware of currently known risks and benefits.

Use of accessories other than those listed and approved for use in this product as original or replacement items may result in increased emissions or decreased immunity.

The total electrical current leakage of all items powered through the incubator, including devices on the outlet strip, must be less than 300uA for 120V AC/ 100V AC systems and less than 500uA for 230V AC systems.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

4

Otherwise, personal injury or equipment damage could occur.

The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consider the use of the accessory in the patient’s vicinity and evidence that the safety certifications of the accessory have been performed in accordance with the appropriate International Electrotechnical commission (IEC) 60601-1 harmonized national standard. Personal injury or equipment damage could occur.

Devices connecting to the serial data port must be compliant with EN 60601-1-2, the EMC requirement for Medical Devices. Failure to do so could result in personal injury or equipment damage.

Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. In addition, portable and mobile RF communications equipment can effect medical electrical equipment.

The equipment shall not be used adjacent to or stack with other devices unless verification of normal operation in the configuration in which it is to be used can be achieved.

Use only Bistos recommended fuel cells for proper operation. Failure to do so could result in personal injury or equipment damage.

Higher incubator relative humidity at any given temperature decreases an infant’s evaporative heat loss, and may cause an increase in the infant temperature. Routinely monitor the infant’s rectal and/or axillary temperature according to the attending physician’s orders or Nursery Standing Orders. Failure to do so could result in personal injury.

Higher relative humidity will, at any given time, decrease an infant’s evaporative water loss, and may cause an increase in infant temperature. This effect is greatest n very low birth-weight, premature infants. The attending physician should prescribe Temperature Control mode, temperature setting, and humidity output level setting. Routinely monitor the infant’s rectal and/or axillary temperature according to the attending physician’s orders or Nursery Standing Orders. Failure to do so could result in personal injury.

Fill the reservoir to the Maximum Filing Limit line. Do not overfill. Water spillage may result, and personal injury could occur.

Use distilled water only (<10 ppm total dissolved solids). The use of sterile water is not acceptable. Equipment damage could occur.

For proper operation of the incubator, use only skin temperature probes from Bistos Co. Ltd. Using other probes could result in personal injury or equipment damage.

Never place the skin temperature probe under the infant or use it rectally.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

5

Personal injury could occur.

BT-500 cannot differentiate between an increase in core temperature with a cold skin (fever) and a low core and skin temperature (hypothermia). The temperature of the infant is monitored separately.

When in skin mode, the skin temperature probe must be in direct contact with the skin to provide accurate monitoring of the infant’s skin temperature. When in skin mode, failure to maintain direct skin contact can result in overheating. Routinely check the infant’s condition for correct sensor attachment, and feel the infant’s skin for signs of overheating.

When an x-ray is taken through the hood, the hood could show up on the x- ray as a radiolucent shadow and could result in incorrect diagnosis.

Do not use in the presence of flammable anesthetics. Personal injury or equipment damage could occur.

Keep matches, and all other sources of ignition, out of the room in which the incubator is located. Textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen. Personal injury or equipment damage could occur.

Small quantities of flammable agents, such as ethyls and alcohol, left in the incubator may cause a fire in connection with oxygen. Personal injury or equipment damage could occur.

A fire and explosion hazard exists when performing cleaning or maintenance procedures in an oxygen-enriched environment. Make sure that oxygen supply is turned Off and the oxygen hose to the incubator is disconnected when performing cleaning and maintenance procedures. Turn off or disconnect oxygen supplies during periods of non-use. Failure to do so could result in personal injury or equipment damage.

If it is necessary to administer oxygen in an emergency, notify the attending physician immediately. Failure to do so could result in personal injury or equipment damage.

Improper use of supplemental oxygen may be associated with serious side effects including blindness, brain damage, and death. The risks vary with each infant. The qualified attending physician should prescribe the method, the concentration, and the duration of oxygen administration.

Administration of oxygen may increase the noise level for the infant within the infant incubator.

An oxygen analyzer shall be used separately when oxygen is delivered to the infant.

Measure the oxygen concentrations to verify delivery of the prescribed oxygen concentration. Failure to do so could result in personal injury or equipment damage.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

6

If the patient’s arterial oxygen levels cannot be maintained when the oxygen control setting is set to maximum, the attending physician should prescribe alternate means of oxygenation. Failure to do so could result in personal injury or equipment damage.

The oxygen concentration inspired by an infant does not accurately

determine the partial pressure of oxygen(pO2) in the blood. When deemed advisable by the attending physician, measure blood pO2 by accepted clinical techniques. Failure to do so could result in personal injury or equipment damage.

Disconnect the incubator from the hospital oxygen source when oxygen is not in use. Failure to do so could result in personal injury or equipment damage.

As oxygen use increases the danger of fire, do not place auxiliary equipment that produces sparks in an incubator. Personal injury or equipment damage could occur.

Use of anesthetic agents can interfere with oxygen analyzer accuracy.

Inspect gas/oxygen service components at regular service intervals for signs of corrosion or damage. Failure to do so could result in personal injury or equipment damage.

A dirty air intake micro filter could affect performance or cause carbon dioxide(CO2) build-up. Ensure that the filter is checked on a routine basis commensurate with local conditions. Particularly, if the unit is used in an unusually dusty environment, more frequent replacements may be necessary. Failure to do so could result in infant injury or equipment damage.

After each change of oxygen flow, allow at least 30 min to achieve new concentrations. Failure to do so could result in personal injury or equipment damage.

Compressed gas cylinders, such as oxygen cylinders, can become hazardous projectiles if the gas is released rapidly due to damage or other causes. Securely fasten the cylinder. Failure to do so could result in personal injury or equipment damage.

Oxygen levels within the incubator hood environment may be affected when the access doors or access panels are opened. Make sure all hood access door gaskets and tubing ports are properly installed. Any open gaps in the incubator hood may reduce the incubator’s internal oxygen. Personal injury could occur.

Make sure all hood access door gaskets and tubing ports are properly installed. Any open gaps in the incubator hood will reduce the incubator’s internal relative humidity. Personal injury or equipment damage could occur.

The use of infant seats, or other accessories within the incubator that can

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

7

alter the airflow pattern, may affect temperature uniformity, temperature variability, the correlation of the incubator temperature reading to center mattress temperature and infant skin temperature. Personal injury could occur.

Phototherapy units located too close to the incubator may affect hood wall temperature, incubator hood temperature, and infant skin temperature. Personal injury of equipment damage could occur.

Phototherapy lamps placed over the top of the incubator hood may interfere with upward travel of the vertical height adjustable stand. To prevent this interference, always remove the phototherapy lamp prior to positioning the stand.

If airflow passages are not kept clear of obstructions, such as blankets and stuffed animals, during clinical usage, patient safety and incubator performance may be compromised.

To avoid overheating the infant due to direct radiation, do not position the incubator in direct sunlight or under other sources of radiant heat.

Do not place surgical covers or blankets over the infant and warm air curtain of side vents simultaneously. This may cause heat-induced injury and burns.

To prevent accidental disconnection, secure all patient leads, infusion lines, and ventilator tubing to the mattress with sufficient excess length to allow for the full range of mattress height adjustment.

Only connect equipment to the serial port that complies with the relevant IEC standard; and use data cables with plastic body connectors.

Do not raise the hood at any time while the infant is in the incubator. Gain access to the infant by the access panels and access doors. Failure to do so could result in personal injury or equipment damage.

When the front access panel (or optional rear) is open, the temperature display may not accurately reflect the incubator temperature. Do not leave the front access panel (or optional rear) open longer than essential. Personal injury could occur.

Positively secure all access panel latches to avoid accidental opening, Failure to do so could result in personal injury or equipment damage.

For infant safety, do not leave the infant unattended when the access panels are open. Personal injury could occur.

Always use two people when moving the incubator and patient together. When moving the incubator within the same floor space, check that the patient is secured safely in the unit and either remove or secure all loose system components to prevent possible patient injury or equipment damage. If the move involves varying floor heights or a complete floor level change (i.e. thresholds, ramps, elevators), remove all items either not being used or not necessary for the move, lower the VHA, IV poles and shelves to

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

8

their lowest position, place all drawers in their locked state, and remove all accessories from the front and rear rail position.

Never place objects taller than the top of the wheel casters beneath the incubator stand. Placement of objects there could interfere with the stability of the vertical height adjustable stand. Personal injury of equipment damage could occur.

To avoid possible tip-over or damage to adjacent carts, IV stands, shelves, etc., keep at least a 12” (30 cm) perimeter area clear around the vertical height adjustable stand.

For optimum incubator stability, always lock all stand wheels, Do not leave the unit unattended when parking on an incline. Failure to do so could result in personal injury or equipment damage.

When raising or lowering the incubator, the operator should ensure that both equipment and appendages are clear of the unit’s travel path. Patient and incubator connections must also be checked before adjusting the incubator height. Never place any objects on top of the drawer assembly and always check before lowering the VH that there is sufficient clearance between the incubator and stand assembly. Do not raise or lower the unit while installing or removing medical gas tanks from the tank holder assembly. Failure to do so could result in personal injury of equipment damage.

The UART port is for debugging purposes only. It does not allow connections with other devices.

Prior to placing the infant in the incubator, pre-warm the incubator to the temperature prescribed by the attending physician, or according to nursing protocol.

Only one monitor shelf should be used per incubator. When using the monitor shelf, always place the monitor in the center of the shelf, ensure that the monitor fits within the border of the shelf, and avoid stacking monitors on the shelf. Personal injury or equipment damage could occur.

Attach the incubator to the stand or the vertical height adjustable stand using the bolts provided. Failure to do so could result in the incubator separating from the stand if sufficiently tilted, particularly with the hood open. Personal injury or equipment damage could occur.

This product has been validated with the accessories and options listed in this manual and found to comply with all relevant safety and performance requirements applicable to the device. It is therefore the responsibility of that person or organization who makes an unauthorized modification, or incorporates an unapproved attachment to the device, to ensure that the system still complies with those requirements.

A pulse oximeter should NOT be used as an apnea monitor

Pulse rate measurement is based on the optical detection of a peripheral

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

9

flow pulse and therefore may not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or substitute for ECG based arrhythmia analysis.

A pulse oximeter is an early warning device. Use lab co-oximeter to completely understand the patient’s condition.

Do not use the MS board pulse oximeter in the presence of flammable anesthetics or other flammable substance in combination with air, oxygen enriched environments, or nitrous oxide.

Do not remove the monitor cover except to replace the battery. An operator may only perform maintenance procedures specifically described in this manual. Refer servicing to Masimo in repair of this equipment.

Leakage current must not exceed 100 microamperes; measure when an external device is connected to the serial port.

Do not use Masimo oximetry sensors during MRI scanning as it could potentially cause burns.

Inaccurate measurements may be caused by incorrect application or use.

Inaccurate measurements may be caused by significant levels of dysfunctional hemoglobin (HbCO or MetHb).

Inaccurate measurements may be caused by intravascular dyes such as indocyanine green or methylene blue.

Inaccurate measurements or loss of pulse signal may be caused by excessive illumination.

Inaccurate measurements may be caused by excessive patient movement.

Inaccurate measurements may be caused by venous pulsation.

Inaccurate measurements or loss of pulse signal may be caused by placement of a sensor on an extremity with a blood pressure cuff, arterial catheter or intravascular line.

The MS board pulse oximeter can be used during defibrillation, but the readings may be inaccurate for a short time.

Loss of pulse signal can occur when the sensor is too tight.

Loss of pulse signal can occur when the patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.

Loss of pulse signal can occur when there is arterial occlusion proximal to the sensor.

Loss of pulse signal can occur when the patient is in cardiac arrest or is in shock.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

10

Use only Masimo sensors for SpO2 measurements.

Tissue damage can occur due to incorrect placement of sensor.

1.3Shock hazards

SHOCK

Unplug the unit from its power source prior to cleaning or maintenance. For

HAZARD

 

units equipped with an uninterruptible power supply(UPS) system, also

 

 

remove the battery pack prior to cleaning or maintenance. Failure to do so

 

 

could result in personal injury or equipment damage.

Some chemical cleaning agents may be conductive and leave a residue that may permit a build-up of conductive dust or dirt. Do not allow cleaning agents to contact electrical components, and do not spray cleaning solutions onto any of these surfaces. Personal injury or equipment damage could occur.

To ensure grounding reliability, plug the AC power cord only into a properly grounded 3-wire hospital-grade or hospital-use outlet. Do not use extension cords. If any doubt exists as to the grounding connection, do not operate the equipment. Personal injury or equipment damage could occur.

Do not expose the unit to excessive moisture that would allow for liquid pooling. Personal injury or equipment damage could occur.

Due to the risk of electrical shock hazard, only qualified personnel with appropriate service documentation should service the unit.

Batteries can present a risk of electric shock. The following precautions should be taken when working on batteries: remove watches, rings or other metal objects; use tools with insulated handles.

The total power of all equipment connected to the convenience outlet strip on the pedestal / stand must me within the electrical requirements shown on the rear of the pedestal / stand. Otherwise, personal injury or equipment damage could occur.

Make sure the Building power source is compatible with the electrical specifications shown on the column of the pedestal / stand and on the incubator. Failure to do so could result in personal injury or equipment damage.

To prevent equipment damage or accidental power disconnections, do not plug an incubator power cord directly to an AC wall socket when the incubator is mounted on a pedestal /stand. Always provide power to the incubator by using the power cord coming directly from the pedestal /stand.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

11

1.4 General precaution on environment

Do not keep or operate the equipment under the environment listed below.

 

Avoid placing in an area

 

 

 

exposed to moisture. Do not

 

Avoid exposure to direct

 

touch the equipment with

 

sunlight

 

wet hand.

 

 

 

 

 

 

 

Avoid placing in an area

 

 

 

where there is a high

 

 

 

variation of temperature.

 

Avoid in the vicinity of Electric

 

Operating temperature

 

 

 

heater

 

ranges from 20°C to 30°C.

 

 

 

 

 

Operating humidity ranges

 

 

 

from 0% to 95%.

 

 

 

 

 

 

 

Avoid placing in an area

 

Avoid placing in an area where

 

where there is an excessive

 

 

 

there is an excessive shock or

 

humidity rise or ventilation

 

 

 

vibration.

 

problem.

 

 

 

 

 

 

 

 

 

Avoid placing in an area

 

Avoid dust and especially

 

where chemicals are stored

 

 

 

metal material into the

 

or where there is in danger

 

 

 

equipment.

 

of gas leakage.

 

 

 

 

 

 

 

 

 

Do not disjoint or

 

Power off when the

 

 

equipment is not fully

 

disassemble the equipment.

 

 

 

installed.

 

BISTOS Co., Ltd. does not

 

 

 

Otherwise, the equipment

 

take responsibility of it.

 

 

 

could be damaged.

 

 

 

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

12

1.5 Symbols

Symbol

Description

 

 

Used to identify safety information. Be well-known this information thoroughly before using

 

BT-500. During the operation, do not disconnect any cable.

 

 

Indicate the warning for hot surface.

 

 

 

 

 

Type BF Applied part

 

 

 

 

 

Refer to operation manual. Read manual before placing the device.

 

 

 

 

Skin temperature sensor #1, to be connected to infant’s abdomen for baby (controlled) mode

 

 

 

Skin temperature sensor #2, to be connected to other than infant abdomen

 

 

 

IPX0

IPX0 Non-protected against ingress of water with harmful effects.

(Device)

 

 

IPX1

IPX1 Protected against the vertically dripping water (Skin temperature sensor_2EA)

 

 

 

IPX2

IPX2 Protected against the dripping water (SpO2 sensor)

 

 

 

 

IPX6

IPX6 Protected against the powerful jetting (Foot switch_2EA)

 

 

 

 

 

Indicates the weight limit

 

 

 

 

 

This symbol indicates the manufacturer.

 

 

 

 

 

This symbol indicates the serial number of the device.

 

 

 

 

This symbol indicates the authorized representative in the European Community of

 

manufacturer.

 

 

 

 

 

This symbol indicates to keep the device dry.

 

 

 

 

This symbol indicates to keep the correct upright position on the transport package.

 

 

 

 

This symbol indicates the device is fragile.

 

 

 

 

This symbol indicates the temperature limitation for operation, transport and storage.

 

 

 

This symbol indicates the humidity limitation for operation, transport and storage.

 

 

 

 

This symbol indicates the packing material is recyclable.

 

 

 

 

 

External Signal IN/OUT Port

 

 

 

 

This symbol indicates the compliance with the essential requirements and provisions of the

 

Medical Device Directive 93/42/EEC as amended by 2007/47/EEC.

 

 

 

 

This symbol indicates to not dispose the device together with unsorted municipal waste(for

 

EU only). The solid bar symbol indicates that mains adapter is put on the market after 13

 

August 2005.

 

 

 

 

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

13

2 Introduction

2.1 General

This chapter provides a general description of the BT-500 infant incubator including.

Brief Device Description

Product Features

Model Configurations

2.2Brief Device Description

A mains electricity (AC-powered) unit designed to provide an enclosed controlled environment to maintain appropriate temperature and humidity levels mainly for premature infants and other newborns who cannot effectively regulate their body temperature. It typically consists of a clear removable plastic hood with a mattress. It typically includes a means to warm the infant such as providing heated air; temperature controls that work automatically either by measuring the air temperature or through a temperature sensor attached to the infant skin; and humidity controls. The device is intended to use in a hospital.

2.3 Intended Use

BT-500 is an infant incubator for non-invasively measuring and showing graphically humidity, air temperature, skin temperature, O2 Module, weight and SpO2. This data is intended to aid the maintaining life of a premature baby or a precocious baby under 2kgs. This device is for use only by trained medical personnel located in hospital. Also this device can be used in the all departments of the hospital which offers a neonatal care service such as NICU(Neonatal Intensive Care Unit), special nursery unit and pediatrics.

2.4Operating Principles

Air and skin temperature measurement and Control: Internal cartridge heater raises the temperature. The infant environmental temperature value is determined by air or skin temperature that is measured by the sensor module. Through the main fan, the air is circulated within the hood and controls the temperature.

Humidity measurement and control: The steam of humidity module vaporizes the water particles. It control the humidity within the hood. The water is sterilized by boiling it to 100 .

O2 Module (Optional): The oxygen concentration in the hood is controlled when the oxygen option was attached.

SpO2 measurement (Optional): The SpO2 of infant can be measured when the option was installed. The probe sensor is applied to the end of an infant’s finger. By measuring the intensity of reflected light which is dependent on the concentration of dissolved oxygen in the blood, the oxygen saturation can be determined.

Pulse rate measurement (PR): During the SpO2 measurement, the light reflected by blood is pulsates by the heartbeat. By measuring this pulsation, the device can determine the pulse rate of the subject.

2.5Essential performance

Air temperature measurement and incubator air temperature control.

Skin temperature measurement and infant Skin temperature control.

Humidity measurement and control inside the incubator.

An alarm occurs if the difference between the control temperature and the reading

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

14

temperature is more than a certain level.

System must remain in a safe condition acc. IEC 60601-1, IEC 60601-2-19

WARNING Do not use the skin temperature control on babies who are in shock or who have high temperatures

2.6 Accessories and options

Picture

Name

 

Description

 

 

Qty

 

Control shell

Hold up

the hood

and

be

composed

with

 

 

instruments and parts that control the temperature

1ea

 

(Standard)

 

and humidity

 

 

 

 

 

 

 

 

 

 

 

 

 

Hood

Made of double framed clear acrylic panel to watch

1ea

 

(Standard)

inside, and to minimize heat loss

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fixed Stand

Movable incubator cradle with wheels

 

1ea

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Basket

Storage of

medical equipment

and items which

1ea

 

(Optional)

infant needs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Basket

 

 

 

 

 

 

 

 

Partition

Partition of Basket

 

 

 

 

1ea

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor

Measures

temperature

and

humidity inside

the

 

 

module

1ea

 

hood and infant’s body temperature

 

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mattress tray

Baby desk with X-ray tray

 

 

 

1ea

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

mattress

Accommodate infant stably with bouncy mattress

1ea

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Skin

 

 

 

 

 

 

 

 

temperature

Measures infant’s skin temperature

 

2ea

 

sensor

 

 

 

 

 

 

 

 

 

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IV-pole

IV hanger.

 

 

 

 

 

1ea

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AC power

AC Power

cord(AC Power cord

for operating

the

 

 

code

1ea

 

equipment)

 

 

 

 

 

(Standard)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

 

BT-500 Operation Manual

 

 

 

 

 

15

 

 

 

 

 

 

 

 

 

 

 

 

 

External LCD

Displays

measured

values

from the control and

 

 

 

 

Monitor

1ea

 

 

 

video of infant inside the hood.

 

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CCD Camera

Takes video of infant inside the hood

 

1ea

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Masimo SpO2

 

 

 

 

 

1ea

 

 

 

sensor probe

Measures infant’s SpO2

 

 

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Masimo

 

 

 

 

 

 

 

 

 

Extension for

Extend sensor cable

 

 

 

1ea

 

 

 

SpO2 sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

IV plate

Plate to place items which infant needs

 

1ea

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Shelf

Plate to place items which infant needs

 

1ea

 

 

 

(Optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lift Stand

Movable

incubator

cradle

with wheels

(VHA-

1ea

 

 

 

(Optional)

Variable Height Adjustable)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weighing

Measures Infant’s weight

 

 

 

 

 

 

Scale

 

 

1ea

 

 

 

(Optional)

 

 

 

 

 

 

 

Notes;

*The built-in air filter requires periodic replacement to maintain clean air. Periodic replacement of the pads according to the maintenance schedule is recommended.

**Oxygen control module is also available as an optional component.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

16

2.7Appearance of BT-500

2.7.1Front View

Control ShellHood

Sensor moduleMoving StandBasket

IV poleIV plate

External monitor

Figure 2-1. Front view

2.7.2 Front View Detail

Console box

Water tank draw

Tilting mechanism handleHood handle

Front access door

Baby desk with X-RAY trayWeighing scale

Compatible mattress

Figure 2-2. Front view details

2.7.3 Rear view

Rear access door

AC power cord & connector

Figure 2-3. Rear view

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

BT-500 Operation Manual

17

2.7.4 Side view

Sensor module, SpO2 sensor & external communication port

Main power switch

Main power AC connectorIncubator handle

Figure 2-4. Left view

Figure 2-5. Right view

2.8 Description of each part

BT-500 is composed with several parts. The control shell is the part which controls the entire device. To measure the infant’s environment, the sensor module is needed. The hood is used to protect an infant from the external environment and maintain the internal environment of hood to best condition.

2.8.1 Control shell

The control shell part consists of console box and water tank.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

Bistos BT-500 User Manual

BT-500 Operation Manual

18

Figure 2-6. Control Shell

The water tank has a 1 Liter capacity. The reservoir permits visual inspection of the water level. It is located in a drawer in the front of the shell. When the drawer is closed and the latching handle is engaged, the reservoir is connected to heater module.

For more information about how to clean, see "Chapter 6 cleanliness and maintenance”

2.8.2 Hood

The hood of BT-500 is an acrylic material. There is Access door in the front, rear and both sides of hood.

Figure 2-8. Hood front view & Access door / side function

WARNING To prevent accidental disconnection, secure all access panel latches. Failure to do so could result in personal injury of equipment damage.

 

 

 

P/N : 500-ENG-OPM-EUR-R12

Bistos Co., Ltd.

2019.07

Loading...
+ 43 hidden pages